K Number
K193377

Validate with FDA (Live)

Date Cleared
2020-11-25

(356 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TopLock Anchor System is intended to secure soft tissue to bone of:
The Shoulder:
Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation, Rotator Cuff Repair, Capsule Repair, Biceps Tenodesis. Deltoid Repair
The Elbow:
Ulnar or Radial Collateral Ligament Reconstruction, Bicep Tendon Reconstruction, Tennis Elbow Repair
The Hand and Wrist:
Scapholunate Ligament Reconstruction, Ulnar / Radial Collateral Ligament Reconstruction, Extensor Tendon Repair, Flexor Tendon Repair
The Knee:
Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Tendon Repair
The Ankle and Foot:
Lateral Ligament Stabilization Repair, Medial Ligament Stabilization Repair.
Achilles Tendon Repair / Reconstruction, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Mid and Rear Foot Reconstruction

Device Description

The TopLock Anchor System consists of a collection of barbed suture anchors, manufactured from medical grade Polyetheretherketone or PEEK (ASTM F2026). The anchors are offered from diameters of 3.5mm to 5.0mm with each size offering two eyelet size options to accommodate suture and suture tape. The size range and configurations are offered for surgical treatment of ligament, tendon and soft tissue pathologies of the shoulder and other joints. The specialized instruments are made primarily of surgical grade stainless steel (ASTM F899).

AI/ML Overview

This document is a 510(k) premarket notification for a medical device (TopLock Anchor System). It describes the device, its indications for use, and a comparison to predicate devices to demonstrate substantial equivalence.

Crucially, this document explicitly states that no clinical studies were performed to prove the device meets acceptance criteria. Instead, non-clinical tests (mechanical testing, cleaning, packaging, sterilization, pyrogenicity) were conducted to demonstrate design equivalence to predicate devices.

Therefore, I cannot provide the detailed information requested in the prompt regarding acceptance criteria, study design, ground truth, or expert involvement for clinical performance. The information provided in the document pertains to non-clinical acceptance criteria and testing for mechanical and material equivalence.

Based on the provided text, here's what can be extracted regarding the device's validation:

Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

Acceptance Criteria (Type of Test)Reported Device Performance
Cleaning ValidationPerformed and met requirements.
Packaging ValidationPerformed and met requirements.
Sterilization ValidationPerformed and met requirements.
Insertion TestingPerformed, showed design equivalence.
Axial Pull-Out TestingPerformed, showed design equivalence.
Cyclic Fatigue TestingPerformed, showed design equivalence.
Post-Fatigue Pullout TestingPerformed, showed design equivalence.
Pyrogenicity (LAL assay)Met recommended maximum endotoxin levels of 20 EU/Device.

Regarding the specific questions about clinical performance and AI-related studies:

  1. A table of acceptance criteria and the reported device performance: Covered above for non-clinical aspects. No clinical performance criteria or data are provided.
  2. Sample size used for the test set and the data provenance: Not applicable to clinical performance as no clinical studies were performed. For non-clinical tests, specific sample sizes are not detailed, but standard validation protocols would have been followed for mechanical and material testing.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for clinical outcomes was not established as no clinical studies were performed.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a surgical implant, not an AI-assisted diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a surgical implant.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable for clinical performance. For non-clinical testing, the "ground truth" would be established engineering and material specifications, and regulatory standards.
  8. The sample size for the training set: Not applicable. No training set for an AI/algorithm was used.
  9. How the ground truth for the training set was established: Not applicable.

In summary, the TopLock Anchor System received 510(k) clearance based on demonstrating "substantial equivalence" to existing predicate devices through non-clinical testing (mechanical, material, and sterility validations), rather than through clinical performance studies. The document explicitly states: "No clinical studies were performed."

{0}------------------------------------------------

November 25, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Fusion Orthopedics, LLC Kolby Black 4135 S. Power Rd., Suite 110 Mesa, Arizona 85212

Re: K193377/S001

Trade/Device Name: TopLock Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: October 28, 2020 Received: October 30, 2020

Dear Mr. Black:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laura C. Rose. Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193377

Device Name TopLock Anchor System

Indications for Use (Describe)

The TopLock Anchor System is intended to secure soft tissue to bone of:

The Shoulder:

Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation, Rotator Cuff Repair, Capsule Repair, Biceps Tenodesis. Deltoid Repair

The Elbow:

Ulnar or Radial Collateral Ligament Reconstruction, Bicep Tendon Reconstruction, Tennis Elbow Repair The Hand and Wrist:

Scapholunate Ligament Reconstruction, Ulnar / Radial Collateral Ligament Reconstruction, Extensor Tendon Repair, Flexor Tendon Repair

The Knee:

Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Tendon Repair

The Ankle and Foot:

Lateral Ligament Stabilization Repair, Medial Ligament Stabilization Repair.

Achilles Tendon Repair / Reconstruction, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Mid and Rear Foot Reconstruction

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date PreparedNovember 24, 2020
Submitted ByFusion Orthopedics, LLC4135 S. Power Rd., Suite 110Mesa, AZ 85212800-403-6876
Primary ContactKolby Black4135 S. Power Rd., Suite 110Mesa, AZ 85212800-403-6876 Telee-mail: FDA@fusionorthopedics.com
Trade NameTopLock Anchor System
SubmissionNumberK193377
ClassificationNameFastener, Fixation, Non-degradable, Soft Tissue
ClassII
Product CodeMBI
CFR Section21 CFR Section 888.3040
Device PanelOrthopedic
Primary PredicateDeviceCore Essence reNOVO Suture Anchors (K071520)
SecondaryPredicate DeviceArthrex PEEK Mini PushLock Anchors (K101679)
ReferenceDevicesCore Essence Orthopaedics, Inc., TAC-tite™ PEEK (K111716)Parcus Medical Twist & Triple Twist PEEK Screw-In Suture Anchors (K120942)Parcus Medical PEEK CF Push-In Anchors (K102162)
DeviceDescriptionThe TopLock Anchor System consists of a collection of barbed suture anchors,manufactured from medical grade Polyetheretherketone or PEEK (ASTM F2026). Theanchors are offered from diameters of 3.5mm to 5.0mm with each size offering two eyeletsize options to accommodate suture and suture tape. The size range and configurations areoffered for surgical treatment of ligament, tendon and soft tissue pathologies of the shoulderand other joints. The specialized instruments are made primarily of surgical grade stainlesssteel (ASTM F899).
Indications forUseThe TopLock Anchor System is intended to secure soft tissue to bone of:The Shoulder:Bankart RepairSLAP Lesion RepairAcromio-Clavicular Separation
Capsule RepairBiceps TenodesisDeltoid Repair
The Elbow:
Ulnar or Radial Collateral Ligament ReconstructionBicep Tendon ReconstructionTennis Elbow Repair
The Hand and Wrist:
Scapholunate Ligament ReconstructionUlnar / Radial Collateral Ligament ReconstructionExtensor Tendon RepairFlexor Tendon Repair
The Knee:
Medial Collateral Ligament RepairLateral Collateral Ligament RepairPosterior Oblique Ligament RepairExtra Capsular ReconstructionIliotibial Band TenodesisPatellar Tendon Repair
The Ankle and Foot:
Lateral Ligament Stabilization RepairMedial Ligament Stabilization RepairAchilles Tendon Repair / ReconstructionMetatarsal Ligament RepairHallux Valgus ReconstructionMid and Rear Foot Reconstruction
MaterialsPolyetheretherketone (PEEK) (ASTM F2026)Ultra High Molecular Weight Polyethylene Suture (UHMWPE) (ASTM F2848)Stainless Steel (ASTM F899)Aluminum (ASTM B209)Nylon
SubstantialEquivalenceClaimed toPredicate DevicesThe TopLock Anchor System is substantially equivalent to the predicate devices in terms ofdesign, materials used, mechanical safety and performances.
Non-clinical TestSummaryValidations were performed on the cleaning, packaging and sterilization of the implants andassociated surgical instruments. Mechanical Testing was performed to show designequivalence including insertion testing, axial pull-out testing, cyclic fatigue testing, and postfatigue pullout testing. Pyrogenicity was evaluated using the Limulus amebocyte lysate(LAL) assay. The testing demonstrated that the subject device meets the recommendedmaximum endotoxin levels of 20 (EU)/Device.
Clinical TestSummaryNo clinical studies were performed.
Conclusions:Non clinical andFusion Orthopedics LLC considers the TopLock Anchor System to be equivalent to thepredicate devices listed above. This conclusion is based upon the devices' similarities in

{4}------------------------------------------------

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.