K Number
K200537
Device Name
Oxxiom
Date Cleared
2020-07-03

(122 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Oxxiom Pulse Oximetry System is a wireless, fully disposable, single-use indicated for measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). It may be used for spot checking, intermittent monitoring, and/or data collection of patients 12 years and older in low acuity settings in facilities such as hospitals, clinics and doctor's offices. It can also be used in home healthcare settings under prescription use. It is not intended for continuous monitoring.

Device Description

Oxxiom Pulse Oximetry System is a wireless, fully disposable, single-use device for measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). lt also measures Pulse Amplitude (PA). It is a prescription device for spot checking, intermittent monitoring, and data collection.

Oxxiom has two main components, the Oxxiom device (hardware device) and the Oxxiom Rx App. Oxxiom also provides tape to apply the device to the finger. Oxxiom also requires the use of an iOS display device, such as an iPhone, iPad, or iPod, that has iOS version 10.1 or later. The iOS device is not provided with the Oxxiom Pulse Oximetry System.

Oxxiom is small, lightweight, and has a 24-hour battery life. The device simultaneously, and wirelessly transfers all collected data through a Bluetooth connection to the Oxxiom App on the iOS device, which analyzes data, displays information, provides warnings as necessary, and stores the information.

Oxxiom has been tested on a subject population with devices placed on the finger. Monitoring at other sites of the body is not recommended. The principle of operation is based upon the noninvasive reflectance of red and infrared light.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Specification)Reported Device Performance
SpO2 Accuracy (Declared Range: 70-100%)
Accuracy: ± 3.5% (Arms)Arms value: 3.4
Pulse Rate Specifications (No Motion)
Accuracy: ± 3 bpmNot explicitly reported for this study, but general specifications for the device state "Accuracy: Patients 12 years and older ± 3 bpm"

2. Sample size used for the test set and the data provenance

  • Sample Size (Clinical Evaluation): 10 subjects
  • Data Provenance: The study was a "comparative, single-center, non-randomized study." While the country of origin isn't explicitly stated, it can be inferred that it likely occurred in the US given the FDA submission. The clinical evaluation describes a prospective study where arterial blood samples were taken to establish a ground truth.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The ground truth for the clinical evaluation was established using CO-Oximetry on arterial blood samples. This is an objective measurement from a medical device, rather than a subjective assessment by human experts. Therefore, no information is provided regarding the number or qualifications of experts for establishing ground truth in this specific context.

4. Adjudication method for the test set

Not applicable. The ground truth was established by CO-Oximetry, which is an objective measurement, not requiring an adjudication method like 2+1 or 3+1 by human experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document describes a clinical evaluation of a pulse oximetry system, not an AI-assisted diagnostic tool. No MRMC study was conducted, nor is there any mention of AI assistance for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the clinical evaluation described is a standalone performance test of the device (Oxxiom Pulse Oximetry System) against a reference standard (CO-Oximetry). It assesses the device's ability to accurately measure SpO2 without human intervention in the measurement process itself, beyond initial setup.

7. The type of ground truth used

The ground truth used for the clinical evaluation was arterial blood samples assessed by CO-Oximetry (SaO2). This is a highly accurate and objective method for determining arterial oxygen saturation.

8. The sample size for the training set

The document does not provide information about a "training set" in the context of machine learning or AI. The described studies are primarily for performance validation of the measurement technology itself, not for training a predictive model.

9. How the ground truth for the training set was established

Not applicable, as no training set (for machine learning/AI) is described in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

True Wearables, Inc. % Amy Fowler Regulatory Counsel Gardner Law, PLLC 423 Main Street Stillwater, Minnesota 55082

Re: K200537

Trade/Device Name: Oxxiom Pulse Oximetry System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: June 2, 2020 Received: June 4, 2020

Dear Amy Fowler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200537

Device Name Oxxiom Pulse Oximetry System

Indications for Use (Describe)

The Oxxiom Pulse Oximetry System is a wireless, fully disposable, single-use indicated for measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). It may be used for spot checking, intermittent monitoring, and/or data collection of patients 12 years and older in low acuity settings in facilities such as hospitals, clinics and doctor's offices. It can also be used in home healthcare settings under prescription use. It is not intended for continuous monitoring.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for True Wearables. The logo consists of a black circle with a white lowercase "t" inside, followed by the words "true wearables" in black, bold font. The "TM" symbol is present twice in the logo, once above the "t" in the circle and once after the word "wearables".

510(k) Summary

According to 21 CFR 807.92

Date Prepared: June 2, 2020

Submitter Information

Submitter:True Wearables, Incorporated
29826 Avenida de Las Banderas, Suite 300
Rancho Santa Margarita, CA 92688
  • Manufacturer: True Wearables, Incorporated 29826 Avenida de Las Banderas, Suite 300 Rancho Santa Margarita, CA 92688
    Establishment Registration Number: 3012234356
Contact:Amy Fowler, Gardner LawRegulatory Counsel for True Wearables, Inc.423 Main Street SouthStillwater, MN 55082
Phone:612-356-9653 (Mobile)949-709-0850 (True Wearables, Inc.)
Fax:NA
E-mail:afowler@gardner.law
Alternate Contact:Marcelo Lamego
CEO, True Wearables, Inc
Phone:949-709-0850
Fax:NA
E-mail:mmlamego@truewearables.com

Name of Device and Classification

Trade Name:Oxxiom® Pulse Oximetry System
Common Name:Pulse Oximetry System
Device Classification:Oximeter, CFR §870.2700, Class II
Product Code(s):DQA

Predicate Device

Predicate:Guardian Angel GA1000 Digital Vital Sign Monitoring System; K16258
--------------------------------------------------------------------------------

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for True Wearables. The logo consists of a black circle with a white lowercase "t" inside, followed by the words "true wearables" in black, bold font. The "TM" symbol is present twice in the logo, once above the "t" in the circle and once after the word "wearables".

True Wearables, Inc. 29826 Avenida de Las Banderas Rancho Santa Margarita CA 92688 (949) 709-0850 www.truewearables.com

Intended Use/Indications for Use

The Oxxiom Pulse Oximetry System is a wireless, fully disposable, single-use device indicated for measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). It may be used for spot checking, intermittent monitoring, and/or data collection of patients 12 years and older in low acuity settings in facilities such as hospitals, clinics, and doctor's offices. It can also be used in home healthcare settings under prescription use. It is not intended for continuous monitoring.

Device Description

Oxxiom Pulse Oximetry System is a wireless, fully disposable, single-use device for measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). lt also measures Pulse Amplitude (PA). It is a prescription device for spot checking, intermittent monitoring, and data collection.

Oxxiom has two main components, the Oxxiom device (hardware device) and the Oxxiom Rx App. Oxxiom also provides tape to apply the device to the finger. Oxxiom also requires the use of an iOS display device, such as an iPhone, iPad, or iPod, that has iOS version 10.1 or later. The iOS device is not provided with the Oxxiom Pulse Oximetry System.

Oxxiom is small, lightweight, and has a 24-hour battery life. The device simultaneously, and wirelessly transfers all collected data through a Bluetooth connection to the Oxxiom App on the iOS device, which analyzes data, displays information, provides warnings as necessary, and stores the information.

Oxxiom has been tested on a subject population with devices placed on the finger. Monitoring at other sites of the body is not recommended. The principle of operation is based upon the noninvasive reflectance of red and infrared light.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for True Wearables. The logo consists of a lowercase "t" inside of a black circle, followed by the words "true wearables" in black font. The trademark symbol is located to the upper right of both the "t" and the word "wearables". The number "K200537" is located in the upper right corner of the image.

True Wearables, Inc. 29826 Avenida de Las Banderas Rancho Santa Margarita CA 92688 (949) 709-0850 www.truewearables.com

Summary of Comparison of Intended Use and Technological Characteristic

Oxxiom Pulse Oximetry System is compared below to the legally marketed predicate device, Guardian Angel GA1000 Digital Vital Sign Monitoring System K162580

CharacteristicSubject DevicePredicate Device
Product NameOxxiom Pulse Oximetry SystemGuardian Angel GA1000 DigitalVital Sign Monitoring System
ManufacturerTrue Wearables, IncTaiwan Aulisa Medical DevicesTechnologies, Inc
RegulationOximeter CFR §870.2700Pulse OximeterOximeter CFR §870.2700Pulse Oximeter
ProductCode(s)DQADQA
Device ClassClass IIClass II
Indications forUseThe Oxxiom® Pulse Oximetry System is awireless, fully disposable, single-usedevice indicated for measuring,displaying, and storing functionaloxygen saturation of arterialhemoglobin (SpO2) and pulse rate (PR).It may be used for spot checking,intermittent monitoring, and/or datacollection of patients 12 years and olderin low acuity settings in facilities such ashospitals, clinics, and doctor's offices. Itcan also be used in home healthcaresettings under prescription use. It is notintended for continuous monitoring.The Guardian Angel GA1000 DigitalVital Sign Monitoring System isindicated for use in measuring anddisplaying functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate of adult andpediatric patients. It is indicated forspot-checking and / or continuousmonitoring of patients during non-motion and under well-perfusedconditions. The intendedenvironment of use is hospital. Thissystem is a reusable device.
Type of UseSpot checkingIntermittent monitoringData collectionSpot checkingContinuous use
MotionNon-motionNon-motion
PatientPopulationPatients 12 years and olderAdultPediatric
PerfusionWellWell
Environmentof UseHospitals, clinics, doctor's offices.Home (under prescription use)Hospitals
CharacteristicSubject DevicePredicate Device
DisposableYesNo
Application SiteFingerFinger
TechnologyRed and Infrared reflectance technologyRed and Infrared transmittance technology
System ComponentsWireless, disposable oximeter device and software App. User to supply iOS deviceSensor Module and Display Unit
Wireless Technology / Data TransmissionBluetooth V4.0Bluetooth V4.0
DisplayNon-dedicated iOS Device DisplayDedicated Display
Software Level of ConcernModerateModerate
SpO2 Specifications (No Motion)Displayed Range: 1-100%Displayed Range: 1-100%
Declared Range: 70-100%Declared Range: 70-100%
Accuracy: Patients 12 years and older± 3.5%Accuracy: Adults/Pediatrics ± 3%
Pulse Rate Specifications (No Motion) ControlsDisplayed Range: 25-250 bpmDisplayed Range: 30-290 bpm
Accuracy: Patients 12 years and older± 3 bpmAccuracy: Adults/Pediatrics ± 3 bpm
Measurement Wavelengths and Output PowerRed 655 nmInfrared 940 nmRed 660 nm @ 1.8 mw nominalInfrared 905 nm @ 2 mw nominal
CharacteristicSubject DevicePredicate Device
BatteryWireless Disposable Oximeter: Nonrechargeable Lithium ManganeseDioxide (LiMnO2), 150 mAh, 3 VSensor Module: 3.7 V Lithium BatteryDisplay Unit: Lithium Battery, ACAdaptor
AlarmAudible and visual pulse rate, oxygensaturation, low battery, out of range,and off patient alarmsAudible and visual pulse rate andoxygen saturation alarms.Low battery and critical battery,disconnection alarms

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for "true wearables™". The logo consists of a lowercase "t" inside of a black circle, followed by the words "true wearables" in black font. The trademark symbol is present next to both the "t" and the word "wearables". The number "K200537" is in the upper right corner of the image.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for True Wearables. The logo consists of a black circle with a white lowercase "t" inside, followed by the words "true wearables" in black, with a "TM" symbol in the upper right corner of both the circle and the word "wearables". The letters are bold and sans-serif.

Comparative analysis of the indication statements between the subject and predicate devices show both similarities and differences. The core indications of measuring and displaying SpO2 and PR is the same for both devices. The patient population for both devices is for adult and pediatric patient populations, but Oxxiom provides age 12 as a specific age limit. Both devices are for prescription use and for use in hospitals, but Oxxiom is indicated for additional environments, including clinics, doctor's offices and home use. While the predicate device is for continuous monitoring and spot checking, the subject device is indicated for low acuity setting spot checking/intermittent use. These differences in indications for use are all within the same intended use and support substantial equivalence. These differences do not affect the safety or efficacy of the subject device when used according to its labeling addresses these differences and provide appropriate instructions for use.

There are some differences in technological characteristics, including the feature of disposable versus reusable and iOS nondedicated display versus dedicated display that do not present different questions of safety and efficacy. Comparison of the similarities and differences between the proposed device and the predicate do not raise new or different questions of safety and effectiveness, when compared to the predicated devices and therefore support the proposed device as substantially equivalent to the predicate devices.

Summary of Performance Testing

Oxxiom Pulse Oximetry System is supported by electromagnetic compatibility, software validation, biocompatibility, clinical testing and other device performance tests to ensure appropriate functionality and demonstrate substantial equivalence to the predicate device. FDA Guidance – Pulse Oximeters – Premarket Notification Submissions [510(k)s] was consulted in determining necessary testing. Software verification and validation testing was done according to FDA Guidance – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for True Wearables. The logo consists of a lowercase "t" inside of a black circle, followed by the words "true wearables" in black, sans-serif font. The letters "TM" are written in superscript next to the "t" and "wearables".

Nonclinical and Clinical Functional and Safety Testing

The following functional and safety testing has been conducted for the OXXIOM device, performed according to the OXXIOM device product requirement specification and quality system to illustrate substantial equivalence.

TestReferenceResult
Electrical SafetyIEC 60601-1IEC 60601-1-2Pass
Clinical Performance TestingISO 80601-2-61:2011Pass
Biocompatibility TestingISO 10993-10:2010ISO 10993-2ISO 10993-5ISO 10993-10ISO 10993-12Pass
EMC TestingIEC 60601-1-2Pass
Alarm TestingIEC 60601-1-8Pass
FCC Testing andCertificationsFCC Part 15 Subpart B Class BFCC Part 15 Subpart CPass
Wireless Coexistence TestingFDA Guidance Radio FrequencyWireless Technology in MedicalDevice (2013)Pass
Software Verification andValidationIEC 62304:FDA Guidance—Guidance forthe Content of PremarketSubmissions for SoftwareContained in Medical DevicesPass

Additional tests performed on the Oxxiom Pulse Oximetry System include the following: Non-standard testing:

    1. Battery capacity/life: Test results for energy capacity and voltage limits verification of the Oxxiom's Li-MnO2 Battery.
    1. Drop testing: Performance/functionality of the Oxxiom pulse oximeter, after being dropped from various heights.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for True Wearables. The logo consists of a lowercase "t" inside of a black circle, followed by the words "true wearables" in black font. The "TM" symbol is located to the upper right of both the circle and the word "wearables".

    1. Shipping conditions: Demonstrate the ability of the Oxxiom pulse oximeter packaging to protect shipped items from foreseeable stresses encountered during shipping.
    1. Shelf Life: Validate the ability of Oxxiom pulse oximeter to withstand storage conditions for 12 months by using an accelerated shelf life study method.
    1. Environmental: Environmental temperature and humidity testing performed on the Oxxiom pulse oximeter, in order to verify and validate the ability of Oxxiom to withstand operational conditions from 5°C - 40°C and 5% - 95% RH.
    1. Water Resistance: Demonstrate the ability of the Oxxiom pulse oximeter to withstand water immersion.
    1. Skin Surface Temperature: Demonstrate the Oxxiom pulse oximeter meets skin surface temperature requirements.
    1. Ambient Light Testing: Demonstrate Oxxiom's performance under different ambient light conditions (incandescent and fluorescent lights, and sunlight light).
    1. Silence, Airplane, Guided Access Modes: Demonstrate Oxxiom operates as intended under the various iOS modes.
    1. Dimensions: Verify Oxxiom's dimensions, weight, outer edges (no sharp edges), and sterility (No visible particles).
    1. Display Range and Waveforms: Oxxiom Pulse Oximetry meets specifications in terms if display ranges (i.e., SpO2: 1-100%, PR: 25-250 BPM, PA: 0.03-20%, Waveforms: graph with auto-adjustable scale).

Clinical Evaluation

A comparative, single-center, non-randomized study was conducted to evaluate SpO2 accuracy and performance of the sensor placed on fingers during non-motion conditions over the range of 70-100% SaO2 arterial blood samples assessed by CO-Oximetry. Arterial blood sampling was used as the basis for comparison.

The demographics of subjects in this study included five males and five females. They range by age (21-37 years), weight (134-230 lbs.), height (62-74"), and BMI (20.7 – 35). For race and ethnicity, the subject pool included two Black/African-American, one Asian, and seven White subjects. One of the ten subjects was of Hispanic/Latino ethnicity and the other nine were not.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the alphanumeric string "K200537" in a simple, sans-serif font. The text is black against a white background. The characters are evenly spaced and clearly legible.

Image /page/10/Picture/1 description: The image shows the logo for True Wearables. The logo consists of a black circle with a white lowercase "t" inside, followed by the words "true wearables" in black, bold font. The letters "TM" are written in superscript next to the circle and after the word "wearables".

True Wearables, Inc. 29826 Avenida de Las Banderas Rancho Santa Margarita CA 92688 (949) 709-0850 www.truewearables.com

The skin pigmentation tones ranged from light to dark, meeting the protocol requirements of having at least 2 (15%) of subjects be of dark pigmented skin.

The Accuracy root mean square (Arms) between measured SpO2 and reference SaO2 met the 3.5% specification for the sensor. Arms is based on statistically distributed measurements, therefore, a sensor/oximeter with n Arms specification of 3.5 is expected to have approximately 68% of the data points fall within that range. The SpO2 performance accuracy evaluated over the SaO2 range of 70-100% under non-motion conditions was shown to have an Arms value of 3.4. This study provides supporting evidence that the SpO2 accuracy performance meets an acceptable specification for the sensor passing an Arms specification of 3.5 under steady state/non-motion conditions over the range of 70-100%.

Conclusion

In this 510(k), the low-risk intended use/indications for use statement and technological characteristics of the Oxxiom Pulse Oximetry System are demonstrated to be substantially equivalent to the Guardian Angel GA1000 Digital Vital Sign Monitoring System. Substantial equivalence is based upon the comparison of labeling, clinical data, and nonclinical/bench testing data. The difference in technological features of the subject device compared to the predicate does not raise different questions of safety and effectiveness. No animal testing was conducted.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).