(185 days)
The physIQ Heart Rhythm and Respiration Module (Version 3.0) is intended for use by a physician or other qualified medical professionals for the calculation of heart rate variability, the detection of atrial fibrillation and determination of respiration rate using ambulatory ECG and triaxial accelerometer data. The physIQ Heart Rhythm and Respiration Module is intended to be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display).
The physIQ Heart Rhythm and Respiration Module may only be used with FDA-cleared, chest-worn biosensors using "wet electrode" technology that capture single-lead digital ECG data at 125Hz or higher and integrated triaxial accelerometer data at 15Hz or higher and that are recorded in a compatible format for analysis.
The physIQ Heart Rhythm and Respiration Module is for use in adult patients in subacute clinical settings for remote patient monitoring. The physIQ Heart Rhythm and Respiration Module is not for use in patients requiring life-supporting or life-sustaining systems or as ECG or respiration alarm devices.
The physIQ Heart Rhythm and Respiration Module (Version 3.0) is a computerized all-software callable function library in the Python programming language that is designed for calculating heart rate and heart rate variability and for detecting atrial fibrillation and determining respiration rate determined by automated analysis of any single electrocardiogram (ECG) channel collected by commercially-available ECG biosensor devices with triaxial accelerometers. The physlQ Heart Rhythm and Respiration Module (3.0) will be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display) that makes calls into the product, most typically via a Python middleware script. The "middleware" accesses the source ECG and triaxial accelerometer data from a customer's data collection system, most likely via its own application programming interface (API), and makes calls to the physIQ Heart Rhythm and Respiration Module (3.0) to input ECG and triaxial accelerometer data for processing into the vital sign outputs of the product. These outputs are returned to the middleware, which may insert these results into a downstream monitoring system for clinical use.
The provided text describes the physIQ Heart Rhythm and Respiration Module (Version 3.0) and its substantial equivalence to a predicate device. However, it does not contain the specific acceptance criteria and detailed study information requested in the prompt.
The document states that "performance testing demonstrates that the physIQ Heart Rhythm and Respiration Module (3.0) meets its intended use and any differences in technological characteristics between the physIQ Heart Rhythm and Respiration Module (3.0) and the predicate device do not raise any new issues and is substantially equivalent to the predicate device."
It also mentions that "Performance testing following guidelines of ANSI/AAMI EC572012: Testing and Reporting Performance Results of Cardiac Rhythm and ST segment Medsurement Algorithms was applied to heart rate variability, and atrial fibrillation algorithms in a previous Traditional 510(k) submission for the physIQ Heart Rhythm and Respiration Module (K183322) predicate device." and that "In this submission, performance of all algorithms including Heartbeat Detector, Heart Rate, Heart Rate Variability, Atrial Fibrillation, and Respiration Rate have been repeated and evaluated using the modified technology. The respiration rate algorithm met its corresponding acceptance criteria and performed comparably to the predicate device."
Based on the provided text, I cannot extract the following information:
- A table of acceptance criteria and the reported device performance. While it mentions that algorithms met acceptance criteria, the specific criteria and reported values are not present.
- Sample size used for the test set and the data provenance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
- Adjudication method for the test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done. (It implies standalone testing as it refers to algorithm performance, but doesn't explicitly state it or provide details).
- The type of ground truth used.
- The sample size for the training set.
- How the ground truth for the training set was established.
Therefore, I can only provide the information that is explicitly stated or strongly implied from the text:
1. A table of acceptance criteria and the reported device performance
- Not explicitly provided. The document states that the respiration rate algorithm met its corresponding acceptance criteria and performed comparably to the predicate device. However, the specific numerical criteria for heart rate variability, atrial fibrillation, and respiration rate, and the reported performance metrics, are not included in this document.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified.
- Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not specified.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not mentioned. The document focuses on the algorithm's performance and substantial equivalence, not human-AI collaboration.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Implied yes. The document repeatedly refers to the "performance of all algorithms" and the "physIQ Heart Rhythm and Respiration Module (3.0)" in isolation, suggesting standalone algorithm testing was performed to assess its functionality and accuracy.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not specified.
8. The sample size for the training set
- Not specified.
9. How the ground truth for the training set was established
- Not specified.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
June 11, 2020
physIQ, Inc George Hides Vice President, Regulatory and Clinical Affairs 300 E. 5th Avenue, Suite 105 Naperville, Illinois 60563
Re: K193415
Trade/Device Name: physIO Heart Rhythm and Respiration Module Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: May 8, 2020 Received: May 11, 2020
Dear George Hides:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193415
Device Name
physIQ Heart Rhythm and Respiration Module (3.0)
Indications for Use (Describe)
The physIQ Heart Rhythm and Respiration Module (Version 3.0) is intended for use by a physician or other qualified medical professionals for the calculation of heart rate variability, the detection of atrial fibrillation and determination of respiration rate using ambulatory ECG and triaxial accelerometer data. The physIQ Heart Rhythm and Respiration Module is intended to be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display).
The physIQ Heart Rhythm and Respiration Module may only be used with FDA-cleared, chest-worn biosensors using "wet electrode" technology that capture single-lead digital ECG data at 125Hz or higher and integrated triaxial accelerometer data at 15Hz or higher and that are recorded in a compatible format for analysis.
The physIQ Heart Rhythm and Respiration Module is for use in adult patients in subacute clinical settings for remote patient monitoring. The physIQ Heart Rhythm and Respiration Module is not for use in patients requiring life-supporting or life-sustaining systems or as ECG or respiration alarm devices.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ----------------------------------------------------------------------------------------------------------------------------- | ---------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Notification K193415
GENERAL INFORMATION
Applicant:
physIQ Inc. 300 E. 5th Avenue Suite 105 Naperville IL 60563 USA Phone: (800) 561-7902
Contact Person:
George Allen Hides Vice President, Regulatory and Clinical Affairs physIQ Inc. 300 E. 5th Avenue Suite 105 Naperville IL 60563 USA Email: george.hides@physiq.com Phone: (312) 654-1010
Trade/Proprietary Name:
physIQ Heart Rhythm and Respiration Module (Version 3.0)
Generic/Common Name:
Electrocardiograph
Classification:
Class II, 21 CFR§870.2340 (Electrocardiograph)
Product Code:
DbS
Predicate Device"
physIQ Heart Rhythm and Respiration Module, Version 2.0 (K183322) physIQ Inc.
Indications for Use:
The physIQ Heart Rhythm and Respiration Module (Version 3.0) is intended for use by a physician or other qualified medical professionals for the calculation of heart rate variability, the detection of atrial fibrillation and determination of respiration rate using ambulatory ECG and triaxial accelerometer data. The physIQ Heart Rhythm and Respiration Module is intended to be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display).
{4}------------------------------------------------
The physIQ Heart Rhythm and Respiration Module may only be used with FDA-cleared, chest-worn biosensors using "wet electrode" technology that capture single-lead digital ECG data at 125Hz or higher and integrated triaxial accelerometer data at 15Hz or higher and that are recorded in a compatible format for analysis.
The physIQ Heart Rhythm and Respiration Module is for use in adult patients in subacute clinical and nonclinical settings for remote patient monitoring. The physIQ Heart Rhythm and Respiration Module is not for use in patients requiring life-supporting or life-sustaining systems or as ECG or respiration alarm devices.
Product Description:
The physIQ Heart Rhythm and Respiration Module (Version 3.0) is a computerized all-software callable function library in the Python programming language that is designed for calculating heart rate and heart rate variability and for detecting atrial fibrillation and determining respiration rate determined by automated analysis of any single electrocardiogram (ECG) channel collected by commercially-available ECG biosensor devices with triaxial accelerometers. The physlQ Heart Rhythm and Respiration Module (3.0) will be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display) that makes calls into the product, most typically via a Python middleware script. The "middleware" accesses the source ECG and triaxial accelerometer data from a customer's data collection system, most likely via its own application programming interface (API), and makes calls to the physIQ Heart Rhythm and Respiration Module (3.0) to input ECG and triaxial accelerometer data for processing into the vital sign outputs of the product. These outputs are returned to the middleware, which may insert these results into a downstream monitoring system for clinical use.
Performance Testing:
The physIQ Heart Rhythm and Respiration Module (Version 3.0) contains a collection of algorithms intended to be applied to ECG data collected by commercially-available ECG biosensor devices with triaxial accelerometers in an ambulatory setting. The collection consists of Heartbeat Detector, Heart Rate, Heart Rate Variability, Atrial Fibrillation and Respiration Rate algorithms. Performance testing following guidelines of ANSI/AAMI EC572012: Testing and Reporting Performance Results of Cardiac Rhythm and ST segment Medsurement Algorithms was applied to heart rate variability, and atrial fibrillation algorithms in a previous Traditional 510(k) submission for the physIQ Heart Rhythm and Respiration Module (K183322) predicate device. There are no FDA-recognized consensus standards to assess the performance of respiration rate algorithms. In the predicate submission for the physIQ Heart Rhythm Module (K183322), performance validation was provided for the determination of respiration rate. In this submission, performance of all algorithms including Heartbeat Detector, Heart Rate, Heart Rate Variability, Atrial Fibrillation, and Respiration Rate have been repeated and evaluated using the modified technology. The respiration rate algorithm met its corresponding acceptance criteria and performed comparably to the predicate device.
Substantial Equivalence:
The physIQ Heart Rhythm and Respiration Module (3.0) has the same intended use to the predicate device, the physlQ Heart Rhythm and Respiration Module (2.0). Both calculate heart rate variability, detects atrial fibrillation and determine respiration rate from a single-lead ECG and triaxial accelerometry. The patient population for both the physlQ Heart Rhythm and Respiration Module (3.0) and the predicate device includes subacute adults who do not require life-supporting or life-sustaining systems or device alarms. Of note, the intended uses of the physIQ Heart Rhythm and Respiration Module (3.0) and the predicate device are to supplement standard of care and not to replace or substitute for
{5}------------------------------------------------
K193415
routine vital signs monitoring. Both the physIQ Heart Rhythm and Respiration Module (3.0) and the predicate have similar technological characteristics as both are all software medical devices and require input of time-series ECG from commercially available devices in a format acceptable for signal processing and algorithm function as well as triaxial accelerometry. Any differences in technological characteristics have been analyzed and addressed through performance validation testing and hazard analysis. Performance testing demonstrates that the physIQ Heart Rhythm and Respiration Module (3.0) meets its intended use and any differences in technological characteristics between the physlQ Heart Rhythm and Respiration Module (3.0) and the predicate device do not raise any new issues and is substantially equivalent to the predicate device.
The following 510(k) Summary is identical in terms of the characterization of device functionality, with the exception of the updated software version number.
| Device Functionality | physIQ Heart Rhythm andRespiration Module(Version 2.0) | physIQ Heart Rhythm andRespiration Module(Version 3.0) |
|---|---|---|
| Comparison | 510(k)Predicate Device | 510(k)Candidate Device |
| Manufacturer | physIQ Inc. | physIQ Inc. |
| 510(k) Number | K183322 | K193415 |
| Classification | Class II,21 CFR §870.2340 | Class II,21 CFR §870.2340 |
| Product Code | DPS | DPS |
| Indications for Use | The physIQ Heart Rhythm andRespiration Module (Version 2.0) is intended for use by a physician or other qualified medical professionals for the calculation of heart rate and heart rate variability, the detection of atrial fibrillation and determination of respiration rate using ambulatory ECG and triaxial accelerometer data. The physIQ Heart Rhythm and Respiration Module supports receiving and analyzing single-lead ECG signals recorded in a compatible format from FDA-cleared ECG biosensor devices using "wet" electrode technology and triaxial accelerometers when assessment of rhythm and respiration rate is desired. The physIQ Heart Rhythm and Respiration Module is for use in adult patients in subacute clinical and non-clinical settings for | The physIQ Heart Rhythm andRespiration Module (Version 3.0) is intended for use by a physician or other qualified medical professionals for the calculation of heart rate and heart rate variability, the detection of atrial fibrillation and determination of respiration rate using ambulatory ECG and triaxial accelerometer data. The physIQ Heart Rhythm and Respiration Module is intended to be integrated by the customer organization into an end-to-end system (biosensor data collection to clinician display).The physIQ Heart Rhythm and Respiration may only be used with FDA-cleared, chest-worn biosensors using "wet electrode" technology that capture single-lead digital ECG data at 125Hz or |
| Device Functionality | physIQ Heart Rhythm andRespiration Module(Version 2.0) | physIQ Heart Rhythm andRespiration Module(Version 3.0) |
| remote patient monitoring. ThephysIQ Heart Rhythm andRespiration Module is not for usein patients requiring life-supporting or life-sustainingsystems or as ECG or respirationalarm devices. | higher and integrated triaxialaccelerometer data at 15Hz orhigher and that are recorded in acompatible format for analysis.The physIQ Heart Rhythm andRespiration Module is for use inadult patients in subacute clinicaland non-clinical settings forremote patient monitoring. ThephysIQ Heart Rhythm andRespiration Module is not for usein patients requiring life-supporting or life-sustainingsystems or as ECG or respirationalarm devices. | |
| Level of Concern | Moderate | Moderate |
| Components | Software only | Software only |
| Interface | Callable application programminginterface (API) | Callable application programminginterface (API) |
| Biosensor requirements | --- | FDA-cleared, chest-worn, "wetelectrode" technologySingle-lead digital ECG data at125Hz or higherTriaxial accelerometer data at15Hz or higher |
| Display | Compatible medical device datasystem (MDDS) display | Compatible medical device datasystem (MDDS) display |
| QRS detection | YES | YES |
| Heart rate non-paced adult | YES | YES |
| Heart rate variability | YES (deterministic based on R-to-R interval derived from QRSdetection) | YES (deterministic based on R-to-R interval derived from QRSdetection) |
| Atrial fibrillation detection | YES | YES |
| Respiration rate | YES | YES |
| ECG morphological analysis | NO (other than QRS location andbeat-to-beat analyses, no ECGmorphological analyses areperformed) | NO (other than QRS location andbeat-to-beat analyses, no ECGmorphological analyses areperformed) |
| Arrhythmia classifications(other than atrialfibrillation) | NO | NO |
| Patient populations | Adult | Adult |
| Device Functionality | physIQ Heart Rhythm andRespiration Module(Version 2.0) | physIQ Heart Rhythm andRespiration Module(Version 3.0) |
| Clinical setting | Subacute (non-life-supporting orlife-threatening systems required) | Subacute (non-life-supporting orlife-threatening systems required) |
| Alarm / Trigger | NO | NO |
{6}------------------------------------------------
{7}------------------------------------------------
Conclusion:
The physIQ Heart Rhythm and Respiration Module (3.0) has the same intended use and patient population and similar technological characteristics as those of the predicate device, the physIQ Heart Rhythm and Respiration Module (2.0). Differences in technological characteristics have been analyzed and addressed through performance validation testing which demonstrated that the physlQ Heart Rhythm and Respiration Module (3.0) meets it intended use and that any differences between the physlQ Heart Rhythm and Respiration Module (3.0) and the predicate devices do not raise any new issues.
Summary:
Based on the information provided and the testing conducted, the physIQ Heart Rhythm and Respiration Module (3.0) is substantially equivalent to the predicate device.
Date: 9 June 2020
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).