K Number
K200968

Validate with FDA (Live)

Date Cleared
2020-12-28

(262 days)

Product Code
Regulation Number
868.5740
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for use in thoracic surgery, bronchospirometry, administration of Endobronchial anesthesia and other uses commonly requiring Endobronchial intubation. The Endobronchial tube is indicated for main stem bronchus intubation and allows for selective inflation or deflation of either lung.

Device Description

ANKOR Endobronchial tube is consisted with double lumen tube with the body including 3 cuff (Bronchus Cuff, Trachea Cuff and ANKOR Cuff) / 2 cuff (Bronchus Cuff and Trachea Cuff), the stylet and elbow connector in order to separate connector for isolating and ventilating one lung during surgical procedures. The different point, between ABC Model and HBC Model, is whether ANKOR Cuff is consisted. ANKOR Cuff, which is a part of the ABC Model only, takes to check correct insertion of the catheter. When user is lack of the experience or is not confidence with operation, ABC Model is intended to use. When the suction is necessary, this device can be possible to be assemble with suction catheter which is 10Fr or 12Fr with PVC (Listing Number: D422226) and this device is packaged with these suction catheter (Listing Number: D422226). The device is made of silicone as a main material and is available in 33Fr to 39Fr depending on the patients who is requiring one-lung isolation under OR and ICU including Non-MR Environment only.

AI/ML Overview

The provided text describes the 510(k) summary for the ANKOR Endobronchial Tube, which is a medical device for tracheal/bronchial differential ventilation. It aims to demonstrate substantial equivalence to a predicate device, not necessarily to prove the device meets specific acceptance criteria for AI performance. As such, the document does not contain information related to AI or a study that proves a device meets acceptance criteria for an AI-powered system.

Therefore, I cannot provide the requested information about acceptance criteria and study data for an AI-powered device based on the provided text. The document focuses on performance testing of the physical medical device and its biocompatibility, sterilization, and material properties, rather than AI performance.

Specifically, the document explicitly states in Section 8:

"Clinical and animal testing were not performed for ANKOR Endobronchial Tube as part of the premarket notification requirements for this 510(k) submission and the subject of this premarket submission, ANKOR Endobronchial Tube, did not require clinical and animal studies to support substantial equivalence."

This reinforces that the approval is for a traditional medical device, not an AI-enabled one, and thus the specific metrics and study designs requested for AI performance are not relevant to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."

December 28, 2020

Insung Medical Co., Ltd. % Dongha Lee Regulatory Affairs Consultant KMC.Inc. Room no. 904, 27, Digital-ro 27ga-gil, Guro-gu Seoul. 08375 Korea, Republic Of

Re: K200968

Trade/Device Name: ANKOR Endobronchial Tube Regulation Number: 21 CFR 868.5740 Regulation Name: Tracheal/Bronchial Differential Ventilation Tube Regulatory Class: Class II Product Code: CBI Dated: November 23, 2020 Received: November 27, 2020

Dear Dongha Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200968

Device Name ANKOR endobronchial tube

Indications for Use (Describe)

This device is intended for use in thoracic surgery, bronchospirometry, administration of Endobronchial anesthesia and other uses commonly requiring Endobronchial intubation.

The Endobronchial tube is indicated for main stem bronchus intubation and allows for selective inflation of either lung.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features green leaves above the word "INSUNG" in a bold, green font. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller, less bold font.

510(k) SUMMARY

This summary of 510(k) –safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: Nov. 23, 2020

1. INFORMATION

1.1 Submitter Information

  • Submitter Name: INSUNG MEDICAL CO., LTD.
  • . Address

: 168, GIEOPDOSI-RO, JIJEONG-MYEON, WONJU-SI, GANGWON-DO, 26354, KOREA

. Telephone Number: +82-70-5077-0118

1.2 Contact Person

  • . Name: Milly (Consultant / KMC, Inc.)
  • . Address: Room no. 1709, 123, Digital-ro 26-gil, Guro-gu, Seoul, 08390, Korea
  • Telephone Number: +82-70-8965-5554 . Fax: +82-2-2672-0579
  • . E-mail: milly@kmcerti.com

2. DEVICE INFORMATION

■ Trade Name / Proprietary Name: ANKOR endobronchial tube
■ Model: ABC33L, ABC35L, ABC37L, ABC39L,ABC33R, ABC35R, ABC37R, ABC39RHBC33L, HBC35L, HBC37L, HBC39L,HBC33R, HBC35R, HBC37R, HBC39R
■ Common Name: Endobronchial Double lumen tube
■ Classification Name: Tube, Tracheal/Bronchial,Differential Ventilation (W/Wo Connector)
■ Product Code: CBI
■ Classification Regulation: 21CFR 868.5740
■ Device Class: Class II
■ Classification Panel: Anesthesiology

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image contains a logo for INSUNG MEDICAL CO., LTD. The logo features the word "INSUNG" in large, bold, green letters. Above the word "INSUNG" are three green leaves. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller letters.

3. PREDICATE DEVICE

Predicate Device
ManufacturerTeleflex Medical
Device Name (Trade Name)Sheridan Endobronchial Tubes
510(k) NumberK180253

4. SUBJECT DEVICE DESCRIPTION

ANKOR Endobronchial tube is consisted with double lumen tube with the body including 3 cuff (Bronchus Cuff, Trachea Cuff and ANKOR Cuff) / 2 cuff (Bronchus Cuff and Trachea Cuff), the stylet and elbow connector in order to separate connector for isolating and ventilating one lung during surgical procedures. The different point, between ABC Model and HBC Model, is whether ANKOR Cuff is consisted.

ANKOR Cuff, which is a part of the ABC Model only, takes to check correct insertion of the catheter.

When user is lack of the experience or is not confidence with operation, ABC Model is intended to use.

When the suction is necessary, this device can be possible to be assemble with suction catheter which is 10Fr or 12Fr with PVC (Listing Number: D422226) and this device is packaged with these suction catheter (Listing Number: D422226).

The device is made of silicone as a main material and is available in 33Fr to 39Fr depending on the patients who is requiring one-lung isolation under OR and ICU including Non-MR Environment only.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features three green leaves at the top left, followed by the word "INSUNG" in large, bold, green letters. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller, thinner letters.

5. INTENDED USE

This device is intended for use in thoracic surgery, bronchospirometry, administration of Endobronchial anesthesia and other uses commonly requiring Endobronchial intubation. The Endobronchial tube is indicated for main stem bronchus intubation and allows for selective inflation or deflation of either lung.

6. SUBSTANTIAL EQUIVALENCE

ANKOR endobronchial tube is substantially equivalent to the predicate device (Endobronchial Tube; K180253). The following table is presented to demonstrate substantial equivalence.

Subject DevicePredicate Device(K180253)S/D
ClassificationNameTube, Tracheal/Bronchial,Differential Ventilation(W/Wo Connector)Tube, Tracheal/Bronchial,Differential Ventilation(W/Wo Connector)S
Product CodeCBICBIS
RegulationNumber21CFR 868.574021CFR 868.5740S
Indications for UseThis device is intended for use inthoracic surgery,bronchospirometry,administration of Endobronchialanesthesia and other usescommonly requiringEndobronchial intubation.The Endobronchial tube isindicated for main stem bronchusintubation and allows for selectiveinflation or deflation of eitherlung.This device is intended for use inthoracic surgery,bronchospirometry,administration of Endobronchialanesthesia and other usescommonly requiringEndobronchial intubation.The Endobronchial tube isindicated for main stem bronchusintubation and allows for selectiveinflation or deflation of eitherlung.S
Environments ofUseHospital - OR and ICUHospital - OR and ICUS
Direction for UsePrescription UsePrescription UseS
Intended UserTrained PhysiciansTrained PhysiciansS
Patient PopulationPatients requiring one-lungisolationPatients requiring one-lungisolationS

S: Same, D: Different

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features the word "INSUNG" in large, bold, green letters. Above the word "INSUNG" are three green leaves. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller letters.

Design FeaturesLumen Shaft, Cuffs, StyletLumen Shaft, Cuffs, StyletS
Sizes (Fr; French)33 to 39 French28 to 41 FrenchD
CuffedYesYesS
RadiopaqueYesYesS
Connection to ventilation source15mm Connector15mm ConnectorS
Single useSingle useSingle useS
Sterilization (Method)Yes(EO Gas)Yes(EO Gas)S
EO and ECH ResidualThe ethylene oxide residual is conform to ISO 10993-7The ethylene oxide residual is conform to ISO 10993-7S
PackageSterile, packed with Elbow connector and suction catheterSterile, packed with connectorD
BiocompatibilityCytotoxicity Test(ISO 10993-5)Sensitization Test(ISO 10993-10)Irritation Test(ISO 10993-10)Particulate Matter Test(ISO 18562-2:2017)Volatile Organic Compound Extraction Test(ISO 18562-3:2017)Cytotoxicity Test(ISO 10993-5)Sensitization Test(ISO 10993-10)Irritation Test(ISO 10993-10)Particulate Matter Test(ISO 18562-2:2017)Volatile Organic Compound Extraction Test(ISO 18562-3:2017)S
DLT MaterialsSiliconPVCD

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features a stylized green leaf design above the word "INSUNG" in large, bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller font.

1 ) Same points between the subject device and the predicate device

  • . Classification Name
    : The proposed classification name of the subject device is "Tracheal/bronchial differential ventilation tube". It is the same classification name both subject device (ANKOR Endobronchial Tube) and the predicate device (K180253).

  • Product Code ●
    : The proposed product code of the subject device is "CBI". It is the same product code both subject device (ANKOR Endobronchial Tube) and the predicate device (K180253).

  • . Regulation Number
    : The regulation number, which is applied to the subject device, is "21CFR868.5740". It is the same regulation number both subject device (ANKOR Endobronchial Tube) and the predicate device (K180253).

  • Indications for Use .
    : The indications for use is "This device is intended for use in thoracic surgery, bronchospirometry, administration of Endobronchial anesthesia and other uses commonly requiring Endobronchial intubation. The Endobronchial tube is indicated for main stem bronchus intubation and allows for selective inflation or deflation of either lung." under both devices (Subject device and predicate device). It is same point.

  • . Environment of Use
    : For environment of use, both devices (subject device and predicate device) is used at hospital in order to safely use the both. It is same point.

  • . Direction for use
    : Both devices are prescription used device. It is same point.

  • . Intended User
    : For intended user, trained physicians shall use the both device in order to safely use the both. It is same fact.

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features a stylized green leaf design above the word "INSUNG" in bold, green letters. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller font size.

  • . Patient Population
    : Bothe device is used for patients requiring one-lung isolation as the patient population in thoracic surgery, bronchospirometry, administration of Endobronchial anesthesia and other uses commonly requiring Endobronchial intubation. It is same point.

  • . Design Features
    : Under the main body of both devices, the design features are consisted with lumen Shaft, Cuff and Stylet. It is same point.

  • Single use
    : Both devices (subject device and predicate device) is single used device which is E.O. Gas Sterilize. It is same point.

  • . Cuffed
    : Both devices have cuffs such as bronchial cuff and tracheal cuff to block the bronchial or tracheal for isolating and ventilating one lung during surgical procedures. It is same point.

  • . Radiopaque line
    : Under the tube (catheter) of both devices, the tubes contain an x-ray opaque line that runs through the tube making them detectable by x-ray. It is same point.

  • . Connection to ventilation source
    : 15mm Connector is used under both devices (subject device and predicate device). The connector could be possible to assemble with ventilation source. It is same point.

  • Single use
    : Both devices (subject device and predicate device) is single used device which is E.O. Gas Sterilize. It is same point.

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features a green and teal leaf design above the word "INSUNG" in large, bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller, less bold font.

  • . Sterilization (Method)
    : Both devices (subject device and predicate device) are single used devices which are E.O. Gas Sterilize. It is same point.

  • . EO and ECH Residual
    : EO Gas Residual of Both devices (subject device and predicate device) are confirmed in accordance with ISO 10993-7. It is same point.

  • . Biocompatibility
    : General Biocompatibility Tests of both devices (subject device and predicate device) are selected in accordance with ISO 10993-1. The same tests, such as following, are conducted.

Test ItemStandard
Cytotoxicity TestISO 10993-5:2009
Sensitization TestISO 10993-10:2010
Irritation TestISO 10993-10:2010

In addition, both device could be possible to be connected with breathing gas supplier. In accordance with "Biocompatibility of breathing gas pathway (ISO 18562-1:2017)" which considers risk related to breathing gas pathway, Particulate Matter Test (Per ISO 18562-2:2017) including CO, CO2 and O3 (Ozone) and Volatile Organic Compound Extraction Test (Per ISO 18562-3:2017) are conducted additionally to assure biocompatibility of subject device.

It leads the subject device is more safety than predicated device that is not assured on the "Biocompatibility of breathing gas pathway".

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features a stylized green leaf design above the word "INSUNG" in large, bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller font, also in green.

  1. Different points between the subject device and the predicate device
  • . Size
    : Although size of the subject device is different with size of predicate device, the subject device is verified by test such as appearance test, dimension test.

And, the subject device is placed on the CE market and Korean Market since 2013 years. As a result, any problems, which influence safety and performance of the subject device, are not founded and received through the customer feedback.

In addition, the size of predicated device includes size of the subject device.

We cannot also found any risks related to the subject device, which has same size with subject device. It leads to safety device and effectiveness size to achieve same indication for use.

Refer to Performance Test Report (IS-20-1026-2) on the Attachment 6

  • Package .
    : Packaging composition is different. It is depending on the manufacturer.

In addition, elbow connector of subject device complies with the ISO 5356-1:2014 and suction catheter is complies with the ISO 8836:2014 to ensure the safety and performance. The result of the test is passed and it leads to ensure the safety and performance.

As a result, there are not any problems, which influence safety of the subject device as well as influence substantial equivalence between the subject device and predicate device.

Refer to Performance Test Report (IS-20-0710) on the Attachment 6

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image contains the logo for INSUNG MEDICAL CO., LTD. The logo features a stylized green leaf design above the word "INSUNG" in bold, dark green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller, lighter font.

  • Materials .
    : As a different point on the material between the subject device and predicate device, there is DLT Materials such as Silicon (Subject device) and PVC (Predicate Device)

Although the subject device is different with predicate device, the subject device is verified by test such as a biocompatibility test according to ISO 10993-1, ISO 18562 and FDA Guidance ("Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process"). And, the endotoxin test is conducted in accordance with USP 39 <85> to ensure safety of the materials.

As a result, there are not any problems, which influence safety of the subject device as well as influence substantial equivalence between the subject device and predicate device.

TestReport No.AttachmentNo.
Bacterial Endotoxin Test20-048384-01-11
Biocompatibility Test(Per ISO 10993-1)MTK-0126521
MGK-201900654
MGK-2019-000089
MGK-2019-000088
Biocompatibility Evaluation Report(Per ISO 18562-1:2017)1001018172-3248567BA2
ISO 18562-2:2017 Test
ISO 18562-3:2017 Test1001018172-32485672

Refer to Biocompatibility Test with on the Attachment 1 and Attachment 2

3) Comparison to predicate device

The subject device is substantially equivalent to the predicate device with respect to indications for use, technology and construction. The differences between the predicate and the subject device is minor and any risks have been mitigated through testing.

{12}------------------------------------------------

Image /page/12/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features three green leaves above the word "INSUNG" in large, bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller, less bold letters.

7. PERFORMANCE DATA

All testing that is required by the required standards has been performed. Non-clinical testing was performed and included standards such as ISO 16628, ISO 5356-1, ISO 5361, ISO 80369-7, ISO 8836. ISO 18562 and ISO 10993-1. The Endobronchial Tube have been found to fall within the required limits of the testing. The test results can be found in both the Biocompatibility and the Performance Testing of this submission. Therefore we have concluded that the subject device is substantially equivalent. A brief summary of tests relied upon to demonstrate substantial equivalence to the predicate can be found in the below:

  • 7.1 Biocompatibility
  1. General Biocompatibility (According to ISO 10993-1)

The biocompatibility tests were performed to protect patients from undue risks arise from biological hazards associated with materials of manufacture and final device. The tests were performed in accordance with the following standards and FDA Guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

No.Test ItemsStandardsCriteriaPass/Fail
1CytotoxicityISO 10993-5:2009Negative Control: reactivity of nonePositive Control: greater than mildPass
2IntracutaneousReactivityISO 10993-10:2010No observe skin responses such asthe erythema, edema and etc.Pass
3SensitizationISO 10993-10:2010No induced any skin reaction in theHartley guinea pigPass
Refer to Biocompatibility Test with on the Attachment 1
-----------------------------------------------------------
TestReport No.AttachmentNo.
Biocompatibility Test(Per ISO 10993-1)MTK-012652
MGK-2019006541
MGK-2019-000089
MGK-2019-000088

{13}------------------------------------------------

Image /page/13/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features green leaves above the word "INSUNG" in large, bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller, less bold letters.

  1. Biocompatibility of Breathing Gas Pathways (ISO 18562-1)

Biocompatibility evaluation of breathing gas pathways is performed in accordance with ISO 18562-1:2017 including ISO 10993-17:2002 to identify which compounds emitted into the airstream of the Endobronchial Double Lumen Tube (ANKOR Endobronchial Tube) are considered chemicals of concern (COCs) and require additional risk review.

No.Test ItemsStandardsCriteriaPass/Fail
1Emission Test ofParticulate Matter(Including CO2, COand Ozone)ISO 18562-2:2017No increase the source gasconcentration of particulates ≤2.5 $\mu$ maerodynamic diameter (PM2.5) morethan 12 µg/m3Pass
No increase the source gas of PM ≤ 10$\mu$ m diameter size (PM10) more than150 µg/m3Pass
2Volatile OrganicCompounds(VOCs) ExtractionTestISO 18562-3:2017Unlikely to result in toxicologicaleffectsPass

Refer to Biocompatibility Test with on the Attachment 2

TestReport No.AttachmentNo.
BiocompatibilityEvaluation Report(Per ISO 18562-1:2017)1001018172-3248567BA2
ISO 18562-2:2017 Test1001018172-3248568P2
ISO 18562-3:2017 Test1001018172-32485672

{14}------------------------------------------------

Image /page/14/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features the word "INSUNG" in large, bold, green letters. Above the word "INSUNG" are three green leaves. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller, thinner letters.

7.2 Risk Management

All risks, related to biocompatibility hazard, function hazard and etc., including device problem founded by TPLC (Total Product Life Cycle),, were identified, evaluated, and controlled in accordance with ISO 14971:2007.

No.ItemsStandardsAttachment No.(Report No.)
1Risk ManagementISO 14971:2007Attachment 4(IM-RMR-09)(IS-RMR-202)

7.3 Sterilization Test

The sterilization test was performed to assess inactivation of the microbiological contaminants, and thereby the device the device is verified as a non-sterile medical device into sterile ones. The tests were performed in accordance with following standards.

No.Test ItemsStandardsAttachment No.(Report No.)
1Microbiological qualityISO 11737-1:2018Attachment 10(IS-SVP-1901)(IS-SVR-02)
2Sterility TestISO 11737-2:2009Attachment 10(IS-SVP-1901)(IS-SVR-02)
3E.O. Gas Sterilization ResidualsISO 10993-7:2008Attachment 10(IS-SVP-1901)(IS-SVR-02)

7.4 Cleanroom Validation

Clean rooms are managed by being based on our work environment management regulations. Clean rooms validation is assured through the verification of clean room validation in accordance with verification method for subject device's safety an hygienically.

No.Test ItemsStandardsAttachment No.(Report No.)
1Particle TestISO 14644-1:2015Attachment 11(IS-CVP-01)(IS-CVR1-01)

{15}------------------------------------------------

Image /page/15/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features the word "INSUNG" in large, bold, green letters. Above the word "INSUNG" are three green leaves. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller, thinner letters.

7.5 Cleaning Validation

Contaminants and residues during the cleaning process were maintained continuously and consistently with acceptable numerical values through the appearance test according to manufacturer SOP and microbial test according to ISO 11737-1:2018

No.Test ItemsStandardsAttachment No.(Report No.)
1Appearance TestManufacturer SOPAttachment 11(IS-WVR-02)
2Microbial TestISO 11737-1:2018Attachment 11(IS-WVR-02)

7.6 Shelf-life Test

The shelf-life test was performed to decide expiration date and to assess a stability of physical properties of their packaging materials within the duration of the proposed shelf-life. The tests were performed in accordance with following standards.

No.Test ItemsStandardsCriteriaPass/FailAttachment No.(Report No.)
1Accelerated Aging TestASTM F 19803 yearsPassAttachment 12(MSK-000970)

According to ISO 11607-1, sterile barrier system and packaging system are established and maintained. The maintenance of this system has been validating for patient safety according to ISO 11607-2

No.Test ItemsStandardsCriteriaPass/FailAttachment No.(Report No.)
1Seal StrengthASTM F88:2000≥1.2(Unit:N/15mm)(After 3yearsAcceleratedaging device)PassAttachment 12(MSK-000970)
2Sterile BarrierSystem Integrity(Dye Penetration)ASTM F1929No leakagePassAttachment 12(MSK-000970)
3Sterility TestKPNo evidence ofmicrobialgrowthPassAttachment 12(MSK-000970)

{16}------------------------------------------------

Image /page/16/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features a stylized green leaf design above the word "INSUNG" in bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller font size.

7.7 Endotoxin Test

Bacterial Endotoxin Test (Kinetic), using the lysate reagent for Endobronchial Double Lumen Tube, was conducted in accordance with USP 39 <85> including USP <161> and "Guidance for industry pyrogen and endotoxin testing: Questions and Answers".

Therefore, the endotoxin concentration of the test article extract used in the test was decided to be less than 0.5 EU/mL.

No.Test ItemsStandardsCriteriaPass/FailAttachment No.(Report No.)
1BacterialEndotoxin TestUSP 39 <85>< 0.5EU/mLPassAttachment 1(20-048384-01-1)

7.8 Performance Test

The following tests were performed to assess effectiveness of the product performance. The tests were performed in accordance with following standards.

No.Test ItemsStandardsCriteriaPass/FailAttachment No.(Report No.)
1Dimension TestISO16628:2008Outside diameter ofbronchial segment :Express the results inmilimetres, asdetermined, to onedecimal place, roundedup to the nearest 0.5mmEffective inside diameter:The determineddimension is expressed inmillimetres rounded tothe nearest 0.2mm.PassAttachment 6(IS-20-1026-2)
2Colour CodingISO16628:2008Bronchial cuff andpilot balloon colourbluePassAttachment 6(IS-20-1026-2)
3SegmentDifferentiationISO16628:2008When viewed from theend of the machine, itmust be clearlydistinguishable fromeach other.PassAttachment 6(IS-20-1026-2)
2Tensile StrengthN/AWhen pulling for 15 seconds with a force of not less than 15 N in the longitudinal direction, except for removable connections such as connectors, there shall be no damage to the connection. (Korean Ministry of Food and Drug Safety Standards)PassAttachment 6(IS-20-1026-2)
3Cuff Leakage TestN/AThere should be no leakage when put into the water.PassAttachment 6(IS-20-1026-2)
4Connector Bonding StrengthISO 5356-1:2014Hold the connector and tube vertically and do not separate from a force of 15 N or less when pulled.PassAttachment 6(IS-20-1026-2)
5Cuff Resting DiameterISO 5361:2016At a pressure of 2 kPa the cuff diameter shall be as follows. within ± 0.5 mm of the nominal value when tested in accordance with ISO 5361 Annex B.PassAttachment 6(IS-20-1026-2)
6Tube CollapseISO 5361:2016The steel ball shall pass freely through the tube when tested according to ISO 5361 Annex CPassAttachment 6(IS-20-1026-2)
7Cuff HerniationISO 5361:2016No part of the inflated cuff shall reach beyond the nearest edge of the bevel when tested according to ISO 5361 Annex DPassAttachment 6(IS-20-1026-2)
8Cuff burstpressure(Cuff BurstEvaluation)N/ADo not damage at90cmH2O pressurePassAttachment 6(IS-20-1026-2)
9Cuff BondPressure(Cuff BondStrength)N/AANKOR cuff : Must notbe damaged at forces of2.0 kgf or less.Bronchial cuff : Must notbe damaged at forces of1.0 kgf or less.Tracheal cuff : Must notbe damaged at forces of2.0 kgf or less.PassAttachment 6(IS-20-1026-2)
10Bond strengthtesting of thejoints ofendobronchialtubeN/ADo not damage at90cmH2O pressurePassAttachment 6(IS-20-1026-2)
11Liquid and AirleakageISO 80369-7:2016No leakage sufficient toform a falling drop ofwaterPassAttachment 6(20-048204-01-1)
12Separation forceof one wayvalveISO 80369-7:2016The conical fitting undertest remain attached tothe test fixturePassAttachment 6(20-048204-01-1)
13Stress crackingISO 80369-7:2016There shall be noevidence of stresscracking of the conicalfittingPassAttachment 6(20-048204-01-1)
14RadiopaqueTestN/AThe outline shall beclearly visible whentaken with the followingconditions.Milliampere: 10MASTube Voltage: 65 to 70KVPPassAttachment 6(IS-20-1026-2)
15DimensionISO8836:2014Distance between focusand sample: 76.2 to 127 cm1. Y-110 : O.D 3.33 ± 0.15mm / minimum ID 2.00mm2. Y-112 : O.D 4.0 ± 0.15mm / minimum ID 2.45mmPassAttachment 6(IS-20-0710)
16Security ofconstruction ofsuction catheter(TensileStrength)ISO8836:2014The connector shall notdamage the catheterconnection at a force ofat least 5 N.PassAttachment 6(IS-20-0710)
17Shaft resistant tonegativepressure ofsuction catheterISO8836:2014Even if the vacuumregulator is blocked, theshaft must be intact.PassAttachment 6(IS-20-0710)

{17}------------------------------------------------

Image /page/17/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features three green leaves at the top, followed by the word "INSUNG" in large, bold, green letters. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in smaller, less bold letters.

{18}------------------------------------------------

Image /page/18/Picture/2 description: The image contains the logo for INSUNG MEDICAL CO., LTD. The logo features the word "INSUNG" in large, bold, green letters. Above the word "INSUNG" are three green leaves. Below the word "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller font.

{19}------------------------------------------------

Image /page/19/Picture/2 description: The image shows the logo for INSUNG MEDICAL CO., LTD. The logo features a green leaf design above the word "INSUNG" in large, bold, green letters. Below "INSUNG" is the text "INSUNG MEDICAL CO., LTD." in a smaller font, also in green. The overall design is clean and professional, suggesting a company focused on health and medical products.

8. Clinical Testing and Animal Testing

Clinical and animal testing were not performed for ANKOR Endobronchial Tube as part of the premarket notification requirements for this 510(k) submission and the subject of this premarket submission, ANKOR Endobronchial Tube, did not require clinical and animal studies to support substantial equivalence.

9. CONCLUSION

Under the comparing substantial equivalence between the subject device (ANKOR Endobronchial Tube) and the predicate device (Endobronchial Tube; K180253), there are the same points such as product code, indications for use, mechanism and structure. Although there are some differences (Size, Package and Raw Material), the performance data

are supported to the safety and effectiveness of the subject device.

In this regard, we conclude that the subject device is substantially equivalent to the predicate device.

§ 868.5740 Tracheal/bronchial differential ventilation tube.

(a)
Identification. A tracheal/bronchial differential ventilation tube is a device used to isolate the left or the right lung of a patient for anesthesia or pulmonary function testing.(b)
Classification. Class II (performance standards).