K Number
K192967

Validate with FDA (Live)

Date Cleared
2020-03-25

(154 days)

Product Code
Regulation Number
888.3660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Medacta Shoulder System - Reverse

Reverse Shoulder Prosthesis

The Reverse Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with a grossly rotator cuff deficient shoulder joint, severe arthropathy failed joint replacement and a grossly rotator cuff deficient shoulder joint.

The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

The glenoid baseplate is intended for cementless application of screws for primary stability.

Short Humeral Diaphysis

The Reverse Shoulder Prosthesis- Short Humeral Diaphysis is indicated for primary total shoulder replacement in patients with grossly deficient rotator cuff shoulder joint with severe arthropathy.

The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

The glenoid baseplate is intended for cementless application with the addition of screws for fixation.

Medacta Shoulder System - Anatomic

Anatomic Shoulder Prosthesis

The Medacta Anatomic Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy or a previously failed joint replacement.

The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary for the device to offer full function in vivo.

The glenoid component is intended for cemented application.

Short Humeral Diaphysis

The Medacta Anatomic Shoulder Prosthesis – Short Humeral Diaphysis is indicated for primary total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary for the device to offer full function in vivo.

The glenoid component is intended for cemented application.

Device Description

The Medacta Shoulder Systems are modular systems intended to be used for shoulder arthroplasty (anatomical or reverse). System components were previously cleared by the FDA under K170910 (anatomic shoulder prosthesis) and K170452 (reverse shoulder prosthesis).

The Long Humeral Diaphysis implants, subject of this 510(k), are implantable devices used to replace the humeral side of the gleno-humeral joint. The product is intended to be used with the Medacta Shoulder System components as an alternative to the Standard Humeral Diaphysis components provided with those systems.

The Medacta Shoulder System Long Humeral Diaphysis couples with the Humeral Reverse Metaphysis (K170452) in the reverse configuration and the cemented and cementless Humeral Anatomic Metaphysis (K170910) in the anatomic configuration. The long humeral diaphysis, provided in 160 mm and 200 mm length options, may be used when additional distal stability in the humeral canal is needed.

The subject devices are manufactured from titanium alloy (Ti6A17Nb) and are provided sterile in 11 sizes per length option.

AI/ML Overview

This device is a medical implant, not an AI/ML device, therefore, the requested information regarding acceptance criteria and study details related to AI/ML performance is not applicable. The provided text is an FDA 510(k) clearance letter and summary for a physical medical device: the Medacta Shoulder System. This document focuses on demonstrating substantial equivalence to a predicate device through material, design, and performance testing, which are different from the criteria typically found in AI/ML device evaluations.

However, I can provide a summary of the performance data that was provided in support of the substantial equivalence determination for this medical device, as described in the document.

Performance Data Provided:

The 510(k) summary lists the following performance data:

  • Sterilization Validation: Per UNI EN ISO 11137-1:2015 and AAMI/ANSI/ISO 11137-2:2013. This ensures the device can be properly sterilized for safe use.
  • Fatigue Testing: Per ASTM F1378-17. This evaluates the device's mechanical durability and its ability to withstand repeated stress over time, simulating conditions within the human body.
  • Cadaver Studies: These studies typically assess surgical technique, implant fit, and preliminary biomechanical performance within an anatomical context.

Please note that none of the specific AI/ML related questions (sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance) are relevant to the information presented in this 510(k) summary for the Medacta Shoulder System.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medacta International, SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street MEMPHIS, TN 38118

Re: K192967

Trade/Device Name: Medacta Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, HSD, KWS, KWT Dated: January 10, 2020 Received: January 13, 2020

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

March 25, 2020

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192967

Device Name Medacta Shoulder System

Indications for Use (Describe) Medacta Shoulder System - Reverse

Reverse Shoulder Prosthesis

The Reverse Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with a grossly rotator cuff deficient shoulder joint, severe arthropathy failed joint replacement and a grossly rotator cuff deficient shoulder joint.

The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

The glenoid baseplate is intended for cementless application of screws for primary stability.

Short Humeral Diaphysis

The Reverse Shoulder Prosthesis- Short Humeral Diaphysis is indicated for primary total shoulder replacement in patients with grossly deficient rotator cuff shoulder joint with severe arthropathy.

The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

The glenoid baseplate is intended for cementless application with the addition of screws for fixation.

Medacta Shoulder System - Anatomic

Anatomic Shoulder Prosthesis

The Medacta Anatomic Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy or a previously failed joint replacement.

The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary for the device to offer full function in vivo.

The glenoid component is intended for cemented application.

Short Humeral Diaphysis

The Medacta Anatomic Shoulder Prosthesis – Short Humeral Diaphysis is indicated for primary total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary for the device to offer full function in vivo.

The glenoid component is intended for cemented application.

{3}------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

2.0 510(k) Summary

I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Sr. Director of Quality and Regulatory, Medacta USA Date Prepared: October 21, 2019 Date Revised: March 18, 2020

II. Device

Device Proprietary Name:Medacta Shoulder System
Common or Usual Name:Shoulder Prosthesis System
Classification Name:Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulation Number:21 CFR 888.3660
Product Code (primary):PHX
Product Code (secondary):HSD, KWS, KWT
Device Classification2

III. Predicate Device

Substantial equivalence is claimed to the following primary predicate device:

  • Zimmer® Trabecular Metal™ Reverse Shoulder System, Non-Porous Humeral Stems, o K122692, Zimmer Incorporated
    The following devices are referenced within the submission:

  • Medacta Anatomic Shoulder Prosthesis, K170910, Medacta International SA ●

  • Medacta Shoulder System, K170452, Medacta International SA ●

  • o Global Unite Anatomic Shoulder, K133834, Depuy

IV. Device Description

The Medacta Shoulder Systems are modular systems intended to be used for shoulder arthroplasty (anatomical or reverse). System components were previously cleared by the FDA under K170910 (anatomic shoulder prosthesis) and K170452 (reverse shoulder prosthesis).

{5}------------------------------------------------

The Long Humeral Diaphysis implants, subject of this 510(k), are implantable devices used to replace the humeral side of the gleno-humeral joint. The product is intended to be used with the Medacta Shoulder System components as an alternative to the Standard Humeral Diaphysis components provided with those systems.

The Medacta Shoulder System Long Humeral Diaphysis couples with the Humeral Reverse Metaphysis (K170452) in the reverse configuration and the cemented and cementless Humeral Anatomic Metaphysis (K170910) in the anatomic configuration. The long humeral diaphysis, provided in 160 mm and 200 mm length options, may be used when additional distal stability in the humeral canal is needed.

The subject devices are manufactured from titanium alloy (Ti6A17Nb) and are provided sterile in 11 sizes per length option.

V. Indications for Use

Medacta Shoulder System – Reverse

Reverse Shoulder Prosthesis

The Reverse Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with a grossly rotator cuff deficient shoulder joint, severe arthropathy or a previously failed joint replacement and a grossly rotator cuff deficient shoulder joint.

The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

The glenoid baseplate is intended for cementless application with the addition of screws for primary stability.

Short Humeral Diaphysis

The Reverse Shoulder Prosthesis- Short Humeral Diaphysis is indicated for primary total shoulder replacement in patients with grossly deficient rotator cuff shoulder joint with severe arthropathy.

The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.

The glenoid baseplate is intended for cementless application with the addition of screws for fixation.

{6}------------------------------------------------

Medacta Shoulder System - Anatomic

Anatomic Shoulder Prosthesis

The Medacta Anatomic Shoulder Prosthesis is indicated for treatment of humeral fractures and for primary or revision total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint, severe arthropathy or a previously failed joint replacement.

The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary for the device to offer full function in vivo.

The glenoid component is intended for cemented application.

Short Humeral Diaphysis

The Medacta Anatomic Shoulder Prosthesis – Short Humeral Diaphysis is indicated for primary total shoulder replacement in patients with an intact or reparable rotator cuff shoulder joint. severe arthropathy. The patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary for the device to offer full function in vivo.

The glenoid component is intended for cemented application.

VI. Comparison of Technological Characteristics

The Medacta Shoulder System Long Humeral Diaphysis and the predicate device share the following characteristics:

  • material of construction;
  • range of sizes;
  • range of lengths; ●
  • fixation; and ●
  • distal shaft shape.

There are no technological differences between the subject and predicate device with respect to the Long Humeral Diaphysis component.

A comparison of key technological features between the subject and predicate device is provided below.

{7}------------------------------------------------

Technological comparison

ParameterSubject DeviceZimmer® TrabecularMetal™ Reverse ShoulderSystem, Non-PorousHumeral Stems (K122692)
Material ofconstructionTitanium alloyTitanium alloy
FixationCementlessCementedCementlessCemented
Distal shaft shapeCircular section, longitudinalgroovesCircular section, longitudinalgrooves

VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination:

  • o sterilization validation per UNI EN ISO 11137-1:2015 and AAMI/ANSI/ISO 11137-2:2013;
  • fatigue testing per ASTM F1378-17 and ●
  • cadaver studies. ●

VIII. Conclusion

The information provided within this submission supports that the Medacta Shoulder Systems are as safe and effective as the predicate device; therefore, the Medacta Shoulder Systems are substantially equivalent to the predicate device.

§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”