K Number
K201354

Validate with FDA (Live)

Device Name
TENS & PMS
Date Cleared
2020-12-11

(204 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TENS(1,3,4,5,6,21): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9126
TENS(3,4,5,6,11,12,16): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7,8,9,10,13,14,15): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9186
TENS(3,4,7,8,11): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,5,6,9,10,12): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model: SM9196
TENS(1015): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1
9): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Device Description

TENS & PMS is a portable and DC 3.7V battery powered multifunction device with multiple models, offering both Transcutaneous Electrical Nerve Stimulation (TENS) and Powered Muscle Stimulation (PMS) qualities in one device. The device is equipped with accessories of electrode pads, electrode cables, a battery charger, and one USB cable. The electrode cables are used to connect the pads to the device; the USB cable is used to connect the charger and the built-in lithium battery. All accessories, including USB cables, electrode pads, electrode cables, chargers can only be changed or replaced by a qualified person.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a TENS & PMS device. This submission aims to demonstrate substantial equivalence to an already legally marketed predicate device (K121719). The document focuses on comparing the technical characteristics and performance of the new device models (SM9079, SM9126, SM9186, SM9196) against the predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are implicitly defined by the safety and performance parameters of the legally marketed predicate device (K121719) and compliance with various recognized standards. The reported device performance is demonstrated through the comparative tables (Table 1, 2, 3, 4) where the new device's parameters are listed alongside the predicate device's, and through the statement of compliance with specific IEC and ANSI/AAMI standards.

Since the tables in the document are already in a comparative format, I will summarize the key parameters and their reported performance for one of the new models (SM9079) against the predicate. The "Remark" column in the original tables already serves as a concise statement of whether the new device meets or is similar to the predicate.

Table: Acceptance Criteria and Reported Device Performance (Example: Model SM9079 vs. Predicate K121719)

Parameter / Acceptance Criteria (Derived from Predicate)Reported Device Performance (New Device SM9079)Meets Criteria?Notes
General Characteristics
ManufacturerHong Qiangxing (Shenzhen) Electronics LimitedYesSame
Intended Use (TENS)Temporary relief of pain for specific body parts due to strainYesSame (Expanded list of programs, but comparable)
Intended Use (PMS)Stimulate healthy muscles to improve performanceYesSame (Expanded list of programs, but comparable)
Type of UseOTCYesSame
Power SourceDC 3.7V lithium batteryYesSame
Method of Line Current IsolationType BFYesSame
Patient Leakage Current (Normal Condition)< 10μAYesSame
Patient Leakage Current (Single Fault Condition)< 50μAYesSame
Average DC current through electrodes< 0.01μAYesSame
Number of Output Modes24SimilarMore treatment programs, all comparable to predicate's mode 3. No new safety/effectiveness issues.
Number of Output Channels2YesSame
Synchronous or Alternating Channel OutputAlternatingSimilarDifferent from predicate's synchronous, but same technology and treatment effect. No new safety/effectiveness issues.
Method of Channel IsolationVoltage transformer IsolationYesSame
Regulated Current or VoltageVoltage controlYesSame
Software/Firmware/Microprocessor ControlSoftwareYesSame
Automatic Overload Trip?NoYesSame
Automatic No-Load Trip?NoYesSame
Automatic Shut Off?YesYesSame
User Override Control?YesYesSame
Indicator (On/Off Status)YesYesSame
Indicator (Low Battery)YesYesSame
Indicator (Voltage/Current Level)YesYesSame
Timer Range10 ~ 60 minutes, 10 min/stepYesSame
BiocompatibilityAll user directly contacting materials comply with ISO10993-5 and ISO10993-10 requirements.YesSame
AccessoriesSelf-adhesive electrodes, electrode wires, Battery charger, USB cableYesSame
Physical Characteristics
Weight102g±5gSimilarDifference in weight and dimensions noted, but deemed not to raise safety/effectiveness issues.
Dimensions6012312mmSimilarDifference in weight and dimensions noted, but deemed not to raise safety/effectiveness issues.
Housing Materials & ConstructionABSYesSame
Output Specifications (Max Values from Table 5 for SM9079 vs. Predicate Mode 3/1/2)
WaveformPulsed, symmetric, biphasicYesSame
ShapeRectangularYesSame
Max Output Voltage (@500Ω)42V±20%SimilarPredicate 42V±10%. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
Max Output Current (@500Ω)84mA±20%SimilarPredicate 84mA±10%. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
Pulse Width190µs (max)SimilarPredicate 100µs. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
Pulse Period40-1070ms (range)SimilarPredicate 9.3~850ms. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
Max Pulse Frequency0.935-25Hz±10% (range)SimilarPredicate 110Hz. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
Net Charge (per pulse)0µC @500Ω; Balanced waveformYesSame
Max Phase Charge (@500Ω)13.30µC@500ΩSimilarPredicate 16.8µC@500Ω. Differences in these parameters allowed as they don't exceed safety limits and passed IEC 60601-2-10 tests.
Max Current Density (@500Ω)0.065mA/cm²YesMeets max current density <2mA/cm² (predicate value not given explicitly here, but stated in Note 4 for SM9126).
Max Average Power Density (@500Ω)2.64mW/cm²YesMeets max average power density <0.25W/cm² (predicate value not given explicitly here, but stated in Note 4 for SM9126).
Compliance with Standards
ANSI AAMI ES60601-1YesYesSame
IEC 60601-2-10YesYesSame
ANSI AAMI IEC 60601-1-2YesYesSame
IEC 62133YesYesSame
ANSI AAMI HA60601-1-11YesYesSame
21 CFR 898 (if applicable)YesYesSame

2. Sample size used for the test set and the data provenance

The document explicitly states that the new devices have "passed testings according to the following standards" and that the "conclusion drawn from the testings are that the new devices are substantially equivalent to the predicate device." These "testings" refer to compliance with various medical device standards (ANSI AAMI ES60601-1, IEC 60601-2-10, ANSI AAMI IEC 60601-1-2, IEC 62133, ANSI AAMI HA60601-1-11).

However, the document does not provide details about a clinical study or a specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective). The submission is based on a comparison of technical specifications and performance against a predicate device and adherence to recognized standards, rather than a separate clinical trial with a defined sample size. The testing appears to be functional and electrical safety testing as per the listed standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Given that a clinical study or human-in-the-loop performance evaluation with a test set requiring expert ground truth is not described in the provided text, information on the number and qualifications of experts for establishing ground truth is not available. The evaluation is based on technical specifications and compliance with standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Since no clinical test set requiring human adjudication is mentioned, there is no information provided on any adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

The device described is a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS). It is a physical therapy device and not an AI-assisted diagnostic or imaging interpretation device. Therefore, a Multi Reader Multi Case (MRMC) comparative effectiveness study focusing on human readers improving with AI assistance would be irrelevant and is not mentioned in this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Again, this is a physical therapy device, not an algorithm, so no standalone algorithm performance evaluation is applicable or mentioned.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this type of device, the "ground truth" for demonstrating substantial equivalence and safety/effectiveness relies on:

  • Compliance with recognized electrical safety and performance standards: This includes specific limits for voltage, current, pulse characteristics, leakage current, and electromagnetic compatibility.
  • Comparison of technical specifications with a legally marketed predicate device: The predicate device itself has established safety and effectiveness. The "ground truth" is that the predicate device is safe and effective when used as intended, and the new device demonstrates similar performance and characteristics to be considered equivalent.
  • Biocompatibility testing: Compliance with ISO10993-5 and ISO10993-10 for materials directly contacting the user.

There is no mention of expert consensus, pathology, or outcomes data from a clinical trial for this specific submission. The "ground truth" is established by adherence to existing regulatory frameworks and engineering standards.

8. The sample size for the training set

As this is not an AI/machine learning device, the concept of a "training set" is not applicable and therefore, no sample size for a training set is provided.

9. How the ground truth for the training set was established

As the concept of a training set is not applicable, no information is provided on how its ground truth would be established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2020

Hong Qiangxing (Shen Zhen) Electronics Limited % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No. 19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China

Re: K201354

Trade/Device Name: TENS & PMS Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX. NUH Dated: September 3, 2020 Received: September 14, 2020

Dear Doris Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201354

Indications for Use (Describe)

Device Name TENS & PMS

Model:SM9079
TENS(1,3,4,5,6,21):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model:SM9126
TENS(3,4,5,6,11,12,16):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7,8,9,10,13,14,15):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model:SM9186
TENS(3,4,7,8,11):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,5,6,9,10,12):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Model:SM9196
TENS(10~15):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1~9):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information:

510(k) Number:K201354
Date:December 11, 2020
Type of 510(k) Submission:Traditional
Basis for 510(k) Submission:New device
Submitter/Manufacturer:Hong Qiangxing(Shen Zhen) Electronics Limited4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'anDistrict, Shenzhen City, Guangdong, China 518126
Contact:Doris Dong[Consultant, from Shanghai CV Technology Co., Ltd.]Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 ChinaE-mail: doris_d@126.comTel: 86 21-31261348 / Fax: 86 21-57712250
2. Device Description:
Proprietary Name:TENS & PMS
Common Name:TENS & PMS
Classification Name:Powered muscle stimulatorTranscutaneous electrical nerve stimulator for pain relief
Regulation Number:21 CFR 890.5850, 21 CFR 882.5890
Product Code:NGX, NUH
Device Class:II
Review Panel:Neurology & Physical Medicine
Device Description:TENS & PMS is a portable and DC 3.7V battery powered multifunctiondevice with multiple models, offering both Transcutaneous Electrical NerveStimulation (TENS) and Powered Muscle Stimulation (PMS) qualities inone device.The device is equipped with accessories of electrode pads, electrode cables,a battery charger, and one USB cable. The electrode cables are used toconnect the pads to the device; the USB cable is used to connect the chargerand the built-in lithium battery. All accessories, including USB cables,electrode pads, electrode cables, chargers can only be changed or replacedby a qualified person.
Indications for use:Model: SM9079TENS(1,3,4,5,6,21):To be used for temporary relief of pain associated with sore and achingmuscles in the shoulder, waist, back, neck, upper extremities (arm), andlower extremities (leg) due to strain from exercise or normal householdwork activities.PMS(2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24):It is intended to be used to stimulate healthy muscles in order to improveand facilitate muscle performance.

{5}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

Model: SM9126

TENS(3.4.5.6.11.12.16):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.

PMS(1,2,7,8,9,10,13,14,15):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Model: SM9186

TENS(3,4,7,8,11):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.

PMS(1,2,5,6,9,10,12):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Model: SM9196

TENS(10~15):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.

PMS(1~9):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

{6}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 3. Working principle

The device uses electrodes to send a specific electric current to the skin. TENS(Transcutaneous electrical nerve stimulation) mode is to relieve pain using electrical stimulation. PMS(Powered Muscle Stimulation) mode is to stimulate healthy muscles in order to improve performance by electrical stimulation.

4. Substantial Equivalence to Predicate device: Basic Unit Characteristics

Table 1-

ParametersNew DevicePredicate DeviceRemark
510(k) NumberK201354K121719--
Device NameTENS & PMSSM TENS & PMS--
ModelSM9079No--
ManufacturerHong Qiangxing (Shenzhen)Electronics LimitedHong Qiangxing (Shenzhen)Electronics LimitedSame
Intended useTENS(1,36,21):To be used for temporary reliefof pain associated with sore andaching muscles in the shoulder,waist, back, neck, upperextremities (arm), and lowerextremities (leg) due to strainfrom exercise or normalhousehold work activities.PMS(2,720,22~24):It is intended to be used tostimulate healthy muscles inorder to improve and facilitatemuscle performance.TENS(1,3,4,5,6):To be used for temporary reliefof pain associated with sore andaching muscles in the shoulder,waist, back, neck, upperextremities (arm), and lowerextremities (leg) due to strainfrom exercise or normalhousehold work activities.PMS(1,2,3,6):It is intended to be used tostimulate healthy musclesinorder to improve and facilitatemuscle performance.Same
Type of useOTCOTCSame
Power Source(s)DC 3.7V lithium batteryDC3.7Vlithium batterySame
- Method of LineCurrent IsolationType BFType BFSame
Patient Leakage--Same
Current
Normal< 10μΑ< 10μΑ
Condition (μΑ)
- Single Fault< 50μA< 50μA
Condition (μΑ)
AverageDC< 0.01μA< 0.01μASame
currentthrough
electrodeswhen
device is on but no
pulses are being
applied (µA)
Number of Output246Similar
ModesNote 1
Number of Output22Same
channels:
AlternatingSynchronousSimilar Note 2
SynchronousOr
Alternating?
- Method ofVoltage transformerVoltage transformerSame
Channel IsolationIsolationIsolation
Regulated CurrentVoltage controlVoltage controlSame
orRegulated
Voltage?
Software/Firmwar
SoftwareSoftwareSame
e/MicroprocessorControl?
AutomaticNoNoSame
Overload Trip?
AutomaticNoNoSame
No-Load Trip?
AutomaticShutYesYesSame
Off?
UserOverrideYesYesSame
Control?
IndicatoOn/OffYesYesSame
rStatus
DisplayLowYesYesSame
Battery
VoltageYesYesSame
/Curren
t Level
TimerRange10 ~ 60 minutes, 1010 ~ 60 minutes, 10Same
(minutes)min/stepmin./step
CompliancewithYes.Yes.Same
VoluntaryAAMI/ANSI ES 60601-1AAMI/ANSI
Hong Qiangxing(Shen Zhen) Electronics Limited4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
Standards?IEC 60601-1-2, IEC60601-2-10, IEC 62133,HA 60601-1-11ES60601-1, IEC 60601-1-2, IEC60601-2-10, IEC62133, IEC 60601-1-11Same
Compliance with21 CFR 898?YesYesSame
Weight (grams)102g±5g140gSimilar Note 1
Dimensions (mm)[W x H x D]6012312mm83429mm
Housing Materials& ConstructionABSABSSame
WaveformPulsed, symmetric,biphasicPulsed, symmetric,biphasicSame
ShapeRectangleRectangleSame
Maximum OutputVoltage (volts)42V±20% @500Ω42V±10% @500ΩSimilar Note 3
Maximum OutputCurrent (specifyunits)84mA±20% @500Ω84mA±10% @500Ω
Pulse width (µsec)190µs100µs
Pulse Period(msec)40-1070ms9.3~850ms
Max. pulsefrequency (Hz) [orRate (pps)]0.935-25Hz±10%110Hz
Net Charge (µCper pulse)0µC @500Ω; Method:Balanced waveform0µC @500Ω; Method: BalancedwaveformSame
Maximum PhaseCharge, (µC)13.30µC@500Ω16.8µC @500ΩSimilar
MaximumAverage Current,(mA)0.78mA0.924mASimilar Note 4
Maximum CurrentDensity, (mA/cm²,r.m.s.)0.065mA/cm² @500Ω0.462mA/cm²
MaximumAverage PowerDensity,(mW/cm²)2.64mW/cm² @500Ω9.702mW/cm²
AccessoriesSelf-adhesive electrodes,electrode wires, Battery charger,USB cableSelf-adhesive electrodes,electrode wires,Battery charger,USB cableSame
BiocompatibilityAll user directlyAll user directlySame

{7}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

Hong Qiangxing(Shen Zhen) Electronics Limited
4F, Jingcheng Building, Xicheng Industrial Zon, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

{8}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

{9}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

contacting materials arecompliance withISO10993-5 and ISO10993-10requirements.contacting materials arecompliance withISO10993-5 and ISO10993-10requirements.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Comparison in details:

Note 1:

The proposed device SM9079 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 3 of the predicate device. The proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 6 are variation waveform of the mode 1 to mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9079 are a little different from predicate device K121719, but these differences will not raise any safety or effectiveness issue.

Note 2:

The output channels of the proposed device is alternating while the predicate device is synchronous. Because the proposed device and predicate device adopt the same fundamental output technology and similar treatment effect. Therefore, this item is considered to be substantially equivalent. Also, the proposed device had passed AAMI / ANSI ES60601-1 and IEC 60601-2-10 test codes, so these differences won't raise any new safety and effectiveness issues.

Note 3:

There are some differences on the pulse width, frequency, pulse period and deviation between proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. Therefore, these differences won't raise any new safety and effectiveness issues.

Note 4:

The maximum average current of proposed device is smaller than that of the predicate device which means the better safety. The maximum current density, maximum average power density have some differences between proposed device and predicate device due to they are calculated by different electrode area. Both of them meet maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectivenessissues.

Final conclusion:

The subject device SM9079 is substantial Equivalent to the predicate device K121719.

Table 2-

ParametersNew DevicePredicate DeviceRemark
510(k) NumberK201354K121719--

{10}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
Device NameTENS & PMSSM TENS & PMS--
ModelSM9126No--
ManufacturerHong Qiangxing (Shenzhen)Electronics LimitedHong Qiangxing (Shenzhen)Electronics LimitedSame
Intended useTENS(36,11,12,16):To be used for temporary reliefof pain associated with sore andaching muscles in the shoulder,waist, back, neck, upperextremities (arm), and lowerextremities (leg) due to strainfrom exercise or normalhousehold work activities.PMS(1,2,710,13~15):It is intended to be used tostimulate healthy muscles inorder to improve and facilitatemuscle performance.TENS(1,3,4,5,6):To be used for temporary relief ofpain associated with sore and achingmuscles in the shoulder, waist, back,neck, upper extremities (arm), andlower extremities (leg) due to strainfrom exercise or normal householdwork activities.PMS(1,2,3,6):It is intended to be used to stimulatehealthy muscles in order to improveand facilitate muscle performance.Same
Type of useOTCOTCSame
Power Source(s)DC 3.7V lithium batteryDC 3.7V lithium batterySame
- Method of LineCurrent IsolationType BFType BFSame
- Patient LeakageCurrent----Same
-- NormalCondition (μΑ)< 10μΑ< 10μΑ
- Single FaultCondition (μΑ)< 50μA< 50μA
Average DC currentthrough electrodeswhen device is onbut no pulses arebeing applied (µA)< 0.01μA< 0.01μASame
Number of OutputModes166SimilarNote 1
Number of Outputchannels:22Same
- Synchronousor Alternating?SynchronousSynchronousSame
- Method ofChannel IsolationVoltage transformer IsolationVoltage transformer IsolationSame
Regulated Currentor RegulatedVoltage?Voltage controlVoltage controlSame
Software/Firmware/SoftwareSoftwareSame

{11}------------------------------------------------

Hồng Qiangxing(Shen Zhen) Electronics Limited4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
MicroprocessorControl?
Automatic OverloadTrip?NoNoSame
Automatic No-LoadTrip?NoNoSame
Automatic Shut Off?YesYesSame
UserOverrideControl?YesYesSame
IndicatorOn/OffStatus?YesYesSame
DisplayLowBattery?YesYesSame
Voltage/CurrentLevel?YesYesSame
Timer(minutes)Range10 ~ 60 minutes, 10 min/step10 ~ 60 minutes, 10 min./stepSame
CompliancewithVoluntaryStandards?Yes.AAMI/ANSI ES 60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, HA 60601-1-11Yes.AAMI/ANSI ES60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, IEC 60601-1-11Same
Compliance with 21CFR 8988?YesYesSame
Weight (grams)55g±5g140gSimilar
Dimensions (mm)[W x H x D]50.3793.410.4mm83429mmNote 1
HousingMaterials& ConstructionABSABSSame
WaveformPulsed, symmetric, biphasicPulsed, symmetric, biphasicSame
ShapeRectangularRectangularSame
Maximum OutputVoltage (volts)46.4V±20% @500Ω42V±10% @500ΩSimilarNote 2
MaximumOutputCurrent (specifyunits)92.8mA±20% @500Ω84mA±10% @500ΩSame
Pulse width (usec)220μs100μsSame
Pulse Period (msec)5-680ms9.3~850msSame
Max. pulsefrequency (Hz) [orRate (pps)]1.47-212.5Hz±10%110HzSame
Net Charge (μC perpulse)0μC @500Ω; Method: Balancedwaveform0μC @500Ω; Method: BalancedwaveformSame
MaximumPhaseCharge, (μC)17.26μC@500Ω16.8μC @500ΩSimilar

{12}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
Maximum AverageCurrent, (mA)2.056mA0.924mASimilarNote 3
Maximum CurrentDensity, (mA/cm²,r.m.s.)0.119mA/cm² @500Ω0.462mA/cm²
Maximum AveragePower Density,(mW/cm²)5.21mW/cm² @500Ω9.702mW/cm²
BiocompatibilityAll user directly contactingmaterials are compliance withISO10993-5 and ISO10993-10requirements.All user directly contacting materialsare compliance with ISO10993-5 andISO10993-10 requirements.Same
AccessoriesSelf-adhesive electrodes,electrode wires, Battery charger,USB cableSelf-adhesive electrodes, electrodewires, Battery charger, USB cableSame

Comparison in details:

Note 1:

The proposed device SM9126 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 1 to mode 3 of the proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 6 are variation waveform of the mode 1 to mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9126 are a little different from predicate device K121719, but these differences are insignificant in the terms of safety or effectiveness.

Note 2:

There are some differences on the maximum output voltage, maximum output current, pulse width, frequency, pulse period, deviation between proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences won't raise any new safety and effectiveness issues.

Note 3:

The maximum current density, maximum average power density have some differences between proposed device and predicate device due to they are calculated by different electrode area. Both of them met maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectiveness issues.

Final conclusion:

The subject device SM9126 is substantial Equivalent to the predicate device K121719.

{13}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 Table 3-

ParametersNew DevicePredicate DeviceRemark
510(k) NumberK201354K121719--
Device NameTENS & PMSSM TENS & PMS--
ModelSM9186No--
ManufacturerHong Qiangxing(Shenzhen) ElectronicsLimitedHong Qiangxing (Shenzhen)Electronics Limited--
Intended useTENS(3,4,7,8,11):To be used for temporaryrelief of pain associatedwith sore and achingmuscles in the shoulder,waist, back, neck, upperextremities (arm), andlower extremities (leg) dueto strain from exercise ornormal household workactivities.PMS(1,2,5,6,9,10,12):It is intended to be used tostimulate healthy musclesin order to improve andfacilitatemuscleperformance.TENS(1,3,4,5,6):To be used for temporaryrelief of pain associated with sore andaching muscles in the shoulder,waist, back, neck, upperextremities (arm), and lowerextremities (leg) due to strainfrom exercise or normalhousehold work activities.PMS(1,2,3,6):It is intended to be used tostimulate healthy muscles inorder to improve and facilitatemuscle performance.Same
Type of useOTCOTCSame
Power Source(s)DC 3.7V lithium batteryDC 3.7V lithium batterySame
- Method of Line Current IsolationType BFType BFSame
- Patient Leakage Current- Normal Condition (μΑ)< 10μA< 10μASame
- Single Fault Condition (μΑ)< 50μA< 50μA
Average DC current throughelectrodes when device is on but nopulses are being applied (µA)< 0.01μA< 0.01μASame
Number of Output Modes126SimilarNote 1
Number of Output channels:22Same
- Synchronous or Alternating?AlternatingSynchronousSimilarNote 2
- Method of Channel IsolationVoltage transformerIsolationVoltage transformer IsolationSame
RegulatedVoltage?Current or RegulatedVoltage controlVoltage controlSame
Software/Firmware/MicroprocessorControl?SoftwareSoftwareSame
Automatic Overload Trip?NoNoSame
Automatic No-Load Trip?NoNoSame
Automatic Shut Off?YesYesSame
User Override Control?YesYesSame
IndicatorOn/Off StatusYesYesSame
DisplayLow BatteryYesYesSame
Voltage/CurrentLevelYesYesSame
Timer Range (minutes)10 ~ 60 minutes, 10min/step10 ~ 60 minutes, 10 min./stepSame
ComplianceStandards?withVoluntaryYes.AAMI/ANSI ES 60601-1Yes.AAMI/ANSI ES60601-1 IECSame
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
IEC 60601-1-2, IEC 60601-2-10, IEC 62133, HA 60601-1-11IEC 60601-1-2, IEC 60601-2-10, IEC 62133, IEC 60601-1-11
Compliance with 21 CFR 8988?YesYesSame
Weight (grams)$50\pm5$ g140gSimilar
Dimensions (mm) [W x H x D]52.99211mm83429mmNote 1
Housing Materials & ConstructionABSABSSame
WaveformPulsed, symmetric, biphasicPulsed, symmetric, biphasicSame
ShapeRectangularRectangularSame
Maximum Output Voltage (volts)$68V\pm20% @500\Omega$$42V\pm10% @500\Omega$Similar
Maximum Output Current (specify units)$136mA\pm20% @500\Omega$$84mA\pm10% @500\Omega$Note 3
Pulse width (μsec)220-230μs100μs
Pulse Period (msec)6.4-901ms9.3~850ms
Max. pulse frequency (Hz) [or Rate (pps)]1.11-155.6Hz110Hz
Net Charge (μC per pulse)0μC @500Ω; Method: Balanced waveform0μC @500Ω; Method: Balanced waveformSame
Maximum Phase Charge, (μC)26μC@500Ω16.8μC @500ΩSimilarNote 4
Maximum Average Current, (mA)5.54mA@500Ω0.924mA
Maximum Current Density, (mA/cm², r.m.s.)0.46mA/cm² @500Ω0.462mA/cm²Same
Maximum Average Power Density, (mW/cm²)31.40mW/cm² @500Ω9.702mW/cm²SimilarNote 4
BiocompatibilityAll user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements.All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements.Same
AccessoriesSelf-adhesive electrodes, electrode wires, Battery charger, USB cableSelf-adhesive electrodes, electrode wires, Battery charger, USB cableSame
Table 4-
ParametersNew DevicePredicate DeviceRemark
510(k) NumberK201354K121719--
Device NameTENS & PMSSM TENS & PMS--
ModelSM9196No--
ManufacturerHong Qiangxing (Shenzhen)Electronics LimitedHong Qiangxing (Shenzhen)Electronics LimitedSame
Intended useTENS(1015):To be used for temporary relief ofpain associated with sore and achingmuscles in the shoulder, waist, back,neck, upper extremities (arm), andlower extremities (leg) due to strainfrom exercise or normal householdwork activities.PMS(19):It is intended to be used to stimulatehealthy muscles in order to improveand facilitate muscle performance.TENS(1,3,4,5,6):To be used for temporary relief ofpain associated with sore and achingmuscles in the shoulder, waist, back,neck, upper extremities (arm), andlower extremities (leg) due to strainfrom exercise or normal householdwork activities.PMS(1,2,3,6):It is intended to be used to stimulatehealthy muscles in order to improveand facilitate muscle performance.Same
Type of useOTCOTCSame
Power Source(s)DC 3.7V lithium batteryDC 3.7V lithium batterySame
- Method of LineCurrent IsolationType BFType BFSame
- Patient LeakageCurrent--Normal Condition (μΑ)< 10μΑSingle Fault Condition (μΑ)< 50μΑ--< 10μΑ< 50μΑSame
Average DC current through electrodeswhen device is on but no pulses arebeing applied (μΑ)< 0.01μA< 0.01μASame
Number of OutputModes156SimilarNote 1
Number of Outputchannels:- Synchronousor Alternating?4Synchronous & Alternating2SynchronousSimilarNote 2
- Method ofChannel IsolationVoltage transformer IsolationVoltage transformer Isolation
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
Regulated Current orRegulated Voltage?Voltage controlVoltage controlSame
Software/Firmware/MicroprocessorControl?SoftwareSoftwareSame
Automatic OverloadTrip?NoNoSame
Automatic No-LoadTrip?NoNoSame
Automatic Shut Off?YesYesSame
UserOverrideControl?YesYesSame
IndicatorOn/OffStatus?YesYesSame
DisplayLowBattery?YesYesSame
Voltage/CurrentLevel?YesYesSame
Timer(minutes)Range60 minutes10 ~ 60 minutes, 10 min./stepSimilarNote 2
CompliancewithVoluntaryStandards?Yes.AAMI/ANSI ES60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, HA 60601-1-11Yes.AAMI/ANSI ES60601-1, IEC60601-1-2, IEC 60601-2-10, IEC62133, IEC 60601-1-11Same
Compliance with 21CFR 8988?YesYesSame
Weight (grams)$100g\pm5g$140gSimilarNote 1
Dimensions(mm)[W x H x D]58mm180mm15.5mm83429mmSimilarNote 1
Housing Materials &ConstructionABSABSSame
WaveformPulsed, symmetric, biphasicPulsed, symmetric, biphasicSame
ShapeRectangleRectangleSame
MaximumVoltage (volts)Output$54V\pm20% @500\Omega$$42V\pm10% @500\Omega$SimilarNote 3
MaximumCurrent(specifyunits)Output$108mA\pm20% @500\Omega$$84mA\pm10% @500\Omega$
Pulse width (μsec)200-210μs100μs
Pulse Period (msec)20.4-620ms9.3~850ms
Max.frequency (Hz) [orRate (pps)]pulse1.613-49.02Hz110Hz
Net Charge (µC perpulse)0µC @500Ω; Method: Balanced waveform0µC @500Ω; Method: Balanced waveformSame
MaximumCharge, (µC)Phase$17\mu C@500\Omega$$16.8\mu C @500\Omega$Similar
Maximum AverageCurrent, (mA)1.88mA0.924mASimilarNote 4
MaximumDensity,r.m.s.)Current(mA/cm2,$0.08mA/cm^2 @500\Omega$$0.462mA/cm^2$
MaximumPower(mW/cm2)AverageDensity,$3.65mW/cm^2 @500\Omega$$9.702mW/cm^2$
BiocompatibilityAll user directly contacting materialsSame

{14}------------------------------------------------

Comparison in details:

Note 1:

The proposed device SM9186 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 3 of the predicate device. The proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 4 to mode 6 are variation waveform of the mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9186 are a little different from predicate device K121719, but these differences are insignificant in the terms of safety or effectiveness.

Note 2:

The output channels of the proposed device is alternating while the predicate device is synchronous. Because the proposed device and predicate device adopt the same fundamental output technology and similar treatment effect. Therefore, this item is considered to be substantially equivalent. Also, the proposed device had passed AAMI / ANSI ES60601-1 and IEC 60601-2-10 test codes, so these differences won't raise any new safety and effectiveness issues.

Note 3:

There are some differences on the maximum output voltage, maximum output current, pulse width, frequency, pulse period, deviation between proposed device and predicate device. Based on the calculation of maximum

{15}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences won't raise any new safety and effectiveness issues.

Note 4:

The maximum phase charge of the proposed device SM9186 is larger than that of the predicate device. Both of them have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. The maximum average power density of proposed device is different from the predicate device due to they are calculated by different electrode area. Both of them meet maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectiveness issues.

Final conclusion:

The subject device SM9186 is substantial Equivalent to the predicate device K121719.

{16}------------------------------------------------

{17}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

are compliance with ISO10993-5 andISO10993-10 requirements.are compliance with ISO10993-5 andISO10993-10 requirements.
AccessoriesSelf-adhesive electrodes, electrode wires, Battery charger, USB cableSelf-adhesive electrodes, electrode wires, Battery charger, USB cableSame

Comparison in details:

Note 1:

The proposed device SM9196 has more treatment programs than the predicate device K121719, all treatment programs of proposed device are comparable to the mode 3 of the predicate device. The proposed device doesn't have treatment program to compare with the mode 6 of the predicate device because predicate device's mode 6 are variation waveform of the mode 1 to mode 3. All of the treatment programs have passed the IEC 60601-2-10 and AAMI / ANSI ES60601-1 test codes. So this difference doesn't raise any safety or effectiveness issue. And the weight, dimensions, appearance of proposed device SM9196 are a little different from predicate device K121719, but these differences are insignificant in the terms of safety or effectiveness.

Note 2:

The proposed device has 4 output channels are alternating & synchronous while the predicate device has 2 output channels and these channels are synchronous. The proposed device gives users more choices. Because the proposed device and predicate device adopt the same fundamental output technology and similar treatment effect. Therefore, this item is considered to be substantially equivalent. Also, the proposed device had passed AAMI / ANSI ES60601-1 and IEC 60601-2-10 test codes, so these differences won't raise any new safety and effectiveness issues.

Note 3:

There are some differences on the maximum output voltage, maximum output current, pulse width, frequency, pulse period, deviation between proposed device and predicate device. Based on the calculation of maximum current density, maximum average power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and these parameters have passed IEC 60601-2-10 test codes. So these differences won't raise any new safety and effectiveness issues.

Note 4:

The maximum current density, maximum average power density have some differences between proposed device and predicate device due to they are calculated by different electrode area. Both of them meet maximum current density <2mA/cm² and maximum average power density <0.25W/cm². Therefore these differences won't raise any new safety and effectiveness issues.

Final conclusion:

The subject device SM9196 is substantial Equivalent to the predicate device K121719.

{18}------------------------------------------------

4. Substantial Equivalence to Predicate device: Output Specifications

Table 5-SM 9079

ParameterNew DevicePredicateNew Device
Mode or program nameProgram 1Program 3Program 4Program 5Program 6Program 21Mode 3Program 2Program 7Program 8
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
Maximum@500Ω34.44234.434.440.840.84238.433.633.6
Output Voltage@2kΩ58.48358.458.470708465.256.856.8
(V)@10kΩ107144107107123123130117104104
Maximum@500Ω68.88468.868.881.681.68476.867.267.2
Output Current@2kΩ29.241.529.229.235354232.628.428.4
(mA)@10kΩ10.714.410.710.712.312.31311.710.410.4
Pulse Width(μs)190190190190190190100190190190
Frequency (Hz)23.81.61324.041.667~25.00.935~7.690.935~251.174.76224.0423.8
For interferential modes only:N/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
-Beat Frequency (Hz)
For multiphasicSymmetricalphasesYesYesYesYesYesYesYesYesYesYes
waveforms
only:Phaseduration190190190190190190100190190190
Net Charge (@500Ω) [uC]0000000000
Maximum Phase Charge (uC)11.09413.8711.3611.3613.3213.2216.812.1211.3511.08
Maximum Current Density (mA/cm²)0.0520.0040.050.0540.020.0650.00490.01160.0510.051
Maximum Power Density, (mW/cm²)1.780.181.81.870.812.640.1030.441.721.70
Burst Mode(a) Pulses perburstNANANANANANANA177949
(b) Bursts persecondNANANANANANANA0.1960.1640.327
(c) Burstduration(seconds)NANANANANANANA3.523.32.05
(d) DutyCycle: Line(b) x Line (c)NANANANANANANA0.690.540.67
ON Time (seconds)NANANANANANANA3.523.32.05
OFF Time (seconds)NANANANANANANA1.582.791
Additional FeaturesN/AN/AN/AN/ANANAN/AN/AN/AN/A
Continue
ParameterNew devicePredicateNew Device
Mode or program nameMode 2Program 9Program 10Program 11Program 12Program 13Program 15Program 16Program 17
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
Maximum@500Ω32.84234.834.834.434.434.434.434.434.4
Output Voltage@2kΩ568458.858.858.458.458.458.458.458.4
(V)@10kΩ103130107107107107107107107107
Maximum@500Ω65.68469.669.668.868.868.868.868.868.8
Output Current@2kΩ284229.429.429.229.229.229.229.229.2
(mA)@10kΩ10.31310.710.710.710.710.710.710.710.7
Pulse Width(us)100190190190190190190190190
Frequency (Hz)2512.5~55.52523.82523.824.392524.0425
For interferential modes only:Beat Frequency (Hz)N/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
For multiphasicwaveformsSymmetricalphasesYesYesYesYesYesYesYesYesYesYes
only:PhaseDuration190100190190190190190190190190
Net Charge (@500Ω) [uC]0000000000
Maximum Phase Charge (uC)10.5616.811.41511.4211.1711.0811.4911.5311.5511.65
MaximumCurrentDensity(mA/cm²)0.0520.23310.0550.0530.0540.0520.0530.0540.0520.054
MaximumPowerDensity,(mW/cm²)1.704.8951.921.831.871.781.831.871.81.87
Burst Mode(a) Pulses perburst130250~111029112912421816176.5149291
(b) Bursts persecond0.1240.0480.0670.1330.1670.0830.1160.2450.1390.067
(c) Burstduration(seconds)5.182011.625.44.969.146.623.066.211.64
(d) DutyCycle: Line(b) x Line (c)0.640.960.780.720.830.760.770.750.860.78
ON Time (seconds)5.182011.625.44.969.146.623.066.211.64
OFF Time (seconds)2.9413.242.1212.961.97113.26
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
Continue

{19}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

{20}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

ParameterNew DevicePredicated Device
Mode or program nameProgram 18Program 19Program 20Program 22Program 23Program 24Mode 1
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangular
Maximum@500Ω34.434.434.43834.434.442
Output Voltage@2kΩ58.458.458.458.458.458.484

{21}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

(V)@10kΩ107107107115107107130
Maximum@500Ω68.868.868.87668.868.884
Output@2kΩ29.229.229.229.229.229.242
Current(mA)@10kΩ10.710.710.711.510.710.713
Pulse Width(us)190190190190190190100
Frequency (Hz)24.524.524.044.76~25.023.823.868
For interferential modes only:N/AN/AN/AN/AN/AN/AN/A
-Beat Frequency (Hz)
For multiphasicSymmetrical phasesYesYesYesYesYesYesYes
waveformsPhase Duration190190190190190190100
only:
Net Charge (@500Ω) [uC]0000000
Maximum Phase Charge (uC)11.3211.4011.5512.2911.0611.5916.8
Maximum Current Density (mA/cm2)0.0530.0530.0520.060.0520.0520.2856
Maximum Power Density, (mW/cm2)1.841.841.82.291.781.785.998
Burst Mode(a) Pulses per burst14112720834-179404415204
(b) Bursts per second0.1370.1620.0950.0980.0450.0450.219
(c) Burst duration (seconds)5.765.178.667.1416.9817.443
(d) Duty Cycle: Line (b) x Line (c)0.790.840.820.70.760.780.66
ON Time (seconds)5.765.178.667.1416.9817.443
OFF Time (seconds)1.5211.963.025.414.781.56
Additional Features (specify, if applicable)N/AN/AN/AN/AN/AN/AN/A

Table 6-SM 9126

ParameterNew DevicePredicateNew Device
Mode or program nameProgram 1Program 8Program 9Program 10Program 13Program 14Mode 1Program 2Program 7Program 15
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
@500Ω@2kΩ@10kΩ@500Ω@2kΩ@10kΩ
MaximumOutput Voltage(V)30.88415930.472.412430.830.830.84.6
MaximumOutput Current(mA)61.64215.960.836.212.461.661.6849.2
4241.230.44.6
8411372.415.6
15916312324
61.682.460.89.2
4256.536.27.8
1316.312.32.4
Pulse Width(µs)220220220220220220220100220220220
Frequency (Hz)52.0840.9840.9840.9841.6752.63689.2638.4657.47
For interferential modes only:-Beat Frequency (Hz)N/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
For multiphasicwaveformsonly:SymmetricalphasesYesYesYesYesYesYesYesYesYes
Phaseduration220220220220220220100220220220
Net Charge (@500Ω) [uC]0000000000
Maximum Phase Charge (uC)13.710.513.713.713.713.716.81510.51.4
Maximum Current Density(mA/cm²)0.1170.0910.0930.0930.0940.1190.28560.030.090.019
Maximum Power Density,(mW/cm²)3.622.782.852.852.93.665.9981.155.210.09
Burst Mode(a) Pulses perburst21252025412752121620434149488
(b) Bursts persecond0.170.0610.120.2300.06250.1730.2190.180.1450.098
(c) Burstduration(seconds)4.0712.76.213.0912.484.133.723.878.5
Cycle: Line(b) x Line (c)
ON Time (seconds)4.0712.76.213.0912.484.133.723.878.5
OFF Time (seconds)1.683.522.31.283.521.661.561.842.981.7
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A

{22}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

{23}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

Continue

ParameterPredicateNew DevicePredicate
Mode or program nameMode 2Program 3Program 4Program 5Program 6Program 11Program 12Program 16Mode 3
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
MaximumOutput Voltage@500Ω4246.430.828.837.64.6114.642
@2kΩ84126846810015.620.815.684
(V)@10kΩ1301611591191532437.224130
MaximumOutput Current@500Ω8492.861.657.675.29.2229.284
@2kΩ42634234507.810.47.842
(mA)@10kΩ1316.115.911.915.32.43.722.413
Pulse Width(μs)100220220220220220220220100
Frequency (Hz)12.5~55.51.4740.983.33~37.041.79~13.51212.5113.657.471.17
For interferential modes only:N/AN/AN/AN/AN/AN/AN/AN/AN/A
Beat Frequency (Hz)
For multiphasicwaveformsonly:SymmetricalphasesPhaseDurationYesYesYesYesYesYesYesYesYes
100220220220220220220220100
Net Charge (@500Ω) [uC]000000000
Maximum Phase Charge (uC)16.817.2613.710.3914.542.024.182.0216.8
Maximum Current(mA/cm²)Density0.23310.0050.0930.0780.0370.0720.0920.0190.0049

{24}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

and the contraction of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution4F,Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126
MaximumPower Density(mW/cm²)4.8950.232.852.251.40.331.010.090.103
Burst Mode(a) Pulses per burst250~1110NANANANANANANANA
(b) Bursts per second0.048NANANANANANANANA
(c) Burst duration(seconds)20NANANANANANANANA
(d) Duty Cycle:Line (b) x Line (c)0.96NANANANANANANANA
ON Time (seconds)20NANANANANANANANA
OFF Time (seconds)1NANANANANANANANA
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/A

Table 7-SM 9186

ParameterNew DevicePredicateNew DevicePredicateNew Device
Mode or program nameProgram 1Program 6Program 9Program 12Mode 1Program 2Program 5Program 10Mode 2Program 3
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
Maximum @500Ω43.84838.442.44243.836.874263
Output @2kΩ577459.67684816613.184116
Voltage (V) @10kΩ14215610915213014912624.2130149
Maximum @500Ω87.69676.884.88487.673.61484126
Output @2kΩ28.53729.8384240.5336.554258
Current @10kΩ (mA)14.215.610.915.21314.912.62.421314.9

{25}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

Pulse Width(μs)Frequency (Hz)For interferential modes only:-Beat Frequency (Hz)FormultiphasicwaveformsSymmetricalphasesPhasedurationNet Charge (@500Ω) [uC]Maximum Phase Charge (uC)Maximum Current Density (mA/cm²)Maximum Power Density, (mW/cm²)Burst Mode(a) Pulses per burst(b) Bursts per second(c) Burst duration (seconds)(d) Duty Cycle: Line (b) x Line (c)ON Time (seconds)220220230220μs100220220230100224
31.25-58.82N/AYes2120.1863.60.673.632.26~58.8234.48~58.8237.03~59.5268N/A8.33~52.6392.59155.612.5~55.51.11
016.70.198.2812000.03420.40.720.4N/AN/AN/AN/AN/AN/A
014.50.219.947110.06412.080.7712.08N/AN/AN/AN/AN/AN/A
014.50.176.653330.1255.60.75.6N/AN/AN/AN/AN/AN/A
015.20.1857.852040.21930.663N/AN/AN/AN/AN/AN/A
016.80.28565.99822530.02342.80.9842.8N/AN/AN/AN/AN/AN/A
016.70.177.416390.0417.70.7317.7N/AN/AN/AN/AN/AN/A
015.10.259.215560.085100.8510N/AN/AN/AN/AN/AN/A
02.60.0840.58250~11100.048200.9620N/AN/AN/AN/AN/AN/A
016.80.23314.895N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
025.070.0050.33N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A

{26}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

OFF Time (seconds)1.88.63.62.41.5616.51.761N/A
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
Continue
ParameterNew DevicePredicate
Mode or program nameProgram 4Program 7Program 8Program 11Mode 3
WaveformBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangular
Maximum Output Voltage (V)@500Ω43.8@2kΩ66@10kΩ128@500Ω59@2kΩ94@10kΩ168@500Ω68@2kΩ117@10kΩ174@500Ω26.8@2kΩ44@10kΩ80@500Ω42@2kΩ84@10kΩ130
Maximum Output Current(mA)@500Ω76@2kΩ33@10kΩ12.8@500Ω118@2kΩ47@10kΩ16.8@500Ω136@2kΩ58.5@10kΩ17.4@500Ω53.6@2kΩ22@10kΩ8@500Ω84@2kΩ42@10kΩ13
Pulse Width(μs)220220220220100
Frequency (Hz)50~92.591.136~59.521.11~92.5992.591.17
For interferential modes only: -Beat Frequency (Hz)N/AN/AN/AN/AN/A
For multiphasic waveforms only:Symmetrical phasesYesPhase Duration220Yes220Yes220Yes220Yes100
Net Charge (@500Ω) [uC]00000
Maximum Phase Charge (uC)16.7221.5269.816.8
Maximum Current Density (mA/cm²)0.2580.260.460.1820.0049
Maximum Power Density, (mW/cm²)9.815.1931.44.880.103
Burst Mode(a) Pulses per burstN/A(b) Bursts per secondN/A(c) Burst duration (seconds)N/A(d) Duty Cycle: Line (b) x Line (c)N/AN/A
ON Time (seconds)N/AN/AN/AN/AN/A
OFF Time (seconds)N/AN/AN/AN/AN/A

{27}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

Additional Features (specify, if applicable)N/AN/AN/AN/AN/A
-----------------------------------------------------------------------

Table 8-SM 9196

ParameterNew DevicePredicateNew Device
Mode or program nameProgram 1Program 2Program 3Program 4Program 6Program 7Mode 1Program 5Program 8Program 9
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
MaximumOutput Voltage(V)@500Ω49.2484842.84847.2424652.450
@2kΩ73.673.67472.874.472.88471.678.8101
@10kΩ118124123120120120130119129133
MaximumOutput Current(mA)@500Ω98.4969685.69694.48492104.8100
@2kΩ36.836.83736.437.236.44235.839.450.5
@10kΩ11.812.412.31212121311.912.913.3
Pulse Width(us)200200200200200200100200200210
Frequency (Hz)47.1747.6247.1748.0849.0248.086848.089.4341.613
For interferential modes only:-Beat Frequency (Hz)N/AN/AN/AN/AN/AN/AN/AN/AN/AN/A
For multiphasicwaveformsonly:SymmetricalphasesYesYesYesYesYesYesYesYesYesYes
Phaseduration200200200200200200100200200210
Net Charge (@500Ω) [uC]0000000000
Maximum Phase Charge (uC)15.8415.2015.2014.515.2014.816.814.616.4816.8
MaximumCurrent(mA/cm²)Density0.070.070.0730.070.080.070.28560.070.020.0027
MaximumPower(mW/cm²)Density3.653.513.482.823.613.435.9983.260.830.14

{28}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

Burst Mode(a) Pulses per burst257238434303152173.1204249.133.47.74
(b) Bursts per second0.1320.1660.0820.120.2440.20.2190.1220.1940.16
(c) Burst duration (seconds)5.4459.26.33.13.635.183.544.8
(d) Duty Cycle: Line (b) x Line (c)0.720.830.750.760.760.710.660.630.690.76
ON Time (seconds)5.4459.26.33.13.635.183.544.8
OFF Time (seconds)2.141.023211.51.5631.621.5
Additional FeaturesN/AN/AN/AN/AN/AN/AN/AN/AN/AN/A

Continue

ParameterPredicateNew DevicePredicate
Mode or program nameMode 2Program 10Program 11Program 12Program 13Program 14Program 15Mode 3
WaveformBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasicBiphasic
ShapeRectangularRectangularRectangularRectangularRectangularRectangularRectangularRectangular
Maximum@500Ω4245.6465447.246.44642
Output@2kΩ84727287.673.670.473.684
Voltage (V)@10kΩ130112118131119117124130
Maximum@500Ω8491.29210894.492.89284
Output@2kΩ42363643.836.835.236.842
Current(mA)@10kΩ1311.211.813.111.911.712.413
Pulse Width(us)100210200200200200210100
Frequency (Hz)12.5~55.549.023.33~47.621.887~15.6348.0848.0847.171.17
For interferential modes only: -Beat FrequencyN/AN/AN/AN/AN/AN/AN/AN/A
(Hz)

{29}------------------------------------------------

Hong Qiangxing(Shen Zhen) Electronics Limited

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

Formultiphasicwaveformsonly:Symmetrical phasesYesYesYesYesYesYesYes
Phase Duration100210200200200200210100
Net Charge (@500Ω) [uC]00000000
Maximum Phase Charge (uC)16.813.614.617.014.814.915.916.8
Maximum Current Density (mA/cm²)0.23310.080.070.030.070.070.070.0049
Maximum Power Density, (mW/cm²)4.8953.423.221.463.433.313.350.103
Burst Mode(a) Pulses per burst250~1110N/AN/AN/AN/AN/AN/AN/A
(b) Bursts per second0.048N/AN/AN/AN/AN/AN/AN/A
(c) Burst duration (seconds)20N/AN/AN/AN/AN/AN/AN/A
(d) Duty Cycle: Line (b) x Line (c)0.96N/AN/AN/AN/AN/AN/AN/A
ON Time (seconds)20N/AN/AN/AN/AN/AN/AN/A
OFF Time (seconds)1N/AN/AN/AN/AN/AN/AN/A
Additional Features (specify, if applicable)N/AN/AN/AN/AN/AN/AN/AN/A

{30}------------------------------------------------

4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126

5. Test summary:

TENS & PMS are safe and effective as the predicate devices cited above. The new devices have passed testings according to the following standards:

  1. ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012

(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD);

  1. IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators:

  2. ANSI AAMI IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances --Requirements And Tests;

  3. IEC 62133 Edition 2.0 2012-12, IEC 62133 Edition 2.0 2012-12 Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications [Including: Corrigendum 1 (2013)]:

  4. ANSI AAMI HA60601-1-11 2015 Medical Electrical Equipment -- Part 1-11: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment (IEC 60601-1-11:2015 MOD);

The conclusion drawn from the testings are that the new devices are substantially equivalent to the predicate device. Furthermore, the new device complies with the recognized standards and performs its intended tasks as well as the legally marketed predicate devices.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).