K Number
K193649

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-05-10

(497 days)

Product Code
Regulation Number
862.1225
Age Range
2 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Yumizen C1200 Creatinine PAP reagent is intended for the quantitative in vitro diagnostic determination of Creatinine in human serum, plasma and urine based on an enzymatic method using a multi- step approach ending with a photometric end-point reaction. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Device Description

Yumizen C1200 Creatinine PAP reagent is intended for the quantitative in vitro diagnostic determination of Creatinine in human serum, plasma and urine based on an enzymatic method using a multi- step approach ending with a photometric end-point reaction.

AI/ML Overview

The Horiba ABX SAS Yumizen C1200 Creatinine PAP device is an in vitro diagnostic intended for the quantitative determination of Creatinine in human serum, plasma, and urine. Its performance was evaluated through various analytical studies to demonstrate substantial equivalence to its predicate device, the ABX Pentra Enzymatic Creatinine CP.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implicit from "within specifications")Reported Device Performance (Yumizen C1200 Creatinine PAP)
Measuring Range
Serum Limit of QuantitationNot explicitly stated, but implies the lowest concentration measurable with acceptable precision.0.11 mg/dL
Serum LinearityNot explicitly stated, but implies correlation across range.0.04 - 19.93 mg/dL
Serum Measuring RangeNot explicitly stated, but likely the linear range with acceptable bias.0.11 - 16.95 mg/dL (up to 50.85 mg/dL with post-dilution)
Urine Limit of QuantitationNot explicitly stated, but implies the lowest concentration measurable with acceptable precision.1.13 mg/dL
Urine LinearityNot explicitly stated, but implies correlation across range.0.00 - 327.60 mg/dL
Urine Measuring RangeNot explicitly stated, but likely the linear range with acceptable bias.3.56 - 175 mg/dL (up to 525 mg/dL with post-dilution)
Accuracy and Precision (Instrument Variability - Serum/Plasma)
Within-run CV (low level)≤ 4.5 %0.5% (Yumizen N Multi Control), 1.4% (Sample 1)
Within-run CV (middle level)≤ 3.4 %0.3% (Yumizen P Multi Control), 0.5% (Sample 2)
Within-run CV (high level)≤ 1.8 %0.3% (Sample 3)
Total CV (low level)≤ 6.0 %1.5% (Yumizen N Multi Control), 2.9% (Sample 1)
Total CV (middle level)≤ 4.5 %1.3% (Yumizen P Multi Control), 2.0% (Sample 2)
Total CV (high level)≤ 2.4 %1.3% (Sample 3)
Accuracy and Precision (Lot to Lot Variability - Serum/Plasma)
Within-run CV (low level)≤ 4.5 %Not reported separately, but "Within-Day (%CV)" for Sample 1 (1.7%) is shown
Within-run CV (middle level)≤ 3.4 %Not reported separately, but "Within-Day (%CV)" for Sample 2 (0.9%) is shown
Within-run CV (high level)≤ 1.8 %Not reported separately, but "Within-Day (%CV)" for Sample 4 (0.4%) is shown
Total CV (low level)≤ 6.0 %4.6% (Sample 1)
Total CV (middle level)≤ 4.5 %2.0% (Sample 2)
Total CV (high level)≤ 2.4 %0.5% (Sample 4)
Accuracy and Precision (Instrument Variability - Urine)
Within-run CV (low level)≤ 4.5 %1.2% (Sample 1)
Within-run CV (middle level)≤ 3.8 %0.8% (Sample 2)
Within-run CV (high level)≤ 3.8 %0.8% (Sample 3)
Total CV (low level)≤ 6.0 %4.2% (Sample 1)
Total CV (middle level)≤ 5.0 %4.3% (Sample 2)
Total CV (high level)≤ 5.0 %3.9% (Sample 3)
Accuracy and Precision (Lot to Lot Variability - Urine)
Within-run CV (low level)≤ 4.5 %Not reported separately, but "Within-Day (%CV)" for Sample 1 (1.1%) is shown
Within-run CV (middle level)≤ 3.8 %Not reported separately, but "Within-Day (%CV)" for Sample 2 (0.9%) is shown
Within-run CV (high level)≤ 3.8 %Not reported separately, but "Within-Day (%CV)" for Sample 3 (0.9%) is shown
Total CV (low level)≤ 6.0 %2.1% (Sample 1)
Total CV (middle level)≤ 5.0 %1.3% (Sample 2)
Total CV (high level)≤ 5.0 %0.9% (Sample 3)
Interferences (Serum/Plasma)Acceptable bias +/-10% of the value without interfering substances.All listed interferents (Hemoglobin, Triglycerides, Total Bilirubin, Direct Bilirubin, Ascorbic Acid, Acetylsalicylic Acid, Ibuprofen, Acetaminophen, N-Acetylcystein, Glucose, Total Protein, Methyldopa, L-Dopa, Calcium Dobesilate) showed no interference higher than +/-10% at the specified concentrations.
Interferences (Urine)Acceptable bias +/-10% of the value without interfering substances.All listed interferents (Hemoglobin, Triglycerides, Direct Bilirubin, Ascorbic acid, N-Acetylcystein, pH) showed no interference higher than +/-10% at the specified concentrations.
Matrix Comparison (Serum/Lithium Heparin Plasma)No significant difference between serum and plasma with heparin specimens.Regression line: Intercept = -0.0281, Slope = 1.0008, r² = 0.995. This indicates no significant difference.
Method Comparison (Serum/Plasma vs. Predicate)High correlation and acceptable agreement (implied by CLSI EP-9A3).Regression line (Passing Bablok): Intercept = -0.0107, Slope = 0.9611, r² = 0.997.
Method Comparison (Urine vs. Predicate)High correlation and acceptable agreement (implied by CLSI EP-9A3).Regression line (Passing Bablok): Intercept = 0.2296, Slope = 0.9772, r² = 0.994.
Reagent Stability (Closed)Stable up to expiry date.12 months (at 2-8°C).
Reagent Stability (Open, On-Board)Stable for a specified period.6 weeks.
Reference Range Verification (Serum/Plasma - Men)Consistent with established literature reference ranges.Normal range: 0.67 - 1.17 mg/dL (consistent with Mazzachi et al. reference).
Reference Range Verification (Serum/Plasma - Women)Consistent with established literature reference ranges.Normal range: 0.51 - 0.95 mg/dL (consistent with Mazzachi et al. reference).

2. Sample sizes used for the test set and the data provenance

  • Measuring Range:
    • Limit of detection and quantitation: Determined according to CLSI guideline EP17-A2. (Specific sample size not provided in the summary but typically involves multiple replicates of low-concentration samples).
    • Linearity: Determined according to CLSI guideline EP06-A. (Specific sample size not provided in the summary but involves multiple concentrations with replicates).
  • Accuracy and Precision:
    • Instrument Variability (Serum/Plasma & Urine): 240 measurements (20x2x2 means 20 replicates for each of 3 samples, on 2 runs for 2 instruments over a certain period - or 20 days x 2 runs/day x 2 instruments). The samples were control materials and native samples.
    • Lot-to-Lot Variability (Serum/Plasma & Urine): 90 measurements (3x5x2x3 implies 3 lots, 5 days, 2 runs/day, 3 samples). The samples were control materials and native samples.
  • Interferences: Not explicitly stated, but typically involves testing known concentrations of interferents in base samples.
  • Matrix Comparison: 84 paired samples (serum and lithium heparin plasma) from single donors.
  • Method Comparison:
    • Serum/Plasma: 103 native human serum samples. Data provenance: collected from CHU Nîmes (University Hospital Center). Retrospective (remnants).
    • Urine: 129 native human urine samples. Data provenance: collected from routine clinical laboratory. Retrospective (remnants).
  • Reagent Stability: Determined according to CLSI guideline EP25-A. (Specific sample size not provided).
  • Reference Range Verification:
    • Serum/Plasma - Men: 45 "normal samples" from a blood bank.
    • Serum/Plasma - Women: 41 "normal samples" from a blood bank.
    • Children & Urine: Verification could not be made due to lack of availability of samples from healthy pediatric patients/healthy people.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not mention the use of experts to establish ground truth for this device, which is a quantitative in vitro diagnostic for creatinine levels. For such devices, ground truth is typically established by:

  • A reference method (e.g., mass spectrometry) for accuracy studies.
  • The established values of control materials.
  • The results from a legally marketed predicate device (as seen in method comparison).
  • Literature values for reference range verification.

4. Adjudication method for the test set

Not applicable. This is a quantitative diagnostic device, not one requiring expert adjudication of results.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a quantitative diagnostic device, not an AI-assisted diagnostic imaging device requiring human reader interpretation. No MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is a standalone in vitro diagnostic device (reagent and instrument system). Its performance described above (e.g., precision, accuracy, linearity, interference) represents the algorithm-only (device-only) performance, without human interpretation of the result influencing the quantitative output.

7. The type of ground truth used

  • Measuring Range, Accuracy, Precision, Interferences, Reagent Stability: Internal specifications, established reference materials (controls), and recognized scientific methods (e.g., spiked samples for linearity and interference).
  • Matrix Comparison: Paired samples from the same donor, with comparison between results from serum and plasma. The expectation is that the creatinine value should be the same across matrices for the same individual.
  • Method Comparison: The predicate device's results (ABX Pentra Enzymatic Creatinine CP) served as the reference for comparison using method comparison studies (Passing Bablok regression).
  • Reference Range Verification: Reference ranges cited in scientific literature (e.g., Mazzachi BC et al., Schlebusch Soldin SJ et al., Roberts WL et al.) were used for verification against measured values in "normal" samples.

8. The sample size for the training set

The document describes performance evaluation studies (validation) rather than a clear "training set" for an algorithm. For a device like this, the "training" usually refers to the development and optimization of the reagent formulation and instrument parameters. The specific sample sizes used for this developmental phase are not detailed in the summary. The provided sample sizes are for the analytical performance studies which are typically considered validation.

9. How the ground truth for the training set was established

As there isn't a "training set" in the context of machine learning, this question isn't directly applicable. For the development of an IVD like this, ground truth would be established during the R&D phase through:

  • Using purified creatinine standards.
  • Comparison with established and highly accurate reference methods (e.g., isotope dilution mass spectrometry, IDMS).
  • Clinical samples with results from well-characterized, clinically accepted methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 10, 2021

Horiba ABX SAS Caroline Ferrer Regulatory Affairs Manager Parc Euromedecine Montpellier Cedex 4, 341184 France

Re: K193649

Trade/Device Name: Yumizen C1200 Creatinine PAP Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine test system Regulatory Class: Class II Product Code: JFY Dated: October 7, 2020 Received: October 9, 2020

Dear Caroline Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

Page

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm, Ph.D. Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193649

Device Name Yumizen C1200 Creatinine PAP

Indications for Use (Describe)

Yumizen C1200 Creatinine PAP reagent is intended for the quantitative in vitro diagnostic determination of Creatinine in human serum, plasma and urine based on an enzymatic method using a multi- step approach ending with a photometric end-point reaction. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K193649

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

1- Date of Summary

Date submitted : October 07th, 2020

2- Company

HORIBA ABX SAS HORIBA MEDICAL Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 France

3- Contact person

Contact Person: Caroline Ferrer (caroline.ferrer@horiba.com) Telephone: + (33) 4 67 14 1843 Fax: + (33) 4 67 14 1517

4- Device Name and Classification

. Devices Names and Intended Uses

Device's namesIntended Use
Yumizen C1200 Creatinine PAP(1300023844/1300023843)Yumizen C1200 Creatinine PAP reagent is intended for thequantitative in vitro diagnostic determination of Creatinine in humanserum, plasma and urine based on an enzymatic method using amulti- step approach ending with a photometric end-point reaction.Creatinine measurements are used in the diagnosis and treatment ofrenal diseases, in monitoring renal dialysis, and as a calculation basisfor measuring other urine analytes.

Devices Classification .

Trade/Proprietary Name:Yumizen C1200 Creatinine PAP
Device Class:Class II / 510(k) required
Classification Name:§862.1225: Creatinine test system
Product Code:JFY
Panel:Clinical Chemistry (75)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

5- Substantial Equivalence Information

The following tables show the similarities and differences and demonstrates substantial equivalence between the candidate device and its predicate device identified below.

a. Predicate Device Name and 510(k) number

Candidate devicePredicate devicePredicateManufacturerPredicate510(k)number
Yumizen C1200 CreatininePAP (Serum/Plasma and Urineapplications)ABX Pentra Enzymatic Creatinine CP(Serum/Plasma and Urine applications)On ABX Pentra 400 / Pentra C400HORIBA ABXSASK110137

The following tables show the similarities and differences and demonstrates substantial equivalence between the candidate device and its predicate device identified below.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is written in large, bold, blue letters. Below the word "HORIBA" is the word "Medical" written in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

b. Yumizen C1200 Creatinine PAP (Serum/Plasma and Urine applications)

i. Comparison with predicate Device : Similarities

ItemPredicate K110137Candidate
Device NameABX Pentra Enzymatic Creatinine CP(A11A01907)Yumizen C1200 Creatinine PAP(1300023843)
Intended UseABX Pentra Enzymatic Creatinine CP reagent is intended for the quantitativein vitro diagnostic determination ofCreatinine in human serum, plasma andurine based on an enzymatic method usinga multi-step approach endingwith a photometric end-point reaction.Creatinine measurements are used in thediagnosis and treatment of renal diseases,in monitoring renal dialysis, and as acalculation basis for measuring other urineanalytesYumizen C1200 Creatinine PAP reagent is intended for the quantitativein vitro diagnostic determination ofCreatinine in human serum, plasma andurine based on an enzymatic methodusing a multi-step approach endingwith a photometric end-point reaction.Creatinine measurements are used in thediagnosis and treatment of renal diseases,in monitoring renal dialysis, and as acalculation basis for measuring otherurine analytes
Reagent formatLiquidSame
MethodEnzymatic colorimetric testSame
MeasurementQuantitativeSame
Manufactured byHORIBA ABX SASSame
Product codeJFYSame
Sample typeSerum, plasmaand UrineSerum,Lithium heparin plasmaUrine
Sample StabilitySerum/Plasma :At 20-25°C :7 daysAt 4-8°C : 7 daysAt -20°C : 3 monthsUrine :At 20-25°C :2 daysAt 4-8°C : 6 daysAt -20°C : 6 monthsSerum/Plasma :At 20-25°C :7 daysAt 4-8°C : 7 daysAt -20°C : 3 monthsUrine :At 20-25°C :2 daysAt 4-8°C : 6 daysAt -20°C : 6 months
Reagent Shelf-lifeStable up to expiry date on the label ifstored at 2-8°C.Stable up to expiry date on the label ifstored at 2-8°C.Store protected fromlight.
ItemPredicate K110137Candidate
Device NameABX Pentra Enzymatic Creatinine CP(A11A01907)Yumizen C1200 Creatinine PAP(1300023843)
InstrumentABX Pentra 400Yumizen C1200 Clinical chemistryAnalyzer
CalibratorsABX Pentra Multical (A11A01652)Yumizen C1200 Multi Cal(1300023891/1300023890)
ControlsABX Pentra N Control (A11A01653A)ABX Pentra P Control (A11A01654)ABX Pentra Urine Control L/H(A11A01674)Yumizen C1200 N Multi Control(1300023938/1300023939)Yumizen C1200 P Multi Control(1300023940/1300023940)Yumizen C1200 Urine Level 1 Control(1300023946)Yumizen C1200 Urine Level 2 Control(130002347)
Packaging22 mL (R1)8 mL ( R2)6x35 mL (R1)6x19 mL (R2)
Analytical RangeMeasuring RangeSerum/Plasma :0.11 - 16.95 mg/dLUrine :3.56 - 282.5 mg/dLMeasuring RangeSerum/Plasma :0.11 - 16.95 mg/dLUrine :3.56 - 175 mg/dL
ReagentOn board Stability30 days6 weeks
Reference rangeSerum/Plasma :Men : 0.62 - 1.10 mg/dLWomen : 0.45 - 0.75 mg/dLUrine :Men :14 - 26 mg/kg/dayWomen :11 - 20 mg/kg/daySerum/Plasma :Men : 0.67 - 1.17 mg/dLWomen : 0.51 - 0.95 mg/dLUrine :Men : 14 - 26 mg/kg/dayWomen :11 - 20 mg/kg/day
Discussion on the analysis differences :1.Instrument: Yumizen C1200 Creatinine PAP is used on Yumizen C1200 analyzer.2. Calibrators&Controls: Yumizen C1200 Creatinine PAP uses Yumizen C1200 Multi Cal,
ItemPredicate K110137Candidate
Yumizen C1200 N Multi Control, Yumizen C1200 P Multi Control, Yumizen C1200 Urine Level 1 Control, and Yumizen C1200 Urine Level 2 Control whereas the predicate uses its own brand calibrator and controls.
3. Packaging is different: depends on cassette capacity.
4. Analytical range: The measuring ranges in urine for Yumizen C1200 Creatinine PAP is reduced when compared to the predicate's. But both allow to cover the reference range of values.
5. On board stability: The on board stability of Yumizen C1200 Creatinine PAP is longer.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is written in large, bold, blue letters. Below it, the word "Medical" is written in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

ii. Comparison with predicate Device: Differences

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, light blue letters, is the word "Medical."

Stability depends on the reagent composition and cassette capacity.

  1. The reference range for serum/plasma is similar.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is written in large, bold, blue letters. Below the word "HORIBA" is the word "Medical" written in a smaller, lighter blue font. The logo is simple and clean, and it is likely used to represent the company in its marketing and branding materials.

6- Special Control/Guidance Document Referenced

a. Standards Followed

The following standards & FDA guidance documents have been used to support this submission:

CLSI Guidelines:

  • . CLSI EP05-A3:Evaluation of Precision of Quantitative Measurement Procedures- Third Edition - October 2014
  • CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement ● Procedures - Second Edition - June 2012
  • CLSI EP09-A3: Measurement Procedure Comparison and Biais Estimation Using Patient ● Samples- Third Edition - August 2013
  • CLSI EP06-A: Evaluation of the Linearity of Quantitative measurement Procedures A ● Statistical Approach - First Edition - April 2003
  • CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical ● laboratory- Third Edition - November 2008
  • CLSI EP25-A : Evaluation of Stability of In Vitro Diagnostic reagents- First Edition- September ● 2009

b. FDA Guidances Followed

  • Guidance for Industry and FDA Staff : Format for Traditional and Abbreviated 510(k)s 2019 ●
  • . Refuse To Accept (RTA) Policy for 510(k) - Guidance for Industry and Food and Drug Administration Staff Document issued on: September 13, 2019.
  • Guidance for Industry and FDA Staff : eCopy Program for Medical Device Submissions - 2015

c. Others Guidances followed

  • Valtec guideline (Vassault et al., Ann. Biol. Clin., 1986, (44), 686-745) ●

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

7- Analytical Performance Characteristics

8.1 Measuring Range

Yumizen C1200 Creatinine PAP .

The limit of detection and quantitation was determined according to the CLSI guideline EP17-A2. The reagent linearity was determined according to CLSI guideline EP06-A. The limit of quantitation and the linearity studies showed measuring range is appropriate.

A Results :

Limit ofdetectionLimit ofquantitationLinearityMeasuring range
Serum0.03 mg/dL0.11 mg/dL0.04 - 19.93 mg/dL0.11 - 16.95 mg/dL
SerumPost-dilutionNANANAuntil 50.85 mg/dL with theautomatic post-dilution
Urine0.16 mg/dL1.13 mg/dL0.00 - 327.60 mg/dL3.56 - 175 mg/dL
UrinePost-dilutionNANANAuntil 525 mg/dL with theautomatic post-dilution

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is written in large, bold, blue letters. Below the word "HORIBA" is the word "Medical" written in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

8.2 Accuracy and Precision

Yumizen C1200 Creatinine PAP ●

Serum/Plasma :

  • Instrument variability: 20x2x2
    Within run precision: CV limits, for the low, middle and high level are respectively 4.5 %, 3.4 % and 1.8 % for serum or plasma.

Total precision: CV limits, for the low, middle and high level are respectively 6.0 %, 4.5 % and 2.4 % for serum and plasma.

SampleNMean(µmol/L)Mean(mg/dL)Within-Run(%CV)Between-Run(%CV)Between-Day(%CV)Between-Instrument(%CV)Total(%CV)
Yumizen C1200N Multi Control240132.881.500.50.70.80.91.5
Yumizen C1200P Multi Control240396.104.480.30.80.50.91.3
Sample 124038.350.431.41.22.20.62.9
Sample 224080.090.910.51.21.30.82.0
Sample 3240786.328.890.30.60.61.01.3

The results are within the specifications.

  • Lot to lot variability: 3x5x2x3 ●
    Within run precision: CV limits, for the low, middle and high level are respectively 4.5 %, 3.4 % and 1.8 % for serum or plasma.

Total precision: CV limits, for the low, middle and high level are respectively 6.0 %, 4.5 % and 2.4 % for serum or plasma.

SampleNMean(µmol/L)Mean(mg/dL)Within-Day(%CV)Between-Day (%CV)Within-Lot(%CV)Between-Lot(%CV)Total(%CV)
YumizenC1200 N MultiControl90136.041.540.60.60.90.00.9
YumizenC1200 P MultiControl90401.674.540.40.40.60.00.6
Sample 19018.920.211.74.34.60.04.6
Sample 29048.920.550.91.72.00.02.0
Sample 390158.801.790.40.60.70.00.7
Sample 490632.517.150.40.30.50.00.5

The results are within the specifications.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, the word "Medical" is in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

Urine :

  • Instruments variability: 20x2x2 ●
    Within run precision: CV limits, for the low, middle and high level are respectively 4.5 %, 3.8 % and 3.8 % for urine.

Total precision: CV limits, for the low, middle and high level are respectively 6.0 %, 5.0 % and 5.0 % for urine.

SampleNMean(µmol/L)Mean(mg/dL)Within-Run(%CV)Between-Run(%CV)Between-Day(%CV)Between-Instrument(%CV)Total(%CV)
Yumizen C1200Urine Level 1Control2406415.6572.500.82.02.12.33.8
Yumizen C1200Urine Level 2Control24013808.68156.041.02.41.32.13.6
Sample 1240638.307.211.21.92.12.84.2
Sample 22401049.5911.860.82.21.73.24.3
Sample 32408422.5395.180.82.21.32.83.9
Sample 424012762.06144.211.22.01.42.33.6

The results are within the specifications.

  • Lot to Lot Variability: 3x5x2x3
    Within run precision: CV limits, for the low, middle and high level are respectively 4.5 %, 3.8 % and 3.8 % for urine.

Total precision: CV limits, for the low, middle and high level are respectively 6.0 %, 5.0 % and 5.0 % for urine.

SampleNMean(umol/L)Mean(mg/dL)Within-Day(%CV)Between-Day (%CV)Within-Lot(%CV)Between-Lot(%CV)Total(%CV)
Yumizen C1200Urine Level 1Control906525.7373.740.80.61.00.01.0
Yumizen C1200Urine Level 2Control9013490.94152.450.80.40.90.20.9
Sample 190497.315.621.11.82.10.02.1
Sample 2901013.6711.450.91.01.30.01.3
Sample 3908328.2194.110.90.40.90.00.9
Sample 49012770.44144.310.70.50.80.00.8

The results are within the specifications.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image displays the logo for HORIBA Medical. The word "HORIBA" is written in a bold, sans-serif font and is colored in a bright blue. Below "HORIBA", the word "Medical" is written in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

8.3 Interferences

The Interferences were determined according to the CLSI guideline EP07-A2. The acceptable bias is defined at +/-10% of the value without interfering substances. The data in the following table represent the highest values for which no interferences higher than +/-10% have been observed.

Serum/Plasma
Hemoglobin290 µmol/L500 mg/dL
Triglycerides5.33 mmol/L466.38 mg/dL
Total Bilirubin351 µmol/L20.54 mg/dL
Direct Bilirubin202 µmol/L11.80 mg/dL
Ascorbic Acid340 µmol/L5.98 mg/dL
Acetylsalicylic Acid3.62 mmol/L65.16 mg/dL
Ibuprofen2.43 mmol/L50.10 mg/dL
Acetaminophen1324 µmol/L20 mg/dL
N-Acetylcystein275 µmol/L27.5 mg/dL
Glucose55 mmol/L985 mg/dL
Total Proteinfrom 1 to 133 g/L.
Methyldopa21.29 µmol/L0.45 mg/dL
L-Dopa20.28 µmol/L0.40 mg/dL
Calcium Dobesilate23.9 µmol/L1.00 mg/dL

. Yumizen C1200 Creatinine PAP

Urine
Hemoglobin290 µmol/L500 mg/dL
Triglycerides4.84 mmol/L423.33 mg/dL
Direct Bilirubin502 µmol/L29.35 mg/dL
Ascorbic acid340 µmol/L5.98 mg/dL
N-Acetylcystein1100 mmol/L110 mg/dL
pHAcidification or alcalinisation do not interfere with this test.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Horiba Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

8.4 Matrix comparison Serum /Lithium Heparin Plasma: Yumizen C1200 Creatinine PAP

Scope of Study

Study of the coagulation effect on the creatinine measurement with the Yumizen C1200 Creatinine PAP reagent.

The goal of this study is to show that coagulation process does not change the creatinine concentration and that the anticoagulant has no interfering action.

Description:

84 paired samples were evaluated on Yumizen C1200 analyser using Yumizen C1200 Creatinine PAP reagent.

For this study, each paired sample (serum and lithium heparin plasma) has been obtained from single donor.

The equation for the regression line using mean square regression was obtained.

Mean square regressionNMinMaxInterceptSloper2
Creatinine PAPSerum(mg/dL)840.3213.94-0.02811.00080.995
Creatinine PAPPlasma(mg/dL)0.3314.77

Conclusion:

The results show there is no significant difference between serum specimens and plasma with heparin specimens → coagulation does not have impact on creatinine determination with Yumizen C1200 Creatinine PAP reagent.

8.5 Method comparison with a predicate device

● Yumizen C1200 Creatinine PAP

Serum/Plasma:

This study has been carried out using recommendations found in the CLSI EP-9A3 guidance. Samples: Anonymous remnants of human serum specimens collected from CHU Nîmes (University Hospital Center). These samples are in the candidate measuring range and predicate measuring range.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, the word "Medical" is in a smaller, lighter blue font. The logo is simple and clean, with a focus on the company name.

103 native samples have been assayed in duplicate, in ascendant order and descendant order on 5 working days.

The equation for the regression line using Passing Bablok was obtained.

Passing BablokNMinMaxInterceptSlopeCorrelation – r²
Creatinine (mg/dL)1030.2215.02-0.01070.96110.997

Urine:

This study has been carried out using recommendations found in the CLSI EP-9A3 guidance. Samples: Anonymous remnants of human urine specimens collected from routine clinical laboratory. These samples are in the candidate measuring range and predicate measuring range.

129 native samples have been assayed in duplicate, in ascendant order and descendant order on 6 working days.

The equation for the regression line using Passing Bablok was obtained.

Passing BablokNMinMaxInterceptSlopeCorrelation – r²
Creatinine (mg/dL)1294.19164.340.22960.97720.994

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

8.6 Reagent Stability

Closed stability 8.6.1

The closed stability was determined according to the CLSI guideline EP25-A.

. Yumizen C1200 Creatinine PAP

Stability before opening: Stable up to the expiry date on the label if stored at 2-8°C.

The Shelf Life of Yumizen C1200 Creatinine PAP is 12 months.

Open stability 8.6.2

The open stability was determined according to the CLSI guideline EP25-A.

On board reagent Stability:

  • . Yumizen C1200 Creatinine PAP : The stability claim after opening, on-Board, is 6 weeks

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, light blue letters, is the word "Medical."

8.7 Reference range

The Reference Ranges were verified according to the CLSI guideline EP28-A3.

Yumizen C1200 Creatinine PAP ●

Serum/Plasma:

트 Men

45 "normal samples" from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.

Study was performed on 2 different working days.

The first replicate result for each subject was compared against reference ranges cited in literature. The verification studies support the following reference ranges which were established through literature.

Normal range

59 -104 umol/L / 0.67 -1.17 mg/dL.

Reference:

Mazzachi BC, Peake MJ, Ehrhard V. Reference range and method comparison studies for enzymatic and Jaffe creatinine assays in plasma and serum and early morning urine. Clin. Lab. (2000) 46: 53-55.

트 Women

41 "normal samples" from blood bank have been assayed with the method in evaluation. Each sample is assayed in duplicates.

Study was performed on 2 different working days. The first result for each subject was compared against reference ranges cited in literature.

The verification studies support the following reference ranges which were established through literature.

Normal range

45 - 84 umol/L / 0.51 - 0.95 mg/dL.

Reference:

Mazzachi BC, Peake MJ, Ehrhard V. Reference range and method comparison studies for enzymatic and Jaffe creatinine assays in plasma and serum and early morning urine. Clin. Lab. (2000) 46: 53-55.

트 Children

Verification of normal values in pediatric samples could not be made due to lack of availability of samples from healthy pediatric patients.

Each laboratory should establish its own reference ranges.

Normal range

Childrenmg/dL \mumol/L
0 – 7 days0.6 – 1.153 – 97

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

1 week – 1 month0.3 – 0.727 – 62
1 – 6 month(s)0.2 – 0.418 – 35
7 – 12 months0.2 – 0.418 – 35
1 – 18 year(s)0.2 – 0.718 – 62

Reference:

Schlebusch Soldin SJ, HicksJM. Pediatric reference ranges. Washington: AACC Press, 1995:50.

Urine:

Verification of normal values in urine could not be made due to lack of availability of urine samples from healthy people.

Each laboratory should establish its own reference ranges.

Normal range Creatinine - Urine (24 hours):

MenWomen
14 - 26 mg/kg/day11 – 20 mg/kg/day
124 - 230 μmol/kg/day97 – 177 μmol/kg/day

Reference:

Roberts WL, McMillin GA, Burtis CA, Bruns DE. Reference Information for the Clinical Laboratory, TIETZ Textbook of Clinical Chemistry and Molecular Diagnostics. 4th Ed; Burtis CA, Ashwood ER, Bruns DE, (Elsevier Saunders eds. St Louis, USA), (2006): 2264

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for HORIBA Medical. The word "HORIBA" is in large, bold, blue letters. Below it, in smaller, lighter blue letters, is the word "Medical."

9. Proposed Labeling

The labeling is written as per the recommendations given in standard EN18113-2. It takes into account the requirements of 21 CFR Part 809.10.

10. Conclusions for Performance Testing

The performance testing data conclude that the safety and effectiveness of the device are not compromised, and that they met all acceptance criteria, demonstrating that each device is substantially equivalent to its predicate device.

§ 862.1225 Creatinine test system.

(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.