K Number
K201604

Validate with FDA (Live)

Date Cleared
2020-12-23

(191 days)

Product Code
Regulation Number
878.4370
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Plasdent Disposable Barrier Sleeves and Barrier Film are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient one time use only.

Device Description

Plasdent Disposable Barrier Sleeves and Barrier Film are made of polyethylene film (PE), and come in various shapes and sizes as dental accessories intended to fit over and cover dental instruments and equipment. The Disposable Barrier Sleeves cover small hand-held dental instruments such as air/water syringes, curing lights, hand pieces, computers, sensors, dental instrument trays, and other similar handheld instruments to provide as a physical barrier or cover during a dental procedure. In other forms, the Disposable Barrier Film Covers are used to cover equipment, such as dental chairs, headrests, X-Ray heads, and other devices. The device is non-sterile, prepackaged, and is disposable, for single patient and one time use only.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA for a medical device: "Plasdent Disposable Barrier Sleeves and Barrier Film." It is primarily focused on demonstrating substantial equivalence to a predicate device through non-clinical performance testing.

It is important to note that this document does NOT describe the validation of an AI/ML powered medical device. It describes the testing of a physical barrier device. Therefore, many of the requested elements for an AI/ML device validation (e.g., sample size for test set with data provenance, number of experts for ground truth, MRMC study, sample size for training set, ground truth for training set) are not applicable or present in this context.

However, I can extract the relevant information for the physical device based on the provided text, focusing on acceptance criteria and performance data from the non-clinical testing.


Acceptance Criteria and Device Performance (for a physical barrier device)

The document presents non-clinical testing results to demonstrate that the Plasdent Disposable Barrier Sleeves and Barrier Film are substantially equivalent to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

TestStandardPurposeAcceptance CriteriaReported Device PerformanceResults Summary
BIOCOMPATIBILITY TESTING:
In-Vitro CytotoxicityANSI/AAMI/ISO 10993-5Evaluate the cytotoxicity of a test article extract.If viability is reduced to <70% of the blank, it has a cytotoxic potential.Text extract 100%: Viab. 79.10%Test extract 75%: Viab. 83.72%Test extract 50%: Viab. 88.91%Test extract 25%: Viab. 94.10%No cytotoxicity potential.
SensitizationISO 10993-10Evaluate the potential of the test article to cause delayed dermal contact sensitization in a guinea pig maximization test.Magnusson and Kligman grades of 1 or greater in the test group indicate sensitization.All dermal reactions for all treatment groups over 24 and 48 hours: 0Not considered a sensitizer in the guinea pig maximization test.
IrritationISO 10993-10Evaluate the potential of the test article to cause skin irritation following application on the skin or rabbits.Cumulative irritation index mean score >0.5 indicate irritation.Overall Test Group Mean: 0Irritation response category of test article classified as Negligible.
PERFORMANCE TESTING:
Tensile StrengthASTM D882-18Determine the tensile properties of Subject device's disposable barrier sleeve compared to that of Predicate device disposable barrier sleeve.Transverse: 35-45 MPaLengthways: 23-28 MPaHL181201: T 39.46, L 26.26HL20190601: T 39.84, L 26.46HL20190605: T 40.32, L 26.52The tensile strength of both products is in the same range.
Puncture ResistanceASTM F1342-05Determine the puncture resistance of Subject device's disposable barrier sleeve compared to that of Predicate device's disposable barrier sleeve.4N-5N20190528: 4.56N20190601: 4.58N20190605: 4.56NThe puncture resistance of both products is in the same range.
Tear ResistanceASTM D1004-13Determine the tear resistance of Subject device's disposable barrier sleeve compared to that of Predicate device's disposable barrier sleeve.4N-6NHL20190605: 5.026NHL20190610: 5.046NHL20190615: 5.032NThe tear resistance of both products is in the same range.
Synthetic Blood PenetrationASTM F1670Evaluate the resistance of test material to penetration by synthetic blood under conditions of continuous liquid contact.Pass determination based on no visual detection of synthetic blood penetration.Material: no synthetic blood penetration seenPass. Test material resistant to synthetic blood penetration.
Synthetic blood penetration at seams and non-continuous componentsASTM F1670Evaluate the resistance of test material at seams to penetration by synthetic blood under conditions of continuous liquid contact.Pass determination based on no visual detection of synthetic blood penetration.Material seams: no synthetic blood penetration seenPass. Test material seams resistant to synthetic blood penetration.
Viral PenetrationASTM F1671Evaluate the barrier performance of test material which are intended to protect against blood borne pathogen hazards.Pass determination based on no visual detection of viral penetration and assay titer value <1.Visual penetration: none seenAssay Titer (PFU/mL): <1Pass. Test material successfully protects against blood borne pathogen hazards.
Viral Penetration at seams and non-continuous componentsASTM F1671Evaluate the barrier performance of test material at seams which are intended to protect against blood borne pathogen hazards.Pass determination based on no visual detection of viral penetration and assay titer value <1.Visual penetration: none seenAssay Titer (PFU/mL): <1Pass. Test material seam successfully protects against blood borne pathogen hazards.
ThicknessASTM D6988-13Determine the thickness of test material.N/A (This is a measurement, not a pass/fail criterion for this test, though it is used for comparison)0.018mmThickness: 0.018mm

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the exact sample sizes (N) for each performance test. It lists results for multiple "samples" (e.g., HL181201, HL20190601, HL20190605 for Tensile Strength) but doesn't quantify the total count per test.
  • Data Provenance: The document does not specify the country of origin for the test data directly. It is inferred to be manufacturer-generated for regulatory submission. All tests appear to be retrospective in the sense that they were conducted in a lab setting, not in a live clinical environment, to characterize the device properties.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. This is a physical device (barrier sleeves/film), not an AI/ML device requiring expert interpretation of images or other data for ground truth establishment. The "ground truth" for these tests comes from objective measurements and standardized test methods (e.g., ASTM, ISO standards).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. Adjudication methods like 2+1 or 3+1 are used for human-reviewed data in clinical studies, typically for establishing ground truth in image analysis or diagnostic contexts. This is a physical device undergoing lab-based performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This type of study relates to AI-assisted human performance, which is not relevant for a physical barrier device. The study is a comparison of the subject device to a predicate device based on material properties and barrier effectiveness.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This applies to AI algorithms. The device itself is a standalone physical product, and its performance was evaluated without human intervention in the loop of its function (i.e., its barrier properties were tested directly).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this device is based on objective measurements and adherence to recognized industry standards (ASTM, ISO) for material properties (tensile strength, puncture resistance, tear resistance, thickness) and barrier performance (synthetic blood penetration, viral penetration), as well as biocompatibility assessments (cytotoxicity, sensitization, irritation). There isn't a "clinical ground truth" in the diagnostic sense.

8. The sample size for the training set:

  • Not Applicable. This is not an AI/ML device, so there is no "training set."

9. How the ground truth for the training set was established:

  • Not Applicable. As there is no training set for an AI/ML device, this question is irrelevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

December 23, 2020

Plasdent Corporation Belen Walayat Corporate Manager 969 Price Street Pomona, California 91767

Re: K201604

Trade/Device Name: Plasdent Disposable Barrier Sleeves and Barrier Film Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape And Drape Accessories Regulatory Class: Class II Product Code: PEM Dated: November 18, 2020 Received: November 27, 2020

Dear Belen Walayat:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 0613012020 See PRA Statement below.

510(k) Number (if known) K201604

Device Name

Plasdent Disposable Barrier Sleeves and Barrier Film

Indications for Use (Describe)

Plasdent Disposable Barrier Sleeves and Barrier Film are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient one time use only.

Model#DescriptionDesigned For
PS201B Tray Sleeves w/ Lock TopDental Instrument Tray, 10-1/2" x 14"
PS202A Tray Sleeves w/ Lock TopDental Instrument Tray, 11-5/8" x 14-1/2"
PS204F Tray Sleeves w/ Lock TopDental Instrument Tray, 7-1/2" x 10-1/2"
PS3800Half Chair Covers or SleevesHalf Dental Chair, 27-1/2"W x 24"L
PS3825Wide Half Chair CoverWide Half Chair, 32"W x 32"L
PS102Half Chair CoverHalf Chair, 24"W x 32"L
PS106Full Chair Covers or SleevesWide Full Dental Chair, 44"W x 54"L
PS3850Full Chair Covers or SleevesX-Long Full Chair, 29"W x 80"L
PS650Headrest CoverHeadrest, 9-1/2"W x 11"W
PS660Headrest CoverLarge Headrest, 10"L x 14"W
PS1250CSticky Wraps (Barrier Film)4"W x 6"L, Clear Color
PS1250BSticky Wraps (Barrier Film)4"W x 6"L, Blue Color
PS400Computer CoversKeyboard cover, 22"W x 14"L
PS400-SComputer CoversSmall Keyboard, 12-1/2"W x 8"L
PF405Computer CoversPC Mouse, Universal
PS410Computer CoversLCD + Keyboard 22"W x 26"L
PS425Computer CoversLaptop, 15-2/3"W x 24"L
PS1100X-ray CoverX-ray 24"W x 32"L
PS1105X-ray CoverX-ray, Universal, 15"W x 26"L
PS328PLite Handle CoverLite Handle for T-Style 5-3/4"W x 4"L
PS320AAir/Water Syringe CoversAir/Water Syringe, 3"W x 10"L
PS3720Air/Water Syringe CoversAir/Water Syringe, Tube 2" Diameter
PS3740Air/Water Syringe CoversAir/Water Syringe, Tube 4" Diameter
PS520Sensor Cover#2 Sensor Cover, 1-5/8"W x 8-3/8"L
PS530Sensor Cover#0 Sensor Cover, 1-3/8"W x 8-3/8"L
PS-SUNI-2X-ray Sensor SheathsFor SUNI Size 2 Compatible TIDI: 20819
PS-DEXIS-2X-ray Sensor SheathsFor DEXIS Size 2 Compatible TIDI: 20999
PS-6100-0X-ray Sensor SheathsFor KODAK 6100 Size 0 Compatible TIDI:20977
PS-6100-1X-ray Sensor SheathsFor KODAK 6100 Size 1 Compatible TIDI:20978
PS-6100-2X-ray Sensor SheathsFor KODAK 6100 Size 2 Compatible TIDI:20979
PS-GXDR-1X-ray Sensor SheathsFor GENDEX/XDR Size 1
PS-GXDR-2X-ray Sensor SheathsFor GENDEX/XDR Size 2
PS-SHICK-1X-ray Sensor SheathsFor SCHICK Size 1 Compatible TIDI: 20824
PS-SHICK-2X-ray Sensor SheathsFor SCHICK Size 2 Compatible TIDI: 20825
PS4550Curing Light SleevesCuring Light Handle, 5"W x 10"L
PS4660Curing Light SleevesComplete Curing Light Handle, 3-1/4"W x12"L
PS-DEMICuring Light SleevesCuring Light for DEMI
PS-LEDCuring Light SleevesLED Curing Light, 3-1/4"W x 12"L
PS700Handpiece CoversLow Speed Long H.P., 1-1/2"W 7-1/2"L
PS710Handpiece CoversLow Speed Contra-Angle, 8"W 3-1/2"L
PS720Handpiece CoversHigh Speed, 1-1/6"W x 7-3/4"L

{3}------------------------------------------------

{4}------------------------------------------------

Type of Use (Select one or both, as applicable)

□ Prescription Use (Part 21 CFR 801 Subpart D)

网 Over-The-Counter Use

(21 CFR 801 Subpart C) CONTINUE ON A SEPARATE

PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17) PublllhiD& Senrkes (301) 443-6740

PSC

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Plasdent Corporation. The logo consists of a green square with a white shape inside, followed by the word "PLASDENT" in green, all caps. Below the word "PLASDENT" are two green lines, and below the lines is the word "CORPORATION" in smaller, green, all-caps letters.

510(K) Summary K201604

I. SUBMITTER: Plasdent Corporation 969 Price Street Pomona, CA 91767 Contact Person: Belen G. Walayat Title: Corporate Manager Phone: (909)620-0289*106 Email: belengw@plasdent.com Summary prepared: 01/30/2019 Summary revised: 08/19/2020

II. DEVICE

Name of Device: Plasdent Disposable Barrier Sleeves and Barrier Film Regulation Number: 21 CFR 878.4370 Common Name: Dental Barriers and Sleeves Classification Name: Surgical drape and drape accessories Regulatory Class: II Product Code: PEM

III. PREDICATE DEVICE

Predicate device: Pac-Dent Barrier Sleeve, Cover-It™ Barrier Film- K151123 Subject Device: Plasdent Disposable Barrier Sleeves and Barrier Film- K201604

IV. DEVICE DESCRIPTION

Plasdent Disposable Barrier Sleeves and Barrier Film are made of polyethylene film (PE), and come in various shapes and sizes as dental accessories intended to fit over and cover dental instruments and equipment. The Disposable Barrier Sleeves cover small hand-held dental instruments such as air/water syringes, curing lights, hand pieces, computers, sensors, dental instrument trays, and other similar handheld instruments to provide as a physical barrier or cover during a dental procedure. In other forms, the

{6}------------------------------------------------

Disposable Barrier Film Covers are used to cover equipment, such as dental chairs, headrests, X-Ray heads, and other devices. The device is non-sterile, prepackaged, and is disposable, for single patient and one time use only.

V. INDICATIONS FOR USE

Plasdent Disposable Barrier Sleeves and Barrier Film are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

Model#DescriptionDesigned For
PS201B Tray Sleeves w/ Lock TopDental Instrument Tray, 10-1/2" x 14"
PS202A Tray Sleeves w/ Lock TopDental Instrument Tray, 11-5/8" x 14-1/2"
PS204F Tray Sleeves w/ Lock TopDental Instrument Tray, 7-1/2" x 10-1/2"
PS3800Half Chair Covers or SleevesHalf Dental Chair, 27-1/2"W x 24"L
PS3825Wide Half Chair CoverWide Half Chair, 32"W x 32"L
PS102Half Chair CoverHalf Chair, 24"W x 32"L
PS106Full Chair Covers or SleevesWide Full Dental Chair, 44"W x 54"L
PS3850Full Chair Covers or SleevesX-Long Full Chair, 29"W x 80"L
PS650Headrest CoverHeadrest, 9-1/2"W x 11"W
PS660Headrest CoverLarge Headrest, 10"L x 14"W
PS1250CSticky Wraps (Barrier Film)4"W x 6"L, Clear Color
PS1250BSticky Wraps (Barrier Film)4"W x 6"L, Blue Color
PS400Computer CoversKeyboard cover, 22"W x 14"L
PS400-SComputer CoversSmall Keyboard, 12-1/2"W x 8"L
PF405Computer CoversPC Mouse, Universal
PS410Computer CoversLCD + Keyboard 22"W x 26"L
PS425Computer CoversLaptop, 15-2/3"W x 24"L
PS1100X-ray CoverX-ray 24"W x 32"L
PS1105X-ray CoverX-ray, Universal, 15"W x 26"L
PS328PLite Handle CoverLite Handle for T-Style 5-3/4"W x 4"L
PS320AAir/Water Syringe CoversAir/Water Syringe, 3"W x 10"L
PS3720Air/Water Syringe CoversAir/Water Syringe, Tube 2" Diameter
PS3740Air/Water Syringe CoversAir/Water Syringe, Tube 4" Diameter
PS520Sensor Cover#2 Sensor Cover, 1-5/8"W x 8-3/8"L
PS530Sensor Cover#0 Sensor Cover, 1-3/8"W x 8-3/8"L
PS-SUNI-2X-ray Sensor SheathsFor SUNI Size 2 Compatible TIDI: 20819
PS-DEXIS-2X-ray Sensor SheathsFor DEXIS Size 2 Compatible TIDI: 20999
PS-6100-0X-ray Sensor SheathsFor KODAK 6100 Size 0 Compatible TIDI: 20977
PS-6100-1X-ray Sensor SheathsFor KODAK 6100 Size 1 Compatible TIDI: 20978
PS-6100-2X-ray Sensor SheathsFor KODAK 6100 Size 2 Compatible TIDI: 20979
PS-GXDR-1X-ray Sensor SheathsFor GENDEX/XDR Size 1
PS-GXDR-2X-ray Sensor SheathsFor GENDEX/XDR Size 2
PS-SHICK-1X-ray Sensor SheathsFor SCHICK Size 1 Compatible TIDI: 20824
PS-SHICK-2X-ray Sensor SheathsFor SCHICK Size 2 Compatible TIDI: 20825
PS4550Curing Light SleevesCuring Light Handle, 5"W x 10"L
PS4660Curing Light SleevesComplete Curing Light Handle, 3-1/4"W x 12"L
PS-DEMICuring Light SleevesCuring Light for DEMI
PS-LEDCuring Light SleevesLED Curing Light, 3-1/4"W x 12"L
PS700Handpiece CoversLow Speed Long H.P., 1-1/2"W x 7-1/2"L
PS710Handpiece CoversLow Speed Contra-Angle, 8"W x 3-1/2"L
PS720Handpiece CoversHigh Speed, 1-1/6"W x 7-3/4"L

{7}------------------------------------------------

{8}------------------------------------------------

VI. TECHNOLOGICAL CHARACTERISTICS COMPARISON

The subject device, Plasdent Disposable Barrier Film, and the predicate device, Pac-Dent Barrier Sleeve, Cover-It™ Barrier Film, have the same intended use, which are used as protective barriers intended to fit over and cover dental instruments and equipment. Both the subject device and predicate device are made of same material, which is polyethylene film (PE). Both are non-sterile, prepackaged, disposable, and are for single patient and one time use only. The main slight difference between the subject and predicate device is the material composition. The polyethylene film of the subject device is composed of 30% low density polyethylene (LDPE) and 70% linear low density polyethylene (LLDPE), while the predicate device is composed of 20% low density polyethylene (LDPE) and 80% linear low density polyethylene (LLDPE). The difference is negligible & within acceptable range.

DEVICESubject DevicePlasdent Disosable BarrierSleeves and Barrier Film(K201604)Primary Predicate DevicePac-Dent Barrier Sleeve,Cover-ItTM Barrier Film(K151123)Comparison
Intended UseTo be used as a barrier for dentalinstruments and equipment.To be used as a barrier for dentalinstruments and equipment.Same
ClassificationProduct CodePEMPEMSame
MaterialPolyethylene filmPolyethylene filmSame
MaterialCompositionLLDPE (70%)LDPE (30%)Blue pigment(item#PS1250B)Adhesive (item#PS1250B,PS1250C)LLDPE(80%)LDPE (20%)Blue pigment(item#100B, C101B)Adhesive (item#C101, C101B)Similar
BiocompatibilityNon-cytotoxicNon-sensitizingNon-irritatingNon-cytotoxicNon-sensitizingNon-irritatingSame
Specificationsand TolerancesPaper backing: noneFilm thickness:0.02-0.06mmTolerance: <0.01mmPaper backing: some of the modelFilm thickness:0.02-0.06mmTolerance: <0.01mmSimilar
MechanicalPropertiesTensile Strength -ASTM D882-Subject device is equivalent topredicate device.Tensile Strength -ASTMD882SamePass
Resistance to Puncture -ASTMF1342-Subject device is equivalentto predicate device.Resistance to Puncture- ASTMF1342SamePass
Tear Strength – ASTM D1424-Subject device is equivalent topredicate device.Tear Strength – ASTMD1424SamePass
Effectiveness of X-Ray and SensorDevices covered with Barrierdevices.SamePass
PerformancePropertiesSynthetic Blood Penetration-PassSynthetic Blood Penetration-PassSamePass
SterilityNon-sterileNon-sterileSame
Single UseSingle use deviceSingle use deviceSame
FDA-RecognizedStandardsASTM F1670ASTM F1670Same
ASTM F1671ASTM F1671
ASTM D882-18ASTM D882-18
ASTM F1342ASTM F1342
ASTM D1004-13ASTM D1004-13
ISO 10993-5ISO 10993-5
ISO 10993-10ISO 10993-10
ISO 6988-13ISO 6988-13

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Plasdent Corporation. The logo consists of a green square with a white outline and a white shape inside. To the right of the square is the word "PLASDENT" in green, with the word "CORPORATION" in smaller green letters below it. The logo is simple and professional, and the green color suggests that the company is environmentally friendly.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Plasdent Corporation. The logo consists of a green square with a white square cut out of the top right corner. Below the square is the word "PLASDENT" in green, with the word "CORPORATION" in smaller green letters below that. The logo is simple and modern, and the green color gives it a sense of freshness and vitality.

VII. SUMMARY OF NON-CLINICAL TESTING

Shown below is the non-clinical testing performed with the subject device can meet the acceptance criteria referenced standard or test methodology shown in the table below:

TESTSTANDARDPURPOSEACCEPTANCE CRITERIARESULTSRESULTS SUMMARY
BIOCOMPATIBILITY TESTING:
In-Vitro CytotoxicityANSI/AAMI/ISO10993-5Evaluate the cytotoxicity of a testarticle extract.If viability is reduced to<70% of the blank, it has acytotoxic potential.Text extract 100%: Viab. 79.10%Test extract 75%: Viab. 83.72%Test extract 50%: Viab. 88.91%Test extract 25%: Viab. 94.10%No cytotoxicity potential.
SensitizationISO 10993-10Evaluate the potential of the testarticle to cause delayed dermalcontact sensitization in a guinea pigmaximization test.Magnusson and Kligmangrades of 1 or greater in thetest group indicatesensitization.All dermal reactions for alltreatment groups over 24 and48 hours: 0Not considered a sensitizerin the guinea pigmaximization test.
IrritationISO 10993-10Evaluate the potential of the testarticle to cause skin irritationfollowing application on the skin orrabbits.Cumulative irritation indexmean score >0.5 indicateirritation.Overall Test Group Mean: 0Irritation response categoryof test article classified asNegligible.
PERFORMANCE TESTING:
Tensile StrengthASTM D882-18Determine the tensile properties ofSubject device's disposable barriersleeve compared to that of Predicatedevice disposable barrier sleeve.Transverse: 35-45 MPaLengthways: 23-28 MPaHL181201: T 39.46, L 26.26HL20190601: T 39.84, L 26.46HL20190605: T 40.32, L 26.52The tensile strength of bothproducts is in the samerange.
Puncture ResistanceASTM F1342-05Determine the puncture resistance ofSubject device's disposable barriersleeve compared to that of Predicatedevice's disposable barrier sleeve.4N-5N20190528: 4.56N20190601: 4.58N20190605: 4.56NThe puncture resistance ofboth products is in the samerange.
Tear ResistanceASTM D1004-13Determine the tear resistance ofSubject device's disposable barriersleeve compared to that of Predicatedevice's disposable barrier sleeve.4N-6NHL20190605: 5.026NHL20190610: 5.046NHL20190615: 5.032NThe tear resistance of bothproducts is in the samerange.
Synthetic BloodPenetrationASTMF1670Evaluate the resistance of testmaterial to penetration by syntheticblood under conditions of continuousliquid contact.Pass determination basedon no visual detection ofsynthetic bloodpenetration.Material: no synthetic bloodpenetration seenPass. Test material resistantto synthetic bloodpenetration.
Synthetic bloodpenetration atseams and non-continuous componentsASTMF1670Evaluate the resistance of testmaterial at seams to penetration bysynthetic blood under conditions ofcontinuous liquid contact.Pass determination basedon no visual detection ofsynthetic bloodpenetration.Material seams: no syntheticblood penetration seenPass. Test material seamsresistant to synthetic bloodpenetration.
Viral PenetrationASTMF1671Evaluate the barrier performance oftest material which are intended toprotect against blood borne pathogenhazards.Pass determination basedon no visual detection ofviral penetration and assaytiter value <1 °Visual penetration: none seenAssay Titer (PFU/mL): <1°Pass. Test materialsuccessfully protects againstblood borne pathogenhazards.
Viral Penetration atseams and non-continuous componentsASTMF1671Evaluate the barrier performance oftest material at seams which areintended to protect against bloodborne pathogen hazards.Pass determination basedon no visual detection ofviral penetration and assaytiter value <1°Visual penetration: none seenAssay Titer (PFU/mL): <1°Pass. Test material seamsuccessfully protects againstblood borne pathogenhazards.
ThicknessASTM D6988-13Determine the thickness of testmaterial.N/A0.018mmThickness: 0.018mm

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Plasdent Corporation. The logo consists of a green square with a white design inside, next to the word "PLASDENT" in green, with the word "CORPORATION" in smaller green letters below it. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains a logo for PLASDENT CORPORATION. The logo consists of a green geometric shape on the left, followed by the text "PLASDENT" in a larger, bolder font. Below "PLASDENT", the word "CORPORATION" is written in a smaller font. The overall design is simple and professional, likely representing a corporate identity.

Clinical Performance Data

No needed for this device.

VIII. CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the Plasdent Disposable Barrier Sleeves and Barrier Films device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, the Pac-Dent Barrier Sleeve cleared under K151123, Class II (21 CFR 878.4370), product code PEM.

§ 878.4370 Surgical drape and drape accessories.

(a)
Identification. A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.(b)
Classification. Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.