K Number
K193487

Validate with FDA (Live)

Device Name
Agilon Strip
Manufacturer
Date Cleared
2020-03-30

(104 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Agilon Strip is indicated for use in voids or gaps of the skeletal system. i.e., the extremities, pelvis, and posteroateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities and pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.

Device Description

Agilon Strip is a flexible, resorbable, wicking, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules bound by type I collagen fibers to facilitate shaping and containment of the implant. The osteoSPAN granules in Agilon Strip are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate, with a thin laver of calcium phosphate throughout its entire porosity.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Agilon Strip," a resorbable calcium salt bone void filler. The submission aims to demonstrate substantial equivalence to a predicate device, Morpheus (K142828).

However, the provided text does not contain any information about acceptance criteria or a study that proves the device meets those criteria in the context of device performance metrics like accuracy, sensitivity, or specificity. Instead, the document focuses on demonstrating substantial equivalence through:

  • Comparison of technological characteristics: Comparing the composition and properties of Agilon Strip to the predicate.
  • Bench testing: Verifying that the addition of collagen does not affect granule chemistry, crystallinity, porosity, pore diameter, and pore interconnectivity, along with sterilization and shelf-life studies.
  • Biocompatibility testing: Following ISO 10993-1 for biological effects such as cytotoxicity, sensitization, systemic toxicity, genotoxicity, and implantation.
  • In vivo testing: Using critically sized defect models and a clinically relevant spinal fusion model to assess bone formation, resorption, remodeling, and fusion rates compared to the predicate and controls.

The "Performance Data" section discusses various tests and their outcomes but does not specify quantitative acceptance criteria for these tests (e.g., "fusion rate must be X% or higher"). The conclusion is based on the device being "substantially equivalent" to the predicate, implying that its performance in these tests was comparable and acceptable, but no explicit thresholds are provided.

Therefore, I cannot populate the requested table and details as the necessary information (explicit acceptance criteria, detailed quantitative performance metrics against those criteria, data provenance, expert qualifications for ground truth, sample sizes for test/training sets with detailed ground truth establishment methods, MRMC studies, or standalone algorithm performance) is not present in the provided text.

The document is a regulatory submission for a bone void filler, not an AI/ML device that would typically have performance metrics like sensitivity, specificity, or accuracy, which necessitate the kind of detailed study design outlined in the prompt (e.g., multi-reader multi-case studies, expert adjudication for ground truth).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is the Department of Health & Human Services logo. The text is in blue, with "FDA" in a larger font size than the rest of the text.

March 30, 2020

Biogennix, LLC % Elaine Duncan President Paladin Medical, Inc. PO Box 560 Stillwater, Minnesota 55082

Re: K193487

Trade/Device Name: Agilon Strip Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: March 3, 2020 Received: March 4, 2020

Dear Ms. Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jesse Muir, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193487

Device Name Agilon Strip

Indications for Use (Describe)

Agilon Strip is indicated for use in voids or gaps of the skeletal system. i.e., the extremities, pelvis, and posteroateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities and pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

I. SUBMITTER

Submitted on behalf of:

Company Name:BIOGENNIX, LLC
Address:1641 McGaw Ave.Irvine, CA 92614
Telephone:949-253-0094
Fax:949-266-5800
by:Elaine Duncan, M.S.M.E., RACPresident, Paladin Medical, IncPO Box 560Stillwater, MN 55082
Telephone:715-549-6035
Fax:715-549-5380
Contact Person:Elaine Duncan
Date Prepared:December 13, 2019

II. SUBJECT DEVICE

Trade Name:Agilon Strip
Common Name(s):Bone void filler, Bone graft substitute, Bone graft extender
Regulation Number:21 CFR 888.3045
Regulation Name:Resorbable calcium salt bone void-filler device
Product Code:MQV
Regulatory Class:II

III. PREDICATE DEVICE

The contents of this submission have demonstrated that Agilon Strip is substantially equivalent to its primary predicate Morpheus (K142828) when used as a bone graft substitute in the extremities and pelvis, and as a bone graft extender in the posterolateral spine.

IV. DEVICE DESCRIPTION

Agilon Strip is a flexible, resorbable, wicking, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules bound by type I collagen fibers to facilitate shaping and containment of the implant.

{4}------------------------------------------------

The osteoSPAN granules in Agilon Strip are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate, with a thin laver of calcium phosphate throughout its entire porosity.

V. INDICATIONS FOR USE

Agilon Strip is indicated for use in voids or gaps of the skeletal system, i.e., the extremities, pelvis, and posterolateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities or pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Product510(k)NumberGranulesCompositionPolymerBinderCollagenOsteo-conductive
MorpheusK132377K142828Calcium Carbonate/Calcium PhosphatecompositeYesNoYes
Agilon StripK193487IdenticalNoYesYes

The function, intended use and technological characteristics of the subject device are substantially equivalent to the predicate device cleared under 510(k) premarket notification K142828.

VII. PERFORMANCE DATA

Biogennix followed the "Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device: Guidance for Industry and FDA, June 2, 2003, as well as the company's design controls and risk analysis procedures to ensure that Agilon Strip is safe and effective for use.

Bench testing of Agilon Strip was performed to verify that the addition of the type 1 collagen does not affect the granule chemistry, crystallinity, porosity, pore diameter, and pore interconnectivity. Sterilization validation and shelf-life aging studies demonstrate the suitability of the packaging system.

A full evaluation of the biocompatibility of Agilon Strip was conducted in accordance with "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." As a tissue/bone

{5}------------------------------------------------

permanent implant device, Agilon Strip satisfied a battery of tests assessing the following biological effects:

  • Cytotoxicity
  • Sensitization ●
  • Intracutaneous Reactivity ●
  • Acute Systemic Toxicity
  • Material-Mediated Pyrogenicity ●
  • Bacterial Endotoxins
  • Subchronic toxicity
  • Genotoxicity ●
  • . Implantation

In vivo testing of Agilon Strip was conducted using a validated, critically sized defect model. Device performance was assessed at multiple time points against the primary predicate and negative controls using histology, histomorphometry, x-ray, and micro-CT analyses. The critical nature of the model was confirmed by the negative controls. No adverse reactions were noted at the implant site or in distant organs; new bone formation, bone remodeling, and implant resorption for the test materials were confirmed with time.

Agilon Strip was also validated in a clinically relevant, single-level posterolateral spinal fusion model. Device performance was evaluated at multiple time points against the primary predicate and positive control using mechanical, histology, histomorphometry, x-ray, and micro-CT analyses. Fusion rates were the same between all treatment groups at each time point and consistent with published literature. No adverse events or device-related failures were noted at the implantation site or distant organs for any implant group. No abnormal inflammation, redness, swelling, or discoloration was noted at any time point. New bone formation on the host transverse bone processes, as well as evidence of resorption and remodeling of the graft materials, were confirmed with time.

Based on the endpoints and results of these studies, Agilon Strip was concluded to be substantially equivalent to the primary predicate osteoSPAN Morpheus.

VIII. CONCLUSIONS

The non-clinical data presented in this submission demonstrates that Agilon Strip is substantially equivalent to the predicate device osteoSPAN Morpheus.

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.