K Number
K193412

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-02-24

(77 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NEXXT MATRIXX® corpectomy devices are indicated for use in the cervical spine (C2-T1) and thoracolumbar spine (T1-L5) in skeletally mature patients to replace a diseased, collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. The NEXXT MATRIXX® corpectomy devices are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. The NEXXT MATRIXX® corpectomy devices are intended for use with autograft or allogenic bone graft comprised of cancellous and/ or corticocancellous bone graft, as an adjunct to fusion and with FDA-cleared supplemental internal fixation.

Device Description

The NEXXT MATRIXX® System is a collection of additively manufactured spacers for cervical, lumbar/lumbosacral and thoracolumbar implantation. The basic shape of these implants is a structural column to provide surgical stabilization of the spine. Each device comprises an external structural frame having a roughened surface (~7μm). The intervening geometric lattices have pores 300-700µm. The inferior/superior aspects of the NEXXT MATRIXX® open devices incorporate a large vertical cavity which can be packed with bone graft material. The inferior/superior aspects of the NEXXT MATRIXX® solid devices are closed and do not permit the packing of bone graft within the implant. The solid devices are only to be used for partial vertebral body replacement. The open and solid devices are available in an assortment of height, length, width and lordotic angulation combinations to accommodate the individual anatomic and clinical circumstances of each patient.

AI/ML Overview

This looks like a 510(k) summary for a medical device called the NEXXT MATRIXX® System, specifically for its corpectomy devices. Unfortunately, the provided document does not contain acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of AI or algorithm performance.

The "Performance Data" section explicitly states: "The modified corpectomy device was evaluated via mechanical testing including included static and dynamic compression and static and dynamic torsion (ASTM F2077), subsidence (ASTM F2267) and expulsion. The results demonstrated the performance of the modified corpectomy is substantially equivalent to the predicate."

This describes physical and mechanical performance testing of the device itself (an implant used in spinal surgery), not the performance of an AI or algorithm. Therefore, I cannot extract the information requested regarding:

  1. A table of acceptance criteria and reported device performance (for AI/algorithm)
  2. Sample size and data provenance for a test set (for AI/algorithm)
  3. Number and qualifications of experts for ground truth (for AI/algorithm)
  4. Adjudication method (for AI/algorithm)
  5. MRMC comparative effectiveness study (for AI/algorithm)
  6. Standalone performance (for AI/algorithm)
  7. Type of ground truth used (for AI/algorithm)
  8. Sample size for the training set (for AI/algorithm)
  9. How ground truth for the training set was established (for AI/algorithm)

The document is about the physical testing and substantial equivalence of a spinal implant device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 24, 2020

Nexxt Spine LLC % Karen E. Warden, PhD President BackRoads Consulting Inc. PO Box 566 Chesterland, Ohio 44026

Re: K193412

Trade/Device Name: NEXXT MATRIXX® System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: PLR, MQP Dated: December 23, 2019 Received: December 26, 2019

Dear Dr. Warden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter. PhD Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K193412

Device Name NEXXT MATRIXX® System

Indications for Use (Describe)

The NEXXT MATRIXX® corpectomy devices are indicated for use in the cervical spine (C2-T1) and thoracolumbar spine (T1-L5) in skeletally mature patients to replace a diseased, collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. The NEXXT MATRIXX® corpectomy devices are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. The NEXXT MATRIXX® corpectomy devices are intended for use with autograft or allogenic bone graft comprised of cancellous and/ or corticocancellous bone graft, as an adjunct to fusion and with FDA-cleared supplemental internal fixation.

Type of Use (Select one or both, as applicable) ☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PRAStaff@fda.hhs.gov

{3}------------------------------------------------

Section 7 – 510(k) Summary

Date:6 December 2019
Sponsor:Nexxt Spine, LLC
14425 Bergen Blvd, Suite B
Noblesville, IN 46060
Office: 317.436.7801
Fax: 317.245.2518
Sponsor Contact:Andy Elsbury, President
510(k) Contact:Karen E. Warden, PhD
BackRoads Consulting Inc.
PO Box 566
Chesterland, OH 44026
Office: 440.729.8457
Proposed Trade Name:NEXXT MATRIXX® System
Common Name:Corpectomy device
Device Classification:Class II
Regulation Name,Regulation Number,Product Codes:Spinal vertebral body replacement device - Cervical, 888.3060, PLRSpinal vertebral body replacement device, 888.3060, MQP
Submission Purpose:The subject 510(k) adds a vertebral body replacement (corpectomy) deviceto the NEXXT MATRIXX® System.
Device Description:The NEXXT MATRIXX® System is a collection of additively manufacturedspacers for cervical, lumbar/lumbosacral and thoracolumbar implantation.The basic shape of these implants is a structural column to provide surgicalstabilization of the spine. Each device comprises an external structuralframe having a roughened surface (~7μm). The intervening geometriclattices have pores 300-700µm.
The inferior/superior aspects of the NEXXT MATRIXX® open devicesincorporate a large vertical cavity which can be packed with bone graftmaterial. The inferior/superior aspects of the NEXXT MATRIXX® soliddevices are closed and do not permit the packing of bone graft within theimplant. The solid devices are only to be used for partial vertebral bodyreplacement. The open and solid devices are available in an assortment ofheight, length, width and lordotic angulation combinations to accommodatethe individual anatomic and clinical circumstances of each patient.
Indications for Use:The NEXXT MATRIXX® corpectomy devices are indicated for use in thecervical spine (C2-T1) and thoracolumbar spine (T1-L5) in skeletally maturepatients to replace a diseased, collapsed, damaged, or unstable vertebralbody due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstructionfollowing corpectomy performed to achieve decompression of the spinalcord and neural tissues in degenerative disorders. The NEXXT MATRIXX®corpectomy devices are also intended to restore the integrity of the spinalcolumn even in the absence of fusion for a limited time period in patientswith advanced stage tumors involving the cervical, thoracic, and lumbarspine in whom life expectancy is of insufficient duration to permitachievement of fusion, with bone graft used at the surgeon's discretion. TheNEXXT MATRIXX® corpectomy devices are intended for use with autograftor allogenic bone graft comprised of cancellous and/or corticocancellousbone graft, as an adjunct to fusion and with FDA-cleared supplementalinternal fixation.
Materials:NEXXT MATRIXX® implants are manufactured from Ti-6Al-4V ELI titaniumalloy (ASTM F3001).
Primary Predicate:NEXXT MATRIXX® System (Nexxt Spine, LLC – K171140)
Additional Predicate:Honour System (Nexxt Spine, LLC – K120345), Capri® Corpectomy CageSystem (K2M Inc. – K180665)
Performance Data:The modified corpectomy device was evaluated via mechanical testingincluding included static and dynamic compression and static and dynamictorsion (ASTM F2077), subsidence (ASTM F2267) and expulsion. Theresults demonstrated the performance of the modified corpectomy issubstantially equivalent to the predicate.
TechnologicalCharacteristics:The modified NEXXT MATRIXX® System corpectomy possesses the sametechnological characteristics as one or more of the predicate devices. Theseinclude:performance (as described above), basic design (additively manufactured structural interbody), material (titanium alloy) and size (dimensions are comparable to those offered by the cleared devices). Therefore the fundamental scientific technology of the modified NEXXTMATRIXX® System corpectomy is the same as previously cleared devices.
Conclusion:The modified NEXXT MATRIXX® System corpectomy possesses the sameintended use and technological characteristics as the predicate devices.Therefore the modified NEXXT MATRIXX® System corpectomy issubstantially equivalent for its intended use.

{4}------------------------------------------------

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.