K Number
K192709
Manufacturer
Date Cleared
2020-12-15

(445 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The Fixone All Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, they may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the healing period, the suture anchor systems stabilize the damaged soft tissue.

Device Description

Fixone All suture anchor consists of one "fix Suture" and two or three non-absorbable Sutures. The nonabsorbable suture is manufactured from UHMWPE and PET fibers. Fixone All suture anchor is implanted using its self-punching option. They are pre-loaded on a handled insertion device is provided sterile, for single use only. This device is medical device. Prescription use only. This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 4 models. It provide non-sterile (user must sterilization before use).

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Fixone All Suture Anchor." This document primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a study proving that the device meets specific acceptance criteria for AI/ML performance.

The document details bench testing for the physical device, including mechanical properties and biocompatibility, but does not describe an AI/ML-based device or a study involving AI/ML performance. Therefore, most of the requested information regarding AI/ML acceptance criteria, test sets, expert ground truth, MRMC studies, or AI algorithm-only performance cannot be extracted from this text.

The "Performance data" section (Section 6) refers to laboratory bench tests for the physical product, not a study evaluating an AI/ML component.

Therefore, I cannot provide the requested information regarding AI/ML validation studies from the given input.

The document states:

  • Device: Fixone All Suture Anchor
  • Purpose: Reattach soft tissue to bone in orthopedic surgical procedures.
  • Nature of evidence provided for FDA clearance: Bench testing (mechanical and biocompatibility) comparing it to a predicate device (K133224, ConMed Corporation / Y-Knot RC All-Suture Anchor w/Two and Three #2).

If this question was mistakenly associated with an AI/ML device, and the intent was to ask about the acceptance criteria and studies for this physical medical device, here's what can be inferred:


Acceptance Criteria and Study for Fixone All Suture Anchor (Physical Medical Device)

Based on the provided FDA 510(k) summary, the device is a physical suture anchor, not an AI/ML diagnostic or therapeutic device. The "acceptance criteria" here refer to demonstrating substantial equivalence through various bench tests.

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemRequirements (Acceptance Criteria)Results (Reported Performance)
Physical/Mechanical Tests
SurfaceWhen examined by normal vision, the surface shall be free from defects.Pass
Nominal size (Exclude suture)When tested by Vernier calipers, micrometer or tapeline, the measurement should be in accordance with standard.MD2018-00210 (Implies it met the standard)
Insertion test (Comparison with predicate)Must perform in a substantially equivalent manner to the predicate device in 10pcf/40pfc Polyurethane block, self-punching, @ 90° implantation angle.Results provided (AJU_M, implying successful comparison with predicate "Yknot")
Pull-out test (Comparison with predicate)Must perform in a substantially equivalent manner to the predicate device in 10pcf/40pfc Polyurethane block, self-punching, @ 90° pull-out angle with MTS Bionix servohydraulic tester.Results provided (D20201006_PT(Js, Yknot), implying successful comparison with predicate "Yknot")
Fatigue test (Comparison with predicate)Must perform in a substantially equivalent manner to the predicate device in 10pcf/40pfc Polyurethane block, self-punching, with same cycle loading and rate.Results provided (Js, Yknot), implying successful comparison with predicate "Yknot")
Biocompatibility Tests
CytotoxicityISO 10993-5 (Tests for in vitro cytotoxicity)Pass
Systemic toxicity testISO 10993-11 (Tests for systemic toxicity)Pass
Pyrogen TestISO 10993-11 (Pyrogen test)Pass
Intracutaneous reactivity testISO 10993-10 (Tests for irritation and skin sensitization)Pass
Maximization sensitizationISO 10993-10 (Maximization test for delayed hypersensitivity)Pass
Genotoxicity testISO 10993-3, OECD 471 (Bacterial reverse mutation test)Pass
Implantation testISO 10993-6 (Tests for local effects after implantation, Annex D test for implantation in bone)Pass
Hemolysis testISO 10993-4 (Selection of tests for interactions with blood)Pass
Subchronic toxicityISO 10993-11 (Tests for systemic toxicity)Pass
Sterilization & Packaging
SterilizationISO 11135:2014 (Ethylene oxide sterilization requirements)ME-SVP-1802 (Implies meeting the standard)
Shelf Life and PackagingISO11607-2 / EN 868-5 / ASTM F1980-16 / ASTM F88-15 / ASTM F1929:1998AJ-SL1702 (Implies meeting the standard)

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: For the mechanical tests (Insertion, Pull-out, Fatigue), the text explicitly states "10pcf/40pfc Polyurethane block." This refers to specific foam densities used for testing bone-like properties. While it mentions "10pcf/40pfc," it doesn't specify the number of samples (individual devices) tested for each condition. For biocompatibility tests, typically multiple replicates are used, but specific numbers are not provided.
  • Data Provenance: The tests were "Bench tests," meaning laboratory-based experimental data. No information about country of origin for the "data" itself, as it's not patient-derived data. The manufacturer is AJU Pharm Co., Ltd. from Korea.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This is for a physical device and involves objective measurements in a lab setting (e.g., force, dimensions, biological reactions based on ISO standards), not expert interpretation of complex data (like medical images).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. Ground truth is established through standardized laboratory testing parameters and measurements, not subjective expert opinion requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI/ML device.

7. The type of ground truth used:

  • "Ground truth" for this device is based on objective physical and chemical measurement standards and comparison to the established performance of a legally marketed predicate device. For instance, the ISO 10993 series for biocompatibility provides the "ground truth" criteria.

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device that requires a "training set."

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

December 15, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

AJU Pharm Co., Ltd. % Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K192709

Trade/Device Name: Fixone All Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: September 20, 2019 Received: September 27, 2019

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K192709

Device Name Fixone All Suture Anchor

Indications for Use (Describe)

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The Fixone All Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, they may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the suture anchor systems stabilize the damaged soft tissue.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[as required by 807.92(c)]

1. Applicant

  • Company : AJU Pharm Co.,Ltd. 1)
    1. Address : A-207, 697, Pangyo-ro, Seongnam-si, Gyeonggi-do, Korea
    1. Tel : 82-31-765-4420
    1. Fax : 82-31-602-7818
  • Prepared date : Sep. 01, 2019 5)
  • ୧) Contact person : Peter Chung, 412-512-8802
    1. Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
    1. Submission date : Mar.11, 2020

2. Device Information

    1. Trade name : Fixone All Suture Anchor
    1. Common name : Non-absorbable Suture Anchor
    1. Regulation name : Fastener, Fixation, Nondegradable, Soft tissue
    1. Product code : MBI
    1. Regulation number : 888.3040
  • ୧) Class of device : Class II
    1. Panel : Orthopedic

3. The legally marketed device to which we are claiming equivalence

K133224, ConMed Corporation / Y-Knot RC All-Suture Anchor w/Two and Three #2

4. Device description

Fixone All suture anchor consists of one "fix Suture" and two or three non-absorbable Sutures. The nonabsorbable suture is manufactured from UHMWPE and PET fibers. Fixone All suture anchor is implanted using its self-punching option. They are pre-loaded on a handled insertion device is provided sterile, for single use only. This device is medical device. Prescription use only.

This device is could used with instrument that manufactured by Aju Pharm Co.,Ltd. It is consist of 4 models. It provide non-sterile (user must sterilization before use).

Product nameInstrument / 4 model codes including Fixone.I.B-Awl450a
Model nameFixone.I.B-Awl450a, Fixone.I.B-Awl550a, Fixone.I.B-Awl450ak, Fixone.I.B-Awl550a
Intended useThe instrument for making a hole in the bone
ManufacturerAjupharm / Korea
SterilizationNon-sterile
Sterilization methodAutoclave / 132°C / 4min / 20 min dry

5. Intended Use :

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The Fixone All Suture Anchors may be used in either arthroscopic or open surgical procedures. After the suture strands are anchored to the bone, they may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. In conjunction with appropriate postoperative immobilization throughout the healing period, the suture anchor systems stabilize the damaged soft tissue.

{4}------------------------------------------------

6. Performance data:

  • Bench test were performed. Bench testing included biocompatibility, mechanical testing 1) including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
Test itemRequirementsResults
SurfaceWhen examined by normal vision, the surface shall be free from defects.Pass
Nominal size (Exclude suture)When tested by Vernier calipers, micrometer or tapeline, the measurement should be in accordance with standard.MD2018-00210
SterilizationISO 11135:2014-Sterilization of health care products - Ethylene oxide -Requirements for development, validation and routine control of asterilization process for medical devicesME-SVP-1802
Shelf Life and PackagingISO11607-2 / EN 868-5 / ASTM F1980-16 / ASTM F88-15 / ASTMF1929:1998AJ-SL1702
Comparison test withpredicate deviceInsertion test10pcf/40pfc Polyurethane block, self-punching /Implantation angle of 90 (vertical)AJU_M
Pull-out test10pcf/40pfc Polyurethane block, self-punching / Pull-out angle of 90 (vertical) with MTS Bionixservohydraulic testerD20201006_PT(Js,
Fatigue test10pcf/40pfc Polyurethane block, self-punching / withsame cycle loading and rateYknot)

1-1) Comparison testing process was based on the test of predicate device. See attached 'Biochemical Comparison_All Suture Anchor' / 'Fixation Strength Biochemical Data'.

    1. Biocompatibility
#Test itemTest method / Test criteriaTest result
1CytotoxicityISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPass
2Systemic toxicity testISO 10993-11, Biological evaluation of medical devices - Part 11: Tests for systemic toxicityPass
3Pyrogen TestISO 10993-11 Test for systemic toxicity, pyrogen testPass
4Intracutaneous reactivity testISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPass
5Maximization sensitizationISO 10993-10, Test for irritation and skin sensitization, Maximization test for delayed hypersensitivityPass
6Genotoxicity testISO 10993-3, Genotoxicity test OECE 471, Bacterial reverse mutation testPass
7Implantation testISO 10993-6, Tests for local effects after implantation, Annex D test methods for implantation in bonePass
8Hemolysis testISO 10993-4, Biological evaluation of medical devices - Part 4: Selection of tests for interactions with bloodPass
9Subchronic toxicityISO 10993-11 Tests for systemic toxicityPass

The performance tests demonstrated that Fixone All Suture Anchor performs in a substantially equivalent manner to the predicate device.

{5}------------------------------------------------

7. Predicate device comparison table

ManufacturerConMed CorporationAJU Pharm Co.,Ltd.Remark
510(k) No.K133224N/A
Indication foruseThe non-absorbable suture anchors areintended to reattach soft tissue to bone inorthopedic surgical procedures. The FixoneAll Suture Anchors may be used in eitherarthroscopic or open surgical procedures.After the suture strands are anchored to thebone, they may be used to reattach softtissue, such as ligaments, tendons, or jointcapsules to the bone. In conjunction withappropriate postoperative immobilizationthroughout the healing period, the sutureanchor systems stabilize the damaged softtissue.The non-absorbable suture anchors areintended to reattach soft tissue to bone inorthopedic surgical procedures. The FixoneAll Suture Anchors may be used in eitherarthroscopic or open surgical procedures.After the suture strands are anchored to thebone, they may be used to reattach softtissue, such as ligaments, tendons, or jointcapsules to the bone. In conjunction withappropriate postoperative immobilizationthroughout the healing period, the sutureanchor systems stabilize the damaged softtissue.Same
ClassificationnameFastener, Fixation, Nondegradable, SoftTissueFastener, Fixation, Nondegradable, SoftTissueSame
Trade nameY-Knot RC All-suture AnchorFixone All Suture AnchorN/A
Model/type-14 model codes including FAJ-1652802ewN/A
AppearanceImage: Y-Knot RC All-suture AnchorImage: Fixone All Suture AnchorSimilar
ProductconfigurationDriverSuture AnchorDriverSuture AnchorSame
MaterialNon-absorbable SutureNon-absorbable SutureSame
Suture
AbsorbableNon-absorbableNon-absorbableSame
Suture diameter0.50~0.599 (USP size 2)0.50~0.599 (USP size 2)Same
MaterialNon-absorbable SutureNon-absorbable SutureSame
SterilizationEO Gas sterilizationAccording to ISO 11135: 2014EO Gas sterilizationAccording to ISO 11135: 2014Same
BiodegradableYesYesSame
Principle ofoperationManual (Self punching)Manual (Self punching)Same
Shelf-life5 years5 yearsSame
PerformancetestingPull-outInsertionFatiguePull-outInsertionFatigueSame

8. Conclusion

The device is investigated for function to compare the operation of function between Fixone All Suture Anchor and predicate devices.

Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.

Therefore, it is concluded that Fixone All Suture Anchor is substantially equivalent to the legally marketed predicate device.

N/A