(266 days)
The DIOP8 surgical Doppler is indicated for use by qualified healthcare practitioners in a sterile condition in the operating theatre for the assessment of vascular blood flow by direct application to the vessel wall.
The DIOP8 is a sterile, single use, high sensitivity probe that assists in the performance of safe surgery. It is a lightweight, 8MHz ultrasound, handheld pencil probe' that has a small diameter tip that enables the detection of blood flow in vessels. The DIOP8 ultrasound probe operates in pulsed-wave mode. It has one circular piezo-electric crystal that is mounted behind the front faceplate and the crystal is used for both transmit and receive functions. The DIOP8 is manufactured at a frequency of 8MHz. The output ultrasonic parameters are fixed during the manufacturing process and cannot be adjusted by the user. A connector system at the end of the cable allows the DIOP8 to be removed from the hand held Doppler. An interconnection adapter houses the bi-directional electronics, which are specific for the device application, and is required as an interface between the main unit and DIOP8. The connector is similar to a conventional connector, but has no external conductive contacts, and is totally sealed, thus preventing the ingress of fluid and contaminants. Within the connector is a small coil that transfers electrical energy to and from a similar, but larger coil in the corresponding socket at the end of the PA8/PA8XS adapter. Transmitted and received electrical signals are passed through this connector, but there is no direct electrical connection between the probe and the PA8/PA8XS adapter. The connector provides electrical isolation between the probe and the PA8/PA8XS adapter, which optimises system electrical safety by limiting patient leakage currents to very low levels. During use, the adapter is securely held in the probe holder that is an integral part of the Doppler main unit. When connected this way, the hand held Doppler unit operates as if a normal VP8 probe is connected.
The requested information is not available in the provided text. The document is a 510(k) summary for a medical device (Disposable Intraoperative Probe - DIOP8) and details the device's characteristics, comparison to a predicate device, and various performance tests conducted. However, it does not include specific acceptance criteria with reported device performance values in a table, nor does it describe a study proving the device meets those criteria with a breakdown of sample sizes, ground truth establishment, expert involvement, or MRMC study results.
The document primarily focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and general performance testing (biocompatibility, mechanical, functional, electrical safety, EMC, environmental, and software).
Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, sample size details for test and training sets, data provenance, number of experts, adjudication methods, MRMC study results, or standalone performance results based on the provided text.
{0}------------------------------------------------
February 19, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Huntleigh Healthcare Ltd. % Mr. Steve Monks QRE Compliance Director 35 Portmanmoor Road Cardiff, CF24 5HN UNITED KINGDOM
Re: K201425
Trade/Device Name: Disposable Intraoperative Probe (DIOP8) Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: Class II Product Code: ITX Dated: December 14, 2020 Received: December 28, 2020
Dear Mr. Monks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201425
Device Name Disposable Intraoperative Probe (DIOP8)
Indications for Use (Describe)
The DIOP8 surgical Doppler is indicated for use by qualified healthcare practitioners in a sterile condition in the operating theatre for the assessment of vascular blood flow by direct application to the vessel wall.
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Huntleigh Healthcare Limited Diagnostic Products Division. The text "Huntleigh Healthcare Limited Diagnostic Products Division" is on the left side of the image. To the right of the text is the word "HUNTLEIGH" in large, blue, sans-serif font.
510(k) Summary K201425 Disposable Intraoperative Probe
| Name & Address: | Huntleigh HealthcareUnit 35, Portmanmoor RoadCardiffCF24 5HNUnited Kingdom |
|---|---|
| Telephone: | +44 (0)2920 485885 |
| Fax: | +44 (0)2920 492520 |
| Prepared: | May 2020 |
| Contact: | Steve Monks - QRE Compliance Director |
{4}------------------------------------------------
Huntleigh Healthcare Limited | | || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || || ||
5.1 Device Information
| Device Name: | Disposable Intraoperative Probe (DIOP8) | ||
|---|---|---|---|
| Regulation Name: | Diagnostic Ultrasonic Transducer | ||
| Classification Name: | Transducer, Ultrasonic, Diagnostic | ||
| Classification | Class | Product Code | Classification Regulation |
| II | ITX | 892.1570 |
5.2 Identification of Legally Marketed Devices to which Equivalence is claimed
| Predicate Device: | IOP8 Intraoperative Probe (K032315) manufactured byHuntleigh Healthcare Ltd., Diagnostic Products Division. | ||
|---|---|---|---|
| Regulation Name: | Diagnostic Ultrasonic Transducer | ||
| Classification Name: | Transducer, Ultrasonic, Diagnostic | ||
| Classification: | Class | Product Code | Classification Regulation |
| II | ITX | 892.1570 |
5.3 Indications for Use
| Indications for Use: | The DIOP8 surgical Doppler is indicated for use by qualified healthcare practitioners in a sterile condition in the operating theatre for the assessment of vascular blood flow by direct application to the vessel wall. |
|---|---|
| ---------------------- | --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
5.4 Description
Description : The DIOP8 is a sterile, single use, high sensitivity probe that assists in the performance of safe surgery. It is a lightweight, 8MHz ultrasound, handheld pencil probe' that has a small diameter tip that enables the detection of blood flow in vessels.
The DIOP8 ultrasound probe operates in pulsed-wave mode. It has one circular piezo-electric crystal that is mounted behind the front faceplate and the crystal is used for both transmit and receive functions. The DIOP8 is manufactured at a frequency of 8MHz. The output ultrasonic parameters are fixed during the manufacturing process and cannot be adjusted by the user.
A connector system at the end of the cable allows the DIOP8
{5}------------------------------------------------
Huntleigh Healthcare Limited =================================================================================================================================================
to be removed from the hand held Doppler. An interconnection adapter houses the bi-directional electronics, which are specific for the device application, and is required as an interface between the main unit and DIOP8. The connector is similar to a conventional connector, but has no external conductive contacts, and is totally sealed, thus preventing the ingress of fluid and contaminants. Within the connector is a small coil that transfers electrical energy to and from a similar, but larger coil in the corresponding socket at the end of the PA8/PA8XS adapter. Transmitted and received electrical signals are passed through this connector, but there is no direct electrical connection between the probe and the PA8/PA8XS adapter. The connector provides electrical isolation between the probe and the PA8/PA8XS adapter, which optimises system electrical safety by limiting patient leakage currents to very low levels. During use, the adapter is securely held in the probe holder that is an integral part of the Doppler main unit. When connected this way, the hand held Doppler unit operates as if a normal VP8 probe is connected.
The DIOP8 is compatible for use with the following Doppler units, all manufactured by Huntleigh Healthcare and have FDA clearance:
Models
- o DISP10XS
- DISP10 ●
- DIPP10 ●
- DISP10XS-DMX ●
- DISP10-D900
- DISP10-SD2 ●
- ISP ●
- ISPXS ●
The DISP10XS model contains 10xDIOP8 probes, a PA8XS adapter and a Pole Clamp (ACC47).
The DISP10 model contains 10xDIOP8 probes, a PA8 adapter and a Pole Clamp (ACC47).
The DIPP10 model contains 10xDIOP8 probes.
The DISP10XS-DMX contains the same as the DISP10XS model. with a DMXR Handheld Doppler added (cleared under K183574).
The DISP10-D900 contains the same as the DISP10 model, with a D900 Handheld Doppler added (cleared under
{6}------------------------------------------------
Huntleigh Healthcare Limited
Diagnostic Products Division
K930200).
The DISP10-SD2 contains the same as the DISP10 model, with a SD2 Handheld Doppler added (cleared under K930200).
The ISP is an Intraoperative Probe Starter Pack and includes a PA8 adapter and Pole Clamp (ACC47).
The ISPXS is an Intraoperative Probe Starter Pack and includes a PA8XS adapter and Pole Clamp (ACC47).
5.5 Comparison of Technological Characteristics with Predicate Device
Comparison of Technological Characteristics with Predicate Device:
Doppler ultrasound is the technological principle for both the subject and predicate device. It is based on a combination of a hand held Doppler unit, an intraoperative probe and a PA8/PA8XS adapter.
The subject and the predicate are based on the same technological elements:
| Technological Element | IOP8 | DIOP8 |
|---|---|---|
| Type of energy used | Ultrasound | Ultrasound |
| Principle used to detectblood flow | Doppler | Doppler |
| Number of piezo-electriccrystals present | 1 | 1 |
| Piezo-electric crystaldiameter | 3 mm | 3 mm |
| Operating mode | Pulsed | Pulsed |
| Pulse repetition frequency | 62.5 kHz | 62.5 kHz |
| Transmit burst duration | 8 µs | 8 µs |
| Receive window duration | 7 µs | 7 µs |
The following technological differences exist between the subject device and predicate device:
| Technological Element | IOP8 | DIOP8 |
|---|---|---|
| Piezo-electric crystal material | PZ27ceramic | NCE51ceramic |
| Probe length | 130 mm | 126 mm |
{7}------------------------------------------------
Diagnostic Products Division |- UNTLEIGH
| Probe diameter | 5 mm | 5 mm |
|---|---|---|
| Probe tube material | Stainlesssteel | ABSthermoplastic |
| Probe faceplate material | FortronPPS | ABSthermoplastic |
Piezo-electric crystal material – PZ27 is manufactured ● by Ferroperm, whereas NCE51 is manufactured by Noliac. The specification references are different, but the material properties are similar. The materials are equivalent in the intraoperative probe application.
- Probe length The DIOP8 is slightly shorter than the . IOP8 (by 4mm). Usability evaluations have verified that there is no impact on probe usability, therefore this difference is insignificant.
- . Probe tube material - The IOP8 tube material is stainless steel, whereas the DIOP8 tube material is ABS thermoplastic. The IOP8 can be re-sterilized using high temperature steam (autoclave method), hence stainless steel was chosen for compatibility. The DIOP8 however, is sterilised once only at the end of the manufacturing process using ethylene oxide (ETO), at a much lower temperature. This allowed the selection of ABS thermoplastic as the material for this component. Usability evaluations have verified that the probe is sufficiently rigid to perform its clinical function. Biocompatibility data is also provided for this material.
- Probe faceplate material The faceplate is the area of . the probe at the front, which is applied to the blood vessel during surgery. The IOP8 faceplate is made from Fortron PPS, which was selected due to the requirements for the probe to be compatible with high temperature steam (autoclave) sterilisation, as PPS is able to withstand high temperatures. The DIOP8 however, is sterilised once only at the end of the manufacturing process using ethylene oxide (ETO), at a much lower temperature. This allowed the selection of ABS thermoplastic as the material for this component. Biocompatibility data is also provided for this material.
5.6 Substantial Equivalence
Substantial The DIOP8 is substantially equivalent to cleared device IOP8 Equivalence: Intraoperative Probe. The DIOP8 has equivalent performance to the IOP8 Intraoperative Probe.
Review of differences:
{8}------------------------------------------------
Huntleigh Healthcare Limited |------------------------------------------------------------------------------------------------------------------------------------------------
5.7 Performance Data
Testing to demonstration substantial equivalence included:
| Testing Conducted | Discussion |
|---|---|
| Biocompatibility Testing-Cytotoxicity-Sensitization-Irritation-Acute SystemicToxicity | The biocompatibility evaluation ofthe device was conducted inaccordance with ISO 10993-1:2009 Biological evaluation ofmedical devices - Part 1:Evaluation and testing within arisk management process. Thesubject device is considered anexternal communicating device,contacting patient tissue (outersurface of blood vessels) for alimited duration of ≤24 hours. |
| MechanicalPerformance Testing | Testing provides verification ofmechanical performance includinggeneral, physical, environment,product labelling, instructions foruse, packaging, durability and lifetesting. |
| Functional PerformanceTesting | Testing included acoustic outputmeasurements completed by a thirdparty laboratory, NPL, inaccordance with both IEC 60601-2-37:2007+A1:2015 and Track 1 ofthe FDA guidance. Testingconfirmed the subject device did notexceed the recommended acousticoutput exposure levels.Additionally, clinical measurementaccuracy testing was completed in-house via a string phantom method. |
| Electrical Safety Testing | Electrical safety testing wasconducted on the product by athird party laboratory, UL. Testingconfirms device complies with IEC60601-1:2005 + A1:2012 |
| EMC Testing | EMC testing was conducted onthe product by a third partylaboratory, Kiwa ComplianceLaboratories. Testing confirms |
| device complies with IEC 60601-1-2:2014. | |
| EnvironmentalPerformance Testing | Testing provides verification of theoperating temperature, humidityand pressure and non-operatingtemperature, humidity, pressure,shock and vibration. |
| Software | The functions of the DIOP8 probeare supported by software, andthe software is previously clearedand installed on the ultrasoundunit that supports the probe. Thesoftware package is identical tothe predicate and no newsoftware code was created for thesubject device. |
{9}------------------------------------------------
5.8 Technologies Summary
Technologies The DIOP8 single use intraoperative Doppler ultrasound Summary: probe uses very similar technology to that of the predicate device; the IOP8 intraoperative Doppler probe. The main difference is that the DIOP8 is single-use, whereas the IOP8 is reusable.
As the DIOP8 is sterilised once only at the end of the manufacturing process using ethylene oxide (ETO), it is manufactured from materials compatible with this process. The IOP8 and DIOP8 operate at the same ultrasound frequency, in the same pulsed operating mode. In use, they provide same functionality; that is to evaluate blood flow within a vessel during surgery. They can be used with the same main Doppler units and PA8 adapters. Both the IOP8 and DIOP8 have the same intended use and indications for use. Where there are material differences, they mainly exist due to the different sterilization requirements, and biocompatibility data is available in this submission for these differences.
5.9 Conclusion
Conclusion:
The data detailed within this submission demonstrates that the DIOP8 is safe, effective and performs as well as the predicate device identified in this summary, which is currently marketed for the same intended use. The device's mechanical and functional performance verification demonstrates that it should perform as intended in the specified user conditions.
N/A