K Number
K200007

Validate with FDA (Live)

Date Cleared
2020-04-25

(114 days)

Product Code
Regulation Number
888.3353
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

  1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
  2. inflammatory degenerative joint disease such as rheumatoid arthritis;
  3. correction of functional deformity; and,
  4. revision procedures where other treatments or devices have failed
    Hydroxyapatite coatings applied to implant surfaces are intended for uncemented arthroplasty.
Device Description

The PROFEMUR® GLADIATOR® Thin Hydroxyapatite-coated (HA) Classic Stems with Collar present a monolithic, tapered-wedge design and a reduced distal stem tip. The 'Thin' description is a relative word MicroPort uses as an internal descriptor to clearly identify the difference in product between the currently cleared US-offered HA coating thickness (65um) and the coating thickness offered in outside US (OUS) markets. The subject stems are coated with a 65um HA coating (ASTM 1185), and the uncoated medial collar, proximal shoulder and neck regions of the stem possess a glass-beaded finish (Ra=0.8μm). The subject stems possess a proximal medial collar designed to assist in discouraging subsidence and maximizing rotational stability. The proximal anterior, posterior, and medial surfaces possess grooves perpendicular to the stem's longitudinal axis and parallel with each other. The proximal grooves are designed to provide a maximized bone-stem contact surface for bone apposition and assist in discouraging stem subsidence. The distal anterior and posterior surfaces of the subject stem possess vertical grooves parallel to the stem's longitudinal axis. The distal grooves are designed to assist the stem in resisting against rotational loads. The subject classic stems possess an oval impaction feature on the proximal shoulder for stem insertion into the femoral canal. The PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are designed for use in total hip arthroplasty.
The PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are manufactured from forged Ti6Al4V alloy (ASTM F620) and come in 10 sizes, with the stem length ranging from 125mm to 175mm. The subject stems features two different neck options: standard (with a CCD neck angle of 135°) and extended/Varus 8° (with a CCD angle of 127°). The PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar were designed to accommodate a wide range of patient anatomy and to provide an optimized fit to adequately restore the biomechanics of the respective hip (including restoration of the natural leg length and head center offset).

AI/ML Overview

The provided document describes a 510(k) premarket notification for a medical device (PROFEMUR GLADIATOR Thin HA Classic Hip Stem with Collar) and focuses on demonstrating its substantial equivalence to previously cleared predicate devices. It does not contain information about a study that proves the device meets acceptance criteria in the context of AI/ML device performance (e.g., diagnostic accuracy, sensitivity, specificity, or human-in-the-loop performance).

Instead, the document details nonclinical testing performed on the hip stem to demonstrate its mechanical properties, safety, and equivalence to existing devices. Therefore, I cannot provide all the requested information, particularly those related to AI/ML device performance studies like sample size for test sets, data provenance, expert ground truth, MRMC studies, or standalone algorithm performance.

Here's an analysis of the information available in the document regarding "acceptance criteria" and "study" in the context of this orthopedic implant:

1. A table of acceptance criteria and the reported device performance:

The document describes various nonclinical tests and their successful completion, implying that passing these tests served as the acceptance criteria for substantial equivalence. However, it does not present a formal table with specific quantitative acceptance criteria alongside exact reported performance values from the tests. Instead, it generally states that the device "met all mechanical testing requirements" and "were successfully performed."

Acceptance Criteria (Implied)Reported Device Performance
Endurance properties (stem endurance) per ISO 7206-4Testing successfully completed, met requirements.
Endurance properties (neck region) per ISO 7206-6Testing successfully completed, met requirements.
Specific requirements for hip-joint replacement implants per EN ISO 21535Testing successfully completed, met requirements.
Magnetic Resonance (MR) Induced Displacement Force per ASTM F2052-15Magnetic Resonance Imaging Safety Analysis successfully completed.
MR Image Artifacts per ASTM F2119-07Magnetic Resonance Imaging Safety Analysis successfully completed.
MR Radio Frequency Induced Heating per ASTM F2182-11aMagnetic Resonance Imaging Safety Analysis successfully completed.
Marking Medical Devices for MR Environment per ASTM F2503-13Magnetic Resonance Imaging Safety Analysis successfully completed.
Distal Fatigue TestingSuccessfully completed, met all mechanical requirements.
Proximal Fatigue TestingSuccessfully completed, met all mechanical requirements.
Range of Motion AnalysisSuccessfully completed, met all mechanical requirements.
Coating Characterization of Hydroxyapatite CoatingSuccessfully completed.
Bacterial Endotoxin Limit (20 EU/device) per ANSI/AAMI ST72 and USP <161>Testing successfully performed, worst-case implants meet the 20 EU/device limit.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the exact number of hip stems or components used for each nonclinical test (e.g., how many stems were subjected to fatigue testing). It refers to "worst-case construct testing" and "worst case implants" for endotoxin testing, suggesting a representative, but unspecified, sample size.
  • Data Provenance: Not applicable in this context. The "tests" are laboratory-based mechanical and biological evaluations, not clinical studies with human patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the evaluation involved nonclinical laboratory testing against pre-defined engineering standards, not expert interpretation of clinical data to establish ground truth.

4. Adjudication method for the test set:

This information is not applicable for nonclinical laboratory testing. The evaluation is based on meeting the quantitative or qualitative criteria of the respective ISO, ASTM, and EN standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The device is a hip implant, not an AI/ML-based diagnostic or assistive technology for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The device is a hip implant, not an AI/ML algorithm.

7. The type of ground truth used:

The "ground truth" in this context refers to the established international and national standards (e.g., ISO 7206 series, EN ISO 21535, ASTM F2052, ASTM F2119, ASTM F2182, ASTM F2503, ANSI/AAMI ST72, USP <161>) against which the hip stem's mechanical, safety, and material properties were evaluated. Compliance with these standards indicates acceptable performance.

8. The sample size for the training set:

This information is not applicable as the device is a physical hip implant. There is no "training set" in the context of an AI/ML algorithm.

9. How the ground truth for the training set was established:

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

April 25, 2020

MicroPort Orthopedics Inc. Allen Mamaril Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002

Re: K200007

Trade/Device Name: PROFEMUR GLADIATOR Thin HA Classic Hip Stem with Collar Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, LPH Dated: December 30, 2019 Received: January 2, 2020

Dear Allen Mamaril:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vesa Vuniqi, MS Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K200007

Device Name

PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar

Indications for Use (Describe)

PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

  1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;

  2. inflammatory degenerative joint disease such as rheumatoid arthritis;

  3. correction of functional deformity; and,

  4. revision procedures where other treatments or devices have failed

Hydroxyapatite coatings applied to implant surfaces are intended for uncemented arthroplasty.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners on the left, inside of which is a white circle containing a red star-like shape. To the right of the square is the text "MicroPort" in a larger, bold blue font, with "Orthopedics" underneath in a smaller blue font. Below that is the text "Full Function, Faster" in a smaller, italicized blue font.

K200007 Page 1/4

510(k) Summary

In accordance with 21 CFR 807.92, this information serves as a summary of information upon which MicroPort Orthopedics claims substantial equivalence for the subject PROFEMUR® GLADIATOR® Thin HA Classic Stem with Collar.

Submitted by:MicroPort Orthopedics Inc.5677 Airline Road,Arlington, TN 38002USAPhone: (901) 290-5175
Date:December 20, 2019
Contact Person:Allen MamarilRegulatory Affairs Specialist IIPhone: (901) 867-4551
Proprietary Name:PROFEMUR® GLADIATOR® Thin HA ClassicStem with Collar
Common Name:Femoral Hip Stem
Classification Name andReference:21 CFR 888.3353 – Hip joint metal/ceramic/ polymersemi-constrained cemented or nonporous uncementedprosthesis21 CFR 888.3358 – Hip joint metal/polymer/metalsemi-constrained porous-coated uncementedprosthesis
Device Class:Class II
Panel Code:Orthopedic/87
Product Code:LZO, LPH
Predicate Device:K112150 PROFEMUR® GLADIATOR® HA HipStem
Reference Device:K110399 PROFEMUR® GLADIATOR® PlasmaClassic Hip StemK191632 PROFEMUR® TL2 StemsK190123 MicroPort CoCr Femoral Heads

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a red star-like shape. To the right of the square is the text "MicroPort" in a larger, bold, blue font, with the word "Orthopedics" underneath it in a smaller, blue font. Below "Orthopedics" is the tagline "Full Function, Faster" in a smaller, blue, italicized font.

K200007

PROFEMUR® GLADIATOR® Thin HA Classic Stem w/ Collar Page 2/4 Traditional 510(k) Tab 007: 510(k) Summary

A. Device Description

The PROFEMUR® GLADIATOR® Thin Hydroxyapatite-coated (HA) Classic Stems with Collar present a monolithic, tapered-wedge design and a reduced distal stem tip. The 'Thin' description is a relative word MicroPort uses as an internal descriptor to clearly identify the difference in product between the currently cleared US-offered HA coating thickness (65um) and the coating thickness offered in outside US (OUS) markets. The subject stems are coated with a 65um HA coating (ASTM 1185), and the uncoated medial collar, proximal shoulder and neck regions of the stem possess a glass-beaded finish (Ra=0.8μm). The subject stems possess a proximal medial collar designed to assist in discouraging subsidence and maximizing rotational stability. The proximal anterior, posterior, and medial surfaces possess grooves perpendicular to the stem's longitudinal axis and parallel with each other. The proximal grooves are designed to provide a maximized bone-stem contact surface for bone apposition and assist in discouraging stem subsidence. The distal anterior and posterior surfaces of the subject stem possess vertical grooves parallel to the stem's longitudinal axis. The distal grooves are designed to assist the stem in resisting against rotational loads. The subject classic stems possess an oval impaction feature on the proximal shoulder for stem insertion into the femoral canal. The PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are designed for use in total hip arthroplasty.

The PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are manufactured from forged Ti6Al4V alloy (ASTM F620) and come in 10 sizes, with the stem length ranging from 125mm to 175mm. The subject stems features two different neck options: standard (with a CCD neck angle of 135°) and extended/Varus 8° (with a CCD angle of 127°). The PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar were designed to accommodate a wide range of patient anatomy and to provide an optimized fit to adequately restore the biomechanics of the respective hip (including restoration of the natural leg length and head center offset).

B. Intended Use

PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

    1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
    1. inflammatory degenerative joint disease such as rheumatoid arthritis;
  • correction of functional deformity; and, 3.

{5}------------------------------------------------

K200007 Page 3/4

Image /page/5/Picture/1 description: The image is a logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners on the left, inside of which is a white circle containing a red four-pointed star. To the right of the square is the text "MicroPort" in a bold, blue font, with "Orthopedics" in a smaller, lighter blue font underneath. Below that is the text "Full Function, Faster" in a smaller, lighter blue font.

    1. revision procedures where other treatments or devices have failed
      Hydroxyapatite coatings applied to implant surfaces are intended for uncemented arthroplasty.

C. Substantial Equivalence Information

The design features and materials of the PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are substantially equivalent to those of the predicate cleared under K112150. The indications of the subject devices are identical to the predicate. The fundamental scientific technology of the subject device has not changed relative to the predicate device. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis of data provided within this Premarket Notification.

D. Nonclinical Testing

The subject PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar were evaluated to demonstrate substantial equivalence to the identified predicate PROFEMUR® GLADIATOR® HA Hip Stem and reference PROFEMUR® GLADIATOR® Plasma Classic Hip Stem. Design verification testing for the subject devices was completed in accordance with:

  • ISO 7206-4 (2010): Implants for surgery - Partial and total hip joint prostheses - Part 4: Determination of endurance properties and performance of stemmed femoral components;
  • ISO 7206-6 (2013) Implants for surgery Partial and total hip joint prostheses Part ● 6: Endurance properties testing and performance requirements of neck region of stemmed femoral components;
  • EN ISO 21535 (2009) Non-active surgical implants Joint replacement implants -● Specific requirements for hip-joint replacement implants.
  • ASTM F2052-15 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in Magnetic Resonance Environment
  • ASTM F2119-07 Standard Test Method for Evaluation of MR Image Artifacts from ● Passive Implants
  • ASTM F2182-11a Standard Test Method for Measurement of Radio Frequency ● Induced Heating on or Near Passive Implants During Magnetic Resonance Imaging
  • ASTM F2503-13 Standard Practice for marking Medical Devices and Other Items for ● Safety in the Magnetic Resonance Environment

{6}------------------------------------------------

K200007

Image /page/6/Picture/1 description: The image shows the logo for MicroPort Orthopedics. The logo consists of a blue square with a white circle inside, and a red star in the center of the circle. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in smaller blue text below it. Below that is the text "Full Function, Faster" in a smaller blue font.

The testing and analyses performed for the subject PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar were:

Page 4/4

  • Distal Fatigue Testing
  • Proximal Fatigue Testing ●
  • Range of Motion Analysis
  • Magnetic Resonance Imaging Safety Analysis ●
  • Coating Characterization of the Hydroxyapatite Coating

The mechanical testing and coating characterization verifies the subject PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar are substantially equivalent to the cleared predicate PROFEMUR® GLADIATOR® HA Hip Stem and cleared reference PROFEMUR® GLADIATOR® Plasma Classic Hip Stem currently on the market and has met all mechanical testing requirements based on the worst-case construct testing.

E. Non-Pyrogenicity Endotoxin Testing

The bacterial endotoxin test, also known as Limulus amebocyte lysate (LAL) test, was performed utilizing worst case implants to verify that the subject implants meet the 20 endotoxin units (EU)/device pyrogen limit specification. The subject PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar were adopted into the current worst case Hip Stem and Neck Endotoxin Family based on the process operations. Testing was successfully performed and it was confirmed that the worst case implants meet the 20 EU/device testing limit for general medical devices that are implanted as outlined in ANSI/AAMI ST72, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing and USP <161>, Medical Devices - Bacterial Endotoxin and Pyrogen Tests.

F. Conclusion

Based on the design features, the use of established well-known materials, feature comparisons, indications for use, and results of the mechanical testing, the subject PROFEMUR® GLADIATOR® Thin HA Classic Stems w/ Collar have shown to be substantially equivalent to the legally marketed devices cited in this summary.

N/A