K Number
K193345

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-07-08

(218 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex DynaNite Compression Plate is intended to be used for fixation such as: LisFranc arthrodesis, mono or bicortical osteotomies in the forefoot, first metatarsophalangeal arthrodesis, Akin osteotomy, midfoot and hindfoot arthrodeses or osteotomies for hallux valgus treatment (Scarf and Chevron), and arthrodesis of the metatarsocuneiform joint to reposition and stabilize metatarsus primus varus.

The Arthrex DynaNite Compression Plate is intended to be used in conjunction with the Arthrex Compression FT Screws.

Device Description

The Arthrex DynaNite Compression Plate is an implant with a number of threaded holes and a central region of articulating arms. Sizes range from approximately 15 to 30mm in hole-to-hole distance. Geometrically, the plate is offered in straight, 'T-shape', and 'X-shape' configurations. The Arthrex DynaNite Compression Plate is sold attached to a clamp delivery device, which constrains the implant in an elongated state, along its long axis. The clamp delivery device is removed after screw fixation is achieved, which in turn causes the plate to return to its original confirmation (shape memory effect), shortening along its long axis and generating compression. The Arthrex DynaNite Compression Plate is sold as sterile and is single-use. The Arthrex DynaNite Compression Plate is to be used with the Arthrex 3.0 mm Titanium VA Locking Screws and Locking Screws cleared under K143614 and K150456.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for a medical device called the "Arthrex DynaNite Compression Plate." This submission confirms that the device is substantially equivalent to legally marketed predicate devices, not an AI/ML device. Therefore, the requested information regarding acceptance criteria and study details for an AI/ML device is not applicable to this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

July 8, 2020

Arthrex Inc. Rebecca R. Homan Senior Regulatory Affairs Associate 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K193345

Trade/Device Name: Arthrex DynaNite Compression Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: June 3, 2020 Received: June 5, 2020

Dear Rebecca R. Homan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193345

Device Name Arthrex DynaNite Compression Plate

Indications for Use (Describe)

The Arthrex DynaNite Compression Plate is intended to be used for fixation such as: LisFranc arthrodesis, mono or bicortical osteotomies in the forefoot, first metatarsophalangeal arthrodesis, Akin osteotomy, midfoot and hindfoot arthrodeses or osteotomies for hallux valgus treatment (Scarf and Chevron), and arthrodesis of the metatarsocuneiform joint to reposition and stabilize metatarsus primus varus.

The Arthrex DynaNite Compression Plate is intended to be used in conjunction with the Arthrex Compression FT Screws.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date PreparedJuly 1, 2020
SubmitterArthrex Inc.1370 Creekside BoulevardNaples, FL 34108-1945
Contact PersonRebecca R. HomanRegulatory Affairs Specialist1-239-643-5553, ext. 73429rebecca.homan@arthrex.com
Name of DeviceArthrex DynaNite Compression Plate
Common Nameplate, fixation, bone
Product CodeHRS
Classification Name21 CFR 888.3030: Single/multiple component metallic bone fixation appliancesand accessories
Regulatory ClassII
Predicate DeviceK161303: ΜΧO dynaMX Compression Plate (Primary Predicate)K090047: Synthes (USA) 1.5mm Mini Fragment LCP System
Reference DeviceK172052: Arthrex DynaNite Nitinol Staple
Purpose of SubmissionThis Traditional 510(k) premarket notification is submitted to obtain clearance forthe Arthrex DynaNite Compression Plate.
Device DescriptionThe Arthrex DynaNite Compression Plate is an implant with a number ofthreaded holes and a central region of articulating arms. Sizes range fromapproximately 15 to 30mm in hole-to-hole distance. Geometrically, the plate isoffered in straight, 'T-shape', and 'X-shape' configurations. The Arthrex DynaNiteCompression Plate is sold attached to a clamp delivery device, which constrainsthe implant in an elongated state, along its long axis. The clamp delivery device isremoved after screw fixation is achieved, which in turn causes the plate to returnto its original confirmation (shape memory effect), shortening along its long axisand generating compression. The Arthrex DynaNite Compression Plate is sold assterile and is single-use. The Arthrex DynaNite Compression Plate is to be usedwith the Arthrex 3.0 mm Titanium VA Locking Screws and Locking Screws clearedunder K143614 and K150456.
Indications for UseThe Arthrex DynaNite Compression Plate is indicated to be used for fixation suchas: LisFranc arthrodesis, mono or bi-cortical osteotomies in the forefoot, firstmetatarsophalangeal arthrodesis, Akin osteotomy, midfoot and hindfootarthrodeses or osteotomies, fixation of osteotomies for hallux valgus treatment(Scarf and Chevron), and arthrodesis of the metatarsocuneiform joint toreposition and stabilize metatarsus primus varus.The Arthrex DynaNite Compression Plate is intended to be used in conjunctionwith the Arthrex Compression FT Screws.
Performance DataStatic Four-Point Bend (ASTM F382) and Four-Point Bend Fatigue (ASTM F382)testing was conducted to demonstrate that the proposed Arthrex DynaNiteCompression Plate performs statistically equivalent to the predicate devicescleared under K090047 and K172052. Compressive Force, TransformationTemperature (ASTM F2082), Cyclic Potentiodynamic Polarization Corrosion(ASTM F2129), and Galvanic Corrosion Resistance (ASTM F2129) testing was alsoconducted.MRI force, torque, and image artifact testing were conducted in accordance with
FDA guidance Testing and Labeling Medical Devices for Safety in the MagneticResonance (MR) Environment, ASTM F2052 Standard Test Method forMeasurement of Magnetically Induced Displacement Force on Medical Devices inthe Magnetic Resonance Environment, ASTM F2119 Standard Test Method forEvaluation of MR Image Artifacts from Passive Implants, ASTM F2182 StandardTest Method for Measurement of Measurement of Radio Frequency InducedHeating Near Passive Implants During Magnetic Resonance Imaging and ASTMF2213 Standard Test Method for Measurement of Magnetically Induced Torqueon Medical Devices in the Magnetic Resonance Environment.
Bacterial Endotoxins Test (BET) was performed on the Arthrex DynaNiteCompression Plate utilizing the Kinetic Chromogenic Method in accordance with
ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, EP 2.6.14. The testingconducted demonstrates that the Arthrex DynaNite Compression Plate meetspyrogen limit specifications.
Cytotoxicity, Sensitization, Irritation, Genotoxicity, Systemic Toxicity,Subchronic/Subacute Toxicity, Implantation and Material Characterization testingwas conducted on the Arthrex DynaNite Compression Plate in accordance withISO 10993-1:2018.
Assessment of physical product attributes including product, design, size, andmaterials as well as the conditions of manufacture and packaging has determinedthat the Arthrex DynaNite Compression Plate does not introduce additional risksor concerns regarding sterilization and shelf-life.
ConclusionThe Arthrex DynaNite Compression Plate is substantially equivalent to thepredicate devices in which the basic design features and intended uses are thesame. Any differences between the proposed device and the predicate devicesare considered minor and do not raise different questions concerning safety oreffectiveness.
The submitted mechanical testing data demonstrates that the Static Four-PointBend and Four-Point Bend Fatigue strength of the proposed device issubstantially equivalent to that of the predicate devices for the desiredindications.
Based on the indications for use, technological characteristics, and the summaryof data submitted, Arthrex Inc. has determined that the proposed device issubstantially equivalent to the currently marketed predicate devices.

{4}------------------------------------------------

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.