K Number
K200022

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-04-03

(88 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedical, abdominal, vascular, cardias, critical care and emergency room procedures.

Device Description

The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System consists of a high voltage (HV) inverter generator, a tube support unit, an X-ray beam limiting device, mobile cart, a detector, operating software, and a tube, and is primarily used in a hospital for diagnosis of diseases in skeletal. respiratory and urinary systems such as the skull, spinal column, chest, abdomen, extremities, and other body parts. This device is not intended to be used for mammography applications.

FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is a solution to produce radiological images of patient during medical operations. This inverter control X-ray unit visualizes the anatomical structure on screen, which is obtained by X-ray fluoroscopy and a flat panel detector. This system can be applied in emergency room, operation room, cast room or etc. of a hospital.

AI/ML Overview

The provided document is a 510(k) summary for the FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System. It focuses on demonstrating substantial equivalence to predicate devices based on technical specifications and adherence to recognized standards.

Therefore, the document does not contain information about acceptance criteria for device performance in a clinical study, or a study specifically designed to prove the device meets such criteria in terms of diagnostic accuracy or impact on human readers.

The information provided describes the device's technical specifications and compares them to predicate and reference devices, along with compliance with relevant industry standards and regulations for medical devices.

Here's a breakdown of what is available and what is not:

1. A table of acceptance criteria and the reported device performance

  • Not Available: The document does not provide a table of acceptance criteria for diagnostic performance outcomes (e.g., sensitivity, specificity, accuracy) or reported performance metrics from a clinical study. The tables present technical specifications of the device and its comparators (e.g., X-ray tube heat capacity, detector DQE, pixel matrix).

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Not Available: There is no mention of a test set, sample size, or data provenance related to a clinical performance study. The testing performed focuses on non-clinical aspects like electrical, mechanical, software, and risk management.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not Available: Since there is no mention of a clinical test set or ground truth establishment, this information is not provided.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not Available: There is no information on any adjudication method for a clinical test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Available: The document describes a fluoroscopic C-arm Mobile X-ray system, which is an imaging device, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers would not be applicable or expected for this device nor is it mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Available: As noted above, this is an imaging acquisition device, not an algorithm, so standalone performance is not relevant in the context of diagnostic algorithms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not Available: There is no mention of ground truth as it pertains to clinical diagnostic performance.

8. The sample size for the training set

  • Not Available: This device is a hardware system, not a machine learning model, so there is no training set in the context of AI/ML.

9. How the ground truth for the training set was established

  • Not Available: See point 8.

What the document does describe (Non-clinical testing for substantial equivalence):

The document details the non-clinical testing performed to establish substantial equivalence (refer to section VIII. Non clinical testing and IX. Conclusion):

  • Standards Adherence: The device successfully underwent testing according to various IEC and EN standards related to medical electrical equipment, including safety, electromagnetic compatibility, and specific requirements for X-ray equipment (e.g., EN 60601-1, EN 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-54).
  • Guidance Document Utilization: The development informed by FDA guidance documents regarding software in medical devices, 510(k) submissions for solid-state X-ray imaging devices, cybersecurity, and pediatric information for X-ray imaging devices.
  • Regulatory Compliance: Conformance to 21 CFR 1020 Subchapter J (Performance Standards for Ionizing Radiation Emitting Products), 21 CFR 1020.30 (Diagnostic x-ray system and their major components), and 21 CFR 1020.32 (Fluoroscopic Equipment).

The conclusion states that based on these tests and comparisons to predicate devices, the FLUSION-9001 system is considered substantially equivalent in its design, mechanical and electrical performance, making it suitable for its intended use.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

April 3, 2020

LiverMoreTech, Inc. % Mr. Dave Kim Regulatory Affairs MTech Group 7707 Fannin Street. Suite 200 HOUSTON TX 77054

Re: K200022

Trade/Device Name: FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA, IZI Dated: December 31, 2019 Received: January 6, 2020

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K200022

Device Name FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System

Indications for Use (Describe)

The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedical, abdominal, vascular, cardias, critical care and emergency room procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K200022

This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: April 1, 2020

I. SUBMITTER

Submitter's NameLivermoretech
Submitter's Address801 North Jupiter Rd, Suite 200Plano TX 75074
Submitter's TelephoneTel: +1-214-257-0113
Contact personJay Kim (jay.kim@aspenstate.com) / RA Manager
Official CorrespondentDave Kim (davekim@mtech-inc.net)
Address7707 Fannin St. Ste 200, V111, Houston, TX 77054
Telephone+713-467-2607

II. DEVICE

Trade/proprietary NameFLUSION-9001 Fluoroscopic C-arm Mobile X-ray System
Regulation NameImage Intensified Fluoroscopic Mobile X-ray System
Regulation Number21 CFR 892.1650
Product CodeOWB, OXO, JAA, IZI
Regulatory ClassClass II

III. PREDICATE DEVICE

510K NumberK032761
ManufacturerUnited Radiology Systems, Inc
Device NameKMC-950
Regulation NameImage Intensified Fluoroscopic Mobile X-ray System
Regulation Number21 CFR 892.1650
Product CodeOWB, OXO, JAA
Regulatory ClassClass II

{4}------------------------------------------------

IV. REFRENCE DEVICE

510K NumberK180473
ManufacturerECOTRON Co., Ltd
Device NameANYVIEW DR SERIES FPD Fluoroscopic Mobile C-Arm
Regulation NameImage Intensified Fluoroscopic Mobile X-ray System
Regulation Number21 CFR 892.1650
Product CodeOWB, OXO, JAA
Regulatory ClassClass II

DEVICE DESCRIPTION: V.

The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System consists of a high voltage (HV) inverter generator, a tube support unit, an X-ray beam limiting device, mobile cart, a detector, operating software, and a tube, and is primarily used in a hospital for diagnosis of diseases in skeletal. respiratory and urinary systems such as the skull, spinal column, chest, abdomen, extremities, and other body parts. This device is not intended to be used for mammography applications.

FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is a solution to produce radiological images of patient during medical operations. This inverter control X-ray unit visualizes the anatomical structure on screen, which is obtained by X-ray fluoroscopy and a flat panel detector. This system can be applied in emergency room, operation room, cast room or etc. of a hospital.

Indications for Use: 21 CFR 807 92 (a) (5) VI.

The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedic, neurological, abdominal, vascular, cardiac, critical care and emergency room procedures.

Comparison Table with the Predicate Device for technological characteristics:

Predicate deviceProposed deviceSE
Model NameKMC-950FLUSION-9001Fluoroscopic C-arm Mobile X-ray System
510(k) NumberK032761K200022
Indications for UseThe KMC-950 is intended toprovide fluoroscopic andradiographic imaging of thepatient during diagnostic,surgical and interventionalprocedures. Clinicalapplications may include butare not limited to digitalsubtraction angiography,The FLUSION-9001fluoroscopic C-arm Mobile X-ray System is intended toprovide fluoroscopic andradiographic imaging of thepatient during diagnostic,surgical and interventionalprocedures. Clinicalapplications may include butSimilar
orthopedic, neurological, abdominal, vascular, cardiac, critical care and emergency room procedures.The system may be used for other RF imaging application at physician's discretion.are not limited to digital subtraction angiography, orthopedic, neurological, abdominal, vascular, cardiac, critical care and emergency room procedures.
X-ray TubeAnode TypeRotatingRotatingSame
Heat Capacity300,000 HU300,000 HUSame
Anode HeatCooling70kHU/min60kHU/minSimilar
Focal size0.3mm / 0.6mm0.3mm / 0.6mmSame
FluoroscopicModekV range40 to 125 kV See below discussion for differences40 to 120 kV See below discussion for differencesSimilar
mA range0.5 to 5 mA. See below discussion for differences0.2 to 8mA. See below discussion for differencesSimilar
Pulse FluoroYesYesSame
ABS functionYesYesSame
DetectorManufacturerImaging Intensifier:Thales (TH9428HP2) Image IntensifierImaging Intensifier:E5830SD-P4A (TOSHIBA)Cleared under K160065
Size9"9"Same
Magnification9" / 6" / 4.5"9" / 6" / 4.5"Same
DQE6565 (@ 0 lp/mm)Same
CameraType1/2" CCD1/2" CCDSame
Active pixels512 x 5121024 x 1024Better
C-armManufacturerGEMSS medicalLivermoretech Inc.
SID950mm1000mmSimilar
Range of C-ramRail Rotation115°135°Wider
Range of theLiner FR-armMovement200 mm200 mmSame
Range of theLinear T-armMovement500 mm500 mmSame
Range ofSwing-armRotation± 12.5°± 15°Wider
Range of Stay-arm Rotation360°190°
CollimatorCollimatorMotor control / rotationMotor control / rotationSame

{5}------------------------------------------------

LIVERMORETECH

801 North JupiterRd, Suite 200, Plano, TX 70574 USA

Tel.: +214-257-0113

{6}------------------------------------------------

Subject DeviceSubject DeviceSubject DeviceReference Device
ModelVIVIX-D1212GPixium® Surgical3030S-A3030DXVVIVIX-D1212G
510(k) NoK180473K182086K132904K180473
Detector typeTFTTFTTFTTFT
Active detector size12 x 12 in12 x 12 in12 x 12 in12 x 12 in
Total pixel matrix2048 x 20481956 x 19561536 x 15362048 x 2048
Pixel pitch145 μm154.0 μm194.0 μm145 μm
Frame rate30 fps @ 2x230 fps30 fps30 fps @ 2x2
MTF1.0 lp/mm60%55%55%60%
DQE0 lp/mm69%75%80%69%

FPD Image Detector Comparison

VII. Discussion of differences

The indications for use, operating principle, technical specifications such as X-ray tube head, image intensifier and FPD of FLUSION-9001 fluoroscopic C-arm Mobile X-ray System are similar to those of the predicate device (K032761) and reference device (K180473). FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System is designed as a set of components (X-ray tube and housing, detector, digital imaging system, collimator, generator etc.) similar to the predicate device KMC-950 (K032761). All digital flat panel detectors available for the subject device have been cleared by FDA as a part of a complete fluoroscopy system (K180473, K182086, K132904). Based on the recognized standard conformity evidences related to electro-, mechanical-, software-, and risk management, Livermoretech certifies that technological characteristics of FLUSION-9001 fluoroscopic C-arm Mobile X-ray System are substantially equivalent to KMC-950, the predicate device. FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System consists of many critical parts same as KMC-950, the predicate device (Collimator, X-ray imager, digital image processing and viewing program, C arm, etc.).

FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System utilizes PVCARM Viewer, a radiological digital image communications system. It provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components provide functions related to simple manipulation, enhancement or quantification of images.

  • PVCARM receives images from a detector, processes and transfers the images and manages patient's information and the images for doctor. PVCARM enables images such as x-ray images to be stored electronically and viewed on screens.

  • PVCARM offers full compliance with DICOM (Digital Imaging and Communications in Medicine) standards to allow the sharing of medical information with other PACS (Picture Archiving and Communication System Server).

{7}------------------------------------------------

Tel.: +214-257-0113

VIII. Non clinical testing

Testing was performed successfully according to the following standards:

  • EN 60601-1:2006/A1:2013 ●
  • EN 60601-1-2:2015 (IEC 60601-1-2: 2014) ●
  • IEC 60601-1-3:2008+A1:2013 ●
  • IEC 60601-2-43:2010 ●
  • IEC 60601-2-54:2009 + A1::2015 ●

Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients:

  • "The Content of Premarket Submissions for Software Contained in Medical Devices" ●
  • "Guidance for the Submission of 510(k) for Solid State X-ray Imaging devices" ●
  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices ● Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014
  • Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff Document issued on November 28, 2017.

All applicable aspects of these guidance documents listed in this 510(k) summary have been addressed.

The device also conforms to the following:

  • 21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products ●
  • 21 CFR 1020.30: Diagnostic x-ray system and their major components ●
  • 21 CFR 1020.32: Fluoroscopic Equipment ●

IX. Conclusion:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided above comparison table, the FLUSION-9001 Fluoroscopic C-arm Mobile Xray System has little difference with its size and user interface as the information in the table. But the system is substantially equivalent to the predicate devices with its design, mechanical and electrical performance as described.

Performance evaluation (test) reports and device inspection report confirmed that the FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System is suitable for its intended use and user instruction of the device.

N/A