K Number
K200699
Date Cleared
2020-05-07

(51 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal(Cardiac) and Peripheral vessel.

The Diagnostic Ultrasound System is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

Device Description

The HS70A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HS70A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS70A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

This document is a 510(k) Premarket Notification for the Samsung Medison HS70A Diagnostic Ultrasound System. It aims to demonstrate substantial equivalence to previously cleared predicate devices.

Summary of Acceptance Criteria and Device Performance:

The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing specific quantitative acceptance criteria or a dedicated study with performance metrics against those criteria for the new device features. Instead, it states that the device was evaluated against existing standards and guidance, and that migrated software functionality was tested to ensure image quality was not compromised.

The comparison table in the document indicates that the HS70A (Under Review) is largely "Same to predicate" for various features, including:

  • Intended Use and Clinical Applications: Identical to predicate devices (Fetal/Obstetrics, Abdominal, Gynecology, etc.).
  • Fundamental Technology: Employs the same as predicate devices.
  • Operating Modes: Identical (B-mode, M-mode, PW Doppler, CW Doppler, Color Doppler, Power Amplitude Doppler, S-Harmonic, 3D imaging, ElastoScan, Combined modes).
  • Scanhead Types and Frequency: Identical.
  • Acoustic Output Display & FDA Limits: Identical.
  • Transmit/Receive Channels: Identical (192).
  • System Characteristics: Mostly identical, with an addition of a 23.8-inch monitor option (predicate had 23-inch).
  • Product Safety Certification: Conforms to IEC 60601-1, CSA C22.2 No.601.1, IEC 60601-2-37.
  • EMC Compliance: Conforms to IEC60601-1-2: 2014(4th Edition).
  • Acoustic Output Display Standard: Track 3.
  • Biocompatibility Compliance: Conforms to ISO10993-1.
  • Software Functionality: Many features are "Same to predicate" (e.g., Q Scan, ClearVision, MultiVision, Auto IMT+, Strain+, S-Detect for Breast/Thyroid, 3D Imaging features, HDVI, RealisticVue, CEUS+, 5D NT, 5D LB, 5D CNS, ElastoScan, E-Thyroid, E-Breast, ArterialAnalysis).

New Features and their Predicate Clearance:

The document highlights the addition of new software features and one new transducer, which are cleared based on their presence in the second predicate device (HERA W9/W10, K192319), indicating that these features themselves have undergone prior FDA review and clearance.

FeatureReference in DocumentPredicate Clearance
IOTA-ADNEXK192319Same (Refer to SE analysis 1)
BiometryAssistK192319Same (Refer to SE analysis 2)
CrystalVueFlowK192319SE (Refer to SE analysis 3)
5D HeartK192319SE (Refer to SE analysis 4)
EV2-10A TransducerK192319Same (Refer to SE analysis 5)

This implies that the acceptance criteria for these specific features were met in their previous 510(k) clearances.

Details of the Study (as described in the document):

The document explicitly states: "The subject of this premarket submission, HS70A, is not required clinical studies to support substantial equivalence."

Instead, the submission relies on:

  1. Non-Clinical Tests: Evaluation for acoustic output, biocompatibility, cleaning and disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety, conforming with applicable medical device safety standards (e.g., ISO 14971, ANSI AAMI ES60601-1, IEC60601-1-2, ISO 10993-1, IEC60601-2-37, NEMA UD 2-2004).
  2. Software Functionality Evaluation: "All migrated software functionality was evaluated using the same test criteria as the predicates for all applicable imaging modes to ensure that migration into a new system design did not compromise image quality with respect to the intended use of each feature." This evaluation followed the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued on May 11, 2005."

Based on the provided text, the following information is not available or not applicable:

  • Sample Size Used for the Test Set and Data Provenance: Not specified for performance of new features or overall substantial equivalence, as clinical studies were not required. The non-clinical evaluations likely used test equipment and phantom data.
  • Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts: Not applicable, as expert-driven ground truth for performance studies is not described.
  • Adjudication Method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not done.
  • Standalone (algorithm only without human-in-the-loop performance) Study: Not explicitly described in terms of separate performance metrics, but the software functionality evaluation would implicitly be a standalone test for the algorithm's output.
  • Type of Ground Truth Used: For non-clinical tests, physical measurements and engineering specifications would serve as ground truth (e.g., Acoustic Output, EMC, Electrical Safety). For the software functionality evaluation, the "same test criteria as the predicates" implies comparison against established performance levels or outputs of the predicate versions.
  • Sample Size for the Training Set: Not applicable, as the submission focuses on substantial equivalence based on existing technology and non-clinical testing, not on developing a new AI/ML model that requires a distinct training set. The software is largely inherited or based on previously cleared modules.
  • How the Ground Truth for the Training Set Was Established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized depiction of the Department of Health & Human Services logo. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

Samsung Medison Co., Ltd. % Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA

Re: K200699

Trade/Device Name: HS70A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: March 13, 2020 Received: March 17, 2020

Dear Ji Yea Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

May 7, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200699

Device Name HS70A Diagnostic Ultrasound System

Indications for Use (Describe)

The Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal(Cardiac) and Peripheral vessel.

The Diagnostic Ultrasound System is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

Type of Use (Select one or both, as applicable) Remediation Use (Part 21 CFR 201.3 Labeling) Over-The-Counter Use (OTC) Remediation Use (Part 21 CFR 201.3 Labeling) Over-The-Counter Use (OTC)
Remediation Use (Part 21 CFR 201.3 Labeling)
Over-The-Counter Use (OTC)

| > Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.

5. 510(K) Summarv: K200699

In accordance with 21 CFR 807.92 the following summary of information is provided:

  • Date Prepared March 13, 2020 1.
    1. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, Republic of Korea

3. Contact Person

Primary contact Ji Yea Lee Regulatory Affairs Specialist Phone: +82.2.2194.1594 Fax: +82.31.8017.9576 Email: iivea722.lee@samsungmedison.com

Secondary Contact Ninad Gujar Director of Regulatory & Quality Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com

4. Proposed Device

  • Common/Usual Name: Diagnostic Ultrasound System and Accessories
  • Proprietary Name: HS70A Diagnostic Ultrasound System
  • Regulation Name: Ultrasound pulsed Doppler imaging system
  • Panel/ Requlatory Class: Radiology / II
  • Product Code: IYN, IYO, ITX
  • Regulation : 21 CFR 892.1550, 892.1560, 892.1570
    1. Predicate Device: HS70A Diagnostic Ultrasound System (K173713) Second Predicate Device:

HERA W9/HERA W10 Diagnostic Ultrasound System (K192319)

Device Description 6.

The HS70A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HS70A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS70A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

7. Indications for Use

The Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound

{4}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/4/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the letters are evenly spaced. The oval is tilted slightly to the right.

imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Sketal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal(Cardiac) and Peripheral vessel.

The Diagnostic Ultrasound System is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.

8. Technology

The HS70A employs the same fundamental scientific technology as its predicate devices.

9. Determination of Substantial Equivalence

The proposed HS70A is similar to the predicate device with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness. A comparison of the proposed device to the currently marketed predicates is provided in the table below.

HS70AAddition of V2.02Remarks
Clinical application- No addition
Operation of modes- No addition
Applied transducers- EV2-10ATransducer 510(k) Clearance EV2-10A K192319
SW Features- Addition of predicate S/W features:: IOTA-ADNEX, BiometryAssist, 5D Heart, CrystalVue Flow- Improvement of cleared S/W functions:: Measurement, Setup&Preferences, Utility, DICOM, 5D Follicle, SonoviewSW Features 510(k) clearance IOTA-ADNEX K192319 BiometryAssist K192319 5D Heart K192319 CrystalVue Flow K192319
HW Features- 23.8 inch Monitor

{5}------------------------------------------------

510(k) Premarket Notification – Traditional

Image /page/5/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed inside a blue, oval-shaped background.

HS70A(Under Review)HS70A(K173713)HERA W9/HERA W10(K192319)Note
Device nameThe subject devicePrimary PredicateSecond Predicate
ManufacturerSAMSUNG MEDISON.SAMSUNG MEDISON.SAMSUNG MEDISON.
Regulation No.892.1550892.1560892.1570892.1550892.1560892.1570892.1550892.1560892.1570
Product CodeIYN, IYO, ITXIYN, IYO, ITXIYN, IYO, ITXSame to predicate
Intended UseThe Diagnostic UltrasoundSystem and transducers areintended for diagnosticultrasound imaging and fluidanalysis of the human body.The Diagnostic UltrasoundSystem and transducers areintended for diagnosticultrasound imaging and fluidanalysis of the human body.The Diagnostic UltrasoundSystem and transducers areintended for diagnosticultrasound imaging and fluidanalysis of the human body.
clinical Application- Fetal/Obstetrics- Fetal/ObstetricsFetal/ObstetricsSame to predicate
- Abdominal- AbdominalAbdominalSame to predicate
- Gynecology- GynecologyGynecologySame to predicate
- Intra-operative- Intra-operativeSame to predicate
- Pediatric- PediatricPediatricSame to predicate
- Small Organ- Small OrganSmall OrganSame to predicate
- Neonatal Cephalic- Neonatal CephalicNeonatal CephalicSame to predicate
- Adult Cephalic- Adult CephalicAdult CephalicSame to predicate
- Trans-rectal- Trans-rectalTrans-rectalSame to predicate
- Trans-vaginal- Trans-vaginalTrans-vaginalSame to predicate
- Trans-esophageal(Cardiac)- Trans-esophageal(Cardiac)Same to predicate
- Muscular-Skeletal(Conventional, Superficial)- Muscular-Skeletal(Conventional, Superficial)Muscular-Skeletal(Conventional, Superficial)Same to predicate
- Urology- UrologyUrologySame to predicate
- Cardiac Adult- Cardiac AdultCardiac AdultSame to predicate
- Cardiac Pediatric- Cardiac PediatricCardiac PediatricSame to predicate
- Peripheral vessel- Peripheral vesselPeripheral vesselSame to predicate
Scanhead Types- Linear Array- Linear ArrayLinear ArraySame to predicate
Device nameHS70A(Under Review)HS70A(K173713)HERA W9/HERA W10(K192319)Note
The subject devicePrimary PredicateSecond Predicate
- Curved Linear Array- Curved Linear ArrayCurved Linear ArraySame to predicate
- Endocavity- EndocavityEndocavitySame to predicate
- Phased Array- Phased ArrayPhased ArraySame to predicate
- Static Probes- Static ProbesStatic ProbesSame to predicate
ScanheadFrequency1.0 ~ 20.0 MHz1.0 ~ 20.0 MHz1.0 ~ 20.0 MHzSame to predicate
Acoustic OutputDisplay & FDALimits- Display Feature for HigherOutput-Track3- MI Output Display- TI Output Display- Display Feature for HigherOutput-Track3- MI Output Display- TI Output Display- Display Feature for HigherOutput-Track3- MI Output Display- TI Output DisplaySame to predicate
Modes of Operation- B-mode- B-modeB-modeSame to predicate
- M-mode- M-modeM-modeSame to predicate
- Pulsed wave (PW)Doppler- Pulsed wave (PW) DopplerPulsed wave (PW) DopplerSame to predicate
- Continuous wave (CW)Doppler- Continuous wave (CW)DopplerContinuous wave (CW)DopplerSame to predicate
- Color Doppler- Color DopplerColor DopplerSame to predicate
- Power Amplitude Doppler- Power Amplitude DopplerPower Amplitude DopplerSame to predicate
- Pulse Inversion HarmonicImaging (S-Harmonic)- Pulse Inversion HarmonicImaging (S-Harmonic)Pulse Inversion HarmonicImaging (S-Harmonic)Same to predicate
- 3D imaging mode- 3D imaging mode3D imaging modeSame to predicate
- ElastoScan Mode- ElastoScan ModeElastoScan ModeSame to predicate
- Combined modes- Combined modesCombined modesSame to predicate
#Transmit Channels192192192Same to predicate
#Receive Channels192192192Same to predicate
System- Beamformer 192- Beamformer 192Beamformer 192Same to predicate
Characteristics- Mobile cart- Mobile cartMobile cartSame to predicate
- LCD Monitor (LEDBacklight unit):- LCD Monitor (LEDBacklight unit):LCD Monitor (LEDBacklight unit):Same to predicate
Device nameHS70A(Under Review)The subject deviceHS70A(K173713)Primary PredicateHERA W9/HERA W10(K192319)Second PredicateNote
23inch, 23.8inch23inch23inch
- 256 gray shades onmonitor- 256 gray shades onmonitor256 gray shades on monitorSame to predicate
- 100-240VAC, 1100VA,- 50/60Hz- 100-240VAC, 1100VA,- 50/60Hz100-240VAC, 1100VA,50/60HzSame to predicate
Product SafetyCertification- IEC 60601-1- IEC 60601-1IEC 60601-1Same to predicate
- CSA C22.2 No.601.1- CSA C22.2 No.601.1CSA C22.2 No.601.1Same to predicate
- IEC 60601-2-37- IEC 60601-2-37IEC 60601-2-37Same to predicate
EMC Compliance- IEC 60601-1-2- IEC 60601-1-2IEC 60601-1-2Same to predicate
Acoustic OutputDisplay Standard- Track 3- Track 3Track 3Same to predicate
BiocompatibilityCompliance- ISO10993-1- ISO10993-1ISO10993-1Same to predicate
SoftwareFunctionality- Q Scan- Q Scan- Q ScanSame to predicate
- ClearVision- ClearVision- ClearVisionSame to predicate
- MultiVision- MultiVision- MultiVisionSame to predicate
- Panoramic- Panoramic- PanoramicSame to predicate
- Needle Mate+- Needle Mate- Needle Mate+Same to predicate
- Auto IMT+- Auto IMT+- Auto IMT+Same to predicate
- Strain+- Strain+Same to predicate
- Stress Echo- Stress EchoSame to predicate
- E-Strain- E-StrainE-StrainSame to predicate
- S-Shearwave- S-ShearwaveSame to predicate
- S-Detect for Breast- S-Detect for BreastS-Detect for BreastSame to predicate
Device nameHS70A(Under Review)The subject deviceHS70A(K173713)Primary PredicateHERA W9/HERA W10(K192319)Second PredicateNote
- S-Detect for Thyroid- S-Detect for Thyroid- S-Detect for ThyroidSame to predicate
- ADVR- ADVR- ADVRSame to predicate
- 3D Imaging (Volume DataAcquisition)- 3D Imaging (Volume DataAcquisition)- 3D Imaging (Volume DataAcquisition)Same to predicate
- 3D Imaging presentation(3D Cine/4D Cine/5D Cine)- 3D Imaging presentation(3D Cine/4D Cine/5D Cine)- 3D Imaging presentation(3D Cine/4D Cine/5D Cine)Same to predicate
-3D Rendering (MPR)-3D Rendering (MPR)-3D Rendering (MPR)Same to predicate
-3D XI (MSV/Oblique View)-3D XI (MSV/Oblique View)-3D XI (MSV/Oblique View)Same to predicate
-3D MXI (Volume Slice,Mirror View)-3D MXI (Volume Slice,Mirror View)-3D MXI (Volume Slice,Mirror View)Same to predicate
- XI Volume CT (VolumeCT)- XI Volume CT (VolumeCT)- XI Volume CT (VolumeCT)Same to predicate
-3D MagiCut-3D MagiCut-3D MagiCutSame to predicate
-Volume Calculation(VOCAL, XI VOCAL)-Volume Calculation(VOCAL, XI VOCAL)-Volume Calculation(VOCAL, XI VOCAL)Same to predicate
- XI STIC- XI STIC- XI STICSame to predicate
- HDVI- HDVI- HDVISame to predicate
- RealisticVue- RealisticVue- RealisticVueSame to predicate
- CEUS+- CEUS+- CEUS+Same to predicate
- 5D Follicle- 5D FollicleSame to predicate
- HDVI- HDVISame to predicate
- 5D NT- 5D NT- 5D NTSame to predicate
- 5D LB- 5D LB- 5D LBSame to predicate
- 5D CNS- 5D CNS- 5D CNSSame to predicate
- Auto NT (Old Name: 2DNT)- Auto NT (Old Name: 2DNT)- Auto NT (Old Name: 2DNT)Same to predicate
- ElastoScan- ElastoScan- ElastoScan+Same to predicate
- E-Thyroid- E-Thyroid- E-ThyroidSame to predicate
- E-Breast- E-Breast- E-BreastSame to predicate
Device nameHS70A(Under Review)The subject deviceHS70A(K173713)Primary PredicateHERA W9/HERA W10(K192319)Second PredicateNote
- ArterialAnalysis- ArterialAnalysisSame to predicate
- IOTA-ADNEX- IOTA-ADNEXSame; refer to SEanalysis 1
- BiometryAssist- BiometryAssistSame; refer to SEanalysis 2
- CrystalVueFlow- CrystalVueFlowSE; refer to SEanalysis 3
- 5D Heart- 5D HeartSE; refer to SEanalysis 4
Transducer- CA1-7A- CA1-7A- CA1-7ASame to predicate
- CA2-8A- CA2-8ASame to predicate
- CF4-9- CF4-9- CF4-9Same to predicate
- E3-12A- E3-12A- E3-12ASame to predicate
- VR5-9- VR5-9Same to predicate
- CA2-9A- CA2-9ASame to predicate
- CA3-10A- CA3-10A- CA3-10ASame to predicate
- EA2-11B- EA2-11B- EA2-11BSame to predicate
- L3-12A- L3-12A- L3-12ASame to predicate
- LA3-16A- LA3-16ASame to predicate
- LA3-16AI- LA3-16AISame to predicate
- LA2-9A- LA2-9A- LA2-9ASame to predicate
- LA4-18B- LA4-18B- LA4-18BSame to predicate
- LM4-15B- LM4-15BSame to predicate
- PE2-4- PE2-4Same to predicate
- PA3-8B- PA3-8B- PA3-8BSame to predicate
- P4-12B- P4-12BSame to predicate
- PA1-5A- PA1-5ASame to predicate
- MMPT3-7- MMPT3-7Same to predicate
- CV1-8A- CV1-8A- CV1-8ASame to predicate
- V5-9- V5-9Same to predicate
Device nameHS70A(Under Review)The subject deviceHS70A(K173713)Primary PredicateHERA W9/HERA W10(K192319)Second PredicateNote
- LV3-14A- LV3-14ASame to predicate
- CW4.0- CW4.0Same to predicate
- CW6.0- CW6.0Same to predicate
- DP2B- DP2BSame to predicate
- DP8B- DP8BSame to predicate
- EV2-10A- EV2-10ASame; refer to SEanalysis 5
CompatibleBiopsy Guide- BP-KIT-029- BP-KIT-029- BP-KIT-029510(k) Exempt
- BP-KIT-030- BP-KIT-030510(k) Exempt
- BP-KIT-041- BP-KIT-041- BP-KIT-041510(k) Exempt
- BP-KIT-043- BP-KIT-043- BP-KIT-043510(k) Exempt
- BP-KIT-053- BP-KIT-053510(k) Exempt
- BP-KIT-054- BP-KIT-054- BP-KIT-054510(k) Exempt
- BP-KIT-055- BP-KIT-055510(k) Exempt
- BP-KIT-057- BP-KIT-057- BP-KIT-057510(k) Exempt
- BP-KIT-058- BP-KIT-058- BP-KIT-058510(k) Exempt
- BP-KIT-059- BP-KIT-059- BP-KIT-059510(k) Exempt
- BP-KIT-060- BP-KIT-060- BP-KIT-060510(k) Exempt
- BP-KIT-065- BP-KIT-065- BP-KIT-065510(k) Exempt
- BP-KIT-066- BP-KIT-066- BP-KIT-066510(k) Exempt
- BP-KIT-068- BP-KIT-068510(k) Exempt
- BP-KIT-069- BP-KIT-069- BP-KIT-069510(k) Exempt
- BP-KIT-071- BP-KIT-071- BP-KIT-071510(k) Exempt
- BP-KIT-075- BP-KIT-075510(k) Exempt
- BP-KIT-083510(k) Exempt
- BP-KIT-085- BP-KIT-085510(k) Exempt
- BP-KIT-086- BP-KIT-086510(k) Exempt
DigitalStorage/Transfer- Foot SwitchECG (defibrillation proof)- Foot SwitchECG (defibrillation proof)- Foot SwitchECG (defibrillation proof)Same to predicate
Device nameHS70A(Under Review)The subject deviceHS70A(K173713)Primary PredicateHERA W9/HERA W10(K192319)Second PredicateNote
Station- Gel Warmer- Clear Track/ Virtual TrackSystem StandWLAN Card- Gel Warmer- Clear Track/ Virtual TrackSystem Stand- WLAN Card- Gel Warmer- Clear Track/ Virtual TrackSystem StandWLAN Card

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the SAMSUNG logo. The logo is in a blue oval shape with the word "SAMSUNG" in white letters. The logo is simple and recognizable.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white letters on a blue oval background. The logo is tilted slightly to the right.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The wordmark is placed inside a blue, oval-shaped background.

{9}------------------------------------------------

Image /page/9/Picture/5 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white letters on a blue oval background. The logo is simple and recognizable.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in a bold, sans-serif font. The word is white and is set against a blue, oval-shaped background. The logo is simple and modern.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the logo is tilted slightly.

510(k) Premarket Notification – Traditional

10. Summary of Non-Clinical Test

The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The HS70A and its applications comply with voluntary standards.

TestStandards and FDA Guidance
Risk ManagementISO 14971 Second edition 2007 Medical devices - Application of risk management to medical devices
Electrical SafetyThe HS70A Ultrasound System with defibrillation-proof ECG electrode was evaluated per the followingstandards.ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012Medical Electrical Equipment - Part 1: General Requirements for basic safety and essential performance.
ElectromagneticCompatibilityIEC60601-1-2: 2014(4th Edition) Medical Electrical Equipment -- Part 1-2: General Requirements ForBasic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances --Requirements And Tests
BiocompatibilityISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation andtesting within a risk management process.
Reprocessing MedicalDevicesFDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labelingissued on March 17, 2015, revised June 9, 2017.
Software/Firmware-driven FunctionalityAll migrated software functionality was evaluated using the same test criteria as the predicates for allapplicable imaging modes to ensure that migration into a new system design did not compromise imagequality with respect to the intended use of each feature.Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued onMay 11, 2005
Ultrasound SafetyMarketing Clearance of Diagnostic Ultrasound Systems and Transducers issued June 27, 2019
IEC60601-2-37:2007 + A1:2015, Particular requirements for the safety of ultrasonic medical diagnostic andmonitoring equipment
NEMA UD 2-2004 (R2009)Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3

{12}------------------------------------------------

510(k) Premarket Notification – Traditional

Image /page/12/Picture/2 description: The image shows the Samsung logo. The logo is a blue oval with the word "SAMSUNG" in white letters inside. The letters are bold and sans-serif. The logo is simple and recognizable.

    1. Summary of Clinical Tests The subject of this premarket submission, HS70A, is not required clinical studies to support substantial equivalence.
    1. Conclusion

SAMSUNG MEDISON CO., LTD. considers the HS70A to be as safe, as effective, and performance is substantially equivalent to the predicate devices.

-END of 510(K) Summary

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)