K Number
K201260

Validate with FDA (Live)

Date Cleared
2020-09-24

(136 days)

Product Code
Regulation Number
878.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.

The Dermal Cooling System is intended to be used by trained healthcare professionals.

Device Description

The Dermal Cooling System is a cryosurgical device used to cool the skin, without the use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lower the temperature of the skin. It is intended for use in a healthcare facility such as a clinic or doctor's office.

The Dermal Cooling System is comprised of the following components:

  • control unit – houses user interface display, the system controller, and the power converters
  • chiller – provides circulating coolant to the handpiece to remove heat from the TEC
  • handpiece - contains the TEC, temperature sensors, the aluminum cooling plate, and user interface elements
  • isolation transformer – isolates system from AC mains power
  • cart – houses the isolation transformer, chiller, and control unit

The control unit initiates the treatment parameters and receives feedback from temperature sensors in the handpiece during the procedure. The user interface provides the mechanism for selection of the treatment plan, initiation of treatment, and display of treatment status.

The non-sterile, reusable handpiece contains the TEC which cools the aluminum contact plate to the treatment temperature, while the circulation of fluid past the TEC allows for the removal of heat. Thermistors affixed to the aluminum contact plate monitor the temperature at the treatment site.

The Dermal Cooling System has three operational modes: cooling off, pre-cool, and treatment. Cooling off represents the resting state; no power is supplied to the TEC in the handpiece. Pre-cool is the state during which the aluminum contact plate in the handpiece is cooled prior to treatment. Treatment is the normal treatment state in which the contact plate temperature is controlled to a pre-defined temperature defined in the treatment plan, as selected from the user interface on the control unit. The minimum temperature is -30°C and the maximum temperature is 40°C; the maximum treatment duration is 300 seconds.

The user interface is the means by which the user can view system status and state of each operational mode. The software provides three primary functions: allowing the user to select a treatment plan; establishing communications with the TEC controller and monitoring and displaying treatment progress.

Use of a TEC with closed-loop temperature feedback for cooling, as provided by the Dermal Cooling System, allows the selected temperature to be precisely and accurately maintained throughout the treatment. The TEC controller in the control unit communicates to the TEC module in the handpiece, maintaining the temperature defined in the treatment plan selected by the user. As heat is removed from the skin through contact with the aluminum plate, it is then transferred through the TEC module to the heat exchanger, which is fluid cooled. The chiller maintains a continuous flow of coolant which is circulated through the heat exchanger to carry heat way from the TEC module. The Dermal Cooling System generates sufficient thermal power to cool the aluminum contact plate to a pre-set temperature over a pre-set time. The temperature of the cooling surface, as measured by the thermistor affixed to the aluminum contact plate, is fed into the TEC controller which modulates power to the TEC such that the surface temperatures are precisely controlled. The Dermal Cooling System has an automatic shut-off feature if the temperature for the TEC and cooling plate are out of range.

The Dermal Cooling System operates from standard 120 VAC (60 Hz) power to provide controlled, active cooling at the skin interface throughout the procedure.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and the study:

The document describes the Dermal Cooling System, a cryosurgical instrument. The information provided focuses on the substantial equivalence determination for an expanded indication, rather than a detailed study proving the device meets acceptance criteria in a traditional sense for a novel device's performance. Instead, it demonstrates that the device already meets performance necessary for the expanded indication by comparing it to predicate devices and inherent capabilities.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

TestTest method/RequirementAcceptance criteriaReported Device Performance
System ValidationSystem performance when subjected to varied contours and body geometriesSystem is able to cool skin's surface to between +2°C and +4°C and maintain temperature for 10 minutesPassed
System ValidationSystem performance when subjected to varied contours and body geometriesHandpiece does not stick to tissue and no visible blanchingPassed

2. Sample size used for the test set and the data provenance

The document does not specify a distinct "test set" with a particular sample size for the System Validation tests. The evaluation appears to be based on bench testing of the device itself rather than a study involving a cohort of subjects. Therefore, information on data provenance (country of origin, retrospective/prospective) is not applicable in the context of human subject testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for the "System Validation" tests was established through direct measurement and observation of the device's physical performance (temperature, sticking, blanching) during bench testing, not through expert interpretation of data.

4. Adjudication method for the test set

Not applicable, as the evaluation was based on direct physical measurements and observations during bench testing, not subjective interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical cryosurgical instrument, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study is not relevant to its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm. Its "standalone" performance is its direct operational capability, as assessed by the bench tests.

7. The type of ground truth used

For the "System Validation" tests, the ground truth was based on direct physical measurements and observations of the device's temperature control capabilities and interaction with simulated tissue (implied by "varied contours and body geometries" and "handpiece does not stick to tissue and no visible blanching"). This is a form of engineering/performance validation rather than a clinical ground truth like pathology or outcomes data.

8. The sample size for the training set

Not applicable. The document does not describe the development of a machine learning model or algorithm that would require a training set. The device's operation is based on established engineering principles (thermoelectric cooling with closed-loop temperature feedback).

9. How the ground truth for the training set was established

Not applicable, as there is no training set for a machine learning model described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 24, 2020

R2 Technologies, Inc. Ms. Ragan Reppond Sr. Director, HR & Corporate Affairs 2603 Camino Ramon, Suite 200 San Ramon, California 94583

Re: K201260

Trade/Device Name: Dermal Cooling System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH, MLY Dated: July 24, 2020 Received: July 27, 2020

Dear Ms. Reppond:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201260

Device Name Dermal Cooling System

Indications for Use (Describe)

The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.

The Dermal Cooling System is intended to be used by trained healthcare professionals.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K201260

I.SUBMITTERR2 Technologies, Inc.2603 Camino Ramon, Suite 200Bishop Ranch 3San Ramon, CA 94583
Contact:Ms. Ragan ReppondSr. Director, HR & Corporate AffairsR2 Technologies, Inc.Phone: 925-378-4400rreppond@r2technologies.com
Date Prepared:July 24, 2020
II.DEVICE
Name of Device:Dermal Cooling System
Common or Usual Name:Cryosurgical unit and accessories
Classification Name:Cryosurgical unit and accessories (21 CFR 878.4350)
Regulatory Class:II
Product Code:GEH, MLY

III. PREDICATE DEVICE

The primary predicate device is the Dermal Cooling System, K171398. The secondary predicate device is the FROZEN C device, K182392.

IV. DEVICE DESCRIPTION

The Dermal Cooling System is a cryosurgical device used to cool the skin, without the use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lower the temperature of the skin. It is intended for use in a healthcare facility such as a clinic or doctor's office.

The Dermal Cooling System is comprised of the following components:

  • . control unit – houses user interface display, the system controller, and the power converters
  • chiller – provides circulating coolant to the handpiece to remove heat from the TEC
  • handpiece - contains the TEC, temperature sensors, the aluminum cooling plate, and user interface elements
  • . isolation transformer – isolates system from AC mains power
  • . cart – houses the isolation transformer, chiller, and control unit

{4}------------------------------------------------

The control unit initiates the treatment parameters and receives feedback from temperature sensors in the handpiece during the procedure. The user interface provides the mechanism for selection of the treatment plan, initiation of treatment, and display of treatment status.

The non-sterile, reusable handpiece contains the TEC which cools the aluminum contact plate to the treatment temperature, while the circulation of fluid past the TEC allows for the removal of heat. Thermistors affixed to the aluminum contact plate monitor the temperature at the treatment site.

The Dermal Cooling System has three operational modes: cooling off, pre-cool, and treatment. Cooling off represents the resting state; no power is supplied to the TEC in the handpiece. Pre-cool is the state during which the aluminum contact plate in the handpiece is cooled prior to treatment. Treatment is the normal treatment state in which the contact plate temperature is controlled to a pre-defined temperature defined in the treatment plan, as selected from the user interface on the control unit. The minimum temperature is -30°C and the maximum temperature is 40°C; the maximum treatment duration is 300 seconds.

The user interface is the means by which the user can view system status and state of each operational mode. The software provides three primary functions: allowing the user to select a treatment plan; establishing communications with the TEC controller and monitoring and displaying treatment progress.

Use of a TEC with closed-loop temperature feedback for cooling, as provided by the Dermal Cooling System, allows the selected temperature to be precisely and accurately maintained throughout the treatment. The TEC controller in the control unit communicates to the TEC module in the handpiece, maintaining the temperature defined in the treatment plan selected by the user. As heat is removed from the skin through contact with the aluminum plate, it is then transferred through the TEC module to the heat exchanger, which is fluid cooled. The chiller maintains a continuous flow of coolant which is circulated through the heat exchanger to carry heat way from the TEC module. The Dermal Cooling System generates sufficient thermal power to cool the aluminum contact plate to a pre-set temperature over a pre-set time. The temperature of the cooling surface, as measured by the thermistor affixed to the aluminum contact plate, is fed into the TEC controller which modulates power to the TEC such that the surface temperatures are precisely controlled. The Dermal Cooling System has an automatic shut-off feature if the temperature for the TEC and cooling plate are out of range.

The Dermal Cooling System operates from standard 120 VAC (60 Hz) power to provide controlled, active cooling at the skin interface throughout the procedure.

V. INDICATIONS FOR USE

The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.

The Dermal Cooling System is intended to be used by trained healthcare professionals.

{5}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Dermal Cooling System ("subject device") and the predicate devices are cryosurgical instruments that deliver extreme cold temperatures for cooling of the skin.

The primary predicate, the Dermal Cooling System, has not been modified for the proposed expanded indication, i.e., temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.

The secondary predicate, the FROZEN C, is included to substantiate the proposed expanded indications for use as the currently cleared indication for the FROZEN C (K182392) includes "temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures". The Dermal Cooling System and the FROZEN C have these same fundamental scientific technological elements:

  • . cooling devices, classified under 21 CRF 878.4350, product code GEH, that have the ability to deliver extreme cold;
  • . comprised of a cooling unit with a handpiece used to cool the skin in the area intended for treatment;
  • . software control of the treatment parameters, display of the treatment information, and monitoring of the temperature at the skin during treatment;
  • capability to provide the treatment conditions at the skin surface for the proposed indication.

The key difference that exists between the Dermal Cooling System and the FROZEN C is:

  • use of a chilled aluminum plate at the treatment site (cooled by a thermoelectric cooler, TEC).
    This difference in cooling delivery, the use of a chilled contact plate versus a refrigerant spray, does not raise new questions of safety and effectiveness with respect to the proposed expanded indication.

VII. PERFORMANCE DATA

Performance data were provided in support of the substantial equivalence determination, as summarized in the table on the following page.

Bench testing was completed to demonstrate the ability of the Dermal Cooling System to meet performance specifications (e.g., cooling characteristics when subjected to varied contours and body geometries).

TestTest method/RequirementAcceptance criteriaResults
SystemValidationSystem performance when subjected to varied contours and body geometriesSystem is able to cool skin's surface to between +2°C and +4°C and maintain temperature for 10 minutesPassed
Handpiece does not stick to tissue and no visible blanchingPassed

No other performance testing was performed for the subject device for this Traditional 510(k) since there was no change to the device design or methods for use required for the proposed expanded indication

{6}------------------------------------------------

and the range of treatment parameters previously cleared for the Dermal Cooling System (K171398) allows the same cooling characteristics at the skin surface as the secondary predicate, FROZEN 2 device.

No preclinical or clinical testing was performed.

VIII. CONCLUSIONS

The Dermal Cooling System that is the subject of this Traditional 510(k) is substantially equivalent to both the primary predicate (Dermal Cooling System) and the secondary predicate (FROZEN C). The subject device and both predicates are Class II devices intended for use to apply extreme cold to cool tissue. The subject Dermal Cooling System and the primary predicate Dermal Cooling System are the same device; there was no change required in the device design or methods for use for this expanded indication. The Dermal Cooling System can deliver the same treatment parameters to the skin surface and provides similar safety features as the secondary predicate FROZEN C device. The difference in the method by which the skin is cooled to the desired temperature between the subject device (contact cooling) and the secondary predicate (spray cooling) does not raise new questions of safety and efficacy for the proposed expanded indication.

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.