K Number
K200713

Validate with FDA (Live)

Device Name
EchoNavigator
Date Cleared
2020-04-09

(22 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EchoNavigator supports the interventionalist and surgeon in treatments where both live Echo guidance are used. The targeted patient population consists of patients with cardiovascular diseases requiring such a treatment.

Device Description

EchoNavigator is a tool that assists the interventionalist and surgeon with image guidance during treatment of cardiovascular disease for which the procedure uses both live X-ray and live Echo guidance. EchoNavigator can be used with compatible Echo-probes and Echo units in combination with compatible Philips interventional X-ray systems.

AI/ML Overview

The provided text does NOT contain information about acceptance criteria or specific study results for the EchoNavigator R3.0.3 device.

The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance studies with acceptance criteria.

Here's a breakdown of what is and is not in the document regarding your request:

What is present in the document:

  • Device Name: EchoNavigator R3.0.3
  • Predicate Device: EchoNavigator R1 (K121781)
  • Indications for Use: "EchoNavigator supports the interventionalist and surgeon in treatments where both live X-ray and live Echo guidance are used. The targeted patient population consists of patients with cardiovascular diseases requiring such a treatment."
  • Technological Characteristics Comparison: The document states that the proposed device has similar technological characteristics to the predicate. It notes the ability to display images from live X-ray and Ultrasound, and an unchanged core algorithm for probe detection. It also lists functionalities available in both devices (Synchronize Image Orientation, Multiple Views, Follow C-arm, Table Side Control, Manual Annotations, Image Capture Export). It mentions modifications like an extended "Multiple Views" functionality with a "Model View" and basic Touch Screen Module control for "Table Side Control."
  • Non-Clinical Performance Data: Mentions "Software verification testing" and "Non-clinical in-house simulated use design validation testing" were performed, and all tests passed. It lists several recognized standards that the device complies with (IEC 62304, IEC 62366-1, IEC 82304-1, ISO 14971, ISO 15223-1, UL 2900-1, IEC 80001-1).
  • Conclusion: States that EchoNavigator R3.0.3 is substantially equivalent to the predicate device in terms of indications for use, technological characteristics, safety, and effectiveness, without raising new safety or effectiveness concerns.

What is NOT present in the document (and therefore cannot be provided in the requested table/answers):

  • Specific Acceptance Criteria: The document does not define quantitative or qualitative acceptance criteria for device performance (e.g., accuracy, precision, sensitivity, specificity, time measurements, etc.). It only states that tests were "passed."
  • Reported Device Performance: No specific performance metrics or values are reported.
  • Sample Size for Test Set: Not mentioned.
  • Data Provenance (country, retrospective/prospective): Not mentioned.
  • Number of Experts and Qualifications for Ground Truth: Not mentioned.
  • Adjudication Method: Not mentioned.
  • MRMC Comparative Effectiveness Study: Not mentioned as being performed.
  • Standalone Performance Study: The document refers to "Software verification testing" and "Non-clinical in-house simulated use design validation testing," which might be considered standalone, but no performance metrics are provided.
  • Type of Ground Truth Used: Not mentioned for any studies.
  • Sample Size for Training Set: Not mentioned (and likely not applicable since it's not described as an AI/ML device that requires a training set in the typical sense; the focus is on software verification and simulated use validation).
  • Ground Truth Establishment for Training Set: Not mentioned.

In summary, the provided text does not contain the detailed performance data, acceptance criteria, study methodologies (like sample sizes, expert involvement, or adjudication), or ground truth specifics that your request asks for. The document is a regulatory submission focused on substantial equivalence based on non-clinical testing and comparison to a predicate, not a detailed performance study report.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. The full name of the agency is "U.S. Food & Drug Administration".

Philips Medical Systems Nederland B.V. % Ioana Ulea Regulatory Affairs Specialist Veenpluis 6 5684 PC Best THE NETHERLANDS

Re: K200713

Trade/Device Name: EchoNavigator R.3.0.3 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: March 16, 2020 Received: March 18, 2020

Dear Ioana Ulea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

April 9, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K200713

Device Name EchoNavigator R3.0.3

Indications for Use (Describe)

EchoNavigator supports the interventionalist and surgeon in treatments where both live Echo guidance are used. The targeted patient population consists of patients with cardiovascular diseases requiring such a treatment.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PHILIPS

510(k) Summary

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.

Date Prepared:March 16th, 2020
Manufacturer:Philips Medical Systems Nederland B.V.Veenpluis 6,5684 PC BestThe NetherlandsEstablishment Registration Number: 3003768277
Primary Contact Person:Ioana UleaRegulatory Affairs Specialist, IGT SystemsPhone: +31 618345875E-mail: ioana.ulea@philips.com
Secondary Contact Person:Michael KoningsDirector Regulatory Affairs IGT SystemsPhone: +31 619864032E-mail: michael.konings@philips.com
Device:
Trade Name:EchoNavigatorRelease Name:R3.0.3Classification Name:Picture archiving and communications systemClassification Regulation:21 CFR, Part 892.2050Classification Panel:RadiologyDevice Class:Class IIProduct Code:LLZTrade Name:EchoNavigatorRelease Name:R3.0.3Classification Name:Picture archiving and communications systemClassification Regulation:21 CFR, Part 892.2050Classification Panel:RadiologyDevice Class:Class IIProduct Code:LLZ
Trade Name:EchoNavigator
Release Name:R3.0.3
Classification Name:Picture archiving and communications system
Classification Regulation:21 CFR, Part 892.2050
Classification Panel:Radiology
Device Class:Class II
Product Code:LLZ
Predicate Device:
Trade Name:EchoNavigator R1Manufacturer:Philips Medical Systems Nederland B.V.510(k) Clearance:K121781 (October 26, 2012)Classification Name:Picture archiving and communications systemClassification Regulation:21 CFR, Part 892.2050Classification Panel:RadiologyDevice Class:Class IIProduct Code:LLZTrade Name:EchoNavigator R1Manufacturer:Philips Medical Systems Nederland B.V.510(k) Clearance:K121781 (October 26, 2012)Classification Name:Picture archiving and communications systemClassification Regulation:21 CFR, Part 892.2050Classification Panel:RadiologyDevice Class:Class IIProduct Code:LLZ
Trade Name:EchoNavigator R1
Manufacturer:Philips Medical Systems Nederland B.V.
510(k) Clearance:K121781 (October 26, 2012)
Classification Name:Picture archiving and communications system
Classification Regulation:21 CFR, Part 892.2050
Classification Panel:Radiology
Device Class:Class II
Product Code:LLZ

{4}------------------------------------------------

PHILIPS

Reference Device:Trade Name:QLAB Advanced Quantification Software
Manufacturer:Philips Ultrasound, Inc.
510(k) Clearance:K181264
Classification Name:Picture archiving and communications system
ClassificationRegulation:21 CFR 892.2050
Classification Panel:Radiology
Device Class:Class II
Product Code:LLZ
Device description:EchoNavigator is a tool that assists the interventionalist and surgeon with image guidance during treatment of cardiovascular disease for which the procedure uses both live X-ray and live Echo guidance. EchoNavigator can be used with compatible Echo-probes and Echo units in combination with compatible Philips interventional X-ray systems.
Indications for Use:EchoNavigator supports the interventionalist and surgeon in treatments where both live X-ray and live Echo guidance are used. The targeted patient population consists of patients with cardiovascular diseases requiring such a treatment.
Technological characteristics:The proposed EchoNavigator R3.0.3 has similar technological characteristics compared to the predicate device.The subject and predicate devices are based on the following same technological characteristics:The ability of the proposed device to display images from the live X-ray and Ultrasound modalities and the core algorithm for probe (transducer) detection is fundamentally unchanged compared to the predicate device. The following functionalities are available in both predicate and proposed device: Synchronize Image Orientation, Multiple Views, Follow C-arm, Table Side Control, Mannual Annotations and Image Capture Export. The following modifications were implemented in EchoNavigator R3.0.3: The Multiple Views functionality provided by the predicate device is extended in the proposed device with the Model View , which is an additional view compared to the Multiple Views available in the predicate device (X-ray, Free, Echo and C-arm). Although the Table Side Control is available in both devices, the proposed device offers basic control from the Touch Screen Module (TSM) in addition to the wireless mouse control on the tray table, as offered by the predicate device. While both devices are allowing the user to identify anatomical structures in one modality and annotate them by Mannual

{5}------------------------------------------------

PHILIPS

Summary of Non-ClinicalPerformanceData:No#StandardNumberand DateStandard NameRecognitionNumber
1.IEC 62304 Edition1.1 2015-06Medical device software –Software life cycle processes13-79
2.IEC 62366-1Edition 1.0 2015-02Medical devices – Part 1:Application of usabilityengineering to medical devices5-114
3.IEC 82304-1Edition 1.0 2016-10Health software – Part 1:General requirements forproduct safety13-97
4.ISO 14971:2007/(R)2010(Corrected 4October 2007)Medical devices – Applicationof risk management to medicaldevices5-40
5.ISO 15223-1 ThirdEdition 2016-11-01Medical devices – Symbols tobe used with medical devicelabels, labelling and informationto be supplied5-117
6.UL 2900-1First Edition 2017Software Cybersecurity forNetwork-Connectable Products,Part 1: General Requirements13-96
7.IEC 80001-1Edition 1.0 2010-10Application of risk managementfor IT-networks incorporating13-38

{6}------------------------------------------------

medical devices – Part 1: Roles,responsibilities and activities
-------------------------------------------------------------------------

Software verification testing has been performed to verify that all System Requirements Specification including all the Privacy and Security requirements have been implemented. Results demonstrated that all executed verification tests were passed. Software Verification data supports the proposed EchoNavigator R3.0.3 relative to the predicate EchoNavigator R1.

Non-clinical in-house simulated use design validation testing has been performed to validate that EchoNavigator R3.0.3 conforms to intended use, claims and user needs. Results demonstrated that all executed validation protocols were passed.

Summary of The EchoNavigator R3.0.3 introduces no new indications for use or Clinical technological characteristics relative to predicate device (K121781) that would Performance require clinical testing. Data: The EchoNavigator R3.0.3 is substantially equivalent to the predicate device Substantial Equivalence EchoNavigator R1 in terms of indications for use, technological characteristics, Conclusion: safety and effectiveness. Additionally, substantial equivalence was demonstrated by non-clinical performance tests provided in this 510(k) premarket notification. These tests demonstrate that EchoNavigator R3.0.3 complies with the user need requirements as well as the requirements specified in the FDA-recognized consensus standards and guidance documents.

Therefore EchoNavigator R3.0.3 is as safe and effective as its predicate device and does not raise any new safety and/or effectiveness concerns.

N/A