(98 days)
KOSMOS is intended to be used by qualified and trained healthcare professionals in the clinical assessment of the cardiac and pulmonary systems and the abdomen by acquiring, displaying, measuring, and storing synchronized ultrasound images, electrocardiogram (ECG) rhythms, and digital auscultation (DA) sounds and waveforms.
With respect to its ultrasound imaging capabilities, KOSMOS is a general purpose diagnostic ultrasound system used in the following clinical applications and modes of operation:
- . Clinical Applications: Cardiac, Thoracic/Lung, Abdominal, Peripheral Vascular, and Image Guidance for Needle/Catheter Placement
- . Modes of Operation: B-mode, Color Doppler, Combined Modes of B+M and B+CD, and Harmonic Imaging
KOSMOS is intended to be used in clinical care and medical education settings on adult and pediatric patient populations.
The device is non-invasive, reusable, and intended to be used on one patient at a time.
KOSMOS consists of a tablet and probe, which connects to the tablet via a cable. Its accessories include a power charger, ECG cable, and binaural headset. KOSMOS can operate on battery or while connected to mains.
The probe face houses an ultrasound transducer and sealed microphones for auscultation. The tablet can display clinical and patient data information including the display of ultrasound images, auscultation and ECG waveforms, and patient data/reports. The tablet also includes speakers for sounds associated with system control and feedback. Additionally, the tablet offers a means of user control with its touchscreen display and buttons.
Although its intended operation is not dependent on Wi-Fi, KOSMOS supports Wi-Fi connectivity for patient data archival.
The tablet can be positioned on a flat surface close to the patient while the user holds the probe with one hand to scan the patient. The user's other hand is then free to interact with the tablet using its touchscreen and buttons. The user can also hold the tablet in one hand and the probe in the other hand to scan the patient.
KOSMOS' ECG capability provides a timing reference with respect to the cardiac cycle as compared with both ultrasound imaging and digital auscultation. Ultrasound imaging, ECG, and DA are all integrated into the probe in a time-synchronized manner.
KOSMOS' 3-lead single-channel ECG allows for the acquisition and display of a single ECG waveform (lead), which can be any one of the Lead II, or Lead III waveforms. One end of the ECG cable connects to the probe via a custom-designed magnetic connector. The other end has three (3) RA/LA/LL leadwires to be connected to usersupplied clip-style electrodes affixed to the patient using the standard RA/LA/LL configuration.
The provided text is a 510(k) premarket notification for the KOSMOS device. It details the device's technical specifications and its comparison to predicate devices to demonstrate substantial equivalence, but it does not contain a study proving the device meets specific acceptance criteria based on clinical performance data.
Section 12 explicitly states: "An assessment of clinical performance data for KOSMOS was not required to support a determination of substantial equivalence." This means that the FDA did not require clinical study data to clear this device for market, and therefore, the document does not include information about acceptance criteria for clinical performance, the reported device performance against those criteria, or details regarding the design and execution of such a clinical study.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and the reported device performance.
- Sample size used for the test set and data provenance.
- Number of experts and their qualifications for ground truth.
- Adjudication method for the test set.
- MRMC comparative effectiveness study details.
- Standalone performance details.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
This information is typically found in clinical study reports, which are
not present in this 510(k) summary given the FDA's "not required" determination for clinical performance data in this instance. The substantial equivalence determination was based on technological characteristics and safety data, as detailed in Section 10 and 11 of the provided document.
{0}------------------------------------------------
March 26, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
EchoNous, Inc. % Ms. Trish Liau Regulatory Affairs Manager 8310 154th Avenue NE, Bldg B., Suite 200 REDMOND WA 98052
Re: K193518
Trade/Device Name: KOSMOS Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, DQD, DPS Dated: February 29, 2020 Received: March 3, 2020
Dear Ms. Liau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193518
Device Name KOSMOS
Indications for Use (Describe)
KOSMOS is intended to be used by qualified and trained healthcare professionals in the clinical assessment of the cardiac and pulmonary systems and the abdomen by acquiring, displaying, measuring, and storing synchronized ultrasound images, electrocardiogram (ECG) rhythms, and digital auscultation (DA) sounds and waveforms.
With respect to its ultrasound imaging capabilities, KOSMOS is a general purpose diagnostic ultrasound system used in the following clinical applications and modes of operation:
- . Clinical Applications: Cardiac, Thoracic/Lung, Abdominal, Peripheral Vascular, and Image Guidance for Needle/Catheter Placement
- . Modes of Operation: B-mode, Color Doppler, Combined Modes of B+M and B+CD, and Harmonic Imaging
KOSMOS is intended to be used in clinical care and medical education settings on adult and pediatric patient populations.
The device is non-invasive, reusable, and intended to be used on one patient at a time.
| Type of Use (Select one or both, as applicable) | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K193518
1. Submitter
EchoNous, Inc. 8310 154th Ave NE Bldg B. Ste 200 Redmond, WA 98052 USA
2. Contact Person
Trish Liau Regulatory Affairs Manager Telephone: (425) 402-4044 E-mail: patricia.liau(@echonous.com
3. Date Prepared
December 17, 2019
-
- Device / Marketing Trade Name KOSMOS (subject to change)
ર. Common / Usual Name
Diagnostic ultrasound system with integrated electronic stethoscope and electrocardiograph
6. Classification
Regulatory Device Class: II Classification Panel: Radiology, Cardiovascular
| Classification Name | 21 CFR Number | Product Code |
|---|---|---|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
| Electronic Stethoscope | 870.1875 | DQD |
| Electrocardiograph | 870.2340 | DPS |
7. Predicate Devices
Primary predicate device: GE Vscan Extend (K180995); Product Code: IYN, IYO, ITX Secondary predicate device: Eko Duo Model E5 (K170874); Product Code: DQD, DPS
8. Device Description
KOSMOS consists of a tablet and probe, which connects to the tablet via a cable. Its accessories include a power charger, ECG cable, and binaural headset. KOSMOS can operate on battery or while connected to mains.
The probe face houses an ultrasound transducer and sealed microphones for auscultation. The tablet can display clinical and patient data information including the display of ultrasound images, auscultation and ECG waveforms, and patient data/reports. The
{4}------------------------------------------------
tablet also includes speakers for sounds associated with system control and feedback. Additionally, the tablet offers a means of user control with its touchscreen display and buttons.
Although its intended operation is not dependent on Wi-Fi, KOSMOS supports Wi-Fi connectivity for patient data archival.
The tablet can be positioned on a flat surface close to the patient while the user holds the probe with one hand to scan the patient. The user's other hand is then free to interact with the tablet using its touchscreen and buttons. The user can also hold the tablet in one hand and the probe in the other hand to scan the patient.
KOSMOS' ECG capability provides a timing reference with respect to the cardiac cycle as compared with both ultrasound imaging and digital auscultation. Ultrasound imaging, ECG, and DA are all integrated into the probe in a time-synchronized manner.
KOSMOS' 3-lead single-channel ECG allows for the acquisition and display of a single ECG waveform (lead), which can be any one of the Lead II, or Lead III waveforms. One end of the ECG cable connects to the probe via a custom-designed magnetic connector. The other end has three (3) RA/LA/LL leadwires to be connected to usersupplied clip-style electrodes affixed to the patient using the standard RA/LA/LL configuration.
9. Intended Use / Indications for Use
KOSMOS is intended to be used by qualified and trained healthcare professionals in the clinical assessment of the cardiac and pulmonary systems and the abdomen by acquiring. processing, displaying, measuring, and storing synchronized ultrasound images, electrocardiogram (ECG) rhythms, and digital auscultation (DA) sounds and waveforms.
With respect to its ultrasound imaging capabilities, KOSMOS is a general purpose diagnostic ultrasound system used in the following clinical applications and modes of operation:
- . Clinical Applications: Cardiac, Thoracic/Lung, Abdominal, Peripheral Vascular, and Image Guidance for Needle/Catheter Placement
- Modes of Operation: B-mode, M-mode, Color Doppler, Combined Modes of B+M ● and B+CD, and Harmonic Imaging
KOSMOS is intended to be used in clinical care and medical education settings on adult and pediatric patient populations.
The device is non-invasive, reusable, and intended to be used on one patient at a time.
Type of Use: Prescription Use (Part 21 CFR 801 Subpart D)
{5}------------------------------------------------
10. Basis for Substantial Equivalence
KOSMOS is substantially equivalent to its predicate devices with regards to intended use, technological characteristics, and safety and effectiveness. A comparison table is provided below.
| Feature | KOSMOS System(This 510(k) submission) | GE Vscan Extend UltrasoundSystem (K180995) | Eko Duo Model E5(K170874) |
|---|---|---|---|
| Intended Use /Indications forUse | KOSMOS is intended to be usedby qualified and trained healthcareprofessionals in the clinicalassessment of the cardiac andpulmonary systems and theabdomen by acquiring, processing,displaying, measuring, and storingsynchronized ultrasound images,electrocardiogram (ECG) rhythms,and digital auscultation (DA)sounds and waveforms.With respect to its ultrasoundimaging capabilities, KOSMOS isa general purpose diagnosticultrasound system used in thefollowing clinical applications andmodes of operation:• Clinical Applications: Cardiac,Thoracic/Lung, Abdominal,Peripheral Vascular, and ImageGuidance for Needle/CatheterPlacement• Modes of Operation: B-mode,M-mode, Color Doppler,Combined Modes of B+M andB+CD, and Harmonic ImagingKOSMOS is intended to be usedin clinical care and medicaleducation settings on adult andpediatric patient populations.The device is non-invasive,reusable, and intended to be usedon one patient at a time.Type of Use: Prescription Use(Part 21 CFR 801 Subpart D) | Vscan Extend is a general purposediagnostic ultrasound imagingsystem for use by qualified andtrained healthcare professionalsenabling visualization andmeasurement of anatomicalstructures and fluid. Its pocket-sized portability and simplifieduser interface enables integrationinto examination and trainingsessions indoors and in otherenvironments described in the usermanual. The information can beused for basic/focused assessmentsand adjunctively with othermedical data for clinical diagnosispurposes during routine, periodicmonitoring, and triage.With the phased array transduceron the sector probe, the specificclinical applications and examtypes include: Cardiac;Abdominal; Renal; OB/GYN;Urology; Fetal, Evaluation ofPresence of Fluid; ImagingGuidance for Needle/CatheterPlacement (e.g. paracentesis,pericardiocentesis, thoracentesis,amniocentesis); PeripheralVascular Imaging (e.g. arteriesand veins); Thoracic/Lung (e.g.pleural motion/sliding, lineartifacts); Adult Cephalic; andPediatrics.With the addition of the lineararray transducer on the single dualheaded probe solution, the specificclinical applications and examtypes are expanded to include:Peripheral vascular imaging (e.g.lower extremity, carotid);Procedure Guidance for Arterial orVenous Vessels (e.g. central lines,upper extremity); Small Organs(e.g. thyroid); Musculoskeletal(Long Bone; Hip, shoulder, elbowand Knee Joints); Evaluation ofPresence of Fluid; Thoracic/Lung(e.g. pleural motion/sliding, lineartifacts); and Pediatrics.Type of Use: Prescription Use(Part 21 CFR 801 Subpart D) | The Eko Model E5 System isintended to be used by healthcareprofessionals to electronicallyamplify, filter, and transfer bodysounds and single-channelelectrocardiogram (ECG)waveforms. The Eko Model E5System also displays ECGwaveforms and phonocardiogramwaveforms on the accompanyingmobile application for storage andsharing (when prescribed or usedunder the care of a physician). Itcan be used to record heart soundsand cardiac murmurs, bruits,respiratory sounds, and abdominalsounds during physicalexamination in normal patients orthose with suspected diseases ofthe cardiac, vascular, pulmonary,or abdominal organ systems. Thedevice can be used on adults andpediatrics.The data offered by the device isonly significant when used inconjunction with physician overread as well as consideration ofother relevant patient data.The device should not be used oninfants weighing less than 10kg.Type of Use: Prescription Use (Part21 CFR 801 Subpart D) |
| Feature | KOSMOS System(This 510(k) submission) | GE Vscan Extend UltrasoundSystem (K180995) | Eko Duo Model E5(K170874) |
| Ultrasound Substantial Equivalence (Technological Characteristics) | |||
| Transducer Types | Phased Array | Phased ArrayLinear Array | |
| ClinicalApplications | Phased Transducer:Anatomy/ Region of Interest:AbdominalPediatricCardiac AdultCardiac PediatricPeripheral VascularThoracic/LungInterventional Guidance:Nonvascular | Phased Transducer (only):Anatomy/ Region of Interest:Fetal – OB/GYNAbdominalPediatricAdult CephalicCardiac AdultCardiac PediatricPeripheral VascularThoracic/PleuralInterventional Guidance:Nonvascular | |
| TransducerFrequency | 1.5 – 4.5 MHz | 1.7 – 3.8 MHz | |
| Modes ofOperation | 2D / B-modeM-modeColor DopplerCombination ModesHarmonic Imaging | 2D / B-modeColor DopplerCombination ModesHarmonic Imaging | |
| PW Doppler | Not available | Not available | Not available |
| CW Doppler | Not available | Not available | Not available |
| 510(k) Track | Track 3 | Track 3 | Track 3 |
| DA and ECG Substantial Equivalence (Technological Characteristics) | |||
| DA PickupSensor andProcessing | Audio microphone + digital signalprocessingSampling Rate: 12.7 kHz | Audio microphone + digital signalprocessingSampling Rate: 4000 Hz | |
| DA Filter Modes | Heart/Midrange (50 – 600 Hz) | Diaphragm (100 – 500 Hz), Bell(20 – 200 Hz, Midrange (50 – 500Hz), Extended ((20 - 2000 Hz) | |
| DA SoundAmplification | Analog gain: 20 dB; Digital gain:user adjustable up to 25 dB | Amplifies up to 60x | |
| DA VolumeControl | Yes; 15 volume steps available | Yes; 12 volume settings | |
| DA AmbientNoise Reduction | Yes | Yes | |
| DA DirectListening | Sounds can be listened to in realtime using a digital-to-analogbinaural headset | Digital-only sound mode | |
| ECG Non-ContinuousMonitoring Leads | 3-lead, single-channel, user-supplied commercial electrodes | Single-channel, 2 stainless steelelectrodes | |
| ECG AnatomicalSites | Chest (torso) and Leg | Chest | |
| ECG Leadwiresand TrunkAssembly | Combines trunk cable and threeleadwires into a single, non-sterile,reusable assembly that forms aconduction channel fortransmitting signals from user-supplied clip-style electrodesaffixed to patient skin to theKosmos Torso (probe) | No | |
| DA and ECGVisualization | Sounds and ECG waveforms canbe visualized and recorded on theKosmos Bridge (tablet) with orwithout an internet connection | Sounds and ECG tracings can bevisualized on a Bluetooth deviceusing the Eko App. The app can beused to visualize waveforms andtracings without an internetconnection; however an internetconnection is necessary to save thedata. | |
| Feature | KOSMOS System(This 510(k) submission) | GE Vscan Extend UltrasoundSystem (K180995) | Eko Duo Model E5(K170874) |
| System Characteristics | |||
| Dimensions andWeight | Handheld tablet display unit(proprietary): 146 x 216 x 59 mm,657 gDisplay: 8"Probe: 150 x 56 x 35 mm, 260 g | Handheld tablet display unit(proprietary): 168 x 76 x 22 mm,321 gDisplay: 12.7 cm, 720 x 1280 pixels resolutionSector probe: 129 x 32 x 25 mm, 85 gDual probe: 129 x 39 x 38 mm, 120 g | Handheld Unit: 119 x 47 x 16 mmWeight: 208 g |
| Power Source | Mains and battery operated(rechargeable lithium ion battery) | Battery operated | Battery operated (rechargeablelithium ion battery) |
| Patient ContactMaterials | Probe Lens: RTV silicone 664Probe Housing: Polysulfone thermoplasticProbe Cemented Joint: RTV silicone 832ECG Leadwires: Thermoplastic urethane | Unknown (information notpublicly available); however,transducer material and otherpatient contact materials arebiocompatible | 6061 machined aluminumenclosureHigh-impact ABS thermoplastic |
| Ingress Protection(IP) Rating | Tablet: IP22Probe: IPX7 | Unit: IP33Probe: IPX7 | IP55 |
| DICOM | Yes | Yes | No |
| WirelessNetworking | Wireless networking (IEEE802.11 b/g/n/ac supported) | Wireless networking (IEEE802.11 b/g/n supported) | Wireless networking (Bluetooth4.0 low-energy) |
{6}------------------------------------------------
{7}------------------------------------------------
Non-Clinical Performance Data 11.
KOSMOS has been designed and evaluated to comply with the following applicable FDArecognized consensus standards. All verification and validation testing for KOSMOS confirms that product specifications are met and are equivalent in design and technological and performance characteristics as the predicate devices.
- ANSI AAMI ES60601-1:2005/(R)2012 ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, ● C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment -Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- ANSI AAMI IEC 60601-2-27:2011(R)2016 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment (limited set of test requirements)
- ANSI AAMI EC53:2013 ECG Trunk Cables And Patient Leadwires (limited set of test ● requirements)
- ANSI AAMI IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
- IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ANSI AAMI IEC 62366-1:2015 Medical devices Part 1: Application of usability engineering ● to medical devices
- IEC 60601-2-37 Edition 2.1 2015 Medical electrical equipment Part 2-37: Particular ● requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- NEMA UD 2-2004 (R2009) Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
- IEC 62359 Edition 2.1 2017-09 CONSOLIDATED VERSION Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related
{8}------------------------------------------------
to medical diagnostic ultrasonic fields
- ANSI AAMI IEC 62304:2006/A1:2016 Medical device software Software life cycle processes . [Including Amendment 1 (2016)]
- . ANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ANSI AAMI ISO 14971:2007/(R)2010 (Corrected 4 October 2007) Medical devices -Applications of risk management to medical devices
12. Clinical Performance Data
An assessment of clinical performance data for KOSMOS was not required to support a determination of substantial equivalence.
13. Conclusion
EchoNous, Inc. considers KOSMOS to be substantially equivalent to its predicate devices with regards to intended use, technological and performance characteristics, and safety and effectiveness.
N/A