K Number
K201223

Validate with FDA (Live)

Date Cleared
2020-06-04

(29 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LZI Tramadol Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of tramadol in human urine at the cutoff value of 100 ng/mL when calibrated against tramadol. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Device Description

The LZI Tramadol Enzyme Immunoassay is a homogeneous enzyme immunoassay with readyto-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available tramadol standard and referred to as tramadol-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, tramadol-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound tramadol-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.

The LZI Tramadol Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit. The LZI Tramadol Enzyme Immunoassay is traceable to a commercially available tramadol standard.

The Ri solution contains mouse monoclonal anti-tramadol antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with tramadol in buffer with sodium azide (0.09 %) as a preservative.

AI/ML Overview

The provided text describes the performance characteristics of the LZI Tramadol Enzyme Immunoassay. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly list "acceptance criteria" for precision, recovery, or method comparison in a consolidated table with specific numerical targets like "sensitivity > X%" or "specificity > Y%". Instead, it presents the raw performance data and concludes that the results are acceptable or show no significant interference. However, using the precision study's implicit goal (correct qualitative classification for samples ± 25% of cutoff) and the method comparison's agreement, and the recovery study's target, we can construct the following:

Acceptance Criterion (Implicit)Reported Device PerformanceStudy
Precision: Samples at 0%, 25%, 50%, 75% of cutoff should be Negative. Samples at 125%, 150%, 175%, 200% of cutoff should be Positive. (For 100% cutoff, some variance is expected due to inherent assay variability near the cutoff).100% Qualitative Agreement for 0, 25, 50, 75, 125, 150, 175, 200 ng/mL samples over 22 days/88 replicates. Samples at 100 ng/mL appropriately showed mixed results.Precision (Qualitative)
Analytical Recovery: Determined average percent recovery for spiked samples to be between 85% and 115% of target concentration.Recovery ranged between 89% - 116%.Analytical Recovery
Method Comparison (Qualitative Accuracy): High percentage agreement with LC/MS for clinical samples. (Implicitly close to 100% after addressing near-cutoff discrepancies).Overall % Agreement: 97.7% (Semi-Quantitative) and 95.3% (Qualitative)Method Comparison - Clinical Samples
Cross-reactivity: Structurally unrelated compounds should not cause false positives or interfere with results at ±25% of cutoff concentrations when present at high concentrations (e.g., 100,000 ng/mL).Most tested compounds showed no cross-reactivity and did not interfere with negative or positive trampadol samples.Cross-reactivity & Structurally Unrelated Compounds
Endogenous/Preservative Interference: Common endogenous and preservative compounds should not cause false positives or interfere with results at ±25% of cutoff concentrations when present at high concentrations (e.g., 1000 mg/dL).Boric Acid interfered at 1% w/v but no other significant interference observed.Endogenous and Preservative Compound Interference
Specific Gravity Interference: No interference observed across a range of specific gravity (1.000 to 1.030) for samples at 0 ng/mL, 75 ng/mL, and 125 ng/mL tramadol.No interference observed.Specific Gravity Interference
pH Interference: No major interference observed between pH 3 to pH 11 for samples at 0 ng/mL, 75 ng/mL, and 125 ng/mL tramadol.No major interference observed.pH Interference

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Test Set: Not clinical samples. Tramadol samples were prepared by spiking a tramadol standard into a pool of negative human urine.
    • For qualitative and semi-quantitative modes: 8 concentrations (0, 25, 50, 75, 100, 125, 150, 175, 200 ng/mL). Each concentration was tested in duplicate, two runs a day for 22 days, totaling 88 replicates per concentration.
  • Analytical Recovery: Not clinical samples. A drug-free urine pool spiked with tramadol at 400 ng/mL was serially diluted. Each dilution was run in 10 replicates.
  • Method Comparison - Clinical Samples: Eighty-six (86) unaltered clinical samples.
    • Provenance: Samples were collected by Lin-Zhi International, Inc. (LZI) and the University of California at San Francisco (UCSF, San Francisco). This indicates a retrospective collection of clinical samples from a US-based origin.
  • Cross-reactivity, Endogenous/Preservative Interference, Specific Gravity, pH Interference: Not clinical samples. Prepared by spiking substances into pooled negative human urine. All samples were tested in duplicates (except for Analytical Recovery, which was 10 replicates).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

N/A. This device is an in-vitro diagnostic assay for tramadol in urine. The ground truth for the clinical samples was established using a "more specific alternative chemical method," specifically LC/MS (Liquid Chromatography/Mass Spectrometry), which is a gold standard analytical technique for drug quantification, not human expert interpretation. For the spiked samples (precision, recovery, interference studies), the ground truth was the known concentration of the spiked substance, not expert consensus.

4. Adjudication Method for the Test Set

N/A. The ground truth for clinical samples was established by LC/MS analysis, which is an objective chemical measurement. There was no human expert adjudication of results, as there would be in an image analysis or diagnostic interpretation study.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This device is an automated enzyme immunoassay, not an AI-assisted diagnostic tool that involves human reader interpretation. Therefore, there is no "human readers improve with AI vs without AI assistance" effect size to report. The study focuses on the analytical performance of the immunoassay itself in comparison to a confirmatory method (LC/MS).

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are standalone (algorithm only) performance evaluations of the LZI Tramadol Enzyme Immunoassay. The assay is automated on a clinical chemistry analyzer, and its results are compared directly to known concentrations or LC/MS results without human interpretative input as part of its primary performance.

7. The Type of Ground Truth Used

  • For Clinical Samples (Method Comparison): LC/MS (Liquid Chromatography/Mass Spectrometry) was used as the confirmatory gold standard for tramadol concentrations.
  • For Spiked Samples (Precision, Analytical Recovery, Cross-reactivity, Interference Studies): The ground truth was the known concentration of tramadol or interfering substance spiked into negative human urine.

8. The Sample Size for the Training Set

N/A. This is an immunoassay, not a machine learning model. It does not have a "training set" in the context of AI/ML. The assay's performance is based on its chemical and enzymatic reactions, calibrated using known standards.

9. How the Ground Truth for the Training Set was Established

N/A. As mentioned above, there is no "training set" for this immunoassay device. The assay uses calibrators (0, 50, 100, 225, and 400 ng/mL tramadol) to establish its calibration curve, which determines the relationship between enzyme activity and tramadol concentration. These calibrators represent known, established concentrations.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue. The logos are placed side by side.

June 4, 2020

Lin-Zhi International, Inc. Bernice Lin VP Operations 2945 Oakmead Village Court Santa Clara, CA 95051

Re: K201223

Trade/Device Name: LZI Tramadol Enzyme Immunoassay Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG Dated: Mav 4, 2020 Received: May 6, 2020

Dear Bernice Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201223

Device Name

LZI Tramadol Enzyme Immunoassay

Indications for Use (Describe)

The LZI Tramadol Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of tramadol in human urine at the cutoff value of 100 ng/mL when callbrated against tramadol. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as gas or liquid chromatography/mass spectrometry (GC/MS or LC/MS) or (2) permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatography and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safetv and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number

K201223

Prepared On

May 4, 2020

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter Name, Address, and Contact:

Lin-Zhi International, Inc. 2945 Oakmead Village Court Santa Clara, CA 95051 Phone: (408) 970-8811 (408) 970-9030 Fax: e-mail: bclin@lin-zhi.com

Contact:Bernice Lin, Ph.D.
VP Operations

Device Name and Classification

Classification Name:Enzyme Immunoassay, Opiates Class II, DJG (91 Toxicology), 21 CFR 862.3650
----------------------------------------------------------------------------------------------------
Common Name:Homogeneous Enzyme Immunoassay
Proprietary Name:LZI Tramadol Enzyme Immunoassay

{4}------------------------------------------------

Legally Marketed Predicate Device(s)

The LZI Tramadol Enzyme Immunoassay (EIA) is substantially equivalent to the ARKTM Tramadol Assay (K182280) manufactured by ARK Diagnostics, Inc. The LZI Tramadol Enzyme Immunoassay is identical or similar to its predicate in terms of intended use, method principle, device components, and clinical performance.

Device Description

The LZI Tramadol Enzyme Immunoassay is a homogeneous enzyme immunoassay with readyto-use liquid reagents. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. The drug-labeled G6PDH conjugate is traceable to a commercially available tramadol standard and referred to as tramadol-labeled G6PDH conjugate. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, tramadol-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug; the unbound tramadol-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm.

The LZI Tramadol Enzyme Immunoassay is a kit comprised of two reagents, an R1 and R2 which are bottled separately but sold together within the kit. The LZI Tramadol Enzyme Immunoassay is traceable to a commercially available tramadol standard.

The Ri solution contains mouse monoclonal anti-tramadol antibody, glucose-6-phosphate (G6P) nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. The R2 solution contains glucose-6-phosphate dehydrogenase (G6PDH) labeled with tramadol in buffer with sodium azide (0.09 %) as a preservative.

{5}------------------------------------------------

Intended Use

The LZI Tramadol Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of tramadol in human urine at the cutoff value of 100 ng/mL when calibrated against tramadol. The assay is designed for prescription use with a number of automated clinical chemistry analyzers.

The semi-quantitative mode is for purposes of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC/MS and LC/MS or (2) permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical result. A more specific alternative chemical method (e.g., gas or liquid chromatograpy and mass spectrometry) must be used in order to obtain a confirmed analytical result. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

{6}------------------------------------------------

Comparison to Predicate Device

The LZI Tramadol Enzyme Immunoassay is substantially equivalent to the ARK™ Tramadol Assay cleared by the FDA under the premarket notification K182280 for its stated intended use.

The following table compares LZI's Tramadol Enzyme Immunoassay with the predicate device.

DeviceSubject DevicePredicate Device (K182280)
CharacteristicsLZI Tramadol Enzyme ImmunoassayARK™ Tramadol Assay
Intended UseSameThe ARK Tramadol Assay is animmunoassay intended for the qualitativeand/or semiquantitative determination oftramadol in human urine at a cutoffconcentration of 100 ng/mL. The assay isintended for use in laboratories withautomated clinical chemistry analyzers.This in vitro diagnostic device is forprescription use only.The semiquantitative mode is for thepurpose of (1) enabling laboratories todetermine an appropriate dilution of thespecimen for confirmation by aconfirmatory method, such asGas Chromatography/Mass Spectrometry(GC/MS) or LiquidChromatography/tandem MassSpectrometry (LC-MS/MS), or (2)permitting laboratories to establish qualitycontrol procedures.The ARK Tramadol Assay provides onlya preliminary analytical test result. Amore specific alternative chemical methodmust be used in order to obtain aconfirmed positive analytical result. GasChromatography/Mass Spectrometry(GC/MS) or LiquidChromatography/tandem MassSpectrometry (LC-MS/MS) is thepreferred confirmatory method. Clinicalconsideration and professional judgmentshould be exercised with any drug testresult, particularly when the preliminarytest result is positive.
Analytetramadoltramadol
Cutoff100 ng/ml100 ng/mL
MatrixUrineUrine
Calibrators Level0, 50, 100, 225, and 400 ng/mL0, 100, 200, 500, and 1000 ng/mL
Controls Level75 ng/mL and 125 ng/mL75 ng/mL and 125 ng/mL
Storage2-8 ℃ until expiration date2-8 ℃ until expiration date

{7}------------------------------------------------

Performance Characteristics Summary:

All validation studies below were conducted on the Beckman Coulter® AU480 Analyzer

Precision: 100 ng/mL Cutoff

The assay was tested in qualitative (ΔΟΣ, mAU) and semi-quantitative (ng/mL) mode using a modified NCCLS-EP5 protocol. Tramadol sample concentrations were prepared by spiking a tramadol standard into a pool of negative human urine at concentrations ±25 %, ±50 %, ±75 %, and ±100 % of the cutoff concentration.

Results shown below were obtained by testing all samples in replicate of two, two runs a day (one in the morning and one in the afternoon) for 22 days on one Beckman Coulter® AU480 automatic clinical analyzer for a total of 88 replicates. Samples were evaluated against the OD of the cutoff calibrator in the qualitative mode and evaluated against the assay's calibration curve in the semi-quantitative mode. One single lot of reagents, calibrators, and controls were used and stored at 2-8℃ when not in use.

TramadolWithin Run (N=22)Total Precision (N=88)
ConcentrationMeanQualitative ResponseMeanQualitative Response
0 ng/mL-1.0--1.0-
25 ng/mL20.2-20.2-
50 ng/mL49.8-49.8-
75 ng/mL74.6-74.6-
100 ng/mL101.5+101.5+
125 ng/mL130.5+130.5+
150 ng/mL162.7+162.7+
175 ng/mL186.2+186.2+
200 ng/mL210.2+210.2+

Semi-Quantitative Precision Analysis Summary: Qualitative Results

Semi-Quantitative Positive/Negative Results:

100 ng/mL Cutoff Result:Within Run (N=22)Total Precision (N=88)
TramadolConcentration% of CutoffNumber ofDeterminationImmunoassayResultNumber ofDeterminationImmunoassayResult
0 ng/mL0 %2222 Negative8888 Negative
25 ng/mL25 %2222 Negative8888 Negative
50 ng/mL50 %2222 Negative8888 Negative
75 ng/mL75 %2222 Negative8888 Negative
100 ng/mL100 %2216 Pos/6 Neg8853 Pos/35 Neg
125 ng/mL125 %2222 Positive8888 Positive
150 ng/mL150 %2222 Positive8888 Positive
175 ng/mL175 %2222 Positive8888 Positive
200 ng/mL200 %2222 Positive8888 Positive

{8}------------------------------------------------

100 ng/mL Cutoff Result:Within Run (N=22)Total Precision (N=88)
TramadolConcentration% of CutoffNumber ofDeterminationImmunoassayResultNumber ofDeterminationImmunoassayResult
0 ng/mL0 %2222 Negative8888 Negative
25 ng/mL25 %2222 Negative8888 Negative
50 ng/mL50 %2222 Negative8888 Negative
75 ng/mL75 %2222 Negative8888 Negative
100 ng/mL100 %2215 Pos/7 Neg8848 Pos/40 Neg
125 ng/mL125 %2222 Positive8888 Positive
150 ng/mL150 %2222 Positive8888 Positive
175 ng/mL175 %2222 Positive8888 Positive
200 ng/mL200 %2222 Positive8888 Positive

Qualitative Positive/Negative Results:

Analytical Recovery:

To demonstrate recovery of the entire assay range, a drug free-urine pool spiked with tramadol at 400 ng/mL was serially diluted. Each sample was run in 10 replicates and the average was used to determine percent recovery compared to the expected target value.

Determined concentration averages were obtained and all averages were ±15 % of the target concentrations. The determined average percent recovery (Determined Concentration Average divided by the Target Concentration) were considered acceptable between 85 – 115 %.

The recovery of tramadol spiked to various concentrations was evaluated and was found to range between 89 % - 116 %.

TargetConcentration(ng/mL)DeterminedConcentration Range(ng/mL)DeterminedConcentration Average(ng/mL)Average% Recovery
400395.2 – 433.3412.7103.2 %
360350.3 - 404.5377.5104.9 %
320316.5 - 371.6346.9108.4 %
280277.6 - 313.0297.8106.4 %
240221.7 – 279.7245.1102.1 %
200202.0 - 215.5210.0105.0 %
160168.2 – 180.2174.5109.1 %
120122.8 – 134.2130.1108.4 %
8078.0 – 84.680.9101.1 %
4035.6 - 43.039.398.2 %
0-4.6 - 1.3-2.2N/A

{9}------------------------------------------------

Method Comparison - Clinical Samples:

A total of eighty-six (86) unaltered clinical samples were tested with the LZI Tramadol Enzyme Immunoassay on the Beckman Coulter® AU480 automated clinical analyzer. Samples were evaluated against the OD of the cutoff calibrator in the qualitative mode and evaluated against the assay's calibration curve in the semi-quantitative mode. All samples were tested in singlet.

All samples were confirmed with LC/MS for tramadol concentrations. Samples were collected by Lin-Zhi International, Inc. (LZI) and and the University of California at San Francisco (UCSF, San Francisco).

TramadolResults100ng/mLCutoffNegative< 50 % of thecutoffconcentrationby LC/MSanalysisNear CutoffNegative(Between 50 %below the cutoffand the cutoffconcentration)Near CutoffPositive(Between thecutoff and 50% above thecutoffconcentration)High Positive(Greater than50 % abovethe cutoffconcentration)%Agreement
Positive001*93397.7 %
Negative207151**097.7 %

Semi-Quantitative Results:

Discrepant samples determined when comparing LC/MS tramadol results with tramadol EIA results on the Beckman Coulter® AU480 automated clinical analyzer.

Sample#LC/MSTramadol(ng/mL)Pos/NegResultLC/MSO-desmethyltramadol(ng/mL)Pos/NegResult
38*71-118+
46**114+57-
  • Discrepant between 50% below cutoff and cutoff concentration (50 - 99.9 ng/mL)

** Discrepant between cutoff and 50% above cutoff concentration (100 - 149.9 ng/mL)

{10}------------------------------------------------

Method Comparison - Clinical Samples (continued):

TramadolResults100ng/mLCutoffNegative< 50 % of thecutoffconcentrationby LC/MSanalysisNear CutoffNegative(Between 50 %below the cutoffand the cutoffconcentration)Near CutoffPositive(Between thecutoff and 50% above thecutoffconcentration)High Positive(Greater than50 % abovethe cutoffconcentration)%Agreement
Positive002*93397.7 %
Negative207141**095.3 %

Qualitative Accuracy Study:

Discrepant samples determined when comparing LC/MS tramadol results with tramadol EIA results on the Beckman Coulter® AU480 automated clinical analyzer.

Sample#LC/MSTramadol(ng/mL)Pos/NegResultLC/MSO-desmethyltramadol(ng/mL)Pos/NegResult
38*71-118++
42*85-56+
46**114+57-
  • Discrepant between 50% below cutoff and cutoff concentration (50 – 99.9 ng/mL)

** Discrepant between cutoff and 50% above cutoff concentration (100 - 149.9 ng/mL)

{11}------------------------------------------------

Cross-reactivity

The cross-reactivity of various potentially interfering drugs were tested by spiking various concentrations of each substance into a pool of negative human urine and then evaluated against the assay's calibration curve in both qualitative and semi-quantitative modes. All samples were tested in duplicates.

The table below lists the concentration of each test compound that gave a response approximately equivalent to that of the cutoff calibrator (as positive) or the maximal concentration of the compound tested that gave a response below the response of the cutoff calibrator (as negative). Compounds tested at high concentration (100,000 ng/mL) with results below the cutoff value were listed as Not Detected (ND). Compounds tested below the high concentration (100,000 ng/mL) that gave a result below the cutoff value were given a "< %" value.

CompoundTestConcentration(ng/mL)% Cross-reactivity
Tramadol100100.00 %
O-desmethyl cis-tramadol HCl16660.24 %
N-desmethyl-cis-tramadol10,0001.00 %
Rac N, O-didesmethyltramadol15,0000.67 %
O-desmethyl tramadol beta-D-glucoronide90111.11 %
Ketamine100,000ND
Dehydronorketamine10,000< 1.00 %
Norketamine20,000< 0.50 %
Phencyclidine (PCP)100,000ND
Venlafaxine100,000ND
O-desmethylvenlafaxine100,000ND

{12}------------------------------------------------

Cross-reactivity (continued)

Structurally unrelated compounds were additionally spiked into pooled negative human urine to described above). These solutions were then split into three portions; one without tramaning two that were further spiked with tramadol standards to a final tramadol concentration of 75 ng/mL (as negative or positive controls, ±25 % of the cutoff concentration, respectively). Samples were then evaluated against the assay's calibration curve in both qualitative and semiquantitative modes. All samples were tested in duplicates. Compounds tested at high concentration (100,000 ng/mL) with results below the cutoff value were listed as Not Detected (ND). Compounds tested below the high concentration (100,000 ng/mL) that gave a result below the cutoff value were given a "< %" value.

Test0 ng/mL-25 % Tramadol Cutoff+25 % Tramadol Cutoff
CompoundConcentrationTramadol(75 ng/mL)(125 ng/mL)
(ng/mL)% CrossResultResult
6-AcetyImorphine100,000NDNegPos
Acetaminophen100,000NDNegPos
Acetylsalicylic Acid100,000NDNegPos
Aliemazine Tartrate100,000NDNegPos
Amitriptyline100,000NDNegPos
Amlodipine Besylate100,000NDNegPos
d-Amphetamine100,000NDNegPos
Amoxicillin100,000NDNegPos
Atorvastatin100,000NDNegPos
Benzoylecgonine100,000NDNegPos
Buprenorphine100,000NDNegPos
Bupropion100,000NDNegPos
Caffeine100,000NDNegPos
Carbamazepine100,000NDNegPos
Cetirizine100,000NDNegPos
Chlorpheniramine100,000NDNegPos
Chlorpromazine100,000NDNegPos
Clomipramine100,000NDNegPos
Codeine100,000NDNegPos
Desipramine100,000NDNegPos
Diphenhydramine100,000NDNegPos
Doxylamine Succinate100,000NDNegPos
CompoundTestConcentration(ng/mL)0 ng/mLTramadol% Cross-25 % Tramadol Cutoff(75 ng/mL)Result+25 % Tramadol Cutoff(125 ng/mL)Result
Duloxetine100,000NDNegPos
Etavirenz100,000NDNegPos
Fentanyl (citrate)100,000NDNegPos
Fluoxetine100,000NDNegPos
Fluphenazine100,000NDNegPos
Gabapentin100,000NDNegPos
Hydrocodone100,000NDNegPos
Hydromorphone100,000NDNegPos
Hydroxyzine Pamoate100,000NDNegPos
Ibuprofen100,000NDNegPos
Imipramine100,000NDNegPos
JWH-073 (SPICE I)100,000NDNegPos
Lisinopril100,000NDNegPos
Loratidine100,000NDNegPos
Lorazepam100,000NDNegPos
Losartan100,000NDNegPos
MDA (3,4-methylenedioxyamphetamine)100,000NDNegPos
MDEA100,000NDNegPos
MDMA (3,4-methylenedioxymethamphetamine)100,000NDNegPos
Meperidine100,000NDNegPos
Metformin100,000NDNegPos
Methadone100,000NDNegPos
d-Methamphetamine100,000NDNegPos
Methapyrilene HCl100,000NDNegPos
Metoprolol100,000NDNegPos
Morphine100,000NDNegPos
Nicotine100,000NDNegPos
Niflumic Acid100,000NDNegPos
CompoundTestConcentration(ng/mL)0 ng/mLTramadol% Cross-25 % Tramadol Cutoff(75 ng/mL)Result+25 % Tramadol Cutoff(125 ng/mL)Result
Nortriptyline100,000NDNegPos
Omeprazole100,000NDNegPos
Oxazepam100,000NDNegPos
Oxycodone100,000NDNegPos
Oxymorphone100,000NDNegPos
Phenobarbital100,000NDNegPos
Prochloreazine100,000NDNegPos
d-Propoxyphene100,000NDNegPos
(1S,2S)-(+)Pseudoephedrine100,000NDNegPos
Quetiapine100,000NDNegPos
Ranitidine100,000NDNegPos
Salbutamol (Albuterol)100,000NDNegPos
Sertraline100,000NDNegPos
THC-COOH(11-Nor-Delta-9-THC-9-carboxylic acid)100,000NDNegPos
l-Thyroxine100,000NDNeg
Trioridazine100,000NDNegPos
(+)Verapamil HCl100,000NDNegPos
Zolpidem1,250< 0.10%NegPos
Zolpidem phenyl-4-carboxylicacid10,000< 0.10%NegPos
Zolpidem-6-carboxylic acid10,000< 0.10%NegPos

Structurally Unrelated Pharmacological Compounds:

{13}------------------------------------------------

Structurally Unrelated Pharmacological Compounds, continued:

{14}------------------------------------------------

Structurally Unrelated Pharmacological Compounds, continued:

It is possible that other substances and/or factors not listed above may interfere with the test and cause false results, e.g., technical or procedural errors

{15}------------------------------------------------

Performance Characteristics Summary, continued:

Beckman Coulter® AU480 Analyzer

Endogenous and Preservative Compound Interference:

Endogenous and Preservative compounds were spiked into pooled negative human urine to desired concentrations. These solutions were then split into three portions; one without tramaning two that were further spiked with tramadol standards to a final tramadol concentration of 75 ng/mL (as negative or positive controls, ±25 % of the cutoff concentration, respectively). Samples were then evaluated against the assay's calibration curve in both qualitative modes. All samples were tested in duplicates.

Interfering SubstanceConcentrationof Compound(mg/dL)0 ng/mLTramadol-25 % Tramadol Cutoff(75 ng/mL)+25 % Tramadol Cutoff(125 ng/mL)
Acetone1000NegNegPos
Ascorbic Acid1500NegNegPos
Bilirubin2NegNegPos
Boric Acid1000NegNegNeg
Calcium Chloride (CaCl2)300NegNegPos
Ciprofloxacin1NegNegPos
Citric Acid (pH 3)800NegNegPos
Creatinine500NegNegPos
Ethanol1000NegNegPos
Galactose10NegNegPos
y-Globulin500NegNegPos
Glucose3000NegNegPos
Hemoglobin300NegNegPos
β-hydroxybutyric Acid100NegNegPos
HSA500NegNegPos
Oxalic Acid100NegNegPos
Potassium Chloride6000NegNegPos
Riboflavin0.3NegNegPos
Urea6000NegNegPos
Uric Acid10NegNegPos
Sodium Azide1000NegNegPos

{16}------------------------------------------------

Performance Characteristics Summary, continued:

Beckman Coulter® AU480 Analyzer

Endogenous and Preservative Compound Interference (continued):

Interfering SubstanceConcentrationof Compound(mg/dL)0 ng/mLTramadol-25 % Tramadol Cutoff(75 ng/mL)+25 % Tramadol Cutoff(125 ng/mL)
Sodium Chloride6000NegNegPos
Sodium Fluoride1000NegNegPos
Sodium Phosphate300NegNegPos

The following endogenous and preservative compounds which showed interference at ±25 % of cutoff concentrations were then spiked into negative urine and at ±50 % of the cutoff concentration (50 ng/mL and 150 ng/mL) for the assay.

Interference was observed with Boric Acid at 1 % w/v. No other significant undesired cross-reactants or endogenous/preservative substance interference was observed.

Interfering SubstanceConcentrationof Compound(mg/dL)0 ng/mLTramadol-50 % Tramadol Cutoff(50 ng/mL)-50 % Tramadol Cutoff(150 ng/mL)
Boric Acid1000NegNegNeg

{17}------------------------------------------------

Performance Characteristics Summary, continued:

Beckman Coulter® AU480 Analyzer

Specific Gravity Interference:

Samples ranging in specific gravity from 1.000 to 1.030 were split into three portions each and either spiked or further spiked to a final tramadol concentration of either 75 ng/mL (as negative or positive controls, ±25 % of the cutoff concentration, respectively). These samples were then evaluated in both semi-quantitative modes. No interference was observed.

SpecificGravityValue0 ng/mLTramadol-25 % Tramadol Cutoff(75 ng/mL)+25 % Tramadol Cutoff(125 ng/mL)
1.000NegNegPos
1.003NegNegPos
1.005NegNegPos
1.006NegNegPos
1.007NegNegPos
1.008NegNegPos
1.009NegNegPos
1.010NegNegPos
1.011NegNegPos
1.012NegNegPos
1.013NegNegPos
1.015NegNegPos
1.018NegNegPos
1.020NegNegPos
1.025NegNegPos
1.028NegNegPos
1.030NegNegPos

{18}------------------------------------------------

pH Interference:

Negative urine and urine spiked with tramadol to the final tramadol concentration of either 75 ng/mL or 125 ng/mL (as negative or positive controls, ±25 % of the cutoff concentration, respectively) were adjusted to the following pH levels and tested by the assay. The pH adjusted solutions were evaluated in both qualitative modes.

No major interference was observed between pH 3 to pH 11. Results are summarized in the following table:

InterferingSubstance0 ng/mLTramadol-25 % Tramadol Cutoff(75 ng/mL)+25 % Tramadol Cutoff(125 ng/mL)
pH 3NegNegPos
pH 4NegNegPos
pH 5NegNegPos
pH 6NegNegPos
pH 7NegNegPos
pH 8NegNegPos
pH 9NegNegPos
pH 10NegNegPos
pH 11NegNegPos

Summary:

The information provided in this pre-market notification demonstrates that the LZI Tramadol Enzyme Immunoasay is substantially equivalent to the legally marketed predicate device for its general intended use. Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available predicate device as confirmed by chromatographymass spectrometry (GCMS), an independent analytical method. The information supplied in this premarket notification provides reasonable assurance that the LZI Tramadol Enzyme Immunoassay is safe and effective for its stated intended use.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).