K Number
K200324

Validate with FDA (Live)

Device Name
NHancer Rx
Date Cleared
2020-08-06

(178 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NHancer Rx Dual Lumen Rapid Exchange catheter with hydrophilic coating is intended to support a guide wire during access of coronary and/or peripheral vasculature and allows for exchange of guide wires and provides a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.

Device Description

The NHancer™Rx is a dual lumen catheter intended to use in the coronary and/or peripheral vasculature. The NHancer™Rx is a 0.014" Rapid Exchange (Rx) guide wire support catheter. The NHancer™Rx has an over the wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal end. The NHancer™Rx has an effective length of 135 cm and is compatible with 0.014"/0.36mm or smaller guide wires. The NHancer™Rx has two depth markings located at 95 cm and 105 cm from the distal tip. The NHancer™Rx has a radiopaque marker identifying the distal end of the catheter and a second radiopaque marker located 8 mm from the distal tip identifies the distal end of the OTW lumen.

To reduce friction in the guide catheter and vasculature, the distal 15 cm of the shaft is fitted with a hydrophilic coating. To reduce friction of the guide wire in the entire lumens of the NHancer Rx are coated with MDX (silicone) coating.

To ease catheter loading into the hemostasis device and the guiding catheter a removable stylet (9) is placed in the OTW lumen

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (NHancer Rx Dual Lumen Rapid Exchange catheter). It details the device's characteristics, intended use, and a comparison to a predicate device, along with a summary of performed studies.

However, this document does not contain information about an AI/ML-based medical device study. Therefore, I am unable to provide a response that directly addresses the specific requirements of your prompt regarding acceptance criteria and a study proving device performance for an AI/ML system.

The "studies" mentioned in the document are performance and biocompatibility tests for a physical medical catheter, not a software algorithm. These tests verify physical properties like flexibility, tensile strength, and material biocompatibility, which are fundamentally different from the performance metrics (e.g., sensitivity, specificity, AUC) and study designs (e.g., MRMC, standalone) associated with demonstrating the effectiveness of an AI/ML device.

Specifically, the document does not include:

  • A table of acceptance criteria and reported device performance for an AI/ML system.
  • Sample sizes, data provenance, or ground truth details for an AI/ML test set.
  • Information on experts, adjudication methods, or MRMC studies for AI/ML.
  • Details on training set size or ground truth establishment for an AI/ML model.

Therefore, paraphrasing or extracting information from the given text to fit your prompt's structure for an AI/ML device would be misrepresentative.

{0}------------------------------------------------

August 6, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

IMDS Operations B.V. Edwin Schulting CEO Ceintuurbaan Noord 150 Roden, 9301 NA NI

Re: K200324

Trade/Device Name: NHancer Rx Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 2, 2020 Received: July 6, 2020

Dear Edwin Schulting:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200324

Device Name NHancer Rx

Indications for Use (Describe)

The NHancer Rx Dual Lumen Rapid Exchange catheter with hydrophilic coating is intended to support a guide wire during access of coronary and/or peripheral vasculature and allows for exchange of guide wires and provides a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

Date Prepared: August 6th, 2020

Submitter's Name / Contact Person
ManufacturerContact Person
IMDS Operations BVFlorence Wagter
Ceintuurbaan Noord 150Director of Quality and Regulatory
9301 NZ Roden, The NetherlandsTel: 0031651453880
Establishment Registration #3007740583Fax: 0031508200231

General Information

Trade NameNHancer Rx
Common/Usual NameDual lumen catheter
Classification NameCatheter, percutaneous
Predicate DeviceK162467, Twin-Pass torque, dual access catheter, Vascularsolutions, Inc.
Reference DeviceK121077, NHancer guidewire support catheter

Device Description

The NHancer™Rx is a dual lumen catheter intended to use in the coronary and/or peripheral vasculature. The NHancer™Rx is a 0.014" Rapid Exchange (Rx) guide wire support catheter. The NHancer™Rx has an over the wire (OTW) lumen that runs the length of the catheter and a rapid exchange (RX) delivery lumen on the distal end. The NHancer™Rx has an effective length of 135 cm and is compatible with 0.014"/0.36mm or smaller guide wires. The NHancer™Rx has two depth markings located at 95 cm and 105 cm from the distal tip. The NHancer™Rx has a radiopaque marker identifying the distal end of the catheter and a second radiopaque marker located 8 mm from the distal tip identifies the distal end of the OTW lumen.

To reduce friction in the guide catheter and vasculature, the distal 15 cm of the shaft is fitted with a hydrophilic coating. To reduce friction of the guide wire in the entire lumens of the NHancer Rx are coated with MDX (silicone) coating.

To ease catheter loading into the hemostasis device and the guiding catheter a removable stylet (9) is placed in the OTW lumen

Intended Use

The NHancer Rx Dual Lumen Rapid Exchange catheter with hydrophilic coating is intended to support a guide wire during access of coronary and/or peripheral vasculature and allows for exchange of guide wires and provides a conduit for the delivery of diagnostic contrast agents. The catheter is not intended for use in the neurovasculature.

Technological Characteristics Comparison

The NHancer Rx is similar in design to the predicate device and both are dual lumen, percutaneous catheters intended to access discrete regions of the coronary and peripheral vasculature, facilitate placement and exchange of guidewires and sub selectively infuse agents.

#ItemNHancer Rxdual lumen guidewire supportcatheterTwin-Pass Torque dual accesscatheter
Model numberNRX14135185201
1Type clinically basedDual lumen guidewire support catheterDual lumen guidewire support catheter
2Shaft materialPolymericPolymeric

{4}------------------------------------------------

3Shaft reinforcementStainless steelStainless steel
4OTW Guidewire lumens11
5Rapid Exchange guidewire lumens11
6Number of guidewire distal exit ports22
7Strain Reliefpresentpresent
8Hub and Luer LockFemale, 6% taper and screw conformNEN EN ISO 80369-7Female, 6% taper and screw conformNEN EN ISO 80369-7
9Radiopaque marker materialTungstenPt/Ir
10Effective Length or Usable Length135 cm135 cm
11Exit Marker location (from tip)95 and 105 cm95 and 105 cm
12Guidewire compatibility0.014 inch0.014 inch
13Maximum injection pressure300 psi300 psi
14Minimum Guiding Catheter size5 Fr5 Fr
15Tip design / shapeStraightStraight
16Tip materialPolymericPolymeric
17Hydrophilic coating distal shaftpresentpresent
18Hydrophilic coating materialPUR / PVPhydrophilic coatingNot known
19Inner lumen coatingHydrophobicNot known
20Removable stylet in OTW lumenYesNo
21Distal shaft designOvalRound

With the exception of dimensional, material and package configuration differences, the NHancer Rx is similar in design and technological characteristics to the predicate device. The dimensional, material and package configuration differences were successfully evaluated in performance tests.

{5}------------------------------------------------

Substantial Equivalence and Summary of Studies

The technological differences between the subject and predicate devices have been evaluated through performance and biocompatibility tests and results did not raise new questions of safety or effectiveness. The NHancer Rx dual lumen catheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design has been verified through the following tests:

    1. Kind resistance/ flexibility
    1. Guide wire insertion
    1. Radiopacity
    1. Distal tip length
    1. Tensile strength
    1. Catheter body burst
    1. Contrast medium flow rate
    1. Leak testing
    1. Effective length
    1. Shaft inner diameter
    1. Outer diameter
    1. Surface coating lubricity
    1. Coating Integrity
    1. Coating Particulate Evaluation
    1. Packaging integrity

Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1:

  • Cytotoxicity ●
  • Sensitization
  • Irritation ●
  • Systemic toxicity
  • Pyrogenicity
  • Hemocompatibility

The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues.

Therefore, the NHancer Rx dual lumen catheter is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).