K Number
K193582

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-03-17

(85 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aesculap SterilContainer™ S2 System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide
  • STERRAD 100NX DUO cycle
  • . STERIZONE VP4

The Aesculap SterilContainer S2 System includes accessories such as silicone mats and organizers, stainless steel baskets, trays, holders, sterilization indicator cards and tamper proof locks.

Device Description

The Aesculap SterilContainer S2 System is a reusable rigid container system intended for sterilization and storage of other medical devices. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide ●
  • STERRAD 100NX DUO cycle
  • STERIZONE VP4 ●

The containers are perforated and made from anodized aluminum and utilize a single-use filter. The SterilContainer S2 System includes accessories such as mats, baskets, trays, instrument holders, organizers, filters, indicator cards and tamper proof locks.

AI/ML Overview

The provided document is a 510(k) summary for a medical device (sterilization container system), not for an AI/ML powered device. As such, the concept of a "test set," "training set," "experts to establish ground truth," "adjudication method," or "MRMC comparative effectiveness study" does not apply to this type of device submission. The performance testing outlined is for sterilization efficacy, simulated use, material compatibility, and cytotoxicity, which are standard for sterilization container systems and do not involve AI/ML components.

Therefore, I cannot extract the information required by your prompt regarding AI/ML device acceptance criteria and study details from this document. The document describes the acceptance criteria and study that proves the device meets those criteria, but these are for a physical sterilization device, not an AI/ML algorithm.

If you are interested in the acceptance criteria and study for the sterilization container itself, I can provide that information based on the document.

However, assuming you are looking for the outlined AI/ML specific information, I must state that the document does not contain it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 17, 2020

Aesculap, Inc. Sierra Mertz Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K193582

Trade/Device Name: SterilContainer S2 System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: December 19, 2019 Received: December 23, 2019

Dear Sierra Mertz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K193582

Device Name

Aesculap SterilContainer S2"™ System

Indications for Use (Describe)

The Aesculap SterilContainer™ S2 System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide
  • STERRAD 100NX DUO cycle
  • . STERIZONE VP4

The Aesculap SterilContainer S2 System includes accessories such as silicone mats and organizers, stainless steel baskets, trays, holders, sterilization indicator cards and tamper proof locks.

The attached table identifies the validated load configurations for each of the modalities.

Sterilization CycleContainer SizeValidated Load Configuration
EtO(130°F, 60 minute exposure,>50% RH 725mg/L gas pressure)Full(1) lumen ( $\geq$ 3mm ID x $\leq$ 400mm L)
Three-QuarterAND(1) lumen ( $\geq$ 3.8mm ID x $\leq$ 370 mm L)
Half
STERRAD 100NX DUO(bottom shelf only)FullFlexible scope ( $\geq$ 1mm ID x $\leq$ 850mm L)
Three-Quarter
Half
STERIZONE VP4Validated Loads 1 & 2(Based on Sterizone Load #7)FullNon Lumened Instruments
Three-Quarter
Half
STERIZONE VP4Validated Load 3(Based on Sterizone Load #8)JS440 (base) +JS489 (lid)(1) Single Channel Flexible Scope ( $\geq$ 1mm ID x $\leq$ 850mm L)OR(1) Dual Channel Flexible Scope ( $\geq$ 1mm ID x $\leq$ 850 mm L and $\geq$ 1 mm ID x $\leq$ 989mm L)
STERIZONEValidated Load 4(Based on Sterizone Load #4)JS440 (base) +JS489 (lid)(1) Semi-rigid dual channel scope ( $\geq$ 0.7mm ID x $\leq$ 500mm L and $\geq$ 1.1mm ID x $\leq$ 500mm L)AND one of the following:(4) Stainless steel lumens( $\geq$ 5.5mm ID x $\leq$ 166mm L; $\geq$ 7mm ID x $\leq$ 105mm L; $\geq$ 7.0mm ID x $\leq$ 227mm L;$\geq$ 7.8mm ID x $\leq$ 198mm L)OR(2) Stainless steel lumens( $\geq$ 4mm ID x $\leq$ 370mm L; $\geq$ 2mm ID x $\leq$ 152mm L)OR(3) Stainless steel lumens( $\geq$ 2.2mm ID x $\leq$ 173mm L; $\geq$ 4.7mm ID x $\leq$ 270mmL ; $\geq$ 4mm ID x $\leq$ 445mm L)

Table 1. SterilContainer S2 System Validated Load Configurations

{3}------------------------------------------------

Table 2. SterilContainer S2 System Load Weights

SterilizationMethodContainer SizeContainerBottom Part #ContainerLid Part #Total Loaded Container(lb)
Full Size - 4 ¼"JS440JS48925
Full Size - 5 ½"JS441
Full Size - 6"JS442
Full Size - 8"JS444
Three-Quarter Size - 4 ¼"JS740JS78925
EtOThree-Quarter Size - 5 ½"JS741
Three-Quarter Size - 6"JS742
Three-Quarter Size - 8"JS744*
Half Size - 4 ½"JS340JS38925
Half Size - 5 ½"JS341
Half Size - 6"JS342
Full Size - 4 ¼"JS440JS48910.97
Full Size - 5 ½"JS441
Full Size - 6"JS442
Full Size - 8"JS444
Three-Quarter Size - 4 ¼"JS740JS78910.04
STERRAD 100NX DUOThree-Quarter Size - 5 ½"JS741
Three-Quarter Size - 6"JS742
Half Size - 4 ½"JS340JS38911.7
Half Size - 5 ½"JS341
Half Size - 6"JS342
Full Size - 4 ¼"JS440JS48925
Full Size - 5 ½"JS441
Full Size - 6"JS442
Full Size - 8"JS444
STERIZONEValidated Loads 1 & 2Three-Quarter Size - 4 ¼"JS740JS78925
Three-Quarter Size - 5 ½"JS741
Three-Quarter Size - 6"JS742
(Based on Sterizone Load #7)Half Size - 4 ½"JS340JS38925
Half Size - 5 ½"JS341
Half Size - 6"JS342
Half Size - 5 ½"JS341
Half Size - 6"JS342
STERIZONEValidated Load 3Full Size - 4 ¼"JS440JS489See load configurationin table 1 above
(Based on Sterizone Load #8)
STERIZONEValidated Load 4Full Size - 4 ¼"JS440JS489See load configurationin table 1 above

*JS744 is for use in Ethylene Oxide only.

{4}------------------------------------------------

Table 3. Sterilization Cycle Compatible Accessories

AccessoriesCompatible withEthylene OxideCompatible withSTERRAD DUOCompatible withSTERIZONE VP4
Stainless Steel baskets,basket lids, and dividersYesYesYes
Instrument Organization System(Silicone and Stainless Steel racks,brackets, holders, and clamps)YesYesYes
Silicone matsYesNoYes
Tamper Evident locks and indicatorcardsYesYesYes

Type of Use (Select one or both, as applicable)

__ Prescription Use (Part 21 CFR 801 Subpart D)


CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) SUMMARY (as required by 21 CFR 807.92)

Aesculap SterilContainer S2 System for Ethylene Oxide, STERIZONE VP4, STERRAD 100NX DUO March 11, 2020

COMPANY:Aesculap, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 291671-
CONTACT:Sierra M. Mertz610-984-9076 (phone)Sierra.mertz@aesculapimplants.com610-791-6882 (fax)
TRADE NAME:Aesculap® SterilContainer™ S2 System
COMMON NAME:Sterilization Container Wrap
CLASSIFICATION NAME:Wrap, Sterilization
REGULATION NUMBER:880.6850
PRODUCT CODE:KCT
DEVICE CLASS:Class II per 21 CFR 880.6850

PREDICATE DEVICES

Aesculap SterilContainer S2 System - K182414 (Primary), SterilContainer S System - K162815 SterilContainer System - K112671

DEVICE DESCRIPTION

The Aesculap SterilContainer S2 System is a reusable rigid container system intended for sterilization and storage of other medical devices. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide ●
  • STERRAD 100NX DUO cycle
  • STERIZONE VP4 ●

The containers are perforated and made from anodized aluminum and utilize a single-use filter. The SterilContainer S2 System includes accessories such as mats, baskets, trays, instrument holders, organizers, filters, indicator cards and tamper proof locks.

{6}------------------------------------------------

INDICATIONS FOR USE

The Aesculap SterilContainer™ S2 System is a reusable rigid sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in the following sterilization modalities:

  • Ethylene Oxide
  • STERRAD 100NX DUO cycle
  • STERIZONE VP4

The Aesculap SterilContainer S2 System includes accessories such as silicone mats and organizers, stainless steel baskets, trays, holders, sterilization indicator cards and tamper proof locks.

The attached table identifies the validated load configurations for each of the modalities.

Table 1. SterilContainer S2 System Validated Load Configurations

Sterilization CycleContainerSizeValidated Load Configuration
EtO(130°F, 60 minute exposure,≥50% RH 725mg/L gaspressure)FullThree-QuarterHalf(1) lumen (≥ 3mm ID x ≤ 400mm L)AND(1) lumen (≥ 3.8mm ID x ≤ 370 mm L)
STERRAD 100NX DUO(bottom shelf only)FullThree-QuarterHalfFlexible scope (≥ 1mm ID x ≤ 850mm L)
STERIZONE VP4Validated Loads 1 & 2(Based on Sterizone Load #7)FullThree-QuarterHalfNon Lumened Instruments
STERIZONE VP4Validated Load 3(Based on Sterizone Load #8)JS440 (base)+ JS489 (lid)(1) Single Channel Flexible Scope (≥1mm ID x ≤ 850mm L)OR(1) Dual Channel Flexible Scope (≥1mm ID x ≤850 mm L and ≥1 mm IDx ≤ 989mm L)
STERIZONEValidated Load 4(Based on Sterizone Load #4)JS440 (base)+ JS489 (lid)(1) Semi-rigid dual channel scope (≥0.7mm ID x ≤500mm L and ≥1.1mmID x ≤500mm L)AND one of the following:(4) Stainless steel lumens(≥ 5.5mm ID x ≤ 166mm L; ≥ 7mm ID x ≤ 105mm L; ≥ 7.0mm ID x ≤227mm L:≥ 7.8mm ID x ≤ 198mm L)OR(2) Stainless steel lumens(≥ 4mm ID x ≤ 370mm L; ≥ 2mm ID x ≤ 152mm L)OR(3) Stainless steel lumens(≥ 2.2mm ID x ≤ 173mm L; ≥ 4.7mm ID x ≤ 270mmL ; ≥ 4mm ID x ≤445mm L)

{7}------------------------------------------------

SterilizationMethodContainer SizeContainerBottom Part#ContainerLid Part #Total LoadedContainer (lb)
EtOFull Size - 4 ¼"JS440JS48925
Full Size - 5 ½"JS441
Full Size - 6"JS442
Full Size - 8"JS444
Three-Quarter Size - 4 ¼"JS740JS78925
Three-Quarter Size - 5 ½"JS741
Three-Quarter Size - 6"JS742
Three-Quarter Size - 8"JS744*
Half Size - 4 ½"JS340JS38925
Half Size - 5 ½JS341
STERRAD 100NX DUOFull Size - 4 ¼"JS440JS48910.97
Full Size - 5 ½"JS441
Full Size - 6"JS442
Full Size - 8"JS444
Three-Quarter Size - 4 ¼"JS740JS78910.04
Three-Quarter Size - 5 ½"JS741
Three-Quarter Size - 6"JS742
Half Size - 4 ¼"JS340JS38911.7
Half Size - 5 ½JS341
Half Size - 6"JS342
STERIZONEValidated Loads 1 & 2(Based on Sterizone Load#7)Full Size - 4 ¼"JS440JS48925
Full Size - 5 ½"JS441
Full Size - 6"JS442
Full Size - 8"JS444
Three-Quarter Size - 4 ¼"JS740JS78925
Three-Quarter Size - 5 ½"JS741
Three-Quarter Size - 6"JS742
Half Size - 4 ¼"JS340JS38925
Half Size - 5 ½JS341
Half Size - 6"JS342
STERIZONEValidated Load 3(Based on Sterizone Load#8)Full Size - 4 ¼"JS440JS489See loadconfiguration in table1 above
STERIZONEValidated Load 4(Based on Sterizone Load#4)Full Size - 4 ¼"JS440JS489

*JS744 is for use in Ethylene Oxide only.

{8}------------------------------------------------

Table 3. Sterilization Cycle Compatible Accessories
AccessoriesCompatible withEthylene OxideCompatible withSTERRAD DUOCompatible withSTERIZONE VP4
Stainless Steel baskets,basket lids, and dividersYesYesYes
Instrument Organization System(Silicone and Stainless Steel racks,brackets, holders, and clamps)YesYesYes
Silicone matsYesNoYes
Tamper Evident locks and indicatorcardsYesYesYes

{9}------------------------------------------------

TECHNOLIGICAL CHARACTERISTICS (compared to predicate devices)

The Aesculap SterilContainer S2 System has the same intended use and the technological characteristics as the Aesculap SterilContainer predicate devices. The subject devices are offered in the same design and same sizes and are made of the same materials as the predicate devices.

Subject Device -AesculapSterilContainer S2(K193582)Primary Predicate -AesculapSterilContainer S2(K182414)Predicate -AesculapSterilContainer(K112671)Predicate -AesculapSterilContainer S(K162815)Discussion
IntendedUseA device intended to beused to enclose anothermedical device that is to besterilizedby a health care provider. Itis intended to allowsterilizationof the enclosed medicaldevice and also to maintainsterility of the encloseddevice until used.A device intended to beused to enclose anothermedical device that is to besterilizedby a health care provider. Itis intended to allowsterilizationof the enclosed medicaldevice and also to maintainsterility of the encloseddevice until used.A device intended to beused to enclose anothermedical device that is to besterilizedby a health care provider. Itis intended to allowsterilizationof the enclosed medicaldevice and also to maintainsterility of the encloseddevice until used.A device intended to beused to enclose anothermedical device that is to besterilizedby a health care provider. Itis intended to allowsterilizationof the enclosed medicaldevice and also to maintainsterility of the encloseddevice until used.The subject deviceand the predicateshave the sameintended use.
SterilizationModalitiesEtOSTERIZONE VP4STERRAD 100NX DUOPrevac SteamGravity SteamSTERRAD: 100S, NXStandard, NXAdvanced, 100NXStandard, 100NX FlexSTERIS: V-PRO 60Lumen, V-PRO 60Non-Lumen, V-PRO 60 Flex, V-PRO maX Lumen, V-PROmaX Non-Lumen, V-PROmaX FlexEtOPreVac SteamGravity SteamIUSSSTERIZONE VP4The subject devicehas beenvalidated usingsterilizationmethods similar tothe predicatedevices.
MaterialContainer: AnodizedaluminumLid: Anodized aluminumGasket: SiliconeContainer: AnodizedaluminumLid: Anodized aluminumGasket: SiliconeContainer: AnodizedaluminumLid: Anodized aluminumGasket: SiliconeContainer: Non-anodizedaluminumLid: Non-anodizedaluminumGasket: SiliconeThe material of thesubject device isidentical to thepredicates.
Filter TypeSingle use: Paper(cellulose)or polypropyleneSingle use: Paper(cellulose) or polypropyleneReusable: PTFESingle use: Paper(cellulose) or polypropyleneReusable: PTFESingle use: polypropyleneThe proposeddevice utilizes anidentical singleuse filter.
ContainerDesignPerforated bottom withperforated lidPerforated bottom withperforated lidPerforated or solid bottomwith perforated lidPerforated bottom withperforated lidThe design of thesubject device isidentical to that ofthe predicatedevices.
Vent toVolume ratio1.02 - 3.41.4 - 3.41.021.4 - 3.4The vent-to-volume ratios ofthe subject devicefall within those ofthe predicates.
SizesFull sizeThree-Quarter SizeHalf SizeFull sizeThree-Quarter SizeHalf SizeThree-Quarter SizeFull sizeThree-Quarter SizeHalf SizeThe size of thesubject device andpredicate devicesare the same.
AccessoriesSilicone mats, baskets,trays, IOS, and racks.Silicone mats, baskets,trays, IOS, and racks.Silicone mats, baskets,trays, IOS, and racks.Silicone mats, baskets,trays, IOS and racks.The subject andpredicate devicesutilize the sameaccessories.

{10}------------------------------------------------

PERFORMANCE TESTING

Non-clinical testing was conducted to verify performance of the subject device. The testing summarized below demonstrates that the subject device meets the established acceptance criteria.

Performance TestingPurposeAcceptance CriteriaResults
Sterilization EfficacyTo determine sterilizationeffectiveness of testdevice after processing ina sterilization cycle.A sterility assurance level (SAL) of10-6 will be achieved post sterilization using theBI overkill method and half cycle validation.Biological indicators must be negative for growthafter incubation period.Pass
Simulated UseTo determine the effectivesterilization of flexiblescopes when used withthe test device.A minimum of 1.0 x 106 spores contained withinorganic soil representative of actual useconditions are killed during defined sterilizationcycle. Biological indicators must be negative forgrowth after incubation period.Pass
Material CompatibilityTo assess effects of fulluse cycles on devicecomponents and theirintended functionality.No degradation or impact to functionality at thecompletion multiple sterilization cycles.Pass
CytotoxicityTo determine the potentialof a test device to causecytotoxicity.Testing completed in accordance with ISO10993-5: 2009 to demonstrate no significantcytotoxic reaction after exposure to sterilant.Using the ISO Elution Method, the response tothe article is not greater than 2 (mild reactivity).Pass

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the Aesculap SterilContainer S2 System is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).