(31 days)
The Parcus Knotless AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:
| Shoulder | Rotator Cuff Repair, Acromioclavicular Separation Repair,Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift orCapsulolabral Reconstruction, Deltoid Repair, SLAP LesionRepair. |
|---|---|
| Knee | Medial Collateral Ligament Repair, Lateral Collateral LigamentRepair, Posterior Oblique Ligament Repair, Extra CapsularReconstruction, Iliotibial Band Tenodesis, Patellar Ligament andTendon Avulsion Repair. |
| Foot/Ankle | Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction,Achilles Tendon Repair, Hallux Valgus Reconstruction, MetatarsalLigament Repair. |
| Elbow | Tennis Elbow Repair, Biceps Tendon Reattachment. |
| Hand/Wrist | Scapholunate Ligament Reconstruction, Ulnar or Radial CollateralLigament Reconstruction, TFCC. |
The Parcus Knotless AP Suture Anchors consist of ribbed anchor bodies in 2.8mm, 4.5mm, and 5.5mm diameters. The anchor body is comprised of Parcus Medical's Advanced Polymer (AP) which is a formulation of ß-TCP and PLGA that has a proven record as a safe and effective absorbable material that has been used in the industry for over 10 years. This material has been developed such that it will maintain the necessary strength to insert successfully and provide the necessary resistance to pull-out long enough for the body to heal and then, over time, break down, be absorbed into the body and eventually replaced by bone. The presence of the osteoconductive ß-TCP will allow for boney replacement at the insertion site as the anchor is absorbed.
Knotless AP Suture Anchors are provided with a single use suture threader and a single use driver. These devices are individually packaged and provided sterile.
The provided document is a 510(k) premarket notification for a medical device called "Parcus Knotless AP Suture Anchors." It primarily focuses on demonstrating substantial equivalence to a predicate device based on material composition, intended use, indications for use, biocompatibility, and mechanical testing.
This document describes a physical medical device (suture anchors) and does NOT contain information about an AI/ML-driven device. Therefore, it is impossible to extract the details requested in points 1-9, which relate specifically to the acceptance criteria and study design for an AI/ML device (e.g., test set sample size, expert ground truth, MRMC studies, standalone performance, training set details).
The "Summary Performance Data" section mentions "pull-out strength and cyclic loading studies" and comparison with "published acceptance criteria," but these are engineering performance metrics for a physical implant, not diagnostic or clinical performance metrics for an AI/ML algorithm.
In summary, the provided text does not describe an AI/ML device or its associated acceptance criteria and validation study as typically found in submissions for such technologies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Parcus Medical, LLC Paul Vagts Director of Regulatory Affairs 6423 Parkland Drive Sarasota, Florida 34243
Re: K192824
Trade/Device Name: Parcus Knotless AP Suture Anchors Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: October 1, 2019 Received: October 2, 2019
Dear Mr. Vagts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laurence D. Coyne, Ph.D. Acting Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K192824 Device Name Parcus Knotless AP Suture Anchors
Indications for Use (Describe)
The Parcus Knotless AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:
| Shoulder | Rotator Cuff Repair, Acromioclavicular Separation Repair,Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift orCapsulolabral Reconstruction, Deltoid Repair, SLAP LesionRepair. |
|---|---|
| Knee | Medial Collateral Ligament Repair, Lateral Collateral LigamentRepair, Posterior Oblique Ligament Repair, Extra CapsularReconstruction, Iliotibial Band Tenodesis, Patellar Ligament andTendon Avulsion Repair. |
| Foot/Ankle | Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction,Achilles Tendon Repair, Hallux Valgus Reconstruction, MetatarsalLigament Repair. |
| Elbow | Tennis Elbow Repair, Biceps Tendon Reattachment. |
| Hand/Wrist | Scapholunate Ligament Reconstruction, Ulnar or Radial CollateralLigament Reconstruction, TFCC. |
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Surgical Innovation | Customer Driv
| 510(k) Summary - K192824 | |
|---|---|
| Submitter: | Parcus Medical, LLC6423 Parkland DrSarasota, FL 34243 |
| Company Contact: | Paul VagtsPhone: (941)755-7965Fax: (941)755-6543 |
| Date Prepared: | October 1st, 2019 |
| Device Trade Name: | Parcus Knotless AP Suture Anchor |
| Common Name: | Suture Anchor |
| Device Class: | Class II |
| Classification Name: | Fastener, Fixation, Biodegradable, Soft Tissue 21 CFR888.3030 - Product Code MAI |
| Predicate Device: | The predicate device is the: |
| • Parcus Medical Twist AP Suture Anchor (K183501)cleared March 15, 2019) |
Device Description:
The Parcus Knotless AP Suture Anchors consist of ribbed anchor bodies in 2.8mm, 4.5mm, and 5.5mm diameters. The anchor body is comprised of Parcus Medical's Advanced Polymer (AP) which is a formulation of ß-TCP and PLGA that has a proven record as a safe and effective absorbable material that has been used in the industry for over 10 years. This material has been developed such that it will maintain the necessary strength to insert successfully and provide the necessary resistance to pull-out long enough for the body to heal and then, over time, break down, be absorbed into the body and eventually replaced by bone. The presence of the osteoconductive ß-TCP will allow for boney replacement at the insertion site as the anchor is absorbed.
Knotless AP Suture Anchors are provided with a single use suture threader and a single use driver. These devices are individually packaged and provided sterile.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for PARCUS MEDICAL, LLC. The word "PARCUS" is in large, bold, blue letters, with a red square to the left of the "P". Below "PARCUS" is the phrase "MEDICAL, LLC." in smaller, blue letters.
Surgical Innovation | Customer Driven
Intended Use:
The Parcus Knotless AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:
| Shoulder | Rotator Cuff Repair, Acromioclavicular Separation Repair, BankartLesion Repair, Biceps Tenodesis, Capsular Shift or CapsulolabralReconstruction, Deltoid Repair, SLAP Lesion Repair. |
|---|---|
| Knee | Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,Posterior Oblique Ligament Repair, Extra Capsular Reconstruction,Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair. |
| Foot/Ankle | Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction,Achilles Tendon Repair, Hallux Valgus Reconstruction, MetatarsalLigament Repair. |
| Elbow | Tennis Elbow Repair, Biceps Tendon Reattachment. |
| Hand/Wrist | Scapholunate Ligament Reconstruction, Ulnar or Radial CollateralLigament Reconstruction, TFCC. |
Substantial Equivalence Summary:
The Parcus Knotless AP Suture Anchors are very similar to the predicate devices in that they are comprised of the same materials, have identical intended use, and have identical indications for use. Pyrogenicity testing was conducted on a representative device from the Parcus AP Suture Anchor product line. The results of this testing met the acceptance criteria described in the FDA recognized standard and therefore it was concluded that the Knotless AP Suture Anchors do not raise any additional concerns regarding pyrogenicity. Extensive biocompatibility and mechanical testing has been conducted on the Knotless AP Suture Anchors or representative devices of the same material and has demonstrated substantial equivalence between the Parcus Knotless AP Suture Anchors and the predicate devices.
Summary Performance Data:
The entire scope of the proposed Knotless AP Suture Anchor product family was considered and worst-case scenarios for various concerns were identified. Devices, or representative devices of the same material, were subjected to biocompatibility and benchtop testing such as pull-out strength and cyclic loading studies. Based on the results of this testing and comparison with the performance of the predicate devices or published acceptance criteria, the Parcus Knotless AP Suture Anchors are determined by Parcus Medical to be substantially equivalent to the predicate devices.
N/A