K Number
K200825

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-06-26

(88 days)

Product Code
Regulation Number
866.5750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NOVEOS Specific IgE Assay is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE Assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an in vitro diagnostic aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories.

Device Description

The NOVEOS Specific IgE Assay is an immunometric, chemilyminescent procedure for the quantitative determination of IgE of known specificity in human serum samples. It employs fluorescent labelled magnetic, streptavidin coated microparticles which are incubated with a biotinylated allergenic capture reagent, patient sample and monoclonal anti-human IgE antibody: horseradish peroxidase conjugate. After a final wash, the resulting complex is incubated with the enzyme substrate and a chemiluminescent signal is generated, the magnitude of which is proportional to the concentration of IgE in the patient sample. The concentration of allergen-specific IgE is determined from a standard curve, which is traceable to the World Health Organization (WHO) reference reagent serum Immunoglobulin E (IqE) 11/234.

AI/ML Overview

This document describes the validation of the NOVEOS Specific IgE (sIgE) assays for Cat Dander (E001, Felis domesticus) and Timothy Grass (G006, Phleum pratense). This is a quantitative in vitro diagnostic device, not an AI/ML powered device. Therefore, many of the typical acceptance criteria for AI models, such as MRMC studies, ground truth establishment by experts, and training set details, are not applicable.

The acceptance criteria for this device are based on its analytical and clinical performance when compared to a legally marketed predicate device (ImmunoCAP Specific IgE Assay) and clinical diagnosis.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the performance metrics reported and the statistical analysis acceptable for FDA clearance of an in vitro diagnostic device of this type. The key performance indicators for a quantitative assay would typically include linearity, precision (imprecision/reproducibility), agreement with a predicate device, and clinical concordance (sensitivity and specificity).

Performance Table for NOVEOS Specific IgE (sIgE) Assays

Performance MetricAcceptance Criteria (Implied)Reported Device Performance (E001 - Cat Dander)Reported Device Performance (G006 - Timothy Grass)
Agreement with PredicateTo demonstrate substantial equivalence, percentages should be high (e.g., >90% for PPA and >95% for NPA, or similar ranges). Precision of CI is also important.PPA: 91.8% (95% CI: 84.6% – 95.8%)NPA: 99.3% (95% CI: 96.1% to 99.9%)(Cut-off: 0.35 kU/L)PPA: 86.4% (95% CI: 78.5% to 91.7%)NPA: 98.5% (95% CI: 94.6% to 99.6%)(Cut-off: 0.35 kU/L)
Clinical PerformanceClinical sensitivity and specificity should demonstrate adequate diagnostic accuracy. Typical acceptance ranges for IVDs might be >70-80% for sensitivity and >90% for specificity, depending on the clinical context.Sensitivity: 75.7% (95% CI 64.5% to 84.2%)Specificity: 100% (95% CI 97.1% to 100%)Sensitivity: 76.2% (95% CI 64.4% to 85.0%)Specificity: 99.2% (95% CI 95.6% to 99.9%)
Imprecision/ReproducibilityTotal CV% should generally be low, typically <10% to 15% across relevant concentrations and study phases (within-run, between-run, between-day, lot-to-lot, site-to-site). Specific ranges depend on the analyte and assay.Total CV% ranges from 7.0-10.5% (E001, within-lab imprecision)Lot-to-lot CV% ranges from 7.6-9.5%Site-to-site reproducibility CV% ranges from 5.2-13.9% (upper bounds for low concentration samples)Total CV% ranges from 5.0-11.1% (G006, within-lab imprecision)Lot-to-lot CV% ranges from 5.2-11.2%Site-to-site reproducibility CV% ranges from 6.0-11.1% (upper bounds for low concentration samples)
LinearityR2 value close to 1.0 (e.g., >0.99) and slope close to 1.0 with intercept close to 0, within the claimed measuring interval.R2: 0.999Slope: 1.00 (95% CI: 0.98 to 1.01)Intercept: -0.37 (95% CI: -0.79 to -0.05)Range: 0.03 - 101.62 kU/LR2: 0.997Slope: 1.02 (95% CI: 0.99 to 1.05)Intercept: 0.58 (95% CI: -0.36 to 1.51)Range: 0.06 - 104.99 kU/L
Detection LimitLoQ (Limit of Quantitation) should be defined and acceptable for clinical use, typically 0.14 kU/L for specific IgE assays to distinguish very low positive from negative (this is often the action limit for clinical use). LoB and LoD values should be reported.LoB: 0.02 kU/LLoD: 0.06 kU/LLoQ: 0.14 kU/LLoB: 0.03 kU/LLoD: 0.06 kU/LLoQ: 0.14 kU/L
Interference/Cross-ReactivityInterference from common endogenous/exogenous substances and cross-reactivity with other related/unrelated allergens or immunoglobulins should be demonstrated to be minimal (e.g., <=15% interference).Less than or equal to 15% interference for listed substances.Non-detectable cross-reactivity with other human immunoglobulins.<=15% inhibition for related/unrelated allergens.Less than or equal to 15% interference for listed substances.Non-detectable cross-reactivity with other human immunoglobulins.<=15% inhibition for related/unrelated allergens.
Reference Range (Expected Values)The device should align with established clinical cut-off values (e.g., <0.35 kU/L for negative) and support the verification of expected values in a normal population.132 healthy subjects tested <0.35 kU/L (all).127 healthy subjects tested <0.35 kU/L (all).
StabilityShelf-life and on-board stability should be demonstrated to support the claimed storage and use conditions.Shelf-life: 18-48 months (accelerated data). Real-time ongoing, currently supports 16 months.On-board: 48 hours to 28 days depending on component.Shelf-life: 18-48 months (accelerated data). Real-time ongoing, currently supports 16 months.On-board: 48 hours to 28 days depending on component.

2. Sample Sizes Used for the Test Set and Data Provenance

The document describes several test sets for different performance evaluations.

  • Agreement with Predicate Device (Comparative Study):
    • E001: 242 clinical samples.
    • G006: 238 clinical samples.
    • Data Provenance: The document does not explicitly state the country of origin. It refers to "clinical samples" and "patient samples" without further geographical details. The studies are assumed to be retrospective as they involve collecting and testing existing clinical samples.
  • Clinical Performance Study:
    • E001: 200 samples (70 samples with allergic status confirmed by skin-prick testing and clinical history, and 130 samples from healthy, non-atopic donors with no reported allergy).
    • G006: 188 samples (63 samples with allergic status confirmed by skin-prick testing and clinical history, and 125 samples from healthy, non-atopic donors with no reported allergy).
    • Data Provenance: Similar to the predicate comparison, the country of origin is not specified, and the studies appear to be retrospective.
  • Reference Range Verification:
    • E001: 132 apparently healthy subjects.
    • G006: 127 apparently healthy subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

For this type of in vitro diagnostic device, "experts" in the context of AI models (e.g., radiologists interpreting images) are not directly applicable.

  • For Agreement with Predicate: The ground truth is the result obtained from the predicate device (ImmunoCAP Specific IgE Assay), which is a previously FDA-cleared and legally marketed device.
  • For Clinical Performance: The ground truth for the clinical study was established by "skin-prick testing and clinical history". This implies that clinical professionals (e.g., allergists, physicians) made the diagnosis of "allergic status" or "non-atopic" based on standard medical procedures and patient history, rather than experts reviewing data specifically for the study.

The number and qualifications of these clinical professionals are not specified in the document, as it's a standard clinical diagnostic process rather than a specific panel of experts assembled for ground truth annotation.

4. Adjudication Method for the Test Set

Adjudication methods like 2+1 or 3+1 are typically used for subjective assessments (e.g., image interpretation by multiple human readers). For this quantitative in vitro diagnostic device:

  • For Agreement with Predicate: There is no specific "adjudication" in the sense of reconciling differing expert opinions. The comparison is directly between the quantitative results of the new device and the predicate device.
  • For Clinical Performance: The "allergic status" or "non-atopic" ground truth was established by clinical diagnosis (skin-prick testing and clinical history). While clinical diagnosis often involves physician judgment, the document does not describe a formal multi-reader adjudication process; it refers to the standard clinical standard of care.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic assay (a laboratory test) that measures a biomarker (specific IgE) in human serum. It is not an AI-powered image analysis device or a device that directly assists human readers in interpreting complex visual data. Therefore, the concept of human readers improving with AI assistance is not applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the primary performance evaluation of this device is standalone. The NOVEOS Specific IgE assay is an automated in vitro diagnostic test performed on the NOVEOS Immunoassay Analyzer. Its performance (quantitative measurement of IgE) is assessed directly through analytical studies (linearity, imprecision, detection limits, interference, cross-reactivity) and comparison to a predicate device/clinical diagnosis, without human intervention in the measurement process itself. The result is a quantitative value used by clinicians in conjunction with other findings.

7. The Type of Ground Truth Used

  • For Agreement with Predicate: The ground truth was based on the results obtained from the legally marketed predicate device (ImmunoCAP Specific IgE Assay). This is a common approach for establishing substantial equivalence for new IVD assays.
  • For Clinical Performance: The ground truth for allergic status was established by clinical diagnosis, specifically "skin-prick testing and clinical history." This is considered an objective clinical standard for diagnosing allergic disorders.

8. The Sample Size for the Training Set

This document describes the validation of an immunoassay, not an AI/ML model that requires "training sets" in the conventional sense. The device is a chemical/biological assay system. Therefore, there is no "training set" in the context of machine learning.

The studies described are for validation/testing of the already developed assay.

9. How the Ground Truth for the Training Set Was Established

As stated above, there is no "training set" for an AI/ML model for this type of device. The assay's "performance characteristics" (e.g., reagent formulations, assay parameters) would have been optimized during the device development phase, which is an engineering/chemistry process, not an AI model training process that uses labeled ground truth data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 26, 2020

Hycor Biomedical Irene Guzman Sr. RA Specialist 7272 Chapman Avenue Garden Grove, California 92841

Re: K200825

Trade/Device Name: NOVEOS Specific IgE (sIgE), Capture Reagent Cat Dander - E001, Felis Domesticus NOVEOS Specific IgE (sIgE), Capture Reagent Timothy Grass - G006, Phleum pratense Regulation Number: 21 CFR 866.5750 Regulation Name: Radioallergosorbent (Rast) Immunological Test System Regulatory Class: Class II Product Code: DHB Dated: March 27, 2020 Received: March 30, 2020

Dear Irene Guzman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ying (Katelin) Mao, Ph.D. Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200825

Device Name

NOVEOS Specific IgE (sIgE), Capture Reagent Cat Dander - E001, Felis domesticus, NOVEOS Specific IgE (sIgE), Capture Reagent Timothy Grass - G006, Phleum pratense

Indications for Use (Describe)

The NOVEOS Specific IgE Assay is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE Assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an in vitro diagnostic aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is prepared in accordance with the requirements of 21 CFR Part 807.92.

Date of Preparation:08-JUNE-2020
Manufacturer:Hycor Biomedical, LLC7272 Chapman AvenueGarden Grove, CA 92841
Contact Person:Irene M. GuzmanSr. RA Specialist7272 Chapman AveGarden Grove, CA 92841(714) 933-3052iguzman@hycorbiomedical.com

Device Name:

NOVEOS Specific IgE (sIqE), Capture Reagent Cat Dander - E001, Felis domesticus

NOVEOS Specific IgE (sIgE), Capture Reagent Timothy Grass - G006, Phleum pratense

Classification

NOVEOS Specific IgE (sIgE) Assay

Product Code DHB

Class II

CFR § 866.5750

Substantial Equivalence to: K051218

ImmunoCAP Specific IgE Assay and ImmunoCAP Specific IgE Conjugate 100 and Conjugate 400

ImmunoCAP Allergen e1, Cat Dander

ImmunoCAP Allergen g6, Timothy Grass

Indications for Use

The NOVEOS Specific IgE Assay is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE Assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an in vitro diagnostic aid in the clinical diagnosis of IgE mediated

{4}------------------------------------------------

allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories.

General Description

Reagents

The IgE Common Kit includes: Diluent A, Conjugate IgE, Substrate A, Substrate B, Fluo Beads™. Other required and recommended reagents include the allergen specific Capture Reagent, IgE Calibrator Set (6 levels -Cal 0.07 IgE, Cal 0.35 IgE, Cal 0.70 IgE, Cal 3.5 IgE, Cal 17.5 IgE, Cal 100 IgE kU/L), Calibrator Antibody IgE, Probe Wash Pack, Wash Buffer Concentrate, Cuvette Wash Pack, IgE Negative Control Pack, and IgE Positive Control Pack.

The liquid ready-to-use reagents demonstrate on-board stability of up to 48 hours for calibrators and controls and from 5 to 28 days for common assay components.

Assay Principle

The NOVEOS Specific IgE Assav is an immunometric, chemilyminescent procedure for the quantitative determination of IgE of known specificity in human serum samples. It employs fluorescent labelled magnetic, streptavidin coated microparticles which are incubated with a biotinylated allergenic capture reagent, patient sample and monoclonal anti-human IgE antibody: horseradish peroxidase conjugate. After a final wash, the resulting complex is incubated with the enzyme substrate and a chemiluminescent signal is generated, the magnitude of which is proportional to the concentration of IgE in the patient sample.

The concentration of allergen-specific IgE is determined from a standard curve, which is traceable to the World Health Organization (WHO) reference reagent serum Immunoglobulin E (IqE) 11/234.

Device Comparison

NOVEOS Specific IgE Assay on the NOVEOS Immunoassay Analyzer is comparable to the predicate device. ImmunoCAP Specific IgE on the ImmunoCAP 100.

{5}------------------------------------------------

Similarities and Differences
AttributeNOVEOS sIgE, E001, G006PredicatePhadia ImmunoCAP K051218
Intended UseThe NOVEOS Specific IgE Assay is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum. NOVEOS Specific IgE Assay is to be used with the NOVEOS Immunoassay Analyzer. It is intended for use as an in vitro diagnostic aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings and is to be used in clinical laboratories.ImmunoCAP Specific IgE is an in vitro quantitative assay for the measurement of allergen specific IgE in human serum or plasma (EDTA or Na-Heparin). ImmunoCAP Specific IgE is to be used with instruments Phadia 100, Phadia 250, Phadia 1000, Phadia 2500 and Phadia 5000. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of IgE mediated allergic disorders in conjunction with other clinical findings, and is to be used in clinical laboratories
Specimen TypeSerumSerum or plasma (EDTA, Na Heparin)
Sample Volume4 µL40 µL
Assay TypeQuantitativeSame
Detection AntibodyHorseradish peroxidase conjugated mouse anti-human IgE monoclonal antibodyβ-Galactosidase-anti-human IgE (mouse monoclonal antibody)
Detection LimitE001LoB: 0.02 kU/LLoD: 0.06 kU/LLoQ: 0.14 kU/LG006LoB: 0.03 kU/LLoD: 0.06 kU/LLoQ: 0.14 kU/LLoB: 0.001 kU/LLoD: 0.02 kU/LLoQ: 0.10 kU/L
Laboratory SettingClinical LaboratorySame
Assay PrinciplesFluorescence adjusted, immunometric, Chemiluminescent assayFluoroenzyme-immunoassay
Solid PhaseMagnetic microparticlesCellulose derivative
Calibrator TraceabilityWorld Health Organization (WHO) reference reagent serum Immunoglobulin E (IgE) 11/234Same
Calibration MethodHeterologous interpolation based on Total IgE calibration curveSame
Number of CalibratorsSixSame
Calibrator Levels0, 0.35, 0.7, 3.5, 17.5 and 100 kU/LSame
Assay Range0.17-100 kU/L0.10-100 kU/L
Reaction Temperature37°CSame
Time to First Result1 hour 45 minutes1 hour 45 minutes to 2 hour 30 minutes depending on model

{6}------------------------------------------------

Data – E001

The following table shows percent agreements between the NOVEOS sIgE and the ImmunoCAP results based on assessment of 242 clinical samples using a cut-off value of 0.35 kU/L:

ImmunoCAP sIgE
Equivocal
NOVEOS sIgEPositive0.22 - 0.31 kU/LNegativeTotal
Positive891191
Negative82141151
Total973142242

Positive percent agreement (PPA): 91.8% (95% CI: 84.6% – 95.8%) Negative percent agreement (NPA): 99.3% (95% Cl: 96.1% to 99.9%)

Data – G006

The following table shows percent agreements between the NOVEOS sIgE and the ImmunoCAP results based on assessment of 238 clinical samples using a cut-off value of 0.35 kU/L:

{7}------------------------------------------------

ImmunoCAP sIgE
NOVEOS sIgEPositiveEquivocal(0.22 - 0.31 kU/L)NegativeTotal
Positive892293
Negative142129145
Total1034131238

Positive percent agreement (PPA): 86.4% (95% Cl: 78.5% to 91.7%) Negative percent agreement (NPA): 98.5% (95% Cl: 94.6% to 99.6%)

Clinical Performance

E001

A clinical study was performed to support the diagnostic performance of the NOVEOS sIgE Assay for Cat Dander, F. domesticus E001. The clinical study comparing NOVEOS sIgE results to the allergic status of n=200 samples from patients was carried out in accordance with CLSI guideline EP9C: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Third Edition. A total of 70 samples with allergic status was confirmed by skin-prick testing and clinical history, and the other 130 samples from healthy, non-atopic donors with no reported allergy. Results are expressed as positive when a sample with a sIqE value is greater than or equal to 0.35 kU/L or negative when a sample with a sIgE value is less than 0.35 kU/L.

NOVEOS sIgEClinical Diagnosis
AtopicNon-atopicTotal
Positive53053
Negative17130147
Total70130200

Clinical Sensitivity: 75.7% (95% Cl 64.5% to 84.2%) Clinical Specificity: 100% (95% CI 97.1% to 100%)

G006

In addition to the ImmunoCAP comparison study, a clinical study was performed to support the diagnostic performance of the NOVEOS sIgE Assay for P. pratense, G006. The clinical study comparing NOVEOS sIgE results to the allergic status of n=188 samples from patients was carried out in accordance with CLSI quideline EP9C: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Third Edition. A total of 63 samples with allergic status was confirmed by skin-prick testing and clinical history, and the other 125 samples from healthy, non-atopic donors with no reported allergy. Results are expressed as positive when a sample with a sIgE value is greater than or equal to 0.35 kU/L or negative when a sample with a sIgE value is less than 0.35 kU/L.

{8}------------------------------------------------

NOVEOS sIgEClinical DiagnosisAtopicNon-atopicTotal
Positive48149
Negative15124139
Total63125188

Clinical Sensitivity: 76.2% (95% Cl 64.4% to 85.0%) Clinical Specificity: 99.2% (95% Cl 95.6% to 99.9%)

Imprecision/Reproducibility

E001

Repeatability and within-laboratory precision were determined in accordance with CLSI guideline EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods: A Statistical Approach and CLSI guideline EP15-A3: User Verification of Precision and Estimation of Bias. Samples were assayed in duplicate replicates in 2 runs per dav for 20 days on 3 NOVEOS Immunoassay Analyzers for a total of 80 replicates per sample. The SD and % CV of the withinrun, between-run, between-day, and total imprecision were calculated for each sample and results are summarized in the following table:

SampleMean(kU/L)Within-RunBetween-RunBetween DayTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)
LoQ180.260.027.1%0.015.0%0.025.9%0.0310.5%
PP468.630.313.6%0.182.1%0.495.7%0.617.0%
Lyphochek PosSample12.590.453.6%0.635.0%0.675.3%1.028.1%
NOVEOS Pos Sample13.620.594.3%0.685.0%0.826.0%1.228.9%
PP2840.811.373.4%1.674.1%2.476.0%3.288.0%

Lot-to-lot imprecision

Lot-to-Lot imprecision was evaluated with three different lots of the NOVEOS, E001, using a panel of serum samples in two replicates per run, two runs per day for twenty days (for a total of 240 replicates per sample). The results are summarized in the following table:

{9}------------------------------------------------

MeanWithin-RunBetween-RunBetween DayTotal
Sample(kU/L)SDCV (%)SDCV(%)SDCV(%)SDCV(%)
LoQ180.260.026.5%0.015.2%0.000.0%0.039.5%
PP468.510.303.5%0.515.9%0.101.2%0.657.6%
Lyphochek Pos Sample12.660.473.7%0.796.3%0.151.2%1.128.8%
NOVEOS Pos Sample13.650.544.0%0.785.7%0.000.0%1.269.2%
PP2841.071.533.7%2.566.2%0.481.2%3.358.2%

Site-to-site reproducibility

Site-to-site reproducibility was evaluated by testing a panel of 4 patient pools (one negative and three positive) and 2 controls (run as samples) at three sites using the same lot of reagent. Each sample was tested in five replicates per run, one run per day for five days on one NOVEOS Immunoassay Analyzer at each site (for a total of 75 replicates per sample). The results are summarized in the following table:

ResultSummaryWithinRun(Repeatability)BetweenDayBetweenSiteReproducibility
PanelNMean(kU/L)SDCVSDCVSDCVSDCV
PP61750.270.027.4%0.013.7%0.0311.1%0.03(.02-0.11)11.1%(7.5%-41.2%)
PP62750.500.036.0%0.012.0%0.0611.9%0.07(0.04-0.20)13.9%(7.9%-39.8%)
PP63751.150.076.1%0.021.7%0.119.6%0.13(0.08-0.37)11.3%(6.6%-32.5%)
NOV7510.640.565.3%0.222.1%0.464.3%0.76(0.54-1.26)7.1%(5.1%-11.8%)
LYP7512.580.524.1%0.221.7%0.322.5%0.65(0.50-0.93)5.2%(4.0%-7.4%)
PP647569.934.907.0%2.012.9%1.832.6%5.60(4.61-7.14)8.0%(6.6%-10.2%)

{10}------------------------------------------------

Imprecision/Reproducibility

G006

Repeatability and within-laboratory precision were determined in accordance with CLSI guideline EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods: A Statistical Approach and CLSI guideline EP15-A3: User Verification of Precision and Estimation of Bias. Samples were assayed in duplicate replicates in 2 runs per day for 20 days on 3 NOVEOS Immunoassay Analyzers for a total of 80 replicates per sample. The SD and % CV of the withinrun, between-run, between-day, and total imprecision were calculated for each sample and results are summarized in the following table:

Mean(kU/L)Within-RunBetween-RunBetween DayTotal
SampleSDCV (%)SDCV (%)SDCV (%)SDCV (%)
LoQ500.190.017.0%0.017.1%0.013.2%0.0210.4%
PP480.370.014.0%0.013.9%0.013.1%0.026.4%
PP256.360.193.0%0.213.3%0.152.4%0.325.0%
Lyphochek PosSample8.860.242.7%0.404.5%0.384.3%0.606.8%
NOVEOS PosSample27.501.154.2%1.866.8%2.137.7%3.0511.1%
PP0644.042.014.6%2.195.0%3.728.4%4.7610.8%

Lot-to-lot imprecision

Lot-to-Lot imprecision was evaluated with three different lots of the NOVEOS, G006, using a panel of serum samples in two replicates per run, two runs per day for twenty days (for a total of 240 replicates per sample). The results are summarized in the following table:

MeanWithin-RunBetween-RunBetween DayTotal
Sample(kU/L)SDCV (%)SDCV (%)SDCV (%)SDCV (%)
LoQ500.200.017.2%0.013.6%0.001.9%0.0210.6%
PP480.370.024.1%0.013.7%0.012.8%0.037.2%
PP256.320.213.3%0.061.0%0.061.0%0.335.2%
Lyphochek PosSample8.830.293.3%0.485.4%0.141.6%0.697.8%
NOVEOS PosSample28.051.214.3%2.087.4%0.080.3%3.1411.2%
PP0645.341.954.3%3.477.7%1.884.2%4.9410.9%

{11}------------------------------------------------

Site-to-site reproducibility

Site-to-site reproducibility was evaluated by testing a panel of 4 patient pools (one negative and three positive) and 2 controls (run as samples) at three sites using the same lot of reagent. Each sample was tested in five replicates per run, one run per day for five days on one NOVEOS Immunoassay Analyzer at each site (for a total of 75 replicates per sample). The results are summarized in the following table:

ResultSummaryWithinRun(Repeatability)BetweenDayBetweenSiteReproducibility
PanelNMean(kU/L)SDCVSDCVSDCVSDCV
PP65740.180.0211.1%0.000.0%0.000.0%0.02(0.01-0.02)11.1%(8.0%-11.2%)
PP66750.420.037.1%0.000.0%0.024.7%0.04(0.03-0.06)9.5%(6.3%-13.5%)
PP67752.160.104.6%0.062.8%0.146.5%0.18(0.12-0.42)8.3%(5.4%-19.4%)
LYP758.460.364.3%0.253.0%0.263.1%0.51(0.38-0.77)6.0%(4.5%-9.1%)
NOV7510.320.535.1%0.272.6%0.484.6%0.76(0.54-1.31)7.4%(5.2%-12.7%)
PP687557.384.027.0%0.761.3%2.033.5%4.56(3.63-6.14)7.9%(6.3%-10.7%)

One result was excluded from the final analysis dataset. The test result had an error flag for insufficient sample volume that was attributed to a bubble; the test was not repeated.

Linearity

Linearity was evaluated in accordance with CLSI quideline I/LA20, Analytical Performance Characteristics, Quality Assurance, and Clinical Utility of Immunological Assays for Human Immunoglobulin E (IgE) Antibodies of Defined Allergen Specificities, Third Edition. For each evaluation, three positive serum samples were each diluted in negative serum pool generating five 2-fold consecutive dilutions. Each diluted sample was tested in replicates of five and the each neat sample was tested in replicates of 15 all within one assay run using three lots of reagents. Dilutions of E001 specific IgE samples with analyte concentrations from 0.03 to 101.62 kU/L, and G006 specific IgE samples with analyte concentrations of 0.06 to 104.99 kU/L, fully encompassing the measuring interval of 0.17 to 100 kU/L were used to calculate the linear regression statistics below.

{12}------------------------------------------------

DilutionRange (kU/L)RegressionEquationSlope(95% CI)Intercept(95% CI)R2
E0010.03 - 101.62$y = 1.00x - 0.37$0.98 to 1.01-0.79 to -0.050.999
G0060.06 - 104.99$y = 1.02x + 0.58$0.99 to 1.05-0.36 to 1.510.997

Interference

Interference testing was carried out in accordance with CLSI guideline EP7, Interference Testing in Clinical Chemistry; Approved Guideline - Third Edition. The following substances show less than or equal to 15% interference with the NOVEOS slgE Assay, E001 & G006.

SubstanceConcentration
Hemoglobin200 mg/dL
Conjugated Bilirubin30 mg/dL
Unconjugated Bilirubin20 mg/dL
Intralipid3000 mg/dL
Biotin3500 ng/mL
Diphenhydramine19.6 µmol/L
Methylprednisolone1000 ng/mL
Ranitidine19.1 µmol/L
Omalizumab0.12 mg/mL
Human Serum Albumin120 g/L
Rheumatoid Factor513 IU/mL

Cross-Reactivity

Cross-reactivity testing for E001 and G006 was carried out in accordance with CLSI guideline EP7, Interference Testing in Clinical Chemistry; Approved Guideline - Third Edition. For each evaluation, all samples were tested in replicates of three within one assay run using three lots of reagents. The crossreactivity with other human immunoglobulins is non-detectable at physiological concentrations of IgA, IgD, IgM and IgG.

{13}------------------------------------------------

Cross-Reactivity (Analytical Specificity)

E001

Specificity of NOVEOS sIgE Assay, E001, was demonstrated by assessing Competitive Inhibition in accordance with CLSI I/LA20, Analytical Performance Characteristics, Quality Assurance, and Clinical Utility of Immunological Assays for Human Immunoglobulin E Antibodies of Defined Allergen Specificities, Third Edition. For E001, all the related (E006. Guinea Pig Epithelia) and unrelated allergens (G013. Extract Velvet Grass: W011. Russian Thistle: and M004. Mucor circinelloides) assessed show ≤15% inhibition to E001.

G006

Specificity of NOVEOS sIgE Assay, G006, was demonstrated by assessing Competitive Inhibition in accordance with CLSI I/LA20, Analytical Performance Characteristics, Quality Assurance, and Clinical Utility of Immunological Assays for Human Immunoqlobulin E Antibodies of Defined Allergen Specificities. Third Edition. For G006, all the related (G202, Corn, Cultivated) and unrelated allergens (W043, Sagebrush, Common; F018, Brazil Nut; and M004, Mucor circinelloides) assessed show ≤15% inhibition to G006.

Detection Limit

E001

Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were estimated in accordance with CLSI guideline EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline - Second Edition. A panel of analyte-free and low IgE samples were assayed on multiple reagent lots and instruments across six days. A total of 60 replicates of analyte-free and 300 replicates of low IgE sample were evaluated from which LoB and LoD were determined to be 0.02 kU/L, and 0.06 kU/L, respectively. The LoQ is defined as the lowest analyte concentration with a within-lab precision of 20%CV. A panel of low analyte samples were assayed in replicates of two in 2 runs per day for 20 days, 80 replicates total. The LoQ was determined to be 0.14 kU/L.

G006

Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) were estimated in accordance with CLSI guideline EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition. A panel of analyte-free and low lgE samples were assaved on multiple reagent lots and instruments across six days. A total of 60 replicates of analyte-free and 360 replicates of low IgE sample were evaluated from which LoB and LoD were determined to be 0.03 kU/L, and 0.06 kU/L, respectively. The LoQ is defined as the lowest analyte concentration with a

{14}------------------------------------------------

within-lab precision of 20%CV. A panel of low analyte samples were assayed in replicates of two in 2 runs per day for 20 days, 80 replicates total. The LoQ was determined to be 0.14 kU/L.

Reference Range

E001

The expected value is negative (<0.35 kU/L) for a specific allergen in a nonatopic person. Each laboratory professional should establish its own expected value. This expected value/reference range of NOVEOS sIgE, E001 in the normal population was verified by testing samples from 132 apparently healthy subjects in a clinical study. All 132 samples were tested below <0.35 kU/L.

G006

The expected value is negative (<0.35 kU/L) for a specific allergen in a nonatopic person. Each laboratory professional should establish its own expected value. This expected value/reference range of NOVEOS slgE, G006 in the normal population was verified by testing samples from 127 apparently healthy subjects in a clinical study. All 127 samples were tested below <0.35 kU/L.

Stability

Shelf life stability: Both an ongoing real-time stability study and an accelerated stability study were performed in accordance with CLSI EP25-A using three lots of NOVEOS sIgE Assay, E001 and G006. The accelerated stability data support the manufacturer's claim of 18 to 48 month unopened shelf-life stability for the individual assay components listed in the table below. The real-time stability study is on-going for E001 and G006. Available data supports 16 month unopened shelf-life stability when stored at 2-8°C per the manufacturer's instruction for use:

Shelf-lifeStability* (2-8°C)
Specific IgE Capture Reagent E00124 months
Specific IgE Capture Reagent G00624 months
IgE Common KitDiluent A48 months
Conjugate IgE18 months
Substrate A and Substrate B24 months
Fluo BeadsTM24 months
IgE Calibrator Set24 months
Calibrator Antibody IgE24 months
OthersProbe Wash Pack24 months

{15}------------------------------------------------

Shelf-lifeStability* (2-8°C)
Wash Buffer Concentrate48 months
Cuvette Wash Pack12 months
ControlsIgE Negative Control Pack48 months
IgE Positive Control Pack24 months

*Results based on accelerated stability data

On-board stability: A real-time stability study using three lots of NOVEOS slgE Assay, E001 and G006 support the on-board stability claim of 48 hours to 28 days for the individual assay components as summarized in the table below.

On-board Stability (2-8°C)
Specific IgE Capture Reagent E00128 days
Specific IgE Capture Reagent G00628 days
IgE Common KitDiluent A14 days
Conjugate IgE14 days
Substrate A and Substrate B14 days
Fluo Beads™14 days
IgE Calibrator Set48 hours
OthersProbe Wash PackN/A
Wash Buffer Concentrate28 days
Cuvette Wash Pack28 days
ControlsIgE Negative Control Pack48 hours
IgE Positive Control Pack48 hours

§ 866.5750 Radioallergosorbent (RAST) immunological test system.

(a)
Identification. A radioallergosorbent immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the allergen antibodies (antibodies which cause an allergic reaction) specific for a given allergen. Measurement of specific allergen antibodies may aid in the diagnosis of asthma, allergies, and other pulmonary disorders.(b)
Classification. Class II (special controls). The device, when intended to detect any of the allergens included in Table 1 in this paragraph, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. Table 1—Class II Exempt Allergens Under § 866.5750—Radioallergosorbent (RAST) Immunological Test Systems
Allergen code Allergen product Source (taxonomical name) Grass Pollens g1 Sweet vernal grass Anthoxanthum odoratum. g3 Cocksfoot grass, Orchard grass Dactylis glomerata. g4 Meadow fescue Festuca elatior. g5 Rye-grass (perennial rye grass) Lolium perenne. g7 Common reed (common reed grass) Phragmites communis. g8 Meadow grass, Kentucky blue (June grass) Poa pratensis. g9 Redtop, Bentgrass Agrostis stolonifera ,Agrostis gigantea (Agrostis alba ). g11 Brome grass Bromus inermis. g12 Cultivated rye (cultivated rye grass) Secale cereale. g13 Velvet grass Holcus lanatus. g14 Cultivated oat (cultivated oat grass) Avena sativa. g15 Cultivated wheat (cultivated wheat grass) Triticum aestivum (Triticum spp.). g16 Meadow foxtail (meadow foxtail grass) Alopecurus pratensis. g17 Bahia grass Paspalum notatum. g24 Wheat grass, Western Agropyron smithii (Elymus smithii ). g30 Bluegrass, annual Poa annua. g70 Wild rye grass Elymus triticoides Elymus condensatus. g71 Canary grass Phalaris arundinacea. g201 Barley, cultivated Hordeum vulgare. g202 Maize, corn (cultivated corn) Zea mays. g203 Salt grass Distichlis spicata. g204 False oat-grass Arrhenatherum elatius. g216 Cyn d 1 Cynodon dactylon. g701 Phl p 1.0102, Phl p 5.0101 Phleum pratense. g702 Phl p 7.0101 Phleum pratense. g703 Phl p 12.0101 Phleum pratense. Weed Pollens w2 Western ragweed Ambrosia psilostachya. w4 False ragweed Ambrosia acanthicarpa (Franseria acanthicarpa ). w5 Wormwood Artemisia absinthium Artemisia annua. w6 Mugwort Artemisia vulgaris. w7 Marguerite, ox-eye daisy Chrysanthemum leucanthemum. w8 Dandelion Taraxacum vulgare ,Taraxacum officinale. w9 Plantain (English), Ribwort Plantago lanceolata. w10 Goosefoot, lamb's quarters Chenopodium album. w11 Saltwort (prickly), Russian thistle Salsola kali (Salsola pestifer ). w12 Goldenrod Solidago virgaurea (Solidago spp.). w13 Cocklebur, common Xanthium commune. w14 Common pigweed (rough pigweed) Amaranthus retroflexus. w15 Scale, Lenscale Atriplex lentiformis. w16 Rough marsh elder Iva ciliate ,Iva annua. w17 Firebush (Kochia) Kochia scoparia. w18 Sheep sorrel Rumex acetosella. w19 Wall pellitory Parietaria officinalis. w20 Nettle (Common stinging nettle) Urtica dioica. w21 Wall pellitory Parietaria judaica. w22 Japanese hop (careless weed) Humulus japonicas (Humulus scandens ). w23 Yellow dock, Yellow dockweed Rumex crispus. w24 Spiny pigweed Amaranthus spinosus. w27 Carnation Dianthus spp. w28 Rose Rosa rugosa. w33 Clover Trifolium pratense. w35 Mexican tea Chenopodium ambrosioides. w36 Rabbit bush Ambrosia deltoidea (Franseria deltoides ). w37 Salt bush, annual Atriplex wrightii. w39 Water hemp, Western Amaranthus rudis (Acnida tamariscina ). w41 Burrobrush Hymenoclea salsola. w42 Poverty weed Baccharis neglecta. w43 Common sagebrush Artemisia tridentata. w45 Alfalfa Medicago sativa. w46 Dog fennel Eupatorium capillifolium. w53 Geranium Geranium spp. w67 Groundsel bush Baccharis halimifolia. w69 Iodine bush Allenrolfea occidentalis. w70 Ragweed, slender Ambrosia confertiflora. w75 Wing scale (wingscale) Atriplex canescens. w82 Careless weed Amaranthus palmeri ,Amaranthus hybridus. w90 Japanese hop Humulus japonicas (Humulus scandens ). w203 Rape (rape pollen) Brassica napus. w204 Sunflower Helianthus annuus. w206 Camomile Matricaria chamomilla. w207 Lupin Lupinus spp. w210 Sugar-beet Beta vulgaris. w211 Par j 2.0101 Parietaria judaica. w231 Art v 1 Artemisia vulgaris (Mugwort). w232 Sal k 1 Salsola kali. w233 Art v 3 Artemisa vulgaris (LTP, Mugwort). w234 Pla l 1 Plantago lanceolata. w235 Che a 1.0101 Chenopodium album. w236 Mer a 1.0101 Mercurialis annua. a753 Art v 1 Artemisia vulgaris (Mugwort weed). Tree Pollens t1 Box-elder (Maple) Acer negundo ,Acer saccharum. t2 Gray alder, speckled alder (alder) Alnus incana. t4 Hazel, hazelnut Corylus avellana ,Corylus americana. t5 American beech (beech) Fagus grandifolia (Fagus americana ). t6 Mountain juniper, Mountain cedar Juniperus ashei (Juniperus sabinoides ). t8 Elm Ulmus americana. t9 Olive Olea europaea. t10 Walnut Juglans californica ,Juglans nigra. t11 Maple leaf sycamore, London plane, Plane tree Platanus acerifolia. t61 Sycamore Platanus occidentalis. t12 Willow Salix caprea ,Salix nigra. t14 Cottonwood (Eastern Cottonwood/Black Cottonwood) Populus deltoides. t15 White ash Fraxinus americana. t16 White pine Pinus strobus. t18 Eucalyptus, gum-tree Eucalyptus globulus (Eucalyptus spp.). t19/t26 Acacia Acacia longifolia (Acacia spp.). t20 Mesquite Prosopis glandulosa/ Prosopis juliflora. t21 Melaleuca, cajeput tree Melaleuca quinquenervia (Melaleuca leucadendron ). t22 Pecan, hickory Carya illinoinensis (Carya pecan ). t23 Italian/Mediterranean/funeral cypress Cupressus sempervirens. t24 Japanese cypress Chamaecyparis obtusa (Chamaecyparis spp.). t25 Ash Fraxinus excelsior. t27 Maple, red Acer rubrum. t29 Acacia Acacia spp. t30 Birch, white Betula populifolia. t32 Willow, black Salix nigra. t33 Ash, Arizona Fraxinus velutina. t35 Cedar, salt Tamarix gallica. t37 Bald cypress (white bald cypress) Taxodium distichum. t38 Elm, Chinese/Siberian Ulmus pumila. t40 Hazelnut tree Corylus americana. t41 White hickory Carya alba (Carya tomentosa ). t42 Oak, red Quercus rubra. t43 Loblolly pine Pinus taeda. t44 Hackberry Celtis occidentalis. t45 Cedar elm Ulmus crassifolia. t47 Juniper, one seed Juniperus monosperma. t48 Pine, lodgepole Pinus contorta. t49 Pine, ponderosa Pinus ponderosa. t50 Beech, European Fagus sylvatica. t51 Tree of Heaven Ailanthus altissima. t52 Western white pine Pinus monticola. t54 Russian olive Elaeagnus angustifolia. t55 Scotch broom Cytisus scoparius. t56 Bayberry Myrica cerifera. t57 Red cedar Juniperus virginiana. t60 Western juniper Juniperus occidentalis. t61 Sycamore Platanus occidentalis. t70 Mulberry (white mulberry) Morus alba. t71 Red mulberry Morus rubra. t72 Queen palm Arecastrum romanzoffiamon. t73 Australian pine Casuarina equisetifolia. t77 Oak mix (red, white, black) Quercus spp. t80 Japanese cypress Chamaecyparis obtusa. t81 Japanese alder Alnus japonica. t83 Mango tree Mangifera indica. t90 Walnut, black Juglans nigra. t96 Poplar, white (poplar) Populus alba. t103/t218 Virginia live oak (live oak) Quercus virginiana. t105 Pepper tree Schinus molle. t110 Orange tree Citrus sinensis. t201 Spruce, Norway spruce Picea abies (Picea excelsa ). t202 Alder, smooth Alnus incana spp.Rugosa (Alnus rugosa ). t203 Horse chestnut Aesculus hippocastanum. t205 Elder Sambucus nigra. t206 Chestnut Castanea sativa. t207 Douglas fir Pseudotsuga menziesii (Pseudotsuga taxifolia ). t208 Linden Tilia cordata. t209 Horn beam Carpinus betulus. t210 Privet Ligustrum vulgare. t211 Sweet gum Liquidambar styraciflua. t212 Cedar Libocedrus decurrens. t213 Pine Pinus radiata. t214 Date palm Phoenix canariensis. t215 Lilac Syringa vulgaris. t217 Pepper tree Schinus molle. t217 Red alder Alnus rubra. t218 Virginia live oak Quercus virginiana. t218 Bayberry (bayberry/sweet gale) Myrica gale. t219 Palo verde Cercidium floridum. t219 Red cedar Juniperus virginiana. t220 Bet v 4 Betula verrucosa (Birch). t221 Bet v 2.0101, Bet v 4 Betula verrucosa (Birch). t222 Cypress (Arizona cypress) Cupressus arizonica. t223 Oil palm Elaeis guineensis. t224 Ole e 1 Olea europaea. t225 Bet v 6 Betula verrucosa (Birch). t226 Cup a 1 Cupressus arizonica. t227 Ole e 7 Olea Europaea. t228 Aspen, quaking Populus tremuloides. t229 Eastern hemlock Tsuga canadensis. t230 Redwood (sequoia) Sequoia sempervirens. t232 Pussy willow Salix discolor. t240 Ole e 9.0101 Olea Europaea. t241 Pla a 1.0101 Platanus acerifolia. t242 Pla a 2 Platanus acerifolia. t243 Pla a 3.0101 Platanus acerifolia. t244 Cor a 1.0103 Corylus avellana. t245 Aln g 1.0101 Alnus glutinosa. t246 Cry j 1 Cryptomeria japonica. t280 Locust tree Robinia pseudoacacia. t401 Brazilian peppertree Schinus terebinthifolius. t402 Mastic tree Pistacia lentiscus. t404 Tree of heaven Ailanthus altissima. t406 Date palm Phoenix dactylifera. a482 Ole e 1 Olea europaea (Olive Oil). Mites d207 Blo t 5.0101 Blomia tropicalis. d208 Lep d 2.0101 Lepidoglyphus destructor. Microorganisms, Molds, Yeast m1 Penicillium chrysogenum (Penicillium notatum ) Penicillium chrysogenum (Penicillium notatum ). m2 Cladosporium herbarum (Hormodendrum ) Cladosporium herbarum (Hormodendrum ). m3 Aspergillus fumigatus Aspergillus fumigatus. m4 Mucor racemosus Mucor racemosus. m5 Candida albicans Candida albicans. m7 Botrytis cinerea Botrytis cinerea. m8 Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes, Helminthosporium interseminatum ) Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes. m9 Fusarium moniliforme (Fusarium proliferatum ) Fusarium moniliforme (Fusarium proliferatum ). m10 Stemphylium botryosum Stemphylium herbarum (Stemphylium botryosum ). m11 Rhizopus nigricans Rhizopus nigricans. m12 Aureobasidium pullulans Aureobasidium pullulans. m13 Phoma betae Phoma betae. m14 Epicoccum purpurascens Epicoccum purpurascens (Epicoccum nigrum ). m15 Trichoderma viride Trichoderma viride. m16 Curvularia lunata Curvularia lunata ,Curvularia specifera ([K923044](https://510k.innolitics.com/device/[K923044](https://510k.innolitics.com/device/K923044)) ). m17 Cladosporium fulvum Cladosporium fulvum. m18 Fusarium culmorum Fusarium culmorum. m19 Aspergillus versicolor Aspergillus versicolor. m20 Mucor mucedo Mucor mucedo. m21 Aspergillus clavatus Aspergillus clavatus. m22 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m23 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m24 Stachybotrys atra Stachybotrys chartarum (Stachybotrys atra ). m24 Paecilomyces spp Paecilomyces spp. m25 Aspergillus versicolor Aspergillus versicolor. m25 Penicillium brevicompactum Penicillium brevicompactum. m26 Cladosporium cladosporioides Cladosporium cladosporioides. m26 Penicillium citrinum Penicillium citrinum. m27 Penicillium spp Penicillium spp. m29 Aspergillus repens Aspergillus repens. m30 Penicillium roqueforti Penicillium roqueforti. m32 Cladosporium cladosporioides Cladosporium cladosporioides. m34 Serpula lacrymans Serpula lacrymans. m36 Aspergillus terreus Aspergillus terreus. m37 Trichophyton mentagrophytes Trichophyton mentagrophytes. m40 Aspergillus amstelodami Aspergillus amstelodami. m43 Saccharomyces Carlsberg Saccharomyces carlsbergensis. m44 Saccharomyces cerevisiae Saccharomyces cerevisiae. m45 Hormodendrum hordei Hormodendrum hordei. m46 Bipolaris spicifera Bipolaris spicifera. m47 Aspergillus nidulans Aspergillus nidulans. m48 Aspergillus oryzae Aspergillus oryzae. m49 Fusarium oxysporum Fusarium oxysporum. m50 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m51 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m53 Microspora canis Microsporum canis (Microspora canis ). m54 Aspergillus flavus Aspergillus flavus. m63 Helminthosporium intersemin Helminthosporium intersemin. m66 Mucor plumbeus Mucor plumbeus. m67 Mycogone Mycogone perniciosa. m68 Nigrospora oryzae Nigrospora oryzae. m69 Rhodotorula Rhodotorula rubra (Rhodotorula mucilaginosa ). m70 Malassezia furfur (Pityrosporum orbiculare ) Malassezia furfur (Pityrosporum orbiculare ). m71 Spondylocladium Spondylocladium spp. m72 Epidermophyton Epidermophyton floccosum. m73 Epicoccum nigrum Epicoccum purpurascens (Epicoccum nigrum ). m80 Staphylococcal enterotoxin A (Sta a SEA ) Staphylococcus aureus. m80 Helminthosporium spp Helminthosporium spp. m81 Staphylococcal enterotoxin B (Sta a SEB ) Staphylococcus aureus. m88 Stemphylium solani Stemphylium solani. m93 Gliocladium fimbriatum Gliocladium fimbriatum. m94 Phycomyces blakesleeanus Phycomyces blakesleeanus. m201 Tilletia tritici (Ustilago nuda, Ustilago tritici ) (Barley smut) Tilletia tritici (Ustilago nuda, Ustilago tritici ). m202 Acremonium kiliense (Cephalosporium acremonium ) Acremonium kiliense (Cephalosporium acremonium ). m203 Trichosporon pullulans Trichosporon pullulans. m204 Ulocladium chartarum Ulocladium chartarum. m205 Trichophyton rubrum Trichophyton rubrum. m207 Aspergillus niger Aspergillus niger. m208 Chaetomium globosum Chaetomium globosum. m209 Penicillium frequentans Penicillium glabrum (Penicillium frequentans ). m209 Stachybotrys chartarum Stachybotrys chartarum (Stachybotrys atra ). m210 Trichophyton mentagrophytes var. goetzii Trichophyton mentagrophytes var. goetzii. m211 Trichophyton mentagrophytes var. interdigitale Trichophyton mentagrophytes var. interdigitale. m211 Oat smut Ustilago avenae. m212 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m212 Geotrichum candidum Geotrichum candidum. m213 Bermuda grass smut Ustilago cynodontis. m214 Johnson grass smut Sphacelotheca cruenta. m215 Corn smut Ustilago maydis. m218 Asp f 1.0101 Aspergillus fumigatus. a3050 Asp r 1 Aspergillus restrictus. m219 Asp f 2 Aspergillus fumigatus. m220 Asp f 3.0101 Aspergillus fumigatus. m221 Asp f 4 Aspergillus fumigatus. m222 Asp f 6.0101 Aspergillus fumigatus. m223 Staphylococcal enterotoxin C (Sta a SEC ) Staphylococcus aureus. m224 Staphylococcal enterotoxin D (Sta a SED ) Staphylococcus aureus. m226 Staphylococcal enterotoxin TSST (Sta a TSST ) Staphylococcus aureus. m227 Malassezia spp. (Pityrosporum spp.) Malassezia spp. (Pityrosporum spp.). m228 Aspergillus flavus. m229 Alt a 1.0101 Alternaria alternata (Alternaria tenuis ). m230 Alt a 6.0101 Alternaria alternata (Alternaria tenuis ). m231 Cla h 8.0101 Cladosporium herbarum (Hormodendrum ). m300 Eurotium spp Eurotium spp. m304 Aspergillus oryzae Aspergillus oryzae. m305 Penicillium brevicompactum Penicillium brevicompactum. m309 Aspergillus terreus Aspergillus terreus. m310 Aspergillus nidulans Aspergillus nidulans. m311 Aspergillus flavus Aspergillus flavus. m312 Aspergillus clavatus Aspergillus clavatus. Epidermal & Animal e6 Guinea pig epithelium Cavia porcellus. e7 Pigeon droppings Columba palumbus ,Columba livia. e25 Chicken serum Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e26 Parrot serum Psittacoidea spp. e62 Camel Camelus dromedaries. e70 Goose feathers Anser anser. e71 Mouse epithelium Mus musculus (Mus spp.). e73 Rat epithelium Rattus norvegicus. e74 Rat urine proteins Rattus norvegicus ,Rattus rattus. e75 Rat serum proteins Rattus norvegicus ,Rattus rattus. e76 Mouse serum proteins Mus musculus (Mus spp.). e77 Budgerigar droppings Melopsittacus undulatus. e78 Budgerigar feathers Melopsittacus undulatus. e79 Budgerigar serum proteins Melopsittacus undulatus. e80 Goat epithelium Capra hircus. e81 Sheep epithelium Ovis aries (Ovis spp.). e82 Rabbit epithelium Oryctolagus cuniculus (Oryctolagus spp.). e83 Swine epithelium Sus scrofa (Sus scrofa domesticus; Sus spp.). e84 Hamster epithelium Cricetus cricetus, Mesocricetus auratus, and Phodopus sungorus. e85 Chicken feathers Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e86 Duck feathers Anas platyrhynchos. e87 Rat epithelium, serum proteins, and urine proteins Rattus norvegicus Rattus rattus. e88 Mouse epithelium, serum proteins, and urine proteins (mouse) Mus musculus (Mus spp.). e89 Turkey feathers Meleagris gallopavo. e90 Budgerigar serum proteins, feathers, and droppings Melopsittacus undulatus. e91 Pigeon serum proteins, feathers, and droppings Streptopelia roseogrisea ,Psittacidae spp. e92 Parrot serum proteins, feathers, and droppings Ara spp. e93 Pigeon serum proteins Streptopelia roseogrisea. e94 Fel d 1.0101 Felis domesticus. a345 Fel d 1 Felis domesticus. e98 Parrot droppings Psittacoidea spp. e101 Can f 1.0101 Canis familiaris (Canis domesticus ). a174 Can f 1 Canis familiaris (Canis domesticus ). e102 Can f 2.0101 Canis familiaris (Canis domesticus ). e196 Parakeet feathers Nymphicus hollandicus. e197 Parakeet droppings Nymphicus hollandicus. e198 Parakeet serum Nymphicus hollandicus. e199 Canary bird serum Serinus canarius. e200 Canary bird droppings Serinus canarius. e201 Canary bird feathers (Canary feathers) Serinus canarius. e202 Reindeer epithelium Rangifer tarandus. e203 Mink epithelium Mustela spp. e204 Bos d 6 Bos domesticus (Bos taurus; Bos spp.). e205 Horse, serum proteins Equus caballus (Equus spp.). e206 Rabbit, serum proteins Oryctolagus cuniculus (Oryctolagus spp.). e208 Chinchilla epithelium Chinchilla laniger. e209 Gerbil epithelium Meriones unguiculatus. e210 Fox epithelium Vulpes vulpes. e211 Rabbit, urine proteins Oryctolagus cuniculus (Oryctolagus spp.). e212 Swine, urine proteins Sus scrofa (Sus scrofa domesticus; Sus spp.). e213 Parrot feathers Ara spp. e214 Finch feathers Lonchura domestica. e215 Pigeon feathers Streptopelia roseogrisea (Streptopelia spp.),Columbia spp. e216 Deer epithelium Dama dama. e217 Ferret epithelium Mustela putorius. e218 Chicken droppings Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e219 Chicken, serum proteins Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e220 Fel d 2, Cat serum albumin Felis domesticus. e221 Can f 3 Canis familiaris (Canis domesticus ) (Dog serum albumin ). e222 Swine serum albumin (Sus s PSA) Sus scrofa (Sus scrofa domesticus; Sus spp.). e225 Lovebird feathers Psittacoidea agapomis. e226 Can f 5.0101 Canis familiaris. e227 Equ c 1.0101 Equus caballus. e228 Fel d 4.0101 Felis domesticus. e230 Equ c 3 Equus caballus. e231 Mus m 1 Mus musculus. Food f9 Rice Oryza sativa. f12 Pea (green pea) Pisum sativum. f15 White bean Phaseolus vulgaris. f19 Cayenne pepper Capsicum frutescens (Capsicum annum ). f21 Sugar cane Saccharum officinarum. f22 Raspberry Rubus idaeus. f26 Pork Sus scrofa (Sus scrofa domesticus; Sus spp.). f29 Watermelon Citrullus lanatus (Citrullus vulgaris ). f31 Carrot Daucus carota. f32 Oyster mushroom Pleurotus ostreatus. f33 Orange Citrus sinensis. f35 Potato Solanum tuberosum. f43 Mother's milk Homo sapiens. f44 Strawberry Fragaria vesca (Fragaria spp.). f45 Yeast, baker's Saccharomyces cerevisiae. f46 Pepper, Red Capsicum annuum. f47 Garlic Allium sativum. f48 Onion Allium cepa. f49 Apple Malus x domestica (Malus spp.). f51 Bamboo shoot Phyllostachys pubescens. f52 Cacao/chocolate Theobroma cacao. f54 Sweet potato Ipomoea batatas. f55 Common millet Panicum miliaceum. f56 Foxtail millet Setaria italica. f57 Japanese millet Echinochloa crus-galli. f58 Pacific squid Todarodes pacificus. f59 Octopus Octopus vulgaris (Octopus spp.). f63 Kefir NA. f67 Parmesan cheese NA. f81 Cheese, cheddar type NA. f82 Cheese, mold type NA. f83 Chicken Gallus domesticus (Gallus gallus domesticus; Gallus spp.). f86 Parsley Petroselinum crispum. f87 Melon Cucumis melo Cucumis melo + Citrullus lanatus. f88 Mutton (lamb) Ovis aries (Ovis spp.). f90 Malt Hordeum vulgare. f92 Banana Musa spp. f93 Cacao Theobroma cacao. f94 Pear Pyrus communis (Pyrus spp.). f97 Yam Dioscorea spp.Dioscorea opposita. f97 Chamomile tea Matricaria chamomilla. f98 Gliadin Triticum aestivum (Triticum spp.). f102 Cantaloupe Cucumis melo var. cantalupensis. f105 Chocolate Theobroma cacao. f109 Cottonseed Gossypium hirsutum. f110 Giant radish Raphanus sativus. f118 Zucchini Cucurbita pepo. f119 Radish Raphanus sativus. f120 Venison Capreolus capeolus. f121 Pinto bean Phaseolus vulgaris. f122 Cheese, American NA. f127 Black-eyed pea Vigna unguiculata. f131 Black Olive Olea europaea. f136 Red beet Beta vulgaris var. conditiva. f139 Goat's Cheese Capra aegagrus. f140 Bran NA. f141 Corn (vegetables) Zea mays. f152 Green bell pepper Capsicum annuum. f155 Brewer's yeast Saccharomyces carlsbergensis. f157 Duck Anas domesticus. f158 Goose Anser anser. f160 Camembert cheese NA. f162 Nectarine Prunus persica var. nucipersica. f163 Kohlrabi Brassica oleracea var. gongylodes. f65 Perch f166 Leek Allium porrum. f170 Cheese (Switzerland) (Swiss cheese) NA. f174 Fig Ficus carica. f177 Cranberry Vaccinium macrocarpon. f179 Raisin Vitis spp. f182 Lima bean Phaseolus lunatus. f198 Flaxseed (bruised grain) Linum usitatissimum. f199 Untreated native milk Bos domesticus (Bos taurus; Bos spp.). f208 Lemon Citrus limon. f209 Grapefruit Citrus paradisi. f210 Pineapple Ananas comosus. f211 Blackberry Rubus fruticosus. f212 Mushroom (champignon) Agaricus hortensis (Agaricus spp.). f213 Rabbit Oryctolagus cuniculus (Oryctolagus spp.). f214 Spinach Spinacia oleracea. f215 Lettuce Lactuca sativa. f216 Cabbage Brassica oleracea var. capitata. f217 Brussels sprouts Brassica oleracea var. gem. f218 Paprika, sweet pepper Capsicum annuum. f219 Fennel seed Foeniculum vulgare. f219 Sage Salvia officinalis. f220 Cinnamon Cinnamomum spp. f221 Coffee Coffea spp. f222 Tea Camellia sinensis. f223 Green olive Olea europaea. f225 Summer squash, pumpkin Cucurbita pepo. f225 Pumpkin Cucurbita maxima. f226 Pumpkin seed Cucurbita pepo. f227 Sugar-beet seed Beta vulgaris. f229 Safflower Seed Carthamus tinctorius. f231 Milk, boiled Bos domesticus (Bos taurus; Bos spp.). f234 Vanilla Vanilla planifolia. f237 Apricot Prunus armeniaca. f241 Gouda cheese NA. f242 Cherry Prunus avium. f244 Cucumber Cucumis sativus. f246 Guar, guar gum Cyamopsis tetragonoloba. f247 Honey NA. f248 Rosemary Rosmarinus officinalis. f254 Plaice Pleuronectes platessa. f255 Plum Prunus domestica ,Prunus americana. f258 Squid Loligo spp. f259 Grape Vitis vinifera (Vitis spp.). f260 Broccoli Brassica oleracea var. italica (Brassica oleracea var. cultivar ). f261 Asparagus Asparagus officinalis. f262 Aubergine, eggplant Solanum melongena. f263 Green pepper Piper nigrum ,Capsicum annuum. f264 Eel Anguilla anguilla. f265 Caraway Carum carvi. f265 Cumin Cuminum cyminum. f266 Mace Myristica fragrans. f267 Cardamon Elettaria cardamomum. f268 Clove Syzygium aromaticum. f269 Basil Ocimum basilicum. f270 Ginger Zingiber officinale. f271 Anise Pimpinella anisum. f272 Tarragon Artemisia dracunculus. f273 Thyme Thymus vulgaris. f274 Marjoram Origanum majorana. f275 Lovage Levisticum officinale. f276 Fennel, fresh Foeniculum vulgare. f277 Dill Anethum graveolens. f278 Bay leaf Laurus nobilis. f279 Chili pepper Capsicum frutescens. f280 Black pepper Piper nigrum. f281 Curry (Santa Maria) NA. f282 Nutmeg Myristica fragrans. f283 Oregano Origanum vulgare. f284 Turkey meat Meleagris gallopavo. f285 Elk/moose meat Alces spp. f286 Mare's milk Equus caballus (Equus spp.). f287 Red kidney bean Phaseolus vulgaris. f288 Blueberry Vaccinium myrtillus (Vaccinium spp.). f289 Date Phoenix dactylifera. f291 Cauliflower Brassica oleracea var. botrytis. f292 Guava Psidium guajava. f293 Papaya Carica papaya. f294 Passion fruit, Maracuja Passiflora edulis (Passiflora spp.). f295 Carambola Averrhoa carambola. f296 Carob Ceratonia siliqua. f297 Gum Arabic Acacia senegal (Acacia spp.). f298 Tragacanth Astragalus spp. f299 Sweet chestnut (chestnut) Castanea sativa. f300 Pinto bean Phaseolus spp. f301 Persimmon (kaki fruit, sharon) Diospyros kaki. f302 Mandarin (tangerine, clementine, satsumas) Citrus reticulata. f305 Fenugreek Trigonella foenum-graecum. f306 Lime Citrus aurantifolia. f307 Hake Merluccius merluccius. f308 Sardine (pilchard) Sardina pilchardus. f310 Blue vetch Lathyrus sativus. f311 Megrim Lepidorhombus whiffiagonis. f315 Green bean Phaseolus vulgaris. f316 Rape seed Brassica napus. f317 Coriander Coriandrum sativum. f318 Jack fruit Artocarpus heterophyllus. f319 Beetroot Beta vulgaris. f320 Crayfish Astacus astacus. f321 Horse meat Equus caballus (Equus spp.). f322 Red currant Ribes sylvestre. f324 Hop (fruit cone) Humulus lupulus. f325 Saffron Colchicum autumnale. f328 Fig Ficus carica. f329 Watermelon Citrullus lanatus. f330 Rose hip Rosa spp. f331 Saffron Crocus sativus. f332 Mint Mentha piperita. f333 Linseed Linum usitatissimum. f336 Jujube Ziziphus jujuba. f336 Wine vinegar Vitis vinifera (Vitis spp.). f337 Sole Solea solea. f337 English sole Parophrys vetulus. f338 Wine, white Vitis vinifera (Vitis spp.). f339 Allspice Pimenta dioica. f339 Wine, red Vitis vinifera (Vitis spp.). f341 Cranberry Vaccinium oxycoccus ,Vaccinium macrocarpon. f342 Olive (black, fresh) Olea europaea. f343 Raspberry Rubus idaeus. f344 Sage Salvia officinalis. f346 Chives Allium schoenoprasum. f347 Quinoa Chenopodium quinoa. f348 Litchi Litchi chinensis. f349 Chum salmon roe Oncorhynchus keta. f358 Artichoke Cynara scolymus. f360 Yogurt NA. f368 Black bass Micropterus dolomieu (Micropterus dolomieui ). f374 Karaya gum Sterculia urens. f375 Horseradish Armoracia rusticana. f377 Maple syrup NA. f379 Okra Abelmoschus esculentus. f382 Beet, sugar Beta vulgaris var. altissima. f401 Loquat Eriobotrya japonica. f402 Fig Ficus carica. f403 Brewer's yeast Saccharomyces cerevisiae. f405 Mint Mentha spp. f406 Arugula Eruca vesicaria. House Dust h1 Greer Labs., Inc NA. h2 Hollister-Stier Labs NA. h6 Japan NA. Venoms & Insects i7 Midge Chironomus yoshimatsui. i8 Moth Bombyx mori ,Heterocera spp. i47 Water flea Daphnia spp. i49 Deer fly Chrysops spp. i51 Black ant Camponotus pennsylvanicus. i54 Flea mix (dog/cat), common flea Ctenocephalides spp. i71 Mosquito Aedes communis ,Aedes spp.and Culex spp. i72 Green nimitti Cladotanytarsus lewisi. i73 Blood worm Chironomus thummi ,Chironomusri parius ,Chironomus spp. i75 European hornet Vespa crabro. i76 Berlin beetle Trogoderma angustum. i77 European paper wasp Polistes dominulus. i78 Fly Musca domestica. i80 Bumblebee Bombus pennsylvanicus. i201 Horse bot fly Gasterophilus intestinalis. i202 Grain weevil Sitophilus granarius. i203 Mediterranean flour moth Ephestia kuehniella (Anagasta kuehniella ). i204 Horse fly Tabanus spp. i205 Bumblebee Bombus terrestris. i208 Api m 1.0101 Apis mellifera. a45 Api m 1 Apis mellifera. i209 Ves v 5.0101 Vespula vulgaris. a670 Ves v 5 Vespula vulgaris. i210 Pol d 5.0101 Polistes dominulus. i211 Ves v 1.0101 Vespula vulgaris. i213 Api m 4 Apis mellifera. i214 Api m 2 Apis mellifera. i215 Api m 3 Apis mellifera. i216 Api m 5 Apis mellifera. i217 Api m 10 Apis mellifera. i220 Bla g 1.0101 Blattella germanica. i221 Bla g 2.0101 Blattella germanica. i222 Bla g 5.0101 Blattella germanica. i223 Bla g 7 Blattella germanica. a46 Api m 2 Apis mellifera. Miscellaneous o1 Cotton, crude fibers Gossypium spp. o3 Cotton (treated) Gossypium spp. o70 Seminal fluid Homo sapiens. o71 Staphylococcus aureus Staphylococcus aureus. o72 Pichia pastoris crude extract customer specific Pichia pastoris. o72 Sperm-sediment Homo sapiens. o73 Pichia pastoris crude extr. vector customer specific Pichia pastoris. o74 Pichia pastoris with vector customer specific Pichia pastoris. o201 Tobacco leaf, tobacco dust Nicotiana tabacum. o202 Artemia salina, fish feed Artemia salina. o203 Tetramin, fish feed NA. o207 Daphnia, fish feed Daphnia spp. o211 Mealworm Tenebrio molitor. o212 Streptavidin Streptomyces avidini. o213 MBP (maltose binding protein) Escherichia coli. o214 CCD; MUXF3 from bromelain Ananas comosus. o72 Enterotoxin A (Sta a SEA) Staphylococcus aureus. o73 Enterotoxin B (Sta a SEB) Staphylococcus aureus. Parasites p1 Ascaris Ascaris suum. p2 Echinococcus Echinococcus granulosus. p3 Schistosoma Schistosoma mansoni. p4 Anisakis (Herring Worm) Anisakis simplex (Anisakis spp.). p5 Toxocara canis Toxocara canis. p10 Ani s 3.0101 Anisakis simplex (Anisakis spp.). p11 Ani s 1 Anisakis simplex (Anisakis spp.). Occupational k4 Threshing dust NA. k5 Flax NA. k7 Hay Dust NA. k8 Hop (hops) Humulus lupulus. k12 Grain mill dust NA. k14 Kapok NA. k20 Sheep's wool (treated) (wool) Ovis aries (Ovis spp.). k21 Sheep's wool (Untreated) Ovis aries (Ovis spp.). k23 Straw Dust NA. k33 Oak NA. k70 Green coffee bean Coffea spp. k71 Castor bean Ricinus communis. k72 Ispaghula Plantago psyllium/Plantago ovata. k73 Silk waste NA. k74 Silk Bombyx mori. k75 Isocyanate TDI (Toluene diisocyanate) NA. k76 Isocyanate MDI (Diphenylmethane diisocyanate) NA. k77 Isocyanate HDI (Hexamethylen diisocyanate) NA. k78 Ethylene oxide NA. k79 Phthalic anhydride NA. k80 Formaldehyde/Formalin NA. k81 Ficus Ficus benjamina (Ficus spp.). k83 Cotton seed Gossypium hirsutum. k84 Sunflower seed Helianthus annuus. k85 Chloramin T NA. k86 Trimellitic anhydride, TMA NA. k87 Asp o 21, alpha-amylase Aspergillus oryzae. k89 Orris root Iris florentina. k99 HSA (Human Serum Albumin) (Hom s HSA) Homo sapiens. k201 Car p 1, Papain Carica papaya. k202 Ana c 2, Bromelain Ananas comosus. k204 Maxatase Bacillus licheniformis. k205 Alcalase Bacillus spp. k206 Savinase, Protease 1 (Bac l Subtilisin) Bacillus spp. k208 Gal d 4, Lysozyme Gallus domesticus (Gallus gallus domesticus; Gallus spp.). k209 Hexahydrophtalic anhydrid NA. k210 Maleic anhydride NA. k211 Methyltetrahydrophtalic anhydrid NA. k212 Abachi wood dust Triplochiton scleroxylon. k213 Pepsin (Sus s Pepsin) Sus scrofa (Sus scrofa domesticus; Sus spp.). k213 TCPA NA. k214 Bougainvillea Bougainvillea spp. k225 Horse radish peroxidase (Arm r HRP) Armoracia rusticana. k226 Ascorbate oxidase (Cuc p ascorbate oxidase) Cucurbita pepo. k301 Flour dust Triticum spp. k501 Savinase customer specific Proprietary knowledge of customer. k502 Lipolase customer specific Proprietary knowledge of customer. k503 Termamyl customer specific Proprietary knowledge of customer. k504 Clazinase customer specific Proprietary knowledge of customer.