K Number
K193588

Validate with FDA (Live)

Device Name
DD Contrast
Manufacturer
Date Cleared
2020-04-16

(115 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DD contrast® veneering ceramics system is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.

Device Description

DD contrast veneering ceramics are dental ceramics for the esthetic individualization of crowns and bridges made from zirconia or lithium disilicate. The DD contrast group consists of various ceramic veneering pastes (already pre-mixed) for coloring, texturing and glazing as well as liquids for adjusting the textures of the paste variants. The DD contrast products are applied to the sintered zirconium dioxide or lithium disilicate restoration and fired in a standard furnace for veneering ceramics. Due to the different consistencies, the pastes can be used to achieve thin shade and glaze layers as well as structure-building areas on the crowns. The veneering ceramics are of type I, class 1b) according to ISO 6872 (FDA Recognition Number 4-251).

AI/ML Overview

The provided document is a 510(k) Premarket Notification for a dental device, specifically DD Contrast veneering ceramics. It does not describe a study involving an AI/ML algorithm or a device that would necessitate reporting on acceptance criteria for AI performance metrics like sensitivity, specificity, or AUC.

Instead, the document focuses on demonstrating substantial equivalence to an already legally marketed predicate device (InSync Ceramic System / MiYO Esthetic System Kit) based on non-clinical tests evaluating material properties and biocompatibility.

Here's an analysis of the provided information, framed to address the prompt's questions where applicable, but highlighting that AI-specific criteria are not relevant to this document:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are established by recognized international standards for dental ceramics and biocompatibility. The "reported device performance" demonstrates compliance with these standards.

Acceptance Criteria (Standard & Requirement)Reported Device Performance (DD Contrast)
Biocompatibility: ISO 10993-1 & ISO 7405 compliantProven to be biocompatible
Ceramic Materials: ISO 6872 (Type 1, Class 1b)Type 1, Class 1b
Flexural strength: $\ge$ 50 MPa (based on predicate and ISO 6872)$\ge$ 50 MPa
Material Composition: Feldspar Ceramics (SiO2, Al2O3, B2O3, K2O, Na2O)Feldspar ceramics, main components: SiO2, Al2O3, B2O3, K2O, Na2O
Form: Pastes and liquidsPastes and liquids
Shades: Different tooth and individualization shadesDifferent tooth and individualization shades

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not specify a "test set sample size" in the context of an AI/ML model. The testing here refers to material characterization and biocompatibility. The provenance is likely from Dental Direkt GmbH's own testing facilities or accredited labs in Germany, as the company is based in Germany. The nature of these tests (material properties) generally implies prospective testing performed on manufactured samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This is not applicable to the provided document. "Ground truth" in the context of this submission refers to the established scientific and regulatory standards for dental materials, not expert consensus on diagnostic images.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This is not applicable. Adjudication methods are relevant for subjective evaluations, typically in clinical readings or when establishing ground truth from multiple human experts for AI model training/testing. The tests performed here (biocompatibility, flexural strength, material composition) are objective and quantitative measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC study is not mentioned and is not relevant to this device. This device is a dental ceramic material, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a physical dental material and not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by recognized international standards for dental materials and biocompatibility (e.g., ISO 10993-1, ISO 7405, ISO 6872). The device is evaluated against the parameters and thresholds defined in these standards.

8. The sample size for the training set

This is not applicable. There is no AI/ML algorithm or training set discussed in this document.

9. How the ground truth for the training set was established

This is not applicable as there is no AI/ML algorithm or training set discussed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 16, 2020

Dental Direkt GmbH Patrick Berz Regulatory Affairs Manager Industriezentrum 106-108 Spenge, 32139 Germany

Re: K193588

Trade/Device Name: DD Contrast Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 15, 2020 Received: January 17, 2020

Dear Patrick Berz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, PhD Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K193588

Device Name DD Contrast

Indications for Use (Describe)

The DD contrast® veneering ceramics system is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Dental Direkt. The logo consists of a stylized cross made up of gray and dark gray squares, with one orange square in the center. To the right of the cross are the words "Dental" in light gray and "Direkt" in dark gray, stacked on top of each other.

Dental Direkt GmbH I Industriezentrum 106 – 108 I D-32139 Spenge

510(k) Summary

Submitter of 510(k)Dental Direkt GmbHIndustriezentrum 106-10832139 Spenge / Germany
Contact PersonMr. Uwe Greitens, CEOPhone:+49 5225 86319-0Fax: +49 5225 86319-99u.greitens@dentaldirekt.de
Regulatory CorrespondentMr. Patrick Berz, Manager Regulatory AffairsPhone:+49 5225 86319-0Fax: +49 5225 86319-99p.berz@dentaldirekt.de
Establishment Registration No.3008347275
Date Prepared2020/04/16
Trade Name of DeviceDD contrast
Common NamePowder, Porcelain
Classification NamePorcelain Powder for clinical use
Product CodeEIH
Regulation Number21 CFR 872.6660
ClassificationClass II
510(k) NumberK193588

{4}------------------------------------------------

Primary Predicate DeviceK111743: Jensen Industries:InSync Ceramic System (Model name: "MiYO – liquidceramic")
Indications for UseThe DD contrast® veneering ceramics system issuitable for the esthetic individualization of monolithicor minimally reduced dental crown and bridgeconstructions made of zirconium dioxide or lithiumdisilicate.
Device DescriptionDD contrast veneering ceramics are dental ceramicsfor the esthetic individualization of crowns andbridges made from zirconia or lithium disilicate. TheDD contrast group consists of various ceramicveneering pastes (already pre-mixed) for coloring,texturing and glazing as well as liquids for adjustingthe textures of the paste variants.The DD contrast products are applied to the sinteredzirconium dioxide or lithium disilicate restoration andfired in a standard furnace for veneering ceramics.Due to the different consistencies, the pastes can beused to achieve thin shade and glaze layers as wellas structure-building areas on the crowns.The veneering ceramics are of type I, class 1b)according to ISO 6872 (FDA Recognition Number 4-251).
Technological CharacteristicsThe technological characteristics of the new productare the same as those of the legally marketed device.Both products are feldspar ceramics fired onto adental ceramic framework made of zirconium dioxideor lithium disilicate. With this veneering technique it ispossible to design a crown or bridge to individualpatient requirements.
Material compositionFeldspar ceramics, main components: SiO2, Al2O3,B2O3, K2O, Na2O

{5}------------------------------------------------

Discussion ofTests PerformedClinical TestsDental Direkt GmbH did not conduct, nor rely upon, clinical tests to determine substantial equivalence as dental ceramics that fall under FDA product code EIH have a long history of safe and effective use in the US.Non-Clinical TestsNon-clinical testing was performed in order to validate the product against the company's specified design requirements according to the following standards:ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (FDA Recognition #4-212) ISO 7405, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry (FDA Recognition #2-258) The DD contrast products are proven to be biocompatible. ISO 6872, Dentistry - Ceramic materials (FDA Recognition #4-251)
SubstancialEquivalenceMaterialThe composition of DD contrast corresponds to that of the predicate device.Physical Properties

DD contrast has physical properties comparable to those of predicates.

{6}------------------------------------------------

Technical comparison

FeaturePredicate DeviceNew Device
Trade nameInSync Ceramic SystemDD contrast
Model nameMiYO Esthetic System KitDD contrast
Indications
Indications for UseStatement"InSync Ceramic System":The pressable ceramicpellets are pressed ontozirconia frames by dentaltechnicians to fabricate fullceramic crowns and theceramic layering porcelainand liquids are used tobuild up the pressedceramic to final toothmorphology and shade.The ceramic layeringporcelain and liquids arealso used in buildingceramic crowns andbridges on titanium andtitanium alloy substructures.Both applications are toprovise protheses formissing / damaged teeth."MiYO Esthetic System Kit":A Type 1, class 1b ceramicfor the coloring, enhancing,and glazing of the followingrestoration types: Zirconiaand Lithium Disilicate.The DD contrast®veneering ceramicssystem is suitable for theesthetic individualization ofmonolithic or minimallyreduced dental crown andbridge constructions madeof zirconium dioxide orlithium disilicate.
Comment on Indications for Use Statement:The indication of the predicate ("InSync CeramicSystem") is comparable to that of the new device, even ifit is written in a very general way and includes differentveneering ceramics. The relevant parts of the indicationdescribe "layering porcelain and liquids" which areapplied to a pressed ceramic veneer which has beenpreviously pressed on a zirconium framework. Evenbetter, however, is the comparability of the correspondingmodel variant (“MiYO Esthetic System Kit”) resulting fromthis group, whose indication is identical to that of the newproduct.
Product for veneering /individualization of dentalframeworksYesYes

{7}------------------------------------------------

FeaturePredicate DeviceNew Device
Framework materialZirconium dioxide andlithium disilicateZirconium dioxide andlithium disilicate
Minimal amount of fires21
For all sizes of dentalframeworksYesYes
Classification acc. to ISO6872Type 1, class 1bType 1, class 1b

{8}------------------------------------------------

FeaturePredicate DeviceNew Device
Technological Characteristics
MaterialFeldspar ceramicsFeldspar ceramics
FormPastes and liquidsPastes and liquids
ShadesDifferent tooth and individualization shadesDifferent tooth and individualization shades
Tested according to ISO 6872YesYes
Flexural strength [MPa]$\ge$ 50$\ge$ 50

Substantial Equivalence Conclusion

Dental Direkt GmbH believes that the new product DD contrast is substantially equivalent to the predicate device when used as instructed by knowledgeable and trained dental personnel. There is no change of the intended use of the new device and no difference in fundamental scientific technology. The new device is made from the same materials as the predicate device. The new device was verified in accordance with the abovementioned FDA recognized standards.

Dental Direkt GmbH therefore believes that DD contrast is substantially equivalent to the legally marketed predicate device.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.