K Number
K201502

Validate with FDA (Live)

Device Name
Vitra 810
Manufacturer
Date Cleared
2020-08-06

(62 days)

Product Code
Regulation Number
886.4390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vitra 810 is intended for retinal photocoagulation, laser trabeculoplasty, transscleral cyclophotocoagulation, transscleral retinal photocoagulation, iridotomy, including the following:
Retinal photocoagulation for the treatment of:
Diabetic retinopathy including:
Nonproliferative retinopathy
Macular edema
Proliferative retinopathy
Retinal Tears, Detachments and Holes
Lattice degeneration
Age-related macular degeneration (AMD) with choroidal neovascularization (CNV)
Retinopathy of prematurity
Sub-retinal (choroidal) neovascularization
Central and Branch Retinal Vein Occlusion
Laser Trabeculoplasty, Iridotomy, Transscleral Cyclophotocoagulation for the treatment of glaucoma, including:
Primary open angle
Closed angle
Refractory Glaucoma (recalcitrant/uncontrolled)

Device Description

The Vitra 810 is a laser system which emits a treatment beam at 810 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye. The laser is used to treat the retinal diseases and glaucoma.
The Vitra 810 is a laser system which emits a treatment beam at 810 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye. The laser is used to treat the anterior and posterior segments of the eye. The laser is particularly well suited for treating the eye because it has minimal effect on transparent tissues and materials. This means that the laser can be efficiently delivered to opaque structures of the eye through the transparent cornea, aqueous humor, lens, and vitreous humor. This allows many conditions to be treated by non-invasive techniques.
Laser energy is delivered to opaque structures within the eye by means of laser delivery systems including slit lamp, indirect ophthalmoscope, operating microscope, cyclophotocoagulation probe and laser probe. The standard delivery system includes a lens system to focus the laser energy and vary the size of the laser spot in the plane of observation of the slit lamp, for example. The laser energy is delivered to the delivery system by the means of a flexible fiber optic. For most procedures, a laser contact lens is used to direct the laser energy to the part of the eye being treated. The contact lens may have mirrors so that laser energy can be delivered to areas of the retina behind the iris, or into the angle so that the trabecular meshwork can be treated. The contact lens also helps to hold the eye open and still so that the laser energy can be delivered effectively.

AI/ML Overview

The Quantel Medical Vitra 810 is an ophthalmic laser system intended for photocoagulation of ocular tissues in the treatment of diseases of the eye, specifically retinal diseases and glaucoma.

Here's an analysis of the acceptance criteria and the study information provided in the document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of quantitative acceptance criteria for device performance. Instead, substantial equivalence is claimed based on comparing the technological characteristics and intended use of the Vitra 810 with two predicate devices. The "Performance Data" section primarily lists compliance with recognized consensus standards related to safety and essential performance, as well as hardware and software validation. These can be considered the implicit acceptance criteria, demonstrating that the device meets established regulatory and technical benchmarks.

Acceptance Criteria CategoryDetails / Reported Device Performance
Intended UsePhotocoagulation of ocular tissue including the retina and pigmented tissue (Matches predicates). Specific indications for use (retinal photocoagulation for diabetic retinopathy, retinal tears, AMD, etc., and laser trabeculoplasty, iridotomy, transscleral cyclophotocoagulation for glaucoma) are identical to the secondary predicate (Iridex IQ 810).
Technological Characteristics (Comparison)Laser Source: Laser Diode (Matches predicates)
Wavelength: 810 nm (Matches predicates)
Treatment Power: 50-3000 mW (depending on delivery device) (Matches predicates)
Laser Safety Class: 4/IV (Matches predicates)
Pulse Duration (Continuous Wave): 10 ms to continuous (up to 60 seconds) (Matches predicates)
Pulse Duration (Subthreshold Mode - Ton): 0.1 ms to 1.0 ms (5% to 35% duty cycle) (Within range of Iridex IQ 810's 0.025 ms to 1.0 ms with 0.5% to 50% duty cycle)
Repeat Interval (Continuous Wave): 50 ms to 1000 ms and one pulse (Similar to Iridex IQ 810's 50 ms to 1000 ms and one pulse; predicate Supra Twin states "Repeat mode: 0.1 - 1 s")
Repeat Interval (Subthreshold Mode - Toff): 0.3 ms to 19 ms (duty cycle: 5% to 35%) (Matches Iridex IQ 810's 1.0 ms to 9.5 ms, 0.5% to 50% duty cycle within acceptable parameters)
Aiming Beam Source: Red Laser Diode, 1 mW maximum (Matches predicates)
Aiming Beam Wavelength: 635 - 650 nm (Matches predicates)
Compliance with StandardsIEC 60601-1:2005 + Corr. 1:2006+Corr. 2:2007+A1:2012 (General requirements for basic safety and essential performance)
IEC 60601-1-2:2014 (Electromagnetic compatibility)
IEC 60601-1-6:2010+A1:2013 (Usability)
IEC 60601-2-22: 2007 (Third Edition) + A1:2012 (Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment)
IEC 60825-1 (Safety of Laser Products – Equipment Classification and Requirements)
Software and Hardware ValidationHardware and software validation activities were performed to ensure the device performed as intended and software documentation appropriate for the major level of concern was provided.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not describe a clinical test set with a specific sample size, nor does it mention data provenance (e.g., country of origin, retrospective/prospective). The assessment for substantial equivalence is based on a comparison of technical specifications and intended use against legally marketed predicate devices, along with compliance to international safety standards. It does not appear to involve a new clinical study with a patient test set in the traditional sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable as the documentation does not describe a clinical study with a test set requiring expert-established ground truth. The device is a medical laser, and its substantial equivalence is determined by engineering and regulatory comparisons, not by diagnostic performance against an expert-derived ground truth.

4. Adjudication Method for the Test Set

This information is not applicable, for the same reasons as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This information is not applicable. The Vitra 810 is a surgical laser device (a treatment tool), not a diagnostic device or an AI-assisted diagnostic tool. Therefore, an MRMC study assessing human reader improvement with AI assistance is irrelevant for this type of medical device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

This information is not applicable. The Vitra 810 is a medical laser system operated by trained medical personnel. It is not an algorithm-only or AI-based device, so standalone algorithm performance testing is irrelevant.

7. The Type of Ground Truth Used

This information is not applicable. As discussed, the substantial equivalence decision is based on technical specifications, intended use, and compliance with safety standards, rather than the performance of a diagnostic algorithm against a clinical ground truth like pathology or outcomes data.

8. The Sample Size for the Training Set

This information is not applicable. There is no mention of a training set as the device is not an AI/ML-based system requiring data for training.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, for the same reasons as point 8.

{0}------------------------------------------------

August 6, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Quantel Medical % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180

Re: K201502

Trade/Device Name: Vitra 810 Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: HQF Dated: June 4, 2020 Received: June 5, 2020

Dear Ms. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201502

Device Name Vitra 810

Indications for Use (Describe)

The Vitra 810 is intended for retinal photocoagulation, laser trabeculoplasty, transscleral cyclophotocoagulation, transscleral retinal photocoagulation, iridotomy, including the following:

· Retinal photocoagulation for the treatment of:

  • o Diabetic retinopathy including:
    • Nonproliferative retinopathy
    • Macular edema
    • Proliferative retinopathy
  • o Retinal Tears, Detachments and Holes
  • o Lattice degeneration
  • o Age-related macular degeneration (AMD) with choroidal neovascularization (CNV)
  • o Retinopathy of prematurity
  • o Sub-retinal (choroidal) neovascularization
  • o Central and Branch Retinal Vein Occlusion

· Laser Trabeculoplasty, Iridotomy, Transscleral Cyclophotocoagulation for the treatment of glaucoma, including:

  • o Primary open angle
  • o Closed angle
  • o Refractory Glaucoma (recalcitrant/uncontrolled)
Type of Use (Select one or both, as applicable)
-------------------------------------------------
☒ Exporting to (Part 374.11(a) Subject to Prior
☐ On-Site Consumption (Part 374.11(a) Subject to

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY K201502

Quantel Medical Vitra 810®

510(k) Owner

Quantel Medical 11 rue Bois Joli CS40015 63808 Cournon D'Auvergne-Cedex France Phone: 33 04 73 745 732 Contact Person: Bruno Pagès

Submission Correspondent:

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180 Phone: 978-207-1245

Date Prepared: August 4, 2020

Trade Name of Device

Vitra 810 ®

Common or Usual Name

Powered Surgical Laser

Classification Name

Powered Laser Surgical Instrument; 21 C.F.R. 886.4390 Class II Product Code: HQF

Primary Predicate Device

Quantel Medical Supra Twin Ophthalmic Laser (K081946) Product Code: HQF

Secondary Predicate Device

Iridex IQ 810 (K071687) Product Code: GEX

{4}------------------------------------------------

Device Description

The Vitra 810 is a laser system which emits a treatment beam at 810 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eve. The laser is used to treat the retinal diseases and glaucoma.

The Vitra 810 is a laser system which emits a treatment beam at 810 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye. The laser is used to treat the anterior and posterior segments of the eye. The laser is particularly well suited for treating the eye because it has minimal effect on transparent tissues and materials. This means that the laser can be efficiently delivered to opaque structures of the eve through the transparent cornea, aqueous humor, lens, and vitreous humor. This allows many conditions to be treated by non-invasive techniques.

Laser energy is delivered to opaque structures within the eye by means of laser delivery systems including slit lamp, indirect ophthalmoscope, operating microscope, cvclophotocoagulation probe and laser probe. The standard delivery system includes a lens system to focus the laser energy and vary the size of the laser spot in the plane of observation of the slit lamp, for example. The laser energy is delivered to the delivery system by the means of a flexible fiber optic. For most procedures, a laser contact lens is used to direct the laser energy to the part of the eye being treated. The contact lens may have mirrors so that laser energy can be delivered to areas of the retina behind the iris, or into the angle so that the trabecular meshwork can be treated. The contact lens also helps to hold the eye open and still so that the laser energy can be delivered effectively.

Intended Use / Indications for Use

The Vitra 810 is intended for retinal photocoagulation, laser trabeculoplasty, transscleral cyclophotocoagulation, transscleral retinal photocoagulation, iridotomy, including the following:

  • Retinal photocoagulation for the treatment of: .
    • o Diabetic retinopathy including:
      • . Nonproliferative retinopathy
      • . Macular edema
      • . Proliferative retinopathy
    • Retinal Tears, Detachments and Holes o
    • Lattice degeneration o
    • Age-related macular degeneration (AMD) with choroidal neovascularization O (CNV)
    • Retinopathy of prematurity O
    • Sub-retinal (choroidal) neovascularization o
    • Central and Branch Retinal Vein Occlusion o
  • Laser Trabeculoplasty, Iridotomy, Transscleral Cyclophotocoagulation for the . treatment of glaucoma, including:
    • Primary open angle o
    • Closed angle o

{5}------------------------------------------------

  • o Refractory Glaucoma (recalcitrant/uncontrolled)

Substantial Equivalence

Quantel Medical believes that the Vitra 810 described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to legally marketed predicate devices that are Class II medical devices. The primary predicate device is the Quantel Medical Supra Twin Ophthalmic Laser cleared by FDA in K081946. The secondary predicate device is the Iridex IO 810 laser system cleared by FDA in K071687. All three devices are intended for use in photocoagulation of ocular tissue including the retina and pigmented tissue. All three devices are prescription devices which are intended to be used by trained medical personnel. Therefore, the Vitra 810 has the same intended use as the identified primary predicate device and an identical indications for use statement to the secondary predicate device and may be found to be substantially equivalent to the predicate devices.

As shown in Table 1, the Vitra 810, the Supra Twin and the Iridex IQ 810 have the same technological characteristics. All three devices are laser diodes with a wavelength of 810 mm. The treatment power ranges from 50-3000 mW for all three devices. The pulse duration is 10 ms to continuous (up to 60 seconds) for continuous wave mode in the Vitra 810, the Supra Twin and Iridex IQ 810. For the Subthreshold mode the pulse duration is 0.1 ms with a 5% to 35% duty cycle which is within the pulse duration range for the Iridex IQ 810 which is 0.025 ms to 1.0 ms (0.5% to 50% duty cycle). All three device use an aiming laser with a wavelength of 635-650 nm with a 1 mW maximum power.

All the systems are compatible with a variety of delivery systems including a slit lamp, indirect ophthalmoscope, operating microscope, laser probes and cyclophotocoagulation probes.

Therefore, the Vitra 810 and the primary predicate device, the Quantel Medical Supra Twin, supported by the Iridex IO810 secondary predicate device. have the same intended use and similar technological characteristics and are therefore substantially equivalent.

{6}------------------------------------------------

Table1

Vitra 810 Substantial Equivalence

MANUFACTURERQUANTEL MEDICALQUANTEL MEDICALIRIDEX
DEVICE NAMEVitra 810Supra TwinIQ 810
510(K)-K081946K071687
PRODUCT CODEHQFHQFGEX
INTENDED USEPhotocoagulation of ocular tissue including the retina and pigmented tissuePhotocoagulation of ocular tissue including the retina and pigmented tissuePhotocoagulation of ocular tissue including the retina and pigmented tissue
INDICATIONS FOR USEThe Vitra 810 is intended for retinal photocoagulation, laser trabeculoplasty, transscleral cyclophotocoagulation, transscleral retinal photocoagulation, iridotomy, including the following: Retinal photocoagulation for the treatment of: Diabetic retinopathy including: Nonproliferative retinopathy Macular edema Proliferative retinopathy Retinal Tears, Detachments and Holes Lattice degeneration Age-related macular degeneration (AMD) with choroidal neovascularization (CNV) Retinopathy of prematurity Sub-retinal (choroidal) neovascularization Central and Branch Retinal Vein810 nm Indications for Use:The Supra Twin Ophthalmic Laser Photocoagulator wavelength 810nm is indicated for use:Photocoagulation or ablation of pigmented tissue within the eye,Transscleral ciliary body ablation (treatment is reserved for patients with chronic glaucoma and those not responding to conventional treatments),Transpupillary photocoagulation,Endophotocoagulation,Treatment of complicated rhegmatogeous, tractional retinal detachments, proliferative vitreoretinopathy, proliferative diabetic retinopathy, macular degeneration, peripheral photocoagulation (recumben patients), transpupillary photocoagulation of choroidalThe Iridex IQ 810 is intended for retinal photocoagulation, laser trabeculoplasty, transscleral cyclophotocoagulation, transscleral retinal photocoagulation, iridotomy, including the following: Retinal photocoagulation for the treatment of: Diabetic retinopathy including: Nonproliferative retinopathy Macular edema Proliferative retinopathy Retinal Tears, Detachments and Holes Lattice degeneration Age-related macular degeneration (AMD) with choroidal neovascularization (CNV) Retinopathy of prematurity Sub-retinal (choroidal) neovascularization Central and Branch Retinal Vein Occlusion
Occlusion• Laser Trabeculoplasty, Iridotomy,Transscleral Cyclophotocoagulationfor the treatment of glaucoma,including:o Primary open angleo Closed angleo Refractory Glaucoma(recalcitrant/uncontrolled)neovasculature, and Age relatedmacular degeneration (AMD)treatments.• Laser Trabeculoplasty, Iridotomy,Transscleral Cyclophotocoagulation forthe treatment of glaucoma, including:o Primary open angleo Closed angle• Refractory Glaucoma(recalcitrant/uncontrolled)
TYPEOF DELIVERYSYSTEMSlit lampIndirect ophthalmoscopeOperating microscopeCyclophotocoagulation probesLaser probesSlit lampIndirect ophthalmoscopeOperating microscopeEndocular probeSlit lampIndirect ophthalmoscopeOperating microscopeCyclophotocoagulation probesLaser probes
LASER SOURCELaser DiodeLaser DiodeLaser Diode
WAVELENGTH810 nm810 nm810 nm
TREATMENT POWER50-3000mW (depending upon deliverydevice)50-3000mW (depending upon deliverydevice)50-3000mW (depending upon deliverydevice)
LASER SAFETY CLASS4/IV4/IV4/IV
PULSE DURATIONContinuous wave mode: 10 ms tocontinuous (up to 60 seconds)Subthreshold (Subliminal) mode (Ton):0.1 ms to1.0 ms (5% to 35% duty cycle)Continuous wave mode: 10 ms to 60secondsRepeat mode: 0.1, 0.15, 0.2, 0.3, 0.4,0.5, 0.6, 0.7 & 1sContinuous wave mode: 10 ms to continuous(up to 60 seconds)Subthreshold mode (Ton): 0.025 ms to1.0 ms(0.5% to 50% duty cycle)TTT mode: 10 seconds to 1800 seconds
REPEAT INTERVALContinuous wave mode: 50 ms to 1000 msand one pulseSubthreshold (Subliminal) mode (Toff):0.3 ms to 19 ms (duty cycle: 5% to 35%)Repeat mode: 0.1 – 1 sContinuous wave mode: 50 ms to 1000 msand one pulseSubthreshold mode (Toff): 1.0 ms to 9.5 ms(duty cycle: 0.5% to 50%)
AIMING BEAM SOURCERed Laser Diode, 1 mW maximumRed Laser Diode, 1 mW maximumRed Laser Diode, 1 mW maximum
AIMINGBEAMWAVELENGTH635 - 650 nm635-650 nm635 - 650 nm
SIZE18 (H) x 19.5 (W) x 30 (D) cm7.1" (H) x 7.7" (W) x 11.8" (D)14.6 cm High x 33 cm Wide x 30.7cm
WEIGHT5.6 kg / 12.3lbs9.5 kg / 21 lbs.-
COOLINGBy Peltier effectBy Peltier effectAir cooled
POWER REQUIREMENT250 VA240 V-

{7}------------------------------------------------

K201502

{8}------------------------------------------------

K201502

{9}------------------------------------------------

Performance Data

Performance testing was conducted in order to demonstrate compliance with recognized consensus standards and to demonstrate substantial equivalence:

  • IEC 60601-1:2005 + Corr. 1:2006+Corr. 2:2007+A1:2012 Medical electrical ● equipment-Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2014: Medical electrical equipment-Part 1-2: General requirements for ● basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests
  • IEC 60601-1-6:2010+A1:2013 Medical electrical equipment-Part 1-6: General . requirements for safety-Collateral Standard Usability
  • . IEC 60601-2-22: 2007 (Third Edition) + A1:2012 Medical Electrical Equipment Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
  • IEC 60825-1 Safety of Laser Products-Part 1: Equipment Classification and . Requirements

Additionally, hardware and software validation activities were performed to ensure the device performed as intended and software documentation appropriate for the major level of concern was provided.

Conclusions

The Vitra 810 and the primary predicate device, the Quantel Medical Supra Twin, supported by the Iridex IQ810 secondary predicate device have the same intended use and similar technological characteristics and are therefore substantially equivalent.

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.