K Number
K201575

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-07-08

(27 days)

Product Code
Regulation Number
892.1720
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.

Device Description

This represents the straightforward combination of three devices: One of three cleared MinXray Portable HF X-ray generators: a) HF120/60H PowerPlus cleared in K040046, (and in K141885) OR b) HF100H+ cleared in K052721 OR C) HF1202 PowerPlus cleared in K153059. d) Plus: A 510(k) cleared (K153058) Digital X-Ray Receptor Panel CareView 1500C X-ray Flat Panel Detector. PLUS: the dicomPACS® software package (Same as our predicate). e) The x-ray generators are portable units which operate from 120/240V 50-60° AC. The generator unit utilizes a high frequency inverter and can be mounted to a tripod or support arm. The usual safety precautions regarding the use of x-rays must be observed by the operator. The digital panel features the Careray flat panel technology in a sleek and compact unit. The portable panel provides digital X-ray image capture for a wide range of applications. The lightweight design, generous imaging area, and fast processing times of the detector make it easy to capture high quality diagnostic images for routine diagnosis, as well as challenging trauma and bedside exams. It's a portable solution for a faster, more streamlined approach to digital radiography. The only difference between this modified device and our predicate devices is the model number of the digital x-ray receptor panel. The predicate panel can communicate either by wireless or wired connection. The subject device communicates by Ethernet only.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a mobile X-ray system (CMDR 2C) seeking substantial equivalence to a predicate device. This submission primarily focuses on hardware equivalence and non-clinical testing, rather than an AI/ML-based diagnostic device requiring extensive performance metrics against a clinical ground truth.

Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML device (e.g., sample sizes for test/training sets, expert adjudication methods, MRMC studies, specific ground truth types for disease detection) are not applicable or not present in this document.

However, I can extract the information related to the device's performance through bench testing and the rationale for claiming substantial equivalence.

Here's an analysis based on the provided document:

1. Table of acceptance criteria and the reported device performance:

Since this FDA submission is for a medical device (mobile X-ray system) and not an AI/ML diagnostic tool, the "acceptance criteria" are related to established performance standards for X-ray imaging devices and demonstrating equivalence to a predicate device. The "reported device performance" is primarily about image quality and compliance with relevant safety and electrical standards.

Acceptance Criteria (Implied from testing)Reported Device Performance
Image Quality: Produce diagnostic quality images.Prototype systems covering all generator/panel combinations were assembled and tested. Using the i.b.a. Test Device DIGI-13 (a device for quality tests at CR and DR systems, e.g., for acceptance tests according to DIN V 6868-58 and constancy tests according to DIN 6868-13), images were obtained from both the predicate and the new digital panel. "The images were evaluated and found to be of diagnostic quality." This implies the new device's image quality is at least equivalent to the predicate and meets diagnostic standards. Specific numerical metrics like resolution (Spatial Resolution), A/D Resolution, MTF, and DQE are provided for the panel, which match those of the predicate device.
Safety and Electrical Compliance: Adherence to relevant safety standards.The completed system complies with DHHS radiation safety standards. It has undergone testing for compliance with: - UL 60601-1 (2005) (Electrical medical device safety) - IEC 60601-1-2 (2007) (Electromagnetic Compatibility) - Additionally, the HF1202H PowerPlus generator meets IEC 60601-2-54: Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy (not applicable to older generators). - Risks and hazardous impacts of the device modification were analyzed by FMEA methodology. Risk control and protective measures were reviewed and implemented. "The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted."
Software Compatibility: New panel compatible with existing software."We verified software compatibility with the new CareView Cw digital panel." The dicomPACS® software was installed on a Dell Inspiron laptop, and proper installation was verified by running the software. A Wi-Fi connection (for the predicate panel) was confirmed, and the X-ray generator was used to take test exposures with a radiographic phantom. "No modifications were necessary to any of the hardware or software other than changing the digital panel."
Substantial Equivalence: Performance is as safe and effective as predicate."The results of bench testing indicate that the new devices are as safe and effective as the predicate devices." The primary difference is the digital X-ray receptor panel, with the new panel communicating via Ethernet only, while the predicate panel could communicate via wireless or wired. However, the panel's core performance characteristics (Pixel Pitch, pixels, Size, Scintillator, Spatial Resolution, A/D Resolution, MTF, DQE) are identical to the predicate panel (which was cited as the predicate in the K150929 submission).

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated as a numerical sample size of "cases" or "patients" in the context of clinical images. The testing involved "Prototype systems covering all generator/panel combinations" and "Several test exposures" with "a radiographic phantom." This suggests a limited number of phantom images rather than a large clinical test set.
  • Data Provenance: The data used for testing (phantom images) would have been generated in-house during the bench testing. The document does not specify country of origin for this testing data. The testing was non-clinical for this specific submission ("Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.").

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable / Not specified. Given that this is a non-clinical bench study focused on physical performance and image quality of an X-ray machine rather than a diagnostic AI algorithm, there was no "ground truth" derived from expert clinical reads of patient images. The evaluation of "diagnostic quality" for phantom images would likely be performed by engineers or physicists familiar with imaging standards, but specific numbers or qualifications are not provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. No adjudication method for expert reads was used as no clinical images or expert reads were part of this substantiation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study was not done. This device is a standard X-ray imaging system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is an X-ray machine; there is no standalone AI algorithm whose performance is being evaluated in this submission.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this submission was based on physical phantom measurements and engineering standards compliance (e.g., compliance with UL, IEC standards, and the ability to produce images "of diagnostic quality" using standard test phantoms like the i.b.a. Test Device DIGI-13). There was no clinical ground truth (expert consensus, pathology, or outcomes data) used in this particular substantial equivalence determination.

8. The sample size for the training set:

  • Not applicable. This submission is for an X-ray imaging device, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. No AI/ML training set was used.

{0}------------------------------------------------

July 8, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling a symbol. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

MinXray, Inc. % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

Re: K201575

Trade/Device Name: CMDR 2C (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB, LLZ Dated: June 10, 2020 Received: June 11, 2020

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201575

Device Name CMDR 2C (Multiple Models)

Indications for Use (Describe)

Intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammography.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: 510(k) Number K201575 MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062, USA Toll Free 1-800-221-2245 (USA & Canada) Date Prepared: June22, 2020 Contact: Keith Kretchmer, President

    1. Identification of the Device: Trade/Device Name: CMDR 2C (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
    1. Equivalent legally marketed device: K191451, MinXray, Inc. Trade/Device Name: CMDR 2CW (Multiple Models) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
    1. Reference device (We employ these cleared devices without modification): Trade/Device Name: K153058 CareView 1500C/CareView 1500L X-ray Flat Panel Detectors; Manufacturer: CareRay Digital Medical Systems Co., Ltd Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
  • Indications for Use (intended use): These digital radiographic systems are intended for use by a qualified/trained 4) physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use.
    1. Description of the Device: This represents the straightforward combination of three devices: One of three cleared MinXray Portable HF X-ray generators:
    • a) HF120/60H PowerPlus cleared in K040046, (and in K141885) OR
    • HF100H+ cleared in K052721 OR b)
    • C) HF1202 PowerPlus cleared in K153059.
    • d) Plus: A 510(k) cleared (K153058) Digital X-Ray Receptor Panel CareView 1500C X-ray Flat Panel Detector.
    • PLUS: the dicomPACS® software package (Same as our predicate). e)

The x-ray generators are portable units which operate from 120/240V 50-60° AC. The generator unit utilizes a high frequency inverter and can be mounted to a tripod or support arm. The usual safety precautions regarding the use of x-rays must be observed by the operator. The digital panel features the Careray flat panel technology in a sleek and compact unit. The portable panel provides digital X-ray image capture for a wide range of applications. The lightweight design, generous imaging area, and fast processing times of the detector make it easy to capture high quality diagnostic images for routine diagnosis, as well as challenging trauma and bedside exams. It's a portable solution for a faster, more streamlined approach to digital radiography. The only difference between this modified device and our predicate devices is the model number of the digital x-ray receptor panel. The predicate panel can communicate either by wireless or wired connection. The subject device communicates by Ethernet only.

{4}------------------------------------------------

6) Substantial Equivalence Chart

CharacteristicPredicate: K191451CMDR 2CW (Multiple Models)CMDR 2C (Multiple Models)
Intended Use:Intended for use by a qualified/trainedphysician or technician on both adult andpediatric subjects for taking diagnostic x-rays. Not for mammographySAME
ConfigurationMobile System with digital x-ray panel andimage acquisition computerSAME
ComputerDell Precision M3530 or Dell Latitude 7424(ruggedized)Dell Precision M3530 or Dell Latitude 7424(ruggedized)
X-rayGenerator(s)(All made byMikasa X-Ray)HF120/60H PowerPlus (K040046) orHF100H+ (K052721) orHF1202H PowerPlus (K153059)SAME
BasicGeneratorCharacteristics120 VAC Line operated:HF120/60H PowerPlus™ produces up to 120kVP and 2.4 kW peakHF100H+: up to 100 kVp and 2 kW peakHF1202H PowerPlus: up to 120 kVp up to3.0 kW peakSAME
CollimatorCollimare LED CollimatorCollimare LED Collimator
GeneratorPhoto(s)The samegenerators willbe used for themodified deviceImage: HF120/60HPPWVImage: HF100H+Image: HF1202 H
Digital X-rayPanel SuppliedK150929 CareView 1500Cw X-ray Flat PanelDetector manufactured by CareRayK153058 CareView 1500C X-ray Flat PanelDetector manufactured by CareRay
PanelPerformancePixel Pitch 154 μm2304 × 2816 pixelsSize 14" x 17"(Cleared in K150929)Csl ScintillatorSpatial Resolution Min. 3.3 line pair/mmA/D Resolution: 16 bit, 65536 grayscaleMTF - 65% @ 1 lp/mmDQE Performance: > 65% @ 0 lp/mmPixel Pitch 154 μm2304 × 2816 pixelsSize 14" x 17"(Cleared in K153058)Csl ScintillatorSpatial Resolution Min. 3.3 line pair/mmA/D Resolution: 16 bit, 65536 grayscaleMTF - 65% @ 1 lp/mmDQE Performance: > 65% @ 0 lp/mmNote: This panel was cited as the predicate inK150929. The panels are identical except for themethod of image communication to the host.
PanelCommunicationIEEE 802.11a/b/g/n (2.4 GHz /5 GHz)Gigabit Ethernet
CharacteristicPredicate: K191451CMDR 2CW (Multiple Models)CMDR 2C (Multiple Models)
Panel PowerSourceDC Adapter or Lithium Ion rechargeablebatteryDC Adapter or Lithium Ion rechargeablebattery
Panel InterfaceEthernet or Wi-Fi wirelessEthernet only
Meets USPerformanceStandardYESYES
PACS softwaredicomPACS® (Cleared with the Toshibapanels in K141440)SAME
Power Source120 V 50/60 Hz AC 20 ampSAME
Digital PanelPower Source120 V 50/60 Hz AC or Lithium IonRechargeable Battery.SAME
Digital PanelImage: Digital PanelImage: Digital Panel
Model Details(All models use O&R Imaging Software with theCareray Digital Panel and a Dell Laptop) (FiveModels) Model Details:(All models use O&R Imaging Software with theCareray Digital Panel and a Dell Laptop) (FiveModels) Model Details:
CMDR.CW.120.60.S for Dell Precision orCMDR.CW.120.60.R for Dell Latitude Laptop orequivalent Uses HF120/60HPPWV Generatorwith CareView 1500Cw and O&R ImagingSoftwareCMDR.C.120.60.S for Dell Precision orCMDR.C.120.60.R for Dell Latitude Laptop orequivalent Uses HF120/60HPPWV Generatorwith CareView 1500Cw and O&R ImagingSoftware
CMDR.CW.100.S for Dell Precision orCMDR.CW.100.Rfor Dell Latitude Laptop orequivalent Uses HF100H+ Generator withCareView 1500Cw and O&R Imaging Software.CMDR.C.100.S for Dell Precision orCMDR.C.100.Rfor Dell Latitude Laptop orequivalent Uses HF100H+ Generator withCareView 1500Cw and O&R Imaging Software.
CMDR.CW.1202.S for Dell Precision Laptop orequivalent orCMDR.CW.1202.R for Dell Latitude Laptop orequivalent Uses HF1202 Generator withCareView 1500Cw and O&R Imaging SoftwareCMDR.C.1202.S for Dell Precision Laptop orequivalent orCMDR.C.1202.R for Dell Latitude Laptop orequivalent Uses HF1202 Generator withCareView 1500Cw and O&R Imaging Software
CMDR 2CW-MIL (Uses Dell Latitude 7424 orequivalent with the HF120/60HPPWV Generatorwith CareView 1500Cw and O&R ImagingSoftwareCMDR 2C-MIL (Uses Dell Latitude 7424 orequivalent with the HF120/60HPPWV Generatorwith CareView 1500Cw and O&R ImagingSoftware

{5}------------------------------------------------

{6}------------------------------------------------

    1. The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and produced diagnostic quality images as good as our predicate generator/panel combination. NO HARDWARE OR SOFTWARE MODIFICATIONS TO ALREADY CLEARED DEVICES WERE REQUIRED TO CREATE THESE NEW MODELS. The main difference from the predicate device is the addition of the CareView 1500 CW flat-panel detector, that replaces the Toshiba and Perkin-Elmer digital imagers used with the predicate system.
    1. Summary of non-clinical testing: Prototype systems covering all generator/panel combinations were assembled and tested. First the dicomPACS® software was installed on the Dell Inspiron laptop computer. The proper installation was verified by running the software. A Wi-Fi connection was confirmed between the PerkinElmer panel and the Dell laptop. The panel was confirmed to be charged and turned on. Then the MinXray HF portable generator was turned on. The MinXray portable X-ray generator was set to generate an exposure. The generator was aimed at the PerkinElmer panel, and a radiographic phantom was placed on the panel. Several test exposures showed that the system was operating properly. No modifications were necessary to any of the hardware or software other than changing the digital panel. The completed system complies with DHHS radiation safety standards currently in effect, and has undergone testing for compliance with UL 60601-1 (2005) (Electrical medical device safety), IEC 60601-1-2 (2007) (Electromagnetic Compatibility). Additionally, the HF1202H PowerPlus generator meets IEC 60601-2-54: Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. The other generators were designed and cleared prior to the isuance of this standard. We verified software compatibility with the new CareView Cw digital panel. The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.

We employed the i.b.a. Test Device DIGI-13, a device for quality tests at CR and DR systems (e.g. for acceptance tests according to DIN V 6868-58 and constancy tests according to DIN 6868-13) to obtain images from both the predicate and the new digital panel. All panel/generator combinations were tested. The images were evaluated and found to be of diagnostic quality.

In recognition of the FDA Guidance on Cybersecurity, we added Cybersecurity precautions to our labeling and obtained Cybersecurity information from our DICOM software supplier Oehm Und Rehbein.

    1. Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.
    1. Conclusion: After analyzing bench and clinical tests, it is the conclusion of MinXray Inc. that the modified Digital Diagnostic X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.