K Number
K192935

Validate with FDA (Live)

Date Cleared
2019-11-15

(29 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CAPERE® Thrombectomy System is indicated for:

  • Non-surgical removal of soft emboli and thrombi from blood vessels.
  • Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel.
    The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
Device Description

The CAPERE® Thrombectomy System primarily consists of an 8Fr Delivery Catheter and Funnel Catheter. The CAPERE® Thrombectomy System is delivered percutaneously via transfemoral or jugular venous access. Once delivered, the System's fine mesh nitinol wire basket is used to capture and mechanically remove emboli and thrombi. The CAPERE® System does not use aspiration to pull out the thrombus but does have a side port in the funnel catheter that allows aspiration or injection of saline or fluids if needed.

AI/ML Overview

The provided text describes the 510(k) clearance for the CAPERE Thrombectomy System. This regulatory submission focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving a specific performance acceptance criteria against a medical outcome (like 'improving diagnostic accuracy'). Therefore, many of the requested categories for AI/diagnostic devices (e.g., multireader-multicase studies, expert consensus for ground truth on a training set) are not applicable to this type of device clearance.

The "acceptance criteria" in this context refer to the device meeting engineering and functional performance specifications, and demonstrating that changes from the predicate device do not introduce new safety or effectiveness concerns.

Here's an analysis based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from text)Reported Device Performance
Compliance with relevant standards (e.g., ISO 10555-1, ISO 594-2) for bond joints affected by increased length.Demonstrated compliance; maintained predicate characteristics.
Simulated clot retrieval without capture basket rupture.Functioned as intended; simulated clot retrieved without rupture.
Simulated clot retrieval without catheter damage.Functioned as intended; no catheter damage.
Simulated clot retrieval without other adverse device effect.Functioned as intended; no other adverse device effects.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not specified in terms of numerical units (e.g., N devices tested). The document states "functional bench testing was conducted" and "Simulated Use Testing was completed." It refers to "the CAPERE® Thrombectomy System" in the singular, but implies testing was performed sufficiently to make general claims. It does not provide the number of devices or iterations tested.
  • Data Provenance: The tests are described as "nonclinical performance testing" and "functional bench testing" / "Simulated Use Testing." This indicates laboratory-based, simulated data, not data from human patients or a specific country of origin. The data is implicitly prospective in the sense that the tests were designed and conducted to evaluate the device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • Not Applicable. This is not a study involving human interpretation of data where expert ground truth is established (e.g., radiology image interpretation). The "ground truth" for these tests is defined by the engineering specifications and the ability of the device to perform its mechanical functions in a simulated environment.

4. Adjudication Method for the Test Set

  • Not Applicable. No human adjudication method (like 2+1, 3+1) is relevant for this type of mechanical and functional performance testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. This is a mechanical device, not an AI/diagnostic software. MRMC studies are not relevant for its clearance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not Applicable. This is a hardware medical device, not an algorithm. Its performance is inherent to its design and function. However, the "Simulated Use Testing" could be considered analogous to "standalone performance" in the sense that it evaluates the device's function without a human patient or a full clinical scenario, but within a simulated environment.

7. The Type of Ground Truth Used

  • Engineering Specifications and Functional Performance Benchmarks: The ground truth for this device's testing is primarily based on predefined engineering and functional performance criteria. For example, "without capture basket rupture," "without catheter damage," and "maintained predicate characteristics" are the benchmarks.

8. The Sample Size for the Training Set

  • Not Applicable. This device is a mechanical thrombectomy system, not an AI/machine learning algorithm that requires a "training set" in the computational sense. The device's design and manufacturing process are informed by engineering principles, not data-driven training.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no training set for an AI algorithm, this question is not relevant. The "ground truth" for the device's design (if you were to interpret it broadly as the basis for its creation) would be established through established medical device design principles, biomechanical understanding of fluid dynamics and clot removal, and experience with predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The FDA logo is in blue.

November 15, 2019

Vascular Medcure, Inc. % Ms. Janice Hogan Regulatory Counsel Hogan & Lovells US LPP 1735 Market Street. Suite 2300 Philadelphia, Pennsylvania 19103

Re: K192935

Trade/Device Name: CAPERE Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEY Dated: October 17, 2019 Received: October 17, 2019

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192935

Device Name CAPERE Thrombectomy System

Indications for Use (Describe) The CAPERE Thrombectomy System is indicated for:

· Non-surgical removal of soft emboli and thrombi from blood vessels.

  • · Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel.
    The CAPERE Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Vascular Medcure, Inc.'s CAPERE® Thrombectomy System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Jeffrey P. DuMontelle Vascular Medcure, Inc. 1500 S. Sunkist Street, Suite H Anaheim, CA 92806 Office: 657-549-5175 714-915-0886 Cell: Date Prepared: November 12, 2019

Name of Device

CAPERE® Thrombectomy System

Common or Usual Name

Embolectomy Catheter

Classification

21 CFR 870.5150, Class II, product code QEY

Predicate Devices

CAPERE® Thrombectomy System (K183179)

Intended Use / Indications for Use

The CAPERE® Thrombectomy System is indicated for:

  • Non-surgical removal of soft emboli and thrombi from blood vessels. O
  • Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. O

The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.

Device Description

The CAPERE® Thrombectomy System primarily consists of an 8Fr Delivery Catheter and Funnel Catheter. The CAPERE® Thrombectomy System is delivered percutaneously via transfemoral or jugular venous access. Once delivered, the System's fine mesh nitinol wire basket is used to capture and mechanically remove emboli and thrombi. The CAPERE® System does not use aspiration to pull out the thrombus but does have a side port in the funnel catheter that allows aspiration or injection of saline or fluids if needed.

Technological Characteristics

The CAPERE® Thrombectomy System has the same technological characteristics as its predicate device CAPERE Thrombectomy Catheter (K183179). Both CAPERE® Thrombectorny Systems have either the same components or substantially equivalent

{4}------------------------------------------------

components (length change). Each device system includes a funnel catheter, delivery catheter, and a mechanism for capturing/removing the soft emboli or thrombi. The predicate and CAPERE® Thrombectomy System both use an 8Fr funnel catheter and a 6Fr. delivery catheter. In addition, both systems use 0.014" guidewires during the procedure. The only change is an increase of 21cm in the working catheter length.

The CAPERE® System includes a Funnel Catheter and a Delivery Catheter. In the same manner as the previously cleared device, the Funnel Catheter and Delivery Catheter are advanced to the therapy site. The funnel of the funnel catheter is unsheathed, and the Delivery Catheter is then advanced passed the obstruction. A nitinol basket connected to the Delivery Catheter is deployed by retracting the outer sheath of the catheter.

To capture the soft emboli or thrombi ("clot"), the deployed basket is retracted while simultaneously being extended to surround and capture the length of the clot. The Delivery Catheter and clot-load is then retracted into the Funnel Catheter where the clot is captured and can be removed.

The basket design in the CAPERE® Thrombectomy Systems is exactly the same (nitinol, 10mm diameter, 11.4cm length) as the basket design in the predicate.

Therefore, the subject CAPERE® Thrombectorny System has the same technological characteristics as its predicate.

Performance Data

The following nonclinical performance testing has been conducted, using the established test methods used in the cleared predicate device, to support the substantial equivalence of the CAPERE® Thrombectomy System to its predicate device. In all instances, the CAPERE® Thrombectomy System functioned as intended under the previously established test methods.

  • . Functional bench testing was conducted (including demonstrated compliance with relevant standards such as ISO 10555-1 and ISO 594-2) on the bond joints affected by the increased length of the CAPERE® Thrombectomy System with respect to the predicate and demonstrated that the CAPERE® Thrombectomy System maintained the predicate characteristics.
  • . Simulated Use Testing was completed to demonstrate that simulated clot can be retrieved under simulated conditions without capture basket rupture, catheter damage, or other adverse device effect and confirmed that the predicate characteristics were maintained.

Substantial Equivalence

The CAPERE® Thrombectomy System has the same intended use and indications for use, technological characteristics, and principles of operation as its predicate devices. Nonclinical testing, including functional testing and simulated use testing demonstrated that the minor differences between the device and predicate (i.e. device length) do not raise new types of safety or effectiveness questions. Thus, the CAPERE® Thrombectomy System is substantially equivalent to the predicate devices.

Conclusion

Vascular Medcure's CAPERE® is an Embolectomy Catheter, Class II device that has been evaluated in nonclinical testing in accordance with FDA's recognized standards and pre

{5}------------------------------------------------

established acceptance criteria. Testing demonstrated that the device performs as intended. The CAPERE® is substantially equivalent to its predicate devices.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).