K Number
K201059

Validate with FDA (Live)

Date Cleared
2020-07-30

(100 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph (Cardiac), Laparoscopic, OB/Gyn and Urology. The Modes of Operation include B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Tissue Harmonic Imaging, Power Doppler Imaging, Directional Power Doppler Imaging, Tissues Doppler Imaging, Pulse Inversion Harmonic Imaging, 3D/4D Imaging mode, Elastography Imaging, Contrast imaging, Panoramic Imaging, Trapezoid Imaging and their combination modes, and the system is intended to be used in a hospital or medical clinic.

Device Description

This SonoScape S60 Elite Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array. This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes. This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler, Power Doppler and Directional Power Doppler Imaging, or the combination of these modes, Contrast lmaging, Elastography, 3D/4D.

AI/ML Overview

The medical device in question is the Sonoscape S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System (K201059).

Based on the provided document, here's the information regarding its acceptance criteria and supporting study:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Standard/Test NameDevice Performance (Result)Details / Remarks
Electrical SafetyIEC 60601-1:2005+A1:2012PassedMet General requirements for basic safety and essential performance.
Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014PassedMet requirements for electromagnetic disturbances.
Acoustic SafetyIEC 60601-2-37:2015PassedMet particular requirements for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment. Acoustic Output Limits: Derated ISPTA: 720mW/cm2 maximum, TIS/TIB/TIC: 6.0 maximum, MI: 1.9 maximum for all listed probes (3C-A, C1-6A, C1-6, C2-9, C322, 6CT-A, 6CI-A, 12L-A, 12L-B, 9L-A, L3-9, L741, L742, 10I2, 12LT-A, 12LI-A, 4P-A, 3P-A, S1-5, VE9-5, 6V1, 6V3, 6V7).
AIUM/NEMA UD 2:2004 (R2009)PassedMet acoustic output measurement standard for diagnostic ultrasound equipment.
BiocompatibilityISO 10993-5:2009ConformedMet tests for in vitro cytotoxicity.
ISO 10993-10:2010ConformedMet tests for irritation and skin sensitization.
Software Verification and ValidationIEC 62304:2006 +A1:2015Passed
Functional Equivalence to Predicate DeviceComparison to SonoScape S60 Series (K172082) & Philips EPIQ 5/7 (K172607) & SonoScape P10 Series (K173058)Substantially EquivalentDemonstrated similar intended uses, compliance with regulations, consistent acoustic output, and similar probes and technical characteristics. Differences in frequency, operation modes (DPDI vs PDI), and functions are deemed not to raise new risks according to the SE Analyses (5, 6, 7).

2. Sample Size Used for the Test Set and Data Provenance:

  • The document states that non-clinical tests (electrical safety, EMC, acoustic, biocompatibility, software verification/validation) were performed.
  • No clinical testing was required for this 510(k) submission.
  • Therefore, there is no specific "test set" in terms of patient data for an AI algorithm as typically understood in such studies. The evaluation focused on engineering and performance parameters against established standards.
  • Data provenance: Not applicable as no clinical data test set was used.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable. Since no clinical test set was utilized and no AI algorithm with independent ground truth establishment was conducted, there were no experts needed to establish ground truth for a clinical test set. The device is an ultrasound system, not an AI diagnostic tool requiring clinical ground truth for performance evaluation in this submission.

4. Adjudication Method for the Test Set:

  • Not applicable. No clinical test set or subjective interpretations requiring adjudication were part of this 510(k) submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

  • No. An MRMC comparative effectiveness study was not performed. This submission is for a general-purpose ultrasound system, not an AI-assisted diagnostic tool that would typically undergo such a study to evaluate reader performance with and without AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • No. This is not an AI algorithm. It is a digital color Doppler ultrasound system. The evaluation was of the system's compliance with safety and performance standards.

7. The Type of Ground Truth Used:

  • For the non-clinical tests, the "ground truth" was based on the requirements and methodologies outlined in the specified international medical device standards (IEC, ISO, AIUM/NEMA). These standards define objective parameters and testing procedures for assessing the safety and performance of ultrasound systems.

8. The Sample Size for the Training Set:

  • Not applicable. This submission does not pertain to an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As there is no AI algorithm training set, no ground truth needed to be established for it.

{0}------------------------------------------------

July 30, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Sonoscape Medical Corp. % Toki Wu Regulatory Affairs Manager Room 201 & 202, 12th Building, Shenzhen Software Park Phase II 1 Keji Middle 2nd Rd, Yuehai Subdistrict, Nanshan District Shenzhen, Guangdong 518057 CHINA

Re: K201059

Trade/Device Name: S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 2, 2020 Received: July 6, 2020

Dear Toki Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201059

Device Name

S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System

Indications for Use (Describe)

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph (Cardiac), Laparoscopic, OB/Gyn and Urology. The Modes of Operation include B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Tissue Harmonic Imaging, Power Doppler Imaging, Directional Power Doppler Imaging, Tissues Doppler Imaging, Pulse Inversion Harmonic Imaging, 3D/4D Imaging mode, Elastography Imaging, Contrast imaging, Panoramic Imaging, Trapezoid Imaging and their combination modes, and the system is intended to be used in a hospital or medical clinic.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K201059

1. Submitter [21 CFR807.92 (a) (1)]

Submitter:SONOSCAPE MEDICAL CORP.
Address:Room 201 & 202, 12th Building, Shenzhen Software Park Phase II,1 Keji Middle 2nd Road, Yuehai Subdistrict, Nanshan District,Shenzhen, 518057, Guangdong, China
Contact Person:Toki Wu
Fax:+86 755 26722850
------------------------

Email: ra@sonoscape.net

Date Prepared July 2, 2020

2. Device [21 CFR807.92 (a) (2)]

Trade Name:S60 Elite Series/S70 Series Digital Color Doppler UltrasoundSystem
Models:S60 Expert, S60 Classic, S60 Nov, S60 Super, S60 Speci, S60Elite, S-Light, S70 Exp, S70, S70 Pro, S70S, S70 VO, S60S, S60N
Common Name:Diagnostic Ultrasound System and Transducers

Classification Regulatory:

CFR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
(Primary)
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX
Classification Panel:Radiology
Device Class:II

3. Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate device within this submission is as follows:

TypeManufacturerDevice510 (K) Number
Primary PredicateSonoScapeS60 Series Digital ColorK172082
DeviceMedical Corp.Doppler Ultrasound System
ReferencePhilipsEPIQ 5/EPIQ 7 DiagnosticK172607

{4}------------------------------------------------

DevicesUltrasound, Inc.Ultrasound System
ReferenceSonoScapeP10 Series Digital Color K173058
DevicesMedical Corp.Doppler Ultrasound System

4. Device Description [21 CFR 807.92(a) (4)]

This SonoScape S60 Elite Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.

This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.

This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler, Power Doppler and Directional Power Doppler Imaging, or the combination of these modes, Contrast lmaging, Elastography, 3D/4D.

Note 1: The S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System is considered as two serial products of S60 Elite Series (including S60 Expert, S60 Classic, S60 Nov, S60 Speci, S60 Speci, S60 Elite, S-Light models) and S70 Series (including S70 Exp, S70, S70 Pro, S70S, S70 VO, S60S, S60N models), but they are the same except partial functions, including the same indications for use, configuration probe, design, hardware, software, mechanic construction, power supply board, main board, specification and etc.

5. Intended Use/Indications for Use [21 CFR 807.92(a) (5)]

5.1 Intended use

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal,

{5}------------------------------------------------

Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

5.2 Indications for use

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

The Modes of Operation include B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Tissue Harmonic Imaging, Power Doppler Imaging, Directional Power Doppler Imaging, Tissues Doppler Imaging, Pulse Inversion Harmonic Imaging, 3D/4D Imaging mode, Elastography Imaging, Contrast imaging, Panoramic Imaging, Trapezoid Imaging and their combination modes, and the system is intended to be used in a hospital or medical clinic.

6. Comparison with the Predicate device [21 CFR 807.92(a) (6)]

S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate device:

TypeManufacturerDevice510 (K) Number
Primary Predicate DeviceSonoScape Medical Corp.S60 Series Digital Color Doppler Ultrasound SystemK172082
Reference DevicesPhilips Ultrasound, Inc.EPIQ 5/EPIQ 7 Diagnostic Ultrasound SystemK172607
Reference DevicesSonoScape Medical Corp.P10 Series Digital Color Doppler Ultrasound SystemK173058

S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System has the same intended uses, complies with the same regulation and safety standards, has the consistent acoustic output levels, has similar probes and has the same technical characteristics as the predicate device legally marketed S60 Series (K172082).

Intended Use Comparison:

Compared with the predicate device S60 Series (K172082), the Subject Device S60 Elite Series/S70 Series has the same intended use.

{6}------------------------------------------------

Subject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60 Super/S60Speci/S60 Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NPrimary Predicate DeviceLegally marketedSonoScape S60 Exp/S60/S60Pro/S60 VO/S60 Maso/S59(K172082)Remark
The Digital Color Doppler Ultrasound is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.Same

Table 2 Intended use Comparison

Technical Characteristics Comparison:

Compared with the predicate device S60 Series (K172082), the Subject Device S60 Elite Series/S70 Series has the similar technical characteristics, including Design, Operation Controls, Display Modes, Operation Modes, Measurement Items, Cine Loop, Operating and Storage Condition. And the differences will not raise new risk and different questions of safety and effectiveness.

Probes Comparison:

Subject device S60 Elite Series/S70 Series has the similar probes as the predicate device SonoScape S60 Series (K172082).

Table 3 a) Probes Comparison

Subject DevicePrimary Predicate DeviceRemark
SonoScape S60 Expert/S60Legally marketed

{7}------------------------------------------------

Classic/S60 Nov/S60 Super/S60SonoScape S60 Exp/S60/S60
Speci/S60 Elite/S-Light/S70 Exp/Pro/S60 VO/S60 Maso/S59
S70/S70 Pro/S70S/S70(K172082)
VO/S60S/S60N
C322 Micro-curved ArrayC322 Micro-curved Array
3C-A Curved Array3C-A Curved Array
C1-6 Curved ArrayC1-6 Curved Array
C1-6A Curved ArrayC1-6A Curved Array
6CT-A Curved Array6CT-A Curved ArraySE
6CI-A Curved Array6CI-A Curved ArrayAnalysis 1
C2-9 Curved ArrayC2-9 Curved Array
C613 Micro-curved ArrayC613 Micro-curved Array
VC2-9 Curved ArrayVC2-9A Curved Array
VC6-2 Micro-curved ArrayVC2-9 Curved Array
6V1 Micro-curved Array6V3 Micro-curved Array
6V3 Micro-curved Array6V3A Micro-curved Array
6V3A Micro-curved Array6V7 Micro-curved Array
6V7 Micro-curved ArrayBCC9-5 Micro-curved Array
BCC9-5 Micro-curved ArrayBCL10-5 Biplane (Micro-curved +SE
BCL10-5 Biplane (Micro-curved +Linear Array)Linear Array)Analysis 1
VE9-5 Micro-curved ArrayVE9-5 Micro-curved Array
C3-10V Micro-curved Array12C-ER Micro-curved Array
C3-10V Micro-curved Array
8P1 Phased Array
4P-A Phased Array8P1 Phased Array
S1-5 Phased Array4P-A Phased ArraySE
7P-A Phased ArrayS1-5 Phased ArrayAnalysis 2
3P-A Phased Array7P-A Phased Array
7P-B Phased Array
1012 Linear Array1012 Linear Array
9L-A Linear Array9L-A Linear Array
12L-A Linear Array12L-A Linear ArraySE
12L-B Linear ArrayVL12-5 Linear ArrayAnalysis 3
13L-A Linear Array12L-B Linear Array
18L-A Linear Array13L-A Linear ArraySEAnalysis 4
12LT-A Linear Array18L-A Linear Array
12LI-A Linear ArrayML3-18 Linear Array
L742 Linear Array12LT-A Linear Array
L741 Linear Array12LI-A Linear Array
L3-9 Linear Array
CWD2.0 CWPWD2.0 Doppler
MPTEE Phased ArrayCWD5.0 CW
MPTEE mini Phased ArrayMPTEE Phased Array
LAP7 Linear ArrayMPTEE mini Phased Array
LAP7 Linear Array

{8}------------------------------------------------

510(k) Submission

SE Analysis 1:

Compared with the predicate device, there are two new Micro-curved Array probes as followed: VC6-2, which is similar with the probe VC2-9 cleared with predicate device SonoScape S60 Series (K172082); and 6V1, which is similar with the probe 6V3 cleared with predicate device SonoScape S60 Series (K172082). All of these new transducers have been cleared in the P10 Series Digital Color Doppler Ultrasound System (K173058) and other series, manufactured by SONOSCAPE MEDICAL CORP.

SE Analysis 2:

Compared with the predicate device, there are two new Phased Array probes as followed: 3P-A, which is similar with the probe 4P-A cleared with predicate device SonoScape S60 Series (K172082). The new probe 7P-B which is similar with the probe 7P-A cleared with predicate device SonoScape S60 Series (K172082). All of these new transducers have been cleared in the P10 Series Digital Color Doppler Ultrasound System (K173058) and other series, manufactured by SONOSCAPE MEDICAL CORP.

SE Analysis 3:

Compared with the predicate device, there are two new Linear Array probes as followed: L742 and L3-9, which is similar with the probe 9L-A cleared with predicate device SonoScape S60 Series (K172082). The new transducers L742 have been cleared in the P10 Series Digital Color Doppler Ultrasound System (K173058) and other series, manufactured by SONOSCAPE MEDICAL CORP., and L3-9 is also legally marketed in S60 Series Digital Color Doppler Ultrasound System (K172082) and other series, manufactured by SONOSCAPE MEDICAL CORP.

{9}------------------------------------------------

SE Analysis 4:

Compared with the predicate device, there is a new CW probes as followed: CWD2.0, which is similar with the probe CWD5.0 cleared with predicate device SonoScape S60 Series (K172082). The new transducers CWD2.0 have been cleared in the P10 Series Digital Color Doppler Ultrasound System (K173058) and other series, manufactured by SONOSCAPE MEDICAL CORP.

Therefore, they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

Functional Comparison of probes

Compared with the predicate device SonoScape S60 Series (K172082), there is new function of contrast imaging in probes (3C-A, C1-6A, C1-6, C2-9, C322, 6CT-A, 6CI-A, 12L-A, 12L-B, 9L-A, L3-9, L741, L742, 10I2, 12LT-A, 12LI-A, 4P-A, 3P-A, S1-5, VE9-5, 6V1, 6V3, 6V7), which is similar with the probes (C6-2, L12-5, S5-1, C10-3v and C10-4ec) cleared with reference device EPIQ 5/EPIQ 7 Diagnostic Ultrasound System (K172607). The clinical application is the same, the performance or frequency is similar, and the difference of these doesn't affect the safety, effectiveness and clinical use.

The engineering drawings of the new function transducers(probes) (3C-A, C1-6, C2-9, C322, 6CT-A, 6CI-A, 12L-A, 12L-B, 9L-A, L3-9, L741, L742, 1012, 12LT-A, 12LI-A, 4P-A, 3P-A, S1-5, VE9-5, 6V1, 6V3, 6V7), and the further comparison are provided as followed.

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 13C-AC6-2/
PhotoImage: 3C-AImage: C6-2/

Table 3 b) Further Comparison for 3C-A Probe

{10}------------------------------------------------

510(k) Submission

ProbeTypeCurved ArrayCurved ArraySame
Frequency1.0-7.0MHz2.0-6.0MHzSEAnalysis 5
Indicationfor useFetal,Other(Ob/GYN),Abdominal,General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompoundImaging,ContrastImaging,ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4DImaging,XRES, ElastographySEAnalysis 7

Table 3 c) Further Comparison for C1-6A Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NPrimary Predicate DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 2C1-6AC6-2/
PhotoImage: C1-6A ProbeImage: C6-2 Probe/
ProbeTypeCurved ArrayCurved ArraySame
Frequency1.0-8.0MHz2.0-6.0MHzSEAnalysis 5

{11}------------------------------------------------

510(k) Submission

Indicationfor useFetal,Other(Ob/GYN)Abdominal,General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI, PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging,ContrastElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 d) Further Comparison for C1-6 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 3C1-6C6-2/
PhotoImage: SonoScape S60 Expert/S60 ProbeImage: EPIQ 5/EPIQ 7 Diagnostic Ultrasound System Probe/
ProbeTypeCurved ArrayCurved ArraySame
Frequency1.0-8.0MHz2.0-6.0MHzSEAnalysis 5
Indicationfor useFetal, Abdominal, Other(Ob/GYN)General purpose abdominal (adult and pediatric, including vascular), bowel, obstetrical, gynecological, prostate andSame

{12}------------------------------------------------

510(k) Submission

interventional applications
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging,Contrast Imaging,ElastographyCompound Imaging,CPA,harmonic Imaging, Contrast Imaging,3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 e) Further Comparison for C2-9 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 4C2-9C6-2/
PhotoImage: Ultrasound probe C2-9Image: Ultrasound probe C6-2/
ProbeTypeCurved ArrayCurved ArraySame
Frequency2.0-13.0MHz2.0-6.0MHzSEAnalysis 5
Indicationfor useFetal, Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6

{13}------------------------------------------------

510(k) Submission

AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging,Contrast ImagingCompound Imaging, CPA,harmonic Imaging, Contrast Imaging,3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 f) Further Comparison for C322 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 5C322C6-2/
PhotoImage: SonoScape S60 Expert/S60 ProbeImage: EPIQ 5/EPIQ 7 Diagnostic Ultrasound System Probe/
ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency2.0-7.0MHz2.0-6.0MHzSEAnalysis 5
Indicationfor useFetal,Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame

{14}------------------------------------------------

FunctionsCompound Imaging, Contrast ImagingCompound Imaging, CPA,THI, harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySE Analysis 7
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Table 3 g) Further Comparison for 6CT-A Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 66CT-AC6-2/
PhotoImage: 6CT-A ProbeImage: C6-2 Probe/
ProbeTypeCurved ArrayCurved ArraySame
Frequency3.0-15.0MHz2.0-6.0MHzSEAnalysis 5
Indicationfor useFetal,Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging,THI,Contrast ImagingCompound Imaging,CPA,harmonic Imaging,ContrastImaging,3D/4DImaging,SEAnalysis 7

{15}------------------------------------------------

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 76CI-AC6-2/
PhotoImage: SonoScape S60 Expert/S60 ProbeImage: EPIQ 5/EPIQ 7 Diagnostic Probe/
ProbeTypeCurved ArrayCurved ArraySame
Frequency3.0-15.0MHz2.0-6.0MHzSEAnalysis 5
Indicationfor useFetal, Abdominal,Other(Ob/GYN)General purpose abdominal(adult and pediatric, includingvascular), bowel, obstetrical,gynecological, prostate andinterventional applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, THI,Contrast ImagingCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 h) Further Comparison for 6CI-A Probe

Table 3 i) Further Comparison for 12L-A Probe

ComparisSubject DeviceReference DeviceRemark
----------------------------------------------------

{16}------------------------------------------------

on ItemsSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)
Probe 812L-AL12-5 50/
PhotoImage: 12L-AImage: L12-5 50/
ProbeTypeLinear ArrayLinear ArraySame
Frequency3.0-17.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid, testes), Musculo-skeletal(Conventional & Superficial), Peripheral vesselBreast, thyroid and superficial small parts; musculoskeletaltendon, abdomen bowel, and vascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2 maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2 maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 j) Further Comparison for 12L-B Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70Reference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{17}------------------------------------------------

510(k) Submission

VO/S60S/S60N
Probe 912L-BL12-5 50/
PhotoImage: 12L-B probeImage: L12-5 50 probe/
ProbeTypeLinear ArrayLinear ArraySame
Frequency3.0-17.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes), Musculo-skeletal(Conventional & Superficial),Peripheral vesselBreast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andvascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 k) Further Comparison for 9L-A Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 109L-AL12-5 50/
PhotoImage: 9L-A probeImage: L12-5 50 probe/

510(k) Summary

page 15 of 29

{18}------------------------------------------------

510(k) Submission

ProbeTypeLinear ArrayLinear ArraySame
Frequency2.0-13.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes), Musculo-skeletal(Conventional & Superficial),Peripheral vesselBreast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andvascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging,3D/4DImaging,XRES, ElastographySEAnalysis 7

Table 3 I) Further Comparison for L3-9 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 11L3-9L12-5 50/
PhotoImage: Ultrasound probeImage: Ultrasound probe/
ProbeTypeLinear ArrayLinear ArraySame
Frequency2.0-13.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes),Musculo-skeletal(Conventional & Superficial),Breast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andSame

{19}------------------------------------------------

Peripheral vesselvascular applications
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 m) Further Comparison for L741 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 12L741L12-5 50/
PhotoImage: Ultrasound probe L741Image: Ultrasound probe L12-5 50/
ProbeTypeLinear ArrayLinear ArraySame
Frequency4.0-16.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes),Musculo-skeletal(Conventional & Superficial),Peripheral vesselBreast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andvascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame

{20}------------------------------------------------

FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Table 3 n) Further Comparison for L742 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 13L742L12-5 50/
PhotoImage: Ultrasound probeImage: Ultrasound probe/
ProbeTypeLinear ArrayLinear ArraySame
Frequency4.0-16.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes), Musculo-skeletal(Conventional & Superficial),Peripheral vesselBreast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andvascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA, harmonic Imaging, Contrast Imaging, 3D/4D Imaging, XRES, ElastographySEAnalysis 7

Table 3 o) Further Comparison for 10I2 Probe

{21}------------------------------------------------

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 141012L12-5 50/
PhotoImage: SonoScape S60 Expert/S60 ProbeImage: EPIQ 5/EPIQ 7 Diagnostic Ultrasound System Probe/
ProbeTypeLinear ArrayLinear ArraySame
Frequency4.0-16.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid, testes), Musculo-skeletal (Conventional & Superficial), Peripheral vesselBreast, thyroid and superficial small parts; musculoskeletal tendon, abdomen bowel, and vascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImagingCompound Imaging, CPA, harmonic Imaging, Contrast Imaging, 3D/4D Imaging, XRES ElastographySEAnalysis 7

Table 3 p) Further Comparison for 12LT-A Probe

Comparison ItemsSubject DeviceReference DeviceRemark
SonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/Legally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)

{22}------------------------------------------------

510(k) Submission

S70/S70 Pro/S70S/S70VO/S60S/S60N
Probe 1512LT-AL12-5 50/
PhotoImage: 12LT-A ProbeImage: L12-5 50 Probe/
ProbeTypeLinear ArrayLinear ArraySame
Frequency4.0-16.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes),Musculo-skeletal(Conventional & Superficial),Peripheral vesselBreast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andvascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImagingCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRES, ElastographySEAnalysis 7

Table 3 q) Further Comparison for 12Ll-A Probe

Comparison ItemsSubject DeviceReference DeviceRemark
SonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)
Probe 1612LI-AL12-5 50/

{23}------------------------------------------------

510(k) Submission

PhotoImage: Ultrasound probeImage: Ultrasound probe/
ProbeTypeLinear ArrayLinear ArraySame
Frequency4.0-16.0MHz5.0-12.0MHzSEAnalysis 5
Indicationfor useSmall Organ (breast, thyroid,testes), Musculo-skeletal(Conventional & Superficial),Peripheral vesselBreast, thyroid and superficialsmall parts; musculoskeletaltendon, abdomen bowel, andvascular applicationsSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImagingCompound Imaging,CPA,harmonic Imaging,ContrastImaging,3D/4DImaging,XRES, ElastographySEAnalysis 7

Table 3 r) Further Comparison for 4P-A Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 174P-AS5-1/
PhotoImage: 4P-A ProbeImage: S5-1 Probe/
ProbeTypePhased ArrayPhased ArraySame

510(k) Summary

page 21 of 29

{24}------------------------------------------------

510(k) Submission

Frequency1.0-6.0MHz1.0-5.0MHzSEAnalysis 5
Indicationfor useAbdominal, Cephalic, CardiacAbdominal, Cephalic, CardiacSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImagingCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRESSEAnalysis 7

Table 3 s) Further Comparison for 3P-A Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 183P-AS5-1/
PhotoImage: 3P-A probeImage: S5-1 probe/
ProbeTypePhased ArrayPhased ArraySame
Frequency1.0-6.0MHz1.0-5.0MHzSEAnalysis 5
Indicationfor useAbdominal, Cephalic, CardiacAbdominal, Cephalic, CardiacSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumSame

510(k) Summary

page 22 of 29

{25}------------------------------------------------

MI: 1.9 maximumMI: 1.9 maximum
FunctionsCompound Imaging, Contrast ImagingCompound Imaging, CPA, harmonic Imaging, Contrast Imaging, 3D/4D Imaging, XRESSEAnalysis 7
Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 19S1-5S5-1/
PhotoImage: SonoScape S60 Expert/S60 S1-5Image: EPIQ 5/EPIQ 7 Diagnostic S5-1/
ProbeTypePhased ArrayPhased ArraySame
Frequency1.0-7.0MHz1.0-5.0MHzSEAnalysis 5
Indicationfor useAbdominal, Cephalic, CardiacAbdominal, Cephalic, CardiacSame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImagingCompound Imaging, CPA,harmonic Imaging,ContrastImaging, 3D/4DImaging,XRESSEAnalysis 7

Table 3 t) Further Comparison for S1-5 Probe

Table 3 u) Further Comparison for VE9-5 Probe

{26}------------------------------------------------

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 20VE9-5C10-3v/
PhotoImage: VE9-5Image: C10-3v/
ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency2.0-13.0MHz3.0-10.0MHzSEAnalysis 5
Indicationfor useTrans-vaginalFetal/OB,Trans-vaginal,Other: GYN/UrologySame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, 3D/4DCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRESSEAnalysis 7

Table 3 v) Further Comparison for 6V1 Probe

Comparison ItemsSubject DeviceReference DeviceRemark
SonoScape S60 Expert/S60Legally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)
Classic/S60 Nov/S60Remark
Super/S60 Speci/S60
Elite/S-Light/S70 Exp/

{27}------------------------------------------------

510(k) Submission

S70/S70 Pro/S70S/S70VO/S60S/S60N
Probe 216V1C10-4ec/
PhotoImage: 6V1 ProbeImage: C10-4ec Probe/
ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency3.0-15.0MHz3.0-10.0MHzSEAnalysis 5
Indicationfor useTrans-rectal, Trans-vaginalFetal/OB,Trans-rectal,Trans-vaginal,Peripheralvessel, Other: GYN/UrologySame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImagingCompound Imaging,CPA,harmonic Imaging,ContrastImaging,3D/4DImaging,XRESSEAnalysis 7

Table 3 w) Further Comparison for 6V3 Probe

Subject DeviceReference Device
Comparison ItemsSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 226V3C10-4ec/

{28}------------------------------------------------

Photo
ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency3.0-15.0MHz3.0-10.0MHzSEAnalysis 5
Indicationfor useTrans-rectal, Trans-vaginalFetal/OB,Trans-rectal,Trans-vaginal,Peripheralvessel, Other: GYN/UrologySame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging, CPA,harmonic Imaging, ContrastImaging, 3D/4D Imaging,XRESSEAnalysis 7

Table 3 x) Further Comparison for 6V7 Probe

Comparison ItemsSubject DeviceSonoScape S60 Expert/S60Classic/S60 Nov/S60Super/S60 Speci/S60Elite/S-Light/S70 Exp/S70/S70 Pro/S70S/S70VO/S60S/S60NReference DeviceLegally marketedEPIQ 5/EPIQ 7 DiagnosticUltrasound System(K172607)Remark
Probe 236V7C10-4ec/
PhotoImage: Ultrasound probeImage: Ultrasound probe/

{29}------------------------------------------------

ProbeTypeMicro-curved ArrayCurved ArraySame
Frequency3.0-15.0MHz3.0-10.0MHzSEAnalysis 5
Indicationfor useTrans-rectal, Trans-vaginalFetal/OB,Trans-rectal,Trans-vaginal, Peripheralvessel, Other: GYN/UrologySame
OperationModeB,THI,M,CFM,PDI,DPDI,PWB,THI,M,CFM,PDI,PWSEAnalysis 6
AcousticOutputLimitsDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumDerated ISPTA: 720mW/cm2maximumTIS/TIB/TIC: 6.0 maximumMI: 1.9 maximumSame
FunctionsCompound Imaging, ContrastImaging, ElastographyCompound Imaging,CPA,harmonic Imaging, ContrastImaging, 3D/4DImaging,XRESSEAnalysis 7

SE Analysis 5:

Although the frequency is different, all of them comply with the requirements of IEC 60601-1 & IEC 60601-1-2 & IEC 60601-2-37 and meet clinical requirements, and no new risk is raised.

SE Analysis 6:

Although the operation mode is different, the new operation mode of DPDI (Directional Power Doppler Imaging) is similar to the PDI mode sequence with direction, and no new risk is raised.

SE Analysis 7:

Although the functions is different, the functions of the subject device are included in the predicated device, and no new risk is raised.

Moreover, compared with predicate device, the subject device (S60 Elite Series/S70 Series) complies with the same regulation and safety standards and has the consistent acoustic output levels.

Summary of the comparison

Compared with the predicate device legally marketed SonoScape S60 Series (K172082), Philips EPIQ 5 and EPIQ 7 Diagnostic Ultrasound Systems (K172607) and SonoScape P10 Series (K173058), the subjective device (S60 Elite Series/S70 Series

{30}------------------------------------------------

Digital Color Doppler Ultrasound System) are all analyzed with the predicate device. The comparison showed that they can be considered Substantially Equivalent in safety and effectiveness. Therefore, there is no new risk raised, and the SE is not affected.

7. Non-Clinical Tests [21 CFR 807.92(b) (1)]

Non-clinical testing to assure compliance with electrical, mechanical, thermal and electromagnetic compatibility safety, acoustic output and biocompatibility were performed and have been found to conform to applicable standards. The S60 Elite Series/S70 Series system has been designed and manufactured to meet the following standards:

IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance [08/20/2012];

IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2 General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests [2014-02];

IEC 60601-2-37:2015 Medical Electrical Equipment-Part 2-37: Particular requirements for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment [Edition 2.1. 2015]:

ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro cytotoxicity [06/01/2009];

ISO 10993-10:2010, Biological Evaluation of Medical Devices- Part 10: Tests for irritation and skin sensitization [08/01/2010];

AIUM/NEMA UD 2:2004 (R2009), Acoustic output measurement standard for diagnostic ultrasound equipment [08/21/2009].

The Digital Color Doppler Ultrasound System is verified through the relevant

Performance testTesting protocols andfail/acceptance criteriaTesting results
Electrical safety testingIEC 60601-1:2005+A1:2012Passed
EMC testingIEC 60601-1-2:2014Passed
Acoustic testingIEC 60601-2-37:2015AIUM/NEMA UD 2:2004 (R2009)Passed
Software Verification andValidationIEC 62304:2006 +A 1:2015Passed

summarized information as followed:

{31}------------------------------------------------

Laboratory tests (including Phantom tests) were conducted to verify that the S60 Elite Series/S70 Series system met all design specifications and the S60 Elite Series/S70 Series system conformed to applicable medical device standards.

8. Clinical Test [21 CFR 807.92(b) (2)]

No clinical testing was required.

9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]

In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SONOSCAPE MEDICAL CORP. concludes that S60 Elite Series/S70 Series Digital Color Doppler Ultrasound System is substantially equivalent to the predicate device with regard to safety and effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.