K Number
K200156

Validate with FDA (Live)

Device Name
Bright Low Flow
Manufacturer
Date Cleared
2020-10-27

(279 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. Class I, III, V (non-stress area, minimally invasive restoration)
  2. Composite resin repair
Device Description

Bright Flow is a light-cured flowable composite resin. It comprises two different types of flowability (Low Flow and High Flow) and 9 shades depending on the intended use, which enables aesthetic and durable outcomes for anterior and posterior composite restorations.

AI/ML Overview

This document is a 510(k) summary for the medical device "Bright Low Flow", a light-cured flowable composite resin manufactured by GENOSS Co., Ltd. It declares that the device is substantially equivalent to a predicate device (K091388 G-aenial Universal Flo) and provides performance data to support this claim.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document presents two tables related to performance criteria: Biocompatibility and Performance Bench Testing.

Biocompatibility Acceptance Criteria and Performance:

No.TestMethodAcceptance criteriaP/F (Result)
1CytotoxicityISO 10993-5None cytotoxicityPass
2IrritationISO 10993-10None oral irritationPass
3SensitizationISO 10993-10None sensitizationPass
4Acute systemic toxicityISO 10993-11None systemic toxicityPass
5GenotoxicityISO 10993-3None genotoxicityPass
6ImplantationISO 10993-6BiocompatiblePass
7Chronic toxicityISO 10993-11No chronic toxicityPass

Performance Bench Testing Acceptance Criteria and Performance:

No.ItemsMethodAcceptance CriteriaResult
1VisualISO 4049No impurities and No specific changesNo impurities and No specific changes
2CapacityISO 4049Capacity error of; Standard Capacity < ±5%Size error of; Standard Size < ±5% H 0.43%, Ø 0.03%
3PackageISO 4049No damageNo damage
4Sensitivity to Ambient LightISO 4049Must be physically uniformUniformity
5Depth of CureISO 4049More than 1.5 mmAverage: 2.9 mm
6ShadeISO 4049Must be shade uniformUniformity
7Color StabilityISO 4049Color should be stableStable
8Flexural StrengthISO 4049More than 80 MPaAverage: 142 MPa
9Water SorptionISO 4049Less than 40 µg/mm3Average: 23 µg/mm3
10SolubilityISO 4049Less than 7.5 µg/mm3Average: 0.9 µg/mm3
11Radio-opacityISO 4049More than Al≥ Al (1.81)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for each of the biocompatibility and performance bench tests. However, the tests are identified by their ISO standard methods (e.g., ISO 10993-5, ISO 4049), implying that the testing was conducted according to these internationally recognized standards, which typically define appropriate sample sizes.

Data Provenance: The document indicates that the submitter is "GENOSS Co., Ltd." located in "Suwon-si, Gyeonggi-do, Korea." This suggests the data was generated in South Korea, or at least commissioned by a South Korean company. The tests are referred to as "Biocompatibility testing on the proposed Bright Low Flow has been completed" and "The proposed Bright Low Flow was evaluated using the following performance bench testing," implying these are prospective tests performed on the device itself for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests performed are laboratory-based physical and chemical property evaluations or biological assays (biocompatibility) rather than interpretations requiring human expert consensus on a test set.

4. Adjudication Method for the Test Set

This information is not applicable as the tests are objective, laboratory-based measurements or biological assays, not interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a "light-cured flowable composite resin" (dental material), not an AI diagnostic or assistive tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a dental material, not an algorithm.

7. The Type of Ground Truth Used

For the biocompatibility tests, the "ground truth" is based on the biological response observed in the specific assays according to the ISO 10993 standards (e.g., cell viability for cytotoxicity, skin reaction for irritation, systemic effects for acute systemic toxicity).

For the performance bench tests, the "ground truth" is the quantitative measurement of the physical/chemical properties of the material (e.g., depth of cure in mm, flexural strength in MPa, water sorption in µg/mm3) as determined by the specified ISO 4049 standard methods. This represents direct measurement of inherent material properties.

8. The Sample Size for the Training Set

This information is not applicable. As a dental material, there is no "training set" in the context of machine learning or AI models. The device's properties are inherent to its formulation and manufacturing process, evaluated through scientific testing.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 27, 2020

Genoss co.,Ltd. Hong Jeon Manager 1F, Gyeonggi R&DB Center, 105 Gwanggyo-ro, Yeongtong-gu, Suwon-si Suwon-si, Gyeonggi-do 16229 South Korea

Re: K200156

Trade/Device Name: Bright Low Flow Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: Class II Product Code: EBF Dated: August 3, 2020 Received: August 3, 2020

Dear Hong Jeon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GENO" are in a dark green color, while the letters "OSS" are in a lighter green color. The font is bold and slightly italicized, giving the word a dynamic appearance.

Indication for use

510(k) Number: K200156

Device Name: Bright Low Flow

Indication for use:

  1. Class I, III, V (non-stress area, minimally invasive restoration)

  2. Composite resin repair

Prescription Use _ √ _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Indication for use

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for GENOSS. The logo is in green and features the word "GENOSS" in a stylized font. The first four letters, "GENO", are in a darker green, while the last two letters, "SS", are in a lighter green. The logo has a modern and clean design.

510(k) Summary K200156

12/20/2019

1. Company

Submitter
NameGENOSS Co., Ltd.
Address1F, Gyeonggi R&DB Center / 226, 2F, GSBC, 105Gwanggyo-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea
Phone/Fax+82-70-7098-7541/ +82-31-888-5105
Contact personHongJun Jeon / RAhjjeon@genoss.com
Summary Date12/20/2019

2. Device Name

Proprietary name: Bright Low Flow Regulation description : Light-cured Flowable Composite Resin Classification name: Tooth shade resin material

3. Predicate Device

K091388 G-aenial Universal Flo

4. Description

Bright Flow is a light-cured flowable composite resin. It comprises two different types of flowability (Low Flow and High Flow) and 9 shades depending on the intended use, which enables aesthetic and durable outcomes for anterior and posterior composite restorations.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first four letters, "GENO", are in a dark green color, while the last two letters, "SS", are in a lighter, lime green color. The letters are connected, giving the word a flowing appearance.

5. Indication for use

    1. Class I, III, V (non-stress area, minimally invasive restoration)
    1. Composite resin repair

6. Technological Characteristics

Bright Low Flow has the similar technological characteristics as the predicate device; main material, indication for use and design. Technological characteristics of Bright Low Flow, Gaenial Universal Flo are as following.

Device nameBright Low FlowG-aenial Universal Flo
ManufacturerGenoss Co., Ltd.GC America, Inc.
510(k) NumberNew DeviceK091388
ClinicalTarget populationHuman toothHuman tooth
PurposeDirect restorationDirect restoration
Site of applicationDentin, EnamelDentin, Enamel
Clinical PerformanceHigh flexural strength and strong bondingHigh flexural strength and strong bonding
BiologicalMaterialsMethacrylate resins, Fillers (Barium glass, Silica)Methacrylate resins, Fillers (Barium glass, Silica)
Chemical SafetyBiocompatibleBiocompatible
SterilizationNon-sterileNon-sterile
Shelf-Life2 years3 years
TechnicalFormPastePaste
Indication for use1) Class I, II, III, V (non-stress area, minimally invasive restoration)2) Composite resin repair1) Direct restorative for Class I, IV, III, II and V cavities.2) Fissure sealant3) Sealing hypersensitive areas4) Repair of (in) direct aesthetic restorations, temporary crown &

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GENO" are in a dark green color, while the letters "SS" are in a lighter green color. The font is bold and sans-serif, and the letters are slightly slanted to the right.

bridge, defect margins when marginsare in enamel5) Blocking out undercuts6) Liner or base
UsePrescriptionPrescription
Depth of cure2.9 mm2.8 mm
Water Sorption23 $\mu g/mm^3$Less than 40 $\mu g/mm^3$
Solubility0.9 $\mu g/mm^3$Less than 7.5 $\mu g/mm^3$
Flexural Strength142 Mpa167 Mpa
Radio-opacity1.81More than Aluminum

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GEN" are in a dark green color, while the letters "OSS" are in a lighter, lime green color. The font is bold and slightly rounded, giving the word a modern and professional appearance.

7. Performance Data

Biocompatibility testing on the proposed Bright Low Flow has been completed. Requirements for biological evaluation of the proposed device were based on FDA recognized concensus standard of ISO10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological bone and tissues with its intended use. The following biocompatibility tests were completed:

(P: Pass, F: Fail)

No.TestMethodAcceptance criteriaP/FReport No.
1CytotoxicityISO 10993-5None cytotoxicityPCDM-18-0057-12
2IrritationISO 10993-10None oral irritationPMTK-2018-000704
3SensitizationISO 10993-10None sensitizationPMTK-2018-000703
4Acute systemic toxicityISO 10993-11None systemictoxicityPMTK-2018-000703
5GenotoxocityISO 10993-3None genotoxicityPMTK-2018-000702
6ImplantationISO 10993-6BiocompatiblePBER-18-027
7Chronic toxicityISO 10993-11No chronic toxicityPBER-18-027

{7}------------------------------------------------

GENO

The proposed Bright Low Flow was evaluated using the following performance bench testing to confirm the performance characteristics:

No.ItemsMethodAcceptanceCriteriaResultFile No.
1VisualISO 4049No impuritiesand No specificchangesNo impuritiesand No specificchangesCDM-18-0057-01
2CapacityISO 4049Capacity error of;Standard Capacity <±5%Size error of;Standard Size < ±5%H 0.43%, Ø 0.03%CDM-18-0057-02
3PackageISO 4049No damageNo damageCDM-18-0057-03
4Sensitivity toAmbient LightISO 4049Must bephysically uniformUniformityCDM-18-0057-04
5Depth of CureISO 4049More than 1.5 mmAverage: 2.9 mmCDM-18-0057-05
6ShadeISO 4049Must be shadeuniformUniformityCDM-18-0057-06
7Color StabilityISO 4049Color should bestableStableCDM-18-0057-07
8FlexuralStrengthISO 4049More than 80 MPaAverage: 142 MPaCDM-18-0057-08
9Water SorptionISO 4049Less than 40 µg/mm3Average: 23 µg/mm3CDM-18-0057-09
10SolubilityISO 4049Less than 7.5 µg/mm3Average: 0.9 µg/mm3CDM-18-0057-10
11Radio-opacityISO 4049More than Al≥ Al (1.81)CDM-18-0057-11

All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Bright Low Flow met the established specifications necessary for consistent performance according to its intended use.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The letters "GENO" are in a dark green color, while the letters "SS" are in a lighter green color. The font is bold and slightly italicized, giving the word a dynamic appearance.

8. Conclusion

Based on the information provided in this premarket notification of GENOSS Co., Ltd. Concludes that Bright Low Flow is acceptable and safe, substantially equivalent to predicate devices.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.