K Number
K193232

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-02-13

(80 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The Emprint™ Ablation System is not intended for use in cardiac procedures.

The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.

The system's optional, 3-D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted traiectory on a computer monitor. The size and shape of the predicted ablation zone relative to the position of the antenna is displayed on the enhanced ultrasound image.

Device Description

The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation platform, was released for commercial distribution in the United States in April 2014. The 510(k) Cleared Emprint™ Ablation System consists of the following components: 1. Emprint™ Ablation Generator (2450 MHz), 2. Emprint™ Percutaneous Antenna (sterile, single use), 3. Emprint™ Ablation Reusable Cable, 4. Emprint™ Ablation Pump. The system also includes the following optional equipment/accessories: 5. Emprint™ Ablation Cart, 6. Ablation Footswitch, 7. Isolation Transformer, 8. Remote Temperature Probe (sterile, single use). The ablation platform must be used with a standard IV bag of sterile, normal saline (not provided with the device).

The Emprint™ SX Ablation Platform with Thermosphere™ Technology is a microwave ablation system with an optional navigation feature. It relies on the same principles of operation, fundamental technology, and performance characteristics as its predicate, the Emprint™ Ablation System with Thermosphere™ Technology (K133821, K163105). The main differences between the Emprint™ SX Ablation Platform with Thermosphere™ Technology and the predicate device include software (vs. manual) control of the generator and pump, as well as the addition of an optional navigation feature. Both systems include a 2.45 GHz microwave generator, a cooling pump, and accessories. A bag/bottle of sterile fluid that connects to the antenna's tubing and to the peristaltic pump is required for proper system operation. The 510(k) Cleared Emprint™ Percutaneous Antenna (K163105) can be used with either system to create predictable, spherical ablation zones in soft tissue in percutaneous, laparoscopic, and intraoperative procedures.

The Emprint™ antenna, the applied part of the Emprint™ ablation system with Thermosphere™ Technology and Emprint™ SX ablation platform with Thermosphere™ Technology that delivers microwave energy to the patient, has a sharp, engineered ceramic tip (called "trocar") at the distal end of the device that allows a user to puncture tissue. The shaft of the disposable antenna is inserted directly into the tissue intended to be ablated. The radiating tip of the antenna shaft delivers microwave energy to the target tissue.

Modifications have been made to the Emprint™ Ablation System and Emprint™ SX Ablation Platform's IFU's and Antenna's. Updates were made to the IFU to reduce the possible trocar detachment post ablation resulting from atypical use. Changes made to the bond curing schedule is to enhance the bond strength between the trocar and the fiberglass shaft. Additionally, the clear non-stick heat shrink was extended further on the antennae to reduce tissue adherence post ablation.

AI/ML Overview

The information provided does not describe a study involving a device that uses AI or reports its performance using metrics such as sensitivity, specificity, or AUC. Instead, it describes a medical device (an ablation system) and modifications made to its antenna and instructions for use. The acceptance criteria and performance data discussed are related to the physical and functional characteristics of the ablation antenna, not AI performance. Therefore, I cannot fulfill the request as it pertains to AI-based device performance and study design.

However, I can extract the acceptance criteria and performance data related to the physical modifications of the Emprint Ablation System with Thermosphere Technology and Emprint SX Ablation Platform with Thermosphere Technology antennae based on the provided text, and present it in a modified format relevant to the device's physical changes.

Accepted Criteria and Device Performance (for physical modifications)

Acceptance CriteriaReported Device Performance
Trocar bond strength greater than the predicate.Verification testing demonstrated the modified antenna's trocar bond strength was greater than the predicate.
No impact on the reliability of the device (thermal ablations in soft tissue during extended use test).Verification testing confirmed that the change did not impact the reliability of the device by performing thermal ablations in soft tissue during the extended use test.
Quantify and compare frictional force between modified antenna and adjustable depth guide.Verification testing quantified and compared the frictional force between the modified antenna and the adjustable depth guide. (Specific quantitative results not provided in this summary, but the test was conducted.)
Quantify and compare insertion force (ex-vivo liver tissue) of the modified antenna relative to cleared devices.Verification testing quantified and compared the insertion force (ex-vivo liver tissue) of the modified antenna relative to the 510(k) cleared devices. (Specific quantitative results not provided in this summary, but the test was conducted.)
Quantify and compare extraction force (ex-vivo liver tissue) of the modified antenna relative to cleared devices.Verification testing quantified and compared the extraction force (ex-vivo liver tissue) of the modified antenna relative to the 510(k) cleared devices. (Specific quantitative results not provided in this summary, but the test was conducted.)
The modifications do not affect the system's essential function, performance, or underlying principles of operation.The modification to the antennae trocar bonding and clear non-stick heat shrink extension have improved parameters such as tensile bond strength, antenna insertion force, and antenna extraction force compared to the current device (predicate) without affecting the design requirements and the essential performance of the antenna and the system as a whole. The antenna changes do not affect the Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology's essential function, performance, or underlying principles of operation. Modifications made to the Antenna's do not alter the Indications for use for the Emprint™ Ablation system and the Emprint™ SX ablation platform.

Study Details (for physical modifications)

  1. Sample size used for the test set and the data provenance: Not explicitly stated for specific tests (e.g., how many antennas were tested for bond strength or insertion force). The verification testing included "thermal ablations in soft tissue during the extended use test" and "ex-vivo liver tissue." The data provenance is implied to be from bench testing conducted by Covidien LLC/Medtronic Inc. (the manufacturer). The study is retrospective in the sense that it evaluates modifications to an existing, cleared device.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable. The ground truth here is related to physical measurements (bond strength, forces, reliability during ablation) demonstrated through bench and ex-vivo testing, not expert consensus on medical images or diagnoses.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. This was bench and ex-vivo testing, not a clinical study requiring adjudication of expert opinions.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe an AI medical device or a MRMC study.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This document does not describe an AI medical device.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" for the performance claims in this submission is derived from bench verification testing (measuring physical parameters like bond strength, frictional forces, insertion/extraction forces) and ex-vivo testing (thermal ablations in soft tissue, ex-vivo liver tissue).

  7. The sample size for the training set: Not applicable. This is not an AI/machine learning device.

  8. How the ground truth for the training set was established: Not applicable. This is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 13, 2020

Covidien LLC Medtronic Inc. Hilla Debby Director Regulatory Affairs 6135 Gunbarrel Ave Boulder, CO 80301

Re: K193232

Trade/Device Name: Emprint Ablation System with Thermosphere Technology, Emprint SX Ablation Platform with Thermosphere Technology Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NEY Dated: January 22, 2020 Received: January 23, 2020

Dear Hilla Debby:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Covidien LLC

Special 510(k) Submission Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology IFU Updates and Antenna Modifications

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: 06/30/2020
Indications for UseSee PRA Statement below.
510(k) Number (if known)

5

K193232

Device Name

Emprint™ Ablation System with Thermosphere™ Technology Emprint™ SX Ablation Platform with Thermosphere™ Technology

Indications for Use (Describe)

The Emprint™ Ablation System with Thermosphere™ Technology:

The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The EmprintTM Ablation System is not intended for use in cardiac procedures.

The Emprint™ SX Ablation Platform with Thermosphere™M Technology:

The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.

The system's optional, 3-D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted traiectory on a computer monitor. The size and shape of the predicted ablation zone relative to the position of the antenna is displayed on the enhanced ultrasound image.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter -

Medtronic Inc 5920 Longbow Dr. Boulder, CO 80301

Contact Person:Hilla Debby
Director of Regulatory Affairs
Phone:97-204-909-7774
Email:Hilla.Debby@medtronic.com
Address:2 Hacarmet St. New Industrial Park POB 258 Yoqneam, IL 20692
Date of Summary Preparation:November 22, 2019

II. Devices

Table 5.1 Devices:

Device Trade NamesEmprint™ Ablation System with Thermosphere™TechnologyEmprint™ SX Ablation Platform withThermosphere™ Technology
Device Common NameMicrowave Ablation System
Classification NameSystem, Ablation, Microwave and Accessories
Regulatory ClassII
Product CodeNEY

III. Predicate devices

This Special 510(k) submission describes changes to the Instructions for Use (IFU) and modifications to the Antenna's previously cleared for the Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology. The modified Emprint™ Ablation System with Thermosphere™ Technology and the modified Emprint™ SX Ablation Platform with Thermosphere™ Technology are substantially equivalent to the predicate devices outlined in Table 5.2.

No reference devices were used in this submission

{4}------------------------------------------------

Table 5.2 Predicate Devices:

Device NameClassification Name and Class510(k) Clearance No.
Emprint™ Ablation System withThermosphere™ TechnologySystem, Ablation, Microwave andAccessories; Class IIK133821
Emprint™ Ablation System withThermosphere™ TechnologySystem, Ablation, Microwave andAccessories; Class IIK163105
Emprint™ SX Ablation Platformwith Thermosphere™ TechnologySystem, Ablation, Microwave andAccessories; Class IIK171358

IV. Device description

The Emprint™ Ablation System with Thermosphere™ Technology, Covidien's microwave ablation platform, was released for commercial distribution in the United States in April 2014.

The 510(k) Cleared Emprint™ Ablation System consists of the following components:

    1. Emprint™ Ablation Generator (2450 MHz)
    1. Emprint™ Percutaneous Antenna (sterile, single use)
    1. Emprint™ Ablation Reusable Cable
    1. Emprint™ Ablation Pump

The system also includes the following optional equipment/accessories:

    1. Emprint™ Ablation Cart
    1. Ablation Footswitch
    1. Isolation Transformer
    1. Remote Temperature Probe (sterile, single use)

The ablation platform must be used with a standard IV bag of sterile, normal saline (not provided with the device).

In 2016 Covidien LLC made changes to improve the usability of the Emprint™ Ablation System's percutaneous antenna, by increasing the stiffness of the antenna's shaft. Like the predicate antenna approved via traditional 510(k) (K133821), the modified antenna is manufactured in 3 lengths (15, 20 and 30 cm) and is uniquely identified by catalog numbers (CA15L2, CA20L2 and CA30L2).

Covidien made the change to accommodate a usability-related preference, expressed by physicians. The modification was not in response to reported device malfunctions or adverse events associated with the current, more-flexible ablation antenna.

This change in stiffness enhances the antenna's usability; and has no effect on the system's essential function, performance or underlying principles of operation.

The Emprint™ SX Ablation Platform with Thermosphere™ Technology is a microwave ablation system with an optional navigation feature. It relies on the same principles of operation, fundamental technology, and performance characteristics as its predicate, the Emprint™ Ablation

{5}------------------------------------------------

System with Thermosphere™ Technology (K133821, K163105).

The main differences between the Emprint™ SX Ablation Platform with Thermosphere™ Technology and the predicate device include software (vs. manual) control of the generator and pump, as well as the addition of an optional navigation feature. Both systems include a 2.45 GHz microwave generator, a cooling pump, and accessories. A bag/bottle of sterile fluid that connects to the antenna's tubing and to the peristaltic pump is required for proper system operation. The 510(k) Cleared Emprint™ Percutaneous Antenna (K163105) can be used with either system to create predictable, spherical ablation zones in soft tissue in percutaneous, laparoscopic, and intraoperative procedures.

The Emprint™ antenna, the applied part of the Emprint™ ablation system with Thermosphere™ Technology and Emprint™ SX ablation platform with Thermosphere™ Technology that delivers microwave energy to the patient, has a sharp, engineered ceramic tip (called "trocar") at the distal end of the device that allows a user to puncture tissue. The shaft of the disposable antenna is inserted directly into the tissue intended to be ablated. The radiating tip of the antenna shaft delivers microwave energy to the target tissue.

Modifications have been made to the Emprint™ Ablation System and Emprint™ SX Ablation Platform's IFU's and Antenna's. Updates were made to the IFU to reduce the possible trocar detachment post ablation resulting from atypical use.

Changes made to the bond curing schedule is to enhance the bond strength between the trocar and the fiberglass shaft.

Additionally, the clear non-stick heat shrink was extended further on the antennae to reduce tissue adherence post ablation.

The Antenna changes mentioned above do not affect the Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology essential function, performance or underlying principles of operation.

Extensive bench verification testing was conducted to ensure the modified Emprint™ Percutaneous Antenna performs as designed and meets essential requirements. The device modification was implemented in compliance with FDA's quality system and design control requirements.

V. Description of Changes

A. INSTRUCTIONS FOR USE (IFU) UPDATE

The following language (see table 5.3 below) was added to the Instructions for Use (IFU) for both the Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology:

"5. When energy delivery to the antenna stops, allow the pump to continue running for five seconds to cool the antenna. Then rotate the antenna shaft to release any adhered tissue. Then reposition or remove the antenna."

{6}------------------------------------------------

Additionally, several minor edits, unrelated to trocar detachment from atypical use, were included in the instructions for use.

ProductCodeDescriptionChange DescriptionName of theDevices510 (k)ClearanceNo.
PT00101561(English)Instructionfor Use(IFU)Added "When energy delivery to the antenna stops, allow the pump to continue running for five seconds to cool the antenna. Then rotate the antenna shaft to release any adhered tissue. Then reposition or remove the antenna." Additionally, several clarifications were included in the instructions for use updates.Emprint™Ablation SystemwithThermosphere™TechnologyK133821andK163105
PT00102817(English)Instructionfor Use(IFU)Added "When energy delivery to the antenna stops, allow the pump to continue running for five seconds to cool the antenna. Then rotate the antenna shaft to release any adhered tissue. Then reposition or remove the antenna." Additionally, several clarifications were included in the instructions for use updates.Emprint™ SXAblation PlatformwithThermosphere™TechnologyK171358
Table 5.3 Lanquage Change Summary for IFU:

B. Bond Cure Schedule Change

Changes have been made to the bond cure schedule in order to increase the bond strength between the trocar and the fiberglass shaft.

{7}------------------------------------------------

ﻥ CLEAR NON-STICK HEAT SHRINK DESIGN CHANGE

The clear non-stick heat shrink was extended further on the antennae to reduce tissue adherence.

VI. Indications for use

Modifications made to the Antenna's do not alter the Indications for use (shown below) for the Emprint™ Ablation system and the Emprint™ SX ablation platform.

The Emprint™ Ablation System with Thermosphere™ Technology:

The Emprint™ Ablation System is intended for use in percutaneous, laparoscopic, endoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The Emprint™ Ablation System is not intended for use in cardiac procedures.

The Emprint™ SX Ablation Platform with Thermosphere™ Technology:

The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.

The system's optional, 3-D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted trajectory on a computer monitor. The size and shape of the predicted ablation zone relative to the position of the antenna is displayed on the enhanced ultrasound image.

VII. Comparison of technological characteristics with the predicate device

The Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology current antennae (predicate device's) and the modified ablation antenna share the same underlying technology. Both devices are used to deliver microwave energy to soft tissue in a highly controlled way. Signal frequency, power and exposure time are controlled by the system's 2450 MHz generator. The design of the radiating portion of the accessory and continuous cooling of the accessory during the ablation procedure determines the shape (near spherical) and size of the resulting ablations.

The modification to the antennae trocar bonding and clear non-stick heat shrink extension have improved parameters such as tensile bond strength, antenna insertion force and antenna extraction force compared to the current device (predicate) without affecting the design requirements and the essential performance of the antenna and the system as a whole.

{8}------------------------------------------------

VIII. Performance data

Non-clinical performance data was provided in support of the substantial equivalence determination.

Bench Verification Testing a)

Verification testing was conducted to:(1) demonstrate the modified antenna's trocar bond strength was greater than the predicate, (2) confirm that the change did not impact the reliability of the device by performing thermal ablations in soft tissue during the extended use test. (3) quantify and compare the frictional force between the modified antenna and the adjustable depth quide. insertion force and extraction force (ex-vivo liver tissue) of the modified antenna relative to the 510(k) cleared devices.

IX. Conclusions

The modified Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology antennae are identical to the predicate device's, except for bond cure schedule change and clear non-stick heat shrink design change. The subject and predicate devices share intended use, technological characteristics, and meet identical performance requirements. The results of comprehensive verification testing raised no new questions of safety or effectiveness. The Emprint™ Ablation System with Thermosphere™ Technology and Emprint™ SX Ablation Platform with Thermosphere™ Technology Antennae (subject device) and the modified antennae are substantially equivalent.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.