K Number
K192903

Validate with FDA (Live)

Date Cleared
2020-01-06

(83 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Device Description

The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode, MV-Flow Mode or as a combination of these modes. The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals. The RS85 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information for the Samsung Medison Co., Ltd. RS85 Diagnostic Ultrasound System.

Important Note: The provided document is a 510(k) summary, which is a premarket notification to the FDA. It declares substantial equivalence to previously cleared devices. Therefore, it primarily compares the new device to predicates rather than presenting de novo acceptance criteria and extensive clinical trial data as might be found in a PMA (Premarket Approval) submission for a novel device or an AI breakthrough.

The document explicitly states: "The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence." This means that acceptance criteria for the device's performance in a clinical setting (e.g., diagnostic accuracy) are not detailed within this particular submission. The "acceptance criteria" presented below are inferred from the substantial equivalence claim and the comparison to predicate devices, focusing on technical and functional equivalence rather than clinical performance metrics.


1. Table of Acceptance Criteria and Reported Device Performance

Since this is a substantial equivalence submission, the "acceptance criteria" largely revolve around demonstrating the new device is functionally, technologically, and safely equivalent to existing cleared devices. The reported device performance is implicitly "meets or is similar to" the predicate devices.

Feature/CriterionAcceptance Criteria (Implied from Substantial Equivalence)Reported Device Performance (Implicitly "Met")Notes
Intended UseSame as predicate devicesSame as predicate devices"The Indications for use of the proposed device is same as primary predicate."
Clinical ApplicationsSame as predicate devicesSame as predicate devicesListed applications are identical to primary predicate (K191115). New (N) or previously cleared (P) status for specific transducers indicates equivalence to other predicates or prior clearances.
Modes of OperationSame as predicate devicesSame as predicate devicesAll listed modes (B, M, PWD, CWD, Color Doppler, Combined, Other) are present and equivalent to primary predicate.
Scanhead TypesSame types and functionality as predicate devicesSame types and functionality as predicate devicesLinear Array, Curved Linear Array, Endocavity, Phased Array, Static Probes.
Scanhead FrequencyWithin range of predicate devices (1.0 ~ 20.0 MHz)1.0 ~ 20.0 MHzIdentical to predicate devices.
Acoustic Output Display & FDA LimitsConform to recognized standards (Track 3) and display MI/TI- Display Feature for Higher Output- Track3 - MI Output Display - TI Output DisplayIdentical to predicate devices.
Electrical SafetyConform to ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010 /(R)2012 and IEC 60601-2-37RS85 evaluated per these standards.Explicitly stated compliance.
Electromagnetic CompatibilityConform to IEC60601-1-2: 2014 (4th Edition)IEC60601-1-2: 2014 (4th Edition)Explicitly stated compliance.
BiocompatibilityConform to ISO 10993-1 Fourth edition 2009-10-15ISO 10993-1 Fourth edition 2009-10-15Explicitly stated compliance.
Reprocessing Medical DevicesConform to FDA Guidance issued March 17, 2015, revised June 9, 2017Evaluated per guidanceExplicitly stated compliance.
Software/Firmware-driven FunctionalityEvaluated using same criteria as predicate for image qualityEvaluated for migration into new system design."All migrated software functionality was evaluated using the same test criteria as the predicate for all applicable imaging modes to ensure that migration into a new system design did not compromise image quality..."
Ultrasound Safety (Acoustic Output Measurement)Conform to IEC60601-2-37:2007 + A1:2015 and NEMA UD 2-2004 (R2009)IEC60601-2-37:2007 + A1:2015, NEMA UD 2-2004 (R2009)Explicitly stated compliance.
System Characteristics (Channels, Monitor, Power)Similar to predicate devices192 Transmit/Receive Channels, 23.8inch (new) / 23inch monitor, 256 gray shades, 100-240VAC, 1100VA, 50/60HzComparison table shows equivalency or slight improvements (e.g., 23.8-inch monitor) that do not alter fundamental safety or effectiveness.
Specific Software FeaturesFunctionally equivalent to corresponding features on predicate/reference devicesNew features (EzHRI, QUS, PureVision), predicate features (ShadowHDR, BiometryAssist, IOTA-ADNEX) added, improved features (ElastoScan+ Mode, MV-Flow Mode, S-Fusion, S-Shearwave Imaging, HQ-Vision, Setup&Preference), and name changes (Panoramic -> Panoramic+, MV-Index -> Vascularity Index, NeedleMate -> NeedleMate+)This implies the new/changed features were tested to ensure they function as intended and do not negatively impact safety or effectiveness compared to predicates they are similar to.
TransducersFunctionally equivalent to previously cleared transducersVarious new (N) and previously cleared (P) transducers listed (e.g., LA2-14A, CA1-7S, PA1-5A, EV2-10A, EA2-11AR, EA2-11AV, EV2-10A) with specific indications.Each transducer is compared to predicate transducers for substantial equivalence. "SE to predicates; refer to SE Analysis X" indicates supporting data.
Biopsy GuidesCompatible and functionally equivalent to predicate devicesBP-KIT-029 etc. (various models)Same or functionally equivalent to predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence."

Therefore:

  • Sample size for test set: Not applicable (no clinical test set as part of this submission for substantial equivalence). The evaluation focused on engineering, software verification, and validation tests comparing the new system to already cleared devices.
  • Data provenance: Not applicable for clinical data. The comparability data for components and technical specifications are likely derived from internal testing and validation processes by Samsung Medison (South Korea) against industry standards and predicate device specifications.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Given the statement that clinical studies were not required, there is no mention of external experts establishing a clinical "ground truth" for a test set in this submission. The "ground truth" for technical and functional validation would be against established engineering specifications, performance standards, and comparison with predicate device outputs. This would involve internal engineering and quality assurance teams.


4. Adjudication Method for the Test Set

Not applicable, as no external clinical test set requiring adjudication by experts is described in this submission.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document explicitly states: "The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence." Furthermore, while the device includes new software features like "EzHRI, QUS(TAI, TSI), PureVision" and improvements to "ElastoScan+ Mode, MV-Flow Mode, S-Fusion, S-Shearwave Imaging", these are presented as enhancements or new capabilities whose safety and efficacy are demonstrated through technical equivalence to existing cleared features or algorithms, not through a comparative effectiveness study involving human readers and AI assistance. Therefore, no MRMC study or effect size for AI assistance is reported here.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document doesn't provide specific details on standalone algorithm performance for features like EzHRI, QUS (TAI, TSI), PureVision, ElastoScan+, MV-Flow, S-Fusion, S-Shearwave Imaging, or BiometryAssist. The submission focuses on demonstrating that these software features (whether new or improved) are substantially equivalent to functions on predicate devices or previously cleared equivalents, or meet established performance criteria during internal testing. It implies that these features are integrated into the diagnostic ultrasound system, functioning as tools for a human operator, rather than performing as a standalone diagnostic algorithm.


7. The Type of Ground Truth Used

For this 510(k) submission, the "ground truth" is primarily based on:

  • Technical and functional specifications: Adherence to established engineering parameters, performance metrics, and compliance with national and international standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, NEMA UD 2).
  • Comparison to predicate devices: The functionality and safety characteristics of the RS85 (including its new and revised features) are demonstrated to be substantially equivalent to those of the predicate devices (RS85 K191115, HERA W10 K182595, HS50/HS60 K181336, and other reference devices like AIXPLORER K173021, Fibroscan K150949, AmCAD-US K162574). This means the performance of the predicate devices implicitly serves as a "ground truth" for what is acceptable.

There is no mention of pathology, expert consensus on clinical cases, or outcomes data as a "ground truth" within this 510(k) summary, as it explicitly states no clinical studies were required.


8. The Sample Size for the Training Set

Not applicable. As no clinical studies were performed, there is no mention of a clinical training set for any AI/software features. If any internal machine learning models are used within the software features (e.g., BiometryAssist, S-Detect), the details of their training data are not provided in this 510(k) summary. The submission focuses on the output of these features being equivalent to predicate devices or meeting internal specifications.


9. How the Ground Truth for the Training Set Was Established

Not applicable, as no clinical training set is described in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

Samsung Medison Co., Ltd Ji Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, 25108 Kr

January 6, 2020

Re: K192903

Trade/Device Name: RS85 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 16, 2019 Received: December 16, 2019

Dear Ji Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K192903

Device Name RS85 Diagnostic Ultrasound System

Indications for Use (Describe)

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) No.:

Device Name: RS85 Diagnostic Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2, 3, 4, 7, 8, 9, 11, 14
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 5, 6, 7, 8, 9, 11, 12, 14, 15
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14, 15
Neonatal CephalicPPPPNote 1Notes 8, 9, 11, 16
Adult CephalicPPPPPNote 1Notes 7
Trans-rectalPPPPNote 1Notes 2, 7, 8, 9, 11, 12, 16, 17
Trans-vaginalPPPPNote 1Notes 2, 7, 8, 9, 11, 12, 16
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14. 16, 17
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14, 16, 17
Intra-luminal
Other (See Note 13)PPPPNote 1Notes 2, 7, 8, 9, 12, 16
Cardiac AdultPPPPPNote 1Notes 4, 7
CardiacCardiac PediatricPPPPPNote 1Notes 4, 7
Trans-esophageal (Cardiac)PPPPPNote 1Notes 4, 7
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

{4}------------------------------------------------

510(k) No.:

Device Name: CA1-7A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11, 14
Abdominal (See Note 10)PPPPPNote 1Notes 2,6,7,8,9,10,11,14,15,16,17
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 6, 7, 8, 9, 11, 14
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Notes 2, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 6, 7, 8, 9, 11, 14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2,6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{5}------------------------------------------------

510(k) No.:

Device Name: CF4-9 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Notes 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Notes 7, 8, 9, 11, 16
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 8, 9, 11
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
  • Note 3: Includes infertility monitoring of follicle development
  • Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{6}------------------------------------------------

510(k) No.:

Device Name: CA2-8A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 7, 8, 9, 10, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{7}------------------------------------------------

510(k) No.:

Device Name: CA3-10A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 10, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11, 16
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2, 7, 8, 9, 11
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{8}------------------------------------------------

510(k) No.:

Device Name: E3-12A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
    Note 8: 3D imaging

  • Note 9: Spatial Compound Imaging

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

{9}------------------------------------------------

510(k) No.:

Device Name: L3-12A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14,16
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14,16
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14,16
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion

Note 15: S-Shearwave

  • Note 16: MV-Flow
    Note 17: S-Shearwave Imaging (2D Shearwave)

{10}------------------------------------------------

Prescription Use (Per 21 CFR 801.109) DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: EA2-11B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 12)PPPPNote 1Note 2, 6, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PY+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode

  • Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwaye

  • Note 16: MV-Flow

Note 17: S-Shearwave Imaging (2D Shearwave)

{11}------------------------------------------------

Prescription Use (Per 21 CFR 801.109) DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: LA2-9A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15,17
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15,17
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging

Note 9: Spatial Compound Imaging

  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging

Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate
    Note 14: S-Fusion

Note 15: S-Shearwaye

Note 16: MV-Flow

Note 17: S-Shearwave Imaging (2D Shearwave)

Note 2: Includes imaging for guidance of biopsy

{12}------------------------------------------------

510(k) No.:

Device Name: LA3-16A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{13}------------------------------------------------

510(k) No.:

Device Name: LA3-16AI for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)PPPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 7, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 7, 9, 11
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{14}------------------------------------------------

510(k) No.:

Device Name: LA4-18B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)NNNNNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Note 2, 5, 6, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14,16
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2,5,6,7,8,9,11,14,16
Musculo-skel. (Superfic.)PPPPNote 1Note 2,5,6,7,8,9,11,14,16
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{15}------------------------------------------------

510(k) No.:

Device Name: LM4-15B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote 1Note 2, 5, 6, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2,5,6,7,8, 9,11,12,16
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{16}------------------------------------------------

510(k) No.:

Device Name: PM1-6A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
  • Note 3: Includes infertility monitoring of follicle development
  • Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{17}------------------------------------------------

510(k) No.:

Device Name: PA3-8B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{18}------------------------------------------------

510(k) No.:

Device Name: PA4-12B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultPPPPPNote 1Note 4, 7
Cardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+C+M, Dual/Ouad. B+C+CW. B+PD+CW. B+E. B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

Note 7: Tissue Harmonic Imaging (THI)

{19}------------------------------------------------

510(k) No.:

Device Name: MMPT3-7 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 12)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote 1Note 4, 7
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{20}------------------------------------------------

510(k) No.:

Device Name: CV1-8A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 4, 7, 8, 9, 10
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{21}------------------------------------------------

510(k) No.:

Device Name: V5-9 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{22}------------------------------------------------

510(k) No.:

Device Name: LV3-14A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{23}------------------------------------------------

510(k) No.:

Device Name: CW6.0 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultP
Cardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

Note 7: Tissue Harmonic Imaging (THI)

{24}------------------------------------------------

510(k) No.:

Device Name: DP2B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultP
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{25}------------------------------------------------

510(k) No.:

Device Name: DP8B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 12)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging&OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

{26}------------------------------------------------

510(k) No.:

Device Name: EV3-10B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

Note 7: Tissue Harmonic Imaging (THI)

{27}------------------------------------------------

510(k) No.:

Device Name: LA2-14A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)NNNNNNote 1Notes 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatricNNNNNNote 1Notes 7, 8, 9, 11
Small Organ (See Note 5)NNNNNNote 1Notes 7, 8, 9, 11,12,16,17
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNNote 1Notes 7, 8, 9, 11,12,16,17
Musculo-skel. (Superfic.)NNNNNNote 1Notes 7, 8, 9, 11,12,16,17
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselNNNNNNote 1Notes 7, 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow
  • Note 17: S-Shearwave Imaging (2D Shearwave)

{28}------------------------------------------------

510(k) No.:

Device Name: CA1-7S for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)NNNNNote 1Notes 2, 4, 7, 8, 9, 11, 14
Abdominal (See Note 10)NNNNNote 1Notes 2,6,7,8,9,10,11,14,15,16,17
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Notes 2, 6, 7, 8, 9, 11, 14
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNote 1Notes 2, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)NNNNNote 1Notes 2, 6, 7, 8, 9, 11, 14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNote 1Notes 2,6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
  • Note 3: Includes infertility monitoring of follicle development
  • Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

{29}------------------------------------------------

510(k) No.:

Device Name: EA2-11AR for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)NNNNNNote 1Note 9
Abdominal (See Note 10)NNNNNNote 1Note 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNNote 1Note 9, 12, 16, 17
Trans-vaginalNNNNNNote 1Note 9, 12,16
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)NNNNNNote 1Note 9, 12, 16, 17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
    Note 8: 3D imaging

  • Note 9: Spatial Compound Imaging

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

  • Note 16: MV-Flow

  • Note 17: S-Shearwave Imaging (2D Shearwave)

{30}------------------------------------------------

510(k) No.:

Device Name: EA2-11AV for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)NNNNNNote 1Note 9
Abdominal (See Note 10)NNNNNNote 1Note 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNNote 1Note 9, 12, 16, 17
Trans-vaginalNNNNNNote 1Note 9, 12, 16
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)NNNNNNote 1Note 9, 12, 16, 17
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
  • Note 3: Includes infertility monitoring of follicle development
  • Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

{31}------------------------------------------------

510(k) No.:

Device Name: EV2-10A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)NNNNNNote 1Note 8, 9
Abdominal (See Note 10)NNNNNNote 1Note 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNNote 1Note 8, 9, 12
Trans-vaginalNNNNNNote 1Note 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)NNNNNNote 1Note 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

{32}------------------------------------------------

510(k) No.:

Device Name: PA1-5A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA 181336; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
  • Note 3: Includes infertility monitoring of follicle development
  • Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: Spatial Compound Imaging
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave
  • Note 16: MV-Flow Note 17: S-Shearwave Imaging (2D Shearwave)

{33}------------------------------------------------

510(K) Summary: K192903

In accordance with 21 CFR 807.92 the following summary of information is provided:

  • Date Prepared October 11, 2019 1.
    1. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
    1. Primary Contact Person Ji Yea Lee Regulatory Affairs Specialist Phone: +82.2.2.2194.1594 Fax: +82.31.8017.9576 Email: jiyea722.lee@samsungmedison.com

Secondary Contact Person Ninad Gujar Director of Regulatory & Quality Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com

    1. Proposed Device
      Common Name: Trade/Device Name: Additional Marketing Name: Regulation Name: Panel/ Regulatory Class: Product Code: Regulation No./Product Code Diagnostic Ultrasound System and Accessories RS85 Diagnostic Ultrasound System RS85 Prestige Diagnostic Ultrasound System Ultrasound pulsed Doppler imaging system Radiology / II IYN, IYO, ITX 892.1550, 892.1560, 892.1570
    1. Predicate Device
    • . Predicates
      • RS85 Diagnostic Ultrasound System (K191115) ; Primary -
      • HERA W10 Diagnostic Ultrasound System (K182595) -
      • HS50/HS60 Diagnostic Ultrasound System (K181336) -
    • Reference
      • AIXPLORER & AIXPLORER Ultimate Ultrasound Diagnostic Systems (K173021) -
      • -Fibroscan (K150949)
      • AmCAD-US (K162574) -
    1. Device Description

The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode, MV-Flow Mode or as a combination of these

{34}------------------------------------------------

modes.

The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals.

The RS85 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    1. Indications for Use
      The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

The Indications for use of the proposed device is same as primary predicate.

    1. Technology
      The RS85 employs the similar fundamental scientific technology as its predicate device(s).
    1. Determination of Substantial Equivalence
      The proposed RS85 is similar to the predicate devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.

A comparison of the proposed device to the currently marketed predicates is provided in the table below.

RS85Addition of V2.00 (RS85 Prestige)Remarks
Clinicalapplication• No addition
Operation ofmodes• No addition
Appliedtransducers• LA2-14A• CA1-7S• PA1-5A• EV2-10A• EA2-11AR• EA2-11ARTransducer 510(k)Clearance LA2-14A NEW CA1-7S NEW PA1-5A K181336 EV2-10A NEW EA2-11AR NEW EA2-11AR NEW
SW Features• Addition of new S/W feature:: EzHRI, QUS(TAI, TSI), PureVision• Addition of predicate S/W features: ShadowHDR, BiometryAssist, IOTA-ADNEX• Improvement of cleared S/W feature: ElastoScan+ Mode(E-Thyroid), MV-Flow Mode,S-Fusion, S-Shearwave Imaging,HQ-Vision, Setup&Preference(EzPrep)• Name change of cleared S/W feature:: Panoramic → Panoramic+MV-Index → Vascularity IndexSW Features 510(k)clearance EzHRI NEW TAI NEW TSI NEW PureVision NEW ShadowHDR, K182595 BiometryAssist K182595 IOTA-ADNEX K182595

{35}------------------------------------------------

RS85Addition of V2.00 (RS85 Prestige)Remarks
NeedleMate → NeedleMate+
HW Features• Addition of 23.8inch Monitor

{36}------------------------------------------------

RS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)PredicateNote
Device nameSAMSUNGSAMSUNGSAMSUNGSAMSUNG
ManufacturerMEDISON.MEDISON.MEDISON.MEDISON.
Regulation No.892.1550892.1560892.1570892.1550892.1560892.1570892.1550892.1560892.1570892.1550892.1560892.1570
Product CodeIYN, IYO, ITXIYN, IYO, ITXIYN, IYO, ITXIYN, IYO, ITX
Intended UseThe ultrasounddiagnostic system andprobes are designed toobtain ultrasoundimages and analyzebody fluids.The ultrasounddiagnostic system andprobes are designed toobtain ultrasoundimages and analyzebody fluids.The ultrasounddiagnostic system andprobes are designed toobtain ultrasoundimages and analyzebody fluids.The ultrasounddiagnostic system andprobes are designed toobtain ultrasoundimages and analyzebody fluids.
clinicalApplication- Fetal/Obstetrics- Abdominal- Gynecology- Intra-operative- Pediatric- Small Organ- Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Trans-esophageal(Cardiac)- Muscular-Skeletal(Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Peripheral vessel- Fetal/Obstetrics- Abdominal- Gynecology- Intra-operative- Pediatric- Small Organ- Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Trans-esophageal(Cardiac)- Muscular-Skeletal(Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Peripheral vessel- Fetal/Obstetrics- Abdominal- Gynecology- Pediatric- Small Organ- Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Muscular-Skeletal(Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Peripheral vessel- Fetal/Obstetrics- Abdominal- Gynecology- Pediatric- Small Organ- Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Muscular-Skeletal(Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Peripheral vesselSame to predicates
Scanhead Types- Linear Array- Linear Array- Linear Array- Linear ArraySame to predicates
RS85(Under Review)RS85(K191115)HERA W10(K182595)HS50/HS60(K181336)
Device nameProposed DevicePrimary PredicatePredicatePredicateNote
- Curved Linear Array- Curved Linear Array- Curved Linear Array- Curved Linear Array
- Endocavity- Endocavity- Endocavity- Endocavity
- Phased Array- Phased Array- Phased Array- Phased Array
- Static Probes- Static Probes- Static Probes- Static Probes
ScanheadFrequency1.0 ~ 20.0 MHz1.0 ~ 20.0 MHz1.0 ~ 20.0 MHz1.0 ~ 20.0 MHz
Acoustic OutputDisplay & FDALimits- Display Feature forHigher Output-Track3- Display Feature forHigher Output-Track3- Display Feature forHigher Output-Track3- Display Feature forHigher Output-Track3
- MI Output Display- MI Output Display- MI Output Display- MI Output Display
- TI Output Display- TI Output Display- TI Output Display- TI Output Display
Modes of Operation- B-mode- B-mode- B-mode- B-mode
- M-mode- M-mode- M-mode- M-mode
- Pulsed wave (PW)Doppler- Pulsed wave (PW)Doppler- Pulsed wave (PW)Doppler- Pulsed wave (PW)Doppler
- Continuous wave(CW) Doppler- Continuous wave(CW) Doppler- Continuous wave(CW) Doppler- Continuous wave(CW) Doppler
- Color Doppler- Color Doppler- Color Doppler- Color Doppler
- Tissue DopplerImaging (TDI)- Tissue DopplerImaging (TDI)- Tissue DopplerImaging (TDI)- Tissue DopplerImaging (TDI)
- Tissue DopplerWave (TDW)- Tissue Doppler Wave(TDW)- Tissue DopplerWave (TDW)- Tissue DopplerWave (TDW)
- Power AmplitudeDoppler- Power AmplitudeDoppler- Power AmplitudeDoppler- Power AmplitudeDoppler
- Pulse Inversion- Pulse Inversion- Pulse Inversion- Pulse Inversion
- Harmonic Imaging(S-Harmonic)- Harmonic Imaging(S-Harmonic)- Harmonic Imaging(S-Harmonic)- Harmonic Imaging(S-Harmonic)
- Directional PowerDoppler (S-Flow)- Directional PowerDoppler (S-Flow)- Directional PowerDoppler (S-Flow)- Directional PowerDoppler (S-Flow)
- Color M-Mode- Color M-Mode- Color M-Mode- Color M-Mode
- 3D/4D imagingmode- 3D/4D imaging mode- 3D/4D imagingmode- 3D ImagingCombination mode
- ElastoScan+ Mode- ElastoScan+ Mode- ElastoScan+ Mode- ElastoScan Mode
- MV-Flow Mode- MV-Flow Mode- MV-Flow Mode
RS85 Diagnostic Ultrasound SystemNote
Device nameRS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)Predicate
- Combined modes- Combined modes- Combined modes- Combined modes-
#Transmit Channels192192192128
#Receive Channels192192192128
SystemCharacteristics:Beamformer 192Beamformer 192Beamformer 192- Beamformer 128
Mobile cartMobile cartMobile cart- Mobile cart
LCD Monitor (LEDBacklight unit):23inch, 23.8inchLCD Monitor (LEDBacklight unit):23inchLCD Monitor (LEDBacklight unit):23inchLCD Monitor (LEDBacklight unit):- 21.5inchSE to predicates; refer to018_Electromagnetic Compatibilityand ElectricalSafety
256 gray shades onmonitor100-240VAC,1100VA,50/60Hz256 gray shades onmonitor100-240VAC,1100VA,50/60Hz256 gray shades onmonitor100-240VAC,1100VA,50/60Hz- 256 gray shades onmonitor- 100-240VAC,1100VA,- 50/60Hz
Product SafetyCertificationIEC 60601-1CSA C22.2 No.601.1IEC 60601-2-37IEC 60601-1CSA C22.2 No.601.1IEC 60601-2-37IEC 60601-1CSA C22.2 No.601.1IEC 60601-2-37- IEC 60601-1- CSA C22.2 No.601.1- IEC 60601-2-37
EMC ComplianceIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2- IEC 60601-1-2Same to predicates
Acoustic OutputDisplay StandardTrack 3Track 3Track 3- Track 3
BiocompatibilityComplianceISO10993-1ISO10993-1ISO10993-1- ISO10993-1
SoftwareFunctionality- Q Scan- ClearVision- MultiVision- Panoramic+- Q Scan- ClearVision- MultiVision- Panoramic- Q Scan- ClearVision- MultiVision- Panoramic- Q Scan- ClearVision- MultiVision- Panoramic
Device nameRS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)PredicateNote
- Needle Mate+- Needle Mate+- Needle Mate+- Needle Mate+
- AutoIMT+- AutoIMT+- AutoIMT+- AutoIMT+
- Strain+- Strain+- Strain+
- StressEcho- StressEcho- StressEcho
- E-Strain- E-Strain- E-Strain- E-Strain
- S-Shearwave- S-Shearwave- S-Shearwave
- S-Detect for Breast- S-Detect for Breast- S-Detect for Breast- S-Detect for Breast
- S-Detect for Thyroid- S-Detect for Thyroid- S-Detect for Thyroid- S-Detect for Thyroid
- S-Fusion- S-Fusion
- Clear Track- Clear Track
- Virtual Track- Virtual Track
- S-3D ArterialAnalysis- S-3D ArterialAnalysis- S-3D ArterialAnalysis
- ADVR- ADVR- ADVR
-3D Imaging (VolumeData Acquisition)-3D Imaging (VolumeData Acquisition)-3D Imaging (VolumeData Acquisition)-3D Imaging (VolumeData Acquisition)
3D Imagingpresentation3D Imagingpresentation3D Imagingpresentation- 3D ImagingpresentationSame to predicates
-3D Cine/4D Cine-3D Cine/4D Cine-3D Cine/4D Cine- 3D Cine/4D Cine
-3D Rendering MPR-3D Rendering MPR-3D Rendering MPR- 3D Rendering MPR
-3D XI MSV/ObliqueView-3D XI MSV/ObliqueView-3D XI MSV/ObliqueView- 3D XI MSV/ObliqueView
- Volume CT- Volume CT- Volume CT- Volume CT
-3D MagiCut-3D MagiCut-3D MagiCut- 3D MagiCut
-Volume Calculation(VOCAL, XIVOCAL)-Volume Calculation(VOCAL, XI VOCAL)-Volume Calculation(VOCAL, XIVOCAL)- Volume Calculation(VOCAL, XIVOCAL)
- XI STIC- XI STIC- XI STIC- XI STIC
- HDVI- HDVI- HDVI- HDVI
- RealisticVue- RealisticVue- RealisticVue- RealisticVue
- NaturalVue- NaturalVue- NaturalVue
RS85 Diagnostic Ultrasound System
Device nameRS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)PredicateNote
- CEUS+- CEUS+- CEUS+- CEUS+
- MV-Flow- MV-Flow- MV-Flow
- SonoSync- SonoSync
- CrystalVue- CrystalVue- CrystalVue- CrystalVue
- ShadowHDR- ShadowHDR
- BiometryAssist- BiometryAssist- BiometryAssist
- IOTA-ADNEX- IOTA-ADNEX
- ElastoScan+- ElastoScan+- ElastoScan+- ElastoScan+Same to predicates
- E-Thyroid- E-Thyroid- E-Thyroid
- E-Breast- E-Breast- E-Breast
- S-Shearwave Imaging- S-Shearwave Imaging
- MV-Flow- MV-Flow- MV-Flow
- HQ-Vision- HQ-Vision- HQ-Vision- HQ-Vision
- EzHRISE to reference;refer to SEAnalysis 1
- TSISE to reference;refer to SEAnalysis 2
- TAISE to reference;refer to SEAnalysis 3
- PureVisionSE to reference;refer to SEAnalysis 4
Transducer- CA1-7A- CA1-7A- CA1-7ASame to predicates
- CF4-9- CF4-9- CF4-9- CF4-9
- CA2-8A- CA2-8A- CA2-8A
- CA3-10A- CA3-10A- CA3-10A- CA3-10A
- E3-12A- E3-12A- E3-12A
RS85 Diagnostic Ultrasound System
Device nameRS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)PredicateNote
- L3-12A- L3-12A- L3-12A
- EA2-11B- EA2-11B- EA2-11B- EA2-11B
- LA2-9A- LA2-9A- LA2-9A- LA2-9A
- LA3-16A- LA3-16A- LA3-16A
- LA3-16AI- LA3-16AI- LA3-16AI
- LA4-18B- LA4-18B- LA4-18B
- LM4-15B- LM4-15B
- PM1-6A- PM1-6A- PM1-6A
- PA3-8B- PA3-8B- PA3-8B- PA3-8B
- PA4-12B- PA4-12B- PA4-12B- PA4-12B
- MMPT3-7- MMPT3-7- MMPT3-7- MMPT3-7
- CV1-8A- CV1-8A- CV1-8A
- V5-9- V5-9- V5-9
- LV3-14A- LV3-14A
- CW6.0- CW6.0- CW6.0
- DP2B- DP2B- DP2B
- DP8B- DP8B- DP8B
- EV3-10B- EV3-10B- EV3-10B
- LA2-14ASE to predicates;refer to SEAnalysis 5
- CA1-7SSE to predicates;refer to SEAnalysis 6
- PA1-5A- PA1-5ASame to predicate
- EV2-10ASE to predicates;refer to SEAnalysis 7
- EA2-11ARSE to predicates;refer to SEAnalysis 8 & note
- EA2-11AV
CompatibleBiopsy Guide- BP-KIT-029- BP-KIT-029- BP-KIT-029- BP-KIT-029Same to predicates
- BP-KIT-030- BP-KIT-030
Device nameRS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)PredicateNote
(Biopsy Needleguide Kit)BP-KIT-043- BP-KIT-043- BP-KIT-043-BP-KIT-043
BP-KIT-053- BP-KIT-053- BP-KIT-053-BP-KIT-053
BP-KIT-054- BP-KIT-054- BP-KIT-054-BP-KIT-054
BP-KIT-055- BP-KIT-055-BP-KIT-055
BP-KIT-057- BP-KIT-057- BP-KIT-057
BP-KIT-058- BP-KIT-058- BP-KIT-058
BP-KIT-059- BP-KIT-059- BP-KIT-059
BP-KIT-060- BP-KIT-060- BP-KIT-060- BP-KIT-060
BP-KIT-065- BP-KIT-065- BP-KIT-065
BP-KIT-066- BP-KIT-066- BP-KIT-066
BP-KIT-067- BP-KIT-067- BP-KIT-067
BP-KIT-068- BP-KIT-068-BP-KIT-068
BP-KIT-069- BP-KIT-069- BP-KIT-069
BP-KIT-071- BP-KIT-071- BP-KIT-071
BP-KIT-075- BP-KIT-075- BP-KIT-075
BP-KIT-076- BP-KIT-076- BP-KIT-076
BP-KIT-077- BP-KIT-077- BP-KIT-077
BP-KIT-078- BP-KIT-078- BP-KIT-078
BP-KIT-079
BP-KIT-080Cleared (K970514)
BP-KIT-081Cleared (K875240)
BP-KIT-082Cleared (K875128)
BP-KIT-083- BP-KIT-083Cleared (K191115)
BP-KIT-085
BP-KIT-086
BP-KIT-088Cleared (K971722)
BP-KIT-089
BP-KIT-090
BP-KIT-094Cleared (K970514)
DigitalStorage/TransferStation- Foot Switch- ECG (defibrillationproof)- Foot SwitchECG (defibrillationproof)- Foot Switch- ECG (defibrillationproof)- Foot Switch- ECG (defibrillationproof)Same to predicates
Device nameRS85(Under Review)Proposed DeviceRS85(K191115)Primary PredicateHERA W10(K182595)PredicateHS50/HS60(K181336)PredicateNote
- Gel Warmer- Gel Warmer- Gel Warmer- Gel Warmer
- Clear Track/ VirtualTrack- Clear Track/ VirtualTrack- Clear Track/ VirtualTrack- Clear Track/ VirtualTrack
System StandSystem StandSystem StandSystem Stand
WLAN Card- WLAN Card- WLAN Card- WLAN Card

{37}------------------------------------------------

510(k) Summary

{38}------------------------------------------------

RS85 Diagnostic Ultrasound S10(k)
RS85 Diagnostic Ultrasound System

510(k) Summary

{39}------------------------------------------------

{40}------------------------------------------------

{41}------------------------------------------------

RS85 Diagnostic Ultrasound System

510(k) Summary

Section 6. Page 9/12

{42}------------------------------------------------

RS85 Diagnostic Ultrasound S10(k)
RS85 Diagnostic Ultrasound System

510(k) Summary

{43}------------------------------------------------

Traditional 510(k) RS85 Diagnostic Ultrasound System

10. Summary of Non-Clinical Test

The proposed device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The RS85 and its applications comply with voluntary standards.

TestStandards and FDA Guidance
Risk ManagementISO 14971 Second edition 2007 Medical devices - Application of risk management to medical devices
Electrical SafetyThe RS85 Ultrasound System with defibrillation-proof ECG electrode was evaluated per thefollowing standards.
ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010 /(R)2012Medical Electrical Equipment - Part 1: General Requirements for basic safety and essentialperformance.
ElectromagneticCompatibilityIEC60601-1-2: 2014(4th Edition) Medical Electrical Equipment -- Part 1-2: General RequirementsFor Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances --Requirements And Tests
BiocompatibilityISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluationand testing within a risk management process.
ReprocessingMedical DevicesFDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods andLabeling issued on March 17, 2015, revised June 9, 2017.
Software/Firmware-drivenFunctionalityAll migrated software functionality was evaluated using the same test criteria as the predicate for allapplicable imaging modes to ensure that migration into a new system design did not compromiseimage quality with respect to the intended use of each feature.

{44}------------------------------------------------

Guidance for the Content of Premarket Submissions for Software Contained in Medical Devicesissued on May 11, 2005
Ultrasound SafetyInformation for Manufacturers Seeking Marketing Clearance of
Diagnostic Ultrasound Systems and Transducers issued September 9, 2008
IEC60601-2-37:2007 + A1:2015, Particular requirements for the safety of ultrasonic medical
diagnostic and monitoring equipment
NEMA UD 2-2004 (R2009)
Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3

11. Summary of Clinical Tests

The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence.

12. Conclusion

SAMSUNG MEDISON considers the RSS5 to be as safe, as effective, and performance is substantially equivalent to the predicate device.

END of 510(K) Summary