(56 days)
The HS40 diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
The HS40 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices.
The HS40 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode. M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a combination of these modes. The HS40 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS40 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
The provided document is a 510(k) Premarket Notification for the Samsung Medison HS40 Diagnostic Ultrasound System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific, novel acceptance criteria through extensive clinical studies.
Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria, specifically concerning the performance of an AI algorithm or an analysis of human reader improvement with AI assistance. The document explicitly states:
- "The subject of this premarket submission, HS40, is not required clinical studies to support substantial equivalence." (Page 26, Section 12)
The main purpose of this submission is to demonstrate that the new HS40 system is as safe and effective as the predicate device (HS40 Diagnostic Ultrasound System K191055, and features migrated from HS50/HS60 Diagnostic Ultrasound System K192505), based on non-clinical tests and technological equivalence.
Here's a breakdown of why each requested point cannot be fully addressed by this document:
-
A table of acceptance criteria and the reported device performance: This document does not present quantitative acceptance criteria for specific diagnostic performance metrics (e.g., sensitivity, specificity for a particular disease detection) that would be proven by a clinical study. Instead, it relies on demonstrating equivalence in technical specifications, safety, and functionality to a previously cleared device.
- However, it does list the clinical applications for the device and its various transducers. These implicitly represent the "performance" capabilities:
- Clinical Applications (HS40 System - Page 3): Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
- Modes of Operation (HS40 System - Page 3): B-mode, M-mode, Pulsed Wave Doppler (PWD), Continuous Wave Doppler (CWD), Color Doppler, and various combined modes. Also mentions S-Harmonic (Harmonic Imaging), Tissue Doppler Imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, 3D imaging, and ElastoScan Mode.
- Added SW features (Page 25): S-Detect for Breast, 2D Follicle, LaborAssist, CrystalVue. These features, specifically S-Detect for Breast and 2D Follicle, hint at automated analysis or AI components, but no performance data for them is provided in this document.
- However, it does list the clinical applications for the device and its various transducers. These implicitly represent the "performance" capabilities:
-
Sample sized used for the test set and the data provenance: Not applicable, as no clinical study or test set with patient data for diagnostic performance evaluation is detailed. The tests performed are non-clinical, related to safety and technical standards (e.g., acoustic output, electrical safety).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical study with ground truth established by experts is detailed.
-
Adjudication method for the test set: Not applicable, as no clinical study requiring adjudication is detailed.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The document does not describe an MRMC study or AI assistance for human readers. While it mentions "S-Detect for Breast" as a new software feature, which typically implies AI-driven detection, the submission does not include any performance data or comparative studies for this feature.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. No standalone algorithm performance study is described.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable, as no clinical study with a ground truth is detailed.
-
The sample size for the training set: Not applicable. This document is a 510(k) for an ultrasound system, not an AI/ML device that would require explicit training set data for its submission. Even if the mentioned "S-Detect for Breast" feature utilizes AI, the details of its training are not part of this specific FDA submission as presented here.
-
How the ground truth for the training set was established: Not applicable for the reasons stated above.
In summary of what the document DOES provide regarding device meeting acceptance criteria:
The "acceptance criteria" here are generally understood as the regulatory requirements for demonstrating substantial equivalence to a predicate device. The study proves the device meets these criteria through:
- Comparison to Predicate (Section 10, Page 25): The document argues that the HS40 is substantially equivalent to its predicate (K191055 and K192505) in:
- Intended use
- Imaging capabilities
- Technological characteristics
- Safety and effectiveness
- Specific examples include sharing imaging modes, modes of operation, and migrating certain software features and transducers from cleared devices.
- Summary of Non-Clinical Tests (Section 11, Page 26): The device has been evaluated and found to conform with applicable FDA guidance and medical device safety standards for:
- Acoustic output
- Biocompatibility
- Software function (without detailing performance metrics for diagnostic features)
- Cleaning and disinfection effectiveness
- Thermal, electrical, electromagnetic, and mechanical safety
- Specific standards cited: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA UD 2-2004. These are the "acceptance criteria" in the context of this specific filing approach – adhering to recognized standards to ensure safety and performance equivalence.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 7, 2020
Samsung Medison Co., LTD % Scully Kim Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-du 25108 REPUBLIC OF KOREA
Re: K200339
Trade/Device Name: HS40 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 11, 2020 Received: February 11, 2020
Dear Scully Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200339
Device Name HS40 Diagnostic Ultrasound System
Indications for Use (Describe)
The HS40 diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
The HS40 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: HS40 Diagnostic Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Note 2, 4, 7, 8, 9, 11 | |
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 11 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 11 | ||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Note 2, 7, 9, 11, 12 | ||
| Neonatal Cephalic | P | P | P | P | Note 1 | Note 8, 9, 11 | ||
| Adult Cephalic | P | P | P | P | P | Note 1 | Note 7 | |
| Trans-rectal | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Trans-vaginal | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 11, 12 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Note 2, 7, 9, 11 | ||
| Intra-luminal | ||||||||
| Other (See Note 13) | P | P | P | P | Note 1 | Note 2, 7, 9, 12 | ||
| Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7, 14 | |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7, 14 |
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 11 |
| Other (spec.) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development -
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients -
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{4}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/4/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The wordmark is set against a blue, rounded, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LA3-16AD for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | |||||||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Note 2, 7, 9, 11 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | P | P | P | P | Note 1 | Note 2, 7, 9, 11 | ||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Note 2, 7, 9, 11, 12 | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Note 2, 7, 9, 11, 12 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Note 2, 7, 9, 11 | ||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac | Cardiac Adult | |||||||
| Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Note 2, 5, 7, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055 ; E= added under Appendix E
Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{5}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/5/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CA2-8AD for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Note 2,4, 7, 9, 11 | |
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 2, 7, 9, 11 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging | Laparoscopic | |||||||
| &Other | Pediatric | P | P | P | P | P | Note 1 | Note 2, 7, 9, 11 |
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Note 2, 7, 9, 11 | |
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| Peripheral | Peripheral vessel | P | P | P | P | P | Note 1 | Note 2, 7, 9, 11 |
| Vessel | Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Note 2: Includes imaging for guidance of biopsy
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{6}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/6/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CF4-9 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | ||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | ||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | ||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
-
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis -
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{7}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/7/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: PN2-4 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | ||||||||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Note 7 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | P | P | P | P | Note 1 | Note 7 | ||
| Trans-rectal (See Note 13) | ||||||||
| Trans-vaginal (See Note 13) | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | P | P | P | P | Note 1 | Note 4, 7, 14 | ||
| Cardiac | Cardiac Pediatric | P | P | P | P | Note 1 | Note 4, 7, 14 | |
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
{8}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/8/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: EVN4-9 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Note 2, 7, 9 |
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 2, 7, 9 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| & Other | Pediatric | |||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | P | Note 1 | Note 2, 7, 9, 12 | |
| Trans-vaginal | P | P | P | P | P | Note 1 | Note 2, 7, 9, 12 | |
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (spec.) (See Note 13) | P | P | P | P | P | Note 1 | Note 2, 7, 9, 12 | |
| Cardiac | Cardiac Adult | |||||||
| Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
- Note 12: ElastoScan Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
Note 2: Includes imaging for guidance of biopsy
{9}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/9/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: VN4-8 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Note 2,4, 7, 8, 9, 11 | |
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 11 | |
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 11 | |
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by K191055; E= added under Appendix E
Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
Note 3: Includes infertility monitoring of follicle development
-
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients -
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI) -
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
- Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Note 2: Includes imaging for guidance of biopsy
{10}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/10/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: V5-9 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Note 2, 7, 8, 9 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Trans-vaginal | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 12 | ||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{11}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/11/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: DP2B for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | |||||||
| Abdominal (See Note 12) | ||||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | P | |||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac | Cardiac Adult | P | ||||||
| Cardiac Pediatric | P | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | ||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{12}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/12/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: C2-8 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Notes 2, 4, 7,8, 9, 11 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Notes 2, 6, 7, 8, 9, 11 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging Note 12: ElastoScan
Note 12: ElastoScan
Note 12: Includes Ural.
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{13}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/13/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: C2-5 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Notes 2, 4, 7, 8, 9, 11 | ||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Notes 2, 6, 7, 8, 9, 11 | ||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging | Laparoscopic | |||||||
| &Other | Pediatric | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| Peripheral | Peripheral vessel | |||||||
| Vessel | Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development -
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{14}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/14/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed inside a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CA2-6BM for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal/Obstetrics ( See Note 3 ) | |||||||||
| Abdominal ( See Note 10 ) | P | P | P | P | Note 1 | Notes 6, 7, 8, 9, 11 | |||
| Intra-operative ( See Note 6 ) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Fetal Imaging& Other | Laparoscopic | ||||||||
| Pediatric | |||||||||
| Small Organ ( See Note 5 ) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | |||||||||
| Musculo-skel. (Superfic.) | |||||||||
| Intra-luminal | |||||||||
| Other ( See Note 13 ) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac | Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Other (spec.) |
N= new indication; P= previously cleared by K191055 ; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
- Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
- Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{15}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/15/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LN5-12 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | ||||||||
| Abdominal (See Note 10) | ||||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Notes 2, 5, 7, 8, 9, 11 | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{16}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/16/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: L5-12/50 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | ||||||||
| Abdominal (See Note 10) | ||||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Notes 2, 5, 7, 8, 9, 11 | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{17}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/17/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: ER4-9 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | ||||||||
| Ophthalmic | ||||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Notes 2, 7, 8 | ||
| Fetal Imaging& Other | Abdominal (See Note 10) | |||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | Note 1 | Notes 2, 7, 8, 12, 13 | ||
| Trans-vaginal | P | P | P | P | Note 1 | Notes 2, 7, 8, 12 | ||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| Peripheral | Peripheral vessel | |||||||
| Vessel | Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{18}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/18/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: L4-7 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal/Obstetrics (See Note 3) | |||||||||
| Abdominal (See Note 10) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Fetal Imaging& Other | Laparoscopic | ||||||||
| Pediatric | |||||||||
| Small Organ (See Note 5) | P | P | P | P | Note 1 | Notes 2, 5, 7, 8, 9, 11 | |||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac Adult | |||||||||
| Cardiac | Cardiac Pediatric | ||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| Peripheral | Peripheral vessel | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | ||
| Vessel | Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
{19}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/19/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LS6-15 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | ||||||||
| Abdominal (See Note 10) | |||||||||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (See Note 5) | N | N | N | N | Note 1 | Notes 2, 5, 7, 8, 9, 11 | |||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Musculo-skel. (Superfic.) | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 | |||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac | Cardiac Adult | ||||||||
| Cardiac Pediatric | |||||||||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | P | P | P | P | Note 1 | Notes 7, 8, 9, 11 | ||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Note 2: Includes imaging for guidance of biopsy
{20}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/20/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: SP3-8 for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | ||||||||
| Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 7 | ||
| Intra-operative (See Note 6) | |||||||||
| Intra-operative (Neuro.) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (See Note 5) | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | P | P | P | P | P | Note 1 | Note 7 | ||
| Trans-rectal (See Note 13) | |||||||||
| Trans-vaginal (See Note 13) | |||||||||
| Trans-urethral | |||||||||
| Trans-esoph. (non-Cardiac) | |||||||||
| Musculo-skel. (Convent.) | |||||||||
| Musculo-skel. (Superfic.) | |||||||||
| Intra-luminal | |||||||||
| Other (See Note 13) | |||||||||
| Cardiac | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7, 14 | |
| Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7, 14 | ||
| Trans-esophageal (Cardiac) | |||||||||
| Other (spec.) | |||||||||
| PeripheralVessel | Peripheral vessel | ||||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
- Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{21}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/21/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: DP8B for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| (Track I only) | (Tracks I & III) | |||||||
| Ophthalmic | Ophthalmic | |||||||
| Fetal Imaging& Other | Fetal/Obstetrics (See Note 3) | |||||||
| Abdominal (See Note 12) | ||||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | P | |||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | ||||||||
| Cardiac | Cardiac Adult | P | ||||||
| Cardiac Pediatric | P | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| Peripheral | Peripheral vessel | P | ||||||
| Vessel | Other (spec.) |
N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
-
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development -
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients -
Note 6: Abdominal organs and peripheral vessel
-
Note 7: Tissue Harmonic Imaging (THI)
-
Note 8: 3D imaging
-
Note 9: MultiVision (Spatial Compound Imaging)
-
Note 10: Includes Renal, Gynecology/Pelvis
-
Note 11: Panoramic imaging
-
Note 12: ElastoScan
-
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{22}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/22/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. The oval is tilted slightly, giving the logo a dynamic appearance.
DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: EV2-10A for use with HS40
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation (*includes simultaneous B-mode) | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track I only) | Specific(Tracks I & III) | B | M | PWD | CWD | ColorDoppler* | Combined*(Spec.) | Other(Spec.) |
| Ophthalmic | Ophthalmic | |||||||
| Fetal/Obstetrics (See Note 3) | P | P | P | P | Note 1 | Note 2, 7,8, 9 | ||
| Abdominal (See Note 12) | ||||||||
| Intra-operative (See Note 6) | ||||||||
| Intra-operative (Neuro.) | ||||||||
| Fetal Imaging& Other | Laparoscopic | |||||||
| Pediatric | ||||||||
| Small Organ (See Note 5) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Trans-vaginal | P | P | P | P | Note 1 | Note 2, 7,8, 9, 12 | ||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Cardiac) | ||||||||
| Musculo-skel. (Convent.) | ||||||||
| Musculo-skel. (Superfic.) | ||||||||
| Intra-luminal | ||||||||
| Other (See Note 13) | P | P | P | P | Note 1 | Note 2, 7, 8, 9, 12 | ||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Trans-esophageal (Cardiac) | ||||||||
| Other (spec.) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (spec.) |
N= new indication; P= previously cleared by FDA K192505; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: MultiVision (Spatial Compound Imaging) Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
{23}------------------------------------------------
510(k) Premarket Notification - Traditional
5. 510(K) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
- Date Prepared Feb 11, 2019 1.
-
- Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
- Primary Contact Person 3. Scully KIM Regulatory Affairs Specialist Phone: +82.2.2194.1312 Fax: +82.31.8017.9576
Email: scully.kim@samsungmedison.com
-
- Secondary Contact Person Ninad Gujar Director of Regulatory & Quality Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com
-
- Proposed Device
- Proprietary Name: HS40 Diagnostic Ultrasound System
- Common Name: System, Imaging, Pulsed Doppler, Ultrasonic System, Imaging, Pulsed Echo, Ultrasonic Transducer, Ultrasonic, Diagnostic
- Classification: 21 CFR 892.1550 Ultrasonic pulsed doppler imaging system 21 CFR 892.1560 Ultrasonic pulsed echo imaging system 21 CFR 892.1570 Diagnostic ultrasonic transducer
- Product Code(s): IYN, IYO, ITX
- Predicate Device 6.
HS40 Diagnostic Ultrasound System (K191055) - Primary Predicate HS50/HS60 Diagnostic Ultrasound System (K192505)
-
- Device Description
The HS40 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode. M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, 3D imaging mode (real-time 4D imaging mode), Elastoscan Mode or as a
- Device Description
{24}------------------------------------------------
510(k) Premarket Notification - Traditional
combination of these modes. The HS40 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS40 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
-
- Intended Use
The HS40 ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids.
- Intended Use
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
The HS40 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices.
-
- Technology
The HS40 employs the same fundamental scientific technology as its predicate device.
- Technology
-
- Determination of Substantial Equivalence
Comparison to Predicate: The HS40 is substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
- Determination of Substantial Equivalence
-
The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis
-
. The proposed HS40 and predicate HS40(K191055) have the same clinical intended use.
-
. The proposed HS40 and predicate HS40(K191055) have the same imaging modes and modes of operation.
-
The proposed HS40 added some of SW features (S-Detect for Breast, 2D Follicle, ● LaborAssist. CrystalVue) that migrated from predicate HS50/HS60 (K192505)
-
. The transducer EV2-10A is added which is previously cleared in the predicate HS50/HS60 (K192505).
-
The proposed HS40 supports Window 10 Operating System. ●
-
. The system is manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
-
. The system has acoustic power levels which are below the applicable FDA limits.
-
The proposed HS40 and predicate HS40 (K191055) have similar capability in . terms of performing measurements, capturing digital images, reviewing and
{25}------------------------------------------------
510(k) Premarket Notification – Traditional
reporting studies.
- . The proposed HS40 and predicate have been designed in compliance with approved electrical and physical safety standards.
-
- Summary of Non-Clinical Test
The device has been evaluated for acoustic output, biocompatibility, software function, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable FDA guidance and medical device safety standards. The HS40 and its applications comply with the following FDA-recognized standards.
| Reference No. | Title |
|---|---|
| IEC 60601-1 | AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012,C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medicalelectrical equipment - Part 1: General requirements for basic safetyand essential performance (IEC 60601-1:2005, MOD) |
| IEC 60601-1-2 | IEC60601-1-2: 2014(4th Edition) , Medical electrical equipment -Part 1-2: General requirements for basic safety and essentialperformance - EMC |
| IEC 60601-2-37 | IEC 60601-2-37 Edition 2.0 2007, Medical electrical equipment –Part 2-37: Particular requirements for the basic safety and essentialperformance of ultrasonic medical diagnostic and monitoringequipment |
| ISO10993-1 | ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices– Part 1: Evaluation and testing within a risk management process |
| ISO14971 | ISO 14971:2007, Medical devices - Application of risk managementto medical devices |
| NEMA UD 2-2004 | NEMA UD 2-2004 (R2009)Acoustic Output Measurement Standard for Diagnostic UltrasoundEquipment Revision 3 |
12. Summary of Clinical Tests
The subject of this premarket submission, HS40, is not required clinical studies to support substantial equivalence.
-
- Conclusion
Since the predicates device and subject device have a similar intended use and key technological features, the non-clinical data support the safety of the device and demonstrate that the HS40 Ultrasound System should perform as intended in the specified use conditions. Therefore, SAMSUNG MEDISON CO., LTD. concludes that the performance of the subject device is as safe and effective, and is therefore substantially equivalent, to the predicate device(s) that are currently marketed for the same intended use..
- Conclusion
END of 510(K) Summary -
N/A