K Number
K200871

Validate with FDA (Live)

Date Cleared
2020-05-04

(33 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aspire MAX 7 - 11F Mechanical Thrombectomy System and Aspirator is indicated for the removal of fresh soft emboli and thrombi from vessels in the peripheral vasculature.

Device Description

The Subject Device includes (1) Aspire Aspirator and (1) MAX 7F - 11F Catheter. Subject Device and the Predicate are single-use, sterile, and short-term use medical devices designed to remove fresh soft emboli, debris, and thrombi from the peripheral vasculature. The Subject Device's Intended Use and Fundamental Scientific Principle are the same as the Predicate. The Predicate and Subject Device are the same in the way they are inserted into the body over a guidewire and advanced to the target arterial or venous peripheral vascular anatomy. The Subject Device and Predicate share the same core Aspirator to create thrombectomy force. MAX 7 - 11F Catheters may be connected to basic syringes and other aspiration pumps. Aspire Mechanical Aspirators may be connected to other catheters. The Subject Device does not add any new materials or manufacturing processes to the manufacturing process.

AI/ML Overview

The provided FDA 510(k) summary for the Aspire MAX 7-11F Mechanical Thrombectomy System and Aspire Mechanical Aspirator primarily focuses on establishing substantial equivalence to a predicate device through non-clinical testing. It explicitly states that clinical testing was not required for this determination. Therefore, a study demonstrating the device meets acceptance criteria through clinical performance (like an MRMC study or standalone AI performance) and an associated table of acceptance criteria with reported performance are not present in this document.

The document discusses non-clinical testing which confirmed the device meets specifications and intended use. However, it does not provide specific acceptance criteria or detailed performance metrics from these non-clinical tests in a format that lends itself to the requested table. The summary states:

"Non-clinical testing in this Submission confirms the Subject Device passes simulated use tracking. simulated aspiration, and simulated thrombectomy in the peripheral vasculature with test units accelerated aged up to 3-years. Accordingly, non-clinical testing confirms the Subject Device meets specifications, intended use, demonstration of claims, and equivalence to Predicate."

Based on the provided document, here's what can be extracted regarding acceptance criteria and the study:

Since the document explicitly states "Clinical testing is not required for the determination of substantial equivalence," and focuses on non-clinical testing, much of the requested information (like sample size for test sets, ground truth establishment, expert qualifications, MRMC studies, standalone AI performance) is not applicable or detailed in this submission.


Regarding the requested information:

1. A table of acceptance criteria and the reported device performance:

The document mentions that non-clinical testing confirmed the device "meets specifications, intended use, demonstration of claims, and equivalence to Predicate." However, specific quantitative acceptance criteria and their corresponding reported performance values from these non-clinical tests are not provided in this public summary. For example, it doesn't state "Aspiration force must be > X N; achieved Y N."

2. Sample size used for the test set and the data provenance:

  • Test set: Not explicitly stated for non-clinical tests in terms of specific numbers of devices or simulated scenarios. It refers to "test units accelerated aged up to 3-years."
  • Data Provenance: Not applicable as this is non-clinical/simulated data, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Ground truth for non-clinical, simulated tests would be based on engineering specifications and physical measurements, not expert human interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable for non-clinical testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. The document explicitly states "Clinical testing is not required for the determination of substantial equivalence." MRMC studies are associated with clinical performance evaluation, particularly for imaging devices involving human interpretation, which is not the case here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a mechanical thrombectomy system, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For the non-clinical tests, the "ground truth" would be the engineering specifications, physical principles, and laboratory measurements confirming the device's mechanical capabilities (e.g., aspiration force, tracking ability, thrombectomy effectiveness in simulated environments).

8. The sample size for the training set:

  • Not applicable. This is not an AI/machine learning device that requires a "training set."

9. How the ground truth for the training set was established:

  • Not applicable.

In summary, the provided document is a 510(k) summary for a mechanical device, not an AI/imaging device. As such, the detailed information typically requested for AI performance studies (like MRMC, standalone AI, training/test sets, expert adjudication) is not found here because it's not relevant to this type of submission. The primary "proof" of meeting acceptance criteria for this device, as presented, is through successful completion of non-clinical, simulated performance tests demonstrating its equivalence to a predicate device and adherence to engineering specifications.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

May 4, 2020

Control Medical Technology Shawn Fojtik President 2757 South 300 West Suite F Salt Lake City, Utah 84115

Re: K200871

Trade/Device Name: Aspire MAX 7 - 11F Mechanical Thrombectomy System and Aspire Mechanical Aspirator

Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEZ Dated: March 25, 2020 Received: April 1, 2020

Dear Mr. Fojtik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200871

Device Name

Aspire MAX 7 - 11F Mechanical Thrombectomy System and Aspire Mechanical Aspirator

Indications for Use (Describe)

The Aspire MAX 7 - 11F Mechanical Thrombectomy System and Aspirator is indicated for the removal of fresh soft emboli and thrombi from vessels in the peripheral vasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K200871

In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of Federal Regulations Part 807 (21 CFR $807), and in particular $807.92, the following summary is provided.

Date Summary was Prepared

April 29, 2020

Submitter

Control Medical Technology ("Submitter") 2757 South 300 West Salt Lake City, Utah 84115 USA Phone (383) 444-2666

FDA Establishment Registration Number

3007282893

Contact

Control Medical Technology 2757 South 300 West Salt Lake City, Utah 84115 USA Shawn Fojtik Phone (383) 444-2666

Device Information

Trade Names:Aspire MAX 7 - 11F Mechanical Thrombectomy System anAspire Mechanical Aspirator
Common Name:Thrombectomy catheter and/or embolectomy catheter
Classification Name:Embolectomy catheter
Product Code:QEZ
Regulation:Class II, 21 CFR 870.5150

Predicate Device

The Subject Device is substantially equivalent to the legally marketed Aspire MAX 5 - 6F Mechanical Thrombectomy System ("Predicate Device(s)").

Trade Names:Aspire MAX 5 – 6F Mechanical Thrombectomy System™
K113757 February 22, 2012.
Common Name:Thrombectomy catheter and/or embolectomy catheter
Classification Name:Embolectomy catheter
Product Code:DXE
Regulation:Class II, 21 CFR 870.5150

Reference Devices

The Subject Device references the Vascular Solutions Pronto XL 8 - 14F Thrombectomy System ("Reference K112571") and the Heraeus DuraSheath K181463 ("Reference K181463"). Reference K112571 is referenced because it is available with 8F - 14F outer diameter components and used for thrombectomy in peripheral vasculature same as the Subject Devices. Reference K181463 is

{4}------------------------------------------------

referenced as the catheter components in the Subject Device are the same outer dimensions, design, and indicated for use in the peripheral vasculature.

Trade Names:Pronto XL 8 – 14F Thrombectomy System K112571September 27, 2011
Common Name:Thrombectomy catheter and/or embolectomy catheter
Classification Name:Embolectomy catheter
Product Code:DXE
Regulation:Class II, 21 CFR 870.5150
Trade Names:Heraeus DuraSheath K181463 December 13, 2018
Common Name:Catheter or Sheath
Classification Name:Introducer, Catheter
Product Code:DYB and DRE
Regulation:Class II, 21 CFR 870.1340

Device Description

The Subject Device includes (1) Aspire Aspirator and (1) MAX 7F - 11F Catheter. Subject Device and the Predicate are single-use, sterile, and short-term use medical devices designed to remove fresh soft emboli, debris, and thrombi from the peripheral vasculature.

Image /page/4/Figure/6 description: The image shows two medical devices, the Aspire Aspirator and the MAX 5-6F Catheter Component. The Aspire Aspirator is a handheld device with a handle and a clear cylinder. The MAX 5-6F Catheter Component is a long, thin tube with a Luer Hub at one end and a Marker Band at the other, with a length of 55 to 135cm and an outer diameter of 5F to 6F.

The Subject Device's Intended Use and Fundamental Scientific Principle are the same as the Predicate. The Predicate and Subject Device are the same in the way they are inserted into the body over a guidewire and advanced to the target arterial or venous peripheral vascular anatomy. The Subject Device and Predicate share the same core Aspirator to create thrombectomy force. MAX 7 - 11F Catheters may be connected to basic syringes and other aspiration pumps. Aspire Mechanical Aspirators may be connected to other catheters. The Subject Device does not add any new materials or manufacturing processes to the manufacturing process.

Image /page/4/Figure/8 description: The image shows the Aspire Aspirator and the MAX 7-11F Catheter Component. The Aspire Aspirator is labeled with K072299, K113757, and K131998. The catheter component has a Luer Hub and a Marker Band, and the catheter is 60 to 90 cm long and 7F to 11F OD.

Indication for Use

Indication: The Aspire MAX 7 - 11F Mechanical Thrombectomy System and Aspire Mechanical Aspirator are indicated for the removal of fresh soft emboli from vessels in the peripheral vasculature. The indication and intended use are same as and substantially equivalent to the Predicate and same as other thrombectomy systems legally marketed under the DXE and QEZ product codes.

{5}------------------------------------------------

Predicate Comparison

Predicate Device K113757Subject Device K200871Comments
Device NameEmbolectomy CatheterEmbolectomy CatheterSame as Predicate.
CommitteeCardiovascularCardiovascularSame as Predicate.
FDA ProductCodeDXEQEZSubject Device product codeupdated per new FDAthrombectomy device productcodes.
CatheterComponentMaterialsMulti-layer nylon Pebax shaft,stainless steel wire braid, andPTFE liner.Multi-layer nylon Pebax shaft,stainless steel wire coil, andPTFE liner with thermoplasticelastomers TPE dilator.Subject Device and Predicatematerials are common multi-layercatheter, metal, and liner designs.
CatheterComponentOuter Diameters5F and 6F7F, 8F, 9F, 10F, 11FSubject Device outer diametersare larger than the PredicateK113757, same as ReferenceK181463, and smaller in outerdiameter than the largestReference K112571. Catheterswith outer diameters up to 14F likReference K112571 are used forthrombectomy in peripheralvasculature.
CatheterComponentLengths55 - 135cm60 - 90cmSubject Device is cathetercomponent lengths are in betweenPredicate K113757 shortest andlongest catheter componentlengths.
Indication for UseThe Aspire MAX 5 – 6FMechanical ThrombectomySystem and Aspire MechanicalAspirator are indicated for theremoval of fresh soft emboliand thrombi from vessels inthe peripheral vasculature.The Aspire MAX 7 – 11FMechanical ThrombectomySystem and Aspire MechanicalAspirator are indicated for theremoval of fresh soft emboliand thrombi from vessels inthe peripheral vasculature.Subject Device is the sameindication of use as the PredicateK113757.
PackagingMaterialsTray, Tyvek Lid, and Boxcommonly used withinterventional devices.Tray, Tyvek Lid, and Boxcommonly used withinterventional devices.Same as Predicate.
SterilizationEthylene Oxide Validated inaccordance withANSI/AAMI/ISO 11135-1 toachieve SAL 10-6Ethylene Oxide Validated inaccordance withANSI/AAMI/ISO 11135-1 toachieve SAL 10-6Same as Predicate.
UseSingle-use, disposableSingle-use, disposableSame as Predicate.

[This space intentionally blank]

{6}------------------------------------------------

The Subiect Device is substantially equivalent to Predicate device used to remove fresh, soft thrombi/emboli from the peripheral vasculature. Substantial equivalence is based on equivalence in:

Science Fundamental Scientific Principle

Device ConstructionDesignFunctionMechanism of ActionManufacturing
Device PerformanceAspirationSingle-use, disposable
LabelingIndication for UseWarningsInstructions for UseBox labels
ManufacturingBiocompatibilitySterilization

Non-Clinical Testing

Non-clinical testing in this Submission confirms the Subject Device passes simulated use tracking. simulated aspiration, and simulated thrombectomy in the peripheral vasculature with test units accelerated aged up to 3-years. Accordingly, non-clinical testing confirms the Subject Device meets specifications, intended use, demonstration of claims, and equivalence to Predicate.

Clinical testing

Clinical testing is not required for the determination of substantial equivalence.

Statement of Substantial Equivalence

The Subject Device is substantially equivalent to the legally marketed Predicate based on comparison of the device classification, fundamental scientific technology, basic operating principle, indication for use, technical characteristics, packaging, and sterilization methods.

Conclusion

The Subject Device is substantially equivalent to the currently marketed Predicate based on comparison of the device classification, fundamental scientific technology, basic operating principle, indication for use, technical characteristics, packaging, and sterilization methods. Testing confirms the suitability of Subject Device for its intended use.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).