K Number
K192916

Validate with FDA (Live)

Date Cleared
2020-12-11

(423 days)

Product Code
Regulation Number
866.5100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NOVA Lite® DAPI dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the qualitative and/or semi-quantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator.

Device Description

The NOVA Lite DAPI dsDNA Crithidia luciliae Kit is an indirect immunofluorescence assay for the qualitative detection and semi-quantitative determination of Anti-dsDNA Antibodies (IgG) in human serum. Samples are diluted 1:10 in PBS and incubated with the antigen substrate (dsDNA on glass microscope slides). After incubation, unbound antibodies are washed off. The substrate is then incubated with antihuman IgG-FITC conjugate. The conjugate contains a DNA-binding blue fluorescent dye, 4',6-diamidino-2phenylindole (DAPI) that is required for NOVA View use. The blue dye is not visible by traditional fluorescence microscope at the wavelength where FITC fluorescence is viewed. Unbound reagent is washed off. Stained slides are read by manual fluorescence microscope or scanned with the NOVA View Automated Fluorescence Microscope. The resulting digital images are reviewed and interpreted from the computer monitor. dsDNA positive samples exhibit an apple green fluorescence corresponding to areas of the substrate where autoantibody has bound.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the NOVA Lite DAPI dsDNA Crithidia luciliae Kit, based on the provided text:

Acceptance Criteria and Reported Device Performance

Test/CharacteristicAcceptance CriteriaReported Device Performance
PrecisionReactivity Grades: Difference between reactivity grades within one run (between replicates) are within ± one reactivity grade. Average reactivity grade difference between any runs is within ± one reactivity grade.For digital image reading, grades were within ± one reactivity grade within one run (within triplicates), and the average grade was no more than one reactivity grade different between runs. (Numerical results provided in table: e.g., Sample 1: 93% positive, Grade range 3-4 for NOVA View; 100% positive, Grade 4 for Manual; 100% positive, Grade 4 for Digital).
Reproducibility (Between Sites/Instruments)Agreement: 90% agreement between operators and between sites.Manual Reading: Qualitative agreement: All samples showed 100% positive/negative agreement across both readers at all three sites for most samples. Sample 4 showed some variability (e.g., Reader 1 Site 1 was 7% negative for a positive sample, others 0%). Digital Reading: Qualitative agreement: All samples showed 100% positive/negative agreement across both readers at all three sites. Operator Agreement (per site): Manual Reading: 99.7% (Site 1), 100.0% (Site 2), 100.0% (Site 3). Digital Reading: 100.0% (Site 1), 100.0% (Site 2), 100.0% (Site 3).
Reproducibility (Between Lots)Qualitative Agreement: Positive, negative, and total agreement ≥ 90%. Grade Agreement: ≥ 90% within ± 1 reactivity grade.Qualitative Agreement: NOVA View: Positive agreement ranged from 91.7% to 100.0%. Negative agreement ranged from 96.4% to 100.0%. Total agreement ranged from 95.0% to 100.0%. Manual: 100% positive, negative, and total agreement. Digital: 92.9% positive agreement, 100% negative agreement, 97.5% total agreement. Grade Agreement: Manual: 100% within ±1 reactivity grade. Digital: 98% within ±1 reactivity grade.
LinearityNot explicitly stated as a pass/fail criterion, but the expectation is that dilutions will follow a predictable pattern.The results show a clear progression of intensity decrease with serial dilution for all three samples across NOVA View, Manual, and Digital interpretations, confirming linearity.
InterferenceGrades obtained on samples with interfering substances are within ± 1 reactivity grade of those obtained on the control samples, spiked with diluent.No interference was detected with hemoglobin (up to 200 mg/dL), bilirubin (up to 100 mg/dL), triglycerides (up to 1,000 mg/dL), cholesterol (up to 224.3 mg/dL), rheumatoid factor (up to 28.02 IU/mL), and various medications (azathioprine, cyclophosphamide, hydroxychloroquine, ibuprofen, methotrexate, methylprednisolone, mycophenolate, naproxen, rituximab, and belimumab) at specified concentrations.
Sample Stability and HandlingNOVA View: Results (positive/negative) do not change category and are not different than the control sample. Manual Reading: Reactivity grades are within ±1 grade of the control sample. Digital Image Interpretation: Reactivity grades are within ±1 grade of the control sample.All samples fulfilled the acceptance criteria at each time point (up to 21 days at 2-8°C, up to 48 hours at room temperature, and up to 3 freeze/thaw cycles) for each condition.
Reagent Stability (Shelf Life)Reactivity grades of all samples/reagent controls run must be within ±1 reactivity grade of the control condition (week 0) for both manual and digital image interpretation for all three lots.The acceptance criteria were successfully met with the accelerated lots tested for a two-year preliminary expiration dating. All samples tested were within ±1 reactivity grade of the control kit. Real-time stability results to date (up to 24, 15, and 19 months for different lots) were within acceptance limits.
Reagent Stability (In-use/Open Vial - Conjugate & Controls)Appearance: Clear liquid, free from foreign matter. Grades: Within ±1 grade from each other. Fluorescence Grading: >3+ for undiluted positive control, 0 for undiluted negative control. Testing: Comparable to control.The acceptance criteria were successfully met for all 8 weeks tested for both conjugate and controls.
Single Well Titer (SWT)Accuracy: SWT is within ± 2 dilution steps of that of the manual end-point titer and the digital titer.Based on 31 samples, 80.6% of SWT results were within ± 1 dilution step of the manual titer, and 83.9% were within ±1 dilution step of the digital titer. Furthermore, 93.3% of SWT results were within ± 2 dilution steps of the manual titer and 93.5% were within ± 2 dilution steps of the digital titer. (Note: 2 out of 31 samples were outside the ±2 dilution step range). Between sites reproducibility study: 100% of SWT results at two external sites were within ± 1 dilution step of the manual titer (14/14 samples), and 92.9% were within ± 1 dilution step of the digital titer (13/14 samples). 100% of SWT results were within ± 2 dilution steps of both manual and digital titers.

Study Details:

  1. Sample sizes used for the test set and the data provenance:

    • Precision Study: 6 samples (2 negative, 2 borderline, 2 positive), each processed in triplicate across 14 runs (2 runs/day for 7 days), resulting in 42 data points per sample.
    • Reproducibility Studies (Between sites/instruments): 10 samples (3 negative, 7 positive), each tested in triplicate, twice a day for 5 days at each of 3 sites. This results in 30 data points per sample per site, or 90 data points per sample across all sites. Total data points for this study: 10 samples * 30 data points/sample * 3 sites = 900 data points.
    • Reproducibility (Between lots): 20 clinically and/or analytically characterized samples, tested in duplicate.
    • Linearity Study: 3 positive samples (high, medium, low), serially diluted from 1:10 up to 1:5120. (Number of replicates not specified for this part, but results are given for each dilution).
    • Interference Study: 3 specimens (one negative, one positive, one strong positive) for each interferent, with interfering substances spiked at three different concentrations in 10% of total specimen volume. Samples assessed in triplicates.
    • Sample Stability and Handling: 3 samples (negative, cut-off, positive), tested in duplicates for various conditions (up to 21 days at 2-8°C, up to 48 hours at room temperature, up to 3 freeze/thaw cycles).
    • Reagent Stability (Shelf-life): 3 lots of the kit, tested over 4 weeks accelerated stability (each week = 6 months real time). Real-time stability data was available up to 24, 15, and 19 months for the respective lots at the time of submission.
    • Reagent Stability (In-use/Open Vial): Not detailed how many units/tests were performed each week for 8 weeks.
    • Clinical Performance (Initial Study): 766 clinically characterized serum samples (391 SLE, 375 other diseases). No explicit country of origin is stated, but given this is an FDA submission for Inova Diagnostics, Inc. in San Diego, California, it is reasonable to infer a US-centric data provenance. The study appears to be retrospective based on "clinically characterized serum samples."
    • Clinical Performance (3 Sites Study): 269 clinically characterized samples tested at three sites. Total for reporting: 100 positive (SLE) and 169 negative (non-SLE) per site. The samples comprise 300 SLE and 507 non-SLE clinical diagnoses in total across the three sites. The data provenance is likely multi-center, potentially within the US. The description "clinically characterized samples" suggests these were collected and diagnosed prior to the study, implying a retrospective nature.
    • Expected Values: 120 samples from apparently healthy subjects (60 females, mean age 41, range 18-73).
    • Comparison with Predicate Device: The same 744 serum samples used in the initial clinical study (391 SLE, 353 other diseases).
    • SWT Validation: 31 positive samples for initial validation. 7 positive samples in the between-sites reproducibility study.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For clinical studies: The ground truth for the clinical studies is stated as "clinically characterized serum samples" and "clinical diagnosis." This implies that the samples were obtained from patients with established diagnoses, likely made by medical professionals (e.g., rheumatologists for SLE patients). The text does not specify the number of experts or their exact qualifications (e.g., "Radiologist with 10 years of experience" is not mentioned, as this is an immunoassay, but rather "clinicians" or "diagnosticians").
    • For analytical studies (Precision, Reproducibility, Linearity, Interference, Stability): The ground truth (Expected Result/Expected Grade) for the control samples or known samples was established by the manufacturer, often based on previous characterization or established laboratory practices. The interpretation of "Manual Reading" and "Digital Reading" results are performed by "trained operators."
  3. Adjudication method for the test set:

    • For analytical results (Precision, Reproducibility, Linearity, Interference, Stability): The text mentions that "Digital images were interpreted and confirmed" in multiple sections (e.g., Linearity, Interference, Sample Stability). For the "Reproducibility Studies (Between sites/instruments)", manual and digital reading was performed by "two operators at each site, to assess between operator reproducibility." The acceptance criteria then focus on agreement percentages between operators. This implies that if disagreements occurred, they were likely adjudicated to reach the "Summary" percentages. However, a specific formal adjudication method like "2+1" or "3+1" is not explicitly stated.
    • For clinical results: The clinical samples were "clinically characterized," meaning their diagnosis served as the ground truth. There's no indication of an adjudication process for these clinical diagnoses within the context of this device study.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not a traditional MRMC comparative effectiveness study involving AI assistance improving human readers. The study compares three modes of interpretation:
      • Manual Reading: Human interpretation using a traditional fluorescence microscope.
      • Digital Reading: Human interpretation of NOVA View generated images on a computer monitor.
      • NOVA View (Software): Automated interpretation by the device's software (algorithm only), which is then confirmed by a trained operator.
    • Therefore, the setup is more of a comparison between manual microscopy, human interpretation of digital images, and the device's automated output. The device itself (NOVA View) is not presented as an AI-assistance tool for human readers but as an alternative interpretation method that still requires human confirmation.
    • The "effect size of how much human readers improve with AI vs without AI assistance" is not directly measured in this context because the "NOVA View" results are the algorithm's output, not a human reader assisted by the algorithm. The "Digital Reading" is human interpretation of the images produced by the NOVA View device, which might be considered an "assisted" or "different modality" reading but not in the typical AI-driven improvement sense.

    Let's look at sensitivity/specificity to show the comparison between manual, digital (human on digital images), and NOVA View (algorithm):

    Initial Clinical Study (N=766)

    • Sensitivity (on SLE):
      • Manual: 48.1% (43.2-53.0)
      • Digital: 48.1% (43.2-53.0)
      • NOVA View: 57.0% (52.1-61.8)
    • Specificity:
      • Manual: 91.2% (87.9-93.7)
      • Digital: 92.3% (89.1-94.6)
      • NOVA View: 88.8% (85.2-91.6)

    Clinical Studies 3 Sites (N=807)

    • Sensitivity (on SLE):
      • Manual Reading: 32.7% (27.6-38.2)
      • Digital Reading: 34.0% (28.9-39.5)
      • NOVA View: 40.0% (34.6-45.6)
    • Specificity:
      • Manual Reading: 95.5% (93.3-97.0)
      • Digital Reading: 95.5% (93.3-97.0)
      • NOVA View: 85.4% (82.1-88.2)

    In both clinical studies, the NOVA View algorithm demonstrates higher sensitivity for SLE detection compared to manual or digital human readings, but lower specificity. This highlights a performance difference, not an improvement of human readers with AI assistance.

  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the "NOVA View" performance reported in the tables (e.g., sensitivity, specificity, qualitative agreements in reproducibility studies) represents the standalone algorithm's performance.
    • The text explicitly states: "All results generated with NOVA View device must be confirmed by a trained operator." This indicates that while the software generates automated classifications, the final clinical interpretation includes a human-in-the-loop for confirmation. The reported performance metrics for "NOVA View" specifically reflect the device's automated output.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Clinical Ground Truth: "Clinical diagnosis" for patient-derived samples (e.g., Systemic Lupus Erythematosus (SLE), Drug Induced Lupus, Infectious Disease, etc.). This is likely based on a combination of clinical findings and serological tests by clinicians. It is stated as "clinically characterized serum samples."
    • Analytical Ground Truth: For the precision, reproducibility, linearity, interference, and stability studies, the "Expected Result" or "Expected Grade" for the tested samples serves as the ground truth. These are typically reference materials or well-characterized samples with known positive/negative status or reactivity grades established by the manufacturer.
  7. The sample size for the training set:

    • The document does not explicitly state the sample size of a training set for the NOVA View algorithm. It describes validation studies (test sets) for the kit and the performance of the NOVA View device, but not how the algorithm itself was developed or trained.
    • What is mentioned is that for the SWT (Single Well Titer) feature, "The SWT function was established using 22 dsDNA positive samples that represent various levels of antibodies." However, this refers to establishing the intensity curves for titer determination, not necessarily a broad 'training set' for the overall positive/negative classification logic or image analysis.
  8. How the ground truth for the training set was established:

    • As the training set size is not stated, neither is the method for establishing its ground truth.
    • For the 22 dsDNA positive samples used to establish SWT intensity curves, it is implied that manual and digital readings (human interpretations) served as comparison points for establishing the LIU (Light Intensity Units) to titer relationship. The validation of SWT compares its output to manual and digital end-point titers.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2020

Inova Diagnostics, Inc. Carolina Auza Supervisor, Research & Development 9900 Old Grove Rd San Diego, California 92131

Re: K192916

Trade/Device Name: NOVA Lite DAPI dsDNA Crithidia luciliae Kit Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear Antibody Immunological Test System Regulatory Class: Class II Product Code: KTL, PIV Dated: November 11, 2020 Received: November 12, 2020

Dear Carolina Auza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ying (Katelin) Mao, Ph.D. Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192916

Device Name NOVA Lite DAPI dsDNA Crithidia luciliae Kit

Indications for Use (Describe)

NOVA Lite® DAPI dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the qualitative and/or semiquantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics, a Werfen Company. The logo consists of a stylized "W" shape in red and blue, followed by the text "Inova Diagnostics" in a sans-serif font. Below the company name is the text "A Werfen Company" in a smaller font size.

510(k) Summary

NOVA Lite® DAPI dsDNA Crithidia luciliae Kit

Page 1 of 19

Table of Contents

Administrative data
Predicate device
Device description
Intended use(s)
Indications for use……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Substantial equivalence
Comparison to predicate device
Analytical performance characteristics
Nomenclature used in the studies
Precision
Reproducibility Studies
Linearity
Interference
Sample Stability and Handling
Reagent Stability
Cut-off, reference range
Clinical performance characteristics
Clinical sensitivity, specificity
Expected values
Comparison with predicate device
System description

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{4}------------------------------------------------

Administrative data

Submitter:Inova Diagnostics, Inc.9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:NOVA Lite® DAPI dsDNA Crithidia luciliae Kit
Scientific contact:Carolina Auza, Supervisor, Research and DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x 1345Fax: 858-863-0025email: cauza@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RAInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900 x 1381Fax: 858-863-0025email: relliott@inovadx.com
Device name (assay kit):Proprietary name: NOVA Lite® DAPI dsDNA Crithidia luciliae KitCommon name: anti-double stranded DNA (dsDNA) kitClassification name: test system, anti-double stranded DNA (dsDNA) antibodies
Regulation Description:Antinuclear antibody immunological test system
Regulation Medical Specialty:Immunology
Review Panel:Immunology
Product Code:KTLPIV
Classification Panel:82- Immunology
Regulation Number:866.5100- Antinuclear Antibody Immunological Test System

{5}------------------------------------------------

Predicate device

NOVA Lite ™ DSDNA, 510(k) number: K880742

Device description

The NOVA Lite DAPI dsDNA Crithidia luciliae Kit is an indirect immunofluorescence assay for the qualitative detection and semi-quantitative determination of Anti-dsDNA Antibodies (IgG) in human serum.

Samples are diluted 1:10 in PBS and incubated with the antigen substrate (dsDNA on glass microscope slides). After incubation, unbound antibodies are washed off. The substrate is then incubated with antihuman IgG-FITC conjugate. The conjugate contains a DNA-binding blue fluorescent dye, 4',6-diamidino-2phenylindole (DAPI) that is required for NOVA View use. The blue dye is not visible by traditional fluorescence microscope at the wavelength where FITC fluorescence is viewed. Unbound reagent is washed off. Stained slides are read by manual fluorescence microscope or scanned with the NOVA View Automated Fluorescence Microscope. The resulting digital images are reviewed and interpreted from the computer monitor. dsDNA positive samples exhibit an apple green fluorescence corresponding to areas of the substrate where autoantibody has bound.

Manual interpretation

A sample is considered positive if specific kinetoplast plus nuclear staining is observed to be greater than the negative control.

NOVA View interpretation

When slides are analyzed by NOVA View, digital images of representative fields of view of the well are captured. These digital images must be reviewed and interpreted from the computer monitor by a trained operator. At the same time when digital images are taken, NOVA View measures the FITC light intensity of the cells that are included in the region. NOVA View reports the measured fluorescence intensity in units of Light Intensity Units (LIU).

NOVA View provides the trained operator with the acquired digital images and the following supportive information:

  • LIU value
  • Negative/positive/indeterminate classification

NOVA View Single Well Titer (SWT)

The Single Well Titer (SWT) is a software application that estimates the endpoint titer (i.e. the highest dilution that produces positive result) for wells with a positive reaction, based on the obtained LIU.

The NOVA Lite® DAPI dsDNA Crithidia luciliae Kit contains the following:

  • dsDNA Crithidia luciliae Slides; 12 wells/slide, with desiccant
  • FITC IgG Conjugate with DAPI, containing 0.09% sodium azide; ready to use.
  • Positive Control: dsDNA; human serum with antibodies to dsDNA antigen, containing 0.09% sodium azide; pre-diluted, ready to use.

{6}------------------------------------------------

  • Negative Control: IFA System Negative Control, diluted human serum with no dsDNA antibodies present, containing 0.09% sodium azide; pre-diluted, ready to use.
  • PBS II (40x) Concentrate, sufficient for making 2000 mL of 1x PBS II.
  • Mounting Medium, containing 0.09% sodium azide
  • Coverslips

Intended use(s)

NOVA Lite® DAPI dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the qualitative and/or semi-quantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator.

Indications for use

Same as Intended use.

Substantial equivalence

The NOVA Lite DAPI dsDNA Crithidia luciliae reagents have the same intended use and assay principle as the predicate device NOVA Lite ™ DSDNA [510(k) K880742]

Similarities
ItemNOVA Lite DAPI Crithidia luciliae KitPredicate Device
Intended useNOVA Lite® DAPI dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the qualitative and/or semi-quantitative determination of anti-double stranded DNA (dsDNA) IgG antibodies in human serum by NOVA View Automated Fluorescence Microscope or manual fluorescence microscopy. The presence of anti-dsDNA can be used in conjunction with other serological and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE). All results generated with NOVA View device must be confirmed by a trained operator.NOVA Lite dsDNA Crithidia luciliae is an indirect immunofluorescent assay for the screening and semi-quantitative determination of anti-double stranded DNA (dsDNA) in human serum. The presence of anti-double stranded DNA can be used in conjunction with other serological tests and clinical findings to aid in the diagnosis of systemic lupus erythematosus (SLE).

Comparison to predicate device

{7}------------------------------------------------

Similarities
ItemNOVA Lite DAPI Crithidia luciliae KitPredicate Device
AnalyteAnti-dsDNA Antibodies (IgG) in human serumAnti-dsDNA Antibodies (IgG) in human serum
Assay methodologyindirect immunofluorescence assayindirect immunofluorescence assay
AntigenCrithidia luciliae cellsCrithidia luciliae cells
Sample typeSerumSerum
Sample dilution1:101:10
ControlsTwo levels of controls: one negative, one positive (dsDNA Positive)Two levels of controls: one negative, one positive (dsDNA Positive)
Storage2-8 °C2-8 °C
Shelf life24 months24 months
Differences
ltemNOVA Lite DAPI Crithidia luciliae KitPredicate Device
Interpretationby manual fluorescence microscopyor with the NOVA View deviceby manual fluorescence microscopy
ConjugateFITC conjugated anti-human IgG (Fcspecific) with added 4',6-diamidino-2-phenylindole (DAPI)FITC conjugated anti-human IgG (Fcspecific)
Additional dye inConjugate4',6-diamidino-2-phenylindole(DAPI)None

Analytical performance characteristics

Nomenclature used in the studies

All studies have been performed by interpreting the results with both manual microscopy and with the NOVA View system.

  • . "Manual" and "Manual reading" refers to results obtained by reading the slides with traditional fluorescence microscope.
  • . "NOVA View" refers to software reported results obtained with the NOVA View Automated Fluorescence Microscope, such as Light Intensity Units (LIU) and positive/indeterminate classification information.
  • . "Digital", "Digital reading" and "Digital image" refers to results obtained by reading NOVA View generated images on the computer monitor.

For statistical calculations, a positive result is presented as "1", and a negative result is presented as "0″. Intensity of the staining is expressed in reactivity grades. Grade 0 is negative; grades 1-4 are weak to strong positive.

  • 4+ Brilliant apple green fluorescence
  • 3+ Bright apple green fluorescence
  • 2+ Clearly distinguishable positive fluorescence
  • 1+ Lowest specific fluorescence that enables the nuclear and/or cytoplasmic staining to be clearly differentiated from the background fluorescence

{8}------------------------------------------------

Precision

To assess the precision performance of the NOVA Lite dsDNA Crithidia luciliae Kit results, a study was performed by processing 2 negative and 2 borderline samples with various intensities, in three replicates, in 14 runs (2 runs per day) for 7 days in triplicate resulting in 42 data points for each sample. The slides were read with NOVA view and digital images were interpreted by the operator as well as being read with manual microscope: i.e. three set of results were generated: NOVA view software interpretation, digital image reading results and manual reading results.

Acceptance criteria: Difference between reactivity grades within one run (between replicates) are within ± one reactivity grade. Average reactivity grade difference between any runs is within ± one reactivity grade.

Results: For both digital images reading, grades were within ± one reactivity grade within one run (within triplicates), and the average grade was no more than one reactivity grade different between runs.

Expected ResultObtained Result
SamplenNegative/Positive (grade)NOVA View resultsManual resultsDigital results
% Negative% PositiveGrade range (0-4+)% Negative% PositiveGrade range (0-4+)% Negative% Positive
Sample 142Positive (3-4)7%93%3-40%100%40%100%
Sample 242Negative (0-1)100%0%0-198%2%0100%0%
Sample 342Borderline (0-2)45%55%0-214%86%1-210%90%
Sample 442Borderline (0-2)100%0%1-20%100%0-269%31%
Sample 542Positive (1-3)64%36%1-30%100%1-214%86%
Sample 642Negative (0)90%10%0100%0%0100%0%

The results are summarized in the table below.

Reproducibility Studies

Between sites/instruments reproducibility

Ten samples (3 negative, 7 positive) were tested in three replicates, twice a day for 5 days at each site (30 data points per sample). Manual and digital reading was performed by two operators at each site, to assess between operator reproducibility.

Acceptance criteria: 90% agreement between operators and between sites

{9}------------------------------------------------

NOVA Lite® DAPI dsDNA Crithidia luciliae Kit

Grades for site 1:

Digital Reactivity GradeManual Reactivity Grade
Sample 1Sample 1
DayRun 1Run 2Run 1Run 2
1444433344
2444433344
3444433344
4444444433
5444443444
Sample 2Sample 2
1444443344
2444444444
3444444444
4444444444
5444444444
Sample 3Sample 3
1444444444
2444444444
3444444444
4444444444
5444444444
Sample 4Sample 4
1322222222
2233222222
3233222222
4233222222
5233222222
Sample 5Sample 5
1112222222
2211122222
3222222222
4222222222
5221122122
Sample 6Sample 6
1111122222
2111111121
3111111111
4111111111
5111111111
Sample 7Sample 7
1444433333
2444433344
Digital Reactivity GradeManual Reactivity Grade
344444333333
444444333333
544444333333
Sample 8Sample 8
100000000000
200000000000
300000000000
400000000000
500000000000
Sample 9Sample 9
100000000000
200000000000
300000000000
400000000000
500000000000
Sample 10Sample 10
100000000000
200000000000
300000000000
400000000000
500000000000

{10}------------------------------------------------

Summary:

Qualitative Agreement (manual reading):

ExpectedResultsManual Reading
Site1Site 2Site 3
Reader 1Reader 2Reader 1Reader 2Reader 1Reader 2
% neg% pos% neg% pos% neg% pos% neg% pos% neg% pos% neg% pos
Samplenpos/neg0%100%0%100%0%100%0%100%0%100%0%100%
130pos0%100%0%100%0%100%0%100%0%100%0%100%
230pos0%100%0%100%0%100%0%100%0%100%0%100%
330pos0%100%0%100%0%100%0%100%0%100%0%100%
430pos7%93%0%100%0%100%0%100%7%93%7%93%
530pos0%100%0%100%0%100%0%100%0%100%0%100%
630pos0%100%0%100%0%100%0%100%0%100%0%100%
730pos0%100%0%100%0%100%0%100%0%100%0%100%
830neg100%0%100%0%100%0%100%0%100%0%100%0%
930neg100%0%100%0%100%0%100%0%100%0%100%0%
1030neg100%0%100%0%100%0%100%0%100%0%100%0%

{11}------------------------------------------------

ExpectedResultsDigital Reading
Site1Site 2Site 3
Samplenpos/negReader 1% negReader 1% posReader 2% negReader 2% posReader 1% negReader 1% posReader 2% negReader 2% posReader 1% negReader 1% posReader 2% negReader 2% pos
130pos0%100%0%100%0%100%0%100%0%100%0%100%
230pos0%100%0%100%0%100%0%100%0%100%0%100%
330pos0%100%0%100%0%100%0%100%0%100%0%100%
430pos0%100%0%100%0%100%0%100%0%100%0%100%
530pos0%100%0%100%0%100%0%100%0%100%0%100%
630pos0%100%0%100%0%100%0%100%0%100%0%100%
730pos0%100%0%100%0%100%0%100%0%100%0%100%
830neg100%0%100%0%100%0%100%0%100%0%100%0%
930neg100%0%100%0%100%0%100%0%100%0%100%0%
1030neg100%0%100%0%100%0%100%0%100%0%100%0%

Qualitative Agreement (digital reading):

Agreement between reader 1 and reader 2 at each site:

% Overall Agreement (Positive/Negative) between operators (per site)
Site 1Site 2Site 3
Manual Reading99.7%100.0%100.0%
Digital Reading100.0%100.0%100.0%

Reproducibility between lots

Lot to lot comparison study was performed on two reagent lots. Twenty clinically and/or analytically characterized samples were tested in duplicate.

The following comparisons were made:

  • . NOVA view output: qualitative agreement.
  • Digital image reading: qualitative agreement and grade agreement .
  • . Manual image reading: qualitative agreement and grade agreement comparison.

Acceptance criteria:

  • . Qualitative positive, negative and total agreement: ≥ 90%
  • . Grade agreement: ≥ 90% within ± 1 reactivity grade

Qualitative Agreement

Qualitative positive agreements in two-way comparisons ranged from 91.7% to 100.0% Qualitative negative agreements in two-way comparisons ranged from 96.4% to 100.0% Qualitative total agreements in two-way comparisons ranged from 95.0%-100.0%

NOVA View Summary Table

NOVA ViewNegative agreement (%)(95% CI)Positive agreement (%)(95% CI)Total agreement (%)(95% CI)
Lot 046700 vs Lot RP00396.4 (82.3-99.4)91.7 (64.6-98.5)95.0 (83.5-98.6)

{12}------------------------------------------------

Manual Summary Table

NOVA ViewNegative agreement (%)(95% CI)Positive agreement (%)(95% CI)Total agreement (%)(95% CI)
Lot 046700 vs Lot RP003100.0 (86.2 – 100.0)100.0 (80.6– 100.0)100.0 (91.2 – 100.0)

Digital Summary Table

NOVA ViewNegative agreement (%)(95% CI)Positive agreement (%)(95% CI)Total agreement (%)(95% CI)
Lot 046700 vs Lot RP003100.0 (87.1-100.0)92.9 (68.5-98.7)97.5 (87.1-99.6)

Grade agreement

All grades (100%) were within ±1 grade from each other for all samples in any pair-wise comparisons for manual and most grades were within ±1 grade from each other for all samples in any pair-wise comparison for digital reading, two comparisons were 100% and one comparison was 98%.

Manual – Agreement +/- 1 reactivity grade = 100%

RP0003
04670001234Total
024000024
1090009
2001001
3000404
4000022
Total24914240

Digital - Agreement +/- 1 reactivity grade = 98%

RP0003
04670001234Total
026000026
1130004
2002002
3000000
4001168
Total27331640

Linearity

The linearity study was performed by serially diluting 3 positive samples (one high positive, one medium positive and one low positive) from 1:10 to 1:5120. These samples were assed with the NOVA Lite DAPI dsDNA Crithidia luciliae Kit and read with the NOVA View, digital images were interpreted and confirmed. All slides were read with manual microscopy. Qualitative and semi-quantitative results (using a scale of 0 (negative) to 4 (strong positive)) were captured for the manual and digital reading.

{13}------------------------------------------------

Sample No.SampleNOVA View (SWT)ManualDigital
1High Positive>=320640640
2Medium Positive402040
3Low Positive202020

Three samples used in the study:

The sample dilutions and associated intensity grade results are summarized in the table below for manual microscopy:

Sample No.10204080160320640128025605120
14444321000
22200000000
31100000000

The sample dilutions and associated intensity grades are summarized in the table below for digital reading:

Sample No.10204080160320640128025605120
14444321000
22210000000
31100000000

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. A set of three specimens were tested (one negative, one positive, one strong positive) using the following Interfering substances (hemoglobin, bilirubin, triglycerides, cholesterol, Rituximab, Methylprednisolone, Cyclophosphamide, Methotrexate, Azathioprine, Ibuprofen, Naproxen, Hydroxychloroquine, Mycophenolate). All interferents were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the NOVA Lite DAPI dsDNA Crithidia luciliae Kit. To assess interference with rheumatoid factor (RF), 10%, 30% and 50% (volume) RF positive sample was added to the test samples. All samples were processed with NOVA Lite DAPI dsDNA Crithidia luciliae kit and read with NOVA View. Digital images were interpreted and confirmed. Moreover, all slides were read by the same operator with manual microscopy. Acceptance criteria for the interference studies were grades obtained on samples with interfering substances are within ± 1 reactivity grade of those obtained on the control samples, spiked with diluent. No interference was detected with hemoglobin up to 200 mg/dL, bilirubin up to 100 mg/dL, triglycerides up to 1,000 mg/dL, cholesterol up to 224.3 mg/dL, rheumatoid factor up to 28.02 IU/mL, azathioprine up to 0.03 mg/ml, cyclophosphamide up to 4.1 mg/mL, hydroxychloroquine up to 0.224 mg/ml, ibuprofen up to 5 mg/mL, methotrexate up to 0.1 mg/mL, methy|prednisolone up to 0.85 mg/ml, mycophenolate up to 0.004 mg/ml, naproxen up to 5 mg/mL, rituximab up to 7.6 mg/mL, and belimumab up to 8 mg/mL.

{14}------------------------------------------------

Sample Stability and Handling

Three samples, encompassing negative, around the cut-off, and positive samples were tested in duplicates for up to 21 days while stored at 2-8°C, up to 48 hours while stored at room temperature, and after repeated freeze/thaw cycles up to 3 cycles. Results were compared to those obtained on control samples (day zero, stored at 2-8°C).

Acceptance criteria:

•NOVA View results of positive or negative do not change category (positive to negative or negative to positive) and are not different than the control sample.

•Manual reading reactivity grades are within ±1 grade of that of the control sample (stored at 2-8°C)

•Digital image interpretation reactivity grades are within ±1 grade of that of the control sample (stored at 2-8°C).

All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples are stored up to 48 hours at room temperature, up to 7 days at 2-8°C and can be subjected to up to 3 freeze/thaw cycles (when samples are stored at or below -20°C).

Reagent Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for up to 4 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on all of the components of NOVA Lite DAPI dsDNA Crithidia luciliae Kit to support stability claim. The components within the kit are as follows: Slides, conjugate, Negative control, positive control, PBSII and Mounting Medium. Each week a new sealed kit was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The reactivity of the kits was calculated for each time point (compared to those obtained with 5 ± 3°C stored kit). All calculations were performed by comparing results of the kit stored at 5 ± 3°C (control) to those stored at 37 ± 3°C (test) for 1, 2, 3, and 4 weeks, where one week is equal to six months at 5 ± 3°C. Comparison of reactivity grades between incubated kits against the control kit within the same lot for manual and digital methods.

Acceptance criteria for two-year preliminary expiration dating: Reactivity grades of all samples/reagent controls run must be within ±1 reactivity grade of the control condition (week 0) for both manual and digital image interpretation for all three lots.

The acceptance criteria were successfully met with the accelerated lots tested. All samples tested against those run on the control kit (week 0) were within ±1 reactivity grade of the control kit.

In-use stability

Conjugate

Stability Claim: stable for 8 weeks after opening when stored at 2-8℃ given that they have not reached the expiration date found on the label.

{15}------------------------------------------------

During assessing open vial stability, conjugate bottle was opened and placed in the refrigerator. It was tested each weeks against a bottle that was left unopened each week using positive and negative controls on dsDNA Crithidia luciliae kits.

Conjugate is considered stable if the following criteria were met:

  • Appearance: Clear liquid, free from foreign matter
  • The grades from each reading are within ±1 grade from each other.
  • . Fluorescence grading of >3+ for undiluted positive control and grading of 0 for undiluted negative control.
  • Testing is comparable to control

The acceptance criteria were successfully met with all 8 weeks tested.

Controls

Stability Claim: stable for 8 weeks after opening when stored at 2-8℃ given that they have not reached the expiration date found on the label.

During assessing open vial stability, control bottles was opened and then closed in the refrigerator. It was tested each week for 8 weeks against a bottle that was left unopened each week using positive and negative controls on dsDNA Crithidia luciliae kits.

Conjugate is considered stable if the following criteria were met:

  • Appearance: Clear liquid, free from foreign matter
  • The grades from each reading are within ±1 grade from each other. ●
  • . Fluorescence grading of >3+ for undiluted positive control and grading of 0 for undiluted negative control.
  • Testing is comparable to control

The acceptance criteria were successfully met with all 8 weeks tested.

Real time stability

Real time stability testing has been scheduled to be performed each of the time points listed below on the NOVA Lite DAPI dsDNA Crithidia luciliae Kit, to verify the 2-year expiration that was assigned based on accelerated stability studies.

At the time of the submission, results were available up to 24 months for lot 1, 15 months for lot 2 and 19 months for lot 3. Complete studies will be done by January 2020.

Acceptance criteria: results should fall within their respective ranges.

All results to date were within the acceptance limits.

Cut-off, reference range

The recommended starting dilution, above which the result is reported as positive and below which the result is reported as negative, is 1:10. The manufacturer suggests performing two-fold dilutions and also recommends that each laboratory establish its own titering protocol.

{16}------------------------------------------------

Clinical performance characteristics

Clinical sensitivity, specificity

To assess clinical performance, a clinical study was performed by Inova Diagnostics on 766 clinically characterized serum samples. The distribution of the cohorts is in the table below:

DiagnosisNNOVA View posManual posDigital pos
Systemic Lupus Erythematosus (SLE)39122357%18848%18848%
Drug Induced Lupus2015%15%15%
Infectious Disease60813%915%813%
Vasculitis3027%00%00%
Primary AntiphospholipidSyndrome20525%525%525%
Sjogren's Syndrome30310%00%00%
Celiac Disease2015%315%00%
Systemic Sclerosis3027%310%310%
Idiopathic Inflammatory Myopathy20420%315%210%
Mixed Connective Tissue Disease20315%315%315%
Crohn's Disease2015%15%15%
Grave's Disease2000%00%00%
Hashimoto's Disease30620%00%310%
Rheumatoid Arthritis3526%26%26%
Autoimmune Hepatitis (AIH)20420%315%15%
Total76626535%22129%21728%

Sensitivity (on SLE) and specificity, calculated on the combined population, are shown below.

N=766Sensitivity (%)(95% CI)Specificity (%)(95% CI)
Manual48.1 (43.2-53.0)91.2 (87.9-93.7)
Digital48.1 (43.2-53.0)92.3 (89.1-94.6)
NOVA View57.0 (52.1-61.8)88.8 (85.2-91.6)

Clinical studies 3 sites

Clinical study was completed at three sites using 269 (clinically characterized samples were tested on the NOVA Lite DAPI dsDNA (Crithidia luciliae) Kit according to its direction insert.

  • a. Clinical Sensitivity/Specificity:
    Sensitivity and specificity for all sites combined (269 x 3 sites= 807 total samples; 100 positive x 3 sites=300; 169 negative x 3 sites= 507)
N=807Clinical Diagnosis
Positive/per total SLENegative/per total non SLE
NOVA View120/300433/507
Manual Reading98/300484/507
Digital Reading102/300484/507

Correlation to clinical diagnosis for each mode at the three sites

{17}------------------------------------------------

NOVA Lite® DAPI dsDNA Crithidia luciliae Kit

PerformanceNOVA ViewManual ReadingDigital Reading
Sensitivity %(95% CI)120/30040.0 (34.6-45.6)98/30032.7 (27.6-38.2)102/30034.0 (28.9-39.5)
Specificity %(95% CI)433/50785.4 (82.1-88.2)484/50795.5 (93.3-97.0)484/50795.5 (93.3-97.0)

Summary: clinical overall agreement for all three sites

Correlation with SLE clinical diagnosis

Sensitivity (95% CI)N=100NOVA ViewManual ReadingDigital Reading
Site 139/10033/10033/100
39.0 (30.0-48.8)33.0 (24.6-42.7)33.0 (24.6-42.7)
Site 243/10033/10034/100
43.0 (33.7-52.8)33.0 (24.6-42.7)34.0 (25.5-43.7)
Site 338/10032/10035/100
38.0 (29.1-47.8)32.0 (23.7-41.7)35.0 (26.4-44.7)

Correlation with differential diagnosis

Specificity (95% CI)N=169NOVA ViewManual ReadingDigital Reading
Site 1143/16984.6 (78.4-89.3)164/16997.0 (93.3-98.7)165/16997.6 (94.1-99.1)
Site 2143/16984.6 (78.4-89.3)159/16994.1 (89.5-96.8)161/16995.3 (90.9-97.6)
Site 3147/16987.0 (81.1-91.2)161/16995.3 (90.9-97.6)158/16993.5 (88.7-96.3)

{18}------------------------------------------------

NOVA Lite® DAPI dsDNA Crithidia luciliae Kit

Site 1Site 2Site 3
DiseaseNNOVA ViewManual ReadingDigital ReadingNOVA ViewManual ReadingDigital ReadingNOVA ViewManual ReadingDigital Reading
SLE10039%33%33%43%33%34%38%32%35%
Specificity Non-SLE
AIH2080%95%95%60%85%85%80%90%85%
APS2085%100%100%95%100%100%90%100%100%
AAV1989%100%100%79%89%95%89%95%89%
CD8100%100%100%88%100%100%88%100%100%
CKD2100%100%100%0%100%100%100%100%100%
COPD978%89%100%89%100%100%100%100%100%
CrD6100%100%100%100%100%100%100%100%100%
HBV2100%100%100%50%100%100%100%100%100%
HCV5100%100%100%80%100%100%100%100%100%
HIV12100%100%100%100%100%100%100%100%100%
RA2080%100%100%100%100%100%80%100%100%
SjS2090%100%100%100%100%100%85%100%100%
SSc2055%85%85%65%75%80%70%75%70%
Syphilis6100%100%100%100%100%100%100%100%100%

Summary: sensitivity/specificity, correlation with clinical diagnosis by disease

SLE= Systemic Lupus Erytematosus; AlH= Autoimmune Hepatitis; APS= Antiphospholipid Syndrome; AAV= ANCA Associated Vasculitis; CD= Celiac Disease; CKD= Chronic Kidney Disease; COPD= Chronic Obstructive Pulmonary Disease; CrD= Chrohns Disease; HBV= Hepatitis B; HCV= Hepatitis C; HIV= Human Immunodeficiency Virus; RA= Rheumatoid Arthritis; SjS= Sjogren's Syndrome; SSc= Systemic Sclerosis.

Expected values

Expected values were analyzed on 120 samples from apparently healthy subjects: 60 females, with mean age of 41 years (range of 18-73).

There were four (3.3%) positive results with manual interpretation and eleven (9.2%) positive results with NOVA View software interpretation and one (0.8%) positive results with digital interpretation with NOVA Lite DAPI dsDNA Crithidia luciliae Kit testing.

{19}------------------------------------------------

Comparison with predicate device

Method comparison was performed with results obtained with the NOVA Lite DAPI dsDNA Crithidia luciliae Kit by NOVA View software interpretation and with the predicate device was performed using the same 744 serum samples (comprising of 391 serum samples from patients with SLE and 353 samples from patients with other diseases) tested in the clinical study (see table above).

Method comparison results are shown below:

n=744Positive agreement (%)95% CI)Negative agreement (%)95% CI)Total agreement (%)
708205 manual vs708215 manual87.8 (82.7-91.5)96.0 (94.0-97.4)93.7
708205 manual vs708215 digital88.0 (82.9-91.7)94.6 (92.3-96.2)92.7
708205 manual vs708215 NOVA View87.0 (81.8-90.9)85.3 (82.0-88.0)85.7
708205 Manual Grade
708215 ManualGrade01234Total
051024101536
11934221076
2213344053
301621634
402183445
Total53174643441744

+/- 2 reactivity grade

99.6%
708205 Manual Grade
708215 Digital Grade01234Total
050719541536
119152610676
2531224953
301142834
410114245
Total53238454386744

+/- 2 reactivity grade

98.4%

System description

NOVA View Automated Fluorescence Microscope and Software

NOVA View Automated Fluorescence Microscope

NOVA View® Automated Fluorescence Microscope is an automated system consisting of a fluorescence microscope and software that acquires, analyzes, stores and displays digital images of stained indirect immunofluorescent slides. It is intended as an aid in the detection and classification of certain antibodies

{20}------------------------------------------------

by indirect immunofluorescence technology. The device can only be used with cleared or approved in vitro diagnostic assays that are indicated for use with the device. A trained operator must confirm results generated with the device.

The NOVA View device has been cleared by the FDA in DEN140039. Subsequently, the use of NOVA View with ANCA modules has been cleared in K161258. Device description and Principle of Operation were described in DEN140039 and K161258 remained unchanged for this submission.

Software

Level of Concern: Moderate.

The NOVA View software version in this submission contains the following changes compared to the version in K161258:

Addition of CLIFT module and CLIFT SWT application.

When CLIFT slides are analyzed by NOVA View, digital images of representative fields of view of the well are captured. At the same time when digital images are taken, NOVA View measures the FITC light intensity of the cells that are included in the region. NOVA View reports the measured fluorescence intensity in units of Light Intensity Units (LIU).

NOVA View provides the trained operator with the acquired digital images and the following supportive information:

  • LIU value

  • Negative/positive classification

NOVA View CLIFT Single Well Titer (SWT) This assay is compatible with NOVA View SWT.

The SWT is a software application that estimates the endpoint titer (e.g., the highest dilution that gives positive result) for wells with a positive reaction with CLIFT, based on the obtained fluorescence intensity. The highest titer that can be differentiated is 1:320. Above this value the NOVA View reports > 320 titer. SWT validation was part of the between sites reproducibility study.

The SWT function was established using 22 dsDNA positive samples that represent various levels of antibodies. Two-fold serial dilutions were made for each sample, starting from 1:10, up to 1:40960. Each dilution was processed on dsDNA Crithidia luciliae slides, and the results were interpreted by NOVA View, digital and manual reading. Results were used to establish the intensity curves based on 4-PL logistic curve fitting. NOVA View uses these built-in curves for the determination of the titer

{21}------------------------------------------------

The validation of the SWT function was performed using 31 positive samples were serially diluted and read using NOVA View and manual microscope (manual end-point titer). SWT was determined for all samples.

Acceptance criteria:

  • . SWT is within ± 2 dilution steps of that of the manual end-point titer and the digital titer.
    Based on 31 samples, 80.6% of SWT results were within ± 1 dilution step of that of the manual titer, and 83.9% were within ±1 dilution step of that of the digital titer, and 93.3% of SWT results were within ± 2 dilution steps of that of the manual titer and 93.5% were within ± 2 dilution steps of that of the digital titer. 2 out of the 31 samples were outside of this range.
SWT, n=31within ± 1 dilution stepwithin ± 2 dilution steps
Manual end-point80.6%93.5%
Digital end-point83.9%93.5%

Additionally, SWT validation was part of the between sites reproducibility study.

Seven positive samples were assayed by all three testing sites (including Inova). 100% (14 out of 14) of SWT results at the two external sites were within ± 1 dilution step of that of the manual titer, and 92.9% (13 out of 14) were within ± 1 dilution step of that of the digital titer; 100% of SWT results were within ± 2 dilution steps of that of both the manual titer and digital titer.

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).