(624 days)
The Nebulizer kit is used to aerosolize medication approved for nebulization and prescribed by a physician for delivery to the airways. The Nebulizer kit is intended for adult consistent with the indication in hospitals and sub-acute institutions.
The nebulizer kit is a small volume jet nebulizer designed to deliver aerosolized medications for inhalation to the respiratory system, the compressed air or oxygen is forced into the nebulizer through a converging nozzle, where it accelerates and emerges at a high velocity, creating a vacuum (Venturi effect), in the process the liquid medicine is converted into small droplets called aerosol by impacting upon a rigid baffle, the aerosol may be used for inhalation treatment of a physician's prescription. The nebulizer kit is single-patient, multi-use; it requires a cleaning and disinfection process before each use, and the service life is for 30 aerosol treatments.
The provided document is a 510(k) summary for a Nebulizer Kit (K192971). It focuses on demonstrating substantial equivalence to a predicate device (VixOne, K800562) rather than proving the device meets acceptance criteria for an AI/ML medical device. As such, many of the requested details, particularly regarding AI/ML related study elements (e.g., sample size for test/training sets, expert consensus for ground truth, MRMC studies, standalone algorithm performance), are not applicable or cannot be extracted from this document.
However, I can extract information related to the device's performance characteristics which were used to demonstrate equivalence to the predicate device, serving as a form of "acceptance criteria" for this type of medical device submission.
Here's how the available information aligns with your request:
Device: Nebulizer Kit (K192971)
Study Purpose: To demonstrate substantial equivalence of the proposed Nebulizer Kit to the predicate device (VixOne Nebulizer, K800562) based on performance characteristics.
1. Table of Acceptance Criteria and Reported Device Performance
For nebulizers, key performance characteristics relate to the aerosolized medication. While specific "acceptance criteria" against numerical targets aren't explicitly stated as pass/fail thresholds in a table, the document implicitly defines acceptance as "not significant statistical difference" compared to the predicate device for key aerosol characteristics.
| Performance Metric (Aerosol Characteristics) | Acceptance Criteria (Implicit) | Proposed Device Performance (Mean ± SD) @ 8L/min Flow | Predicate Device Performance (Mean ± SD) @ 8L/min Flow | Proposed Device Performance (Mean ± SD) @ 4L/min Flow | Predicate Device Performance (Mean ± SD) @ 4L/min Flow |
|---|---|---|---|---|---|
| Total Mass (µg) | No significant statistical difference from predicate device | Albuterol: 1146.80±142.12 | Albuterol: 1309.33±137.13 | Albuterol: 1031.51±113.44 | Albuterol: 1104.12±72.02 |
| Ipratropium: 228.18±12.40 | Ipratropium: 213.81±15.03 | Ipratropium: 199.78±15.60 | Ipratropium: 216.84±10.63 | ||
| Cromolyn: 6307.49±802.07 | Cromolyn: 7172.85±626.18 | Cromolyn: 5466.01±567.02 | Cromolyn: 5987.10±715.04 | ||
| Budesonide: 467.07±34.11 | Budesonide: 501.02±23.65 | Budesonide: 430.79±18.55 | Budesonide: 452.60±22.52 | ||
| Particle Size - MMAD (µm) | No significant statistical difference from predicate device | Albuterol: 2.80±0.23 | Albuterol: 2.55±0.15 | Albuterol: 4.71±0.11 | Albuterol: 4.52±0.13 |
| Ipratropium: 2.14±0.22 | Ipratropium: 2.07±0.20 | Ipratropium: 3.21±0.17 | Ipratropium: 3.19±0.24 | ||
| Cromolyn: 2.66±0.20 | Cromolyn: 2.41±0.15 | Cromolyn: 4.25±0.18 | Cromolyn: 4.09±0.10 | ||
| Budesonide: 3.32±0.18 | Budesonide: 3.11±0.18 | Budesonide: 4.97±0.14 | Budesonide: 4.84±0.15 | ||
| Geometric Standard Deviation (GSD) | No significant statistical difference from predicate device | Albuterol: 2.49±0.16 | Albuterol: 2.51±0.05 | Albuterol: 2.42±0.05 | Albuterol: 2.40±0.07 |
| Ipratropium: 2.81±0.06 | Ipratropium: 3.01±0.16 | Ipratropium: 2.76±0.16 | Ipratropium: 2.71±0.11 | ||
| Cromolyn: 2.48±0.09 | Cromolyn: 2.56±0.09 | Cromolyn: 2.52±0.01 | Cromolyn: 2.48±0.05 | ||
| Budesonide: 2.27±0.08 | Budesonide: 2.20±0.12 | Budesonide: 2.53±0.13 | Budesonide: 2.59±0.15 | ||
| Respirable Fraction (% 0.5-5µm) | No significant statistical difference from predicate device | Albuterol: 70.76±1 | Albuterol: 73.02±2 | Albuterol: 52.17±1 | Albuterol: 54.06±1 |
| Ipratropium: 71.32±3 | Ipratropium: 68.86±1 | Ipratropium: 63.63±3 | Ipratropium: 64.3±3 | ||
| Cromolyn: 72.33±3 | Cromolyn: 73.52±3 | Cromolyn: 55.96±2 | Cromolyn: 57.79±1 | ||
| Budesonide: 68.09±2 | Budesonide: 71.82±4 | Budesonide: 49.57±1 | Budesonide: 50.55±0.1 | ||
| Respirable Mass (µg 0.5-5µm) | No significant statistical difference from predicate device | Albuterol: 810.89±94.73 | Albuterol: 956.45±107.82 | Albuterol: 538.94±70.97 | Albuterol: 596.14±35.20 |
| Ipratropium: 162.63±10.34 | Ipratropium: 147.16±11.58 | Ipratropium: 126.97±10.16 | Ipratropium: 139.41±11.84 | ||
| Cromolyn: 4576.38±739.65 | Cromolyn: 5273.70±512.70 | Cromolyn: 3056.41±297.19 | Cromolyn: 3460.11±416.01 | ||
| Budesonide: 318.22±27.49 | Budesonide: 359.25±7.21 | Budesonide: 213.50±8.65 | Budesonide: 228.69±9.93 |
Overall Conclusion regarding acceptance: The document states: "the parameters of total mass, MMAD, GSD, respirable fraction and respirable Mass are analyzed and compared at a 95% confidence level, there is not significant statistical difference between the two devices." This indicates the proposed device "meets the requirements of its pre-defined acceptance criteria" by being statistically comparable to the predicate for these key characteristics.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (for performance testing): While specific numerical sample sizes for each drug and flow rate are not explicitly stated as "n=X", the phrase "For each device, four drugs... were tested respectively" implies a repeated measures design. The presence of mean and standard deviation values indicates multiple measurements were taken for each characteristic. The "variability of intra-sample and inter-sample was tested for the proposed devices," further suggesting a robust, though unspecified, sample size for these technical measurements.
- Data Provenance: The testing was conducted in accordance with relevant standards (CDRH Guidance Document, USP 40<601>, ISO 27427:2013). The manufacturer is Excellentcare Medical (Huizhou) Ltd. in China, implying the data likely originated from their testing facilities or a contracted lab. The study is inherently prospective in the sense that controlled laboratory tests were conducted on the device for the purpose of this submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
- Not Applicable. This is a device performance test (aerosol characteristics) for a hardware device, not an AI/ML study requiring expert annotations for ground truth.
4. Adjudication Method for the Test Set:
- Not Applicable. See point 3.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No. This is not an AI/ML medical device study.
6. Standalone (Algorithm Only) Performance:
- Not Applicable. This is a hardware device, not an algorithm.
7. Type of Ground Truth Used:
- Laboratory Measurements / Defined Standards. The "ground truth" for the device's performance is established by the results of the aerosol characterization tests (e.g., measuring particle size, total mass) conducted according to recognized industry standards (USP, ISO, FDA guidance). The comparison to the predicate device validates these measurements against an already cleared product.
8. Sample Size for the Training Set:
- Not Applicable. This device does not have a "training set" in the context of AI/ML.
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. See point 8.
Summary of what the document does provide:
The document details the physical and performance characteristics of a Nebulizer Kit and compares them to a legally marketed predicate device. The "acceptance criteria" were met by demonstrating "not significant statistical difference" in critical aerosol characteristics (Total Mass, MMAD, GSD, Respirable Fraction, Respirable Mass) at two different flow rates for four different drugs, when compared to the predicate device. This approach is standard for 510(k) submissions of non-AI/ML devices to establish substantial equivalence.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Excellentcare Medical (Huizhou) Ltd. % Elly Xu Consultant Manager Shenzhen Joyantech Consulting Co., Ltd. NO. 55 Shizhou middle road, Nanshan District Shenzhen, Guangdong 518000 China
Re: K192971
Trade/Device Name: Nebulizer Kit Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: May 26, 2021 Received: June 10, 2021
Dear Elly Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon L. Blakely, Ph.D. Assistant Director (Acting) DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K192971
Device Name Nebulizer Kit
Indications for Use (Describe)
The Nebulizer kit is used to aerosolize medication approved for nebulization and prescribed by a physician for delivery to the airways. The Nebulizer kit is intended for adult consistent with the indication in hospitals and sub-acute institutions.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
510k number: K192971
1. Administrative Information
| Date of Summary prepared | May 26, 2021 |
|---|---|
| Manufacturer information | Company name:Excellentcare Medical (Huizhou) Ltd. |
| Company address:Shatou Industrial Zone. Yuanzhou Town. 516123Huizhou Guangdong, China | |
| Contact person: Chunyu Zhao | |
| Phone: +86-752-6358312 | |
| Fax: +86-752-6358899E-mail: bradleyzhao@aliyun.com | |
| Submission Correspondent | Shenzhen Joyantech Consulting Co., Ltd.Address: 1713A, 17th Floor, Block A, ZhongguanTimes Square, Nanshan District, Shenzhen |
| Contact person: James Tsai | |
| E-Mail: james_tsai@cefda.com; | |
| field@cefda.com | |
| Establishment registrationnumber | |
| 2. Device Information | |
| Type of 510(k) submission: | Traditional |
| Trade Name: | Nebulizer Kit |
| Classification name: | Nebulizer |
| Classification: | II |
| Review Panel: | Anesthesiology |
| Product Code: | CAF |
| Regulation Number: | 868.5630 |
| 3. Predicate Device Information | |
| Trade Name | Disposable hand held nebulizer (VixOne) |
| 510(k) Number | K800562 |
| Classification name | Nebulizer |
| Classification: | II |
| Review Panel: | Anesthesiology |
| Product code | CAF |
| Regulation No. | 868.5630 |
{4}------------------------------------------------
4. Device Descriptions
The nebulizer kit is a small volume jet nebulizer designed to deliver aerosolized medications for inhalation to the respiratory system, the compressed air or oxygen is forced into the nebulizer through a converging nozzle, where it accelerates and emerges at a high velocity, creating a vacuum (Venturi effect), in the process the liquid medicine is converted into small droplets called aerosol by impacting upon a rigid baffle, the aerosol may be used for inhalation treatment of a physician's prescription. The nebulizer kit is single-patient, multi-use; it requires a cleaning and disinfection process before each use, and the service life is for 30 aerosol treatments.
5. Intended Use/ Indications for Use
The Nebulizer kit is used to aerosolize medication approved for nebulization and prescribed by a physician for delivery to the airways. The Nebulizer kit is intended for adult patients consistent with the indication for aerosol medication in hospitals and sub-acute institutions.
6. Comparisons to predicate device
The subject Nebulizer kit and VixOne™ Nebulizer (K800562) are similar in purpose, function, core technology and method of operation, the following table provides a comparison of the two devices.
| Items | Proposed Device | Predicate Device (K800562) | Remark | |
|---|---|---|---|---|
| Intendeduse/Indication for Use | The nebulizer kit is designed toaerosolize medication approvedfor nebulization and prescribedby a physician for delivery to theairways. The Nebulizer kit isintended for adult consistent withthe indication for aerosolmedication in hospitals and sub-acute institutions. | A handheld, pneumatic nebulizerdesigned to aerosolizeprescription drugs for inhalationby a patient. Its use is indicatedwhenever a healthcareprofessional administers orprescribes medical aerosolproducts to a patient using asmall volume nebulizer | Same | |
| Technology | Pneumatic Jet Nebulizer | Pneumatic Jet Nebulizer | Same | |
| Operationprinciple | Venturi | Venturi | Same | |
| Environment of use | Hospital, clinic | Hospital, clinic | Same | |
| PatientPopulation | Adult | Adult and pediatric | Similar | |
| Singlepatient useonly | Yes | Yes | Same | |
| AnatomicSite | Mouth | Mouth | Same | |
| Components | Nebulizer, A-type mouth piece,T-piece, Corrugated tube, andAir tube | Nebulizer, Mouth piece, T-piece,Corrugated tube, and Air tube | Same | |
| Items | Proposed Device | Predicate Device (K800562) | Remark | |
| Nebulizer(integral) | Same | |||
| Photos ofcomponents | Nebulizer(component) | Same | ||
| Nebulizerkit | Same | |||
| Aerosolization | Continuous during inhalationand exhalation | Continuous during inhalationand exhalation | Same | |
| Type ofdevice | Disposable, prescription only,non-sterile | Disposable, prescription only,non-sterile | Same | |
| Flow rate | 4-8L/min | 4-10L/min | Different | |
| MaximumFill Volume | 12 mL | 10mL | Different | |
| Materials | Polypropylene (PP),Polystyrene (PS),Polyethylene (PE),Ethylene-vinyl acetatecopolymer (EVA),Polyvinyl chloride (PVC) | Not publicly available | Different | |
| Type of gassource | Compressed air or oxygen | Compressed air or oxygen | Same | |
| Biocompatibility | Conform to FDA guidance ofBiological evaluation of medicaldevices | Conform to FDA guidance ofBiological evaluation of medicaldevices | Same | |
| Aerosolparticlecharacteristics | The aerosol particle characteristics tests are carried out for theproposed device and predicate device, the parameters of total mass,MMAD, GSD, respirable fraction and respirable Mass are analyzedand compared at a 95% confidence level, there is not significantstatistical difference between the two devices. | Different |
{5}------------------------------------------------
{6}------------------------------------------------
The proposed device of Nebulizer kit and the predicated device of VixOne™ Nebulizer are similar in purpose, function, core technology and method of operation, both the proposed device and predicated device have the similar patient population, type of gas source, biocompatibility performance, and there is not significant statistical difference in aerosol particles characteristics between nebulizer kit and VixOne. Only minor differences exist between the proposed device and predicate device.
7. Brief discussions of the nonclinical tests
The proposed device conforms to the following standards:
7.1 Biocompatibility evaluation
The biocompatibility evaluations of the proposed device were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" and FDA biocompatibility guidance, the proposed devices will be classified as external communication devices-tissue contact, and considering the cumulative exposure, the contact time will be permanent. The testing standards include the following:
-ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
-ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization
-ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
-ISO 10993-3: 2014 Biological evaluation of medical devices-Part 3: tests for genotoxicity, carcinogenicity and reproductive toxicity
-ISO 10993-6: 2016 Biological evaluation of medical devices - Part 6: Tests for local effects after implantation
-ISO 10993-17: 2002 Biological evaluation of medical devices-Part 17: Establishment of allowable limits for leachable substances
-ISO 10993-18: 2020 Biological evaluation of medical devices-Part 18: Chemical characterization of medical device materials within a risk management process
7.2 Dry Gas Pathway Testing
Testing pertaining to the dry gas pathway and associated risk assessments/conclusions were conducted by an independent source. Testing included the assessments of Emissions of volatile organic compounds (VOCs), Fine particles (particulate matter PM2.5). The testing standards include the following:
-ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
-ISO 18562-2: 2017 Tests for emissions of volatile Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
-ISO 18562-3: 2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds (VOCs)
7.3 Performance data
Aerosol characterization tests for the proposed devices and predicate devices were conducted in accordance with the relevant sections of the CDRH Guidance Document "Reviewer Guidance for Nebulizer, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH-1993), USP 40<601> Aerosols, Nasal Sprays, Metered-Dose Inhalers, and Dry
{7}------------------------------------------------
Powder Inhalers, and ISO 27427:2013 Anaesthetic and respiratory equipment-Nebulizing systems and components. For each device, four drugs: Albuterol sulfate (2.5mg/3.0ml), lpratropium bromide (0.5mg/2.5ml), Cromolyn sodium (20mg/2ml) and Budesonide suspension (1mg/2ml) were tested respectively. The comparison items include Mass Median Aerodynamic Diameter (MMAD), Geometric Standard Deviation (GSD), Respirable mass, Respirable Fraction and Total Mass. Also the variability of intra-sample and inter-sample was tested for the proposed devices.
| Aerosolcharacteristics | Aerosol particle characteristics | |
|---|---|---|
| Proposed Device (K192971) | Predicate Device (K800562) | |
| Total Mass (µg) | 1146.80± 142.12 (Albuterol sulfate)228.18±12.40 (Ipratropiumbromide)6307.49±802.07 (Cromolynsodium)467.07±34.11 (BudesonideSuspension) | 1309.33± 137.13 (Albuterol sulfate)213.81±15.03 (Ipratropiumbromide)7172.85±626.18 (Cromolynsodium)501.02±23.65 (BudesonideSuspension) |
| Particle Size-MMAD (µm) | 2.80± 0.23 (Albuterol sulfate)2.14±0.22 (Ipratropium bromide)2.66±0.20 (Cromolyn sodium)3.32±0.18 (BudesonideSuspension) | 2.55± 0.15 (Albuterol sulfate)2.07±0.20 (Ipratropium bromide)2.41±0.15 (Cromolyn sodium)3.11±0.18 (BudesonideSuspension) |
| GeometricStandardDeviation (GSD) | 2.49± 0.16 (Albuterol sulfate)2.81±0.06 (Ipratropium bromide)2.48±0.09 (Cromolyn sodium)2.27±0.08 (BudesonideSuspension) | 2.51± 0.05 (Albuterol sulfate)3.01±0.16 (Ipratropium bromide)2.56±0.09 (Cromolyn sodium)2.20±0.12 (BudesonideSuspension) |
| Respirablefraction (% 0.5-5µm) | 70.76± 1 (Albuterol sulfate)71.32±3 (Ipratropium bromide)72.33±3 (Cromolyn sodium)68.09±2 (Budesonide Suspension) | 73.02± 2 (Albuterol sulfate)68.86±1 (Ipratropium bromide)73.52±3 (Cromolyn sodium)71.82±4 (Budesonide Suspension) |
| Respirable Mass(µg 0.5-5) | 810.89± 94.73 (Albuterol sulfate)162.63±10.34 (Ipratropiumbromide)4576.38±739.65 (Cromolynsodium)318.22±27.49 (BudesonideSuspension) | 956.45± 107.82 (Albuterol sulfate)147.16±11.58 (Ipratropiumbromide)5273.70±512.70 (Cromolynsodium)359.25±7.21 (BudesonideSuspension) |
7.3.1 Aerosol particle characteristics (supplied air flow rate for nebulizer: 8L/min)
| 7.3.2 Aerosol particle characteristics (supplied air flow rate for nebulizer: 4L/min) | ||
|---|---|---|
| Aerosolcharacteristics | Aerosol particle characteristics | |
| Proposed Device (K192971) | Predicate Device (K800562) | |
| Total Mass (µg) | 1031.51± 113.44 (Albuterol sulfate) | 1104.12± 72.02 (Albuterol sulfate) |
| 199.78±15.60 (Ipratropiumbromide) | 216.84±10.63 (Ipratropiumbromide) | |
| 5466.01±567.02 (Cromolynsodium) | 5987.10±715.04 (Cromolynsodium) | |
| 430.79±18.55 (BudesonideSuspension) | 452.60±22.52 (BudesonideSuspension) | |
| Particle Size-MMAD (µm) | 4.71±0.11 (Albuterol sulfate) | 4.52± 0.13 (Albuterol sulfate) |
| 3.21±0.17 (Ipratropium bromide) | 3.19±0.24 (Ipratropium bromide) |
{8}------------------------------------------------
| 4.25±0.18 (Cromolyn sodium) | 4.09±0.10 (Cromolyn sodium) | |
|---|---|---|
| 4.97±0.14 (Budesonide | 4.84±0.15 (Budesonide | |
| Suspension) | Suspension) | |
| GeometricStandardDeviation (GSD) | 2.42± 0.05 (Albuterol sulfate) 2.76±0.16 (Ipratropium bromide) 2.52±0.01 (Cromolyn sodium) 2.53±0.13 (Budesonide Suspension) | 2.40± 0.07 (Albuterol sulfate) 2.71±0.11 (Ipratropium bromide) 2.48±0.05 (Cromolyn sodium) 2.59±0.15 (Budesonide Suspension) |
| Respirablefraction (% 0.5-5μm) | 52.17± 1 (Albuterol sulfate) 63.63±3 (Ipratropium bromide) 55.96±2 (Cromolyn sodium) 49.57±1 (Budesonide Suspension) | 54.06±1 (Albuterol sulfate) 64.3±3 (Ipratropium bromide) 57.79±1 (Cromolyn sodium) 50.55±0.1 (Budesonide Suspension) |
| Respirable Mass(µg 0.5-5) | 538.94± 70.97 (Albuterol sulfate) 126.97±10.16 (Ipratropium bromide) 3056.41±297.19 (Cromolyn sodium) 213.50±8.65 (Budesonide Suspension) | 596.14± 35.20(Albuterol sulfate) 139.41±11.84 (Ipratropium bromide) 3460.11±416.01 (Cromolyn sodium) 228.69±9.93 (Budesonide Suspension) |
7.4 Cleaning and Disinfection Validation
Microbiological efficiency control tests were conducted in order to validate the nebulizer cleaning and disinfection methods in the IFU. Testing involved validation of a manually cleaning method and a chemical disinfection method. All testing concluded that the nebulizer could be cleaned and disinfected effectively by use of the methods stated in the IFU. The references include the following:
-FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling.
-ASTM E1837-96 (Reapproved 2014): Standard Test Method to Determine Efficacy of Disinfection Processes for Reusable Medical Devices.
8. Conclusions
Based on comparison information and testing data, including biocompatibility tests and aerosol particle performance tests, the nebulizer kit meets the requirements of its pre-defined acceptance criteria and intended use. Therefore, the proposed device is substantially equivalent to the predicate devices.
§ 868.5630 Nebulizer.
(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).