K Number
K192715

Validate with FDA (Live)

Date Cleared
2020-01-07

(102 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CRF Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue.

Device Description

The CRF Radiofrequency Ablation System consists of an RF generator, active electrode, grounding pad, and peristallic punp for electrode cooling. This device is designed to cal tissue heating at the tip of the electrode causing coagulation and ablation of tissue. The CRF Generator supplies radiofrequency energy at 480 kHz while monitoring both the tissue imperature at the CRF electrode tip. The CRF Generator monitors the power, resistance, current and temperature during energy delivery to the paient.

The CRF Electrodes are sterile, single-use, hand-held for use in radiofrequency ablation procedures. The electrodes are surgically invasive, transient devices in patient contact typically for less than one hour. Cooling of the electrodes is provided by chilled sterile saline or water that is pumped through the electrode and out via outflow tubing. This is an enclosed system within the electrode and the saline does not contact. The CRF Generator automatically monitors impedance and temperature and adjusts RF output accordingly. The CRF Radiofrequency Ablation System is used in a healthcare facility or hospital setting.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (CRF Radiofrequency Ablation System), not an AI/ML device. Therefore, the traditional acceptance criteria and study designs relating to AI/ML performance metrics (such as sensitivity, specificity, AUC, MRMC studies, and expert adjudication as ground truth) are not applicable here.

The document describes the device's substantial equivalence to a predicate device based on non-clinical data.

However, to address the prompt's structure as best as possible within the context of the provided document, I will reframe some of the categories to fit the information given, focusing on the comparisons and testing performed for this physical medical device.

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present "acceptance criteria" in the traditional sense of numerical thresholds for AI/ML performance (e.g., sensitivity > X%). Instead, the "acceptance" for this 510(k) submission relies on demonstrating substantial equivalence to a predicate device through various non-clinical tests and comparing technological characteristics. The reported device performance is presented as compliance with standards or successful demonstration of equivalence in specific tests.

Acceptance Criteria (Demonstration for Equivalence)Reported Device Performance
Technological Characteristics Comparison
Principle of Operation: Identical to predicateIdentical
Generator Electrical Classification: IdenticalIdentical
Generator Output Frequency: IdenticalIdentical (480 kHz)
Generator Waveform: IdenticalIdentical (Sine wave)
Generator Waveform Crest Factor: IdenticalIdentical (1.4)
Generator Impedance: IdenticalIdentical (Up to 1000 ohms)
Generator Temperature Monitor: IdenticalIdentical (Thermocouple)
Generator Footswitch control: IdenticalIdentical (Available)
Max number of Electrodes: IdenticalIdentical (Up to 3)
Temp probe sensor: IdenticalIdentical (Type T Thermocouple)
Temp Probe & Introducer length/diameter: IdenticalIdentical (20cm; 0.9, 1.3mm)
Active electrode material: IdenticalIdentical (Stainless Steel)
Electrode insulation material: IdenticalIdentical (Polyester)
Electrode sterilization: IdenticalIdentical (ETO Sterilization)
Electrode shelf life: IdenticalIdentical (4 years)
Electrode usage: IdenticalIdentical (Single Use)
Pump type: IdenticalIdentical (Peristaltic pump)
Tubing Set configuration: IdenticalIdentical (Two-piece set)
Tubing Set material: IdenticalIdentical (PVC tubing)
Tubing Set sterilization: IdenticalIdentical (ETO Sterilization)
Tubing Set shelf life: IdenticalIdentical (4 years)
Tubing Set usage: IdenticalIdentical (Single Use)
Ground Pad material: IdenticalIdentical (Conductive hydrogel)
Ground Pad usage: IdenticalIdentical (Single Use)
Ground Pad shelf life: IdenticalIdentical (2 years)
Ground Pad regulatory clearance: IdenticalIdentical (510(k) clearance)
Non-Clinical Performance Testing
Mechanical testing: Compliance with IEC 60601-1 & IEC 60601-2-2, internal test requirementsPerformed to ensure compliance
Electrical testing: Compliance with IEC 60601-1 & IEC 60601-2-2Performed to ensure compliance
Electromagnetic compatibility: Compliance with IEC 60601-1-2Completed to ensure compliance
Benchtop ablation testing: Substantially equivalent ablation performance to predicateComparative ablation performed using ex vivo tissue model to demonstrate substantial equivalence
Usability testing: Compliance with IEC 62366 & IEC 60601-1-6Performed to verify and validate usability requirements
Biocompatibility testing: Compliance with ISO 10993-1, including cytotoxicity, systemic toxicity, irritationAll CRF devices demonstrated compliance with listed requirements
Sterilization validation testing: SAL of 10^-6 in accordance with ISO 11135Carried out using overkill method and half-cycles, providing SAL of 10^-6
Packaging and shelf life testing: Performance (sterile packaging) & accelerated/real-time aging (shelf life)Conducted based on ASTM F1980-16 for accelerated aging. Real-time aging will support results.
Software validation testing: Compliance with IEC 62304 and FDA Guidance (2005)Conducted in accordance with standards and guidance

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not applicable in the context of an AI/ML device for which this question is typically posed. For this medical device, various non-clinical tests were conducted:
    • Mechanical, Electrical, EMC: Implied to be on a sufficient number of device units to establish compliance with standards. Details on the exact number of units are not provided but are typically part of a detailed testing report not fully included in this summary.
    • Benchtop Ablation: Performed using an "ex vivo tissue model." The number of tissues or ablation cycles is not specified.
    • Usability, Biocompatibility, Sterilization, Packaging/Shelf Life, Software: Applied to the device and its components.
  • Data Provenance: Not applicable in the AI/ML sense. All data appears to be from prospective non-clinical testing conducted specifically for this submission by the manufacturer or authorized labs (e.g., benchtop tests, lab tests for electrical safety, biocompatibility, etc.). Country of origin of data is not specified but would typically be where the manufacturing and testing facilities are located.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is not an AI/ML device requiring expert-established ground truth for image interpretation or diagnosis. Ground truth for this device is established through compliance with engineering standards (e.g., electrical safety, mechanical requirements) and observed physical outcomes in benchtop models (e.g., ablation performance). The "experts" involved are likely engineers, quality assurance personnel, and regulatory affairs specialists who conduct and interpret the non-clinical tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This device does not involve a diagnostic task requiring human-expert adjudication for a test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, and no MRMC study was conducted. No human "readers" are involved with the device in this capacity.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm. Therefore, "standalone" performance for an algorithm is not relevant. The device operates with a human operator (surgeon/clinician) in a "human-in-the-loop" manner by design, but its performance is evaluated in non-clinical settings without directly assessing human improvement from an AI perspective.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's assessment of substantial equivalence relies on:

  • Compliance with codified U.S. and international standards (e.g., IEC 60601 series, ISO 11135, ASTM F1980-16, ISO 10993-1).
  • Direct measurement and observation of physical characteristics and performance in laboratory and benchtop settings (e.g., electrical parameters, ablation zone size in ex vivo tissue, material properties, sterilization efficacy).
  • Comparison to the established characteristics and performance of a legally marketed predicate device.

8. The sample size for the training set

Not applicable. This device does not involve AI/ML algorithms that require a "training set."

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth established for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG" and "ADMINISTRATION".

January 7, 2020

Cambridge Interventional LLC Michael Arnold President 78 Cambridge Street Burlington, Massachusetts 01803

Re: K192715

Trade/Device Name: CRF Radiofrequency Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 12, 2019 Received: November 19, 2019

Dear Michael Arnold:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K192715

Device Name

CRF Radiofrequency Ablation System

Indications for Use (Describe)

The CRF Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

CRF Radiofrequency Ablation System

Submitter:

Cambridge Interventional LLC 78 Cambridge Street Burlington MA 01803

Telephone: 617-365-6899 Contact Name: Michael Arnold Email: marnold@cambint.com Date prepared: November 12, 2019

Device Name and Classification

Model Name: CRF Radiofrequency Ablation System Common Name: Electrosurgical Cutting and Coagulation Device and Accessories Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Regulation Number: 21 CFR 878.4400 Product Code: GEI Device Class: II

Name of Predicate Device

Valleylab Cool-tip™ Switching Controller (K070446) - Product Code GEI, Class II

Name of Reference Device

Valleylab Cool-tip™ Radiofrequency Ablation System (K052796) - Product Code GEI, Class II

Description of Device

The CRF Radiofrequency Ablation System consists of an RF generator, active electrode, grounding pad, and peristallic punp for electrode cooling. This device is designed to cal tissue heating at the tip of the electrode causing coagulation and ablation of tissue. The CRF Generator supplies radiofrequency energy at 480 kHz while monitoring both the tissue imperature at the CRF electrode tip. The CRF Generator monitors the power, resistance, current and temperature during energy delivery to the paient.

The CRF Electrodes are sterile, single-use, hand-held for use in radiofrequency ablation procedures. The electrodes are surgically invasive, transient devices in patient contact typically for less than one hour. Cooling of the electrodes is provided by chilled sterile saline or water that is pumped through the electrode and out via outflow tubing. This is an enclosed system within the electrode and the saline does not contact. The CRF Generator automatically monitors impedance and temperature and adjusts RF output accordingly. The CRF Radiofrequency Ablation System is used in a healthcare facility or hospital setting. The table below lists components for the CRF Radiofrequency Ablation System.

Catalog No.Description of ComponentProvided Sterile?
CRFCRF Radiofrequency GeneratorNo
CRF-C1020RF Cannula, 10 cm, 2 cm TipYes
CRF-C1030RF Cannula, 10 cm, 3 cm TipYes
CRF-C1510RF Cannula, 15 cm, 1 cm TipYes
CRF-C1520RF Cannula, 15 cm, 2 cm TipYes
CRF-C1530RF Cannula, 15 cm, 3 cm TipYes
CRF-C2020RF Cannula, 20 cm, 2 cm TipYes
CRF-C2030RF Cannula, 20 cm, 3 cm TipYes
CRF-C2040RF Cannula, 20 cm, 4 cm TipYes
CRF-C2050RF Cannula, 20 cm, 5 cm TipYes

{4}------------------------------------------------

Catalog No.Description of ComponentProvided Sterile?
CRF-C2520RF Cannula, 25 cm, 2 cm TipYes
CRF-C2530RF Cannula, 25 cm, 3 cm TipYes
CRF-E10RF Electrode, Cooled, Blunt Tip, 10 cmYes
CRF-E15RF Electrode, Cooled, Blunt Tip, 15 cmYes
CRF-E20RF Electrode, Cooled, Blunt Tip, 20 cmYes
CRF-E25RF Electrode, Cooled, Blunt Tip, 25 cmYes
CRF-N1005RF Electrode, Non-Cooled, Sharp Tip, 10cm, 0.5cm TipYes
CRF-N1510RF Electrode, Non-Cooled, Sharp Tip, 15cm, 1cm TipYes
CRF-N2020RF Electrode, Non-Cooled, Sharp Tip, 20cm, 2cm TipYes
CRF-N2520RF Electrode, Non-Cooled, Sharp Tip, 25cm, 2cm TipYes
CRF-TNTemperature Probe, CannulaYes
CRF-TPTemperature Probe, Blunt ProbeYes
CRF-U1020RF Electrode, Cooled, Sharp Tip, 10cm, 2cm TipYes
CRF-U1030RF Electrode, Cooled, Sharp Tip, 10cm, 3cm TipYes
CRF-U1507RF Electrode, Cooled, Sharp Tip, 15cm, 0.7cm TipYes
CRF-U1510RF Electrode, Cooled, Sharp Tip, 15cm, 1cm TipYes
CRF-U1520RF Electrode, Cooled, Sharp Tip, 15cm, 2cm TipYes
CRF-U1530RF Electrode, Cooled, Sharp Tip, 15cm, 3cm TipYes
CRF-U2020RF Electrode, Cooled, Sharp Tip, 20cm, 2cm TipYes
CRF-U2030RF Electrode, Cooled, Sharp Tip, 20cm, 3cm TipYes
CRF-U2050RF Electrode, Cooled, Sharp Tip, 20cm, 5cm TipYes
CRF-U2525RF Electrode, Cooled, Sharp Tip, 25cm, 2.5cm TipYes
CRF-U2535RF Electrode, Cooled, Sharp Tip, 25cm, 3.5cm TipYes
CRF-FSFootswitchNo
CRF-PUMPPumpNo
CRF-A3ECable, CRF electrodes to CRF generatorNo
CRF-AMECable, CRF electrode to Cool-tip generatorNo
CRF-AMGCable, CRF-GP to Cool-tip generatorNo
CRF-SCGStorage Case for CRF GeneratorNo
CRF-SCPStorage Case for PumpNo
CRF-TUBETubing Set for PumpYes
CRF-GBCIntroducer for CRF-CYes
CRF-GBUIntroducer for CRF-UYes
CRF-GPGround Pad, DisposableNo

Indications for Use

The CRF Radiofrequency Ablation System is intended for use in percutaneous, intraoperative coagulation and ablation of tissue.

The Indications for Use statement is consistent with the predicate Cool-tip™ device, but without specific indications.

Comparison to Predicate Device

Compared to the predicate Cool-tip™ device, the CRF Radiofrequency Ablation System has similar technological characteristics and operating principle. There are no significant design or performance specifications that affect safety of the subject device. There is substantial equivalence between the subject device and the predicate with respect to indications and intended use. The CRF Radiofrequency Ablation System is substantially equivalent to the legally marketed predicate device with respect to indications for use and technology characteristics. The table below presents side-by-side comparisons for each major component for the subject and predicate device.

{5}------------------------------------------------

ParameterSubject DevicePredicate DeviceComparison
DeviceCRF Radiofrequency Ablation SystemValleylab Cool-tip™ SwitchingController (K070446)N/A
IndicationThe CRF Radiofrequency AblationSystem is indicated for use inpercutaneous, laparoscopic andintraoperative coagulation and ablationof tissue.The Cool-tip™ RF Switching Controlleris to be used with the Cool-tip™ RFSystem and is intended for use inpercutaneous, laparoscopic andintraoperative coagulation and ablationof tissue, such as partial or completeablation of nonresectable liver tumors,and osteoid osteoma tumors withinbone.Different
Principle of OperationProvide RF energy to patient-contactingTC Electrode for tissue ablation.Provide RF energy to patient-contactingTC Electrode for tissue ablation.Identical
Classification878.4400878.4400Identical
Device ClassGEIGEIIdentical
Generator ElectricalClassificationClass 1, IPX0, Type BFClass 1, IPX0, Type BFIdentical
Generator Output Frequency480 kHz480 kHzIdentical
Generator waveformSine waveSine waveIdentical
Generator waveform crest factor1.41.4Identical
Generator Rated PowerUp to 370wUp to 200wDifferent
Generator ImpedanceUp to 1000 ohmsUp to 1000 ohmsIdentical
Generator TimerUp to 15:00 minUp to 30:00 minDifferent
Generator Temperature MonitorThermocoupleThermocoupleIdentical
Generator Footswitch controlAvailableAvailableIdentical
TC Electrode length10, 15, 20, 25cm10, 14.4, 15, 20, 25cmDifferent
TC Electrode tip exposure0.5, 0.7, 1, 2, 2.5, 3, 3.5, 4, 5cm0.7, 1, 2, 2.5, 3, 4cmDifferent
TC Electrode tip diameter0.9, 1.5, 1.8mm1.5mmDifferent
TC Electrode CoolingMechanismElectrode cooling is provided by sterilewater, which is pumped through theinflow tubing into the Electrode shaft,and out through the outflow tubing.Electrode cooling is provided by sterilewater, which is pumped through theinflow tubing into the Electrode shaft,and out through the outflow tubing.Identical
Maximum number of Electrodesconnected simultaneouslyUp to 3Up to 3Identical
Temp probe sensorType T ThermocoupleType T ThermocoupleIdentical
Temp Probe & Introducer length20cm20cmIdentical
Temp Probe & Introducerdiameter0.9, 1.3mm0.9, 1.3mmIdentical
Active electrode materialStainless SteelStainless SteelIdentical
Electrode insulation materialPolyesterPolyesterIdentical
Electrode sterilizationETO SterilizationETO SterilizationIdentical
Electrode shelf life4 years4 yearsIdentical
Electrode usageSingle UseSingle UseIdentical
Pump typePeristaltic pumpPeristaltic pumpIdentical
Pump specification80-120 ml/min80-140 ml/minDifferent
Tubing Set ConfigurationTwo-piece set (inflow and outflowcomponents)Two-piece set (inflow and outflowcomponents)Identical
Tubing Set MaterialPVC tubingPVC tubingIdentical
Tubing Set sterilizationETO SterilizationETO SterilizationIdentical
Tubing Set shelf life4 years4 yearsIdentical
Tubing Set usageSingle UseSingle UseIdentical
Ground Pad materialConductive hydrogelConductive hydrogelIdentical
Ground Pad usageSingle UseSingle UseIdentical
Ground Pad shelf life2 years2 yearsIdentical
Ground Pad regulatory clearance510(k) clearance (K000079)510(k) clearance (K030697)Identical

{6}------------------------------------------------

K192715

Summary of Non-Clinical Data

Performance testing of the device was completed to demonstrate substantial equivalence of the CRF Radiofrequency Ablation System to the identified predicate device. Non-clinical animal testing of the device was not required to establish substantial equivalence. The device components were subjected to the following verification and validation testing, as required:

• Mechanical testing: Mechanical verification testing was performed on subject devices to ensure compliance with current IEC 60601-1 and IEC 60601-2-2 standards and the Company's internal test requirements.

· Electrical testing: Electrical verification testing was conducted for the subject device to ensure compliance with current IEC 60601-1 and IEC 60601-2-2 electrical requirements.

· Electromagnetic compatibility: Electromagnetic compatibility testing was completed for the subject device to ensure compliance with the current IEC 60601-1-2 standard.

· Benchtop ablation testing. Comparative ablation verification testing was performed using an ex vivo tissue model to demonstrate the substantially equivalent ablation performance of the subject and predicate devices.

• Usability testing: Testing was performed to verify and validate the usability requirements of the subject devices as a system in compliance with IEC 62366 and IEC 60601-1-6. Elements captured included normal use errors, and testable requirements for primary operating functions and frequent use scenarios.

· Biocompatibility testing: Biocompatibility testing of patient contacting CRF devices was performed in accordance with the FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" including cytotoxicity, systemic toxicity and irritation testing. All CRF devices demonstrated compliance with listed requirements.

· Sterilization validation testing: ETO sterilization validation for CRF devices was carred out using the overkill method and halfcycles in accordance with the current ISO 11135 standard to provide a SAL of 10-6.

• Packaging and shelf life testing: Sterile packaging performance testing was conducted for the subject devices in their final packaging. Shelf life testing for sterile subject devices and packaging was conducted based on an accelerated aging protocol in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. Real-time age testing will be performed to add to and support the results of the accelerated age testing.

· Software validation testing:

The CRF radiofrequency generator includes software developed according to the Company's software development process in compliance with the current IEC 62304 standard. Software verification testing was conducted in accordance with the FDA Gudance for Content of Premarket Submissions for Software Contained in Medical Devices (2005) and IEC 62304.

Summary of Clinical Data

Clinical testing of the device was not required to establish substantial equivalence.

Conclusion

The CRF Radiofrequency Ablation System is substantially equivalent to the predicate device cleared in K070446. The subject and predicate devices share the same fundamental scientific technology, including principles of operation and mechanism of action. Differences in design and technological characteristics between the subject device do not raise any new questions of safety and effectiveness. The intended use of the subject device has not changed. The comparison of technological characteristics, non-clinical performance data, electrical safety and electromagnetic compatibility testing demonstrates that the CRP Radiofrequency Ablation System is substantially equivalent to the previously cleared predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.