K Number
K200169

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-05-14

(112 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VFS1 Bipolar Electrosurgical Generator is indicated for use in micro, macro, and endoscopic bipolar irrigation, cutting and coagulating of tissue and the coagulation of blood vessels in all types of surgery.

Device Description

The VFS1 Bipolar Electrosurgical Generator (referred to as generator or VFS1 Generator) is designed to cut and coagulate living human tissue during surgical procedures. The system provides 1 MHz Radio Frequency (RF) energy that is user/operator selectable based on the surgical procedure undertaken. The generator is configured for bipolar output and is earth isolated, to minimize leakage current, thus providing increased patient comfort and safety. The system will be provided with safety features that are continuously monitored and will provide warnings and/ or alerts when critical parameters exceed specified limits.

The VFS1 Generator is designed to cut and coagulate living tissue during surgical procedures. The system includes a bipedal footswitch. The generator is equipped with a tone generator to provide audible feedback each time cutting, or coagulation is performed. The generator incorporates a variable-rate irrigation pump to provide saline flow during procedures.

The VFS1 Generator is compatible with a variety of reusable bipolar products; i.e., insulated mirror finish bipolar forceps, disposable nonstick bipolar instruments as well as disposable bipolar cords and tubing products.

Items that are included in the system are:

  • VFS1 generator -CO 840
  • . Footpedal- CO 841
AI/ML Overview

The provided text describes the 510(k) premarket notification for the "VFS1 Bipolar Electrosurgical Generator." This document outlines the device's technical specifications, intended use, and comparison to predicate devices, along with summaries of non-clinical performance testing.

However, the nature of this document (a 510(k) summary for an electrosurgical generator) means that the types of "acceptance criteria" and "study" details requested in your prompt (which relate to AI/Machine Learning models for diagnostic purposes, involving concepts like sample size for test/training sets, expert ground truth, MRMC studies, etc.) are not applicable to this medical device.

Electrosurgical generators are physical devices that perform cutting and coagulation using electrical energy. Their performance is evaluated through engineering tests, electrical safety standards compliance, and bench/ex vivo testing on tissue, rather than the statistical performance metrics used for AI diagnostic tools.

Therefore, I cannot extract the information you've requested about acceptance criteria and studies (as defined by your prompt's context) from this document.

For clarity, here's why each point in your request is not present or applicable:

  1. Table of acceptance criteria and reported device performance: The document implicitly states that tests "met specifications" or were "acceptable." There isn't a table comparing a specific performance metric against a quantitative acceptance threshold in the way an AI diagnostic model would have (e.g., "sensitivity > 90%"). Instead, it confirms functional compliance.
  2. Sample size for the test set and data provenance: Not applicable. The "test set" here refers to hardware components and ex vivo tissue samples, not patient data for an AI model.
  3. Number of experts and their qualifications for ground truth: Not applicable. Ground truth for an electrosurgical device is about its physical effects on tissue, assessed typically by engineers and potentially surgeons, not by interpretation of images/data by multiple experts.
  4. Adjudication method for the test set: Not applicable. There's no subjective interpretation requiring adjudication.
  5. MRMC comparative effectiveness study: Not applicable. This is for evaluating changes in human reader performance with AI assistance, which is irrelevant for an electrosurgical generator.
  6. Stand-alone (algorithm only) performance: Not applicable. This is a hardware device, not a standalone algorithm.
  7. Type of ground truth used: Ex vivo testing on tissue samples to demonstrate cutting and coagulation effect. This is a physical outcome, not a diagnostic classification.
  8. Sample size for the training set: Not applicable. There is no AI model being trained.
  9. How ground truth for the training set was established: Not applicable for the same reason as above.

In summary, the provided document describes a traditional medical device (electrosurgical generator), and the requested information about acceptance criteria and studies are phrased in the context of AI/Machine Learning diagnostic devices. These frameworks are fundamentally different.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 14, 2020

FHC, Inc. % Craig Pagan, Consultant C2C Development, LLC 1135 W NASA Blvd, Suite 500 Melbourne, Florida 32901

Re: K200169

Trade/Device Name: VFS1 Bipolar Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 10, 2020 Received: April 14, 2020

Dear Craig Pagan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K200169

Device Name

VFS1 Bipolar Electrosurgical Generator

Indications for Use (Describe)

The VFS1 Bipolar Electrosurgical Generator is indicated for use in micro, and endoscopic bipolar irrigation, cutting and coagulating of tissue and the coagulation of blood vessels in all types of surgery.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY SECTION 5

  1. Submitter's name, address, telephone number, a contact person, and the date the summary was prepared:

FHC, Inc. 1201 Main Street Bowdoin, ME-04287 Tel: 207-666-5651 Fax: 207-666-8292

Contact: Kelly Moeykens

Preparation Date: April 6th, 2020

2. Name(s) of the Device:

Proprietary/Trade Name:VFS1 Bipolar Electrosurgical Generator
Common Name:Bipolar Electrosurgical Generator or Electrosurgical Generato
Regulation Number21 CFR 878.4400
Regulation Name:Electrosurgical device
Classification PanelGeneral and Plastic Surgery
Product Code:GEI
Regulatory Class:II

3. Legally Marked Predicate Device to which the submitter claims substantial equivalence:

Trade Name:Malis Bipolar 2000 Electrosurgical Generator
510(k) #:K050364
Product Code:GEI
Decision Date:March 9th, 2005
Trade Name:MALISTM Irrigation Module 1000
510(k) #:K033499
Product Code:GEI
Decision Date:November 28th, 2003

{4}------------------------------------------------

4. Description of device:

The VFS1 Bipolar Electrosurgical Generator (referred to as generator or VFS1 Generator) is designed to cut and coagulate living human tissue during surgical procedures. The system provides 1 MHz Radio Frequency (RF) energy that is user/operator selectable based on the surgical procedure undertaken. The generator is configured for bipolar output and is earth isolated, to minimize leakage current, thus providing increased patient comfort and safety. The system will be provided with safety features that are continuously monitored and will provide warnings and/ or alerts when critical parameters exceed specified limits.

The VFS1 Generator is designed to cut and coagulate living tissue during surgical procedures. The system includes a bipedal footswitch. The generator is equipped with a tone generator to provide audible feedback each time cutting, or coagulation is performed. The generator incorporates a variable-rate irrigation pump to provide saline flow during procedures.

The VFS1 Generator is compatible with a variety of reusable bipolar products; i.e., insulated mirror finish bipolar forceps, disposable nonstick bipolar instruments as well as disposable bipolar cords and tubing products.

ltems that are included in the system are:

  • VFS1 generator -CO 840
  • . Footpedal- CO 841
C0840Valley Forge ScientificVFS1 BipolarElectrosurgicalGenerator
C0841Electrosurgical FootPedal

{5}------------------------------------------------

5. Statement of Indications for Use:

The VFS1 Bipolar Electrosurgical Generator is indicated for use in micro, macro, and endoscopic bipolar irrigation, cutting and coagulating of tissue and the coagulation of blood vessels in all types of surgery.

6. Intended Use

The VFS1 Bipolar Generator is intended for use in surgical procedures for cutting tissue and coagulating blood vessels. Procedures must be performed by a surgeon with experience in the use of electrosurgical devices and generators.

TechnologicalCharacteristicsValley Forge Scientific(division of FHC Inc.)VFS1 ElectrosurgicalGeneratorValley Forge ScientificCorp.Malis 2000 BipolarElectrosurgical SystemPREDICATEMalis Irrigation Module1000
ModalitiesElectrosurgical Cuttingand CoagulationIrrigationElectrosurgical Cuttingand CoagulationIrrigation
Environment ofUseHospitals and ClinicalSettings whereelectrosurgery is used.Hospitals and ClinicalSettings whereelectrosurgery is used.Hospitals and ClinicalSettings whereelectrosurgery is used.
Power SourceAC Voltage from wallreceptacleAC Voltage from wallreceptacleAC Voltage from wallreceptacle
Device (System)Components- Generator- Foot pedal- Power Cord- Cord/Tubing Set- IrrigationModule(check)- Generator- Foot pedal- Power Cord- Cord/Tubing Set- Peristaltic Pump- Foot Pedal
LCD Display, displays:LCD Display, displays:
Cut and Coag PowerSettingsCut and Coag PowerSettings
Power UnitsPower Units
Display ScreenDetailsTone Volume andBrightness SettingsTone and Voice Volumeand Brightness SettingsBuilt in Numeric display
System Errors andWarningsSystem Errors andWarnings
Irrigation SettingAbsent

7. Comparison of Technological Characteristics to Predicate Devices

{6}------------------------------------------------

TechnologicalCharacteristicsValley Forge Scientific(division of FHC Inc.)VFS1 ElectrosurgicalGeneratorPREDICATEValley Forge ScientificMalis 2000 BipolarElectrosurgical SystemPREDICATEMalis Irrigation Module1000
Pump Head TypePeristalticN/APeristaltic
Motor TypeDC electric motorN/ADC electric motor
IrrigationDelivery MethodTubing set with industrystandard connectorsN/ATubing set with industrystandard connectors
Patient interfaceBipolar forcepsN/ABipolar forceps
Liquid SourceSaline bagN/ASaline bag
Delivery RateSlow drip, medium drip,fast dripN/ASlow drip, medium drip,fast drip
RF Output ModeBipolarBipolarN/A
RF Output PowerRangeCut: 0 to 100 WattsCoag: 0 to 50 WattsCut: 0 to 300 WattsCoag: 0 to 50 WattsN/A
RF OutputWaveformsCut: 1 MHz sinewaveCoag: 1 MHz dampedaperiodicCut: 1 MHz sinewaveCoag: 1 MHz dampedaperiodicN/A
Output PowerControlsRotary SwitchesPushbutton SwitchesRotary SwitchesRotary Switches
Power ActivationFoot pedalFoot pedal and FingerswitchesFoot pedal

8. Performance Data

Non-clinical: Basic mechanical and functional capabilities were tested for the subject electrosurgical generator to determine substantial equivalence to the predicate device. This also includes a Comparibility Study of the subject device and the predicate device. Report Refer to section 18 for details.

The following is a summary of components and testing performed:

. VCP-09-103 – VFS1 Hardware Verification Testing

Testing and inspection performed on the VFS1 Electrosurgical Generator, to verify compliance with the Product Development Specification, DSI-09-100. Testing included:

  • о Input Voltage
  • Input Frequency o
  • Input Protection o

{7}------------------------------------------------

  • Output Waveform O
  • RF Power Output O
  • RF Output vs Load Resistance O
  • O RF Overcurrent Protection
  • Over-dosage Protection O
  • Irrigation O
  • Real Time Clock O
  • Mechanical O
  • Operating Conditions O
  • Storage Conditions O
  • O Cooling
  • Transportation O
  • Shock O
  • Cleaning O
  • Front Panel Connections O
  • Rear Panel Connections O
  • Power Cord O
  • O Front Panel Controls
  • O Rear Panel Controls
  • Display O

All testing met specifications.

VCP-09-109 – VFS1 Comparibility Study Report ●

Testing was performed on the VFS1 Electrosurgical Generator to verify proper operation of the generator with all the components and accessories working together as a system. Testing included Ex Vivo testing for Cut and Coag using steak, liver and kidney tissue samples. All testing was found to be acceptable.

● ROH-09-100 - VFS1 ROHS Verification Report

RoHS Compliance was verified for all components and subassemblies of the VFS1 Surgical Generator, Model CO840.

VDP-09-100 – Shipping Test Report ●

Testing was performed to confirm that the external packaging configuration for the VFS1 Electrosurgical Generator provides a high probability of safe and intact arrival at the destination by confirming that the integrity of the packaging and its components withstand simulated shipping and handling stress and by demonstrating that the shipping and handling stresses have no adverse effect on the VFS1 Generator performance. Testing was performed per ISTA-3A Shipping Test and met all requirements.

Clinical: No clinical data was used to determine substantial equivalence.

{8}------------------------------------------------

9. Conclusion

Based on the non-clinical performance data performed comparing the VFS1 Electrosurgical Generator to the predicate devices (Malis 2000 Bipolar Electrosurgical System and the Malis Irrigation Module 1000), it is concluded that the device is as safe, as effective, and performs as well as or better than the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.