K Number
K193497

Validate with FDA (Live)

Date Cleared
2021-01-05

(385 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Fixone Biocomposite Interference Screws are indicated for fixation of bone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.

Device Description

The Fixone Biocomposite Interference Screws are intended to be used for fixation of bone-tendon-bone or soft tissue grafts during anterior or posterior cruciate ligament reconstruction surgery. This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 6 models. It provide non-sterile (user must sterilization before use).

AI/ML Overview

The provided text describes a medical device called "Fixone Biocomposite Interference Screw" and outlines its clearance process with the FDA. It includes a 510(k) summary focusing on bench testing and biocompatibility.

Here's an analysis of the acceptance criteria and study information, based on the provided document:

1. A table of acceptance criteria and the reported device performance

Test ItemAcceptance CriteriaReported Device Performance
External surfaceASTM F2502 and USP<28> (Implied: device meets these standards)Pass
MeasurementWhen tested by vernier calipers, the tolerance should be within ±5%Pass
Tensile strengthThe average strength of 10 stands is not less than 100N.Pass
Extractable colorCompare with the matching solution: any color present is not more intense than of the appropriate matching solution.MD2014-00216 (Implied: Pass)
Insertion torqueTested with ASTM F2502-Standard specification and test methods for bioabsorbable plates and screws for internal fixation implants (Implied: meets specified performance within this standard)Pass
Fixation strength (Push-out test)Tested with ASTM F2502-Standard specification and test methods for bioabsorbable plates and screws for internal fixation implants; Tested subject and predicate device under same condition for the comparison (Implied: comparable to predicate)AJU-M20201006-01 (Implied: Pass or equivalent to predicate)
Extraction test pHThe difference should be 1.5 and less.Pass
Potassium permanganate reducing substancesThe difference of the consumption of potassium permanganate should be 2.0 mL and less.Pass
Residue after evaporationRecord the weight of the residue should be 1.0mg and less.Pass
Heavy metalsAny brown color produced within 10 minutes in the tube containing the extract of the prepared sample does not exceed that in the tube containing the standard lead solution.MD2014-00132 (Implied: Pass)
UV spectrum (250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 and less.Pass
Property (Visual inspection)When observing it with the naked eye, test solution should be clear and have no foreign particles.Pass
Performance Testing of Fixone® Biocomposite Anchor (Pull-out test)Pull-out test by immersion time of saline solution to evaluate two bioabsorbable suture anchors. [Absorption time (soaking time) : 4 weeks, 6 weeks, 12 weeks and 26 weeks] (Implied: comparable to predicate, within acceptable limits)Ajum20170428-01 (Implied: Pass or equivalent to predicate)
ASTM F1839-08 (Polyurethane Foam for Orthopaedic Devices)(Implied: Compliance with this standard for material properties)Pass
Biocompatibility Tests:
CytotoxicityISO 10993-5 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicityPass
Acute systemic toxicity testISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
Intracutaneous (intradermal) reactivity testISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
Maximization test for delayed hypersensitivityISO 10993-10 Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
Bacterial reversion mutation testISO 10993-3 Genotoxicity test OECE 471, Bacterial reverse mutation testPass
Mammalian erythrocyte micronucleus testISO 10993-3 Genotoxicity test OECE 471, Bacterial reverse mutation testPass
Implantation testISO 10993-6 Tests for local effects after implantation, Annex D test methods for implantation in bonePass
Bioabsorbable screws testASTM F2502 Standard specification and test methods for bioabsorbable plates and screws for internal fixation implantsPass
Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11- Test for systemic toxicityPass

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily details bench testing for mechanical, physical, and biocompatibility properties. It does not mention a clinical test set or human subject data.

  • Sample Size (for bench tests): The "Tensile strength" test mentions an "average strength of 10 stands," indicating a sample size of at least 10 for that specific test. Other tests, such as those related to "Extractable color" and "Residue after evaporation," refer to "a quantity of suture, equivalent to not less than 250mg" and standard protocols, but specific sample counts for each bench test are generally not provided.
  • Data Provenance: The bench tests were conducted by AJU Pharm Co., Ltd. in Korea (per the applicant information). These are laboratory tests, not clinical data from patients. They are inherently retrospective in the sense that they are conducted on manufactured devices in a controlled environment.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This device approval is based on bench testing and biocompatibility assessments, not on studies requiring human experts to establish "ground truth" for diagnostic or clinical performance. The "ground truth" for the tests performed are the defined standards and specifications (e.g., ASTM, ISO standards, quantified limits for chemical properties).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As noted above, this is not a diagnostic or clinical performance study that would involve expert adjudication of results. The results are quantitative measurements against predefined criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (an interference screw for ACL/PCL reconstruction) and not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is established by:

  • Industry Standards: ASTM F2502, ASTM F1839-08, various ISO 10993 series for biocompatibility.
  • Pharmacopoeial Standards: USP<28> (for external surface).
  • Defined Quantitative Limits: Specific values for pH difference, potassium permanganate consumption, residue after evaporation, maximum absorbance, and tensile strength (e.g., "not less than 100N").
  • Comparison to Predicate Device: The fixation strength (push-out test) directly compares the subject device to the predicate device under the same conditions. This implies the predicate device's established performance serves as a comparative ground truth.

8. The sample size for the training set

Not applicable. This is not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set.

{0}------------------------------------------------

January 5, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

AJU Pharm Co., Ltd. % Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K193497

Trade/Device Name: Fixone Biocomposite Interference Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: December 1, 2020 Received: December 8, 2020

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193497

Device Name Fixone Biocomposite Interference Screw

Indications for Use (Describe)

Fixone Biocomposite Interference Screws are indicated for fixation of bone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

BR

PSC Publishing Services (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

[as required by 807.92(c)]

1. Applicant

    1. Company : AJU Pharm Co.,Ltd.
  • Address : A-207, 697, Pangyo-ro, Seongnam-si, Gyeonggi-do, Korea 2)
  • Tel : 82-31-765-4420 3)
  • Fax : 82-31-602-7818 4)
    1. Prepared date : Dec. 01, 2019
  • ୧) Contact person : Peter Chung, 412-512-8802
    1. Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
    1. Submission date : Nov. 30, 2020
  • ರು) Submission type : Traditional
    1. Prior related submissions : K192484, B-type anchor

2. Device Information

  • Trade name : Fixone Biocomposite Interference Screw 1)
    1. Common name : Screw, Fixation, Bone
  • Regulation name : Smooth or threaded metallic bone fixation fastener 3)
    1. Product code : HWC
    1. Regulation number : 888.3040
    1. Class of device : Class II
    1. Panel : Orthopedic

3. The legally marketed device to which we are claiming equivalence

K032224, Smith&nephew, Inc. / Smith & Nephew BIORCI HA Screw

4. Device description

The Fixone Biocomposite Interference Screws are intended to be used for fixation of bone-tendon-bone or soft tissue grafts during anterior or posterior cruciate ligament reconstruction surgery.

This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 6 models. It provide non-sterile (user must sterilization before use).

Product nameInstrument / 6 model codes including Fixone.I.I-T8090
Intended useTo make a thread and insert a screw in the hole
ManufacturerAJU Pharm Co.,Ltd. / Korea
CharacteristicThis device is provided as Non-sterile and Reusable device.This device is provided non-sterile. This device is user-sterilized.
Sterilization methodAutoclave / 134°C/ 20 min
MaterialGrip : Aluminum (Anodized) / Shaft : Stainless Steel (SUS304)
Product codeLXH – Orthopedic manual surgical instrument (Class I)
Pilot Hole4.5 ~ 5.5 mm

5. Intended Use :

Fixone Biocomposite Interference Screws are indicated for fixation of bone-tendon-bone or soft tissue grafts during Anterior and/or Posterior Cruciate Ligament (ACL/PCL) reconstruction.

6. Performance data:

{4}------------------------------------------------

  • Bench test were performed. Bench testing included biocompatibility, mechanical testing, and sterility 1) testing including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
Test itemRequirementsResults
External surfaceASTM F2502 and USP<28>
MeasurementWhen tested by vernier calipers, the tolerance should be within ±5%Pass
Tensile strengthThe average strength of 10 stands is not less than 100N.
Extractable colorWeigh a quantity of suture, equivalent to not less than 250mg, and place in aconical flask containing 1.0mL of water for each 10mg of the sample. Close theflaskm and allow it to stand at (37±0.5)°C for 24 hours. Coll, decant the waterfrom the suture, and compare it with the matching solution : any color present isnot more intense than of the appropriate matching solution.MD2014-00216
Insertion torqueTested with ASTM F2502-Standard specification and test methods forbioabsorbable plates and screws for internal fixation implantsPass
Fixation strength(Push-out test)Tested with ASTM F2502-Standard specification and test methods forbioabsorbable plates and screws for internal fixation implantsTested subject and predicate device under same condition for the comparisonAJU-M20201006-01
Extraction test
pHThe difference should be 1.5 and less.
Potassium permanganatereducing substancesThe difference of the consumption of potassium permanganateshould be 2.0 mL and less.
Residue after evaporationRecord the weight of the residue should be 1.0mg and less.Pass
Heavy metalsAny brown color produced within 10 minutes in the tube containingthe extract of the prepared sample does not exceed that in the tubecontaining the standard lead solutionMD2014-00132
UV spectrum(250nm~350nm)Maximum absorbance between 250 to 350 nm should be 0.1 andless.
PropertyWhen observing it with the naked eye, test solution should be clearand have no forign particles.
Performance Testing of Fixone ® Biocomposite Anchor (Pull-out test)
Pull-out test by immersion time of saline solution to evaluate two bioabsorbable suture anchors.[Absorption time (soaking time) : 4 weeks, 6 weeks, 12 weeks and 26 weeksAjum20170428-01
ASTM F1839-08, Standard specification for RGID Polyurethane Foam for Use as a Standard Materialfor Testing Orthopaedic Devices and Instruments

2) Biocompatibility

Screw

#Test itemTest method / Test criteriaTestRep No.
----------------------------------------------------------

{5}------------------------------------------------

Result
1CytotoxicityISO 10993-5 Biological evaluation of medical devices - Part5:Tests for in vitro cytotoxicityPassMD2014-00216
2Acute systemic toxicitytestISO 10993-11 Biological evaluation of medical devices -Part 11: Tests for systemic toxicityPassMD2014-00132
3Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPassMD2014-00132
4Intracutaneous(intradermal) reactivity testISO 10993-10 Biological evaluation of medical devices - Part 10:Tests for irritation and skin sensitizationPassMD2014-00132
5Maximization test fordelayedhypersensicivityISO 10993-10 Test for irritation and skin sensitization,Maximization test for delayed hypersensitivityPassMD2014-00216
6Bacterial reversmutation testISO 10993-3 Genotoxicity test OECE 471, Bacterial reversemutation testPassMD2014-00133
7Mammalianerythrocytemicronucleus testISO 10993-3 Genotoxicity test OECE 471, Bacterial reversemutation testPassMD2014-00133
8Implantation testISO 10993-6 Tests for local effects after implantation, Annex Dtest methods for implantation in bonePassMD2014-00104
9Bioabsorbable screwstestASTM F2502 Standard specification and test methods forbioabsorbable plates and screws for internal fixation implantsPassMD2014-00104
10Subchronic toxicity testISO 10993-11 Biological Evaluation of Medical Devices Part 11-Test for systemic toxicityPassD00021847001-800565.1

The performance tests demonstrated that Fixone Biocomposite Interference Screw performs in a substantially equivalent manner to the predicate device.

7. Predicate device comparison table

The Fixone Biocomposite Interference Screw has the similar device characteristics as the predicated device, the Smith & Nephew BIORCI HA Screw; intended use, material, composition, design and use concept are similar.

The Fixone Biocomposite Interference Screw has been subjected to extensive safety, performance, and product validations prior to release. Safety tests have been performed to ensure the devices comply to applicable industry and US regulations.

An extensive review of pertaining to the safety and biocompatibility of Fixone Biocomposite Interference Screw has been conducted.

9. Conclusion

The device is investigated for function to compare the operation of function between Fixone Biocomposite Interference Screw and predicate devices.

Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.

Therefore, it is concluded that Fixone Biocomposite Interference Screw is substantially equivalent to the legally marketed predicate device.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.