K Number
K201250

Validate with FDA (Live)

Device Name
SQuEEZ Software
Manufacturer
Date Cleared
2020-12-18

(221 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SQuEEZ Software is an image analysis software application for cardiac Computer Tomography (CT) studies to assist cardiologists and radiologists in assessing function when producing a cardiac evaluation. The software calculates a Stretch Quantifier for Endocardial Engraved Zones (SQuEEZ) value to highlight and color code the wall motions of the heart. Tools are provided to display regional motion properties of the heart.

Device Description

Cardiowise SQuEEZ Software is a post processing image analysis software application for cardiac Computer Tomography (CT) studies to assist cardiologists and radiologists in assessing left ventricle function when producing a cardiac evaluation. SQuEEZ Software enables a quantitative evaluation of the regional myocardial strain using the data from an ultra-high-resolution cardiac CT scan that requires an acquisition time of as short as a single heartbeat. This is done by calculating a Stretch Quantifier for Endocardial Engraved Zones (SQuEEZ) value to highlight and color code the wall motions of the heart. The resulting analysis is graphically presented in threedimensional color contour maps or regional graphical representations that indicate the level of strain the left ventricle experiences during the cardiac cycle.

AI/ML Overview

The provided text does not contain the detailed acceptance criteria and study results in the format requested. While it mentions performance data and studies, it lacks specific numerical acceptance criteria, the reported device performance against those criteria, and explicit details about the ground truth establishment, expert qualifications, and sample sizes for the test and training sets.

Here's a breakdown of what can be extracted and what is missing:

1. Table of Acceptance Criteria and Reported Device Performance:

  • Missing: The document does not provide a table with specific, measurable acceptance criteria (e.g., minimum accuracy, precision, or correlation thresholds) and the corresponding reported performance values from the device during a validation study.

2. Sample size used for the test set and the data provenance:

  • Mentioned/Implied:
    • "digital phantom study": No sample size given. Likely synthetically generated data.
    • "Comparison to tagged MRI (magnetic resonance imaging) in acutely infarcted dogs": This is an animal study, not human clinical data. No specific sample size (number of dogs) is mentioned.
    • "peer-reviewed publication describing a study of SQuEEZ values in human subjects with normal left ventricle function": This is a human study, but it's not explicitly stated as the test set for the regulatory submission, nor is its sample size provided in this document. The provenance (country of origin, retrospective/prospective) is also not stated for any of these.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Missing: This information is not provided for any of the mentioned studies. There is no mention of experts establishing ground truth for the digital phantom study or the animal study. For the human study, no details about expert involvement or qualifications are given.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Missing: No information on adjudication methods is present.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Missing: The document does not describe an MRMC study comparing human readers with and without AI assistance, nor does it provide an effect size for such a study. The device is described as "image analysis...to assist cardiologists," which implies aiding human readers, but no comparative effectiveness study with human readers is detailed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Implied: The "digital phantom study" and the "comparison to tagged MRI in acutely infarcted dogs" likely represent standalone performance, as they evaluate the algorithm's output against a reference without direct human intervention in the interpretation process of the SQuEEZ values. However, specific performance metrics are not given.

7. The type of ground truth used:

  • Mentioned/Implied:
    • Digital phantom study: Ground truth would likely be the known, simulated properties of the phantom.
    • Comparison to tagged MRI in acutely infarcted dogs: Ground truth is "tagged MRI," which is a validated imaging technique for assessing myocardial motion.
    • Peer-reviewed publication in human subjects: The document states this publication describes "normal left ventricle function," implying a ground truth of "normal function" based on clinical assessment, but the specific method of establishing this normal function (e.g., expert consensus, other imaging, pathology, outcomes data) is not detailed.

8. The sample size for the training set:

  • Missing: The document solely focuses on performance testing and does not mention the training set size or how the model was developed/trained.

9. How the ground truth for the training set was established:

  • Missing: No information provided regarding the training set or its ground truth.

In summary, the document states what kinds of studies were done (digital phantom, animal, human publication) to support substantial equivalence but does not provide the quantitative details of the acceptance criteria and performance results that would typically be found in such a study description.

{0}------------------------------------------------

December 18, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Cardiowise, Inc. % Pierre Bounaud, Ph.D. Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave, Suite B-257 SAN DIEGO CA 92110

Re: K201250

Trade/Device Name: SQuEEZ Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: November 23, 2020 Received: November 25, 2020

Dear Dr. Bounaud:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K201250

Device Name

SQuEEZ Software

Indications for Use (Describe)

The SQuEEZ Software is an image analysis for cardiac Computer Tomgraphy (CT) studies to assist cardiologists in assessing function when producing a cardiates a Stretch Quantifier for Endocardial Engraved Zones (SQUEZ) value to highlight and color code the wall motions of the heart. Tools are provided to display regional motion properties of the heart.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D) [ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displaysa currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Applicant/Submitter

Company Name:Cardiowise, Inc.
Phone Number:(479) 571-2592
Company Street Address535 W. Research Center Blvd., Suite 135
Fax Number:(479) 571-8814
City:Fayetteville
State:AR
Country:USA
Zip Code:72701

2. Contact Person

Full Name:John Coats
Job Title:CEO
Phone:(240) 405-8312
Email:jcoats@cardiowiseinc.com

3. Correspondent Information

Full Name:Pierre Bounaud, Ph.D.
Job Title:Principal Consultant
Phone:(619) 458-9547
Email:pbounaud@acknowledge-rs.com

4. Date of Preparation

Date of Preparation:

December 15, 2020

5. Device Information

Trade NameSQUEEZ Software
Common or Usual NamePicture archiving and communications system.
Classfication Name21 CFR 892.2050
Regulatory Class2
Product CodeLLZ

{4}------------------------------------------------

6. Predicate Device

Predicate Type510(k) NumberDevice NameManufacturer
Primary DeviceK112531VITREA CTVITAL IMAGES,
MYOCARDIALINC.
ANALYSIS

7. Device Description

Cardiowise SQuEEZ Software is a post processing image analysis software application for cardiac Computer Tomography (CT) studies to assist cardiologists and radiologists in assessing left ventricle function when producing a cardiac evaluation. SQuEEZ Software enables a quantitative evaluation of the regional myocardial strain using the data from an ultra-high-resolution cardiac CT scan that requires an acquisition time of as short as a single heartbeat. This is done by calculating a Stretch Quantifier for Endocardial Engraved Zones (SQuEEZ) value to highlight and color code the wall motions of the heart. The resulting analysis is graphically presented in threedimensional color contour maps or regional graphical representations that indicate the level of strain the left ventricle experiences during the cardiac cycle.

8. Intended Use/Indications for Use

The SQuEEZ Software is an image analysis software application for cardiac Computer Tomography (CT) studies to assist cardiologists and radiologists in assessing function when producing a cardiac evaluation. The software calculates a Stretch Quantifier for Endocardial Engraved Zones (SQuEEZ) value to highlight and color code the wall motions of the heart. Tools are provided to display regional motion properties of the heart.

9. Comparison of Technological Characteristics with Predicate ·

SQuEEZ Software is substantially equivalent to the predicate device based on the information summarized here:

The subject device has the same intended use as the predicate device cleared in K112531, i.e. software application for post-processing analysis of radiological images to assist in the assessment of the regional function of the heart muscle (myocardium).

The subject device and the device cleared in K112531 both use cardiac CT studies as inout for post-processing. The device cleared in K112531 achieves its intended use by evaluating heart tissue density as an indicator of wall thickening. Transmural wall thickening happens during the ventricular relaxation-contraction process, thus indirectly representing heart wall motions. The predicate device includes tools (segmentation, color-coding, polar maps) to visualize hypo/hyperdense areas of the myocardium. The subject device creates an endocardial mesh to directly track the left ventricle movements through the same ventricular relaxation-contraction process. Form the endocardial mesh, the subject device derives a calculated value, SQuEEZ, representing local motion of the endocardial surface. The subject device includes similar visualization tools for its SQuEEZ values as the predicate device (i.e. segmentation, color-coding, polar maps).

Based on the testing performed, including software verification and validation, digital

{5}------------------------------------------------

phantom study, and animal performance testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed SQuEEZ Software are assessed to be substantially equivalent to the predicate device.

10. Performance Data

The following tests were performed to demonstrate safety based on current industry standards:

  • . Software testing per IEC 62304
  • Non-clinical performance testing including: .
    • o Digital phantom study
    • o Comparison to tagged MRI (magnetic resonance imaging) in acutely infarcted dogs for noninvasive regional myocardial function analysis

The results of these tests indicate that SQuEEZ Software is substantially equivalent to the predicate device.

Additionally, a peer-reviewed publication describing a study of SQuEEZ values in human subjects with normal left ventricle function was provided to support a determination of substantial equivalence for the subject device.

11. Conclusion

Based on the testing performed, including software testing, digital phantom study, and non-clinical performance testing in animals, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed SQuEEZ Software are assessed to be substantially equivalent to the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).