(408 days)
The ExoAtlet-II is intended to perform ambulatory functions in rehabilitation under the supervision of a trained physical therapist for the following population with upper extremity motor function at least 4/5 in both arms:
· Individuals with spinal cord injuries at levels T4 to L5
- · Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D)
The therapist must complete a training program prior to use of the device.
The device is not intended for sports or stair climbing.
The ExoAtlet-II is intended to perform ambulatory functions in rehabilitation institutions, hospitals, or similar institutions under the supervision of a healthcare professional (HCP) such as a trained physical therapist. The ExoAtlet-II is indicated for individuals with spinal cord injuries at levels T4-L5 and individuals with spinal cord injuries at levels of C7-T3. All users must have upper extremity motor function of at least 4/5 in both arms. The ExoAtlet-II is intended for multiple users in a clinical setting (i.e., gait lab or rehabilitation center) under supervision. The ExoAtlet-II includes a battery powered exoskeleton, an ExoCrutch, and a Tablet PC.
ExoAtlet-II is designed to be operated with a trained physical therapist who has received preliminary training on the Exoskeleton operation. This therapist will be familiar with the rules of the use and operational documentation. Training is conducted in a designated medical facility that is representative of the indicated environments (e.g., hospital, outpatient physical therapy clinic or similar). The training identifies safe environments of device use and uses all of the safety features found in the subject device. The training manual will be provided with the user manual along with a trainee itinerary and a trainee certification exam. The trainee needs to pass 80% of the exam questions to obtain certification.
Here's a breakdown of the acceptance criteria and study information for the ExoAtlet-II device based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it describes general performance characteristics and notes differences from the predicate device that were evaluated for safety and effectiveness through testing.
However, based on the "Comparison of Technological Characteristics" table and the "Summary of Clinical Testing" section, we can infer some key performance aspects and their reported outcomes:
| Feature/Metric | Acceptance Criteria (Inferred from Predicate/Safety Evaluation) | Reported Device Performance (ExoAtlet-II) |
|---|---|---|
| Indications for Use | Safe and effective for specified SCI populations. | Intended for individuals with spinal cord injuries at T4-L5 and C7-T3 (ASIA D) with 4/5 upper extremity motor function. |
| Mobility Ability | Sit, Stand, Walk, Turn (predicate) | Sit, Stand, Walk (turning capability not included, evaluated via clinical testing). |
| Walking Speed | ~2 km/hr (predicate) | 1.5 km/hr (slightly slower, evaluated via clinical testing, does not pose increased safety risk). |
| Grade of Inclination | 1.15° (predicate) | 0-5° (greater inclination, evaluated via clinical testing for safety). |
| Range of Motion | Hips: 135° flex to 20° ext; Knees: 130° flex to 0° ext; Ankles: 10° flex to 10° ext (predicate) | Hips: 108° flex to 21° ext; Knees: 112° flex to 2° ext; Ankles: 15° flex to 0-15° ext (larger range, evaluated via clinical testing for safety/effectiveness). |
| Crutch Use | Safe and effective | Evaluated in compliance with ISO 11334-1:2007. |
| Failsafe Feature | Safely manage failure event | Knees and pelvis become locked (different from predicate, evaluated via clinical testing for safety). |
| Electrical Safety | Compliance with IEC 60601-1 | Complies with IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1: 2012. |
| EMC Compatibility | Compliance with IEC 60601-1-2 | Complies with IEC 60601-1-2: 2014. |
| Software Integrity | Compliance with IEC 62304 | Complies with IEC 62304:2006. |
| Battery Safety/Performance | Safe, functional battery | Tested in compliance with IEC 62133: 2012. 4 hours continuous usage, 4 hours charge time. Evaluated electrical safety, EMC, and clinical testing. |
| Durability | Expected life comparable to predicate | 5 years (longer than predicate's 4 years), evaluated via durability testing. |
| 10 Meter Walk Test | Successful completion | 19 of 20 subjects successfully completed. |
| 6-Minute Walk Test (6 MWT) | Successful completion | 19 of 20 subjects successfully completed (1 unable due to physical fitness). |
| User Comfort Survey | Positive user experience | 18 of 20 subjects rated comfort 8-10 out of 10; 2 subjects rated 6 out of 10. |
| Skin Assessment (AEs) | No adverse events related to biocompatibility | None found (abrasions, dry flaky skin, cuts, eschar, rash) before, during, or after use. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical Study): 20 subjects (n=20)
- Data Provenance: The study was described as an "open label, non-randomized clinical study." The document does not specify the country of origin but implies it was internal to ExoAtlet given "ExoAtlet conducted an... clinical study." The study is prospective as it was conducted to assess the safety and effectiveness of the device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not mention the use of experts to establish a "ground truth" for a test set in the traditional sense of diagnostic accuracy. Instead, the clinical study involved subjects using the device under the supervision of a "trained physical therapist." The therapist's role is to supervise, and their training and certification are mentioned as a requirement for device use. The study monitors objective performance (walk tests) and subjective feedback (comfort surveys, skin assessments) directly from the subjects and observations.
4. Adjudication Method for the Test Set
Not applicable. The clinical study described is a performance and safety study involving a device for rehabilitation, not a diagnostic study requiring adjudication of expert interpretations for ground truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic device, but a powered exoskeleton for rehabilitation. Therefore, an MRMC study and effects on human reader performance are not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
Not applicable. The ExoAtlet-II is a physical device intended to be used by individuals under the supervision of a trained physical therapist. It is inherently a "human-in-the-loop" device by design and intended use. There is no mention of a standalone algorithm performance test.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The clinical study assessed the device's safety and effectiveness directly through:
- Performance metrics: Successful completion of the 10 Meter Walk Test and 6-Minute Walk Test.
- User feedback: Comfort ratings via surveys.
- Medical observation: Skin assessments for adverse events.
- Overall safety and functionality: Absence of new safety issues or ineffectiveness compared to predicates.
These are direct measures of the device's functional performance and user impact, rather than a "ground truth" established through external pathological or expert consensus for a diagnostic task.
8. The Sample Size for the Training Set
The document does not provide information about a "training set" in the context of machine learning or AI. The training described is for the human therapist who uses the device.
- Training for Therapists: Not a "training set" for the device's algorithm, but a program for healthcare professionals. The therapist completes a training program and must pass 80% of an exam for certification.
9. How the Ground Truth for the Training Set Was Established
As noted above, there is no AI/ML training set mentioned. The "training set" for therapists involves instructions, operational documentation, and an exam, but this is human training, not algorithm training data.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name in a lighter shade of blue.
July 16, 2021
ExoAtlet Asia Co. Ltd. Allison Komiyama, PhD Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave. Suite B-257 San Diego, California 92110
Re: K201473
Trade/Device Name: ExoAtlet-II Regulation Number: 21 CFR 890.3480 Regulation Name: Powered Lower Extremity Exoskeleton Regulatory Class: Class II Product Code: PHL Dated: June 16, 2021 Received: June 16, 2021
Dear Dr. Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Tushar Bansal -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are evenly spaced, and the overall impression is clean and legible.
For Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201473
Device Name ExoAtlet-II
Indications for Use (Describe)
The ExoAtlet-II is intended to perform ambulatory functions in rehabilitation under the supervision of a trained physical therapist for the following population with upper extremity motor function at least 4/5 in both arms:
· Individuals with spinal cord injuries at levels T4 to L5
- · Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D)
The therapist must complete a training program prior to use of the device.
The device is not intended for sports or stair climbing.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for EXOATLET. The logo is in a sans-serif font, with the "EXO" portion in a bright blue color and the "ATLET" portion in black. The letters are all capitalized and have a modern, clean design.
DATE PREPARED
July 16, 2021
MANUFACTURER AND 510(k) OWNER
ExoAtlet Asia Co. Ltd. 4582 S. Ulster Street, Suite 205 Denver, CO 80237 Telephone: Official Contact: Bret Bostock, CEO ExoAtlet US
REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Michelle Rubin-Onur, Ph.D. AcKnowledge Regulatory Strategies, LLC +1 (619) 458-9547 Telephone: Email: akomiyama@acknowledge-rs.com, mrubin@acknowledge-rs.com Website: https://www.acknowledge-rs.com/
DEVICE INFORMATION
| Proprietary Name/Trade Name: | ExoAtlet-II |
|---|---|
| Common Name: | Powered Exoskeleton |
| Regulation Number: | 21 CFR 890.3480 |
| Class: | Class II |
| Product Code: | PHL |
| Premarket Review: | Neurology |
| Review Panel: | Neuromodulation and Physical Medicine Devices (DHT5B) |
PREDICATE DEVICE IDENTIFICATION
The ExoAtlet-II is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---|---|---|
| K143690 | Ekso™ (version 1.1) and Ekso GT™ (version 1.2) /Ekso Bionics, Inc. | ✓ |
The predicate devices have not been subject to a design related recall.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for EXOATLET. The logo is written in a sans-serif font, with the first three letters, "EXO," in blue. The remaining letters, "ATLET," are in gray. The logo is simple and modern.
DEVICE DESCRIPTION
The ExoAtlet-II is intended to perform ambulatory functions in rehabilitation institutions, hospitals, or similar institutions under the supervision of a healthcare professional (HCP) such as a trained physical therapist. The ExoAtlet-II is indicated for individuals with spinal cord injuries at levels T4-L5 and individuals with spinal cord injuries at levels of C7-T3. All users must have upper extremity motor function of at least 4/5 in both arms. The ExoAtlet-II is intended for multiple users in a clinical setting (i.e., gait lab or rehabilitation center) under supervision. The ExoAtlet-II includes a battery powered exoskeleton, an ExoCrutch, and a Tablet PC.
ExoAtlet-II is designed to be operated with a trained physical therapist who has received preliminary training on the Exoskeleton operation. This therapist will be familiar with the rules of the use and operational documentation. Training is conducted in a designated medical facility that is representative of the indicated environments (e.g., hospital, outpatient physical therapy clinic or similar). The training identifies safe environments of device use and uses all of the safety features found in the subject device. The training manual will be provided with the user manual along with a trainee itinerary and a trainee certification exam. The trainee needs to pass 80% of the exam questions to obtain certification.
INDICATIONS FOR USE
The ExoAtlet-II is intended to perform ambulatory functions in rehabilitation under the supervision of a trained physical therapist for the following population with upper extremity motor function at least 4/5 in both arms:
- Individuals with spinal cord injuries at levels T4 to L5
- Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D)
The therapist must successfully complete a training program prior to use of the device. The device is not intended for sports or stair climbing.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
ExoAtlet believes that the ExoAtlet-II is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design and dimensions as the devices cleared in K143690. The subject device has the same intended use and similar technological characteristics (i.e., energy source, maximum user weight, and type of surface the device can function on) to the devices cleared in K143690. The ExoAtlet-II has undergone testing to ensure that any differences in technological characteristics (i.e., ability of user mobility, walking speed, grade of inclination, and range of motion) do not affect safety and effectiveness when compared to the predicate devices.
{5}------------------------------------------------
EXOATLET
| Subject Device | Predicate Device | ||
|---|---|---|---|
| ExoAtlet Asia Co. Ltd. | Ekso Bionics, Inc. | Statement of | |
| Equivalence | |||
| ExoAtlet-II | Ekso™ (version 1.1) and Ekso GT™ (version 1.2) | ||
| K201473 | K143690 | ||
| Indications for | The ExoAtlet-II is intended to perform ambulatory functions in | The Ekso™ (version 1.1) and Ekso GT™ (version 1.2) are intended | Different to the |
| Use | rehabilitation institutions under the supervision of a trained | to perform ambulatory functions in rehabilitation institutions | predicate. The subject |
| physical therapist for the following population with upper | under the supervision of a trained physical therapist for the | device is not intended to | |
| extremity motor function at least 4/5 in both arms: | following population with upper extremity motor function of at | be used by individuals | |
| Individuals with spinal cord injuries at | least 4/5 in both arms: | with hemiplegia whereas | |
| levels T4 to L5 | ● Individuals with hemiplegia due to stroke | the predicate device is. | |
| Individuals with spinal cord injuries at | ● Individuals with spinal cord injuries at levels T4 to | Both the subject and | |
| levels of C7 to T3 (ASIA D) | L5 | predicate device includea training program prior | |
| The therapist must successfully complete a training program | ● Individuals with spinal cord injuries at levels ofC7 to T3 (ASIA D). | to use of the device. | |
| prior to use of the device. The device is not intended for sports | The therapist must complete a training program prior to use of | Both devices are not | |
| or stair climbing. | the device. The devices are not intended for sports or stair | intended for sports or | |
| climbing. | stair climbing. | ||
| Product Codes / | PHL / 21 CFR 890.3480 | PHL / 21 CFR 890.3480 | Identical to the |
| Regulation | predicate. No impact on | ||
| Number | safety or effectiveness. | ||
| Regulation | Powered lower extremity exoskeleton | Powered lower extremity exoskeleton | Identical to the |
| Description | predicate. No impact on | ||
| safety and effectiveness. | |||
| Technical Characteristics | |||
| Components | ExoskeletonExoCrutchTablet PC | ● Exoskeleton● Mobile application | Different to the |
| predicate device. Clinical | |||
| evidence demonstratesthat there is no impact | |||
| on safety and | |||
| effectiveness. | |||
| Exoskeleton | 25 kg | 23 kg | Different to the |
| Weight | predicate device. | ||
| Performance and clinical | |||
| testing demonstrate that | |||
| there is no impact on | |||
| safety and effectiveness. | |||
| Body Coverage | Worn over legs and upper body with rigid torso | Worn over legs and upper body with rigid torso | Identical to the |
| predicate device. No | |||
| impact on safety and | |||
| effectiveness. | |||
| Size of | Adjustable upper leg, lower leg, and hip width | Adjustable upper leg, lower leg, and hip width , control unit | Identical to the |
| Components | integrated into the torso | predicate device. No | |
| impact on safety and | |||
| effectiveness. | |||
| Ability of UserMobility | |||
| SitStandWalk | SitStandWalkTurn | Different from thepredicate device. TheExoAtlet-II does notturn. Effectiveness of thesit, stand, and walkfunctions wereevaluated throughclinical testing. | |
| Mobility Aid | Crutch, Walker | Crutch, Walker | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| Energy Source | Rechargeable lithium batteries 25.55V, 5700mAh, 4 hours ofcontinuous usage per charge | Rechargeable lithium batteries 48.1V, 30A peak current, 1 hour ofcontinuous usage per charge | Different from thepredicate device. Thebattery of the ExoAtlet-IIcan be charged forlonger. Safety concernsregarding battery safetywere evaluated throughelectrical safety andEMC testing. |
| Maximum UserWeight | 100 kg (220 lbs) | 100 kg (220 lbs) | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| User Height | 1.60-1.90 m | 1.58-1.88 m | Different to thepredicate device.Performance and clinicaltesting demonstrate thatthere is no impact onsafety and effectiveness. |
| Walking Speed | 1.5 km/hr | ~ 2 km/hr | Different from thepredicate device. TheExoAtlet-II speed isslightly slower and doesnot pose an increasedsafety risk; the reducedspeed of the device wasevaluated throughclinical testing. |
| Grade ofInclination | 0-5º | 1.15º | Different from thepredicate device. TheExoAtlet-II can handle agreater grade ofinclination. Safetyconcerns regarding theincreased inclinationwas evaluated throughclinical testing. |
| Type of Surface | Smooth, cement, carpet | Smooth, cement, carpet | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| Range of Motion | Hips: 108º flexion to 21º extension Knees: 112º flexion to 2º extension Ankles: 15º flexion to 0-15º extension | Hips: 135º flexion to 20º extension Knees: 130º flexion to 0º extension Ankles: 10º flexion to 10º extension | Different from thepredicate device. TheExoAtlet-II has a largerrange of motion in thehips, knees, and ankles.Safety and effectivenesswere evaluated throughclinical testing. |
| Fall Detectionand Mitigation | None | None | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| Failsafe Feature | Yes, in the event of a failure-knees and pelvis become locked | Yes, in the event of a failure-knees become locked and hips free | Different from thepredicate device. TheExoAtlet-II locks theknees and pelvis. Safetyconcerns regarding thislocking were evaluatedthrough clinical testing.The therapist trainingprogram also educatesthe therapist on what todo under thiscircumstance |
| Operatingtemperature | 0°C to 35°C | -12°C to 40°C | Different from thepredicate device. TheExoAtlet-II can beoperated at atemperature that iswithin the predicate'soperating range. Anypossible effect on devicefunction was evaluatedthrough electrical safetyand EMC testing, as wellas performance testing. |
| Battery ChargeTime | 4 hours | 1 hour | Different from thepredicate device. TheExoAtlet-II can remaincharged for longer.Battery safety andeffectiveness wereevaluated throughelectrical safety andEMC testing, and clinicaltesting. |
| Expected useablelife | 5 years | 4 years | Different from thepredicate device. TheExoAtlet-II has a slightlylonger shelf life. Thedifference in shelf lifewas evaluated throughdurability testing. |
| Testing | |||
| PerformanceTesting | Software Electrical safety and electromagnetic compatibility Durability Clinical | Software Electrical safety and electromagnetic compatibility Durability Clinical | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| Electrical safetyTesting | IEC 60601-1 | IEC 60601-1 | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| EMCCompatibilitytesting | IEC 60601-1-2 | IEC 60601-1-2 | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| Clinical | Yes | Yes | Identical to thepredicate device. Noimpact on safety andeffectiveness. |
| Training Program | Yes | Yes | Identical to the predicate device. No impact on safety and effectiveness. |
| Certification Program | Yes | Yes | Identical to the predicate device. No impact on safety and effectiveness. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "EXOATLET" in a stylized font. The "EXO" portion of the word is in blue, while the "ATLET" portion is in black. The font is sans-serif and has a modern, minimalist look.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for EXOATLET. The logo is in a sans-serif font. The letters "EXO" are in blue, while the letters "ATLET" are in black.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "EXOATLET" in a stylized font. The "EXO" portion of the word is in a light blue color, while the "ATLET" portion is in black. The font is sans-serif and has a modern, minimalist look.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "EXOATLET" in a stylized font. The "EXO" portion of the word is in a bright blue color, while the "ATLET" portion is in a light gray color. The font is modern and sleek, with sharp angles and clean lines. The logo appears to be for a sports or fitness-related company.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for "EXO ATLET". The word "EXO" is in a blue color, while the word "ATLET" is in black. The font is sans-serif and modern. The logo is simple and clean.
SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the ExoAtlet-II.
Bench Testing: The elbow crutches provided in the ExoAtlet-II were testing in compliance to:
-
. ISO 11334-1:2007 Assistive products for walking manipulated by one arm -Requirements and test methods — Part 1: Elbow crutches
Software Verification: The software development and testing were executed in compliance to: -
. IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1: 2012 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
-
IEC 62304:2006 Medical device software Software life cycle processes ●
Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to:
- IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1: 2012 Medical electrical equipment -. Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 60601-1-6:11 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- CISPR 11: 2015 (Class A, Group 1) Industrial, scientific and medical equipment - Radiofrequency disturbance characteristics - Limits and methods of measurement
- . IEC 61000-3-2: 2014 Electromagnetic compatibility (EMC) – Part 3-2: Limits – Limits for harmonic current emissions (equipment input current ≤ 16 A per phase)
- . IEC 61000-3-3: 2013 Amendment 1 - Electromagnetic compatibility (EMC) - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public lowvoltage supply systems, for equipment with rated current ≤ 16 A per phase and not subject to conditional connection
- IEC 62368-1: 2014 (Second Edition) Audio/video, information and communication technology equipment - Part 1: Safety requirements
- IEC 62133: 2012 (Second Edition) Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary lithium cells, and for batteries made from them, for use in portable application – Part 2: Lithium systems
SUMMARY OF CLINICAL TESTING
ExoAtlet conducted an open label, non-randomized clinical study to assess the safety and effectiveness of the ExoAtlet-II in subjects, 18 years and older, with spinal cord injuries in the T4-L5 or C7-T4 range and type D neurological disorders according to the ASIA scale (n=20).
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the text "510(k) Summary" in a large, bold font. The text is black and appears to be centered on a white background. The text is likely a title or heading for a document or presentation.
Image /page/11/Picture/1 description: The image shows the word "EXOATLET" in a stylized font. The "EXO" portion of the word is in a bright blue color, while the "ATLET" portion is in a dark gray color. The font is sans-serif and appears to be custom-designed, with unique shapes for some of the letters.
19 of the 20 subjects successfully completed the 10 Meter Walk Test and the six-minute walk test (6 MWT). One subject was unable to complete the 6 MWT due to insufficient physical fitness. The surveys found that all of the subjects were able to use the subject device and on a scale from 1-10 (worst to best), 18 of the 20 subjects rated their comfort from eight out of ten to a ten out of ten. Two subjects rated their comfort a six out of ten. Additionally, none of the skin assessments found any adverse events related to biocompatibility such as abrasions, dry flaky skin, cut, eschar, or rash before, during, or after use of the ExoAtlet-II.
CONCLUSION
Based on the testing performed, including clinical testing, software validation, and electrical safety testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed ExoAtlet-II are assessed to be substantially equivalent to the predicate devices.
N/A