K Number
K192847

Validate with FDA (Live)

Device Name
AUTOLIGN
Manufacturer
Date Cleared
2020-07-21

(292 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AUTOLIGN is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the AUTOLIGN requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Device Description

The AUTOLIGN is stand-alone software which utilizes images of the patient's intraoral anatomy from intraoral cameras and/or desktop laboratory scanners to create a 3D virtual dental model that can be used in the same manner as a traditional physical dental model. AUTOLIGN facilitates the segmentation and editing of the 3D virtual digital model as well as analysis which can be used in secondary orthodontic treatment planning. The software allows for measurement- including Bolton analyses. The models and analysis produced by the proposed Software can be exported to an orthodontic laboratory or directly to orthodontic appliance manufacturers for use in orthodontic treatment planning and design of orthodontic appliances. The material used for vacuum pressing (thermoforming) the final device appliances of sequential aligners on the production cast is DURAN®, but those which satisfy all the requirements specified on the Autoling User Manual can also be used.

AI/ML Overview

The provided text describes the AUTOLIGN device and its comparison to predicate devices, but it does not contain specific acceptance criteria, a detailed study proving the device meets those criteria, or the requested information about sample sizes, ground truth establishment, or clinical study details.

The document is a 510(k) summary for the AUTOLIGN device, which is a software tool for orthodontic treatment planning and design. Its purpose is to demonstrate substantial equivalence to legally marketed predicate devices, not to present a detailed clinical validation study with specific performance metrics.

Here's what can be extracted and what is missing based on your request:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated in terms of quantitative performance metrics for the device itself (e.g., accuracy, sensitivity, specificity for specific measurements or analyses). The "acceptance" is implicitly that the software functions as intended and is as safe and effective as the predicate device.
  • Reported Device Performance: The document only states that "The software passed the testing and performed per its intended use." No specific performance metrics or quantitative results are provided.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not mentioned.
  • Data Provenance: Not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of Experts: Not mentioned.
  • Qualifications of Experts: Not mentioned.
  • Ground Truth Establishment: Not mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • An MRMC study is not mentioned. The study described is a non-clinical software verification and validation, along with a comparison to predicate devices. There is no mention of human readers or AI assistance in a comparative effectiveness context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The document implies standalone testing as it states "The AUTOLIGN underwent software, hardware, integration, verification and validation testing... The software passed the testing and performed per its intended use." However, specific standalone performance metrics (e.g., algorithm accuracy for segmenting teeth, or measuring arch length) are not provided. The device itself is described as "stand-alone software."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not mentioned.

8. The sample size for the training set

  • Not mentioned, as there is no specific mention of a machine learning model or training set in the context of the device's validation.

9. How the ground truth for the training set was established

  • Not mentioned (for the same reason as point 8).

Summary of what is available from the text:

  • Device Name: AUTOLIGN
  • Regulatory Status: Class II medical device (K192847)
  • Intended Use: Front-end device for managing orthodontic models, systematic inspection, detailed analysis, treatment simulation, and virtual appliance design based on 3D models of patient dentition.
  • Non-clinical Tests: Software, hardware, integration, verification, and validation testing were performed according to FDA guidance.
  • Result of Non-clinical Tests: "The software passed the testing and performed per its intended use."
  • Predicate Device Comparison: The report extensively compares the AUTOLIGN's indications for use and technological characteristics to three predicate devices (Ortho System™, Orchestrate 3D, CEREC Ortho Software) to establish substantial equivalence.

In conclusion, the provided text focuses on regulatory approval through substantial equivalence to predicate devices and general software validation, rather than a detailed clinical study with quantitative acceptance criteria and performance metrics for the AUTOLIGN itself.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 21, 2020

Diorco Co., Ltd. % Edward Park CEO Radios LLC 4408 Tortuga Ln McKinney, Texas 75070

Re: K192847

Trade/Device Name: Autolign Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN, LLZ Dated: June 16, 2020 Received: June 22, 2020

Dear Edward Park:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192847

Device Name AUTOLIGN

Indications for Use (Describe)

The AUTOLIGN is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the AUTOLIGN requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - Traditional 510(k)

Submitter Information

Company Name:DIORCO Co., Ltd.
Company Address:4F, 26, Yeongil-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16705, Korea
Company Phone:+82-70-5030-3037
Company Fax:+82-31-205-0942
Contact Person:HWI JOON PARK, Official Correspondent for DIORCO Co., Ltd.
Date Summary Prepared:Aug 30, 2019

Device Identification

Trade / Proprietary name:AUTOLIGN
510(k) Number:K192847
Regulation Number:872.5470
Regulation Name:Orthodontic Plastic Bracket
Classification:Class II
Product Code:PNN, LLZ

Predicate Device

Primary predicate device: Ortho System™ (3Shape A/S) - K171634 Reference device: Orchestrate 3D (Orchestrate 3D) - K181112 Reference device: CEREC Ortho Software (Dentsply Sirona) - K171122

Device Description

The AUTOLIGN is stand-alone software which utilizes images of the patient's intraoral anatomy from intraoral cameras and/or desktop laboratory scanners to create a 3D virtual dental model that can be used in the same manner as a traditional physical dental model. AUTOLIGN facilitates the segmentation and editing of the 3D virtual digital model as well as analysis which can be used in secondary orthodontic treatment planning. The software allows for measurement- including Bolton analyses. The models and analysis produced by the proposed Software can be exported to an orthodontic laboratory or directly to orthodontic

{4}------------------------------------------------

appliance manufacturers for use in orthodontic treatment planning and design of orthodontic appliances. The material used for vacuum pressing (thermoforming) the final device appliances of sequential aligners on the production cast is DURAN®, but those which satisfy all the requirements specified on the Autoling User Manual can also be used.

Indications for Use

The AUTOLIGN is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.

The use of the AUTOLIGN requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Comparison of Indication for Use Statements

The following table compares Indications for Use Statements between the Autolign and the three predicate devices, i.e. Ortho System™, Orchestrate 3D Orthodontic Software, and CEREC Ortho Software. Similar to the other predicate devices, the Autolign is used for managing patient and case base data, collection, alignment, measurement and analysis of study material, treatment of simulation, and virtual appliance design.

AUTOLIGNOrtho System™Orchestrate 3DOrthodontic SoftwareCEREC OrthoSoftware
The AUTOLIGN isOrtho System™ isThe Orchestrate 3DOrthodontic Software isCEREC Ortho Software
intended for use as aintended for use as aindicated for use as ais intended for use with
medical front-end devicemedical front-end devicefront-end software toolimage data acquired
providing tools forproviding tools forfor management offrom handheld intra oral
management ofmanagement oforthodontic models,3D cameras and desktop
orthodontic models,orthodontic models,systematic inspection,laboratory scanners to
systematic inspection,systematic inspection,detailed analysis,create 3D virtual models
detailed analysis,detailed analysis,treatment simulation andto be used for data
treatment simulation andtreatment simulation andvirtual appliance designacquisition and modeling
virtual appliance designvirtual appliance designoptions, including dentalanalysis for orthodontic
options (Export ofoptions (Custom Metalcasts, which may bepatients and conditions.
Models, IndirectBands, Export ofused for sequentialThe CEREC Ortho
Bonding TransferModels, Indirectaligner trays or retainers.Software 3D model data
Media) based on 3DBonding Transfercan be exported to

{5}------------------------------------------------

models of the patient'sMedia) based on 3DThese applications areorthodontic design
dentition before the startmodels of the patient'sbased on 3D models ofsoftware to aid in the
of an orthodonticdentition before the startthe patient's dentitiondesign of orthodontic
treatment. It can also beof an orthodonticbefore the start of anappliances.
applied during thetreatment. It can also beorthodontic treatment. It
treatment to inspect andapplied during thecan also be applied
analyze the progress oftreatment to inspect andduring the treatment to
the treatment. It can beanalyze the progress ofinspect and analyze the
used at the end of thethe treatment. It can beprogress of the
treatment to evaluate ifused at the end of thetreatment. It can be used
the outcome is consistenttreatment to evaluate ifat the end of the
with the planned/desiredthe outcome is consistenttreatment to evaluate if
treatment objectives.with the planned/desiredthe outcome is consistent
The use of thetreatment objectives.with the planned/desired
AUTOLIGN requires theThe use of the Orthotreatment objectives.se
user to have theSystem™ requires theof the Orchestrate 3D
necessary training anduser to have therequires the user to have
domain knowledge innecessary training andthe necessary training
the practice ofdomain knowledge inand domain knowledge
orthodontics, as well asthe practice ofin the practice of
to have received aorthodontics, as well asorthodontics, as well as
dedicated training in theto have received ato have received a
use of the software.dedicated training in thededicated training in the
use of the software.use of the software.

Comparison of Technological Characteristics

Both AUTOLIGN and 3Shape Ortho System™ Software (K171634) are stand-alone software module. They also import digital patient scan files. All the digital data format used for AUTOLIGN are those which the predicate device is already supporting. Like the other predicate devices, AUTOLIGN can be used to design dental cast.With respect to the AUTOLIGN, the 3D virtual model and finalized analysis can be exported to support orthodontic diagnosis and treatment planning of tooth movement. It supports STL files.

Comparison ItemAUTOLIGNOrtho System™Orchestrate 3DOrthodonticSoftwareCEREC OrthoSoftware
Stand-alone softwaremoduleOOOO
Imports scanned imageof patientOOOO
Can be used to designdental castsOOOO
Useful for diagnosis,treatment planning, andCAD designOOOO

{6}------------------------------------------------

Comparison ItemAUTOLIGNOrtho System™Orchestrate 3DOrthodonticSoftwareCEREC OrthoSoftware
Virtual planning of toothmovementOOOUnknown
Supports .stl filesOOOO
Managing patient and case base data
Creating, editing, deletingand copying patient dataOOOO
Creating, editing, deletingand copying case dataOOOO
Collection of study material
Surface scan for intra-oralscannerXOOO
Surface scan from STLfileOOOO
CT image dataXO (DICOM)XX
2D overlayO (JPG, BMP)O (PNG, JPG,BMP)XX
Creation of virtual 3Dvirtual dental modelsOOOO
Alignment of study material
Aligning surface scan orCT imageO (but CT imageis not available)OXO
Aligning cephalometricimagesOOXO
Alignment of 2D overlays(e.g. ideal arch)OOXO
Ability to check/adjustDICOM visibilityXOXO
DICOM scansegmentationXXXO
Measuring study material
2D measurement toolboxOOXO
3D measurement toolboxXOXO
Analyzing study material
Definition of dental Archshape & lengthOOOO
Wire lengthXOXO
Tooth widthmeasurementsOOXO
Comparison ItemAUTOLIGNOrtho System™Orchestrate 3DOrthodonticSoftwareCEREC OrthoSoftware
Tooth and gingivaseparation/segmentationOUnknownUnknownO
Bolton's analysisOOXO
Space analysisOOXO (Nance andMoyer spaceanalysis)
Overjet/overbiteXOOUnknown
Occlusal mappingOOOO
Treatment simulation
2D simulationXOXX
3D simulationOOOX
Virtual appliance design
Orthodontic appliancesearchOOXX
Orthodontic appliancevirtual preparationOOOX
Orthodontic appliancedesignOOOX
Orthodontic applianceexportOOOX

{7}------------------------------------------------

Most comparison items between the subject device and the primary predicate device are identical. However, the other items, which are not identical to the primary predicate device, such as CT image data availability, Ability to check/adjust DICOM visibility, 3D measurement toolbox, Wire length analyzing study material, and 2D treatment simulation, are identical to the other two reference devices.

Non-clinical Tests

The AUTOLIGN underwent software, hardware, integration, verification and validation testing in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005). The software passed the testing and performed per its intended use.

The AUTOLIGN does not include any physical device, accessory, or component with patient contacting intended use. Therefore, no sterilization, or shelf life analyses were included in support of substantial equivalence. However, we tested the compatibility of accessories from other manufacturers such as 3D scanners, CAD/CAM production machines, dental vacuum forming machines, and dental cast material sheets to mitigate concerns with biocompatibility of the proposed device.

{8}------------------------------------------------

Conclusion

Based on comparison of indications for use, scientific concept, technological features, technical data, performance testing, and software validation testing, the AUTOLIGN is found to be as safe and as effective as the primary predicate device. And All the results have been reviewed and approved, showing the AUTOLIGN to be substantially equivalent to the primary predicate device.

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.