K Number
K200329

Validate with FDA (Live)

Date Cleared
2021-04-22

(437 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IDCAM Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients esthetics and chewing function. IDCAM implants are intended for single or multiple unit restorations on splinted applications. They are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period.

Device Description

The ID CAM Dental Implant is a tapered conical implant system with two designs. IDCAM M (with mini-threads), and IDCAM ST. IDCAM ST has large flat threads at the top and sharp threads at the bottom and comes in diameters of 3.7, 3.9, 4.2 and 5.2 mm. Lengths of 9.4, 11.4, and 14.4 mm are available. The IDCAM M has mini-threads at the top and with sharp threads below. It comes in 4.2, and 5.2 mm diameter with lengths of 8.5,9.4, 11.4, and 14.4mm. The bottom of both IDCAM designs is convex.

All implants and abutments are made of ASTM F136 Ti 6Al 4V ELI. The implants have a grit blasted and acid etched surface.

ID CAM abutments for single unit restorations are all attached by Morse taper fit as well as a screw. Morse taper cone straight abutments come in 3.6 and 4.2mm diameter with gingival heights of 1.4, 3, and 5mm. Morse taper cone shouldered straight abutments come in 5.4mm diameter at the upper platform with a 3.6mm diameter at the platform and gingival heights of 1.3, 2.2, 3.2, and 5mm. Morse taper angled cone abutments come in 7°, 15°, and 23° angles in diameters of 3.6 and 4.2 and gingival heights of 1.4 (7° only) or 1.63 (15° and 23°), 3, and 5mm. Morse taper angled cones abutments with shoulder come in 7°, 15°, and 23° angles in a diameter of 5.4 at the upper platform with a 3.6mm diameter at platform bottom and gingival heights of 0.9, 3, and 5mm in 7°, and gingival heights of 0.9, 2.35, 3.30, and 5.03mm in 15°, and 23°.

PLAN abutments in 5.4mm diameter at shoulder top with a 3.6mm diameter at shoulder bottom and come in gingival heights of 1.5 and 3mm and there are straight, 15°, and 23° versions. Plan abutments can be used for single or multiple unit restorations but are not used for removable prostheses.

Straight IDUnit abutments, angled IDUnit abutments, IDLoc abutments, and ball attachments are permanent threaded abutments which are for multi-restorations only. IDLOC is 3.6mm diameter at platform and comes in gingival heights of 2.5, and 7.5 mm. Ball attachments are in 3.5 diameter and come in gingival heights of 1, 2.5, 4, or 6mm. Straight IDUnits are in 3.6mm diameter with 4.9mm diameter at top of shoulder and come in gingival heights of 1, 2.5, 4 or 6 mm. 17° angled IDUnits come in 3.6mm diameter with 4.9mm diameter at top of shoulder and in a gingival heights of 1.35,3.02 or 5 mm. 30° angled IDUnits come in 3.6mm diameter with a 4.9mm diameter at top of shoulder and in a gingival heights of 1,3.01 or 5 mm. Straight and angled IDUnit mounted dentures can only be removed by the dentures mounted on IDLoc or ball attachments can be removed by the patient.

Provisory abutments are temporary use abutments which allow placement of a temporary restoration. These come in 3.6mm diameter and varieties of nonrotational (4.8mm diameter at top of shoulder, gingival height of 1.5mm and post height of 7.5mm), nonrotational tall (4mm diameter at top of shoulder, gingival height of 1.5mm and post height of 12.5mm), rotational (gingival height of 1.5mm and post height of 7.5mm) and provisory IDUnit (4.9mm diameter at the base and a post height of 10.5mm).

Healing caps are temporary use abutments used during the healing phase which come in different gingival heights shapes in order to account for tissue thickness differences and space differences. Cylindrical shape healing caps come in diameters of 3.2mm with heights of 3.5 and 5mm, 4 & 5 mm with heights of 2, 4, 6, and 8mm. Conical profile healing caps come in 3.6mm diameter at the platform with a 6mm cone top diameter and 4mm gingival height, or a 3.6mm diameter at platform with a 6mm cone top diameter and gingival height of 6mm. There is also a healing cap for the IDUnit.

AI/ML Overview

This document is a 510(k) Premarket Notification summary for the IDCAM Dental Implants. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving the device meets specific performance acceptance criteria based on clinical outcomes or AI performance.

Therefore, many of the requested details about acceptance criteria, study design, sample sizes, expert involvement, and ground truth for an AI/device performance study are not applicable to this type of regulatory submission for a dental implant.

Here's an analysis of the provided text in relation to your request:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria with numerical performance data in the way you'd expect for an AI or diagnostic device assessing clinical outcomes (e.g., sensitivity, specificity).
Instead, it demonstrates substantial equivalence to predicate devices for a physical implant. The "acceptance criteria" here are more about meeting recognized standards and being comparable to legally marketed devices.

Acceptance Criterium (from common implant testing practices)Reported Device Performance (from document)
Material CompositionAll implants and abutments are made of ASTM F136 Ti 6Al 4V ELI. (Same as predicate: Ti6Al4V)
BiocompatibilitySurface analysis and cytotoxicity testing per ISO 10993-5 was done to show the surface treatment does not adversely change the biocompatibility of the materials.
Mechanical Strength/Fatigue (ISO 14801)IDCAM implants exhibited a run out limit equivalent to other implant systems. (Comparable to predicate: Run out limit is comparable to other implant systems.)
Surface TreatmentGrit blasted and acid etched surface (SLA). (Predicate: SLA or RBM)
Sterilization (Implants)Sterilization validation was conducted on the implants per ISO 11137-1,-2,and -3. Provided sterile by gamma irradiation. (Same as predicate)
Sterilization (Abutments)Abutment steam sterilization validation was done per ISO 17665-2 and ANSI/AAMI ST79. Provided non-sterile with instructions for user to sterilize them. (Same as predicate)
Package Integrity & AgingPackage integrity testing and accelerated aging were conducted per ISO 11607-1 and -2.
Endotoxin TestingEndotoxin testing according to USP 161 was conducted.
Intended UseIDCAM Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices... (Same as predicate)
Design/ConnectionsTapered conical implant system with two designs. Conical connection. (Comparable to predicate: conical)

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify a "test set" sample size in terms of patient data or clinical images. For mechanical testing (like ISO 14801), the sample sizes would be a small number of physical implant components tested in a lab, not patient data. These details are typically in a separate test report, not summarized in the 510(k) summary itself.
  • Data Provenance: The data comes from in-vitro (lab-based) physical and chemical testing of the device components, not from clinical studies on patients. Therefore, concepts like "country of origin of the data," "retrospective," or "prospective" are not applicable here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. "Ground truth" in this context refers to the defined properties of the materials and the performance under standardized mechanical tests, not clinical diagnoses made by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this relates to expert review of clinical data, which was not the basis of this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (dental implant), not an AI diagnostic software. No human reader studies were conducted or required for this type of submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is established by:

  • Adherence to recognized international standards (e.g., ISO 14801 for fatigue testing, ISO 10993-5 for cytotoxicity, ISO 11137 for sterilization).
  • Material specifications (ASTM F136).
  • Comparison to the established, legally marketed predicate device's characteristics and performance under these same standards.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI model. There is no concept of a "training set" in this context.

9. How the ground truth for the training set was established

Not applicable.


Summary of the Study Proving Device Meets "Acceptance Criteria" (Substantial Equivalence):

The "study" in this context is a series of engineering and biocompatibility tests designed to demonstrate that the IDCAM Dental Implants are substantially equivalent to a legally marketed predicate device (SpiralTech Dental Implant System Ultimate Conical K170372, and reference predicates Cortex K090709 and K163385).

  • Objective: To show that the new device is as safe and effective as the predicate device, and does not raise new questions of safety or effectiveness.
  • Methodology:
    • Indications for Use Comparison: The Indications for Use statement for IDCAM Dental Implants was compared directly to that of the predicate device and found to be "same."
    • Material Comparison: The device material (ASTM F136 Ti 6Al 4V ELI) was confirmed to be identical to the predicate device. Biocompatibility testing per ISO 10993-5 was performed.
    • Design Comparison: Detailed design specifications (diameters, lengths, connections, screw types, abutments, etc.) were compared. Minor differences in specific sizes were acknowledged but deemed not to change substantial equivalence given similar use, materials, and technology.
    • Performance Testing (In-Vitro):
      • Dynamic fatigue testing according to ISO 14801: This standard evaluates the endurance limit of endosseous dental implants under cyclic loading. The reported performance was that IDCAM implants "exhibited a run out limit equivalent to other implant systems," specifically comparable to the predicate.
      • Surface Modification Information: Provided per FDA Guidance Document for Endosseous Dental Implants and Abutments.
      • Sterilization Validation: Conducted per ISO 11137-1,-2,and -3 for implants and ISO 17665-2 and ANSI/AAMI ST79 for abutments.
      • Package Integrity Testing and Accelerated Aging: Per ISO 11607-1 and -2.
      • Endotoxin Testing: Per USP 161.
  • Conclusion: Based on these comparisons and performance tests demonstrating equivalence in indications for use, materials, design, and fatigue performance, the FDA determined that the IDCAM Dental Implants are substantially equivalent to the predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

Implants Diffusion International % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172

Re: K200329

Trade/Device Name: IDCAM Dental Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: March 18, 2021 Received: March 23, 2021

Dear Angela Blackwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200329

Device Name IDCAM Dental Implants

Indications for Use (Describe)

IDCAM Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patients and chewing function. IDCAM implants are intended for single or multiple unit restorations on splinted applications. They are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading, These implants can also be used for loading after a conventional healing period.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k Summary April 21, 2021 IDCAM Dental Implants K200329

Name and address: Implants Diffusion International 23 rue Emile Zola Montreuil France 93100 Contact Person: Rony Boukhris Phone Number: +33 (0) 148707048 Name of device: IDCAM Dental Implants Classification Name: Endosseous dental implants CFR: 21 CFR 872.3640 Primary Product Code: DZE Secondary Product Code: NHA

Submission Contact:

Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com

Device Description: The ID CAM Dental Implant is a tapered conical implant system with two designs. IDCAM M (with mini-threads), and IDCAM ST. IDCAM ST has large flat threads at the top and sharp threads at the bottom and comes in diameters of 3.7, 3.9, 4.2 and 5.2 mm. Lengths of 9.4, 11.4, and 14.4 mm are available. The IDCAM M has mini-threads at the top and with sharp threads below. It comes in 4.2, and 5.2 mm diameter with lengths of 8.5,9.4, 11.4, and 14.4mm. The bottom of both IDCAM designs is convex.

All implants and abutments are made of ASTM F136 Ti 6Al 4V ELI. The implants have a grit blasted and acid etched surface.

ID CAM abutments for single unit restorations are all attached by Morse taper fit as well as a screw. Morse taper cone straight abutments come in 3.6 and 4.2mm diameter with gingival heights of 1.4, 3, and 5mm. Morse taper cone shouldered straight abutments come in 5.4mm diameter at the upper platform with a 3.6mm diameter at the platform and gingival

{4}------------------------------------------------

heights of 1.3, 2.2, 3.2, and 5mm. Morse taper angled cone abutments come in 7°, 15°, and 23° angles in diameters of 3.6 and 4.2 and gingival heights of 1.4 (7° only) or 1.63 (15° and 23°), 3, and 5mm. Morse taper angled cones abutments with shoulder come in 7°, 15°, and 23° angles in a diameter of 5.4 at the upper platform with a 3.6mm diameter at platform bottom and gingival heights of 0.9, 3, and 5mm in 7°, and gingival heights of 0.9, 2.35, 3.30, and 5.03mm in 15°, and 23°.

PLAN abutments in 5.4mm diameter at shoulder top with a 3.6mm diameter at shoulder bottom and come in gingival heights of 1.5 and 3mm and there are straight, 15°, and 23° versions. Plan abutments can be used for single or multiple unit restorations but are not used for removable prostheses.

Straight IDUnit abutments, angled IDUnit abutments, IDLoc abutments, and ball attachments are permanent threaded abutments which are for multi-restorations only. IDLOC is 3.6mm diameter at platform and comes in gingival heights of 2.5, and 7.5 mm. Ball attachments are in 3.5 diameter and come in gingival heights of 1, 2.5, 4, or 6mm. Straight IDUnits are in 3.6mm diameter with 4.9mm diameter at top of shoulder and come in gingival heights of 1, 2.5, 4 or 6 mm. 17° angled IDUnits come in 3.6mm diameter with 4.9mm diameter at top of shoulder and in a gingival heights of 1.35,3.02 or 5 mm. 30° angled IDUnits come in 3.6mm diameter with a 4.9mm diameter at top of shoulder and in a gingival heights of 1,3.01 or 5 mm. Straight and angled IDUnit mounted dentures can only be removed by the dentures mounted on IDLoc or ball attachments can be removed by the patient.

Provisory abutments are temporary use abutments which allow placement of a temporary restoration. These come in 3.6mm diameter and varieties of nonrotational (4.8mm diameter at top of shoulder, gingival height of 1.5mm and post height of 7.5mm), nonrotational tall (4mm diameter at top of shoulder, gingival height of 1.5mm and post height of 12.5mm), rotational (gingival height of 1.5mm and post height of 7.5mm) and provisory IDUnit (4.9mm diameter at the base and a post height of 10.5mm).

Healing caps are temporary use abutments used during the healing phase which come in different gingival heights shapes in order to account for tissue thickness differences and space differences. Cylindrical shape healing caps come in diameters of 3.2mm with heights of 3.5 and 5mm, 4 & 5 mm with heights of 2, 4, 6, and 8mm. Conical profile healing caps come in 3.6mm diameter at the platform with a 6mm cone top diameter and 4mm gingival height, or a 3.6mm diameter at platform with a 6mm cone top diameter and gingival height of 6mm. There is also a healing cap for the IDUnit.

Indications for Use: IDCAM Dental Implants are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth , in order to restore patients esthetics and chewing function. IDCAM implants are intended for single or multiple unit restorations on splinted applications. They

{5}------------------------------------------------

are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period.

Testing Summary: Dynamic fatigue testing according to ISO 14801 was conducted to determine the abutments and implants are strong enough for their intended use. IDCAM implants exhibited a run out limit equivalent to other implant systems. Surface modification information was provided per the FDA Guidance Document for Endosseous Dental Implants and Abutments. Surface analysis and cytotoxicity testing per ISO 10993-5 was done to show the surface treatment does not adversely change the biocompatibility of the materials are ones common in dental implant systems and are appropriate materials for devices used with dental implants. Implants and abutments are made of titanium alloy which meets ASTM F136. Sterilization validation was conducted on the implants per ISO 11137-1,-2,and -3. Abutment steam sterilization validation was done per ISO 17665-2 and ANSI/AAMI ST79. Package integrity testing and accelerated aging were conducted per ISO 11607-1 and -2. Endotoxin testing according to USP 161 was conducted.

Primary Predicate Device: SpiralTech Dental Implant System Ultimate Conical K170372 Reference Predicates: Cortex K090709 and K163385

Substantial Equivalence:

Implant SystemComparison TableID CAM DentalImplants (SubjectDevice)SpiralTech DentalImplant SystemK170372Ultimate Conical
Indications for UseIDCAM DentalImplants areendosseous implantsintended to besurgically placed in theupper or lower jawarches to providesupport for prostheticdevices, such as anartificial tooth , inorder to restorepatients esthetics andchewing function.The Spiraltech DentalImplants areendosseous implantsintended to besurgically placed inthe upper or lowerjaw arches to providesupport for prostheticdevices, such as anartificial tooth, inorder to restorepatient's esthetics andchewing function.Spiraltech implantsare intended for single
IDCAM implants areor multiple unit
intended for single orrestorations on
multiple unitsplinted or non-
restorations onsplinted applications.
splinted or non-The implants ESi
splinted applications.Dynamic and Ultimate
are intended for
They are intended forimmediate loading
immediate loadingwhen good primary
when good primarystability is achieved,and with appropriate
stability is achieved,occlusive loading.
and with appropriateThese implants [along
occlusive loading.with Premium and
These implants canOne Piece] can also be
also be used forused for loading after
loading after aa conventional healing
conventional healingperiod.
period.Solo One Piece 3.0
and 3.3 implants,
Ultimate (conical) 3.0
implants, and ESi
(conical) 3.0 implants
are intended to
replace a lateral
incisor in the maxilla
and/or a central or
lateral incisor in the
mandible. Mandibular
central and lateral
incisors must be
splinted if using two
or more 3.0 and/or3.3 implants adjacent
to one another.
MaterialTi6Al4VTi6Al4V
IDCAM M
Diameter of Implants4.2, 5.23.0, 3.5, 4.3, 5.0, 6.0
Implant Lengths8.5, 9.4, 11.4, 14.48, 10, 11.5, 13, 15(3.0mm diameter notin 8 length)
Surface TreatmentSLASLA or RBM
Sterilization of ImplantsProvided sterile by gamma irradiationProvided sterile by gamma irradiation
Sterilization of abutmentsProvided non-sterile with instructions for user to sterilize themProvided non-sterile with instructions for user to sterilize them
ConnectionConicalConical
Spiral Implant DesignIDCAM MUltimate Conical
ISO 14801 Fatigue TestRun out limit is the same or higher than those of other implant systemsRun out limit is comparable to other implant systems.
IDCAM ST
Diameter of Implants3.7, 3.9, 4.2, 5.23.0, 3.5, 4.3, 5.0, 6.0
Implant Lengths9.4, 11.4, 14.48, 10, 11.5, 13, 15 (3.0mm diameter not in 8 length)
Surface TreatmentSLASLA or RBM
Sterilization of ImplantsProvided sterile by gamma irradiationProvided sterile by gamma irradiation
Sterilization of abutmentsProvided non-sterile with instructions for user to sterilize themProvided non-sterile with instructions for user to sterilize them
ConnectionConicalConical
Spiral Implant DesignIDCAM STUltimate Conical
ISO 14801 Fatigue TestRun out limit is the same or higher than those of other implant systemsRun out limit is comparable to other implant systems.

IDCAM Dental Implants are substantially equivalent to SpiralTech Dental Implant System in indications for use, materials, design, and fatigue performance,

{6}------------------------------------------------

{7}------------------------------------------------

Conical ImplantPartsIDCAM DentalImplantsSpiralTech DentalImplant SystemConicalConnection PartsCortex K090709and K163385ConicalConnectionParts
Cover screwCover screw forIDCAM 3.6mmdiameterCover screw NPand RP

{8}------------------------------------------------

Abutment screwIDCAM andIDUnit 3.6mmabutment screwsNP and RPabutment screws
Multi-UnitAbutments*IDUnit 3.6mmdiameter atplatform 4.9mmdiameter at topof shoulderabutments ingingival heights of1,2.5, 4, and 6mmNP, RP, and WPMulti-unitabutments inheights of 1, 2, 3,4, and 5 mm
17° Multi-UnitAbutments *IDUnit 17° Multi-Unit Abutments3.6mm diameterat platform4.9mm diameterat top of shoulderin gingival heightsof 1.35,3.02, 5mmNP and RP 18°Multi-UnitAbutmentsin heights of 1, 2,3, 4, 5 mm
30° Multi-UnitAbutments *IDUnit 30° Multi-Unit Abutments3.6mm diameterat platform4.9mm diameterat top of shoulderin gingival heightsof 1,3.01, 5 mmNP and RP 30°Multi-UnitAbutmentsin heights of 1, 2,3, 4, 5 mm
Ballattachments*3.5 mm diameterat platform ingingival heights of1, 2.5, 4, and6mmNP and RP Ballattachments inheights of1,2,3,4,5, and6mm
Healing Caps NPCylindricalHealing CapNarrow 3.2mmdiameter in3.5,5.5 mmgingival heightsNP Healing Cap in2,3,4,5 and 6 mmheight
Healing Caps RPCylindricalHealing CapRegular 4mmRP Healing Cap in2,3,4,5, and 6mm height
diameter in 2,4, 6and 8 mmgingival heights
Healing Cap WPCylindricalHealing Cap Wide5mm diameter in2, 4, 6, 8mmgingival heights5.8mm Healingcap in 3,4,5,6mm
Conical ProfileHealing CapRegular3.6mm diameterat platform 4mmcone topdiameter and 6mm gingivalheightNP Healing Cap in2,3,4,5 and 6 mmheight
Conical ProfileHealing Cap Wide3.6mm diameterat platform 6mmcone topdiameter and 6mm gingivalheightNP Healing Cap in2,3,4,5 and 6 mmheight
Healing Cap forMulti-UnitIDUnit healingcapTitanium healingcap for multi-unit 5.0mmdiameter 4.8mmheight
StraightAbutmentMorse taper conestraightabutment3.6,4.2mmdiameter atplatformgingival heights of1.4, 3,5 mmRP Flat StandardAbutment withheight of 6,7,9and 11 mm
WP ShoulderAbutmentMorse tapershoulderedstraight coneabutment 5.4mmdiameter atupper platform3.6 mm diameterat platformbottom gingivalNP and RP Flatwith straightshoulderabutment inheights of 1,2,and 3 mmWP anatomicabutment5.8mm diameterin heights of 1, 2,3, and 4mm

{9}------------------------------------------------

{10}------------------------------------------------

NP and RP 7°Abutmentheights of 1.3,2.23.2, 5 mmHeight betweenplatform top andbottom 3.7 +gingival heightMorse taper 7°angled coneabutment3.6 and 4.2mmdiameter withgingival heights of1.4,3,5 mmNP and RP 15°Abutment withheights of 1,2,3mmNP and RP 15°AnatomicAbutment withheights of 1,2,3,4 mm
WP 7° Abutmentwith shoulderMorse taper 7°angled coneabutment withshoulder5.4 mm diameterat upper platform3.6mm diameterat platformbottomgingival heights of0.9,3,5 mmHeight betweenplatform top andbottom 3.7 +gingival heightWP 15°AnatomicAbutment withheights of 1,2,3,4 mm
NP and RP 15°AbutmentMorse taper 15°angled coneabutment3.6 and 4.2mmdiameter withgingival heights of1.63, 3, 5 mmNP and RP 15°Abutment withheights of 1,2,3mmNP and RP 15°AnatomicAbutment withheights of 1,2,3,4 mm
WP 15°Abutment withshoulderMorse taper 15°angled coneabutment withshoulder5.4 mm diameterat upper platform3.6mm diameterat platformbottomWP 15°AnatomicAbutment withheights of 1,2,3,4 mm
NP and RP 23°Abutmentgingival heights of0.9, 2.35, 3.30,5.03 mmHeight betweenplatform top andbottom 3.7 +gingival heightNP and RP 25°Abutment withheights of 1,2,and 3mmNP and RP 25°AnatomicAbutment withheights of 2,3, 4mm
Morse taper 23°angled coneabutment3.6 and 4.2mmdiameter gingivalheights of 1.63, 3,5 mm
WP 23°Abutment withshoulderMorse taper 23°angled coneabutment withshoulder5.4 mm diameterat upper platform3.6mm diameterat bottom ofplatformgingival heights of0.9, 2.35, 3.3,5.03 mmHeight betweenplatform top andbottom 3.7 +gingival heightWP 25°AnatomicAbutment withheights of 1,2,3,4 mm
LocatorAbutments *IDLoc attachment3.6mm diameterat platformigingival heights of2.5,4, 5.5, and 7.5mmIPI NP and RP inheights of1,2,3,4,5 and 6mm
WPStraightAbutment withshoulderPlan straightabutment withshoulder 5.4mmdiameter atshoulder top3.6mm diameterat shoulderbottomWP anatomicabutment5.8mm diameterin heights of 1, 2,3, and 4mm
gingival heights of1.5 and 3mm
WP 15° angledabutment withshoulderPlan 15° angledabutment withshoulder 5.4mmdiameter atshoulder top3.6mm diameterat shoulderbottomgingival heights of1.5 and 3mmWP 15°AnatomicAbutment withheights of 1,2,3,4 mm
WP 23° angledabutment withshoulderPlan 23° angledabutment withshoulder 5.4mmdiameter atshoulder top3.6mm diameterat shoulderbottomgingival heights of1.5 and 3mmWP 25°AnatomicAbutment withheights of 1,2,3,4 mm
TemporaryAbutments3.6mm diameterprovisionalabutments inrotational(gingival height1.5, post height7.5), non-rotational(gingival height1.5, post height7.5, 4.8mmdiameter at topof shoulder), non-rotational tall(gingival height1.5mm, postheight 12.5mm,4.0mm diameterat top ofshoulder) andIDUnit (4.9 mmNobel BiocareK161435Temporary SnapAbutmentsengaging, non-engaging, andmulti-unit
diameter at base, post height 10.5)

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

*These models of abutments are not for single crown use.

Conclusion:

IDCAM Dental Implants are substantially equivalent to SpiralTech Dental Implant System. They both have the same indications for use, are of the same material, and have conical connections. Performance testing demonstrates substantial equivalence to the identified predicate devices. The abutments, healing caps, and angled abutments are offered in similar designs and heights. Any abutments not found in the predicate device system are found in the reference device system. The design and size differences between predicate or reference devices and the subject devices are only minor differences in geometry and size so given the use, materials and technology is the same they do not change the substantial equivalence.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.