K Number
K201560

Validate with FDA (Live)

Date Cleared
2021-08-31

(447 days)

Product Code
Regulation Number
892.2070
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Auto Lung Nodule Detection is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 10 to 30 mm in size. It is designed to aid the physician to review the PA chest radiographs of adults as a second reader and be used as part of S-Station, which is operation software installed on Samsung Digital X-ray Imaging systems. Auto Lung Nodule Detection cannot be used on the patients who have lung lesions other than abnormal nodules.

Device Description

Auto Lung Nodule Detection is a Computer-Aided Detection (CADe) device that is designed to perform CAD processing in Chest X-ray images for indication of locations for high nodule probability, which has an effective detection sizes from 10 mm to 30 mm.

Auto Lung Nodule Detection receives images acquired with SAMSUNG Digital X-ray Imaging Systems as an input and identifies suspected nodules, and then sends information of suspected nodules to the visualization part of S-Station, which is installed on all kinds of SAMSUNG Digital X-ray Imaging Systems, to generate output images with circular marks. The CAD performed images, are displayed on the screen by S-Station without defeat of original images and used as a second reader only after the initial read is completed.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Auto Lung Nodule Detection device, based on the provided FDA 510(k) Premarket Notification:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes the clinical study's objective as demonstrating that the device improves human readers' nodule detection performance. While specific numerical acceptance criteria (e.g., minimum sensitivity, maximum FPPI) are not explicitly called out in a "table of acceptance criteria," the clinical performance testing section clearly states that the results have demonstrated that all readers' nodule detection performances using the proposed device have increased with statistical significance. This implicitly defines the acceptance criteria: a statistically significant improvement in nodule detection performance metrics.

Metric (Implicit Acceptance Criteria)Reported Device Performance
SensitivityIncreased with statistical significance (with ALND assistance)
False Positives per Image (FPPI)Improved (implicitly, as performance increased)
JAFROC Figure of Merit (FOM)Increased with statistical significance (with ALND assistance)

2. Sample Size and Data Provenance

  • Test Set Sample Size: Not explicitly stated. The document mentions "a test dataset containing both normal and diseased images."
  • Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as a "clinical evaluation," implying real patient data. It is not specified if the data was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not explicitly stated.
  • Qualifications of Experts: Not explicitly stated. However, given the context of a CADe device for radiological use, these would typically be board-certified radiologists or pulmonary specialists.

4. Adjudication Method for the Test Set

The document mentions that "Readers were asked to mark their region of nodule suspicion on the images while also providing confidence scores on each decision they have made." It then states, "After independent reading, readers were allowed to adjust their confidence scores after reviewing the ALND's detection results." This describes a workflow, but does not explicitly describe the ground truth adjudication method (e.g., consensus reading by multiple experts, pathology confirmation). The term "independent reading" before ALND review suggests individual reads prior to any potential adjudication or consensus if it occurred.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Yes, an MRMC comparative effectiveness study was conducted.

  • Effect Size of Improvement: The document states, "The results have demonstrated that all readers' nodule detection performances using the proposed device have increased with statistical significance." While specific numerical effect sizes (e.g., percentage increase in sensitivity, reduction in FPPI, or change in JAFROC FOM) are not provided, the key finding is the statistical significance of the improvement when radiologists used ALND as an assistant tool.

6. Standalone (Algorithm Only) Performance

The document does not explicitly report standalone performance of the algorithm without human-in-the-loop. The clinical evaluation focuses on the human-in-the-loop performance (with ALND assistance vs. without).

7. Type of Ground Truth Used

The type of ground truth used is not explicitly stated. It can be inferred that it involved expert consensus or a gold standard based on follow-up, but the document does not specify if it was pathology, follow-up imaging, or expert consensus.

8. Sample Size for the Training Set

The sample size for the training set is not provided in this document.

9. How Ground Truth for the Training Set was Established

The method for establishing ground truth for the training set is not provided in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 31, 2021

Samsung Electronics Co., Ltd. % Jaesang Noh Senior Professional, Regulatory Affairs 129, Samsung-ro. Yeongtong-gu Suwon-si, Gyeonggi-do 16677 REPUBLIC OF KOREA

Re: K201560

Trade/Device Name: Auto Lung Nodule Detection Regulation Number: 21 CFR 892.2070 Regulation Name: Medical image analyzer Regulatory Class: Class II Product Code: MYN Dated: July 15, 2021 Received: July 20, 2021

Dear Jaesang Noh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K201560

Device Name Auto Lung Nodule Detection

Indications for Use (Describe)

The Auto Lung Nodule Detection is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 10 to 30 mm in size. It is designed to aid the physician to review the PA chest radiographs of adults as a second reader and be used as part of S-Station, which is operation software installed on Samsung Digital X-ray Imaging systems. Auto Lung Nodule Detection cannot be used on the patients who have lung lesions other than abnormal nodules.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.

510(k) Premarket Notification - Traditional

Section 5: 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92

    1. Date: August 26, 2021

Submitter 2.

  • Company Name: SAMSUNG ELECTRONICS Co., Ltd. A.
  • B. Address: Address: 129, Samsung-ro, Yeongtong-qu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea

3. Primary Contact Person

  • Name: JAESANG NOH A.
  • B. Title: Regulatory Affairs, Senior Professional
  • ﻥ Phone Number: +82-2-2193-2444
  • D. FAX Number: +82-31-200-6401
  • ட் E-Mail: jaesang.noh@samsung.com

Secondary Contact Person 4.

  • Name: Ninad Gujar A.
  • B. Title: Vice President, Regulatory Affairs & Quality Control
  • Phone Number: 978-564-8503 ﻥ
  • D. FAX Number: 978-560-0602
  • ட E-Mail: nqujar@samsungneurologica.com

Proposed Device 5.

  • A. Trade Name: Auto Lung Nodule Detection
  • B. Device Name: Auto Lung Nodule Detection
  • C. Common Name: Medical image analyzer
  • D. Classification Name: Medical image analyzer
  • ட Product Code: MYN
  • F. Regulation: 21 CFR 892.2070

Predicate Device 6.

  • A. Manufacturer: Riverain Medical Group, LLC
  • Trade Name: ClearRead Detect B.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.

510(k) Premarket Notification - Traditional

  • ﻥ Classification: Medical image analyzer
  • D. Product Code: MYN
  • ட 510(k) Number or PMA Number: P000041
  • F. Decision Date: July 12, 2001
Predicate Device
Trade NameClearRead Detect
Classification NameMedical image analyzer
Product CodeMYN
Regulation21 CFR 892.2070
PMA#P000041
Decision DateJuly 12, 2001

*The product code MYN has been reclassified from Class II since February 21, 2020.

7. Device Description

Auto Lung Nodule Detection is a Computer-Aided Detection (CADe) device that is designed to perform CAD processing in Chest X-ray images for indication of locations for high nodule probability, which has an effective detection sizes from 10 mm to 30 mm.

Auto Lung Nodule Detection receives images acquired with SAMSUNG Digital X-ray Imaging Systems as an input and identifies suspected nodules, and then sends information of suspected nodules to the visualization part of S-Station, which is installed on all kinds of SAMSUNG Digital X-ray Imaging Systems, to generate output images with circular marks. The CAD performed images, are displayed on the screen by S-Station without defeat of original images and used as a second reader only after the initial read is completed.

Intended Use 8.

The Auto Lung Nodule Detection is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 10 to 30 mm in size. It is designed to aid the physician to review the PA chest radiographs of adults as a second reader and be used as part of S-Station, which is operation software installed on Samsung Digital X-ray Imaging systems. Auto Lung Nodule Detection cannot be used on the patients who have lung lesions other than abnormal nodules.

9. Summary of Technological characteristic of the proposed device compared with the predicate device

Samsung believes that the proposed device is substantially equivalent to the predicate device because some differences in the design and features is considered low risk and do not raise new questions on safety and effectiveness of the proposed device for its intended use based on the non-clinical and clinical testing.

A. Comparing with Predicate Device

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, oval background.

510(k) Premarket Notification - Traditional

Predicate DeviceProposed Device
Device NameClearRead DetectAuto Lung NoduleDetection
ManufactureRiverain Medical Group,LLCSamsung Electronics
PMA NumberPMA P000041-
Indication for useCLEARREAD DETECT isa computer-aided detection(CAD) system intended toidentify and mark regionsof interest (ROIs) ondigitized frontal chestradiographs. It identifiesfeatures associated withsolitary pulmonary nodulesfrom 9 to 30 mm in size,which could representearly-stage lung cancer.The device is intended foruse as an aid only after thephysician has performedan initial interpretation ofthe radiograph.The Auto Lung NoduleDetection is computer-aided detection software toidentify and mark regionsin relation to suspectedpulmonary nodules from 10to 30 mm in size. It isdesigned to aid thephysician to review the PAchest radiographs of adultsas a second reader and beused as part of S-Station,which is operation softwareinstalled on SamsungDigital X-ray Imagingsystems. Auto Lung NoduleDetection cannot be usedon the patients who havelung lesions other thanabnormal nodules.
Intended usersPhysicianPhysician
Intended body partChestChest
Imaging modalityX-rayX-ray
Key featureIdentification of lungnodulesIdentification of lungnodules
Technologyheuristic decision rules,artificial neural network, andfuzzy logicMachine learning
Operating systemsStandard PC/WindowsStandard PC/Windows
InputImagetypeDICOMDICOM
ApplicableProtocolsChest PA/APChest PA
OutputOutputtypeROI marked on theduplicated input imageInformation for ROI to bemarked on the duplicatedinput image
Markertype/sizeCircular/AdjustableCircular/Fixed
ReportThe number of findingsThe number of nodulemarkers
Reader workflowSecond reader workflowSecond reader workflow

10. Safety and Effectiveness Information

Software design description, hazard analysis, and labeling information are provided in support of

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. The oval is tilted slightly upwards from left to right.

510(k) Premarket Notification - Traditional

this premarket notification for the proposed device. The device labeling contains instructions for use with cautions to provide for safe and effective use of the device. The results of the hazard analysis with appropriate risk controls indicate the proposed device is of moderate level of concern, as per the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Device" issued on May 11, 2005.

11. Software Verification and Validation

Non-clinical Testing A.

Non-clinical tests were conducted for Auto Lung Nodule Detection during the device development in accordance with Samsung verification and validation process complied with the FDA Quality System Regulations, ISO 13485 requirements, and the following standards.

  • -ISO14971:2007, Medical devices - Application of risk management to medical devices (2nd Ed.)
  • IEC62304:2006, Medical devices Software life cycle processes -
  • -IEC62366-1:2015, Medical device – Part1: Application of usability engineering to medical devices

Verification and validation activities for Auto Lung Nodule Detection were conducted to provide evidences that the design meets user needs and intended use and application specification. The testing results support that all the software specifications have met the acceptance criteria and the claims of substantial equivalence.

B. Clinical Performance Testing

Clinical evaluation was performed to validate the clinical efficacy of Samsung Auto Lung Nodule Detection (ALND) in helping radiologists find pulmonary nodules on digital chest radiographs. In this clinical study, nodule detection performances of human readers were measured using a test dataset containing both normal and diseased images. Readers were asked to mark their region of nodule suspicion on the images while also providing confidence scores on each decision they have made. After independent reading, readers were allowed to adjust their confidence scores after reviewing the ALND's detection results. Nodule detection performances before and after ALND were measured via sensitivity, false positives per image (FPPI), and jackknife alternative free response receiver operating characteristic (JAFROC) figure of merit (FOM). The results have demonstrated that all readers' nodule detection performances using the proposed device have increased with statistical significance. Therefore, the proposed device could provide potential assistance for radiologists in the interpretation and detection of pulmonary nodules when used as an assistant tool.

12. Conclusions

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the text "SAMSUNG ELECTRONICS Co., Ltd." The text is in a bold, sans-serif font. The text is black and the background is white.

Image /page/7/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue oval background. The oval is tilted slightly to the right.

510(k) Premarket Notification - Traditional

Results of all conducted testing were found acceptable in supporting the claim of substantial equivalent to the predicate device.

N/A