K Number
K192939

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-12-27

(70 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Volumetric Specimen Imager is a Cabinet x-ray system that is used to provide two and three dimensional digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

Device Description

The Clarix Imaging Volumetric Specimen Imager (VSI) is a portable, fully shielded cabinet X-ray system that provides high resolution 2D and 3D radiographic images of surgically excised specimens. The VSI system consists of an image-acquisition device and software for image reconstruction, visualization, and archival. Enabled by optimized imaging geometry and iterative image reconstruction algorithm, VSI provides superior spatial and contrast resolution for visualizing specimens with soft tissues or boney structures. VSI is designed to comply with 21 CFR 1020.40 and DICOM standards.

AI/ML Overview

The provided text is a 510(k) summary for the Clarix Imaging Volumetric Specimen Imager (VSI). It describes the device, its intended use, and a comparison with a predicate device. However, it explicitly states that clinical data was not required to demonstrate substantial equivalence to the predicate device. Therefore, the document does not contain the detailed information requested regarding acceptance criteria and a study proving the device meets those criteria through clinical performance.

The "Non-clinical Performance Data" section lists several tests performed, but these are primarily for electrical safety, electromagnetic compatibility, compliance with regulations and standards (IEC, CFR, DICOM), phantom imaging, and user testing. The "Tissue Verification" listed here would likely refer to non-clinical verification of image quality for tissue, not a clinical study involving human patients or complex diagnostic assessments.

Given the information provided in the 510(k) summary, I cannot fill out the requested table or describe a study that proves the device meets specific acceptance criteria based on clinical performance. The 510(k) pathway for this device relied on demonstrating substantial equivalence to a predicate, not on new clinical performance data to establish specific diagnostic accuracy or efficacy metrics.

Therefore, many of the requested fields about ground truth, expert readers, MRMC studies, and effect sizes cannot be answered from this document. If such a study were performed, its details would be found in a separate clinical study report, not typically summarized in this manner for a 510(k) submission where clinical data was explicitly not required for substantial equivalence.

{0}------------------------------------------------

December 27, 2019

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Clarix Imaging % Xiao Han, Ph.D. CEO 2242 West Harrison Street, Suite 201 CHICAGO IL 60612

Re: K192939

Trade/Device Name: Volumetric Specimen Imager Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MWP Dated: October 16, 2019 Received: October 18, 2019

Dear Dr. Han:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192939

Device Name Volumetric Specimen Imager

Indications for Use (Describe)

The Volumetric Specimen Imager is a Cabinet x-ray system that is used to provide two and three dimensional digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K192939(As required by 21 CFR 807.92)
Date Summary Prepared:October 16, 2019
Company Name:As required by 807.92(a)(1)Clarix Imaging Corporation
Xiao Han
CEO, Clarix Imaging
2242 West Harrison Street Suite 201
Chicago, IL 60612
(872) 760-3788
xiao.han@clariximaging.com
Device Name:As required by 807.92(a)(2)Device/Trade Name:Volumetric Specimen Imager
Device Common Name:Specimen X-ray System/Cabinet X-ray System
Regulation Number:21 CFR 892.1680
Regulation Name:Stationary X-ray System
Class:II
Product Code:MWP
Predicate Device(s):As required by 807.92(a)(3)Faxitron VisionCT, K173309
Device Description:As required by 807.92(a)(4)The Clarix Imaging Volumetric Specimen Imager (VSI) is aportable, fully shielded cabinet X-ray system that provideshigh resolution 2D and 3D radiographic images of surgicallyexcised specimens. The VSI system consists of an image-acquisition device and software for image reconstruction,visualization, and archival. Enabled by optimized imaginggeometry and iterative image reconstruction algorithm, VSIprovides superior spatial and contrast resolution forvisualizing specimens with soft tissues or boney structures.VSI is designed to comply with 21 CFR 1020.40 and DICOMstandards.
Statement of Intended Use:As required by 807.92(a)(5)The Volumetric Specimen Imager is a Cabinet x-ray systemthat is used to provide two and three dimensional digital x-ray images of harvested specimens from various anatomicalregions in order to provide rapid verification that the correcttissue has been excised during the biopsy procedure. Doingthe verification directly in the same room or nearby enables

{4}------------------------------------------------

cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.

Comparison of Technological Characteristics with Predicate Devices: As required by 807.92(a)(6)

As required by 807.92(a)(b)
DeviceProposed DeviceVolumetric Specimen ImagerPredicate DeviceFaxitron VisionCT, K173309
Detector
• Array Size• 1488 x 1148• 1548 x 1548
• Pixel Size• 99 x 99 μm• 99 x 99 μm
• Resolution• 5 lp/mm• 5 lp/mm
X-ray Tube
• Voltage Range• 4-60 kV• 15-50 kV
• Current• Up to 1.0 mA• Up to 1.0 mA
• Target Material• Tungsten• Tungsten
Exposure
• Time Range• Up to 300 seconds• 0 to 900 seconds

The proposed and predicate device only differ in minor technological details sourced from the devices using different x-ray tubes and detectors. Overall both the proposed and predicate devices have the specifications necessary to deliver imaging safely and effectively.

Non-clinical Performance Data:As required by 807.92(b)(1)Non-clinical performance testing was performed to verify substantial equivalence to predicate device, Faxitron VisionCT (K173309). Testing included:• Electrical safety: IEC 61010-1, IEC 61010-2-091, IEC 61010-2-101
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
  • Electromagnetic compatibility: IEC 61326-1, IEC ● 61326-2-6
  • Compliance with 21 CFR 1020.40 ●
  • Compliance with DICOM Standards ●
  • Phantom imaging ●
  • User testing ●
  • Tissue Verification .

Clinical data was not required to demonstrate substantial Assessment of Clinical Data: As required by 807.92(b)(2) equivalence to predicate device.

for Overall Conclusions: Based on the indications use, technological As required by 807.92(b)(3) characteristics, and performance results, the Volumetric Specimen Imager has been shown to be substantially equivalent to the predicate and is safe and effective for its intended use.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.