K Number
K201406

Validate with FDA (Live)

Device Name
Picowon
Manufacturer
Date Cleared
2020-07-22

(55 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Picowon is indicated for tattoo and benign pigmented lesions removal.

Device Description

The Picowon laser system is Alexandrite laser system. The Picowon laser system consists of an Alexandrite laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains Alexandrite laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed, in the special adjustable holders composed the laser cavity. And Picowon does not contain any external wired and/or wireless communication interfaces (Wired: USB, ethernet, SD, CD, RGA, etc. or Wireless: Wi-Fi, Bluetooth, RF, inductive, Cloud, etc.).

The laser energies generated by the medium of Alexandrite is converted to the heat energy once it gets to human skin surface and used for tattoos and benign pigmented lesions removal. The combination of wavelength, pulse duration and energy fluence are disrupting the tattoo dye or pigment particles under the skin without harming the surrounding tissue. The fragment particles eventually surface and fade as the epidermal layer or the skin is renewed. The system delivers laser energy at a wavelength of 755 nm. The output of the laser is delivered to the treatment area through an articulated Arm with a handpiece. A trigger (Foot Switch) controls the delivery of pulses. The treatment parament parameters and other functions operated by software on the graphical user interface.

AI/ML Overview

This 510(k) premarket notification describes a laser surgical instrument (Picowon) and compares it to a predicate device (Picosure™ workstation) to establish substantial equivalence.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly define acceptance criteria in terms of numerical performance metrics (e.g., sensitivity, specificity, accuracy) for clinical outcomes, as it's a laser surgical instrument rather than a diagnostic AI device. Instead, the "acceptance criteria" are implied by the comparison to the predicate device and compliance with established electrical, EMC, and laser safety standards. The "device performance" refers to the Picowon meeting the specifications similar to or safer than the predicate.

CharacteristicAcceptance Criteria (Implied by Predicate)Reported Proposed Device Performance (Picowon)
K NumberK121346K201406
ManufacturerCynosure, Inc.WON TECH Co., Ltd.
ModelPicosure™ workstationPicowon
Product CodeGEXGEX
Indications for UseTattoo and benign pigmented lesions removal.Tattoo and benign pigmented lesions removal.
Laser TypeFlash lamp-excited, Solid state Alexandrite laserFlash lamp-excited, Solid state Alexandrite laser
Wavelength755nm755nm
Maximum Average Fluence6.37 J/cm²6.37 J/cm²
Pulse Duration450-900 ps600-800 ps
Repetition Rate1-10 Hz1-5 Hz
Spot SizeZoom 2-6 mm, Fixed 2,3,4,6,8,10 mmZoom 2, 3, 3.5, 4, 5, 5.5, 6, 7, 8 mm
Safety and EffectivenessDemonstrated by predicate deviceDemonstrated through non-clinical testing for compliance with standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, software validation, biocompatibility, and in-house performance tests.

2. Sample size used for the test set and the data provenance

No clinical studies were conducted or deemed necessary for the Picowon device, as explicitly stated in the "Clinical Test Summary" section: "No clinical studies were considered necessary and performed." Therefore, there is no "test set" in the context of patient data for clinical performance evaluation. The testing involved non-clinical bench tests and in-house performance tests related to electrical, electromagnetic, and laser characteristics.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. As no clinical studies were performed, there was no clinical "ground truth" to be established by experts for a test set.

4. Adjudication method for the test set

Not applicable. No clinical test set was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a laser surgical instrument, not an AI diagnostic or assistance tool for human readers. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The Picowon is a physical medical device, a laser system, not an algorithm, and therefore does not have a standalone algorithm-only performance.

7. The type of ground truth used

For the technical performance aspects, the "ground truth" was established by engineering specifications, international safety standards (IEC, AAMI), and the performance characteristics of the legally marketed predicate device. For instance, laser output accuracy, pulse width, and repetition rate were verified against predefined limits derived from design specifications and predicate comparison.

8. The sample size for the training set

Not applicable. This device is a laser surgical instrument, not an AI/ML algorithm that requires a training set of data.

9. How the ground truth for the training set was established

Not applicable. There was no training set for an AI/ML algorithm.

{0}------------------------------------------------

July 22, 2020

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a symbol with three stylized human figures. The FDA logo is on the right and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

WON TECH Co., Ltd. Jake Yu Staff of Regulatory Affairs 64 Techno 8-Ro, Yuseong-gu Daejeon, 34028 Korea, Republic Of

Re: K201406

Trade/Device Name: Picowon Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 25, 2020 Received: May 28, 2020

Dear Jake Yu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201406

Device Name Picowon

Indications for Use (Describe)

The Picowon is indicated for tattoo and benign pigmented lesions removal.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-7836-6921 / Fax: +82-70-934-9491

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)] July 16, 2020

2. Submitter's Information [21 CFR 807.92(a)(1)]

  • Name of Sponsor: WON TECH Co., Ltd.
    • 64 Techno 8-Ro, Yuseong-gu, Daejeon, - Address: Republic of Korea, 34028
  • Contact Name: Jake Yu/ Staff of Regulatory Affair
    • Telephone No.: +82-70-7836-6921 -
    • -Fax No.: +82-70-934-9491
    • Email Address: regulatory(@wtlaser.com -
  • Name of Manufacturer: . Same as Sponsor
    • -Address: Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|

Trade NamePicowon
Common NamePowered Laser Surgical Instrument
Device Classification NameLaser surgical instrument for use in general and plastic surgery and indermatology (21 CFR 878.4810, Product Code GEX)
Regulation Number21 CFR 878.4810
Classification Product CodeGEX
Device ClassClass II
510k Review PanelGeneral & Plastic Surgery

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow, followed by the text "WON" in red and "TECH" in gray. The logo is simple and modern, and the colors are bright and eye-catching.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republi Tel: +82-70-7836-6921 / Fax: +82-70-934-9491

Identification of Predicate Device(s) [21 CFR 807.92(a)(3)| 4.

The identified predicate devices within this submission are shown as follow:

Primary Predicate device

510(k) Number:K121346
Applicant:CYNOSURE, INC.
Classification Name:Laser surgical instrument for use in generaland plastic surgery and in dermatology(21 CFR878.4810, Product Code GEX)
Trade Name:Picosure™ workstation

Description of the Device [21 CFR 807.92(a)(4)] న్.

The Picowon laser system is Alexandrite laser system. The Picowon laser system consists of an Alexandrite laser head, a power supply, a cooling system, a delivery system and other electrical components. The laser head contains Alexandrite laser medium, and two high-intensity xenon flash lamps enclosed together into the water cooling housing and two reflected mirrors fixed, in the special adjustable holders composed the laser cavity. And Picowon does not contain any external wired and/or wireless communication interfaces (Wired: USB, ethernet, SD, CD, RGA, etc. or Wireless: Wi-Fi, Bluetooth, RF, inductive, Cloud, etc.).

The laser energies generated by the medium of Alexandrite is converted to the heat energy once it gets to human skin surface and used for tattoos and benign pigmented lesions removal. The combination of wavelength, pulse duration and energy fluence are disrupting the tattoo dye or pigment particles under the skin without harming the surrounding tissue. The fragment particles eventually surface and fade as the epidermal layer or the skin is renewed. The system delivers laser energy at a wavelength of 755 nm. The output of the laser is delivered to the treatment area through an articulated Arm with a handpiece. A trigger (Foot Switch) controls the delivery of pulses. The treatment parament parameters and other functions operated by software on the graphical user interface.

Indications for Use [21 CFR 807.92(a)(5)] 6.

The Picowon is indicated for tattoo and benign pigmented lesions removal.

Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)] 7.

There are no significant differences in the technological characteristics of this device compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the Picowon and the predicate device:

Proposed DevicePredicate DeviceSE decision
K NumberK201406K121346-

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow, followed by the text "WON" in red and "TECH" in gray. The logo is simple and modern, and the colors are bright and eye-catching.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic o Tel: +82-70-7836-6921 / Fax: +82-70-934-9491

Proposed DevicePredicate DeviceSE decision
ManufacturerWON TECH Co., Ltd.Cynosure, Inc.-
ModelPicowonPicosure™ workstation-
Product CodeGEXGEXSame
Indications for UseThe Picowon is indicated fortattoo and benign pigmentedlesions removal.The Picosure™ workstation isindicated for tattoo and benignpigmented lesions removal.Same
Laser TypeFlash lamp-excited, Solidstate Alexandrite laserFlash lamp-excited, Solid stateAlexandrite laserSame
Wavelength755nm755nmSame
Maximum AverageFluence6.37 J/cm²6.37 J/cm²Same
Pulse Duration600-800 ps450-900 psSimilar, in the rangeof pulse duration ofpredicate device,more specific
Repetition Rate1-5 Hz1-10 HzSimilar, less thanpredicate device,safer
Spot SizeZoom 2, 3, 3.5, 4, 5, 5.5, 6, 7,8 mmZoom 2-6 mm,Fixed 2,3,4,6,8,10 mmSame

The pulse duration of Picocare is in the range of the pulse duration of predicate device. So the tissue reaction has no difference between them. And the repetition rate is related with only treatment time, but not effectiveness. Therefore the key differences between Picowon and the predicate device are Pulse duration and Repetition rate which do not raise any new safety and effectiveness issues. The Picowon and predicate device have the same wavelength, output power and indications for use.

Verification and validation activities were conducted to establish the performance and safety characteristics of the Picowon. The results of these activities show that there are no any new safety and effectiveness issues. Therefore, the Picowon is considered substantially equivalent to the predicate device.

Non-Clinical Test Summary [21 CFR 807.92(b)(1)]

  1. Electrical Safety, Electromagnetic Compatibility and Performance.

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a white "W" inside of it. To the right of the circle, the text "WONTECH" is written in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-7836-6921 / Fax: +82-70-934-9491

StandardsNo.StandardsOrganizationStandard TitleVersionPublicationYear
ES60601-1AAMIMedical Electrical Equipment - Part 1: GeneralRequirements for Basic Safety and Essential PerformanceEdition 3.12012
60601-1-2IECMedical Electrical Equipment - Part 1-2: GeneralRequirements for Safety - Collateral Standard:Electromagnetic Compatibility - Requirements and TestsEdition 42014
60601-2-22IECMedical Electrical Equipment - Part 2-22: ParticularRequirements for Basic Safety and Essential Performanceof Surgical, Cosmetic, Therapeutic and Diagnostic LaserEquipmentEdition 32012
60825-1IECSafety of Laser Products - Part 1: EquipmentClassification, and RequirementsEdition 32014

2) Software Validation

The Picowon contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.

3) Biocompatibility

As an accessory, Picowon handpiece tip was designed not contacting to patient skin for the laser treatment effectiveness and aiming beam indicates a treatment laser radiation zone.

4) In-House performance test

In-House performance test for Picowon was conducted for verifying the accuracy of main performance specifications such as irradiation range (spot size), energy density, laser output accuracy, guide beam output limitation, laser output stability, operation check of the safety devices such as beam shutters, deviation of focal position of treatment guide light, variable range of pulse width and accuracy of repetitive frequency.

Clinical Test Summary [21 CFR 807.92(b)(2)]

No clinical studies were considered necessary and performed.

8. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd., concludes that the Picowon is substantially equivalent in safety and effectiveness to the predicate device as described herein.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.