K Number
K200423

Validate with FDA (Live)

Date Cleared
2021-09-22

(580 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

Device Description

The Model XW-LY001 Electric wheelchair is powered by Li-ion battery, driven by DC motor. Users control direction and adjust speed by controller. It is suitable to be used in low speed, good road condition and small slope. The electric wheelchair consists of two foldable armrests, a backrest, a seat cushion, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, a Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller. The subject device has 8 inch front wheel and 12.5 inch rear wheel. The motor of electric wheelchair is DC24V 200W; the battery is 25.2V 10.4Ah Li-ion battery; the charger is DC 24V, 2A. Max. loading can not be over than 125Kgs, and its maximum speed is 6km/h. The braking time is about 2-3s,and the braking distance is ≤1.5m

AI/ML Overview

The provided FDA 510(k) summary (K200423) for the Electric Wheelchair (XW-LY001) details non-clinical tests to demonstrate substantial equivalence to a predicate device (K170787). This document does not describe a study involving an AI device or a comparative effectiveness study with human readers. Therefore, several aspects of your request related to AI performance, human reader improvement, and ground truth establishment for AI models cannot be addressed from the given text.

Based on the provided information, here's a breakdown of the acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are its compliance with various ISO and IEC standards for wheelchairs and the demonstration of substantial equivalence to a predicate device. The reported device performance is presented as compliance with these standards and a comparison of its characteristics to the predicate device.

Acceptance Criteria (Compliance with Standards)Reported Device Performance (Compliance Statement)
ISO 7176-1:2014 (Static stability)Complies with the standard
ISO 7176-2:2017 (Dynamic stability)Complies with the standard
ISO 7176-3:2012 (Brakes effectiveness)Complies with the standard
ISO 7176-4:2008 (Energy consumption)Complies with the standard
ISO 7176-5:2008 (Dimensions, mass, maneuvering)Complies with the standard
ISO 7176-6:2018 (Max speed, accel, decel)Complies with the standard
ISO 7176-7:1998 (Seating and wheel dimensions)Complies with the standard
ISO 7176-8:2014 (Static, impact, fatigue strengths)Complies with the standard
ISO 7176-9:2009 (Climatic tests)Complies with the standard
ISO 7176-10:2008 (Obstacle-climbing ability)Complies with the standard
ISO 7176-11:2012 (Test Dummies)Complies with the standard
ISO 7176-13:1989 (Coefficient of Friction)Complies with the standard
ISO 7176-14:2008 (Power and control systems)Complies with the standard
ISO 7176-15:1996 (Information disclosure)Complies with the standard
ISO 7176-16:2012 (Resistance to ignition)Complies with the standard
ISO 7176-21:2009 (Electromagnetic compatibility)Complies with the standard
ISO 7176-22:2014 (Set-up Procedures)Complies with the standard
ISO 7176-25:2013 (Batteries and chargers)Complies with the standard
IEC 62133:2012 (Secondary cells and batteries)Complies with the standard
ISO 10993-1:2018 (Biocompatibility)Complies with the standard
ISO 10993-5:2009 (Cytotoxicity)Complies with the standard
ISO 10993-10:2010 (Irritation and Skin Sensitization)Complies with the standard

The "study that proves the device meets the acceptance criteria" is described as "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device." (page 5, "7.0 _Non-Clinical Test Conclusion"). These tests demonstrated compliance with the listed ISO and IEC standards.

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of data or AI model evaluation. The tests conducted were non-clinical performance and safety tests of the physical wheelchair device. The provenance of the data is not applicable in this context as it refers to physical testing results against engineering standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not an AI device or a diagnostic device where expert ground truth would be established for image or data interpretation. The "ground truth" here is the pass/fail criteria defined by the ISO and IEC standards.

4. Adjudication method for the test set

Not applicable, as there is no human adjudication of a test set in the context of this device's non-clinical safety and performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic or interpretation device. No MRMC study was performed or is relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used

The "ground truth" for the device's performance is adherence to the specifications and performance metrics outlined in the various international standards (ISO and IEC) for wheelchairs and medical devices. These standards define the acceptable range or criteria for parameters like static stability, dynamic stability, braking effectiveness, energy consumption, dimensions, maximum speed, strength, climatic performance, obstacle climbing, power/control systems, electromagnetic compatibility, battery safety, and biocompatibility.

8. The sample size for the training set

Not applicable, as this is not an AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI device that requires a training set or ground truth establishment for AI models.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" and the words "U.S. Food & Drug Administration".

September 22, 2021

Yongkang Dingchang Industry & Trade Co., LTD. % Boyle Wang Official correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 608, No. 738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K200423

Trade/Device Name: Electric Wheelchair (XW-LY001) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: August 11, 2021 Received: August 24, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

For Heather Dean, Ph. D Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200423

Device Name Electric Wheelchair (XW-LY001)

Indications for Use (Describe)

The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K200423

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name: YONGKANG DINGCHANG INDUSTRY & TRADE CO., LTD. Address: NO.99, HUAXI ROAD, XICHENG STREET, YONGKANG CITY, ZHEJIANG PROVINCE, 321300 CHINA Tel: 86- 13429055889 Fax: 86- 579-87321558 Contact: Tony Chen Date of Preparation: Sep.21.2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Electric Wheelchair Common name: Powered wheelchair Classification name: Wheelchair, Powered Model(s): XW-LY001

3.0 Classification

Production code: ITI Regulation number: 21 CFR 890.3860 Classification: Class II Panel: Physical Medicine

4.0 Predicate Device Information

Manufacturer: Nanjing Jin Bai He Medical Apparatus Co., Ltd. Device: Powered Wheelchair DYW30A(D09)

{4}------------------------------------------------

510(k) number: K170787

5.0 Indication for Use Statement

The device is a motor-driven, and indoor transportation vehicle with the intended use to provide mobility to a disabled or an elderly person limited to a seated position.

6.0 Device Description

The Model XW-LY001 Electric wheelchair is powered by Li-ion battery, driven by DC motor. Users control direction and adjust speed by controller.

It is suitable to be used in low speed, good road condition and small slope.

The electric wheelchair consists of two foldable armrests, a backrest, a seat cushion, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, a Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller.

The subject device has 8 inch front wheel and 12.5 inch rear wheel.

The motor of electric wheelchair is DC24V 200W; the battery is 25.2V 10.4Ah Li-ion battery; the charger is DC 24V, 2A.

Max. loading can not be over than 125Kgs, and its maximum speed is 6km/h. The braking time is about 2-3s,and the braking distance is ≤1.5m

7.0 _Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ISO 7176-1:2014 Wheelchairs – Part 1: Determination of static stability.

ISO 7176-2:2017 Wheelchairs - Part 2:Determination of dynamic stability of electrically powered wheelchairs.

ISO 7176-3:2012 Wheelchairs – Part 3: Determination of effectiveness of brakes.

ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range.

ISO 7176-5:2008 Wheelchairs – Part 5: Determination of dimensions, mass and manoeuvring space.

ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs.

ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions.

ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths.

ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs.

ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of

{5}------------------------------------------------

electrically powered wheelchairs.

ISO 7176-11:2012 Wheelchairs - Part 11 Test Dummies

ISO 7176-13:1989 Wheelchairs - Part 13 Determination of Coefficient of Friction of Test Surfaces.

ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods; ISO 7176-15:1996 Wheelchairs – Part 15: Requirements for information disclosure, documentation and labeling.

ISO 7176-16:2012 Wheelchairs – Part 16: Resistance to ignition of postural support device.

ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers

ISO 7176-22:2014 Wheelchairs - Part 22 Set-up Procedures

ISO 7176-25:2013 First edition 2013-07-15Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs

IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications

Biocompatibility of patient-contacting parts

Statement for Biocompatibility Certification

The Joystick knob, Joystick Gaiter, Enclosure Moulding(s), and Keypad of the Electric Wheelchair, XW-LY001 in its final finished form are identical to the Joystick knob, Joystick Gaiter, Enclosure Moulding(s), and Keypad of the Y207 Electric Wheelchair (JIANGSU INTCO MEDICAL PRODUCTS CO., LTD., K202482, clearance date 03/18/2021) in formulation, processing, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents). The patient-contacting materials of Electric Wheelchair, XW-LY001 have the same nature of tissue contact and contact duration as the Y207 Electric Wheelchair.

Other patient-contacting material are carried out biocompatibility assessment in accordance with IS0 10993-1: 2018, including :

Cytotoxicity per ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

Irritation and Skin Sensitization per ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

8.0 Clinical Test Conclusion

{6}------------------------------------------------

Clinical testing was not required for this submission.

9.0 Comparison to Predicate Device and Conclusion

The technological characteristics, features, specifications, materials, mode of operation, and intended use of the model XW-LY001 Electric Wheelchair is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

CharacteristicSubject DevicePredicate DeviceJustification forSubstantial Equivalence
Proprietary name,Electric WheelchairPowered Wheelchair--
ManufacturerYONGKANGDINGCHANG INDUSTRY& TRADE CO., LTD.Nanjing Jin Bai HeMedical Apparatus Co.,Ltd.
Model(s)XW-LY001DYW30A(D09)--
510(k) numberK200423K170787--
Frame MaterialAluminium alloyAluminium alloySame
Frame Design/StyleFoldable/The device consists of afoldable and non-rigid typeof power wheelchair basewith rear drive and 2casters in the front andtwo anti-tippers in the rear.Foldable/The device consists of afoldable and non-rigid typeof power wheelchair basewith rear drive and 2casters in the front andtwo anti-tippers in the rear.Similar
Folding MechanismA foldable seat frames(The backrest could befolded to seat)A foldable seat frames(The backrest could befolded to seat)Same
Product StructureConsists of two foldablearmrests, a backrest, aseat cushion, a foldableframe, two rear drivingwheels with hubmotor/electromagneticbrake assemblies, twopivoting casters, a Li-ionbatteries, an off-boardbattery charger, a controlpanel, and an electricmotor controller.Consists of two foldablearmrests, a seat belt, abackrest, a seat cushion, afoldable frame, two reardriving wheels with hubmotor/electromagneticbrake assemblies, twopivoting casters, two Li-ionbatteries, an off-boardbattery charger, a controlpanel, and an electricmotor controllerThe subject device has noseat belt. Relatedprecautions and warningsfor use the device and theinformation of operatingsurface and max. slopehave clearly defined in theuser manual.
Length37.4"37.4"Minor differences inthe dimensions willnot impact the safetyand effectiveness ofthe substantialequivalence.
Width25.6"22.6"
Height38.4"36.2"
Seat Dimensions
Width18.5"Not publicly available/
Depth16.9"Not publicly available/
Height19.3"Not publicly available/
Folded Dimensions
Length30.5"Not publicly available/
Width25.6"Not publicly available/
Height13.8"Not publicly available/
Wheelchair Weight
With batteries27.8kg / 61 lbs51.8" lbs. / 23.5 kgThe difference willnot raise any newsafety andeffectivenessconcerns
Without batteries25.8kg / 55.6 lbsNot publicly available/
ControllerPG DRIVESTECHNOLOGY LTD.,newVSiChangzhou BillonElectronic Appliance Co.,Ltd., WS-1Although differentcontrolleris used, both the controlsystem, including thejoystick controller, theelectromagnetic brakesandthe user interface aresimilar.The joystick controls thedirections and speed ofmovement, and when thejoystick is released, thepowered wheelchair willslow down to stop and thebrakes will automaticallyre-engage.The controller alsoprovides the battery statusdisplaying and abnormalcondition displaying. Bothof the control systems areevaluated according to
Drive Style(e.g., rear, mid, front)Direct drive on the rear wheelsDirect drive on the rear wheelsSame
Motor TypeBrushed DC motorBrushless DC motorThe minor differences will not impact the safety and effectiveness of the substantial equivalence.
Motor OutputDC 24V x 200W*2 pcs24 VDC *250W *2 pcs
BatteriesKunshan Hi-FortuneHealth Products Co.,LtdJiangsu Feng Chi GreenPower Supply Co., Ltd.The minor differences will not impact the safety and effectiveness of the substantial equivalence.
Quantity1pcs2 pcs
TypeFH2410ITP2406
ChemistryLithium-ionLithium-ion
Range per Charge15.5 miles (25km)11.2 miles (18 km)There is a larger cruising range for the subject device.
Charger Type(On-board/Off-board/Carry-on)Off-boardOff-boardSame
Input/Output PowerDC 25.2 V, 10.4 Ah24 VDC, 6 AhThe minor differences will not impact the safety and effectiveness of the substantial equivalence.
BrakeSmart ElectromagneticbrakeSmart ElectromagneticbrakeSame
Minimum brakingdistance and time$≤$ 1.5 m,0.8s1mOur minimum braking distance is a little bigger thanthe predicate device. There is no new safety and effectiveness concerns raised due to the difference
Max speed
Forward3.75 mph (6 km/h)3.75 mph (6 km/h)Same
Reverse1.79 mph(2.88 km/h)1.86 mph (3.0 km/h)The devices are evaluated according to standard ISO 7176-6:2018, so the different will not impact the safety and effectiveness.
Rear Wheels Size12.5" x 2.25"(320 x 57 mm)12.5" x 2.25"(320 x 57 mm)Same
Quantity22Same
Tire Pressure (ifpneumatic)2kgf/m²PU solid tire/
Castors Size8"x1.97"(200 x 50mm)7"x1.77" (180 x 45 mm)Larger sizes of frontwheels bring steadierpivoting function thanpredicate device.
Quantity22Same
Tire Pressure (ifpneumatic)PU solid tirePU solid tireSame
Anti-tip Wheels2"x0.4"(48 x10mm)Not publicly available/
Removable (Yes/No)NoNoSame
Maximum OccupantMass125kg/ 276 lbs.120kg/ 264 lbs.Large loading weightmeans more convenientfor the User.
Curb Climbing ability1.97 "(50mm)1.36" (34.5 mm)The larger height in theobstacle climbing willnot impact the safetyand effectiveness of thesubject device.
Ground clearance25mmNot publicly available/
Minimum TurningRadius≤42.2"(≤1.2m)32.8" ( 833 mm)The difference in theturning radius will bringmore convenience when itturns. The difference willnot raise any new safetyandeffectivenessconcerns.
Maximum Incline10 degrees8 degreesLarger safe operationalincline of subject bringmore convenient for theuse environment
FootplatesPlasticNot publicly available/
Back UpholsteryOxford clothPVC VinylBiocompatibility evaluationhas been carried outper ISO 10993-1.There are no new safetyand effectivenessconcerns due to thedifference.
Armrest TypeUnadjustableNot publicly available/
Operating surface &environmentIndoorIndoorSame
Additional AccessoryElectronic items: 1 yearNot publicly available/

Table1- General Device Characteristics Comparison Table

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Warranty

Frame: 5 years

Summary of substantial equivalence discussion:

The XW-LY001 Electric wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ાડભ 7176-5:2008, ISO 7176-6:2001, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16:2012, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013,IEC 62133:2012, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2010. The intended uses for both devices are the same. Mainframes of two devices are folded by way of front and rear close, and frame materials all meet the Tensile Strength, Yield Load, and Elongation tests. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2001, ISO 7176-10:2008. The biocompatibility of the Predicate device and Subject device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2010. The flame retardant test of the seat cushion/back cushion and armrest of both subject device and predicate device is carried out according to the ISO 7176-16 test. Therefore, both devices are assured to be under the same safety level. In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

10.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.

N/A