K Number
K192738

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-11-26

(60 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ETEST® is a manual, quantitative technique for determination of antimicrobial susceptibility of non-fastidious Gram-negative and Gram-positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in ug/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.

Delafloxacin has been shown to be active against the aerobic microorganisms listed below according to the FDA label for this antimicrobial agent.

ETEST® DFX can be used to determine the MIC of Delafloxacin against the following microorganisms:

Active both in vitro and in clinical infections:

Gram-positive bacteria:

  • · Staphylococcus aureus (including methicillin-resistant and methicillin-susceptible strains)
  • Staphylococcus haemolyticus
  • Staphylococcus lugdunensis
  • Enterococcus faecalis

Gram-negative bacteria:

  • Pseudomonas aeruginosa
Device Description

ETEST® is a thin, inert and non-porous plastic strip carrying on one side the MIC reading scale in ug/mL, and on the other side a predefined antibiotic gradient.

When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacterial growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. The MIC value is read from the scale in terms of ug/mL at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip.

ETEST® Delafloxacin contains a range of delafloxacin from 0.002 to 32 u2/mL.

AI/ML Overview

This document describes the performance of the ETEST® Delafloxacin (DFX) device, a manual, quantitative technique for determining antimicrobial susceptibility. The study aims to demonstrate that the device is substantially equivalent to a predicate device (ETEST® Telavancin (TLA)) and meets pre-defined acceptance criteria based on established guidance and standards.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are derived from the "FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued on August 28, 2009" and "CLSI M100-S29 January 2019". While specific numerical targets for Essential Agreement (EA) and Category Agreement (CA) are not explicitly stated as "acceptance criteria" percentages in the text, the performance data presented in the tables are implied to meet these criteria, as the conclusion states the device demonstrated "acceptable performance" and "substantially equivalent performance". The data strongly suggests typical FDA acceptance thresholds for AST devices, which often require high percentages.

Performance MetricAcceptance Criteria (Implied by FDA Guidance/CLSI)Reported Device Performance (ETEST® Delafloxacin)
Essential Agreement (EA)High percentage (e.g., typically >90%)
Staphylococcus aureus96.5%
Staphylococcus haemolyticus100.0%
Staphylococcus lugdunensis100.0%
Enterococcus faecalis100.0%
Pseudomonas aeruginosa98.5%
Category Agreement (CA)High percentage (e.g., typically >90%)
Staphylococcus aureus93.0%
Staphylococcus haemolyticus93.5%
Staphylococcus lugdunensisNot applicable (no FDA breakpoints established)Not applicable
Enterococcus faecalis96.1%
Pseudomonas aeruginosa95.5%
ReproducibilityHigh percentage (e.g., typically 100% for best-case/worst-case)Best-case: 100%Worst-case: 100%
Quality ControlResults within range > 95% of the times testedResults within range > 95% of the times tested

2. Sample size used for the test set and the data provenance

The document mentions "fresh and stock clinical isolates, as well as a set of challenge strains" were used for external evaluations. However, the specific sample sizes (number of isolates) for the test set are not provided in the document.

The data provenance implies a multi-center study ("External evaluations were conducted") and the use of clinical and challenge strains, suggesting a mix of real-world (clinical) and engineered (challenge) samples. The country of origin for the data is not explicitly stated, but the submitter's address is in France. The study appears to be prospective in nature for data collection, as it describes conducting evaluations for the purpose of this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The ground truth for antimicrobial susceptibility testing (AST) devices is typically established through a reference method.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document. For AST devices, adjudication as typically understood for image-based diagnostic AI is not directly applicable. Instead, the "ground truth" is established by a well-defined and accepted reference method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done or mentioned. This device is a manual, quantitative technique for antimicrobial susceptibility testing, not an AI-assisted diagnostic tool that would typically involve human reader performance improvement with AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This refers to the performance of the ETEST® Delafloxacin device itself. The data presented in Table 1 ("Performance Characteristics for ETEST® Delafloxacin") represents the standalone performance of the device compared to the reference method. There is no "human-in-the-loop" component in the sense of an AI interpreting results and a human reviewing them; rather, a human reads the MIC value directly from the ETEST strip.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the performance study was established by the CLSI M07-A11 January 2018 broth microdilution reference method. This is considered the gold standard for determining Minimum Inhibitory Concentrations (MICs) in microbiology.

8. The sample size for the training set

The document does not specify a training set sample size. ETEST® products are physical diagnostic devices (strips), not machine learning algorithms that require a "training set" in the computational sense. The "training" for such devices would involve extensive research and development, and the performance is validated through studies like the one described.

9. How the ground truth for the training set was established

As there is no "training set" in the context of an AI algorithm, this question is not applicable. The development and optimization of the ETEST® system itself would have relied on established microbiological principles and validated methods for determining antimicrobial concentrations and their effects on bacterial growth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 26, 2019

BioMerieux SA Sophie Quiblier Regulatory Affairs Specialist 376. Chemin de L'Orme Marcy L'Etoile, 69280 Fr

Re: K192738

Trade/Device Name: ETEST Delafloxacin (DFX) (0.002-32 ug/mL) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: JWY Dated: September 26, 2019 Received: September 27, 2019

Dear Sophie Quiblier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar, Ph.D. (ABMM) Chief. General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a yellow and green bottom half. The word "BIOMÉRIEUX" is written in white letters across the blue top half of the circle.

ETEST® Delafloxacin (DFX) (0.002-32 µg/mL)

A. 510(k) Submission Information:

Submitter's Name:bioMerieux SA
Address:376 Chemin de l'Orme
69280 Marcy-l'Etoile, FRANCE
Contact Person:Sophie Quiblier
Regulatory Affairs Specialist
Phone Number:+33 (0)4 78 87 75 71
Date of Preparation:September 26th, 2019

B. Device Name:

Formal/Trade Name:ETEST® Delafloxacin (DFX)
(0.002 – 32 µg/mL)
Classification Name:21 CFR 866.1640
Manual Antimicrobial Susceptibility Test SystemsProduct Code: JWY
Common Name(s):ETEST® Delafloxacin; ETEST® DFX
C. Predicate Device:ETEST® Telavancin (TLA)(0.002 – 32 µg/mL) (K180936)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for bioMérieux, a French multinational biotechnology company. The logo is a circle divided into two sections. The top half of the circle is a solid dark blue color, and the text "BIOMÉRIEUX" is written in white, sans-serif font in the center of the blue section. The bottom half of the circle is a gradient that transitions from yellow to green.

D. Device Description:

ETEST® is a thin, inert and non-porous plastic strip carrying on one side the MIC reading scale in ug/mL, and on the other side a predefined antibiotic gradient.

When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacterial growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. The MIC value is read from the scale in terms of ug/mL at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip.

ETEST® Delafloxacin contains a range of delafloxacin from 0.002 to 32 u2/mL.

E. Intended Use:

ETEST® is a manual, quantitative technique for determination of antimicrobial susceptibility of non-fastidious Gram-negative and Gram-positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in ug/mL) of different antimicrobial agents against microorganisms tested on agar media after overnight incubation.

Delafloxacin has been shown to be active against the aerobic microorganisms listed below according to the FDA label for this antimicrobial agent.

ETEST® DFX can be used to determine the MIC of Delafloxacin against the following microorganisms:

Active both in vitro and in clinical infections:

Gram-positive bacteria:

  • · Staphylococcus aureus (including methicillin-resistant and methicillin-susceptible strains)
  • Staphylococcus haemolyticus
  • Staphylococcus lugdunensis
  • Enterococcus faecalis

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for BIOMÉRIEUX. The logo is a circle with a blue top half and a yellow-green bottom half. The word "BIOMÉRIEUX" is written in white letters in the center of the blue half of the circle.

Gram-negative bacteria:

  • Pseudomonas aeruginosa

F. Summary of the technological characterictics of the new device in comparison to those of the predicate device.

The similarities and differences of ETEST® Delafloxacin (DFX) when compared to the predicate device, ETEST® Telavancin (TLA)(K180936) are described in the table below:

Test DevicePredicate Device
Similarities
ETEST® Delafloxacin (DFX)(0.002-32 µg/mL)ETEST® Telavancin (TLA)(0.002-32 µg/mL)
Intended UseETEST® is a manual, quantitative technique for determination of antimicrobial susceptibility of non-fastidious Gram-negative and Gram-positive aerobic bacteria and fastidious bacteria. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC, in µg/mL) of different antimicrobial agents against microorganisms tested on agar media after overnightETEST® is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is
incubation.Delafloxacin has been shown tobe active against the aerobicmicroorganisms listed belowaccording to the FDA label forthis antimicrobial agent.used to determine the MinimumInhibitory Concentration(MIC), in μg/mL, of differentantimicrobial agents againstmicroorganisms as tested onagar media using overnightincubation.
ETEST® DFX can be used todetermine the MIC ofDelafloxacin against thefollowing microorganisms:Telavancin has been shown tobe active against the Gram-positive aerobicmicroorganisms listed belowaccording to the FDA label forthis antimicrobial agent.
Active both in vitro and inclinical infections:Staphylococcus aureus (including methicillin-resistant and methicillin-susceptible strains) Staphylococcus haemolyticus Staphylococcus lugdunensis Enterococcus faecalis Pseudomonas aeruginosaActive both in vitro and inclinical infections:Staphylococcus aureus (including methicillin resistant isolates) Enterococcus faecalis: (vancomycin-susceptible only)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a blue semi-circle on top of a green semi-circle. The company name, "BIOMÉRIEUX", is written in white, sans-serif font inside the blue semi-circle.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for bioMérieux. The logo consists of a blue semi-circle at the top, with the word "BIOMÉRIEUX" in white, sans-serif font centered within the semi-circle. The font is bolded and the logo is simple and clean.

Clinical &ChallengePerformanceDataAcute Bacterial Skin and SkinStructure Infections (ABSSSI)indication
Staphylococcus aureus(methicillin resistant andmethicillin-susceptibleisolates):EA = 96.5%CA = 93.0%Staphylococcus aureus:EA = 98.4%CA = 97.9%
Staphylococcus haemolyticus:EA = 100%CA = 93.5%• Enterococcus faecalis:(vancomycin-susceptibleonly)EA = 91.6%CA = 97.6%
Staphylococcus lugdunensisEA = 100%CA : Not applicable
Enterococcus faecalis:EA = 100%CA = 96.1%
Pseudomonas aeruginosa:
EA = 98.5%
CA = 95.5%
Community Acquired BacterialPneumonia (CABP) indication
Staphylococcus aureus(methicillin-susceptibleisolates):
EA = 97.3%
CA = 91.8%
Pseudomonas aeruginosa:
EA = 98.5%
CA = 95.5%
ReproducibilityBest-case: 100%Best-case: 100%
Worst-case: 100%Worst-case: 100%
Quality ControlResults within range > 95% ofthe times tested.Results within range > 95% ofthe times tested.
Meets GuidanceDocumentPerformanceRequirementsYesYes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a blue semi-circle at the top, with the company name "BIOMÉRIEUX" in white, sans-serif font inside the semi-circle. Below the blue semi-circle is a gradient of green and yellow.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for bioMérieux. The logo consists of a blue semi-circle at the top, with the company name "BIOMÉRIEUX" in white, sans-serif font inside the semi-circle. Below the blue semi-circle is a gradient of yellow and green.

Differences
AntimicrobialAgentDelafloxacinTelavancin
Claimed speciesStaphylococcus aureus (including methicillin-resistant and methicillin-susceptible strains) Staphylococcus haemolyticus Staphylococcus lugdunensis Enterococcus faecalis Pseudomonas aeruginosaStaphylococcus aureus (including methicillin resistant isolates) Enterococcus faecalis : (vancomycin-susceptible only)

G. Performance Overview

ETEST® Delafloxacin (DFX) (0.002-32 ug/mL) demonstrated substantially equivalent performance when compared with the CLSI M07-A11 January 2018 broth microdilution reference method, following rules as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued on August 28, 2009 and following specifications as defined in CLSI M100-S29 January 2019.

This Premarket Notification (510/k)) presents data in support of ETEST® Delafloxacin (DFX) (0.002-32 ug/mL) for: Staphylococcus aureus (including methicillin-resistant and methicillinsusceptible strains), Staphylococcus haemolyticus, Staphylococcus lugdunensis, Enterococcus faecalis and Pseudomonas aeruginosa.

External evaluations were conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to establish the performance of ETEST® Delafloxacin (DFX) (0.002-32 µg/mL) by comparing with the CLSI broth microdilution reference method.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for a company called BIOMÉRIEUX. The logo is a circle with a blue top half and a yellow-green gradient bottom half. The company name is written in white letters in the blue portion of the circle.

ETEST® Delafloxacin (DFX) (0.002-32 ug/mL) demonstrated acceptable performance as presented in Table 1 below:

% Essential Agreement(EA)a)% Category Agreement(CA)
Staphylococcus aureus (methicillinresistant and methicillin-susceptibleisolates) b), c)96.593.0
Staphylococcus haemolyticus b)100.093.5
Staphylococcus lugdunensis b)100.0Not applicable d)
Enterococcus faecalis b), c)100.096.1
Pseudomonas aeruginosa b)98.595.5

Table 1: Performance Characteristics for ETEST® Delafloxacin

Reproducibility and Quality Control demonstrated acceptable results.

Notes:

a) EA = % of MIC values within ± 1 dilution of the reference method.

b) The optional inoculator and ETEST® strip applicator can be used for plate inoculation and applying ETEST® strips onto agar media. In the ETEST® Delafloxacin clinical studies, swabs and the Inoculator RETRO C80™ were used for plate inoculation/streaking and forceps and the Vacuum Pen NEMA C88™ were used for ETEST® strip application.

c) ETEST® Delafloxacin MIC values tended to be in exact agreement or one doubling dilution higher when testing Staphylococcus aureus and Enterococcus faecalis compared to the reference broth microdilution method.

d) Category Agreement is not calculated because Delafloxacin breakpoints for S. lugdunensis were not established by the FDA.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image is a logo for BIOMÉRIEUX. The logo is a circle with a blue top half and a yellow-green gradient bottom half. The word "BIOMÉRIEUX" is written in white letters in the center of the blue half of the circle.

Conclusion:

The performance data presented in this submission support a substantial equivalence decision. ETEST® Delafloxacin (DFX) (0.002-32 µg/mL) is substantially equivalent to ETEST® Telavancin (TLA) (0.002-32 µg/mL) (K180936).

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).