K Number
K200533

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-04-27

(56 days)

Product Code
Regulation Number
866.3393
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The binx health io CT/NG Assay, when tested using the binx health io Instrument, is a fully automated, rapid, qualitative test intended for use in point-of-care or clinical laboratory settings for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA by polymerase chain reaction. The binx health io CT/NG Assay is intended for use with female vaginal swab specimens, collected either by a clinician or self-collected by a patient in a clinical setting, or male urine specimens, as an aid in the diagnosis of symptomatic or asymptomatic Chlamydia trachomatis and/or Neisseria gonorrhoeae infection. For a symptomatic male patient with a chlamydia negative test result, further testing with a laboratory-based molecular test is recommended.

Device Description

The binx health io CT/NG Assay System (the "binx io System", "binx io CT/NG Assay" or "the System") is a rapid qualitative in vitro diagnostic system consisting of the following:

  1. The binx io Instrument for running the Cartridge (the "Instrument")
  2. The binx io CT/NG Cartridge (the "CT/NG Cartridge", "Cartridge" or "Cartridges"), that contains all the necessary reagents to perform the binx io CT/NG Assay (the "Assay") on the binx io Instrument
  3. A single-use, fixed-volume transfer pipet (packaged with the Cartridge) for transferring the sample to the Cartridge
  4. A Male Urine Collection Kit consisting of sample Collection tube containing preservation medium and a transfer pipet (the "Male Urine Collection Kit")

All reagents are contained in the Cartridge as a combination of liquid reagents in blister packs and dried reagents. The Instrument is a small, desk top, fully-integrated instrument that uses air pressure to open and close valves on the binx io CT/NG Cartridge which, in turn, controls the movement of solutions within the Cartridge; the Instrument takes full control of the Cartridges once they are inserted, and no further user interaction is required. The operation of the Instrument is designed to be simple and intuitive; a user follows simple instructions on the graphical user interface (GUI) screen to load the Cartridge onto the Instrument with no further interaction required.

The Male Urine Collection Kit consists of a tube containing a small volume of preservative medium and a Urine Transfer Pipet. To obtain a sample suitable for use on the CT/NG Cartridge, a 20-30 mL first-catch urine sample is collected by a patient in a suitable sterile vessel. The Urine Transfer Pipet provided with the Male Urine Collection Kit is used to transfer a sufficient volume of urine (nominally 4 mL in 2x 2 mL steps) into the collection tube such that the total volume falls between the two indicated lines on the collection tube.

Once a sample has been correctly collected, the required sample volume (0.5 mL) is transferred from the sample collection tube to the Cartridge using the Sample Transfer Pipet provided with the Cartridge. The Cartridge has a visual sample addition indicator window which turns from light to dark to show the user that a sample has been added to the Cartridge.

The Cartridge has three main Assay steps: sample preparation to isolate and purify target DNA; ultra-rapid polymerase chain reaction (PCR), which amplifies specific regions of DNA from the target organisms; and a proprietary electrochemical detection method to identify the presence of amplified DNA.

When the specimen is added to the Cartridge, it is automatically mixed with a lysis solution to disrupt the cells present and release DNA which also rehydrates the Internal Process Control (IPC) sample. DNA extraction takes place and the eluted DNA is transferred to a homogenization chamber.

The DNA in solution is transferred into two separate amplification chambers and reconstitutes the dried PCR reagents as it enters the chambers which are located over the Instrumentcontrolled PCR heater. Ultra-rapid PCR is carried out using sequence-specific primers for CT, NG (two separate genomic targets) and the IPC.

Following amplification, the amplified target DNA is transferred from each PCR chamber into two separate detection chambers (four detection chambers in total) which contain a carbonbased screen-printed electrode. When the target is present and amplified, the target-specific probes and amplicon hybridize. The electrochemical labels are cleaved using a double-strand specific exonuclease. The cleaved electrochemical label diffuses to the electrode surface generating an electrical current that can be measured at a distinct voltage in nano Amps (nA) for each electrochemical label used.

The presence of a measurable peak to a fixed cut-off parameter for each target returns a qualitative result without the need for any user interpretation or calculations.

AI/ML Overview

The binx health io CT/NG Assay and System is a fully-automated, rapid, qualitative test for the detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) DNA. The following information details its acceptance criteria and the study that demonstrated its performance.

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance (Male Urine) - Total
Chlamydia trachomatis (CT)
SensitivityNot explicitly stated, implied to be high for regulatory clearance92.5% (95% CI: 86.4% - 96.0%)
SpecificityNot explicitly stated, implied to be high for regulatory clearance99.3% (95% CI: 98.4% - 99.7%)
Neisseria gonorrhoeae (NG)
SensitivityNot explicitly stated, implied to be high for regulatory clearance97.3% (95% CI: 90.7% - 99.3%)
SpecificityNot explicitly stated, implied to be high for regulatory clearance100.0% (95% CI: 99.5% - 100.0%)

Note: The document provides confidence intervals for sensitivity and specificity, indicating high performance was expected for regulatory approval.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 922 fully evaluable male urine specimens.
  • Data Provenance: Prospective, multi-center study conducted at nine investigational sites throughout the U.S.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number or specific qualifications of experts used for establishing ground truth for the test set. However, the ground truth (Composite Infected Status - CIS) was established using a concordance of three legally marketed predicate devices, run on validated laboratory systems. These predicate devices are themselves validated diagnostic tools, implying a high level of accuracy typically overseen by qualified laboratory personnel.

4. Adjudication Method for the Test Set

The adjudication method used to establish the Composite Infected Status (CIS) for the test set was defined as follows:

  • A patient was considered infected if at least two out of the three reference tests were positive.
  • A patient was considered not infected if at least two out of the three reference tests were negative.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The study evaluated the standalone performance of the binx io CT/NG Assay against a Composite Infected Status derived from multiple reference NAATs, rather than assessing human reader improvement with AI assistance.

6. Standalone Performance Study

Yes, a standalone performance study was conducted. The clinical performance data presented (sensitivity and specificity) are for the binx io CT/NG Assay and System operating independently, without a human-in-the-loop for interpretive decisions, other than initial sample collection and loading. The system provides qualitative results (e.g., "CT Detected") without user interpretation.

7. Type of Ground Truth Used

The type of ground truth used was a Composite Infected Status (CIS). This CIS was established by comparing the results of the binx io CT/NG Assay to three legally marketed and clinically validated reference nucleic acid amplification tests (NAATs):

  • Hologic Aptima Combo 2 (AC2) Chlamydia/Gonorrhea Assay run on Panther
  • BD ProbeTec Chlamydia trachomatis (CT) Q* and BD ProbeTec Neisseria gonorrhoeae (GC) Q* assays run on the Viper XTR™
  • Roche cobas CT/NG v2.0 test run on the cobas 4800 System

8. Sample Size for the Training Set

The document does not explicitly provide the sample size for a separate "training set" in the context of clinical performance evaluation. The clinical study described is a prospective validation study against a defined ground truth for market clearance purposes. Analytical studies (e.g., LoD, inclusivity, exclusivity) generally use controlled laboratory samples for method development and validation, which can be considered analogous to "training" or "development" data in an algorithm context, but a distinct clinical training set size is not reported.

9. How the Ground Truth for the Training Set Was Established

Given that a specific clinical "training set" with established ground truth is not detailed, the ground truth for the analytical performance studies (which would inform development) was established through controlled laboratory methods. For example:

  • Limit of Detection (LoD): Determined using cellular CT and NG material with quantified genome equivalents (GE/mL).
  • Analytical Reactivity (Inclusivity/Exclusivity): Evaluated against known concentrations of various CT serovars, NG strains, and panels of potentially cross-reactive microorganisms or human genomic DNA. These "ground truths" are based on the known identity and concentration of the spiked organisms or substances.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 27, 2020

binx health Inc Sarah Kalil Advisor 77 North Washington Street, 5th Floor Boston, Massachusetts 02114

Re: K200533

Trade/Device Name: binx io CT/NG Assay and binx io CT/NG System Regulation Number: 21 CFR 866.3393 Regulation Name: Nucleic Acid Detection System For Non-Viral Microorganism(S) Causing Sexually Transmitted Infections Regulatory Class: Class II Product Code: QEP, LSL, MKZ, NSU Dated: March 2, 2020 Received: March 2, 2020

Dear Sarah Kalil:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Steven Gitterman, M.D., Ph.D. Deputy Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) SUMMARY

SUBMITTER NAME:binx health, Inc.
SUBMITTER ADDRESS:77 N. Washington Street,5th FloorBoston, MA 02114USA
CONTACT PERSON:Sarah KalilRegulatory Advisorbinx health, Inc.Phone: (646) 847 8573Email: sarah.kalil@mybinxhealth.com
DATE PREPARED:March 2, 2020
DEVICE TRADE NAME:binx io CT/NG Assay
CLASSIFICATION:21 C.F.R. § 866.3393 (Nucleic acid detection system fornon-viral microorganism(s) causing sexually transmittedinfections)Class II
PRODUCT CODE:QEP
SUBSEQUENTCODES:PRODUCT NSU
REVIEW PANELMicrobiology
REQUESTED CLIACLASSIFICATION:Moderate Complexity
PREDICATE DEVICES:binx health io CT/NG Assay for female vaginal swab(K191352)
INTENDED USE:The binx health io CT/NG Assay, when tested using thebinx health io Instrument, is a fully automated, rapid,qualitative test intended for use in point-of-care or clinicallaboratory settings for the detection of Chlamydiatrachomatis and Neisseria gonorrhoeae DNA bypolymerase chain reaction. The binx health io CT/NGAssay is intended for use with female vaginal swabspecimens, collected either by a clinician or self-collectedby a patient in a clinical setting, or male urine specimens,as an aid in the diagnosis of symptomatic or asymptomaticChlamydia trachomatis and/or Neisseria gonorrhoeaeinfection. For a symptomatic male patient with a chlamydianegative test result, further testing with a laboratory-basedmolecular test is recommended.

{3}------------------------------------------------

Device Description:

The binx health io CT/NG Assay System (the "binx io System", "binx io CT/NG Assay" or "the System") is a rapid qualitative in vitro diagnostic system consisting of the following:

    1. The binx io Instrument for running the Cartridge (the "Instrument")
    1. The binx io CT/NG Cartridge (the "CT/NG Cartridge", "Cartridge" or "Cartridges"), that contains all the necessary reagents to perform the binx io CT/NG Assay (the "Assay") on the binx io Instrument
    1. A single-use, fixed-volume transfer pipet (packaged with the Cartridge) for transferring the sample to the Cartridge
    1. A Male Urine Collection Kit consisting of sample Collection tube containing preservation medium and a transfer pipet (the "Male Urine Collection Kit")

All reagents are contained in the Cartridge as a combination of liquid reagents in blister packs and dried reagents. The Instrument is a small, desk top, fully-integrated instrument that uses air pressure to open and close valves on the binx io CT/NG Cartridge which, in turn, controls the movement of solutions within the Cartridge; the Instrument takes full control of the Cartridges once they are inserted, and no further user interaction is required. The operation of the Instrument is designed to be simple and intuitive; a user follows simple instructions on the graphical user interface (GUI) screen to load the Cartridge onto the Instrument with no further interaction required.

The Male Urine Collection Kit consists of a tube containing a small volume of preservative medium and a Urine Transfer Pipet. To obtain a sample suitable for use on the CT/NG Cartridge, a 20-30 mL first-catch urine sample is collected by a patient in a suitable sterile vessel. The Urine Transfer Pipet provided with the Male Urine Collection Kit is used to transfer a sufficient volume of urine (nominally 4 mL in 2x 2 mL steps) into the collection tube such that the total volume falls between the two indicated lines on the collection tube.

Once a sample has been correctly collected, the required sample volume (0.5 mL) is transferred from the sample collection tube to the Cartridge using the Sample Transfer Pipet provided with the Cartridge. The Cartridge has a visual sample addition indicator window which turns from light to dark to show the user that a sample has been added to the Cartridge.

The Cartridge has three main Assay steps: sample preparation to isolate and purify target DNA; ultra-rapid polymerase chain reaction (PCR), which amplifies specific regions of DNA from the target organisms; and a proprietary electrochemical detection method to identify the presence of amplified DNA.

When the specimen is added to the Cartridge, it is automatically mixed with a lysis solution to disrupt the cells present and release DNA which also rehydrates the Internal Process Control (IPC) sample. DNA extraction takes place and the eluted DNA is transferred to a homogenization chamber.

The DNA in solution is transferred into two separate amplification chambers and reconstitutes the dried PCR reagents as it enters the chambers which are located over the Instrumentcontrolled PCR heater. Ultra-rapid PCR is carried out using sequence-specific primers for CT, NG (two separate genomic targets) and the IPC.

Following amplification, the amplified target DNA is transferred from each PCR chamber into two separate detection chambers (four detection chambers in total) which contain a carbonbased screen-printed electrode. When the target is present and amplified, the target-specific

{4}------------------------------------------------

probes and amplicon hybridize. The electrochemical labels are cleaved using a double-strand specific exonuclease. The cleaved electrochemical label diffuses to the electrode surface generating an electrical current that can be measured at a distinct voltage in nano Amps (nA) for each electrochemical label used.

The presence of a measurable peak to a fixed cut-off parameter for each target returns a qualitative result without the need for any user interpretation or calculations.

INTERNAL PROCESS CONTROL

The Assay incorporates a positive IPC which is processed along with a patient sample and therefore is exposed to the same testing steps as the sample from DNA extraction and purification through to detection.

The IPC verifies all aspects of the Assay process have functioned as expected. In an Assay where CT and/or NG is not detected, the IPC is measured by the Instrument to ensure it is within an acceptable range to validate a negative result. If it is outside the acceptable range the binx io Instrument will return an "Assay Invalid" message and no result will be displayed or recorded against that specimen. If it is within the acceptable range the "CT Not Detected" and/or "NG Not Detected" result will be displayed and recorded by the Instrument.

ASSAY OUTCOMES

Qualitative results are provided to the user in text format only. Assay results are displayed with the Specimen ID and Assay type. To maintain patient confidentiality, the Patient ID (if one has been entered) will not be displayed on the same screen as the Assay result.

The results shown below are the only results the Instrument will return following completion of a test.

Assay ResultInterpretation of Assay Result
CT Not DetectedChlamydia trachomatis target DNA was not detected in the patientspecimen.The IPC passed specification. This is a valid result
CT DetectedChlamydia trachomatis target DNA was detected in the patientspecimen.This is a valid result.
NG Not DetectedAt least one of the DNA targets for Neisseria gonorrhoeae was notdetected in the patient specimen.The IPC passed specification. This is a valid result.
NG DetectedBoth DNA targets for Neisseria gonorrhoeae were detected in thepatient specimen.This is a valid result.
Assay InvalidThe presence or absence of the DNA targets for Chlamydiatrachomatis and/or Neisseria gonorrhoeae could not be ascertainedin the patient specimen.If a CT or NG Not Detected result is obtained, the IPC was outside theacceptable range indicating a failure in the Assay process.

{5}------------------------------------------------

The Assay should be repeated using the same patient specimen.
User AbortedA user cancelled the Assay. No result is given.
ErrorAn internal fault occurred that terminated the Assay before it finished.

PERFORMANCE EVALUATION

ANALYTICAL PERFORMANCE

Analytical testing was performed to evaluate the performance of the binx io CT/NG Assay using the following studies:

ANALYTICAL SENSITIVITY - LIMIT OF DETECTION

Studies were carried out to determine the analytical limit of detection (LoD) of the binx io CT/NG Assay using cellular CT and NG material, for which the genome equivalents (GE)/mL were quantified. Two cartridge lots were used for each estimate of LoD to enable a lot to lot reproducibility comparison.

At least five separate input concentrations were used to cover a wide range (0.01-99%) of detection rates and each input concentration was tested with at least 20 replicates. A probit regression analysis was used to model the 'CT Detected' rate and identify the concentration level that demonstrated a detection rate of 95%. The LoD was then verified for each CT serovar and each NG strain, using a further total of 40 Cartridges per serovar/strain per Cartridge lot using two further preparations of the claimed LoD generated by two different operators. The LoD for each CT serovar and NG strain was set as the highest value generated from the two reagent lots and of the two tested serovars/strains.

OrganismGE/mLIFU/mLCFU/mL
CT serovar E (ATCC-VR-348B)485.36.6N/A
CT serovar F (ATCC-VR-346)769.30.3N/A
NG strain ATCC 49226125.6N/A1.1
NG strain ATCC 700825212.3N/A2.5

ANALYTICAL REACTIVITY (INCLUSIVITY)

Analytical reactivity of additional CT serovars and NG strains was evaluated in the LoD studies described above.

CT serovars A, B, Ba, C, D, I, J, L2, nvCT were detected at 384.7 GE/mL. Serovars G, H, K, L1, L3 were detected at 769.3 GE/mL in ≥19/20 replicates.

Thirty additional NG strains (including two fluoroquinolone resistant isolates) and the reported detectable level was confirmed by testing three replicates at or near the LoD. Sixteen strains were detected at 212.3 GE/mL in three out of three replicates. Thirteen strains were detected at 1,061.0 GE/mL in ≥19/20 replicates. The remaining strain (NG California 201304 #1) was detected in 17/20 replicates and was therefore subjected to a Probit LoD study which yielded an LoD of 553.38 GE/mL, confirmed by a verification study (40/40 replicates detected).

{6}------------------------------------------------

ANALYTICAL REACTIVITY (EXCLUSIVITY)

A panel of 62 species was investigated for cross-reactivity. A panel of microorganisms and H. sapiens were assessed using cultured organisms at a concentration of 1 x 10º CFU/mL for bacteria or 1 x 10 PFU/mL for viruses, or at a concentration of 2 ng/mL of genomic DNA generated by reverse transcription as available. Two further species were evaluated in silico by bioinformatic analysis of the genetic targets used in the binx health io CT/NG Assay against the published genome sequences for these organisms. In silico analysis concluded that neither of these organisms would be detected by the Assay.

All isolates were reported as CT Not Detected/NG Not Detected with the exception of Corynebacterium xerosis which gave a single NG positive result and Herpes Simplex virus 2 which gave a single CT positive result from 20 replicates in male urine and may therefore be cross-reactive with the NG and CT analytes, respectively.

Bacteriodes fragilis*Neisseria meningitidis Serogroup D*
Bacteriodes ureolyticus*Neisseria meningitidis Serogroup W135*
Clostridium perfringens*Neisseria meningitidis Serogroup Y*
Corynebacterium genitaliumNeisseria cinerea
Corynebacterium xerosisNeisseria denitrificans
Escherichia coliNeisseria elongata (4)
Gardnerella vaginalis*Neisseria gonorrhoeae*
Haemophilus ducreyi*Neisseria flava
Herpes simplex virus 1*Neisseria flavescens (3)
Homo sapiens*Neisseria lactamica (3)
Human papilloma virus 16*Neisseria mucosa (4)
Kingella dentrificansNeisseria perflava (2)
Kingella kingaeNeisseria polysaccharea
Lactobacillus acidophilusNeisseria sicca (4) *
Lactobacillus brevisNeisseria subflava (2)
Lactobacillus jenseniiTrichomonas vaginalis
Lactobacillus lactisUreaplasma urealyticum*
Moraxella lacunataUreaplasma parvum*
Staphylococcus epidermidisAtopobium vaginae*
Streptococcus agalactiaeBifidobacterium longum*
Candida albicansBVAB-2†
Candida glabrataEnterococcus faecalis
Candida parapsilosisHerpes Simplex Virus 2*
Chlamydia pneumoniae*Klebsiella pneumoniae
Chlamydia psittaci*Megasphaera type 1†
Mycoplasma genitalium*Mobiluncus curtisii*
Mycoplasma hominis*Mobiluncus mulieris*
Neisseria meningitidis Serogroup A*Peptostreptococcus anaerobius*
Neisseria meningitidis Serogroup B*Proteus mirabilis
Neisseria meningitidis Serogroup C*Pseudomonas aeruginosa
-Staphylococcus aureus
-Chlamydia trachomatis*

Microorganisms tested in the binx io CT/NG Assay

(n) number of strains tested

*Organisms tested with genomic DNA (2 ng/mL)

† In silico analysis

{7}------------------------------------------------

ANALYTICAL SPECIFICITY - INTERFERENCE

The analytical performance of the io CT/NG Assay was evaluated in the presence of a panel of potentially interfering substances that may be found in male urine specimens. The substances were diluted to the concentrations shown in the table below and spiked into negative pooled male urine matrix. The substances were tested in the absence of CT and NG (negative) and at 2x LoD of both CT serovar F (ATCC VR-346) and NG strain ATCC 49226. No interference was observed with the exception of one substance, leukocytes, which produced one false negative result for Neisseria gonorrhoeae out of 20 replicates tested.

Interfering SubstanceConcentration / Details
Human blood1% (v/v)
Seminal fluid5% (v/v)
Mucin0.5% (w/v)
Albumin (BSA)10 mg/mL
Glucose10 mg/mL
Bilirubin0.2 mg/mL
Leukocytes1x106 cells/mL
Progesterone7 mg/mL
β-Estradiol0.25% (v/v) (Estracecream)
Paracetamol3.2 mg/mL
Aspirin40 mg/mL
Acidic UrinepH 4.0
Alkaline UrinepH 9.0
Azithromycin0.8 mg/mL
Doxycycline3.6 mg/mL

Interfering substances tested in the binx io CT/NG Assay with male urine

Microbial Interference

The performance of the CT/NG Assay was evaluated when 2x LoD of both CT serovar F (ATCC VR-346) and NG strain ATCC 49226 were spiked into negative pooled male urine matrix, aliquots of which were subsequently spiked with a panel of ten microorganisms at a concentration of 1 x 105 CFU/mL. No interference was observed and an expected result of CT, NG Detected was obtained in all cases.

Panel of organisms used for microbial interference testing with male urine

Organism
Corynebacterium xerosis
Escherichia coli
Lactobacillus acidophilus

{8}------------------------------------------------

Lactobacillus brevis
Lactobacillus jensenii
Lactobacillus lactis
Staphylococcus epidermidis
Streptococcus agalactiae
Candida albicans
Candida glabrata

PRECISION - REPRODUCIBILITY

CT and NG organisms were seeded into male urine matrix at concentrations representing low positive (1x LoD), moderate positive (3x LoD) and high positive (4.15 x 10°GE/mL CT or 8.4 x 10° GE/mL NG) samples. Negative (non-seeded) pooled male urine samples were also included. The resulting panel of 11 pooled male urine samples were tested three times per day for five consecutive days by two operators at three sites (11 specimens x 3 replicates x 5 days x 3 sites x 2 operators). io CT/NG Assays were performed according to the Assay procedure. The rate of agreement for male urine matrix samples with expected CT and NG results for each panel member is shown below:

Summary of reproducibility results in male urine: percent agreement by study site

PanelNo.SampleAnalyteSite 1%agreementSite 2%agreementSite 3%agreement% TotalAgreement
1CT: NegCT100.0%(30/30)100.0%(30/30)96.7%(29/30)98.9%(89/90)
NG: High PositiveNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
2CT: NegCT100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
NG: Mod. PositiveNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
3CT: NegCT96.7%(29/30)100.0%(30/30)100.0%(30/30)98.9%(89/90)
NG: Low PositiveNG93.3%(28/30)96.7%(29/30)93.3%(28/30)94.4%(85/90)
4CT: High PositiveCT100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
NG: NegNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
5CT: Mod. PositiveCT96.7%(29/30)100.0%(30/30)100.0%(30/30)98.9%(89/90)
NG: NegNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
6CT: Low PositiveCT93.3%(28/30)83.3%(25/30)96.7%(29/30)91.1%(82/90)
NG: NegNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
7CT: High PositiveCT100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
NG: High PositiveNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
8CT: High PositiveCT100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
NG: Low PositiveNG96.7%(29/30)93.3%(28/30)96.7%(29/30)95.6%(86/90)
9CT: Low PositiveCT100.0%(30/30)100.0%(30/30)93.3%(28/30)97.8%(88/90)

{9}------------------------------------------------

9NG: High PositiveNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)
10CT: Low PositiveCT86.7%(26/30)100.0%(30/30)93.3%(28/30)93.3%(84/90)
NG: Low PositiveNG90.0%(27/30)100.0%(30/30)93.3%(28/30)94.4%(85/90)
11CT: NegCT93.3%(28/30)100.0%(30/30)100.0%(30/30)97.8%(88/90)
NG: NegNG100.0%(30/30)100.0%(30/30)100.0%(30/30)100.0%(90/90)

Neg = Negative CT High Positive = 4.15 x 106 GE/mL Low Positive = 1x LoD NG High Positive = 8.40 x 105 GE/mL Mod. Positive = 3x LoD

SAMPLE STORAGE AND STABILITY

Specimen stability studies were carried out to determine the length of time samples can be stored prior to testing on the binx io CT/NG Assay.

Six batches of collection kits were tested: three that were within expiry and three batches that were expired (12 months (expiry date) +32 days). Of the three batches that were expired, each were stored at 4°C and at 30°C.

The overall agreement for both positive and negative samples was 100% after storage for 25 hours at 25°C or 8 days at 2-8°C irrespective of whether the collection kits were within expiry date or past the expiry of 12 months and had been stored within the labeled temperature or at 4°C or 30°C. The study indicated that male urine samples are stable for up to 24 hours at 25°C and 7 days at 2-8°C prior to testing with the Assay.

OPERATIONAL ENVIRONMENT

A study was carried out to verify the performance of Instruments and Cartridges when run beyond the extremes of typical ambient temperatures and at high and low levels of relative humidity. Performance of the binx io Instruments was evaluated by placing the Instruments in validated and monitored environmental chambers held at a range of temperatures and humidity levels that were outside the typical normal Instrument operating range.

EnvironmentalTemperature(°C)EnvironmentHumidity(RH%)Expected results:Positive samples% correctresultsExpected results:Negative samples% correctresults
9°C40%CT DetectedNG Detected100%CT Not DetectedNG Not Detected100%
37°C40%CT DetectedNG Detected100%CT Not DetectedNG Not Detected100%
22°C83%CT DetectedNG Detected100%CT Not DetectedNG Not Detected100%
30°C83%CT DetectedNG Detected100%CT Not DetectedNG Not Detected100%
20°C15%CT DetectedNG Detected100%CT Not DetectedNG Not Detected100%

Summary of operational environment testing

Open pack stability

A study was carried out to verify the performance of the Instrument and Cartridge when subjected to a range of temperature and humidity levels.

{10}------------------------------------------------

The study was carried out using CT/NG Cartridges tested with 4x LoD for CT serovar F (ATCC-VR-346) and NG strain ATCC 49226 spiked into eNAT buffer. Cartridges were removed from their packaging and CT and NG spiked into eNAT buffer (positives) or eNAT buffer (negatives) were added. Cartridges were placed in a controlled and monitored incubator for 1, 2, 3, 4, 5 and 6 hours at +30°C. In addition, a set of Cartridges were loaded with a sample and incubated for six hours at +30°C at low (≤20% Relative Humidity (RH)) and high humidity (≥60% RH) using a monitored environmental chamber.

All samples gave the expected results and verified that the CT/NG Assay generates the correct results when a sample is loaded into a Cartridge and stored at 30°C in both high and low humidity conditions following up to six hours storage.

Cartridge performance when run immediately from 2-8°C storage

A study was carried out using Cartridges that had been stored at 2-8°C for a minimum of 12 hours using samples that had either been stored at room temperature or stored refrigerated at 2-8°C for a minimum of 12 hours. All replicates generated correct results.

Shipping stability

The performance of the Assay was assessed following simulated shipping conditions in order to demonstrate performance after undergoing two temporary storage and shipping cycles. This study was carried out using eNAT samples spiked with 4x LoD for positives, or eNAT buffer for negatives. All samples tested generated correct results.

ISTA 3A testing

An ISTA (International Safe Transit Association) 3A study was carried out by an accredited test site followed by subsequent inspection and test performance. The study used a total of 50 Cartridges within a shipper of five cartons containing 10 CT/NG Cartridges each. No damage was observed to the packaging or Cartridges and all Cartridges tested with 4x LoD CT/NG spiked into eNAT buffer for positives and eNAT buffer for negatives generated correct results.

Cross contamination

A study was conducted to demonstrate that the CT/NG Cartridge prevents run-to-run cross contamination when negative samples containing pooled vaginal matrix were run following very high CT/NG double positive samples (containing 2.26 x 10° GE/mL CT and 1.18 x 10′ GE/mL NG). The study consisted of four separate Instruments, with 50 Cartridges run per Instrument, alternating between negative samples and very high CT/NG double positive samples (200 Cartridges run across all Instruments, comprising 100 negative and 100 very high positive). All negative samples were correctly detected as CT, NG Not Detected and all positive samples were correctly identified as CT, NG Detected.

Internal Control Function

A study was carried out to demonstrate the IPC function. The objective was to demonstrate that a Cartridge lacking internal control DNA would report an Assay Invalid result. A panel of conditions was tested including Cartridge manufactured specifically with no IPC present. All samples tested generated the expected results including the Cartridges with no IPC which delivered the expected Assay Invalid result.

All experiments carried out to evaluate analytical performance met established acceptance criteria.

{11}------------------------------------------------

CLINICAL PERFORMANCE

A prospective, multi-center study was carried out to evaluate the performance of the binx io CT/NG Assay with specimens collected at nine investigational sites throughout the U.S. The Assay was compared to the (i) Hologic Aptima Combo 2 (AC2) Chlamydia/Gonorrhea Assay run on Panther, (ii) BD ProbeTec Chlamydia trachomatis (CT) Q*, and BD ProbeTec Neisseria gonorrhoeae (GC) Q* assays run on the Viper XTR™, and (iii) Roche cobas CT/NG v2.0 test run on the cobas 4800 System. The three reference tests were used to form a Composite Infected Status (CIS) where a patient was considered if at least two out of the three reference tests were positive and not infected if at least two out of the three reference tests were negative.

Urine samples were prospectively collected from men without symptoms of infection from a variety of clinical venues in the United States. Clinical sites included sexually transmitted disease clinics, family planning clinics and HIV clinics.

Site personnel that carried out testing using the CT/NG Assay were, in the vast majority (94% overall), point-of-care personnel trained in the use of the binx io CT/NG System, but not trained or experienced in general laboratory testing procedures.

There were 1,157 total participants enrolled into the study of which eight were ineligible or withdrew consent and 227 participants were excluded due to deviations to the study protocol.

A total of 922 male urine specimens were fully evaluable. Of the total number of male urine specimens collected, 614 were from asymptomatic participants and 308 were from symptomatic participants.

A total of 120 eligible specimens were classified as infected for CT, of which 60 were symptomatic and 60 were asymptomatic. A total of 802 participants were classified as not infected for CT. of which 248 were symptomatic and 554 were asymptomatic. A total of 74 participants were classified as infected for NG, of which 62 were symptomatic and 12 were asymptomatic. A total of 848 participants were classified as not infected for NG, of which 246 were symptomatic and 602 were asymptomatic.

The median age of participants was 28, ranging from 17 to 76 years.

Positivity rates for the binx io CT/NG assay observed in the clinical study spanned a range across clinical sites from 0.0%-24.7% (average 12.7%) for chlamydia and 0.0%-22.2% (average 7.8%) for gonorrhea.

CTNG
Site No.Type ofClinicUSStateTotalNo.SamplesNo.Positive bybinx%PositivityRateNo.Positiveby binx% PositivityRate
1STDIN892224.7%1618.0%
2STDMD125118.8%129.6%
3STDLA2583814.7%187.0%
4HIVMS12216.7%00.0%
5FamilyPlanningTX1271915.0%64.7%
6FamilyPlanningPA33412.1%13.0%
7STDAL59915.3%711.9%

binx io CT/NG Assay Positivity Rates in Males (observed during the Clinical Study)

{12}------------------------------------------------

8HIVCA181116.1%95.0%
9STDNC900.0%222.2%
10STDMS2913.4%13.4%
Total92211712.7%727.8%

Composite Infected Status for Chlamydia trachomatis by symptom status for male urine samples

CISComparator Systembinx io®Symptom StatusTotal
NAAT1NAAT2NAAT3SxAsx
NI----243542785
NI--+-O33
NI-+--000
NI+---011
NI--IND-213
NI-IND--101
NIIND---123
NI---+15
NI--++000
NI-+-+000
NI+--+000
NI--IND+000
NI-IND-+000
NIIND--+000
NI---IND000
NI--+IND000
NI-+-IND000
NI+--IND000
NI--INDIND000
NI-IND-IND000
NIINDl-IND000
Total Not Infected248554802
l++++5456110
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++-+000
l+-++000
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+++000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++IND+000
l+IND++000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------IND+++101
l+++-33
l++--101
l+-+-101
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++-011
l++IND-000
l+IND+-000
lIND++-000
l+++IND000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++-IND000
l+-+IND000
l-++IND000
l++INDIND000
l+IND+IND000
lIND++IND000
Total Infected6060120
CIS = Comparator Infected StatusNI = Not Infectedl = InfectedIND = Indeterminate

CIS = Comparator Infected Status Sx = Symptomatic

NI = Not Infected Asx = Asymptomatic

Composite Infected Status for Neisseria gonorrhea by symptom status for male urine samples

CISComparator Systembinx io®Symptom StatusTotal
NAAT1NAAT2NAAT3SxAsx
NI----241594835

{13}------------------------------------------------

NI--+-145
NI-+--000
NI+---123
NI--IND-213
NI-INDl-101
NIIND---011
NI---+000
NI--++000
NI-+-+000
NI+--+000
NI--IND+000
NI-IND-+000
NIINDl-+000
NI---IND000
NI--+IND000
NI-+-IND000
NI+--IND000
NI--INDIND000
NI-IND-IND000
NIIND--IND000
Total Not Infected246602848
-++++દેત્ત્વ10દિવે
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++-+0O0
l+-++000
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+++000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++IND+000
l+IND++000
lIND+++213
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+++-O11
l++--000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+-+-101
l-++-000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++IND-000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+IND+-000
lIND++-000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+++IND000
l++-IND000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+-+IND000
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------++IND000
l++INDIND000
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+IND+IND000
lIND++INDO00
Total Infected621274
CIS = Comparator Infected StatusNI = Not Infectedl = InfectedIND = Indeterminate

Sx = Symptomatic

Asx = Asymptomatic

Results from the CT/NG Assay were compared to the CIS for the determination of sensitivity and specificity. Sensitivity and specificity for CT and NG with male urine, by symptom status, are shown below.

Clinical Performance of the binx health io CT/NG Assay Against CIS for Chlamydia trachomatis with male urine specimens

Symptom StatusTotal NTPFNTNFPSensitivity(95% CI)Specificity(95% CI)
Asymptomatic614564549593.3%(84.1% -97.4%)99.1%(97.9% -99.6%)
Symptomatic308555247191.7%(81.9% -96.4%)99.6%(97.8% -99.9%)
Total9221119796692.5%(86.4% -96.0%)99.3%(98.4% -99.7%)

N= Number of specimens TP = True Positive FN = False Negative TN = True Negative FP = False Positive
Confidence Intervals (CI) for Sensitivity and specificity: Wilson Score

{14}------------------------------------------------

Clinical Performance of the binx health io CT/NG Assay Against CIS for Neisseria gonorrhoeae with male urine specimens

Symptom StatusTotal NTPFNTNFPSensitivity(95% CI)Specificity(95% CI)
Asymptomatic614111602091.7%(64.6% - 98.5%)100.0%(99.4% - 100.0%)
Symptomatic308611246098.4%(91.4% - 99.7%)100.0%(98.5% - 100.0%)
Total922722848097.3%(90.7% - 99.3%)100.0%(99.5% - 100.0%)

N= Number of specimens TP = True Positive FN = False Negative TN = True Negative FP = False Positiv ence Intervals (CI) for Sensitivity and specificity: Wilson Score Method

Overall clinical performance

TargetOverall SensitivityOverall Specificity
CT111/120 (92.5%)796/802 (99.3%)
NG72/74 (97.3%)848/848 (100.0%)

Rate of Invalid Results

In cases where the Assay returned a Test Invalid result, the patient specimen was retested. The result of the retest was used in the analysis. A Test Invalid result was reported in 15 of the 937 cartridges tested (1.6%) in this study. If three successive assays returned a Test Invalid result on a single specimen this was recorded as indeterminate. Of the 922 specimens for which a CIS could be determined, none of the specimens generated a final indeterminate result.

Positive and Neqative Predictive Values

The sensitivity and specificity of the binx health io CT/NG Assay when used with male urine, was used to calculate the hypothetical positive predictive values (PPV) and negative predictive values (NPV) at a range of hypothetical prevalence rates.

Hypothetical PPV and NPV: Male urine specimens

Chlamydia trachomatisNeisseria gonorrhoeae
HypotheticalPrevalencebinx CT/NGSens./Spec.% PPV95% CI% NPV95% CIbinx CT/NGSens./Spec.% PPV95% CI% NPV95% CI
1%Sensitivity=92.5%57.2%33.3%-78.1%99.9%99.4%-100.0%Sensitivity=97.3%100.0%70.0-100.0%100%99.5%-100.0%

{15}------------------------------------------------

Chlamydia trachomatisNeisseria gonorrhoeae
HypotheticalPrevalencebinx CT/NGSens./Spec.% PPV% NPVbinx CT/NGSens./Spec.% PPV% NPV
95% CI95% CI95% CI95% CI
2%Specificity=99.3%72.9%99.8%Specificity=100.0%100%99.5%
52.7%-86.7%99.3%-100.0%82.4%-100.0%99.5%-100.0%
5%87.4%99.6%100%99.9%
75.3%-94.1%98.9%-99.9%92.1%-100%
10%93.6%99.2%100%99.7%
86.6%-97.1%98.3%-99.6%95.9%-100%
15%95.9%98.7%100%99.5%
91.0%-98.2%97.6%-99.3%97.2%-100%
20%97.1%98.1%100%99.3%
93.4%-98.7%96.9%-98.9%97.9%-100%
25%97.8%97.5%100%99.1%
94.8%-99.1%96.1%-98.5%98.3%-100%

SUBSTANTIAL EQUIVALENCE

ItemPredicate Device:binx io CT/NG Assay andbinx io CT/NG System(K191352)Proposed Device:binx io CT/NG Assay andbinx io CT/NG System
Regulation866.3393866.3393
RegulationSpecialtyMicrobiologyMicrobiology
Device ClassClass IIClass II
TechnologyAutomated multiplex polymerasechain reaction with electrochemicaldetectionSame
Intended UseThe binx health io CT/NG Assay,when tested using the binx health ioInstrument, is a fully-automatedrapid, qualitative test intended foruse in point-of-care or clinicallaboratory settings for the rapiddetection of Chlamydia trachomatisand Neisseria gonorrhoeae DNA infemale vaginal swab specimenscollected either by a clinician orThe binx health io CT/NG Assay,when tested using the binx healthio Instrument, is a fully automated,rapid, qualitative test intended foruse in point-of-care or clinicallaboratory settings for the detectionof Chlamydia trachomatis andNeisseria gonorrhoeae DNA bypolymerase chain reaction. Thebinx health io CT/NG Assay is
self-collected by a patient in aclinical setting, to aid in thediagnosis of symptomatic orasymptomatic infection in femalepatients with Chlamydiatrachomatis and/or Neisseriagonorrhoeae .intended for use with femalevaginal swab specimens, collectedeither by a clinician or self-collected by a patient in a clinicalsetting, or male urine specimens,as an aid in the diagnosis ofsymptomatic or asymptomaticChlamydia trachomatis and/orNeisseria gonorrhoeae infection.For a symptomatic male patientwith a chlamydia negative testresult, further testing with alaboratory-based molecular test isrecommended.
IndicationsSymptomatic and asymptomaticfemale patientsSymptomatic and asymptomaticfemale patientsSymptomatic and asymptomaticmale patients
Assay TargetDNA from Chlamydia trachomatisand/or Neisseria gonorrhoeaeSame
SpecimenTypesClinician-collected female vaginalswabsClinician-collected female vaginalswabs
Patient-collected female vaginalswabs (in a clinical setting)Patient-collected female vaginalswabs (in a clinical setting)Patient-collected male urinespecimens (in a clinical setting)
CT AnalyteTargetCT Genomic DNACT Genomic DNA
NG AnalyteTargetNG Genomic DNANG Genomic DNA
Collection KitVaginal Swab Specimen CollectionKitMale Urine Specimen CollectionKitVaginal Swab Specimen CollectionKit
Nucleic AcidExtractionYesYes
Assay ResultsQualitativeQualitative
Assay ControlsInternal Process ControlExternal controls available but notsuppliedInternal Process ControlExternal controls available but notsupplied
InstrumentSystembinx health io®binx health io®

{16}------------------------------------------------

§ 866.3393 Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s).

(a)
Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.