K Number
K201476

Validate with FDA (Live)

Date Cleared
2020-08-28

(86 days)

Product Code
Regulation Number
862.1358
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DreaMed Advisor Pro is a decision-support software intended for assisting healthcare professionals in the management of patients with Type 1 diabetes who:

· use insulin pumps as their insulin delivery therapy;

· monitor their glucose levels using continuous glucose monitoring (CGM) and/or Self-Monitoring Blood Glucose (SMBG) meter;

• are above the age of 6; and

· use rapid acting U-100 insulin analogs in their pump

DreaMed Advisor Pro is indicated for use by healthcare professionals when analyzing CGM, SMBG and pump data to generate recommendations for optimizing a patient's insulin pump settings for basal rate, Carbohydrate Ratio (CR), and Correction Factor (CF); without considering the full clinical status of a particular patient. DreaMed Advisor Pro does not replace clinical judgement.

Device Description

DreaMed Advisor Pro is a software device that is designed to provide insulin therapy adjustment recommendations to physicians to assist in the management of diabetes for patients with Type 1 diabetes using an insulin pump, a continuous glucose monitoring (CGM) system and self-management blood glucose meter (SMBG).

The DreaMed Advisor Pro gathers and analyzes information inputted through qualified Diabetes Management Systems (DMS), which collects biological input information from various diabetes devices. Diabetes device information required and used by DreaMed Advisor Pro includes glucose readings (either CGM sensor readings and/or capillary blood glucose measurements), insulin dosing logs, and meal data during daily routine care.

Following data collection and analysis, the DreaMed Advisor Pro generates results containing summary data and recommendations for adjustments to the patient's insulin therapy parameters, including basal insulin delivery rate(s), insulin to carbohydrate ratio and correction factor (insulin sensitivity). DreaMed Advisor Pro may also advise behavioral changes. Results are sent to a qualified Diabetes Management Systems, which displays results to physicians and a report provided by DreaMed Diabetes. The physician can approve, reject or change the recommendations and issue the updated treatment plan to the patient.

AI/ML Overview

The provided text is a 510(k) summary for the DreaMed Advisor Pro, which is a software device intended to assist healthcare professionals in managing Type 1 diabetes. It primarily describes the device, its intended use, and argues for its substantial equivalence to a previously cleared version of the same device (K191370).

However, the document does not contain the detailed information requested about acceptance criteria, specific study results, sample sizes for test and training sets, expert qualifications, adjudication methods, or MRMC comparative effectiveness studies. It simply asserts that the updated DreaMed Advisor Pro is "as safe and effective" as the previously cleared one, with the only change being an expansion of the age range to include patients over 65.

Therefore, I cannot populate the table and answer all the questions based on the provided text. I will indicate where the information is missing.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Metric, Threshold)Reported Device Performance
Not provided in the document. The document asserts substantial equivalence to a predicate device rather than detailing specific performance criteria or the results of a new study proving these.Not provided in the document. No specific performance metrics or thresholds are given, nor are results of a study designed to meet such criteria.

2. Sample size used for the test set and the data provenance

  • Sample size for test set: Not provided in the document.
  • Data provenance: Not provided in the document. No specific study data is presented, so the origin (country, retrospective/prospective) of any potential test data is not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not provided in the document. Since no specific test set study is detailed, there's no information on ground truth establishment or expert involvement.


4. Adjudication method for the test set

Not provided in the document.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not provided in the document. The submission focuses on substantial equivalence to a previous version of the same AI device, not on comparing human readers with and without the AI.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document describes the device as "decision-support software intended for assisting healthcare professionals," and states that "DreaMed Advisor Pro does not replace clinical judgement." This implies human-in-the-loop operation. No standalone (algorithm-only) performance is described or implied to have been studied for this 510(k) submission.


7. The type of ground truth used

Not provided in the document. No specific ground truth methodology is mentioned as a study proving performance is not detailed in this submission.


8. The sample size for the training set

Not provided in the document.


9. How the ground truth for the training set was established

Not provided in the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 28, 2020

DreaMed Diabetes Ltd Inbal Beinglass Peled, Director RA/QA 5 Mota Gur St Petah Tikva, 4952701 Israel

Re: K201476

Trade/Device Name: DreaMed Advisor Pro Regulation Number: 21 CFR 862.1358 Regulation Name: Insulin therapy adjustment device Regulatory Class: Class II Product Code: QCC Dated: June 3, 2020 Received: June 3, 2020

Dear Inbal Beinglass Peled:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201476

Device Name DreaMed Advisor Pro

Indications for Use (Describe)

DreaMed Advisor Pro is a decision-support software intended for assisting healthcare professionals in the management of patients with Type 1 diabetes who:

· use insulin pumps as their insulin delivery therapy;

· monitor their glucose levels using continuous glucose monitoring (CGM) and/or Self-Monitoring Blood Glucose (SMBG) meter;

• are above the age of 6; and

· use rapid acting U-100 insulin analogs in their pump

DreaMed Advisor Pro is indicated for use by healthcare professionals when analyzing CGM, SMBG and pump data to generate recommendations for optimizing a patient's insulin pump settings for basal rate, Carbohydrate Ratio (CR), and Correction Factor (CF); without considering the full clinical status of a particular patient. DreaMed Advisor Pro does not replace clinical judgement.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY DreaMed Advisor Pro

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

DreaMed Diabetes Ltd. 5 Mota Gur St. Petah Tikva 4952701 lsrael

Contact Person: Inbal Beinglass Peled Email: inbal.peled@dreamed.ai Phone: +972-52-3642760 Date Prepared: August 28, 2020

Name of Device and Name/Address of Sponsor

DreaMed Advisor Pro DreaMed Diabetes Ltd. 5 Mota Gur Street Petah Tikva 4952701 Israel

Common or Usual Name

Insulin Therapy Adjustment Device

Classification

Class II, 21 CFR 862.1358, Product Code: QCC

Predicate Devices

DreaMed Advisor Pro, K191370

Intended Use / Indications for Use

DreaMed Advisor Pro is a decision-support software intended for assisting healthcare professionals in the management of patients with Type 1 diabetes who:

  • . use insulin pumps as their insulin delivery therapy;
  • . monitor their glucose levels using continuous glucose monitoring (CGM) and/or Self-Monitoring Blood Glucose (SMBG) meter;
  • are above the age of 6; and
  • . use rapid acting U-100 insulin analogs in their pump

DreaMed Advisor Pro is indicated for use by healthcare professionals when analyzing CGM, SMBG and pump data to generate recommendations for optimizing a patient's insulin pump settings for basal rate, Carbohydrate Ratio (CR), and Correction Factor (CF); without considering the full clinical status of a particular patient. DreaMed Advisor Pro does not replace clinical judgement.

{4}------------------------------------------------

Device Description

DreaMed Advisor Pro is a software device that is designed to provide insulin therapy adjustment recommendations to physicians to assist in the management of diabetes for patients with Type 1 diabetes using an insulin pump, a continuous glucose monitoring (CGM) system and self-management blood glucose meter (SMBG).

The DreaMed Advisor Pro gathers and analyzes information inputted through qualified Diabetes Management Systems (DMS), which collects biological input information from various diabetes devices. Diabetes device information required and used by DreaMed Advisor Pro includes glucose readings (either CGM sensor readings and/or capillary blood glucose measurements), insulin dosing logs, and meal data during daily routine care.

Following data collection and analysis, the DreaMed Advisor Pro generates results containing summary data and recommendations for adjustments to the patient's insulin therapy parameters, including basal insulin delivery rate(s), insulin to carbohydrate ratio and correction factor (insulin sensitivity). DreaMed Advisor Pro may also advise behavioral changes. Results are sent to a qualified Diabetes Management Systems, which displays results to physicians and a report provided by DreaMed Diabetes. The physician can approve, reject or change the recommendations and issue the updated treatment plan to the patient.

Substantial Equivalence

The Advisor Pro is as safe and effective as the previously cleared Advisor Pro. The Advisor Pro has the same technological characteristics and principles of operation as the previously cleared Advisor Pro.

With respect to the indication for use we conclude that the differences in the indication for use between the previous and updated Advisor Pro in allowing patients over the age of 65 to use the device do not affect the safety and effectiveness of the device. This is because based on literature and feedback from clinicians, patients do not receive different insulin therapy management based on their age alone but rather it is based on their overall health and any such considerations are already included in the current labeling. Thus, physicians treating patients over the age of 65 who are otherwise healthy and meet all other indications, device specifications and are not contraindicated could use the device. Anyone else not falling under the indications, falling under the contraindications or not meeting device specifications would not be treated using the device. The indications for use differences between the Advisor Pro and its predicate devices raise no new issues of safety or effectiveness. Thus, the Advisor Pro is substantially equivalent to its predicate device.

Conclusions

DreaMed Diabetes believes that the changes as described in this 510(k) submission, do not present additional safety or effectiveness concerns for the DreaMed Advisor Pro, which is a modification of the legally marketed DreaMed Advisor Pro (K191370).

§ 862.1358 Insulin therapy adjustment device.

(a)
Identification. An insulin therapy adjustment device is a device intended to incorporate biological inputs, including glucose measurement data from a continuous glucose monitor, to recommend insulin therapy adjustments as an aid in optimizing insulin therapy regimens for patients with diabetes mellitus.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) A complete description of the required data inputs, including timeframe over which data inputs must be collected and number of data points required for accurate recommendations;
(ii) A complete description of the types of device outputs and insulin therapy adjustment recommendations, including how the recommendations are generated;
(iii) Robust data demonstrating the clinical validity of the device outputs and insulin therapy recommendations;
(iv) A robust assessment of all input data specifications, including accuracy requirements for continuous glucose monitors and other devices generating data inputs, to ensure accurate and reliable therapy adjustment recommendations. This assessment must include adequate clinical justification for each specification;
(v) A detailed strategy to ensure secure and reliable means of data transmission to and from the device, including data integrity checks, accuracy checks, reliability checks, and security measures;
(vi) Robust data demonstrating that users can understand and appropriately interpret recommendations generated by the device; and
(vii) An appropriate mitigation strategy to minimize the occurrence of dosing recommendation errors, and to mitigate the risk to patients of any residual dosing recommendation errors to a clinically acceptable level.
(2) The device must not be intended for use in implementing automated insulin dosing.
(3) Your 21 CFR 809.10(b) labeling must include:
(i) The identification of specific insulin formulations that have been demonstrated to be compatible with use of the device;
(ii) A detailed description of the specifications of compatible devices that provide acceptable input data (e.g., continuous glucose monitors, insulin pumps) used to provide accurate and reliable therapy adjustment recommendations;
(iii) A detailed description of all types of required data (inputs) and dosing recommendations (outputs) that are provided by the device; and
(iv) A description of device limitations, and instructions to prevent possible disruption of accurate therapy adjustment recommendations (e.g., time zone changes due to travel).